UT Level 2
UT Level 2
3. In an homogenous and isotropic elastic medium such as low carbon steel, sound velocity
A) decreases with distance from source
B) varies with direction
C) is constant in all directions
D) depends on frequency
4. The amount of time between two compressions, or two rarefactions of an elastic wave is called
A) wavelength
B) period
C) frequency
D) velocity
7. The elastic wave that has particle motion parallel to the direction of wave propagation is called
A) longitudinal wave
B) compression wave
C) density wave
D) all of the above
1
ESTestMaker Questions
10. In bending waves (plate wave mode) particles in the middle zone of the plate vibrate
A) in longitudinal mode
B) in shear mode
C) in Rayeigh mode
D) not at all
11. In the Lamb wave called a dilational wave, particles in the middle zone of the plate vibrate
A) in longitudinal mode
B) in shear mode
C) in Rayleigh mode
D) not at all
12. If one sound beam passes through another with the same wavelength but moving in the opposite
direction, the result on the local particle displacement will be,
A) a change in amplitude
B) a change in mode
C) a change in frequency
D) no change
2
ESTestMaker Questions
18. The ratio of sound velocity in water to the longitudinal velocity of sound in steel is very nearly
A) 1:1
B) 1:2
C) 1:4
D) 1:5
20. The ratio of the incident sound pressure to the reflected sound pressure is called the
A) acoustic impedance
B) acoustic intensity
C) coefficient of reflection
D) coefficient of transmission
21. The ratio of the incident sound pressure to the transmitted sound pressure is called the
A) acoustic impedance
B) acoustic intensity
C) coefficient of reflection
D) coefficient of transmission
3
ESTestMaker Questions
26. If the Sine of a refracted angle is calculated to be 0.707, the refracted angle will be
A) 36 degrees
B) 45 degrees
C) 60 degrees
D) undetermined from the given information
27. A shear wave travelling from steel to water incident on the boundary at 10 degrees will give a refracted
shear wave in water with an angle of
A) 0 degrees
B) 5 degrees
C) 20 degrees
D) none of the above
29. The critical angle occurs when the sine of the refracted angle equals
A) 0.707
B) 0.846
C) 1.000
D) 1.414
4
ESTestMaker Questions
30. At a solid to free boundary, an obliquely incident longitudinal wave from the solid can result in, at most,
A) a reflected longitudinal wave only
B) a reflected longitudinal and reflected shear wave
C) a refracted longitudinal long wave
D) a reflected longitudinal and reflected shear and refracted longitudinal wave
31. At a liquid/solid boundary with an obliquely incident longitudinal wave from the liquid the result could at
most be
A) a reflected longitudinal wave only
B) a refracted longitudinal wave only
C) a reflected longitudinal, and a refracted long wave
D) a reflected long and refracted long and shear wave
32. Given V (water) = 1.5 mm/us and V (steel) longitudinal velocity = 5.0 mm/us and shear velocity
3.2mm/us, what is the second critical angle of an incident longitudinal wave from steel.
A) 19 degrees
B) 27 degrees
C) 36 degrees
D) none of the above
33. A shear wave polarized in the plane of incidence impinging on a free boundary at right angles to the
boundary will result in
A) a reflected shear wave only
B) a reflected shear wave and reflected longitudinal wave
C) reflected longitudinal wave only
D) a surface wave
34. Maximum surface wave energy is obtained when the angle of incidence is
A) 0 degrees
B) 60 degrees
C) 90 degrees
D) just over the critical angle
36. The relationship for acoustic pressure of spherical waves P=P1/d (where P1 is the pressure at a known
distance and d is distance), implies
A) an inverse proportionality to distance
B) infinite sound pressure at the source
C) both a and b
D) none of the above
5
ESTestMaker Questions
37. Spherical and cylindrical waves reflecting off a plane surface preserve shape but not
A) velocity
B) wavelength
C) frequency
D) acoustic pressure distribution
38. The advantage of inspecting a thick solid cylinder with a waterpath two times the cylinder radius instead
of just the cylinder radius is
A) increased sensitivity
B) reduced entry noise
C) smaller near zone
D) more uniform pressure through the cylinder
39. The ratio of the diameter of an oscillator to the wavelength it generates gives the
A) acoustic velocity
B) near zone length
C) acoustic impedance of the oscillator
D) number of interference maxima and minima
40. For practical purposes, the equation for the near zone distance can be approximated by (where λ =
wavelength and D = probe diameter)
A) D /λ
B) D squared / 4λ
C) D squared minus l squared/4λ
D) (D-4l)/λ^2)
41. Given an "X-cut" normal beam contact probe on steel, the probe is 12 mm diameter and has a
frequency of 2 MHz. The approximate near zone length is (V steel = 6.0mm/us)
