This Year 2000 Readiness Disclosure Statement revises, supplements, and/or restates important
information previously disseminated in our Year 2000 Readiness Disclosure Statement dated February
12, 1999 (the “February 1999 Statement”). Revised and supplemental paragraphs and sections are so
noted in this new Year 2000 Readiness Disclosure Statement. Those systems described in this
Statement as “year 2000 compliant” are year 2000 compliant in accordance with the attached Year
2000 Compliance Limited Warranty.
‚A parameter save/load function performed using the Parameter Floppy Disk after December 31,
1999 will result in the date stamp indicating the “Last Modified” status as the year “1900”.
DSD/00001/7194.5
REPLY systems with operating software versions below version 2.4 cannot be upgraded using the
special software disk.
SUPPLEMENTAL INFORMATION: After December 31, 1999, a four-digit year must be used to
define the date in the OLYMPUS AU560 “SET DATE & ROUND” and OLYMPUS REPLY “SET
DATE & CAROUSEL” menus. If existing data indexes have a two-digit year entry they must be
searched separately from those data indexes having a four-digit year entry.
The OLYMPUS AU800 and OLYMPUS AU1000 utilize a number of different Hewlett-Packard
computers. If the computer installed with a system is a Hewlett-Packard Vectra VL 5/133 or higher
clock speed unit, no corrective action will be required. OLYMPUS AU800 and OLYMPUS AU1000
units which utilize these higher clock speed computers are year 2000 compliant. Within the
OLYMPUS AU800 and OLYMPUS AU1000 software program, on systems utilizing Hewlett-
Packard computers with slower clock speeds (all 386, 486, 5/75, and 5/120 units), the display date,
data index, and printout date appear to change properly from “12/31/1999” to “01/01/2000”.
However, upon powering off, the computer’s internal clock will revert to 1980 when the system is
next powered on. Therefore, those Hewlett-Packard computers with slower clock speeds will require
manual updating of the current date by using the MS-DOS “DATE” command to set the date to
2000 upon initial use AFTER 12:00 a.m. on January 1, 2000. No action should be taken prior to
12:00 a.m. on January 1, 2000. Please follow the Year 2000 Conversion Instructions and Test
Procedures attached to this Year 2000 Readiness Disclosure Statement. This update MUST be
performed on all Hewlett-Packard computers with clock speed slower than the Hewlett-Packard
Vectra VL 5/133.
SUPPLEMENTAL INFORMATION:
¬ The OLYMPUS AU800 and OLYMPUS AU1000 incorporate features that allow data to be
stored on floppy disk. The “OFFLINE FLOPPY DATA OUTPUT” menu stores data in a form
that can be used offline for functions such as spreadsheet applications. When performing the
“OFFLINE FLOPPY DATA OUTPUT” function using a Data Floppy Disk initialized after
January 1, 2000, program execution will be interrupted. As a result, program status information
may appear on the PC monitor. Patient data residing in the system hard drive is not affected and
operation can be easily resumed by re-booting the system.
It is recommended that a sufficient quantity of Data Floppy Disks be initialized prior to January
1, 2000 in order to support offline floppy data output for a period of one month. In the event
DSD/00001/7194.5
2
that there are no initialized Data Floppy Disks available upon transition to January 1, 2000, the
system date can be set to a date prior to January 1, 2000 in the “AUXILIARY/CURRENT
DATE AND TIME” menu. Disks can then be initialized using the “FLOPPY INITIALIZE”
menu. Should this procedure be required, the date must be returned to the then-current date and
time prior to analysis in the “AUXILIARY/CURRENT DATE AND TIME” menu. In addition,
pre-initialized Data Floppy Disks will be available from Olympus after January 15, 2000. An
updated version of software will be provided at no charge by June 1, 2000.
- Re-using a Data Floppy Disk containing data stored after December 31, 1999 will result in an
interruption in program execution. As a result, program status information may appear on the
PC monitor. Patient data residing in the system hard drive is NOT affected. Furthermore,
operation can be easily resumed by re-booting the system. Disks can be re-used; however, disks
containing data stored after December 31, 1999 must be initialized using the “FLOPPY
INITIALIZE” menu and using a system date earlier than January 1, 2000. An updated version of
software will be provided at no charge by June 1, 2000.
® The “DATA DOWN LOAD TO FD” menu saves data onto a floppy disk in a form that will
allow review or editing on the OLYMPUS AU800 and OLYMPUS AU1000 systems. This
method will function using disks initialized before and after January 1, 2000. This is the
recommended action for data archive purposes.
¯ Floppy disk file date stamps will be recorded on the disk as “53” during the year 2000. In
addition, years subsequent to 2000 will be displayed as “54”, “55”, etc. Information stored on
the disks will not be affected. File date stamps are displayed only when viewed outside the
AU800/AU1000 application. There will be no software update available to resolve this
condition; however, this condition has no impact on system functionality.
