AXIOM Aristos Image System
VB21 and MX, VX, VX Plus, TX, FX, FX Plus System SW;
© Siemens, 2011
© Siemens, 2011
All documents may only be used for rendering
services on Siemens Healthcare Products. Any
document in electronic form may be printed
once. Copy and distribution of electronic docu-
ments and hardcopies is prohibited. Offenders
will be liable for damages. All other rights are re-
served.
2011
2 Copyright / Version / Disclaimer
1Copyright / Version / Disclaimer
Copyright
“© Siemens, 2011“ refers to the copyright of a Siemens entity such as Siemens Aktienge-
sellschaft - Germany, Siemens Shenzhen Magnetic Resonance Ltd. - China, Siemens
Shanghai Medical Equipment Ltd. - China, Siemens Medical Solutions USA Inc. - USA,
Siemens Healthcare Diagnostics Inc. - USA and/or Siemens Healthcare Diagnostics Prod-
ucts GmbH - Germany.
Document Version
Siemens reserves the right to change its products and services at any time.
In addition, manuals are subject to change without notice. The hardcopy documents corre-
spond to the version at the time of system delivery and/or printout. Versions to hardcopy
documentation are not automatically distributed. Please contact your local Siemens office
to order current version or refer to our website http://www.healthcare.siemens.com.
Disclaimer
Siemens provides this documentation “as is“ without the assumption of any liability under
any theory of law.
The installation and service of equipment described herein requires superior understand-
ing of our equipment and may only be performed by qualified personnel who are specially
trained for such installation and/or service.
1 _______ Requirements___________________________________________________ 6
Manufacturer’s Note . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
How to use these Instructions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Screen Shots & Graphics in this Document . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Required Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Protection of Electrostatically-sensitive Components . . . . . . . . . . . . . . . . . . . . . . . . . 7
Required Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
General remarks. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Special Tools / Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
For Collimator SW Download . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Collimator Software Download . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Log files. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Backup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Uninstalling System Management (MNP) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Booting the COBR Disk . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Installation with the Clone CD or DVD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Setting or Checking the IP Address . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Checking the Time Zone and Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Resetting Regional and Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Installing the Aristos Service SW on the Radis PC . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Generating the new XCU Floppy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Installation of the VB21 Patch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Installation of DICOM Attrib. DAC Configurator (only if used) . . . . . . . . . . . . . . . . . . . . 33
RADIS, Initial Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Import Licenses for additional Functions (only if new functions are to be added) . . . . . 35
RADIS, Final Setup, Setting Firewall, Security, and Radis Log . . . . . . . . . . . . . . . . . . . 36
If needed: Deactivating the Serial Mouse at COM 1 (only for Aristos MX / VX /TX) . . . 37
Updating the OT 16-bit Collimator N FD II (P/N 10092604 only) to VC00N . . . . . . . . . . 41
Downloading the new XCU Floppy (only for Aristos MX / VX /TX) . . . . . . . . . . . . . . . . . 43
Updating the Aristos FX System SW (only for FX/FX Plus) . . . . . . . . . . . . . . . . . . . . . . 44
Downloading the FX VA04 SW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Downloading the new XCU Floppy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Checking and Configuring the new SW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Checking and Activating System Management (only if remote is configured) . . . . . . . . 49
Activating System Management . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Restoring the Customer OGPs (for all systems) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Checking for RIS Match in the OGP Editor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Ortho (necessary only when installing for the first time). . . . . . . . . . . . . . . . . . . . . . . . . 54
General Remarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Creation of a repair plan (usually by USC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
What to do before Reinstallation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
On-site Work Steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Log files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Perform a virus scan if a virus is suspected (see the repair plan, if available). . . . . . 71
Backup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Booting the COBR Disk. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Reinstallation with a Recovery Medium . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Checking the latest VB21 Patch according to the Repair Plan . . . . . . . . . . . . . . . . . . . . 78
Installing the Aristos Service SW on the Radis PC. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Restore of the syngo and Radis Configuration, if the restored Image is older than the Back-
ups . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Check for the latest Licenses (according to the Repair Plan) . . . . . . . . . . . . . . . . . . . . . 81
Checking and Activating System Management (only if remote is configured) . . . . . . . . 82
Activating System Management . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
Restoring the Customer OGPs (only if the OGPs are newer than in the Recovery Image)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Checking for RIS Match in the OGP Editor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Generating Backups if changes have been made. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
Creating a current Recovery Image, if changes have been made . . . . . . . . . . . . . . . . . 88
Labeling the Recovery Image(s) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
Final check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
Informing the Customer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
Reporting Errors by the Customer or CSE (for information only) . . . . . . . . . . . . . . . 92
General Remarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
Creation of a repair plan (usually by USC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
What to do prior to Reinstallation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
Collimator Software Download (only if coming from VB10) . . . . . . . . . . . . . . . . . . . . 95
On-site Work Steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
Retrieve Log Files from a VB10x PC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
Screen shots and graphics used in this document should be viewed as examples only.
They cannot describe all possible input combinations; they have an informative character
and are used only for better orientation.
The actual parameters that must be configured can be found in the text and the tables and
depend on the requirements and configuration (items included in the shipment) of the indi-
vidual system.
Requirements 1.1
Required Tools 0
Safety Information 0
The “Safety Information“ contained in the system manual and in the particular unit instruc-
tions must be observed.
Integrated circuits and circuit boards on which MOS components are installed call for par-
ticularly careful handling because of their highly sensitive electrical structures and because
of their extremely high input resistance. The ESD regulations that apply for electrostati-
cally-sensitive components must be observed.