A) 4mm
B) 12mm
C) 24mm
D) 48mm
42. A transducer has a near field in water of 35 mm. When used in contact on steel the near zone will be
about
A) 47 mm
B) 35 mm
C) 18 mm
D) 9 mm
6
ESTestMaker Questions
44. Sin(gamma) = 1.2 lambda / D (where lambda is wavelength and D is the crystal diameter) is the
equation for
A) the first critical angle
B) interference minima in the near zone
C) half angle of divergence in the Fraunhoffer zone
D) optimum energy angle for shear waves
46. A rectangular probe, 4mm X 8mm, will have its maximum half angle of divergence
A) in the 4mm direction
B) in the 8mm direction
C) in no particular orientation
D) constant in all directions
47. Given a 10 X 10mm square probe of 5 MHz fundamental frequency, compared to a 10mm diameter 5
MHz probe, the near zone of the square probe would be
A) longer
B) shorter
C) the same
D) independent of wavelength
48. In ultrasonic testing, actual point focusing of sound is not possible due to
A) damping effects
B) divergence in the far zone
C) assymetric vibrations
D) diffraction effects
49. Huygen's wavelet theory and Fresnel zones can be used to predict
A) focal length
B) maxima and minima points in a sound beam
C) radius of curvature in a focused probe
D) fundamental frequency
50. AVG (or DGS in English) diagrams compare flaw signal amplitudes to
A) side drilled holes
B) flat bottomed holes
C) a theoretical maximum
D) DAC's
7
ESTestMaker Questions
51. Flaws oriented obliquely to the incident sound beam may be better detected using
A) high gain
B) higher frequency
C) focused probes
D) separate transmit and receive probes
53. A surface can be considered smooth if its irregularities are not more than ________ wavelength.
A) 1
B) 1/3
C) 1/10
D) 1/100
55. The effect of oxides or other non metallic inclusions within a defect is to
A) amplify the signal
B) split the signal by mode conversion
C) reduce the signal amplitude
D) none of the above
56. As the pulse length of the excitation voltage is shortened the transmitted pulse
A) frequency spectrum broadens
B) frequency spectrum shortens
C) increases energy output
D) increases penetration ability
57. In general, the frequency content of an ultrasound beam has a larger proportion of high frequencies in
its spectrum
A) on axis
B) off axis
C) in the far zone
D) in the free zone
8
ESTestMaker Questions
61. If 3 dB gain was removed from a 100% FSH signal the resulting signal would be __________ FSH.
A) 97
B) 94
C) 85
D) 71
62. If a signal is dropped from 100% FSH to 32% FSH, the number of dB gain removed from the receiver is
A) 14
B) 10
C) 6
D) 4
9
ESTestMaker Questions
67. For a given metal of a given grain size, attenuation in the cast form of the metal relative to the worked
form is usually
A) greater
B) less
C) identical
D) unpredictable
68. The effect utilized by ultrasonic transducer crystals is the _________ effect.
A) piezoelectric
B) pyroelectric
C) ferroelectric
D) hall
70. An X-cut quartz crystal is primarily used to generate the ________ wave mode.
A) longitudinal
B) transverse
C) Rayleigh
D) Lamb
71. A typical voltage range for driving (exciting) piezoelectric crystals would be
A) 50 to 100 mV
B) 50 to 100 V
C) 50 to 100 kV
D) 100 to 1000 V
10
ESTestMaker Questions
73. The efficiency of the piezoelectric effect is rated by K33 which is the
A) piezoelectric modulus
B) electromechanical coupling factor
C) deformation constant
D) Young's modulus
74. After a voltage excitation of the piezoelectric crystal, the amplitude of its oscillations decrease by the
quantity determined by the
A) Q factor
B) damping coefficient
C) characteristic frequency
D) resonance frequency
78. Although rarely used in modern day NDT transducers, quartz has the advantage of
A) chemical inertness
B) high coupling coefficient
C) high dielectric constant
D) all of the above
11
ESTestMaker Questions
84. Optical methods used in ultrasonic testing such as Schlieren diffraction and interferometer methods are
used in
A) transmission techniques only
B) reception techniques only
C) both transmission and reception
D) optical methods are not used in ultrasonics
12
ESTestMaker Questions
87. Using the pulse echo method with a 0 degree probe having separate transmit and receive crystals, the
pattern on a CRT of a flat steel plate would appear as
A) evenly spaced multiples
B) randomly spaced multiples
C) a single spike from the opposite wall
D) an uninterrupted base line
88. Phantom echoes of multiples when testing thick specimens are a result of
A) insufficient suppression
B) modulation from the power supply
C) PRF set too high
D) too much gain
90. When an ultrasonic machine is equipped with this option, the pulse energy and pulse length can be
adjusted
A) receiver fine grain control
B) swept gain
C) time corrected gain
D) damping
91. The voltage of a received ultrasonic signal at the machine's receiver is typically
A) 1/1000 to 1 volt
B) 10 to 100 volts
C) 100 to 1000 volts
D) not measurable
92. In order to adequately amplify received signals an ultrasonic machine's receiver amplifier must have a
gain of about
A) 6 to 12 dB
B) 20 to 40 dB
C) 80 to 100 dB
D) 100 to 200 dB
13
ESTestMaker Questions
94. The zero time indicating zero depth in the tested sample when using a contact normal beam probe
A) corresponds to the rising edge of the main bang
B) corresponds to the falling edge of the main bang
C) is determined indirectly by calibration blocks
D) none of the above
96. A great disadvantage of the old quartz crystal transducers that used the work piece as one of the
electrodes was
A) a fluctuating dead zone
B) insufficient damping
C) decreasing sensitivity with use
D) an increase in frequency with use
97. A contact angle beam probe used in fixed housings occasionally requires
A) transformer tuning
B) machining or replacing of the wear face
C) new couplant between crystal and plastic wedge
D) all of the above
98. Wedges used to introduce refracted waves into a test specimen are made of
A) copper
B) brass
C) plastics
D) all of the above can be used
99. In order that one can operate above the critical angle of reflection of longitudinal waves, wedge material
for angle beams are
A) grooved on the front face
B) always made of polystyrene
C) chosen with a longitudinal velocity greater than the shear velocity of the test piece
D) chosen with a longitudinal velocity less than the shear velocity of the test piece
100. The layer of highly absorbing material sometimes bonded to the top and front of an angle probe's
wedge is called a (an)