° The impact of the aforementioned conditions on overall system operation is limited and only
applies to the operating conditions described above.
The OLYMPUS AU600 utilizes a number of different Hewlett-Packard personal computers. If the
computer installed with a system is a Hewlett-Packard Vectra VL 5/133 or higher clock speed unit,
no corrective action will be required. OLYMPUS AU600 units which utilize these higher clock speed
computers are year 2000 compliant. Within the OLYMPUS AU600 software program, on systems
utilizing Hewlett-Packard computers with slower clock speeds (5/75 and 5/120 units), the display date
with two-digit year code located in the upper right-hand portion of the screen changes from
“12/31/99” to “01/01/00”. The date display is for convenience purposes only and does not affect
the system’s functionality. However, the data index and printout dates within the system program
software contain a four-digit year code. Consequently, upon powering off, the computer’s internal
clock will revert to 1980 when the system is next powered on. Therefore, those Hewlett-Packard
computers with slower clock speeds will require manual updating of the current date by using the MS-
DOS “DATE” command to set the date to 2000 upon initial use AFTER 12:00 a.m. on January 1,
2000. No action should be taken prior to 12:00 a.m. on January 1, 2000. Please follow the Year
2000 Conversion Instructions and Test Procedures attached to this Year 2000 Readiness Disclosure
Statement. This update MUST be performed on all Hewlett-Packard computers with clock speed
slower than the Hewlett-Packard Vectra VL 5/133.
SUPPLEMENTAL INFORMATION:
DSD/00001/7194.5
3
• Offline Output and Data floppy disk file date stamps will be recorded on the disks as “53” during
the year 2000. In addition, years subsequent to 2000 will be displayed as “54”, “55”, etc.
Information stored on the disks will not be affected. File date stamps are displayed only when
viewed outside the AU600 application. This condition has no impact on system functionality.
‚ In the “Periodic Maintenance” menu, the “Next” date, indicating the next scheduled due date for
a maintenance item, will be displayed one day late when February 29, 2000 falls between the
“Executed” date (i.e., the date periodic maintenance was last performed) and the “Next” date
(i.e., the calculated date the next periodic maintenance is due). Manual tracking of periodic
maintenance must be conducted under these conditions. The “Periodic Maintenance” function is
provided as a convenience and, therefore, in no way impacts system functionality except as
described above.
ƒ Integrity of patient data is not affected by the aforementioned conditions. An updated version of
software will be provided at no charge by June 1, 2000.
- In the “Periodic Maintenance” menu, the color of maintenance items should change to yellow
when the then-current date coincides with the “Next” maintenance due date. The color should
change to red when the current date exceeds the “Next” maintenance due date. However, when
the “Next” maintenance due date is February 29, the “Next” maintenance due date will be
displayed in yellow on March 1 and red on March 2. Manual tracking of periodic maintenance
must be conducted under these conditions. The “Periodic Maintenance” function is provided as a
convenience and, therefore, in no way impacts system functionality except as described above.
® In the “Periodic Maintenance” menu, the grid corresponding to the date when maintenance was
last performed should be marked with an “o” symbol and the “Next” maintenance due date should
be displayed in yellow in the “Grid Display” menu. However, symbols indicating last performed
(“o”) and “Next” due (yellow) dates for the month of January will not be displayed in the “Grid
Display” menu from December 1 through December 30. Manual tracking of periodic
maintenance must be conducted under these conditions. The “Periodic Maintenance” function is
provided as a convenience and, therefore, in no way impacts system functionality except as
described above.
¯ In the “Grid Display” menu, the color of grids for maintenance items should change to yellow
when the then-current date coincides with the “Next” maintenance due date. However, when the
“Next” maintenance due date is February 29, the grids for both February 29 and March 1 will be
displayed in yellow. In addition, printing of the “Grid Display” using the “Report/Schedule”
menu will assign an “*” mark (indicating maintenance due date) for both February 29 and March
1. Manual tracking of periodic maintenance must be conducted under these conditions. The
“Grid Display” function is provided as a convenience and, therefore, in no way impacts system
functionality except as described above.
DSD/00001/7194.5
4
° The “Cuvette Status” menu displays a yellow color to alert the operator when 8 or more days
have elapsed since the last W2/Photocal operation was performed. When a W2/Photocal is
performed between February 22 and February 29, the yellow color alert will not be displayed until
9 days following the date of such W2/Photocal operation. The scheduling of W2/Photocal
operations during this period must be manually tracked.
± The printout generated by the “Report/Schedule” menu should include the “o” and “*” symbols
for the “Executed” maintenance date and the “Next” maintenance due date, respectively.
