Fundamentals On CPC and SMPS CPC-003-US-web
Fundamentals On CPC and SMPS CPC-003-US-web
Fundamentals On CPC and SMPS CPC-003-US-web
FUNDAMENTALS OF CONDENSATION
PARTICLE COUNTERS (CPC) AND
SCANNING MOBILITY PARTICLE SIZER™
(SMPS™) SPECTROMETERS
APPLICATION NOTE CPC-003 (US)
Introduction
When using aerosol measurement instruments, it is important to have a basic understanding of the
fundamental physics and calculations behind their operation. Condensation Particle Counters
(CPCs) are used to measure total number concentration of an aerosol, while Scanning Mobility
Particle Sizer (SMPS) spectrometers are used to measure size distributions of aerosols.
_____________________
TSI, TSI logo, and Aerosol Instrument Manager are registered trademarks of TSI Incorporated.
SMPS is a trademark of TSI Incorporated.
3772 CPC Counting Efficiency
100%
90%
80%
Counting Efficiency
70%
60%
50%
40%
30%
20%
10%
0%
1 10 100
Diameter (nm)
Total particle counts from the CPC for a certain sample time (usually 1s) are then divided by the
product of sample time and flow rate to determine the particle concentration as shown below:
[ ]
[ ] [ ]
Equation 1
It is important to remember that the particle concentration measured by the CPC is the total
number concentration of all particles that the CPC can detect. For example, using the curve shown
in Figure 2, 50% of 10 nm particles will be detected by a 3772 CPC. Because the CPC alone has no
ability to differentiate between particles of different sizes, the total number concentration cannot
be corrected using the CPC efficiency curve.
A DMA column consists of a center rod and an outer grounded tube. A high voltage is applied to the
center rod. The difference in voltage between the rod and the tube generates an electric field. The
aerosol sample is drawn into the space between the center rod and the outer tube and drawn
toward the bottom of the column. As the aerosol sample flows from the top to the bottom of the
column, the charged particles are either forced to the outer tube or drawn to the inner rod by the
electric field depending on their polarity. The uncharged particles pass through the column,
unaffected by the electric field.
-2-
Particles with a higher electrical mobility are drawn
towards the inner rod at a faster rate than those with a
lower electrical mobility. As seen in Figure 3, since the
particles flow toward the bottom of the column at a
constant rate, the electrical mobility of a particle
together with the strength of the electric field determine
where the particle will impact the center rod. For a
given voltage, particles with a high electrical mobility
will impact the rod sooner than those with a low
electrical mobility. There is a slit near the base of the
rod. Particles that would normally impact the rod in this
location pass through the slit and exit the DMA;
therefore, for a given voltage, only particles with a
certain electrical mobility will pass through the slit.
Thus, a DMA is used to classify polydisperse aerosol
particles based on electrical mobility since the voltage
on the rod can be directly related to the electrical
mobility of the particles exiting the DMA. In an SMPS
system, a classifier scans across a voltage range while a
CPC counts the classified particles, generating a size
distribution.
Aerosol Neutralization
An aerosol neutralizer standardizes the sample aerosol by subjecting the sample aerosol to ionic
atmosphere, leaving the aerosol with a stationary state charge distribution. Historically, a
Boltzmann equilibrium distribution has been used to approximate the charge state of a neutralized
aerosol; however, this approximation assumes equilibrium and neglects image forces and the
increase in ion concentration at the surface of the particle and is only applicable in the particle
diameter range of 0.2 to 2 µm.
The Fuchs limiting sphere charge distribution model (1963) accounts for these deviations from
equilibrium and is widely accepted; however, it must be solved numerically due to its complexity.
Wiedensohler (1988) developed an approximation for the stationary state charge distribution of an
aerosol in a bipolar ionic atmosphere, accurate below 1 micrometer. The Boltzmann distribution,
Fuchs distribution, and Wiedensohler approximation for charge probability as a function of size are
shown in Figure 4 through Figure 6 for 0-2 charges.
-3-
Figure 4: Equilibrium 0 charge probability as a function of particle diameter
As seen in Figure 4, the majority of particles in the submicron range have no charge. Figure 5 shows
that in the range of particle sizes measured by the SMPS spectrometer, 1-20% of particles will have
1 charge and that above 200 nm, roughly 10% of particles are doubly charged.