A) cap
B) backing
C) anechoic trap
D) none of the above
14
ESTestMaker Questions
102. The purpose of the 50mm diameter perspex insert in the IIW block is to
A) check constancy of probe sensitivity
B) establish probe exit point
C) calibrate for range
D) both a and b
104. Pulse echo instruments without an image tube or A-scan display are
A) not useful in NDT
B) not yet available
C) used for wall thickness measurements
D) not portable
106. When flaw echo signals are recorded so as to display a plan view of the test piece the presentation is
called
A) A-scan
B) B-scan
C) C-scan
D) D-scan
15
ESTestMaker Questions
109. Uniform and strongly adhering films of paint or oxides are not removed for ultrasonic testing if
A) sand blasting cannot remove them
B) formed on curves surfaces
C) their presence does not interfere with the test
D) sufficiently smooth
110. Due to its cost and toxicity _________ is not used as a couplant in contact testing,
A) SAE oil
B) heavy diesel oil
C) mercury
D) methyl-cellulose solution
111. Water jet techniques with water paths of 100mm are restricted to the through transmission methods
because
A) sound cannot travel upstream for pulse-echo testing
B) too much water is required
C) of disturbing echoes behind the surface echo
D) none of the above
112. A narrow high intensity ultrasound beam capable of finding very small flaws has its disadvantage in
A) sizing the flaw
B) characterizing the flaw
C) requiring a small scanning grid
D) none of the above
113. An effective maximum range for a 5MHz 10mm diameter probe is around ________ in steel.
A) 50mm
B) 100mm
C) 200mm
D) 300mm
114. A minimum distance of a probe from an edge of the test piece is maintained to eliminate side wall
interference. This distance is inversely proportional to the
A) soundpath
B) wavelength
C) frequency
D) beam spread
16
ESTestMaker Questions
115. When testing a 30mm diameter, 500 mm long shaft from the flat end of the shaft using longitudinal
waves from a 20 mm diameter 2 MHz probe, numerous signals are seen on the screen after 500mm.
These are
A) ghost images
B) secondary echoes
C) internal thread indications
D) none of the above
116. Touching the reflection point of a normally incident transverse wave with an oily finger will
A) reduce the signal by about 2 dB
B) reduce the signal by about 6 dB
C) increase the signal by about 2 dB
D) have no effect on signal amplitude
117. When using a refracted angle of 60° on a 20 mm thick plate the soundpath at the half skip distance is
A) 28mm
B) 34mm
C) 40mm
D) 58mm
118. To calculate skip distance in a flat plate of thickness 'd' for a refracted angle 'X' we use the formula
A) (cos X)/d
B) (sin 2X)/d
C) 2d/cos X
D) 2d (tan x)
119. To inspect cylindrical pieces with transverse waves in the circumferential direction, maximum depth is
obtained using
A) 35° refracted angle
B) 45° refracted angle
C) 70° refracted angle
D) any of the above is adequate if the diameter ratio is sufficient
17
ESTestMaker Questions
123. Changes in flaw echo shape on the video display are used to determine
A) flaw depth
B) flaw type
C) stand-off
D) all of the above
124. A defect that is not likely to be located by ultrasonic testing of a forging blank is
A) segregation
B) flake cracks
C) non-metallic inclusions
D) cold cracks
125. In order to introduce a shear mode into cylindrical shaped specimens when immersion scanning in the
circumferential direction you need only
A) use Y-cut crystals in the transducer
B) switch from pulse-echo to through transmission
C) off-set the probe from normal incidence
D) none of the above
126. When immersion inspecting small diameter tubing (<15mm diameter) from the inside of the tube with a
normal beam probe radiating axially towards a mirror mounted ahead of it, the wave mode used for
inspection in the tube is
A) longitudinal
B) transverse
C) either a or b depending on mirror orientation
D) none of the above, mirrors are not used
128. Relatively low ultrasonic frequencies are used in long range sonar units because
A) whales and submarines are big targets
B) higher frequencies cause whales to attack
C) sea water attenuates sound at 60 dB/km
D) all of the above
18
ESTestMaker Questions
130. The time required for a wave packet of ultrasound to go from 10% to 90% of its maximum amplitude is
called
A) pulse length
B) PRF
C) rise time
D) fall time
131. The time duration for an ultrasonic wave packet to have its trailing edge drop from 90% to 10% of its
maximum amplitude is called
A) pulse length
B) PRF
C) rise time
D) fall time
133. In ferroelectric materials, the units which contain a net electric polarization are called
A) cells
B) bi-poles
C) mono-poles
D) domains
19
ESTestMaker Questions
136. A net dipole moment is found in ferroelectric ceramics such as PZT when
A) heated above the curie temperature
B) the domains are frozen and aligned
C) used under water
D) use on highly retentive steel
138. In which of the following properties, considered when dealing with transducer materials, does quartz
excel in?
A) receiving constant
B) transmitting constant
C) dielectric constant
D) coupling coefficient
141. Internally reflected waves within a lightly loaded resonant transducer (i.e. acoustic impedance on either
side is lower than the transducer's)