However, from December 31 through January 30, the printout generated by the
“Report/Schedule” menu will not include the “o” and “*” symbols for the portion of the month
of January after the then-current date. Information prior to and including the then-current date
will be printed. Manual tracking of periodic maintenance must be conducted under these
conditions. The “Report/Schedule” function is provided as a convenience and, therefore, in no
way impacts system functionality except as described above.
² The print function for maintenance activities will indicate a status of 31 days in all months.
Please disregard maintenance marks for all dates that do not exist (e.g., February 30, April 31).
This print function is provided as a convenience and, therefore, in no way impacts system
functionality except as described above.
***
Included with this Year 2000 Readiness Disclosure Statement are Year 2000 Conversion Instructions
and Test Procedure documents for the REPLY, AU560, AU600, AU800, AU1000, AU5200, and
PK7200 systems. If you have any questions, please do not hesitate to contact Olympus’s Technical
Services Department at (800) 223-0130.
***
Olympus appreciates your continued patronage and support, and we look forward to serving you in
the future.
Glenn Mann
DSD/00001/7194.5
5
Director of Service
Olympus America Inc.
Diagnostic Systems Division
Notice: This Statement contains proprietary information that is protected by copyright. All rights
are reserved. No part of this document may be photocopied, reproduced, or translated to another
language without the prior written consent of Olympus America Inc.
DSD/00001/7194.5
6
YEAR 2000 COMPLIANCE LIMITED WARRANTY
OF
OLYMPUS AMERICA INC.
DIAGNOSTIC SYSTEMS DIVISION
EXCEPT FOR THE LIMITED WARRANTY SET FORTH ABOVE, OAI-DSD MAKES NO
AND DISCLAIMS ALL OTHER REPRESENTATIONS AND WARRANTIES, WHETHER
EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED
WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE,
OR CONCERNING ANY PATENTS OR TECHNOLOGY. OAI-DSD DOES NOT
WARRANT THAT SOFTWARE WILL BE ERROR-FREE AND/OR UNINTERRUPTED.
MOREOVER, OAI-DSD DOES NOT WARRANT THE YEAR 2000 COMPLIANCE OF
PRODUCTS MANUFACTURED BY THIRD PARTIES OR PRODUCTS NOT SHIPPED
FROM OAI-DSD INVENTORY. OAI-DSD DISCLAIMS RESPONSIBILITY AND
LIABILITY FOR ENSURING THE YEAR 2000 COMPLIANCE OF INTEGRATED
SYSTEMS COMPRISED OF OLYMPUS AND NON-OLYMPUS PRODUCTS. ALL
GUARANTIES, WARRANTIES, CONDITIONS, AND REPRESENTATIONS
WHATSOEVER, EITHER EXPRESS OR IMPLIED, WHETHER ARISING UNDER ANY
STATUTE, LAW, COMMERCIAL USAGE OR OTHERWISE, ARE HEREBY EXPRESS
EXCLUDED. IN NO EVENT SHALL OAI-DSD BE LIABLE FOR ANY INDIRECT,
INCIDENTAL, CONSEQUENTIAL OR SPECIAL DAMAGES OF ANY KIND
(INCLUDING WITHOUT LIMITATION LOSS OF PROFITS OR LOSS OF USE),
WHETHER OR NOT OAI-DSD SHALL BE OR SHOULD BE AWARE OF THE
POSSIBILITY OF SUCH POTENTIAL LOSS OR DAMAGE.
Representations and warranties made by any person, including dealers and representatives of
OAI-DSD, which are inconsistent with, in conflict with, or in addition to, the terms of this
DSD/00001/7194.5
Limited Warranty, shall not be binding upon OAI-DSD unless reduced to writing and approved
by an expressly authorized officer of Olympus America Inc.
DSD/00001/7194.5
2
Year 2000 Conversion Instructions and Test Procedure
OLYMPUS REPLYand OLYMPUS AU560 Automated Chemistry Analyzers
Note: If you wish to do a year 2000 test of the instrument please see page 5.
The following items are required to perform this procedure on the Reply/AU560.
1. Confirm that POWER ON switch on the front of the DPR is not illuminated and the analyzer
and DPR are off.
2. Move the BOOT switch on the rear side of the DPR from “WD” to “FD”. When moving the
switch, please pull out on the switch. If the switch is not pulled, it cannot be moved (FIG.12-
¬-).
3. Switch ON the LOCAL POWER switch on the rear side of the DPR (FIG.12-¯). Wait until
the DPR displays the message ”Please insert system floppy disk & hit any key then restart disk
mount”.
4. Confirm that Write Protect is off on the Program Floppy Disk 1/3, insert it into the FD drive
and press the Enter ↵ key.