Required Parts 0
The table provides an overview of the required CDs/DVDs for different software installation
situations. These are:
• New installation (from scratch), e.g. upgrade (retrofit) from VB10... to VB21C and
higher
• Re-installation, e.g. because the existing software crashed
• Latest patch beginning with VB21C
VA20C COBR x x
Aristos RADIS VB21 Patch CD x x x
Aristos Recovery CDs (blank) x
Aristos I.S. VB21A 460 Clone CD x
from initial installation: recorded x
recovery CDs
Stand system SW, and backups x x
Tab. 2 PC M410
VA20C COBR x x
Aristos RADIS VB21 Patch CD x x x
Aristos Recovery CDs (blank) x
from initial installation: recorded x
recovery CDs
Aristos I.S. VB21A M410 Clone CD x
Stand system SW, and backups x x
Tab. 3 PC M430
VA20C COBR x x
Aristos RADIS VB21 Patch CD x x x
Aristos Recovery DVD (blank) x
from initial installation: recorded x
recovery DVDs
Aristos I.S. M430 VB21A Clone x
DVD
Stand system SW, and backups x x
VA20C COBR x x
Aristos RADIS VB21 Patch CD x x x
Aristos Recovery DVDs (blank) x
from initial installation: recorded x
recovery DVDs
Aristos I.S. M440 VB21A Clone x
DVD
Stand system SW, and backups x x
Tab. 5 PC M460
This procedure describes the steps necessary to install the Radis VB21 SW on the AXIOM
Aristos Image System for the 1st time, e.g. only if VB10 or VB20 was on the system
before.
• The following is an outline of some of the work steps that will be covered in this chapter:
Tab. 6 Work Steps
Check the Med CS Intranet/Extranet for the latest version of this document (For Service ->
Product Information) or contact your local Siemens office if you do not have access to the
Siemens Intranet/Extranet.
To avoid extra work, check the XCS SSW to determine whether the collimator SW is lower
than VC00N. If so, follow the following work steps. If not, skip this paragraph.
• Go to the download location on the CS Home page: For Service > Product Informa-
tion > AX Systems > Radiography > Direct Technique > Software.
• The required download file is the 10092710_OT_II_VC00N.exe and is located under
the description Collimator G052G.
• Double-click on it and unzip it to a folder of your choice on the C or D drive of your ser-
vice PC.
• You will find two files in the folder: CollimatorSWLoader.exe and
SW_D21_TBN_OT_II_VC00N.H86. Keep them there for later use.
Preparations 2.2
Log files 0
1. Unlock the access to Windows® with Tab - Del - Transfer image (“+”).
2. Press the Windows® key, and then Run; enter radissysinfo > OK.
3. Press Start.
5. The log file will be created and saved to the selected directory.
8. Copy the compressed log file (2/Fig. 5 / p. 15) to a USB stick or similar medium, and
keep it ready in case it is requested by RSC/HSC/QIB/DCU.
¹ The log files look like this: “Log<date>_<time>.zip“
9. If a customer complaint has to be filed, send the file along with the complaint to Health-
care AX Complaint.
Backup 0
The following configurations have to be saved manually because they are related to Win-
dows® and therefore are not backed up with the syngo backup.
• Make a note of the service key:
- Options > Service > Local Service...
Service Key
• Make a note of the host name (= computer name) and the IP addresses:
- Options > Service > Local Service... > Configuration > TCP/IP LAN
- Home > close the window.
IP address
Subnet mask
Default gateway
Customer Configuration:
- > Options > Configuration.. > Regional and Language...
- “Regional Options”: Make a note of the date and time format.
- “ Languages > Details...”: Make a note of the language and keyboard.
Date Format
Time Format
Language
Keyboard (input locales)
Network Speed
Select the Advanced tab card, read the value for Speed & Duplex or Link
Speed & Duplex and make a note of it.
The following work steps describe how to create backups of “SW Settings 02” on CD, “SW
Settings, Calibration” on CD, “Customer OGPs” on floppy (or USB stick) (the other pack-
ages do not have to be saved because there are newer versions with the new VB21 SW).
• To do this, go to:
> Options > Service > Local Service > Backup & Restore
Server name
TCP/IP address
• Options > Service > local Service > Configuration > System Mgmt.
Deselect all check boxes in the Agent Control window and start with Go in the Action
Bar.
Uninstalling System Management (MNP) in VB10x:
• The following screen displays the restore progress (Fig. 16 / p. 25). This will take a
while ...
- If the medium set consists of more than one CD/DVD, you will be prompted to insert
the next medium.
¹ Follow the instructions on the screen.
- When the restore has ended, a success message is shown in a new screen
(Fig. 17 / p. 25).
- Close the success window with <OK>.
- Remove the Ghost image from the drive and close the tray.
¹ The window will close and the PC starts to reboot.
• After restart, enter only the host name (= computer name) that was noted before.
Do NOT change the password.
• Go to Start > Settings > Control Panel > Switch to Classic View > Network Con-
nections.
• Highlight “Local Area Connection” and right mouse-click on “Properties”.
• Highlight “Internet Protocol (TCP/IP)” and click on “Properties”.
• Check or set “Use the following IP address”. (There must be a fixed IP address and
NOT DHCP.) Insert the correct IP address (noted down before).
• Confirm by clicking on “OK”.
• Set the network speed to the value noted down before:
Without selecting anything, click on Configure.
Select the Advanced tab card and set the value for Speed & Duplex or Link Speed
& Duplex to the value previously noted.
• Go to Start > Settings > Control Panel > Date and Time.
• Go to the “Time Zone” tab card and set the correct time zone.
• Place a checkmark in “Automatically adjust clock for daylight saving changes”.
• Confirm by clicking on “Apply”.
• Go to the “Date & Time” tab card and set the correct date and time.
• Confirm by clicking on “OK”.
