Datasheet
SUP-pH6.0
Technical Specification
Performance:
·Measure Range: PH(0-14);ORP(-1000~+1000Mv)
·Resolution: PH: 0.01 ORP: 1Mv
·Stability: PH:≤0.02PH/24H; ORP: ≤3Mv/24H
·Accuracy: PH: ±0.02PH; ORP:±1Mv
·Input impedance: ≥10^12
·Temperature compensation: NTC 10K or
PT1000 -10-130℃
·Power Supply: AC220V±10%, 50Hz or DC 24V
Output Signal:
·Communication:RS485, MODBUS-RTU
·Transmission signals: 4-20 mA
·Process alarm:High/Low process alarms, selected from pH, ORP+
·Relay output: AC250V, 3A
Other:
·Panel Front Size (mm) : 96(W) x 96(H) x Approx.112 (D)
·Panel Cutout Size (mm): 92(W) x 92(H) x Approx.112 (D)
·Converter weight : Approx. 1 kg
·Display: LCD with back lights in English
·Calibration: Semi-automatic 3 points calibration using pre-configured buffer tables 4,
7& 10, or 4, 6.18& 9.18
·Color: Black
·Ambient temperature : -20 to +55ºC (-5 - 130 ºF).
·Storage temperature : -30 to +70ºC (-20 - 160 ºF).
·Humidity : 10 to 90% RH at 40ºC (100 ºF)
Features
·Easy operation
·NEMA enclosure for field mounting and panel mounting
·Automatically Temperature Compensation
·Directly switchable to PH or ORP
·Large LCD display with background lighting
·PH or ORP sensors can be connected thanks to the sensor supply integrated in the
output
·Using the setup program: user-friendly programming
·4-20mA analog output
·RS485 communication
·Relay output
Product description
The device is designed for use on site. A rugged housing protects the electronics
and the electrical connections from corrosive environmental conditions . As an
alternative, the device can also be installed in a control panel.
pH6.0 pH controller
1, Temperature:Compensation temperature
2, Analyst output:Analyst output
3,Measured value:Real-time measurements value
4, High alarm:High alarm
5, Low alarm:Low alarm
6,ESC:Check related warning status on the “monitoring page”
Return to previous level page in the up& down level page linked to
“menu page”
7, Menu: Enter the MENU on the “monitoring page”
Exit the MENU on the “menu page”
8, Choose :“MOVE RIGHT” and “MOVE DOWN”
9, Enter:Enter the sub-menu or confirm modification on the “menu page”
Instrument wiring
Wiring diagram
Identification of terminal
·INPUT: Measuring terminal of the electrode
·REF: Reference terminal of the electrode
·A: Temperature compensation terminal A,NTC10K and PT1000 connect here
·B: Temperature compensation terminal B,NTC10K and PT1000 connect here
·C: Temperature compensation terminal C, PT1000 three-wire temperature
grounding, PT1000 two-wire need to be short-connected to TEMPB, not
NTC10K.
·NC: Unidentified
·RS485 (A +): RS485 communication interface A +
·RS485 (B -): RS485 communication interface B-
·4-20mA (+): 4-20mA output end+
·4-20mA (-): 4-20mA output end-
·AC220V (L): AC220V Fire Wire
·AC220V (N): AC220V zero line
·HO: high alarm normally open relay
·HC: high alarm normally closed relay
·LO: low alarm normally open relay
·LC: low alarm normally closed relay
·COM: common
Display
pH calibration
The characteristic of a pH
electrode will change with time due
to electrode coating and aging. And
even a pH electrode would be stable
over time, pH electrodes cannot be
produced with identical
characteristics.
PH Calibration: Enter the PH calibration screen, the first PH electrode into the
4.00PH standard solution, standing for a moment, to be stable after the show, press
the 【ENT】 , then the PH electrode into the 6.86PH standard solution Set the PH
electrode into the 9.18PH standard solution, put it aside for a moment, after the test
is stable, press 【ENT】, the display shows the success of the calibration, the whole
process of PH calibration completes.
PH Modification: The measured PH can be modified between 2 PH values.
ORP Calibration: enter the ORP calibration screen, the first ORP electrode into the
86mV standard solution, standing for a moment, to be shown after the stability,
press the【ENT】, then the ORP electrode into the 256mV standard solution, static Set
a moment, after the display is stable, press the【ENT】display calibration is successful,
ORP calibration process is over.
ORP Modification: The measured ORP can be modified between 300mV.
Temperature correction: You can correct the temperature of the automatic
temperature compensation, the correction range is ± 20.0 ℃.
Multi-point calibration
To achieve the best possible accuracy, the calibration should cover the range of
the desired measurement values. If the readings go beyond the calibrated range, the
pH meter assumes linearity and simply extrapolates the value to be displayed. The
true value may be slightly different.
More advanced pH meters will let the user calibrate at three, four or five and
even higher numbers of pH values. A multi-point calibration mean, in comparison to a
two-point calibration, that you can calibrate your pH tester on both sides of the zero
point (pH 7.00). This will expand your pH measurement range without the need of
recalibrating.
Ordering Code
pH Probe
C:Plastic pH sensor with NTC 10K (0-60℃) Power Supply
D: Glass pH sensor with NTC 10K (0-80℃) N: 220VAC
E:High Tmeperature pH sensor with NTC O: 24VDC
10K (0-130℃)
F: Made in Germany pH sensor
E+H (0-80℃)
G:Made in Germany pH sensor Medium
Jumo (0-80℃) P: Normal water
H: Plastic ORP sensor with NTC 10K Q: Waste water
(0-60℃) R: Nutrient solution
I: Glass ORP sensor with NTC 10K (0-80℃) S: Galvanic plating
T: Cooling water
Cable Length U: Drinking water
J:5 meters V: Others:
K: 10 meters
J:15 meters
Order code: A
A
Order example: B C J L N P
Address: 5th floor, Building 4, Address: 2 Venture Address: Göttinger Straße.59 Address: No 3, Jalan Emas
Singapore-Hangzhou Science & Drive #11-30 Vision 30449 Hannover Niedersachsen Jaya 1, Taman Industries