5. Wait until the DPR displays the following menu (FIG.1).
# SELECT No.>
---------------------------------------------------------------------
1 INITIALIZE MEDIA
2 PASSWORD SET
3 COMMUNICATION
4 MAINTENANCE
5 HISTORY
6 COPY FILES
7 SYSTEM UP
1: 2: 3: 4: 5: 6: 7: 8:EXIT
FIG.1
7. Wait until the DPR displays the following menu (FIG.2). Only Item 1 displays at this time.
DSD/00001/7194.5
3
*** SYSTEM UP ***
# SELECT No.>
---------------------------------------------------------------------
1 DATE & TIME SET
(2 SYSTEM PARAMETERS )
(3 OPTION UNIT )
(4 DATA FILE INITIALIZE )
(5 PARAMETER INITIALIZE )
1: 2: 3: 4: 5: 6: 7: 8:PRINT
FIG.2
8. Type S and the Enter ↵ key, B and the Enter ↵ key and M and the Enter ↵ key on the
keyboard. Items 2 through 5 will now be displayed on the above menu.(FIG.2)
9. Type 2 and the Enter ↵ key on the keyboard. Wait until the DPR displays the following menu
(FIG.3).
# SELECT No.>
---------------------------------------------------------------------
1 DATE MARK 19
2 AREA AMERICA
3 CYCLE 60
4 PERIOD
ROUND-1 9 SEC.
ROUND-2 9 SEC.
ROUND-3 9 SEC.
ROUND-4 9 SEC.
5 INQUIRY
ROUTINE & EMERGENCY BARCODE
ACAL SEQUENCE
QC SEQUENCE
6 SYSTEM PASSWORD
1: 2: 3: 4: 5: 6: 7: 8:PRINT
FIG.3
DSD/00001/7194.5
4
*** COPY FILES ***
# SELECT No.>
---------------------------------------------------------------------
1. FILE COPY
2. FILE CREATE
1: 2: 3: 4: 5: 6: 7: 8:PRINT
FIG.4
14. Type 1 and the Enter ↵ key on the keyboard. Wait until the DPR displays the menu shown in
FIG.5.
# SELECT No.>
---------------------------------------------------------------------
DEV. FILE NAME DEV. FILE NAME DEV. FILE NAME
1.MKP.SYS : :
2.M510SYS.FIL : :
: : :
: : :
15. Type 2 and the Enter ↵ key on the keyboard. Wait until file number 2 (M510SYS.FIL) has
been copied.
16. Press the OPS. Menu on the keyboard. The DPR displays the menu shown in FIG.1.
17. Press the function key F8 (Exit) on the keyboard. Wait until the DPR displays the message ”BE
SURE TO SET BOOT SWITCH ON WD SIDE AND PRESS RESET SWITCH”.
18. Remove the Program Floppy Disk 1/3.
19. Switch OFF the LOCAL POWER switch on the rear side of the DPR (FIG12- ¯).
20. Wait 10 seconds and switch ON the LOCAL POWER button on the rear side of the DPR
(FIG12- ¯). Wait until the DPR displays the message ”Please insert system floppy disk & hit
any key then restart disk mount”.
21. Insert the MU5907 SPECIAL TOOL for YEAR 2000 floppy disk into the FD drive and press
the Enter ↵ key on the keyboard. Wait until the DPR displays the following messages (FIG.6).
DSD/00001/7194.5
5
CP/M-86 Version 1.0
CP/M86=Version 1.0
A>
FIG.6
22. Type GO2000 and press the Enter ↵ key on the keyboard. (see FIG.7)
A>GO2000
FIG.7
23. Wait until the DPR displays the following messages (Fig.8).
A>GO2000
A>
FIG.8
Note: If any errors are displayed during steps 21 to 23, refer to the Appendix Error List.
24. Remove the MU5907 SPECIAL TOOL for YEAR 2000 floppy disk.
25. Move the BOOT switch on the rear side of the DPR from “FD” to “WD”. When moving the
switch, please pull out on the switch. If the switch is not pulled, it cannot be moved (FIG.12-
¨).
26. Switch OFF the LOCAL POWER button on the rear side of the DPR (FIG12- ¯).
27. Power on the equipment using the POWER ON switch on the front of the DPR.
28. Execute preparation check after program loading completes. The Start Condition menu is
displayed.
29. In the Start Condition menu, check the following:
a. Check that the date and time displayed on the top right of the screen is correct.
b. Execute data index change entering = for date. Check that the date of the new index
indicates the correct date.
DSD/00001/7194.5
6
30. No further action is required.
Use the following steps to test the change to the year 2000 and then reset to the current date.
The following items are required to perform this procedure on the Reply/AU560.
31. Confirm that POWER ON switch on the front of the DPR is not illuminated and the analyzer
and DPR are off.