• Insert the software CD-Rom according the Compatibility List for the AXIOM Aristos
MX/VX, TX or FX/FX Plus into the CD drive.
• Open “Windows® Explorer”, select the CD-Rom drive and double-click on the
“start.exe” file. The Installer window opens.
• Click on the link under "Service PC (+ Download Software) ...".
• If queried, choose Full Install.
• Choose Continue > OK.
• For Aristos FX/FX Plus: Insert a blank, formatted floppy into the floppy drive of the
RADIS PC and click on "VA04B" (or higher) under "XCU floppy Software of Aristos FX".
A copy routine is started and the floppy is created. When doing this, observe the sys-
tem messages and, where applicable, acknowledge them.
• For Aristos MX/VX: Insert a blank, formatted floppy into the floppy drive of the RADIS
PC and click on "VC01B" (or higher) under "XCU floppy Software of Aristos MX".
A copy routine is started and the floppy is created. When doing this, observe the sys-
tem messages and, where applicable, acknowledge them.
• For Aristos VX Plus: Insert a blank, formatted floppy into the floppy drive of the RADIS
PC and click on "VA00C" (or higher) under "XCU floppy Software of Aristos VX Plus".
A copy routine is started and the floppy is created. When doing this, observe the sys-
tem messages and, where applicable, acknowledge them.
• For Aristos TX: Insert a blank, formatted floppy into the floppy drive of the RADIS PC
and click on "VA00M" (or higher) under "XCU floppy Software of Aristos TX".
A copy routine is started and the floppy is created. When doing this, observe the sys-
tem messages and, where applicable, acknowledge them.
• Label the created floppy appropriately. Keep it for later use.
• Close the Service SW windows and remove the CD and floppy.
To upgrade from the initial VB21A or B version to VB21C or higher, follow the steps below.
• Insert the Aristos RADIS VB21C Patch or higher CD into the CD drive.
• Navigate to the CD/DVD drive with the Windows® Explorer.
• Double-click on ParamUpdate.exe.
• The patch installation will begin; installation may take 15 ...20 sec.
• Click on OK.
• If the system is using a DAC, install the DAC program from the "DICOM Attribute Con-
fig. VA20A", P/N 7264435 or higher.
• Follow the "DICOM Attrib. Configurator" Instructions, AX00-000.816.01.02… or higher.
• To do this:
- Go to Options > Service > Local Service... > Configuration
- > Next
- > Licensing
- Insert the data medium with the new License.dat (it is possible that the new license
file still needs to be renamed to “license.dat”).
- Click on “Browse..” and go to the new “license.dat”.
- Click on “Open” > “Import License file” and confirm with “OK”.
- Check in the Licenses Window (above) if the added license is shown and has in X in
the “Avail” column.
- Select “Home” and close the window.
- Reboot the Radis PC: Options > End Session... > Restart System... > Yes
RADIS, Final Setup, Setting Firewall, Security, and Radis Log 2.13
• Unlock Windows® by pressing “Tab + Del. + Send (Num field)” and open the Win-
dows® Explorer.
• Navigate to C:\post_clone_tools\Postclone.bat.
• Double-click on C:\post_clone_tools\Postclone.bat.
> OK.
• Connect all cables.
In some cases, the mouse cursor sticks in the top right corner when the RADIS PC is con-
nected to the XCU; this is because the RADIS PC thinks that an RS232 mouse is con-
nected to its RS232 interface and interprets the signals from the XCU as a "mouse
command". To prevent this, the RS232 mouse must be disabled on the RADS PC.
• Select Start > Run > "devmgmt.msc" using the "Windows® key", the arrow keys and
the Enter key.
Fig. 20:
• After starting the Device Manager, select the "Mice and other pointing devices"
hardware components and select the context menu with the right mouse button.
• Connect the serial XCU service cable to the Radis PC.
• Select "Scan for hardware changes".
- If the serial cable is connected, the serial mouse driver will be found and the mouse
cursor will begin to move by itself.
- If the “mice and other pointing devices” was not found, repeat “scan for hard-
ware changes” several times.
Fig. 21:
• Press Alt + Tab to activate the Device Manager window.
• Use the up/down arrow keys to navigate to "Mice and other pointing devices" (use
left/right arrow keys to open the sub-menus).
Fig. 22:
• Select "Microsoft Serial Ball Point".
• Press Enter.
• Press Tab until "Device Usage" is selected.
• Use the up/down arrow keys to select "Do not use this device (disable)".
Fig. 23:
• The mouse for this hardware profile will now be deactivated for COM 1 and the Device
Manager will display a red cross in front of the deactivated component.
Fig. 24:
• Loosen the two No. 2 Allen screws (1/Fig. 25 / p. 41). Swing out the front panel of the
collimator.
If there is no FX, skip this chapter and go to (Checking and Activating System Management
(only if remote is configured) / p. 49).
NOTICE Before starting a download, disable the screen saver and any
power-saving features which might shut down the PC unex-
pectedly! This update takes approx. 30...45 min. and should
be done parallel to the Radis update using the Service PC.
Screen shots do not always show the correct software ver-
sions.
¹ Follow chapter headlines, notes and hints closely!
• Connect the Service PC to the gateway master (use the short adapter cable) or use the
Radis PC.
If using the Service PC: Make sure that the latest FX Service SW is installed.
• If using the Radis PC, go to C:\SSW\Aristos\Aristos.exe to start the Aristos FX SSW:
- Log in with your name. The password is "_Siemens" (case-sensitive!).
• Perform a backup with SSW Aristos FX > Backup/Restore > Backup ...by pressing
the backup button.
• In the SSW, select Aristos FX > Backup/Restore > Download (1/Fig. 28 / p. 45).
• Check whether the clutch power settings are correct according to Speed Info 21929:
Set the clutch power to 0%....