A) nullify resonance
B) change the resonant frequency
C) increase the maximum oscillation amplitude
D) are totally internally refracted
142. For continuous wave testing, rise time of the transducer oscillation relative to the applied driving voltage
is
A) longer than fall time
B) shorter
C) about the same
D) not important
20
ESTestMaker Questions
143. The first oscillation of a piezoelectric element hit by an alternating voltage pulse is
A) higher than all other oscillations in that pulse
B) lower than all other oscillations in that pulse
C) never higher than the second oscillation in that pulse
D) higher or lower than the second oscillation depending on the front loading
146. Nickel and its compounds as well as cobalt and ferrites differ from other magnetostrictive materials in
that they
A) expand in applied magnetic fields
B) shrink in applied magnetic fields
C) are anti-magnetostrictive
D) are also piezoelectric
148. In a standing wave, the points of zero pressure change are called
A) null points
B) nodes
C) anti-nodes
D) the still zone
149. A lens shape that would result in a divergent beam in optics gives a convergent beam in ultrasonics
because
A) electromagnetic wavelengths are shorter
B) electromagnetic waves are faster in solids than liquids
C) mechanical waves are usually faster in solids than liquids
D) none of the above
21
ESTestMaker Questions
150. Ultrasonic transducers used in ultrasonic cleaning apparatus are operating at a frequency of
approximately
A) 25 kHz
B) 200 kHz
C) 0.5 MHz
D) 50 MHz
151. The purpose of ultrasonic vibrations added to the tip of a soldering gun when soldering aluminum or
magnesium is to
A) eliminate the need for penetrant inspection
B) make a fine grain structure with high shear strength
C) break through the oxide coating
D) all of the above
153. In the old traditional oscilloscope the timebase or baseline is displayed as a horizontal line by means of
A) line focus electronics
B) line focus optical lenses
C) a light spot moving across phosphor
D) none of the above
154. In the old traditional oscilloscope increasing the pulse repetition frequency will result in
A) decreasing sensitivity
B) increased resolution
C) altering the probes' frequency output
D) brightening the baseline
155. If it is difficult to resolve two defects separated by 0.5 cm, resolution may be improved by
A) increasing gain
B) increasing soundpath
C) decreasing pulse length
D) decreasing probe frequency
22
ESTestMaker Questions
157. Elastic constants such as Young's modulus and the bulk modulus can be calculated from ultrasonic
measurement of
A) attenuation
B) reflection coefficients
C) acoustic velocities
D) frequency domains
160. Linearity of an ultrasonic machine's timebase should be within _____ of the total screen range.
A) 1mm
B) 2mm
C) 1%
D) 10%
161. Which item is not always marked on all standard transducers sold for NDT purposes
A) frequency
B) exit point
C) crystal dimensions
D) serial number
163. When evaluating the performance of a probe-instrument combination, far field resolution is limited by
A) the probes' functional frequency
B) instrument gain
C) available suppression
D) the test surface of the calibration block
23
ESTestMaker Questions
164. The arbitrary amplitude response from flat bottomed or side drilled holes is used to establish sensitivity
level because
A) sensitivity is optimised
B) resolution is optimised
C) results can be reproducible
D) both a and b
165. Viscosity of couplant chosen for a contact ultrasonic inspection will usually depend on
A) probe size
B) nominal frequency
C) surface condition
D) sensitivity required
166. In longitudinal wave inspection of the parent metal adjacent to a weld, minimum sensitivity is usually set
such that
A) it is 6 dB over the 80% FSH level for the appropriate side drilled hole
B) it is 20 dB over the 80% FSH level for the appropriate side drilled hole
C) backwall echo is 100% FSH in the absence of defects
D) none of the above
167. When doing an ultrasonic weld inspection, the material thickness, weld preparation configuration, defect
type and orientation are the main considerations for
A) probe dimensions
B) frequency
C) probe angle
D) all of the above
170. When calibrating an ultrasonic instrument for range, the maximum distance of interest should not be
less than
A) 100 mm
B) 200 mm
C) one half the horizontal scale
D) two thirds the horizontal scale
24
ESTestMaker Questions
172. In the DGS (AVG German) system of defect sizing, the diagram relates to soundpath distance to the
_________ to obtain the relative distance.
A) probe size
B) near-field length
C) stand-off distance
D) depth of defect
173. Where geometry permits in weld inspection, the distance you need to move the probe back from the
weld to ensure 100% volume inspection is
A) 1/2 skip from the weld centre-line
B) 1 full skip from the weld centre-line
C) 1/2 skip from the edge of the heat affected zone
D) 1 full skip from the HAZ edge
174. During manual scanning, detection of obliquely oriented defects may be improved by
A) using a lower frequency
B) longitudinal wave inspections
C) adding swivel motion to the angle beam probe
D) all of the above
175. Shear-wave probe angle for contact testing of plate between 10mm and 100 mm thick is
A) 45 and 60 degrees
B) determined by probe frequency chosen
C) dependant on plate thickness, position and nature of defect
D) none of the above
176. Given a flat plate 25mm thick, the full skip range distance with a 60 degree probe is
A) 35mm
B) 50mm
C) 100mm
D) not possible to determine
177. The half skip range on a 16mm thick plate with a 70% probe is
A) 24mm
B) 33mm
C) 47mm
D) 64mm
25
ESTestMaker Questions
178. When using the 6dB or 20 dB drop technique, a defects' vertical extent is deduced by __________
motion of the probe
A) lateral
B) traversing
C) rotational
D) orbital
179. A sharp signal with large amplitude response seen from one side of a weld but not the other, having a
sharp drop off with both rotational and orbital motion would indicate the presence of
A) porosity
B) lack of fusion
C) slag
D) cracking
180. The use of ultrasonic methods to check fillet welds is usually restricted to determining the extent of
penetration. This is best facilitated by
A) a guidance system to hold the probe at a fixed distance
B) c-scan presentations
C) 45 and 60 degree probes in tandem
D) holographic techniques
182. The amplitude reference line used to compare flaw response to reference hole response at varying
depths is abbreviated
A) ARL
B) AVG
C) DAC
D) DGS
183. If you are drawing a DAC for an inspection range of 200mm and your response from the 3/8 node
reference hole at 125mm has already dropped to 10% FSH you will have to use
A) a new calibration block
B) larger reference holes
C) the DGS (AVG) system
D) a split DAC
26
ESTestMaker Questions
185. If coupling conditions on a test piece are better than on a calibration block the transfer value will be
A) greater than zero
B) less than zero
C) not used
D) 1 (one)
188. The most pronounced effects of beam deviation by surface roughness are had when using
A) immersion testing
B) low frequency transducers
C) 70 degree contact probes
D) normal beam probes
189. The result of increasing the temperature of a test piece from 10°C to 30°C is
A) increase sensitivity
B) decrease the refracted angle
C) decrease the acoustic velocity
D) all of the above
190. The ability to discern individual defects separate from the initial interface signal is called
A) dead zone limit
B) near surface resolution
C) surface acuity
D) damping
27
ESTestMaker Questions
192. The maximum amplitude response from the 1.5mm diameter hole in the IIW Block #1 is used to
determine
A) longitudinal wave sensitivity
B) shear wave sensitivity
C) dead zone distance
D) both a and b
195. The purpose of the small V2 or DIN block (IIW Block #2) is
A) range calibration
B) determining exit point of probe
C) determining actual refracted angle
D) all of the above
196. The 5mm diameter through hole in the small DIN or V2 block is used for checking
A) shear wave sensitivity