32. Move the BOOT switch on the rear side of the DPR from “WD” to “FD”. When moving the
switch, please pull out on the switch. If the switch is not pulled, it cannot be moved (FIG.12-
¬-).
33. Switch ON the LOCAL POWER switch on the rear side of the DPR (FIG.12-¯). Wait until
the DPR displays the message ”Please insert system floppy disk & hit any key then restart disk
mount”.
34. Confirm that Write Protect is off on the Program Floppy Disk 1/3, insert it into the FD drive
and press the Enter ↵ key.
35. Wait until the DPR displays the (FIG.1) menu.
36. Type 7 and the Enter ↵ key on the keyboard.
37. Wait until the DPR displays the (FIG.2) menu. Only Item 1 displays at this time.
38. Type S and the Enter ↵ key, B and the Enter ↵ key and M and the Enter ↵ key on the
keyboard. Items 2 through 5 will now be displayed on the (FIG.2) menu.
39. Type 2 and the Enter ↵ key on the keyboard. Wait until the DPR displays the (FIG.3) menu.
40. Type 1 and the Enter ↵ key on the keyboard.
41. Type 20 and the Enter ↵ key on the keyboard.
42. Press the Menu key. The DPR displays the menu shown in FIG.2.
43. Type 1 and the Enter ↵ key on the keyboard. The DPR displays the following menu (FIG.9).
# SELECT No.>
---------------------------------------------------------------------
1 YEAR XX
2 MONTH XX
3 DAY XX
4 HOUR XX
5 MINUTE XX
6 SECOND XX
1: 2: 3: 4:SET 5: 6: 7: 8:
FIG.9
DSD/00001/7194.5
7
46. Type 2 and the Enter ↵ key on the keyboard.
47. Type 01 for the month and the Enter ↵ key on the keyboard.
48. Type 3 and the Enter ↵ key on the keyboard.
49. Type 01 for the day and the Enter ↵ key on the keyboard.
50. Press F4 (Set) to set the date and time. The date and time displayed in the upper right hand
corner of the screen will reflect your changes.
51. Press the OPS Menu key on the keyboard. The DPR displays the menu shown in FIG.1.
52. Type 6 and the Enter ↵ key on the keyboard. Wait until the DPR displays the (FIG.4) menu.
53. Type 1 and the Enter ↵ key on the keyboard. Wait until the DPR displays the (FIG.5) menu.
54. Type 2 and the Enter ↵ key on the keyboard. Wait until file number 2 (M510SYS.FIL) has
been copied.
55. Press the OPS. Menu on the keyboard. The DPR displays the menu shown in FIG.1.
56. Press the function key F8 (Exit) on the keyboard. Wait until the DPR displays the message ”BE
SURE TO SET BOOT SWITCH ON WD SIDE AND PRESS RESET SWITCH”.
57. Remove the Program Floppy Disk 1/3.
58. Switch OFF the LOCAL POWER switch on the rear side of the DPR (FIG12- ¯).
59. Wait 10 seconds and switch ON the LOCAL POWER button on the rear side of the DPR
(FIG12- ¯). Wait until the DPR displays the message ”Please insert system floppy disk & hit
any key then restart disk mount”.
60. Insert the MU5907 SPECIAL TOOL for YEAR 2000 floppy disk into the FD drive and press
the Enter ↵ key on the keyboard. Wait until the DPR displays the message from FIG.6.
61. Type GO2000 and press the Enter ↵ key on the keyboard. (see FIG.7)
62. Wait until the DPR displays the message from Fig.8.
Note: If any errors are displayed during steps 60 to 62, refer to the Appendix Error List.
63. Remove the MU5907 SPECIAL TOOL for YEAR 2000 floppy disk.
64. Move the BOOT switch on the rear side of the DPR from “FD” to “WD”. When moving the
switch, please pull out on the switch. If the switch is not pulled, it cannot be moved (FIG.12-
¨).
65. Switch OFF the LOCAL POWER button on the rear side of the DPR (FIG12- ¯).
66. Power on the equipment using the POWER ON switch on the front of the DPR.
67. Execute preparation check after program loading completes. The Start Condition menu is
displayed.
68. In the Start Condition menu, check the following:
a. Check that the date and time displayed on the top right of the screen is correct.
b. Execute data index change entering = for date. Check that the date of the new index
indicates the correct date.
Note: Measurement done in the current index with the year set to 2000 will remain on the hard
drive and may cause some confusion.
When testing is complete use the following steps to return to the current date.