To do so, enter the Aristos FX SSW in developer mode (a developer password for 2010
can be found in the Knowledge Base).
Click on SSW Aristos FX > Development > GW Config > Mechanical offset.
Click on Read. Change all values to 0 in the Clutch Power column.
Click on Save.
Switch the entire system off and on and check the function.
• Wait until the SW has downloaded. After the program has downloaded, an audible sig-
nal is sounded (TOOT-TA-TOOT) (three beeps).
• Remove the floppy.
• Reboot the XCU by unplugging and plugging in the "XCU X1" power connector.
• Switch the system off completely (hold the switch at the control cabinet for more than 3
sec.) and switch on again.
• Click on the buttons Update Cmp Ver and Update IF Ver. Check the software version
of all components (Fig. 30 / p. 47) against the provided the Compatibility List (the
screen shot is just an example!).
• To test the Connectivity, click on the Test button. Since this is a very time-critical action,
it is possible that a connection will not be immediately established. Because of this, sev-
eral attempts should be made.
¹ The server must send back a message, e.g.:
NOTE If outside NAT is configured, the test function will not work
and the message “...not alive” will be displayed.
The new server name and IP address will then have to be reg-
istered with the local administrator and the SRS Help Desk.
Despite this, continue to the next step of the installation.
NOTE If the text under Status is underlined, a link to the error mes-
sage is overlaid. See (1/Fig. 36 / p. 51).
• Activate a test patient in RIS with several studies and check whether the studies match
after opening the patient on the image system.
Ortho (necessary only when installing for the first time) 2.20
Enter the correct distance between the Ortho support and the detector:
• Options>Service>Local Service>Radis Tools>Image Proc. Params.
• Enter the correct value in “Ortho Support Detector Distance” and “Save”.
NOTE A table with FX Ortho OGPs can be found in the last chapter
of these instructions, “Useful Information and Error Correc-
tion”.
• Configure a network printer (Fig. 40 / p. 55) only if you want to print to a print server
(Fig. 59 / p. 60) and (Fig. 60 / p. 60).
• Select Create a new port and Standard TCP/IP Port (Fig. 41 / p. 55).
Fig. 44:
Fig. 43:
• Select the printer manufacturer and type or have the printer’s floppy to load the driver.
Fig. 52:
Fig. 51:
• Click on Finish.
• For printer setup, click on Properties (properties depend on the type of printer; images
and configuration may differ from this description)(Fig. 52 / p. 58).
• Click on Next.
• In the menu that comes up, enter the printer’s name and path (Fig. 60 / p. 60) (the
screen shot is in German, please apply accordingly).
• Continue with (Fig. 41 / p. 55).
• The following work steps describe how to create backups. This is necessary only if
some changes were made since the installation. To save time, save packages only
where changes have been made.
• To do this, go to:
> Options > Service > Local Service > Backup & Restore > Backup
- Insert the backup CD / diskette / USB stick
- Select Drives: [-R-]CD-R; select the following packages one after the other.
- In each case, press GO and wait for the Ready messages.
- SW Settings02
- SW Settings Calibration
- IQAP_Database
- Customer_OGP_DBs
- Clinical Assurance
- Insert a blank diskette, labeled "Customer OGP Diskette (Date)". Go back to the
Local Service main menu. Select Radis Tools > Organ Programs > Backup. Wait
for the "Successfully..." message.
- Insert a floppy (blank or the site structure backup floppy) into the floppy drive. Go
back to Radis Tools. Click on Service Software. Click on OK in the first message.
Observe the second message. Log in to the XCS SSW with User name and OK. Pro-
duce a backup of the site structure with Data > Backup to disk
- Go to the Aristos FX / FX Plus SSW: SSW Aristos FX > Backup/Restore > Backup
and wait for the completion message.
• Generate a recovery CD / DVD for the C and D drives by following the work steps
below. Destroy the clone medium. Do not leave it at the customer site.
l) A new window comes up (Fig. 62 / p. 63) displaying the backup process of the sys-
tem partition. This will take a while ...
m) At the end of the backup routine, the COBR Disk Backup window appears
(Fig. 63 / p. 63) indicating backup success.
n) Remove the CD/DVD from the drive and close the tray. Label it according to the next
paragraph.
o) Confirm the COBR Disk Backup window with <OK>(Fig. 63 / p. 63).
p) Go back to step c, select (D:) Med_Data (3/Fig. 61 / p. 62) and repeat the steps.
Then continue with step q.
q) In the COBR Disk Backup mask, press <Quit> (Fig. 61 / p. 62).
¹ The window will close and a command window is displayed (Fig. 64 / p. 64).
Perform a function check of the system in accordance with the modifications that have been
made.
• The customer or the CSE can easily report problems with the system online.
- Under Help, select Report Error.
-
This procedure describes the steps necessary to reinstall the Radis VB21 SW on the
AXIOM Aristos Image System when a VB21 was installed before, e.g. after a system
crash.
• Here is a guideline of what will be covered (among other things) in this chapter:
Tab. 7 Work Steps
Creating a repair plan is the best way to ensure that all necessary steps are performed in
an efficient way and that multiple visits after setup of the PC are avoided.
• Identify the SW version before the crash (possibly consider open patches or new ver-
sions - > to save time) .
List all performed updates and remote updates available in the update handling tool /
country SAP (notifications)] and the sequence of updates that need to be installed.
• Check whether a virus protection contract exists -> if yes, download the latest released
scan engine + latest virus pattern from the Intranet.
• Options & Licenses:
Check whether there is a distribution list (ROD Remote Options Distribution) [current
on-site license file + ROD] using the remote connection.
• Download all needed remote updates, incl. UI (by CSE (on laptop HD) or by USC (pre-
pare DVD etc. if applicable))
• If a virus is suspected, check the system using the bootable virus scanner.
