TOC L Quick Reference Guide PT
TOC L Quick Reference Guide PT
3
Instrument Start-Up
4
Instrument Start-Up
1. Open the TOC sample table editor and open a new sample table
2. Select the correct hardware configuration
3. Press ‘OK’
5
Instrument Start-Up
Press ‘Connect’ to start initializing the instrument
6
Instrument Start-Up
It Takes approximately 30-45 minutes for the instrument to stabilize and come to a ready status.
Push the ‘Monitor’ button to view the instrument status in real time. When all parameters are ok the
instrument is ready to start running samples.
9
Software Administration Functions
10
Software Administration Functions
If audit trail is enabled, log browser allows you to
search the audit trails.
11
Software Administration Functions
12
Software Administration Features
Allows you to customize
LIMS Settings
13
Daily Checks
14
Daily Checks
16
Replacing Catalyst
After the combustion tube has been removed the catalyst can be removed and washed or if it is
consumed can be replaced with new. Quartz combustion tubes can be rinsed with DI water and reused
until they break.
17
Replacing Catalyst
18
Replacing Catalyst
19
Replacing Catalyst
20
Replacing Catalyst
TOC-LC*N TOC-LC*H
21
Replacing Catalyst
22
Replacing the Halogen Scrubber
23
Replacing the Syringe / Plunger Tip
24
Replacing the Syringe / Plunger Tip
25
Replacing the Syringe / Plunger Tip
26
Replacing the TC Slide Injection O-rings
Open the lid on the top of the TOC and locate the
TC injector block.
27
Replacing the TC Slide Injection O-rings
28
Replacing the TC Slide Injection O-rings
29
Replacing the IC Slide Injection O-rings
Open the front door of the TOC and locate the 1. Remove the thumb screw connecting
sliding IC sample injector the PTFE slider to the motor drive.
30
Replacing the IC Slide Injection O-rings
31
Replacing the Rotor in the 8-port Valve
32
Replacing the Rotor in the 8-port Valve
33
Replacing the CO2 Scrubber
Disconnect the flexible tubing from the old CO2 scrubber and dispose of the scrubber per
your local waste disposal regulations. Follow instructions below for installing the new
scrubber.
34
Replacing the IC Vessel Pump Head
35
Replacing the ASI Rinse Pump Head
36
Replacing the TC Injection Tubing
37
Replacing the TC Injection Tubing
38
Replacing the TC Injection Tubing
39
Replacing the ASI Needle
40
Replacing the Detector Gasket
2. Loosen the two screws on the back of the TNM-L that hold the cover on.
41
Replacing the Detector Gasket
42
Replacing the Ozone Killing Catalyst
43
Replacing the Ozone Killing Catalyst
44
Replacing the Ozone Killing Catalyst
45
Removing the Covers
Section 1 / 2 46
Removing the Covers
Remove the screws and carefully
remove the rear cover first.
CAUTION
Section 2 / 2 47
Replacing the Catalyst
Section 1 / 7 48
Replacing the Catalyst
Section 2 / 7 49
Replacing the Catalyst
Section 3 / 7 50
Replacing the Catalyst
Section 4 / 7 51
Replacing the Catalyst
1. Remove the sampling port
CAUTION
Section 5 / 7 52
Replacing the Catalyst
Section 6 / 7 53
Replacing the Catalyst
CAUTION
Section 7 / 7 54
Replacing the O-rings
Section 1 / 3 55
Replacing the O-rings
Section 2 / 3 56
Replacing the O-rings
Section 3 / 3 57
Creating a Hardware Configuration
58
Creating a Hardware Configuration
59
Creating a Hardware Configuration
60
Creating a Hardware Configuration
61
Creating a Calibration Curve
1. Push the calibration curve list button
2. Push the ‘New’ button
3. Select the correct hardware configuration
62
Creating a Calibration Curve
63
Creating a Calibration Curve
64
Creating a Calibration Curve
65
Creating a Calibration Curve
67
Creating a Calibration Curve
68
Creating a Method
1. Push the ‘Method’ button to show the 3. Select the correct hardware
list of saved methods. configuration.