DSD/00001/7194.5
8
73. Wait until the DPR displays the (FIG.1) menu.
74. Type 7 and the Enter ↵ key on the keyboard.
75. Wait until the DPR displays the (FIG.2) menu. Only Item 1 displays at this time.
76. Type S and the Enter ↵ key, B and the Enter ↵ key and M and the Enter ↵ key on the
keyboard. Items 2 through 5 will now be displayed on the (FIG.2) menu.
77. Type 2 and the Enter ↵ key on the keyboard. Wait until the DPR displays the (FIG.3) menu.
78. Type 1 and the Enter ↵ key on the keyboard.
79. Type 19 and the Enter ↵ key on the keyboard.
80. Press the Menu key. The DPR displays the menu shown in FIG.2.
81. Type 1 and the Enter ↵ key on the keyboard. The DPR displays the (FIG.9) menu.
82. Type 1 and the Enter ↵ key on the keyboard.
83. Type the current year and the Enter ↵ key on the keyboard.
84. Type 2 and the Enter ↵ key on the keyboard.
85. Type the current month and the Enter ↵ key on the keyboard.
86. Type 3 and the Enter ↵ key on the keyboard.
87. Type the current day and the Enter ↵ key on the keyboard.
88. Press F4 (Set) to set the date and time. The date and time displayed in the upper right hand
corner of the screen will reflect your changes.
89. Press the OPS Menu key on the keyboard. The DPR displays the menu shown in FIG.1.
90. Type 6 and the Enter ↵ key on the keyboard. Wait until the DPR displays the (FIG.4) menu.
91. Type 1 and the Enter ↵ key on the keyboard. Wait until the DPR displays the (FIG.5) menu.
92. Type 2 and the Enter ↵ key on the keyboard. Wait until file number 2 (M510SYS.FIL) has
been copied.
93. Press the OPS. Menu on the keyboard. The DPR displays the menu shown in FIG.1.
94. Press the function key F8 (Exit) on the keyboard. Wait until the DPR displays the message ”BE
SURE TO SET BOOT SWITCH ON WD SIDE AND PRESS RESET SWITCH”.
95. Remove the Program Floppy Disk 1/3.
96. Switch OFF the LOCAL POWER switch on the rear side of the DPR (FIG12- ¯).
97. Wait 10 seconds and switch ON the LOCAL POWER button on the rear side of the DPR
(FIG12- ¯). Wait until the DPR displays the message ”Please insert system floppy disk & hit
any key then restart disk mount”.
98. Insert the MU5907 SPECIAL TOOL for YEAR 2000 floppy disk into the FD drive and press
the Enter ↵ key on the keyboard. Wait until the DPR displays the message from FIG.6.
99. Type GO2000 BACK and press the Enter ↵ key on the keyboard. (see FIG.10)
Note: A space exists between “GO2000” and “BACK”
A>GO2000 BACK
FIG.10
DSD/00001/7194.5
9
A>GO2000 BACK
A>
FIG.11
Note: If any errors are displayed during steps 98 to 100, refer to the Appendix Error List.
100. Remove the MU5907 SPECIAL TOOL for YEAR 2000 floppy disk.
101. Move the BOOT switch on the rear side of the DPR from “FD” to “WD”. When moving
the switch, please pull out on the switch. If the switch is not pulled, it cannot be moved
(FIG.12-¨).
102. Switch OFF the LOCAL POWER button on the rear side of the DPR (FIG12- ¯).
103. Power on the equipment using the POWER ON switch on the front of the DPR.
104. Execute preparation check after program loading completes. The Start Condition menu is
displayed. In the Start Condition menu, check the following:
a. Check that the date and time displayed on the top right of the screen is correct.
b. Execute data index change entering = for date. Check that the date of the new index
indicates the correct date (19xx).
105. No further action is required.
DSD/00001/7194.5
10
Rear side of the DPR
WD
‡A
Â
‡@
BOO
À BOOT
WD
,eD
FD Á
FD
‡B
DPR LOCAL
5A POWER
FUSE
ON
OFF
‡C
FIG.9
Ã
FIG.12
DSD/00001/7194.5
11
Year 2000 Conversion Instructions and Test Procedure
OLYMPUS AU1000 Automated Chemistry Analyzer
1. Press the Power switch on the analyzer and then press the “Q” key while the * is displayed in
the upper left hand corner of the screen. The Dos C:> prompt will be displayed.
2. Type “date” and press enter↵. Follow the prompts to enter the correct date within the year
2000 and press enter↵.
3. Press the Em. Stop switch to remove power from the analyzer.
4. Press the Power On switch and then the Warm Up switch to restore power and load the
program on the analyzer. Check that the correct date is displayed in the upper right corner of
the screen.
Use the following steps to test for the change to the year 2000 and then reset to the current date.
5. Press the Power switch on the analyzer and then press the “Q” key while the * is displayed in
the upper left hand corner of the screen. The Dos C:> prompt will be displayed.