B) longitudinal wave sensitivity
C) beam angle
D) all of the above
197. When setting sensitivity of a longitudinal probe using the IIW blocks
A) total number and size of last echo are used
B) response from the side drilled hole is used
C) both a and b can be used
D) IIW blocks are never used to set sensitivity
198. The dB difference between a signal 25% FSH and 100% FSH is
A) 6 dB
B) 8 dB
C) 10 dB
D) 12 dB
28
ESTestMaker Questions
199. To reduce the influence of incident angle when evaluating beam characteristics the preferred target is
A) a cylindrical target at right angles to beam axis
B) a spherical target of tungsten
C) a flat bottom hole
D) vee notches
200. Given that maximum acoustic intensity occurs at the focal point, the focal point for a flat 100mm
diameter 4 MHz probe in water is
A) 10mm inside the probe damping
B) the probe face
C) the end of the near zone
D) none of the above
201. In the AVG (DGS) diagrams originated by Krautkramer in 1959, the reduced range is given
A) as a fraction of the near zone
B) in millimeters only
C) as a ratio of the probe diameter
D) where the amplitude equals the ideal backwall echo
202. In immersion testing, defect sizing is improved in the flaw traverse technique measuring probe
movement to 6 or 20 dB drop levels by use of
A) lower frequency probes
B) focused probes
C) Y-cut transducer elements
D) oil coupling instead of water
205. When a flaw has a size that is less than the wavelength of ultrasound impinging on it,
A) the AVG (DGS) system must be used for sizing
B) dimensions are determined using the 20 dB drop method
C) it is not possible to determine the flaws' shape
D) it cannot be detected
29
ESTestMaker Questions
206. Signal averaging, correlation, and filtering are techniques used in ultrasonic systems to
A) extract weak signals from incoherent noise
B) improve resolution
C) characterize defects for type
D) none of these techniques are used in ultrasonic testing
207. The process where by a recurrent signal is extracted from incoherent noise is called
A) amplitude modulation
B) frequency modulation
C) signal averaging
D) filtering
208. Receiver noise must often be filtered out of a test system. Receiver amplifier noise increases
proportionally to
A) the square root of the bandwidth
B) the inverse square of the bandwidth
C) attenuation
D) temperature
209. A group of ultrasonic transducers arranged in some form of geometrical pattern with individual
transducers use sequentially as transmitters, receivers or both is called a
A) sequenced array
B) switched pulser
C) C-scan array
D) multi-element probe
211. In order to determine a particular material parameter by ultrasonics you would need to find its
A) critical angle
B) acoustic velocity and acoustic impedance
C) attenuation
D) any or all of the above depending on the parameter sought
212. Increasing a focused probe's focal length can have the advantage of increasing focal depth, but this is
achieved at the expense of
A) increasing attenuation
B) increasing focal spot size
C) reducing resolution
D) both b and c
30
ESTestMaker Questions
218. Even under ideal conditions, the electro mechanical coupling coefficient (k) will not exceed
A) 0.1
B) 1.0
C) 10
D) 100
219. If the product of the relative transmitter and receiver efficiencies of quartz is 1, which of the following
piezomaterials would have a product less than 1?
A) PZT
B) ZnO
C) PVDF
D) none of the above
31
ESTestMaker Questions
220. A dual crystal probe using PZT as a transmitter and PVDF as a receiver, as compared to just PZT used
in the send-receive mode would be
A) about 7 times more efficient
B) half as efficient
C) hampered by low signal to noise ratio
D) useless
221. The purpose of using metal powder in epoxies as probe backing material is to
A) increase acoustic impedance of the backing to match the ceramic piezoelements' acoustic impedance
B) provide an electric conductor to the back electrode
C) decrease bandwidth
D) none of the above
223. The result of a cylindrical shaped lens on the front of an ultrasonic probe is a(n)
A) point focused beam
B) line focused beam
C) asymptotic beam
D) negatively focused beam
32
ESTestMaker Questions
227. The most common method used to determine the nature of a probes' sound field is by
A) Fast Fourier transforms
B) impedance curves
C) side drilled holes
D) using two transducers, one transmitting and the other receiving
228. An angle beam probe is checked for refracted angle on an IIW and found to be 44 degrees. The IIW
block temperature is 10°C Moving to the 40° C test piece the refracted angle will be
A) more
B) less
C) the same
D) not possible to know
230. The time interval in microseconds between the first and last instant at which the value of the pulse
reaches 10% of its peak amplitude is considered the
A) pulse length
B) 20 dB drop length
C) dead zone
D) rise time
33
ESTestMaker Questions
234. For a given range, deviations of + or - 2° in refracted angle will result in greatest depth error for
A) 70° probes
B) 60° probes
C) 45° probes
D) error will be constant at all angles
236. A discontinuity whose shape, size location or properties make it detrimental to the useful service of the
product in which it occurs or which exceeds the accept/reject criteria is a(n)
A) defect
B) slag inclusion
C) crack
D) anomaly
237. A response or evidence of a response in non-destructive testing that requires interpretation is called
A) an indication
B) a defect
C) a flaw
D) signal-to-noise ratio
238. The component of ultrasonic wave attenuation resulting from conversion of mechanical energy to heat is
called
A) acoustic absorption
B) coupling
C) divergence
D) acoustic impedance
239. The property which determines acoustic transmission across a boundary between two media is termed
A) acoustic impedance
B) electric impedance
C) conductivity
D) transmissivity
240. Coupling two media to provide optimum transfer of ultrasonic energy between them is
A) a code requirement
B) acoustic impedance matching
C) best accomplished by dry coupling
D) all of the above
34
ESTestMaker Questions
241. For a given ultrasonic beam impinging on a surface, the angle of incidence, the angle of reflection and
the normal to that surface are
A) always equal
B) never equal
C) found by the inverse squared law
D) in the same plane
242. Loss of acoustic energy in a material due to scatter, absorption and dispersion is called
A) suppression
B) attenuation
C) damping
D) all of the above
243. A curve showing the relationship of amplitude to distance actually travelled to reflectors of the same
area is usually referred to as a(n)
A) AVG curve
B) DAC curve
C) DGS curve
D) NDT curve
244. The ultrasonic pulse received from the geometric boundary of a body normal to the beam axis is termed
A) backwall echo
B) a multiple reflection
C) the normal echo
D) a defect
245. An ultrasonic display in rectangular coordinates where distance or time of flight is represented in one
direction and probe displacement represented on the other and reflected pulses as bright marks on a
dark background (or vise versa) is called a(n)
A) A-scan
B) B-scan
C) C-scan
D) tomograph
246. Noise generated by the transmitting transducer which is produced in the receiving transducer of a dual
crystal probe is called
A) signal from noise effect
B) cross-talk
C) acoustic pick-up
D) FM cross-over
247. An assembly of ultrasonic crystals mounted so as to behave as though it were a single crystal is called
a(n)
A) acoustic mat
B) crystal mosaic
C) phased array
D) linear array
35
ESTestMaker Questions
248. The time interval between the initial pulse and the initiation of the time base sweep is termed
A) range
B) time of flight
C) programmed off-set
D) delay
249. The frequency at which the overall response of an ultrasonic pulse-echo flaw detection system is
maximum is the
A) dominant frequency
B) resonance frequency
C) nominal frequency
D) anti-resonance frequency
250. The lowest acoustic frequency which will cause a condition of resonance to be established in a given
material of given thickness is the _________ frequency.