• Check the external removable devices (CD, DVD, USB stick, USB disk) if a virus is sus-
pected.
• Unplug the network cable (virus protection).
• Install the latest Ghost image.
• Install all required updates and remote updates.
• If a virus protection contract exists, install the latest virus pattern.
• Install the remote option (ROD), if required.
• Plug in the network cable only after all remote updates and virus patterns (VP) have
been installed.
Preparations 3.2
NOTE Please consider the steps created in the repair plan. Prior to
a new installation, all patients must be archived from the
local database, because the database is over-written during
the new installation.
Log files 0
4. Copy the compressed log file (2/Fig. 69 / p. 71) to a USB stick or similar medium, and
keep it ready in case it is requested by RSC/HSC/QIB/DCU.
¹ The log files look like this: “Log<date>_<time>.zip“
5. If a customer complaint has to be filed, send the file along with the complaint to Health-
care AX Complaint.
Perform a virus scan if a virus is suspected (see the repair plan, if available). 0
Backup 0
The following configurations have to be saved manually because these are not part of the
syngo backup.
Service Key
The following work steps describe how to create backups of all syngo settings on CD, “Cus-
tomer OGPs” on floppy (or USB stick).
• To do this, go to:
> Options > Service > Local Service > Backup & Restore > Command > Backup
- Insert the backup CD or USB stick.
- Under Command > Backup > Drives: [-R-]CD-R, select the following packages :
- SW Settings02
- SW Settings Calibration
- IQAP_Database
- Customer_OGP_DBs
- Clinical Assurance
- Press GO for each one and wait for the Ready messages.
- Go back to the Local Service > Configuration menu. Click on Next until you come
to the actual Configuration menu.
• If System Management is configured, manually save the settings:
Go back to the Local Service > Configuration menu.
Make a note of the settings for System Mgmt.:
Server name
TCP/IP address
Local Host NAT
Network adapter
• The following screen displays the restore progress (Fig. 75 / p. 77). This will take a
while ...
- If the medium set consists of more than one CD/DVD, you will be prompted to insert
the next medium.
¹ Follow the instructions on the screen.
- When the restore has ended, a success message is shown in a new screen
(Fig. 76 / p. 77).
- Close the success window with <OK>.
- Remove the FD recovery image from the drive and close the tray.
¹ The window will close and the PC will start to reboot.
Checking the latest VB21 Patch according to the Repair Plan 3.5
To upgrade from the initial VB21A or B version to VB21C or higher, follow the steps below.
• Insert the Aristos RADIS Patch VB21C CD into the CD drive.
• Navigate to the CD/DVD drive with the Windows® Explorer.
• Double-click on ParamUpdate.exe.
• The patch installation will begin; installation may take 15 ...20 sec.
• Click on OK.
• Insert the software CD-Rom according the Compatibility List for the AXIOM Aristos
MX/VX, TX or FX/FX Plus into the CD drive.
• Open “Windows® Explorer”, select the CD-Rom drive and double-click on the
“start.exe” file. The Installer window opens.
• Click on the link under "Service PC (+ Download Software) ...".
• If queried, choose Full Install.
• Choose Continue > OK.
Check for the latest Licenses (according to the Repair Plan) 3.8
• To do this:
- Go to Options > Service > Local Service... > Configuration
- > Next
- > Licensing
- Insert the data medium with the new License.dat (it is possible that the new license
file still needs to be renamed to “license.dat”).
- Click on “Browse..” and go to the new “license.dat”.
- Click on “Open” > “Import License file” and confirm with “OK”.
- Select “Home” and close the window.
- Reboot the Radis PC: Options > End Session... > Restart System... > Yes
• To test the Connectivity, click on the Test button. Since this is a very time-critical action,
it is possible that a connection will not be immediately established. Because of this, sev-
eral attempts should be made.
¹ The server must send back a message, e.g.:
NOTE If outside NAT is configured, the test function will not work
and the message “...not alive” will be displayed.
The new server name and IP address also have to be regis-
tered with the local administrator and SRS Help Desk.
Despite this, continue to the next step of the installation.
NOTE If the text under Status is underlined, a link to the error mes-
sage is overlaid. See (1/Fig. 81 / p. 84).
Restoring the Customer OGPs (only if the OGPs are newer than in
the Recovery Image) 3.10
• Activate a test patient in RIS with several studies and check whether the studies match
after opening the patient on the image system.
l) A new window comes up (Fig. 85 / p. 89) displaying the backup process of the sys-
tem partition. This will take a while ...
m) At the end of the backup routine, the COBR Disk Backup window appears
(Fig. 86 / p. 89) indicating backup success.
n) Remove the CD/DVD from the drive and close the tray. Label it according to the next
paragraph.
o) Confirm the COBR Disk Backup window with <OK>(Fig. 86 / p. 89).
p) Go back to step c, select (D:) Med_Data (3/Fig. 84 / p. 88) and repeat the steps.
Then continue with step q.
q) In the COBR Disk Backup mask, press <Quit> (Fig. 84 / p. 88).
¹ The window will close and a command window is displayed (Fig. 87 / p. 90).
Perform a function check of the system in accordance with the modifications that have been
made.
• The customer or the CSE can easily report online any system problems that may occur.
- Under Help, select Report Error.
-
• The following is a guideline of what will be covered (among other things) in this chapter:
Tab. 8 Work Steps
Creation of a repair plan is the best way to ensure that all necessary steps are performed
efficiently and that multiple visits after setup of the PC are avoided.
• Identify the SW version before the crash (possibly consider open patches or new ver-
sions - > to save time).
If going from VB10 to the new PC (latest version of VB21), additional collimator SW has
to be downloaded from Intranet.