2. Push the ‘New’ button. 4. Select ‘Next’
69
Creating a Method
3. Select ‘Next’.
70
Creating a Method
4. Push ‘Next’
71
Creating a Method
72
Creating a Method
73
Creating a Method
4. Push ‘Next’
74
Creating a Method
75
Creating a Method
76
Creating a Method
77
Creating Control Samples
1. Push the ‘Control Samples’ button to show the list of control samples
2. Push the ‘New’ button to create a new control sample
3. Select the correct hardware configuration from the drop down list
4. Press ‘Next’
78
Creating Control Samples
3. Press ‘Next’
79
Creating Control Samples
5. Push ‘Next’
80
Creating Control Samples
4. Press ‘Next’
81
Creating Control Samples
Press ‘Next’
82
Creating Control Samples
Press ‘Next’
83
Creating Control Samples
Recovery %
Note: when ‘Spiked’ is selected when you insert the control into
the sample table, two lines in the table will be entered. The
first line is for the un-spiked sample and the second line is for
the spiked sample. The result of the sample spike is calculated
as the difference between these two values.
85
Creating Control Samples
86
Creating Control Samples
87
Creating Control Samples
88
Creating Control Samples
89
Creating Control Samples
Select ‘Finish’ to save the control sample template to the list of control
samples.
90
Creating a Schedule
91
Creating a Schedule
92
Creating a Schedule
3. Push ‘Save’
93
Creating a Schedule
A schedule file can then be inserted into a sample table. Schedule files are useful for labs where the
same samples are run often.
94
Creating a Sample Table
1. Push the ‘Sample Table’ button to show the list of saved sample tables.
2. Push the ‘New’ button to open a new table
3. Select the correct hardware configuration from the drop down list
4. Press ‘OK’
95
Creating a Sample Table
Push the ‘Calibration Curve’ button to display the saved calibration curves and drag
and drop the desired Calibration curve into the sample table
96
Creating a Sample Table
Highlight the next row in the table and right click on the mouse. From the drop
down list select ‘Insert – Sample’
97
Creating a Sample Table
4. Push ‘Open’
5. Push ‘Next’
98
Creating a Sample Table
5. Press ‘Next’.
99
Creating a Sample Table
100
Creating a Sample Table
101
Creating a Sample Table
Press ‘Next’
102
Creating a Sample Table
103
Creating a Sample Table
104
Creating a Sample Table
Add as many sample to the table as needed. Note that multiple samples can be added to
the table at once by selecting ‘Insert-Multiple Samples’
105
Creating a Sample Table
Push the ‘Control Sample’ button to show the list of saved control samples.
Drag and drop the appropriate control sample into the sample table.
106
Creating a Sample Table
107
Creating a Sample Table
108
Starting a Sample Run
109
Starting a Sample Run
4. Push ‘Start’
110
Starting a Sample Run
111
Editing a Sample Table
112
Viewing Sample Results
Sample results can be found in the ‘Result’ column of the sample table. Note that samples that are
using a Method with multiple analysis types show results for all measurements in the ‘Result’ column.
113
Viewing Sample Results
114
Viewing Sample Results
115
Viewing Sample Results
116
Viewing Sample Results
117
Viewing Sample Results
118
Viewing Sample Results
119
Viewing Calibration Curves
The TOC-L software will create two different types of calibration curve files.
120
Viewing Calibration Curves
121
Viewing Calibration Curves
OR
3. Push ‘Open’
122
Viewing Calibration Curves
123
Creating a Sample Report
124
Creating a Sample Report
To customize a report select ‘File’ then ‘Page Setup’. In the page setup window select the items to be
displayed in the report.