6. Type “date” and press enter↵. Follow the prompts to enter the date you wish within the year
2000 and press enter↵.
7. Press the Em. Stop switch to remove power from the analyzer.
Press the Power On switch and then the Warm Up switch to restore power and load the program
on the analyzer. Check that the date you entered is displayed in the upper right corner of the screen.
When testing is complete, use the following steps to return to the current date.
9. Press the Power switch on the analyzer and then press the “Q” key while the * is displayed in
the upper left hand corner of the screen. The Dos C:> prompt will be displayed.
10. Type “date” and press enter↵. Follow the prompts to enter the current date and press enter↵.
11. Press the Em. Stop switch to remove power from the analyzer.
Press the Power On switch and then the Warm Up switch to restore power and load the program
on the analyzer. Check that the current date is displayed in the upper right corner of the screen. No
further action is required.
DSD/00001/7194.5
12
Year 2000 Conversion Instructions and Test Procedure
OLYMPUS AU5200 Automated Chemistry Analyzer
1. From the Ops Menu select 21. Auxilliary Menu then select 8. Current Date and Time.
2. In this menu press the Enter key for the Year, Month, Day, Hour, Minute and Day of Week.
This menu will have the updated date already.
Use the following steps to test for the change to the year 2000 and then reset to the current date.
4. From the Ops Menu select 21. Auxilliary Menu then select 8. Current Date and Time.
5. In this menu type 1999 for the Year, 12 for the Month, 31 for the Day, 23 for the Hour, 59
for the Minute and Friday for the Day of Week.
7. The date in the upper right hand corner of the screen will switch to 01/01/1900.
9. In this menu press the Enter key for the Year, Month, Day, Hour, Minute and Day of Week.
This menu will have the updated date already.
10. Press the Yes key at the # All Ok ? (Yes,No) > __ prompt. The display will now be
correct.
When testing is complete use the following steps to return to the current date.
12. In this menu type the current date, time, hour, minute and day of the week for the Year,
Month, Day, Hour, Minute and Day of Week prompts.
13. Press the Yes key at the # All Ok ? (Yes,No) > __ prompt. Confirm the correct date and
time is displayed in the upper right hand corner of the display.
DSD/00001/7194.5
13
Year 2000 Conversion Instructions and Test Procedure
OLYMPUS AU600 Automated Chemistry Analyzer
12. Press the Power switch on the analyzer and then press the “Q” key while the * is displayed in
the upper left hand corner of the screen. The System Maintenance Menu will be displayed.
13. Type “14” and press enter↵ to select option 14. Go to DOS.
14. Type “date” and press enter↵. Follow the prompts to enter the correct date within the year
2000 and press enter↵.
15. Press the Em. Stop switch to remove power from the analyzer.
Press the Reset switch and then the Power switch to restore power and load the program on the
analyzer. Check that the correct date is displayed in the upper right corner of the screen. No further
action is required.
Use the following steps to test for the change to the year 2000 and then reset to the current date.
16. Press the Power switch on the analyzer and then press the “Q” key while the * is displayed in
the upper left hand corner of the screen. The System Maintenance Menu will be displayed.
17. Type “14” and press enter↵ to select option 14. Go to DOS.
18. Type “date” and press enter↵. Follow the prompts to enter the date you wish within the year
2000 and press enter↵.
19. Press the Em. Stop switch to remove power from the analyzer.
Press the Reset switch and then the Power switch to restore power and load the program on the
analyzer. Check that the date you entered is displayed in the upper right corner of the screen.
When testing is complete, use the following steps to return to the current date.
21. Press the Power switch on the analyzer and then press the “Q” key while the * is displayed in
the upper left hand corner of the screen. The System Maintenance Menu will be displayed.
22. Type “14” and press enter↵ to select option 14. Go to DOS.
23. Type “date” and press enter↵. Follow the prompts to enter the current date and press enter↵.
DSD/00001/7194.5
14
24. Press the Em. Stop switch to remove power from the analyzer.
Press the Reset switch and then the Power switch to restore power and load the program on the
analyzer. Check that the current date is displayed in the upper right corner of the screen. No further
action is required.
DSD/00001/7194.5
15
Year 2000 Conversion Instructions and Test Procedure
OLYMPUS PK7200 Automated Microplate System
14. Power on the instrument using the Power On switch on the front of the DPR.
***OPERATION MENU***
# SELECT No.>
_____________________________________________________________
15. From the Operation Menu select F5: DIAG then press the Yes key at the DIAG ? Yes or No
prompt.