List all performed updates and remote updates available in the update handling tool /
country SAP (notifications). List the sequence of updates needed to be installed.
• Check whether a virus protection contract exists -> if yes, download the latest released
scan engine + latest virus pattern from the Intranet.
• Options & Licenses:
Check whether there is a distribution list via remote (current license file on-site + ROD
Remote Options Distribution)
• Download all required remote updates incl. UIs (by CSE (on laptop HD) or by USC (pre-
pare DVD etc., if applicable))
To avoid extra work, check whether the collimator SW is lower than VC00N using the XCS
SSW. If so, follow the work steps below. If not, skip this paragraph.
• Go to the download location on the CS Home page: For Service > Product Informa-
tion > AX Systems > Radiography > Direct Technique > Software.
• The required download file is the 10092710_OT_II_VC00N.exe and is located under
the description Collimator G052G.
• Double-click on it and unzip it to a folder of your choice on the C or D drive of your ser-
vice PC.
• You will find two files in the folder: CollimatorSWLoader.exe and
SW_D21_TBN_OT_II_VC00N.H86. Keep them there for later use.
• If a virus is suspected, check the system using the bootable virus scanner.
• Check the external removable devices (CD, DVD, USB Stick, USB disk) if a virus is sus-
pected.
• Consider using the escalation routine.
What to do in case of a Virus Infection This should be implemented in CB Doc with a link
to the Knowledge Base.
• Unplug the Net cable (virus protection).
Preparations 4.2
1. Unlock the access to Windows® with Tab - Del - Transfer image (“+”).
2. Press the Windows® key, and then Run; enter radissysinfo > OK.
3. Press Start.
5. The log file will be created and saved to the selected directory.
8. Copy the compressed log file (2/Fig. 94 / p. 100) to a USB stick or similar medium, and
keep it ready in case it is requested by RSC/HSC/QIB/DCU.
¹ The log files look like this: “Log<date>_<time>.zip“
9. If a customer complaint has to be filed, send the file along with the complaint to Health-
care AX Complaint.
AXIOM Aristos Image System AXB4-320.816.08.03.02 Page 100 of 134 © Siemens, 2011
05.11 CS PS XP RFU
PC Replacement: SW Installation and Settings 101
© Siemens, 2011 AXB4-320.816.08.03.02 Page 101 of 134 AXIOM Aristos Image System
05.11 CS PS XP RFU
102 PC Replacement: SW Installation and Settings
4. Copy the compressed log file (2/Fig. 97 / p. 102) to a USB stick or similar medium, and
keep it ready in case it is requested by RSC/HSC/QIB/DCU.
¹ The log files look like this: “Log<date>_<time>.zip“
5. If a customer complaint has to be filed, send the file along with the complaint to Health-
care AX Complaint.
Backup 0
The following configurations have to be saved manually because these are Win-
dows®-related configurations and are not part of the syngo backup.
• Make a note of the service key:
- Options > Service > Local Service...
Service Key
AXIOM Aristos Image System AXB4-320.816.08.03.02 Page 102 of 134 © Siemens, 2011
05.11 CS PS XP RFU
PC Replacement: SW Installation and Settings 103
• Make a note of the host name (= computer name) and the IP addresses:
- Options > Service > Local Service... > Configuration > TCP/IP LAN
- Home > close the window.
IP address
Subnet mask
Default gateway
Customer Configuration:
- > Options > Configuration.. > Regional and Language...
- “Regional Options”: Make a note of the date and time format.
- “ Languages > Details...”: Make a note of the language and keyboard.
Date Format
Time Format
Language
Keyboard (input locales)
Network Speed
© Siemens, 2011 AXB4-320.816.08.03.02 Page 103 of 134 AXIOM Aristos Image System
05.11 CS PS XP RFU
104 PC Replacement: SW Installation and Settings
AXIOM Aristos Image System AXB4-320.816.08.03.02 Page 104 of 134 © Siemens, 2011
05.11 CS PS XP RFU
PC Replacement: SW Installation and Settings 105
Select the Advanced tab card, read the value for Speed & Duplex or Link
Speed & Duplex and make a note of it.
The following work steps describe how to create backups of “SW Settings02”,“ SW Settings
Calibration” and “Customer OGPs” on floppy (or USB stick).
The other packages do not have to be saved because there are newer versions with the
new VB21 SW.
• To do this, go to:
> Options > Service > Local Service > Backup & Restore
© Siemens, 2011 AXB4-320.816.08.03.02 Page 105 of 134 AXIOM Aristos Image System
05.11 CS PS XP RFU
106 PC Replacement: SW Installation and Settings
Server name
TCP/IP address
Local Host NAT
Network adapter
AXIOM Aristos Image System AXB4-320.816.08.03.02 Page 106 of 134 © Siemens, 2011
05.11 CS PS XP RFU
PC Replacement: SW Installation and Settings 107
© Siemens, 2011 AXB4-320.816.08.03.02 Page 107 of 134 AXIOM Aristos Image System
05.11 CS PS XP RFU
108 PC Replacement: SW Installation and Settings
AXIOM Aristos Image System AXB4-320.816.08.03.02 Page 108 of 134 © Siemens, 2011
05.11 CS PS XP RFU
PC Replacement: SW Installation and Settings 109
© Siemens, 2011 AXB4-320.816.08.03.02 Page 109 of 134 AXIOM Aristos Image System
05.11 CS PS XP RFU
110 PC Replacement: SW Installation and Settings
NOTE Before switching on the PC, make sure that only power, the
mouse keyboard and the monitor are connected. Do not yet
connect the network cable, fiber optic cables or the XCS ser-
vice cable.