125
Creating a Sample Report
126
Solid Sample Module Analysis
Section 1 / 8 127
Solid Sample Module Analysis
Section 2 / 8 128
Solid Sample Module Analysis
Section 3 / 8 129
Solid Sample Module Analysis
Section 4 / 8 130
Solid Sample Module Analysis
Section 5 / 8 131
Solid Sample Module Analysis
Once a calibration curve for solid samples has been
created, samples can be entered into the sample
table just like for liquid samples.
Section 6 / 8 132
Solid Sample Module Analysis
After sealing the sample boat into the sample port chamber you need
to wait 3-5 minutes before starting analysis. This is the time it takes
for the air trapped in the system to work through the detector. Failure
to do so will result in false peaks from the CO2 in the air
compromising measurement of the sample.
Section 7 / 8 133
Solid Sample Module Analysis
Section 8 / 8 134
Total Nitrogen Module (TNM-L)
With the optional TNM-L total nitrogen module, the analysis of the
total nitrogen in a sample can be performed. Please note that it can
only measure TOTAL nitrogen and cannot speciate out the different
forms of nitrogen present.
135
Total Nitrogen Module (TNM-L)
136
Total Nitrogen Module (TNM-L)
137
Total Nitrogen Module (TNM-L)
138
Total Nitrogen Module (TNM-L)
When samples are measured using a method for both carbon and nitrogen, the instrument will
perform one injection and measure both carbon and nitrogen simultaneously. Results for both
analysis will be displayed in the ‘Result’ of the Sample Table.
139
Total Nitrogen Module (TNM-L)
140
Pharmaceutical Water Testing
141
Pharmaceutical Water Testing
142
Pharmaceutical Water Testing
143
Pharmaceutical Water Testing
144
Pharmaceutical Water Testing
145
Pharmaceutical Water Testing
146
Pharmaceutical Water Testing
147
Pharmaceutical Water Testing
148
Pharmaceutical Water Testing
149
Pharmaceutical Water Testing
150
Pharmaceutical Water Testing
151
Pharmaceutical Water Testing
152
Pharmaceutical Water Testing
153
Pharmaceutical Water Testing
154
Pharmaceutical Water Testing
155
Pharmaceutical Water Testing
156
Pharmaceutical Water Testing
157
Pharmaceutical Water Testing
158
Cleaning Validation with the SSM
159
Cleaning Validation with the SSM
160
No Peaks
Probably the most common issue with the TOC-L is when user’s see no peaks
during analysis or no instrument response. This problem it almost always
caused by one of two things:
161
No Peaks
To check to make sure sample is being injected into the combustion tube, start a sample analysis and open the
front door of the TOC and watch what the instrument does.
Watch the 8-port valve and the syringe to make sure the instrument is pulling in sample and then injecting it into
the combustion tube via port-7 of the 8-port valve. You can tell which port the valve is on by watching the “O”
mark on the valve. Whichever port the “O” lines up with is the port it is currently on.
Open the lid on top and locate the top of the combustion tube that
sticks up out of the furnace. Make sure that when the slider moves
forward over the combustion tube you see sample being injected into
the combustion tube. If no sample is being pulled up and injected into
the combustion tube:
162
No Peaks
To perform a leak check, pinch the “L” tubing going to the CO2 scrubber blocking any flow of gas that may be going
through that tube. While pinching the tube, check for bubbling in the drain pot on the left hand side of the instrument.
If there is NO bubbling in the drain pot then there is a gross leak of carrier gas somewhere in the system. If
there IS bubbling in the drain pot then there is no leak of carrier gas in the system.
163
No Peaks
The most common spots for leaks of carrier gas are in or around the combustion
tube. If it is determined there is a leak of carrier gas:
1. shut off the furnace and allow it to cool before proceeding.