***OPERATION MENU***
# SELECT No.>
_____________________________________________________________
1 DIAGNOSTIC (DPR)
2 DIAGNOSTIC (DPR)
3 DIAGNOSTIC (DPR)
4 DIAGNOSTIC (DPR)
5 MAINTENANCE (DPR)
6 MAINTENANCE (ANL)
7 MANUAL MEASURE
1: 2: 3:WARM 4: 5: 6: 7:END 8:
# SELECT No.>
DSD/00001/7194.5
16
_____________________________________________________________
1 DATE MARK 19
2 DATE & TIME SET
3 CLEANING DRIVE
1: 2: 3: 4: 5: 6: 7: 8:
17. In the Maintenance (DPR) menu select 1, Date Mark, then press enter ↵.
19. The date and time will be properly displayed and updated in the upper right corner of the CRT.
No further action is required.
Use the following steps to test for the change to the year 2000 and then reset to the current date.
20. Power on the instrument using the Power On switch on the front of the DPR.
***OPERATION MENU***
# SELECT No.>
_____________________________________________________________
21. From the Operation Menu select F5: DIAG then press the Yes key at the DIAG ? Yes or No
prompt.
***OPERATION MENU***
# SELECT No.>
_____________________________________________________________
DSD/00001/7194.5
17
1 DIAGNOSTIC (DPR)
2 DIAGNOSTIC (DPR)
3 DIAGNOSTIC (DPR)
4 DIAGNOSTIC (DPR)
5 MAINTENANCE (DPR)
6 MAINTENANCE (ANL)
7 MANUAL MEASURE
1: 2: 3:WARM 4: 5: 6: 7:END 8:
# SELECT No.>
_____________________________________________________________
1 DATE MARK 19
2 DATE & TIME SET
3 CLEANING DRIVE
1: 2: 3: 4: 5: 6: 7: 8:
23. In the Maintenance (DPR) menu select 1, Date Mark, then press enter ↵.
26. Enter the Year, Month, Date, Hour, Minute, and Second you wish within the year 2000.
Press the F4, Set, key.
The date and time you entered will be displayed in the upper right corner of the CRT.
When testing is complete, use the following steps to return to the current date.
27. From the Operation Menu select F5: DIAG then press the Yes key at the DIAG ? Yes or No
prompt.
***OPERATION MENU***
# SELECT No.>
_____________________________________________________________
1 DIAGNOSTIC (DPR)
DSD/00001/7194.5
18
2 DIAGNOSTIC (DPR)
3 DIAGNOSTIC (DPR)
4 DIAGNOSTIC (DPR)
5 MAINTENANCE (DPR)
6 MAINTENANCE (ANL)
7 MANUAL MEASURE
1: 2: 3:WARM 4: 5: 6: 7:END 8:
# SELECT No.>
_____________________________________________________________
1 DATE MARK 19
2 DATE & TIME SET
3 CLEANING DRIVE
1: 2: 3: 4: 5: 6: 7: 8:
29. In the Maintenance (DPR) menu select 1, Date Mark, then press enter ↵.
32. Enter the current Year, Month, Date, Hour, Minute, and Second. Press the F4, Set key.
The date and time you entered will be displayed in the upper right corner of the CRT. No further
action is required.
DSD/00001/7194.5
19
Year 2000 Conversion Instructions and Test Procedure
OLYMPUS AU800 Automated Chemistry Analyzer
25. Press the Power switch on the analyzer and then press the “Q” key while the * is displayed in
the upper left hand corner of the screen. The Dos C:> prompt will be displayed.
26. Type “date” and press enter↵. Follow the prompts to enter the correct date within the year
2000 and press enter↵.
27. Press the Em. Stop switch to remove power from the analyzer.
28. Press the Power On switch and then the Warm Up switch to restore power and load the
program on the analyzer. Check that the correct date is displayed in the upper right corner of
the screen.
Use the following steps to test for the change to the year 2000 and then reset to the current date.
29. Press the Power switch on the analyzer and then press the “Q” key while the * is displayed in
the upper left hand corner of the screen. The Dos C:> prompt will be displayed.
30. Type “date” and press enter↵. Follow the prompts to enter the date you wish within the year
2000 and press enter↵.
31. Press the Em. Stop switch to remove power from the analyzer.
Press the Power On switch and then the Warm Up switch to restore power and load the program
on the analyzer. Check that the date you entered is displayed in the upper right corner of the screen.
When testing is complete, use the following steps to return to the current date.
33. Press the Power switch on the analyzer and then press the “Q” key while the * is displayed in
the upper left hand corner of the screen. The Dos C:> prompt will be displayed.
34. Type “date” and press enter↵. Follow the prompts to enter the current date and press enter↵.
35. Press the Em. Stop switch to remove power from the analyzer.
Press the Power On switch and then the Warm Up switch to restore power and load the program
on the analyzer. Check that the current date is displayed in the upper right corner of the screen. No
further action is required.
DSD/00001/7194.5
20