Datasheet
SUP-DY2900
Product description:
oxygen molecules is calculated. basin for a highly efficient biological cleaning process
• Water treatment:
Oxygen measurement for status monitoring of drinking water
• Fish farming:
Oxygen measurement and regulation for optimum living and
growth conditions
Features
Measuring principle
• Sensor design:
(fluorescence layer).
– The sensor optics are directed at the underside of the fluorescence layer.
• There is an equilibrium between the oxygen partial pressure in the medium and that in
– If the sensor is immersed in the medium, the equilibrium is established very quickly.
• Measuring process:
– The sensor optics send green light pulses to the fluorescence layer.
– The Oxygen-sensitive molecules instantly becomes excited and releases red light.
– The duration and intensity of the response signals is directly dependent on the oxygen
– If the medium is free from oxygen, the response signals are long and very intense.
• Measurement result:
– The sensor returns a signal that is in proportion to the oxygen concentration in the
medium.
– The medium temperature and air pressure are already taken into account calculated in
the sensor.
– In addition to the standard values of concentration, saturation index and partial pressure,
the sensor also returns a raw measured value in μs. The value corresponds to the decay
time of the fluorescence and is approx. 20 μs in air and approx. 60 μs in media free from
oxygen.
Technical Specifications:
Resolution 0.01mg/L,0.1%,1hPa
Accuracy ±1.5%FS
Repetition ±0.5%FS
Temperature
NTC 10K/PT1000
measurement type
Auto A/manual H -10 - 60℃ Resolution: 0.1℃ Correction
Measuring range accuracy: ±0.5℃
Accuracy 0.1%FS
Installation methods
Open a 92.5 * 92.5(mm) installation hole on the instrument cabinet or installation panel
(the dimension is 100*100*150mm). Insert the instrument into the installation hole and
latch on the butterfly clasp, as shown below.