1. Remove the combustion tube and check to make sure it is not cracked or broken and replace if necessary.
164
High Baseline Position
A high baseline is any baseline position above 200 mV. When the baseline is above 200 mV the instrument will give a
‘NOT READY’ status and will not allow a run to be started. At this point action needs to be taken to lower the baseline
below 200 mV. There are several things that can cause a high baseline including:
1. A leak of carrier gas. The TOC measures CO2 so if there is a leak of carrier gas, no clean gas is flowing through the
detector so it becomes filled with ambient air. The instrument will then measure the CO2 in ambient air resulting
in a high baseline position.
2. A dirty detector cell. The detector cell is gold lined for maximum transmission of light to the detector. Over time
the gold lining can become dirty, corroded, or even scratched, leading to a high baseline. In this scenario the
baseline will usually creep up over time and when it gets to 200 mV, will no longer give the ‘READY’ status. Please
contact Shimadzu for technical assistance.
3. Contaminated carrier gas. Carrier gas that contains high levels of CO2 or other hydrocarbons can lead to a high
baseline. If using compressed cylinders, it is possible to get the occasional bad cylinder of air. If using a TOC gas
generator, check the filters and scrubbers on the generator and replace if necessary.
4. A faulty detector or light source. This is the least common scenario as the detector and light source are quite
robust typically lasting for ten years or more. This scenario usually results in a baseline being so high or so low
that it cannot even be seen on the background monitor page. Please contact Shimadzu for technical assistance.
165
Poor Reproducibility
It is common in TOC analysis to perform multiple injections for each sample and then average those results together
for the final result. Replicate injections should have low CV (RSD) values, and low SD values. When these values
become large, action should be taken.
166
Poor Reproducibility
There are many things that can cause poor reproducibility including:
1. A worn out syringe and/or plunger tip
167
Poor Recovery of Control Samples / Check Standards
1. Bad or consumed catalyst. Remove the combustion tube and replace the catalyst.
2. Clogged ASI needle or ASI tubing. Check the needle and tubing and replace if necessary.
3. Worn out syringe or syringe plunger tip. Check and replace if necessary.
6. Improperly prepped Calibration Standards. Check the calibration standards and re-prep if necessary.
Then re-run the calibration curve.
168
Poor Recovery of Control Samples / Check Standards
High recovery of Control Samples or Check Standards is much less common than low recovery however
High recovery could be caused by:
1. Instrument contamination. If instrument contamination is suspected, remove and clean these parts:
⚫ The ASI needle
⚫ The syringe
⚫ The 8-port valve rotor
⚫ The combustion tube
⚫ The TC injection tubing
2. Instrument Carry-over. If Carry-over between sample is suspected, increase the number of washes
and/or ASI needle rinses.
3. Improperly prepped Calibration Standards. Check the calibration standards and re-prep if necessary.
Then re-run the calibration curve.
169
High Blank Values
1. Contamination in the instrument. If instrument contamination is suspected, remove and clean these
parts:
⚫ The ASI needle
⚫ The syringe
⚫ The 8-port valve rotor
⚫ The combustion tube
⚫ The TC injection tubing
2. Contaminated DI water. Replace the dilution water bottle and ASI rinse bottle with freshly collected
DI water.
3. Contaminated acid. Replace the acid with a new or freshly prepped acid.
4. Not enough ASI needle rinses or syringe /flow line washes. Click to the next pages for instructions on
where to increase these values.
170
High Blank Values
In the TOC-L software, ‘Washes’ refers to the number of times the instrument washes the syringe, valve, and flow lines with
SAMPLE prior to analysis. These values can be changed by opening either an existing calibration curve file or method file.
171
High Blank Values
In the Hardware Settings, the number of times the auto-sampler rinses the needle with DI
rinse water can be set. Increase this value for really dirty samples or if needle contamination
and carryover is suspected.
172
ERROR: Syringe Pulse Count / Syringe Pump
173
ERROR: 8port Valve Pulse Count / 8port Valve Position
Either of these error messages signifies an issue with the 8-port Valve
drive in the TOC.