A) fundamental
B) parallel
C) series
D) anti-resonance
251. A probe which incorporates separate transmit and receive crystals in the same housing is called a
A) dual crystal probe
B) phased-array probe
C) delta probe
D) normal probe
253. The display of remnant reflections originating from previously transmitted pulses due to too high a PRF
are called
A) transients
B) trip signals
C) ghost echoes
D) none of the above
36
ESTestMaker Questions
255. The boundary between any two media of different acoustic impedances is termed a(n)
A) interface
B) divide
C) interlude
D) front
256. A wave that is propagated in thin material, & whose velocity is dependant on frequency and material
thickness.
A) Lamb wave
B) Rayleigh wave
C) shear wave
D) new wave
257. The repeated reflections of ultrasonic pulses from between surfaces or discontinuities within a body are
A) ghost echoes
B) sing-around
C) multiple echoes
D) wrap-around
259. The method whereby a discontinuity is detected and evaluated using reflected pulses.
A) pulse-echo method
B) through transmission
C) bulk wave testing
D) crack tip diffraction
260. The number of times per second that an ultrasonic transducer is excited to produce a pulse is
A) determined by the crystal thickness
B) a function of the cable capacitance
C) the pulse repetition frequency
D) all of the above
261. The gain or attenuation setting at which indications are assessed is the ________ level.
A) reference
B) scan
C) threshold
D) bell
37
ESTestMaker Questions
262. Instrument settings which relate a reference echo of reproducible amplitude with which other instrument
settings relating to a discontinuity echo are compared is the
A) scanning level
B) threshold level
C) reference sensitivity
D) overall system gain
264. The relationship between amplitudes of an indication in A-scan presentation and the magnitude of the
corresponding received signals is used to determine
A) vertical linearity
B) manual scanning speed
C) longitudinal/shear mode energy ratios
D) none of the above
265. The device which incorporates one or more ultrasonic crystals mounted inside a liquid filled flexible tire
is commonly called a
A) rubber tester
B) rotating head probe
C) wheel probe
D) plate probe
266. Which pulse method of ultrasonic testing uses only a single crystal?
A) pulse-echo
B) pitch-catch
C) through-transmission
D) all of the above
38
ESTestMaker Questions
269. Examination of a metal's elastic properties by increasing the angle of incidence until either the
longitudinal or shear mode disappears is called
A) wave mode analysis
B) spectral analysis
C) specular analysis
D) critical-angle analysis
270. When doing immersion critical-angle analyses the pitch-catch technique is used. The receiver probe is
used to study the
A) reflected longitudinal wave
B) reflected transverse wave
C) transmitted longitudinal wave
D) Rayleigh waves
271. Non-contacting ultrasonic coupling, i.e. coupling across an air gap has its advantage in
A) avoiding fluid contamination of test piece
B) testing at elevated temperatures is possible
C) avoiding transducer wear on rough surfaces
D) all of the above
272. When performing a "normal beam" inspection of 60mm thick plate by immersion technique, the
minimum waterpath to avoid waterpath multiples in the first metal wall thickness display is
A) 15mm
B) 30mm
C) 60mm
D) 240mm
273. If the longitudinal waves are required at an angle other than zero degrees from the normal in the test
piece
A) the incident angle must be less than the first critical angle
B) the incident angle must be less than the second critical angle
C) the incident angle must be between the first and second critical angles
D) you are out of luck-it can't be done
274. Although no longer commonly used for thickness testing of steel plate, resonance testing still finds
application in
A) flaw detection in integrated circuits
B) locating nonbond areas in rubber to metal
C) shadow technique ultrasonic testing
D) all of the above
275. Ultrasonic systems similar to those used in nondestructive testing are used in
A) ultrasonic welding
B) diagnostic medicine
C) de-watering units
D) ultrasonic machining
39
ESTestMaker Questions
277. When information is presented as a B-scan on an oscilloscope, intensity (or amplitude) of a signal is
indicated by
A) strobe effects for signals over a threshold amplitude
B) digital readout on the corner of the screen
C) brightness of the spot on the scope
D) none of the above
278. When measuring crack depth using shear-wave soundpaths, amplitude drop and probe travel, you need
in addition to various distances and the refracted angles
A) a transmitting and a receiving transducer
B) the beamwidth in the vertical plane
C) maximum amplitude with respect to a side drilled hole
D) maximum amplitude with respect to a flat bottom hole
279. For a given flaw, with a physical vertical extent of 11mm, the probe requiring the maximum forward
travel to locate the flaw ends would be the
A) 0°
B) 45°
C) 60°
D) 70°
282. Ultrasonic methods used to monitor or detect fatigue cracks are based on
A) attenuation effects
B) amplitude of reflection energy
C) both a and b are used
D) none of the above
40
ESTestMaker Questions
283. An advantage of using ultrasonic surface waves over other wave modes is
A) good sensitivity to subsurface defects
B) the ease of focusing the beam
C) test piece contours are followed without loss due to direction changes
D) none of the above
284. The weld defect that results when a weld puddle solidifies from the outer edges and causes stresses
sufficient to produce cracking is
A) crater cracking
B) underbead cracking
C) transverse cracking
D) puddle jumping
285. The purpose of removing a weld cap by grinding it flush with the base metal in preparation for ultrasonic
testing is to
A) reduce scatter from redirection of beam
B) improve volumetric coverage
C) increase sensitivity to near surface defects
D) all of the above
286. When testing tubular products with ultrasonics it is essential to ensure a fixed alignment between the
probe and workpiece since a slight shift of tube axis could result in
A) a large change in incidence angle
B) damaging the probe
C) differential losses in frequency content
D) none of the above
287. In immersion ultrasonic testing of tubular products, line focused probes have an advantage over point
focused probes in that
A) higher frequencies are available
B) higher inspection rates are possible
C) higher resolution is achieved
D) higher sensitivity is achieved
288. When inspecting tubing using ultrasonic immersion methods in the tube mill
A) the tube is rotated under the ultrasonic probes
B) the probes are rotated around the tube
C) both a or b can be arranged
D) none of the above
41
ESTestMaker Questions
291. Which of the following non-metals would more commonly be tested by ultrasonics in the kiloherz range
of frequencies?