-4-
SMPS Spectrometer Calculations
Aerosol Instrument Manager® software uses the DMA voltage, the CPC concentration, the CPC
efficiency curve, and the Wiedensohler equations to calculate the aerosol size distribution:
An aerosol neutralizer is used to apply a stationary-state charge distribution on the particles.
The electrostatic classifier is used to select particles of a single electrical mobility
The CPC counts the selected particles
The voltage on the DMA determines the electrical mobility of the particles that pass through the
DMA at a given time
The Wiedensohler approximation is used to determine the fraction of particles at a certain size
that are singly charged
The CPC efficiency curve is used to determine the fraction of particles of that size that are
detectable by the CPC
This charge efficiency fraction and CPC detectable fraction are combined with the fraction of
particles measured by the CPC to calculate the total concentration of particles of each size. These
size-resolved particle concentrations are combined to generate an aerosol size distribution.
Complicating Factors
There are several factors that affect the SMPS spectrometer concentrations that need to be
considered.
The first factor is that due to the Fuchs stationary-state charge distribution, only a small fraction
(approximately 1-20%) of particles are singly charged and make it through the DMA. Thus, any
variability in particle count or charging can be amplified depending on the size of the particle. This
is especially important at small particle sizes where only a small percentage of the particles
are charged and at very low concentrations where there is a higher uncertainty in counting.
demonstrates the dependence of counting uncertainty on size, where is the
√
relative counting uncertainty and C is the particle count (not concentration). Thus, at lower counts,
the uncertainty is higher.
Equation 2
√
When comparing total concentration as measured by a CPC to the total concentration as measured
by an SMPS spectrometer, concentrations must only be compared over the detectable size range of
the CPC. When measuring very small particles near the d50 of the CPC, the CPC may not count all of
the particles in this range; the SMPS spectrometer will compensate for the CPC efficiency curve
because it is able to resolve particle size.
Other factors that may or may not be relevant depending on particle size are diffusion loss, multiple
charges on the particles, and the shape of the particles.
Very small particles may be lost due to diffusion to the walls of the tubing in the sample system, the
tubing in the classifier, the DMA, the tubing between the classifier and the CPC and prior to the CPC
detector. The Diffusion loss correction option in Aerosol Instrument Manager Software takes into
account the flow rate and dimensions of the aerosol transfer path in the classifier and CPC and
corrects the data based on the number of particles that are theoretically lost due to diffusion in the
system (see application note SMPS-001).
As seen in Figure 4 through Figure 6, especially for larger particles, a significant fraction of particles
may carry multiple charges. Aerosol Instrument Manager software can take these probabilities into
account and make a correction for multiply charged particles.
-5-
Particle Detectors
Aerosol Instrument Manager software assumes all of the particles measured are spherical.
However, some aerosols, such as soot, are composed of chain aggregates made up of smaller
particles. To correct for the variation in electromobility in chain aggregates compared to spherical
particles, the aggregate mobility analysis tool within Aerosol Instrument Manager software can be
used (see application note SMPS-002). This is especially important when calculating the surface
area, volume and mass of the particles. For many particles, the assumption that the particles are
spheres is either valid or is a close approximation to the shape of the particle. For chain aggregates,
such as soot, this assumption is clearly not valid and the aggregate mobility analysis tool can result
in an improved calcuation of these parameters.
Summary
An SMPS system and a CPC have two separate purposes. A CPC is designed to measure total particle
number concentration, while an SMPS spectrometer is used to measure a quantitative size
distribution of a polydisperse aerosol. Because of the differences in principles between the
operation of an SMPS spectrometer and a CPC, the total particle concentration calculated using an
SMPS spectrometer differs slightly from the total particle concentration measured by a CPC,
especially at low concentrations.
Key References
N. A. Fuchs, On the Stationary Charge Distribution on Aerosol Particles in a Bipolar Ionic
Atmosphere, Geophys. Pure Appl. 1963, 56, 185-193
Wiedensohler, An Approximation of the Bipolar Charge Distribution for Particles in the
Submicron Size Range, J. Aerosol. Sci. 1988, 19, 387-389
SMPS Spectrometer Bibliography
Aerosol Statistics Lognormal Distributions and dN/dlogDp (Application note PR-001)
Scanning Mobility Particle Sizer (SMPS) Spectrometer Diffusion Loss Algorithm (Application
note SMPS-001)
Scanning Mobility Particle Sizer (SMPS) Spectrometer Nanoparticle Aggregate Mobility Analysis
Software Module (Application note SMPS-002)
More information can be found at www.tsi.com.