174
ERROR: TC / IC Slider Position
175
ERROR: ASI Arm Vertical Pulse Count / ASI Arm Vertical
Position
Either of these error messages signifies an issue with the ASI arm in the auto-sampler in relation to the up
and down movement.
3. Contact Shimadzu Service. Removing the covers to get down to the motor drives in the ASI is difficult
and requires a trained Shimadzu Service Engineer.
176
ERROR: ASI Arm Horizontal Pulse Count / ASI Arm
Horizontal Position
Either of these error messages signifies an issue with the ASI arm in the auto-sampler in relation to the forward and
backward movement.
3. Contact Shimadzu Service. Removing the covers to get down to the motor drives in the ASI is difficult and
requires a trained Shimadzu Service Engineer.
177
ERROR: ASI Tray Pulse Count / ASI Tray Position
Either of these error messages signifies an issue with the sample tray in the auto-sampler.
3. Contact Shimadzu Service. Removing the covers to get down to the motor drives in the ASI is difficult and
requires a trained Shimadzu Service Engineer.
178
Mechanical Check
Mechanical check allows users to check all the mechanical functions of the TOC-L. It can be very useful
for troubleshooting and diagnosing mechanical issues. To access the mechanical check, select
‘Instrument’ > ‘Maintenance’ > ‘Mechanical Check’.
179
Mechanical Check
180
Mechanical Check
181
Troubleshooting TNM-L Issues
Section 1 / 5 182
Troubleshooting TNM-L Issues
Section 2 / 5 183
Troubleshooting TNM-L Issues
⚫ The TNM-L is downstream of the TOC-L. Sample passes through the TOC-L where carbon Is measured and then enters the TNM-
L where nitrogen is measured.
⚫ When troubles arise with nitrogen measurement such as no peaks or poor reproducibility it is first a good idea to check the
carbon measurement.
⚫ If carbon measurement is normal and looks good then the issue with the TNM-L measurement lies within the TNM-L so
troubleshoot the TNM-L
⚫ If carbon measurement is not good then the issue with TNM-L measurement is likely a result of something within the TOC-L
so troubleshoot the TOC-L
Section 3 / 5 184
Troubleshooting TNM-L Issues
⚫ Check the hours on the Ozone Generator. The ozone generator does have a lifespan and if no ozone is being generated then
there will be no peaks for nitrogen measurement. Replacement of ozone generator is performed by Shimadzu Service
Engineer. Please contact your local Shimadzu Office.
Section 4 / 5 185
Troubleshooting TNM-L Issues
⚫ Check black tubing that connects the detector of the TOC-L to the reactor on the TNM-L and make sure the connection is
good at the TNM-L reactor.
⚫ Pull apart the TNM-L reactor assembly and inspect and/or replace the detector gasket.
Section 5 / 5 186
Troubleshooting SSM Issues
Section 1 / 2 187
Troubleshooting SSM Issues
Section 2 / 2 188
Principles of Analysis
Section 1 / 4 189
Principles of Analysis
Section 2 / 4 190
Principles of Analysis
Section 3 / 4 191
Principles of Analysis
Section 4 / 4 192
TOC-LC*H
Section 1 / 4 193
TOC-LC*H + TNM-L + Ext. Sparge Kit
Section 2 / 4 194
TOC-LC*N
Section 3 / 4 195
TOC-LC*N + TNM-L + Ext. Sparge Kit
Section 4 / 4 196
Preparation of Standard Solutions
Section 1 / 8 197
Storage of Standards
Section 2 / 8 198
TC/NPOC Standard Prep
Section 3 / 8 199
IC Standard Prep
Section 4 / 8 200
TN Standard Prep
Section 5 / 8 201
TC/TN Mixed Standard Prep
Section 6 / 8 202
Hydrochloric Acid Solution
Section 7 / 8 203
IC Reagent (Phosphoric Acid Solution)
Section 8 / 8 204