A) wood
B) concrete
C) rubber
D) all of the above
293. Ultrasonic velocities of glasses such as quartz, fused silica and pyrex are
A) not possible to determine
B) available for only the longitudinal mode
C) about the same as steel
D) about the same as water
294. Ultrasonics has been successfully used in testing rubber tires for determining
A) porosity location
B) de-lamination location
C) the state of cure
D) all of the above
295. Ultrasonic holography has an advantage over other imaging techniques in that it provides
A) total information from a single pulse
B) a 3-dimensional display of defects
C) maximum inspection speed
D) the best sensitivity of all imaging systems
296. Non-destructive testing is one of many applications of low-intensity ultrasound. Which of the following is
not an example of low-intensity ultrasound application.
A) acoustic holography
B) ultrasonic fish-finders
C) ultrasonic cleaning
D) ultrasonic flowmeters
42
ESTestMaker Questions
297. The weak emission of light that sometimes accompanies cavitation in high intensity ultrasonic fields is
called
A) sonoluminescence
B) fluorescence
C) bioluminescence
D) ultraluminescence
298. Diagnostic medical application of low intensity ultrasound has its advantage in
A) its ability to detect phenomena X-rays cannot
B) destroying gall stones
C) generating therapeutic heat
D) all of the above
300. Ultrasound intensity used in low intensity diagnostic testing (such as nondestructive testing) is on the
order of
A) 100µW/cm²
B) 100mW/cm²
C) 100kW/cm²
D) 100MW/cm²
43
ESTestMaker Questions
308. If a procedure fails to take into consideration a significant variable, such as temperature, the result will
be
A) unreliable inspection results
B) missed defects
C) lack of repeatability
D) over sensitivity
309. If it is possible to automate a previously manually performed scan, you can always expect
A) faster inspections
B) more accurate results
C) improved flaw detection reliability
D) all of the above
310. When dealing with ultrasonic nondestructive testing the letters PZT stand for
A) Paul Zamphir Tait, the discoverer of the piezo-ceramics
B) lead zirconate titanate
C) phosphoric zirconate tungstenate
D) none of the above
44
ESTestMaker Questions
311. The most commonly used backing used in ultrasonic transducers used in NDT is
A) air
B) spur's epoxy
C) epoxy resin filled with tungsten powder
D) brass
312. The material added to epoxy resin to increase acoustic impedance of the probe backing is usually
A) Rochelle salt
B) iron filings
C) tungsten powder
D) PbO (lead oxide)
315. In time of flight diffraction techniques, the first wave to arrive at the receiving probe is the _________
wave.
A) Rayleigh
B) diffracted
C) lateral
D) reflected
316. The longitudinal blurring of ultrasonic B-scan images that results due to ultrasonic beam width can be
reduced by
A) using a lower frequency probe
B) synthetic aperture focusing technique (saft)
C) increasing the soundpath
D) using immersion methods.
317. Using a contact probe with lucite wedge designed to produce a 70° shear wave in steel, which material
could you not inspect using shear waves generated from this probe ( V lucite = 2.68mm/µsec V steel
(transverse) = 3.2 mm/µsec)?
A) Aluminum (Vl = 6.32mm/µsec Vt = 2.49 mm/µsec)
B) Brass (Vl = 4.28mm/µsec Vt = 2.03 mm/µsec)
C) nickel (Vl = 5.63 mm/µsec Vt = 2.96 mm/µsec)
D) titanium carbide (Vl = 5.63 mm/µsec Vt = 5.16 mm/µsec)
45
ESTestMaker Questions
318. Given a wedge meant to produce a 60° refracted shear wave in steel (Vl = 5.66 Vt = 3.12 mm/µsec),
which material would you be able to also use this wedge for a 60° refracted shear wave?
A) aluminium 2117T4 (Vl=6.50 Vt=3.12mm/µsec)
B) beryllium (Vl=12.9 Vt=8.88mm/µsec)
C) iron (Vl=5.9 Vt=2.32 mm/µsec)
D) nickel (Vl=5.66 Vt=2.96mm/µsec)
319. The angle of incidence from lucite Vl=2.76mm/µsec to produce a 60° refracted shear wave in steel
(Vl=5.9 Vt=3.2mm/µsec) is
A) 23.1°
B) 34.6°
C) 48.3°
D) 58.1°
321. The inverse sine of the ratio of the acoustic velocity in the incident medium to refracting medium's
acoustic velocity gives
A) the first critical angle
B) the second critical angle
C) the third critical angle
D) the critical angle, which critical angle depends on the mode velocity used
322. The inverse sine of the ratio of the _____ velocities gives the first critical angle.
A) refracted long. and incident shear
B) incident long. and refracted long.