AXIOM Aristos Image System AXB4-320.816.08.03.02 Page 110 of 134 © Siemens, 2011
05.11 CS PS XP RFU
PC Replacement: SW Installation and Settings 111
© Siemens, 2011 AXB4-320.816.08.03.02 Page 111 of 134 AXIOM Aristos Image System
05.11 CS PS XP RFU
112 PC Replacement: SW Installation and Settings
• Insert the software CD-Rom according the Compatibility List for the AXIOM Aristos
MX/VX, TX or FX/FX Plus into the CD drive.
• Open “Windows® Explorer”, select the CD-Rom drive and double-click on the
“start.exe” file. The Installer window opens.
• Click on the link under "Service PC (+ Download Software) ...".
• If queried, choose Full Install.
• Choose Continue > OK.
AXIOM Aristos Image System AXB4-320.816.08.03.02 Page 112 of 134 © Siemens, 2011
05.11 CS PS XP RFU
PC Replacement: SW Installation and Settings 113
Checking the latest VB21 Patch according to the Repair Plan 4.7
To upgrade from the initial VB21A or B version to VB21C or higher, follow the steps below.
• Insert the Aristos RADIS Patch VB21C CD into the CD drive.
• Navigate to the CD/DVD drive with the Windows® Explorer.
• Double-click on ParamUpdate.exe.
• The patch installation will begin; installation may take 15 ...20 sec.
• Click on OK.
© Siemens, 2011 AXB4-320.816.08.03.02 Page 113 of 134 AXIOM Aristos Image System
05.11 CS PS XP RFU
114 PC Replacement: SW Installation and Settings
• If the system is using a DAC, install the DAC program from the "DICOM Attribute Con-
fig. VA20A", P/N 7264435 or higher.
• Follow the "DICOM Attrib. Configurator" Instructions, AX00-000.816.01.02… or higher.
AXIOM Aristos Image System AXB4-320.816.08.03.02 Page 114 of 134 © Siemens, 2011
05.11 CS PS XP RFU
PC Replacement: SW Installation and Settings 115
• Go to “Options > Service > Local Service” and log in with the current service key.
• Insert the backup CD / diskette / USB stick
• Select Backup & Restore > Command > Backup > Drives: [-R-]CD-R; select the fol-
lowing packages :
• SW Settings02
• SW Settings Calibration
• IQAP_Database
• Only if coming from VB21: Customer_OGP_DBs
• Only if coming from VB21: Clinical Assurance
• Press GO for each one and wait for the Ready messages.
• Go back to the Local Service > Configuration menu. Click on Next until you come to
the actual Configuration menu.
• If there is an newer “defect map” on site, it must be restored.
For this, go to RADIS Tools > Import FD Calib. Data.
Follow the description on the screen.
• Select Home.
• Close the window.
© Siemens, 2011 AXB4-320.816.08.03.02 Page 115 of 134 AXIOM Aristos Image System
05.11 CS PS XP RFU
116 PC Replacement: SW Installation and Settings
Check for the latest Licenses (according to the Repair Plan) 4.10
• To do this:
- Go to Options > Service > Local Service... > Configuration
- > Next
- > Licensing
- Insert the data medium with the new License.dat (it is possible that the new license
file still needs to be renamed to “license.dat”).
- Click on “Browse..” and go to the new “license.dat”.
- Click on “Open” > “Import License file” and confirm with “OK”.
- Select “Home” and close the window.
- Reboot the Radis PC: Options > End Session... > Restart System... > Yes
AXIOM Aristos Image System AXB4-320.816.08.03.02 Page 116 of 134 © Siemens, 2011
05.11 CS PS XP RFU
PC Replacement: SW Installation and Settings 117
© Siemens, 2011 AXB4-320.816.08.03.02 Page 117 of 134 AXIOM Aristos Image System
05.11 CS PS XP RFU
118 PC Replacement: SW Installation and Settings
• To test the Connectivity, click on the Test button. Since this is a very time-critical action,
it is possible that a connection will not be immediately established. Because of this, sev-
eral attempts should be made.
¹ The server must send back a message, e.g.:
NOTE If outside NAT is configured, the test function will not work
and the message “...not alive” will be displayed.
The new server name and IP address also have to be regis-
tered with the local administrator and SRS Help Desk.
Despite this, continue to the next step of the installation.
AXIOM Aristos Image System AXB4-320.816.08.03.02 Page 118 of 134 © Siemens, 2011
05.11 CS PS XP RFU
PC Replacement: SW Installation and Settings 119
© Siemens, 2011 AXB4-320.816.08.03.02 Page 119 of 134 AXIOM Aristos Image System
05.11 CS PS XP RFU
120 PC Replacement: SW Installation and Settings
NOTE If the text under Status is underlined, a link to the error mes-
sage is overlaid. See (1/Fig. 108 / p. 119).
AXIOM Aristos Image System AXB4-320.816.08.03.02 Page 120 of 134 © Siemens, 2011
05.11 CS PS XP RFU
PC Replacement: SW Installation and Settings 121
Restoring the Customer OGPs (only if the OGPs are newer than in
the Recovery Image) 4.12
• Activate a test patient in RIS with several studies and check whether the studies match
after opening the patient on the image system.
© Siemens, 2011 AXB4-320.816.08.03.02 Page 121 of 134 AXIOM Aristos Image System
05.11 CS PS XP RFU
122 PC Replacement: SW Installation and Settings
• The following work steps describe how to create backups for all “Customer OGPs” set-
tings on floppy (or USB stick) and Configuration Data on CD (or USB stick)
• To do this, go to:
> Options > Service > Local Service > Backup & Restore
- Insert the backup CD / USB stick
- Under Command > Backup > Drives: [-R-]CD-R, select the following packages:
- SW Settings02
- Customer_OGP_DBs
- IQAP_Database
- Clinical Assurance
- SW Settings Calibration
- Press GO for each one and wait for the Ready messages.
- Press Home > Radis Tools > Organ Programs.
- Select A:\ for diskette (recommended) or the drive for USB stick under “Drive
Select”.
- Press “Backup” to save the OGPs.
- Close the window and press “Home”.
AXIOM Aristos Image System AXB4-320.816.08.03.02 Page 122 of 134 © Siemens, 2011
05.11 CS PS XP RFU
PC Replacement: SW Installation and Settings 123
© Siemens, 2011 AXB4-320.816.08.03.02 Page 123 of 134 AXIOM Aristos Image System
05.11 CS PS XP RFU
124 PC Replacement: SW Installation and Settings
l) A new window comes up (Fig. 112 / p. 124) displaying the backup process of the
system partition. This will take a while ...
m) At the end of the backup routine, the COBR Disk Backup window appears
(Fig. 113 / p. 124) indicating backup success.
n) Remove the CD/DVD from the drive and close the tray. Label it according to the next
paragraph.
o) Confirm the COBR Disk Backup window with <OK>(Fig. 113 / p. 124).
p) Go back to step c, select (D:) Med_Data (3/Fig. 111 / p. 123) and repeat the steps.
Then continue with step q.
q) In the COBR Disk Backup mask, press <Quit> (Fig. 111 / p. 123).
¹ The window will close and a command window is displayed (Fig. 114 / p. 125).
AXIOM Aristos Image System AXB4-320.816.08.03.02 Page 124 of 134 © Siemens, 2011
05.11 CS PS XP RFU
PC Replacement: SW Installation and Settings 125
© Siemens, 2011 AXB4-320.816.08.03.02 Page 125 of 134 AXIOM Aristos Image System
05.11 CS PS XP RFU
126 PC Replacement: SW Installation and Settings
Perform a function check of the system in accordance with the modifications that have been
made.
AXIOM Aristos Image System AXB4-320.816.08.03.02 Page 126 of 134 © Siemens, 2011
05.11 CS PS XP RFU
PC Replacement: SW Installation and Settings 127
• The customer or the CSE can easily report problems with the system online.
- Under Help, select Report Error.
-
© Siemens, 2011 AXB4-320.816.08.03.02 Page 127 of 134 AXIOM Aristos Image System
05.11 CS PS XP RFU
128 Useful Information and Correction of Errors
5-
For Information Only
5Useful Information and Correction of Errors
• The customer or the CSE can easily report problems with the system online.
- Under Help, select Report Error.
-
If there are display problems with an EIZO monitor connected to Radis via the Digital Video
Interface (DVI) in the Celsius M410 (gray housing), carry out these settings:
AXIOM Aristos Image System AXB4-320.816.08.03.02 Page 128 of 134 © Siemens, 2011
05.11 CS PS XP RFU
Useful Information and Correction of Errors 129
• Unlock Windows® by pressing “Tab + Del. + Send (Num field)”, open the Explorer and
go to Control Panel; press GO; if necessary, select "classical view" in the window
that comes up; select: the Display > Settings tab card.
- Color quality: Highest (32-bit)
- Screen resolution: 1280 x 1024 pixels
- Click on "Advanced" > Monitor > Set Screen refresh rate: 60 Hz
- Close all windows
Using XP Explorer 0
Unlock Windows® by pressing “Tab + Del. + Send (Num field)” and open the Explorer.
Configuring the CDs/DVDs as Offline Devices for Export of 12-bit Images (only if
needed) 0
This feature can be used only if the CD/DVD burner is configured in the Local Service con-
figuration.
• Go to Options > Local Service... > Configuration
• Make sure that "DICOM Offline Devices" is enabled.
© Siemens, 2011 AXB4-320.816.08.03.02 Page 129 of 134 AXIOM Aristos Image System
05.11 CS PS XP RFU
130 Useful Information and Correction of Errors
AXIOM Aristos Image System AXB4-320.816.08.03.02 Page 130 of 134 © Siemens, 2011
05.11 CS PS XP RFU
Useful Information and Correction of Errors 131
• After reboot, go to Options > Local Service... "RADIS Tools" > "CD Recording".
© Siemens, 2011 AXB4-320.816.08.03.02 Page 131 of 134 AXIOM Aristos Image System
05.11 CS PS XP RFU
132 Useful Information and Correction of Errors
• Unlock Windows® by pressing “Tab + Del. + Send (Num field)” and open the Explorer.
• Go to "D:\Images\Exported14bitImages”
• Delete all files in this directory.
• Close all windows.
The reason for a failed installation could be that fragments from the old installation were not
completely deleted. Following a failed installation, a low level formatting of both hard drives
should be performed prior to making a second attempt. To do this:
• During the reboot, press the "CTRL + A" keys.
• Confirm the selection.
• Under "Options", select "SCSI Disk Utilities" with the arrow keys.
• Confirm the selection with Enter.
• Select the 1st hard drive (SCSI ID #0) in the table.
• Confirm the "Format Disk" selection with Enter.
• Confirm the safety inquiry with "Yes".
• Repeat the procedure for the 2nd hard drive (SCSI ID #1) and then exit the Adaptec
tool.
AXIOM Aristos Image System AXB4-320.816.08.03.02 Page 132 of 134 © Siemens, 2011
05.11 CS PS XP RFU
Changes to Previous Version 133
6Changes to Previous Version 6-
Chapters 2, 3 and 4: included backup and restore of the “IQAP Database”.
© Siemens, 2011 AXB4-320.816.08.03.02 Page 133 of 134 AXIOM Aristos Image System
05.11 CS PS XP RFU
134 Changes to Previous Version
AXIOM Aristos Image System AXB4-320.816.08.03.02 Page 134 of 134 © Siemens, 2011