C) incident long. and refracted shear
D) incident long. and reflected long.
323. Given water, density=1g/cc Vl=1.5mm/µsec and lead, density=11.4g/cc Vl=2.16mm/µsec. The second
critical angle of sound moving from water to lead is
A) 7.8°
B) 14.4°
C) 28°
D) not possible to determine from this information
324. Given the acoustic velocities Vl=1.5mm/µsec for water and Vl=2.16 Vt=0.70 mm/µsec for lead. What is
the second critical angle for a longitudinal wave from water to lead.
A) 43.9°
B) 33.9°
C) 27.5°
D) none of the above
46
ESTestMaker Questions
325. In the welding process, removal of weld metal and base metal from the opposite side of a welded joint
to ensure complete penetration upon welding from that side is called
A) de-welding
B) beveling
C) back gouging
D) joint preparation
328. In welding processes, a preplaced filler metal which is fused into the root of a joint and thereby becomes
part of the weld is a
A) chill
B) chaplet
C) consumable insert
D) covered electrode
329. A crack occurring in the depression at the termination of a weld bead is called a
A) hot tear
B) termination crack
C) heat check crack
D) crater crack
330. In a welded joint, the minimum distance from the root to the weld face is the
A) hypotenuse
B) leg
C) effective throat
D) length of weld
47
ESTestMaker Questions
332. In welding, the area of base metal melted as determined on a cross-section of the weld is the
A) heat affected zone
B) fusion zone
C) dead zone
D) twilight zone
333. The portion of base metal that has not been melted but whose mechanical properties or microstructure
have been altered by heat of welding or cutting is called the
A) fusion zone
B) heat affected zone
C) dead zone
D) twilight zone
334. In a welding process using inert gas as a part of the operation, the purpose of the inert gas is to
A) react with the filler metal
B) provide a protective atmosphere over the weld puddle
C) increase the heat at the arc
D) harden the weld metal deposited
335. The protrusion of weld metal beyond the toe, face or root of a weld is called
A) overlap
B) undercut
C) pass
D) bead
337. Cracking occuring in the heat affected zone that does not usually extend to the surface of the metal is
A) transverse cracking
B) underbead cracking
C) laminal cracking
D) post weld heat treatment cracking
48
ESTestMaker Questions
339. When inspecting a curved surface with a contact probe it is advised to use
A) circular probes
B) rectangular probes
C) large probes
D) small probes
340. When inspecting rough surfaces, such as castings, coupling efficiency can be improved by
A) using larger dimensioned probes
B) increasing receiver gain
C) using lower frequency probes
D) increasing damping
341. For contact testing of thin steel plate (5mm) it is recommended to use shear waves at 80°. A significant
problem to contend with however is
A) a large fraction of energy is converted to surface waves
B) interference from side lobes
C) identifying defects
D) locating the position of defects
342. You are required to scan the full volume of an unground tee joint weld on 100mm thick plate. Scanning
is to be done from one side of the weld prep plate using a 45° probe. What is the scan distance you
must move from the tee plate surface to ensure 100% coverage of the weld having a leg of 30mm?
A) 142mm
B) 230mm
C) 283mm
D) 300mm
343. You are to scan an unground 50mm thick butt weld from one side of the weld only. The plate is 20mm
thick and full volumetric coverage of the weld plus 20mm heat affected zone is needed. Weld face is
20mm wide. What is the maximum scan distance from the weld centre-line for this coverage with a 60°
probe?
A) 40mm
B) 68mm
C) 98mm
D) 120mm
344. A sharp narrow signal whose amplitude remains constant when orbited and drops off quickly with probe
rotation is probably
A) lack of fusion
B) an isolated pore
C) slag
D) a transverse crack
49
ESTestMaker Questions
345. A ragged, cluster of individual spikes is located and determined to be a defect. Sound path varies,
amplitudes vary with rotation and orbiting with the probe. But the defect does not drop off completely
when orbited. It is most likely a
A) crack
B) slag inclusion
C) porosity cluster
D) lack of fusion
346. A sharp narrow defect signal is located during a standard A-scan of a weld. Rotation and orbit of the
defect cause it to drop off quickly with lateral probe motion the signal remains constant in both
amplitude and sound path. It is most likely a
A) slag inclusion
B) crack
C) lack of fusion
D) pore
347. Given a butt weld in a 55mm thick plate, ground flush, you locate a large sharp narrow reflector with a
soundpath of 154mm and exit point 128mm from the weld centreline using a 45° probe. If it has length
of 20mm the flaw is likely
A) lack of penetration
B) mis-match
C) lack of fusion
D) suck back
349. When performing a manual pulse-echo contact scan to evaluate the root area of a weld, a helpful aid is
a(n)
A) level 1 technician
B) straight edged probe guide
C) electronic gate
D) threshold alarm
350. Given an immersion probe with a focal length of 50mm in water, about what depth in steel would this
probe focus if positioned with normal incidence 40mm over a steel plate (using water couplant)?
A) 2.5 mm
B) 5.0 mm
C) 10 mm
D) 40 mm
351. The probe used in the 30-70-70 mode converted method uses
A) 3 separate elements mounted at different angles
B) 2 separate elements mounted at different angles
C) a single element mounted to induce a 70° refracted Longitudinal wave
D) a single element mounted to induce a 70° refracted Shear wave
50
ESTestMaker Questions
352. The indirect shear wave in the 30-70-70 mode conversion method is formed off the
A) 70° Long. wave
B) creeping wave
C) direct shear wave
D) damping material on the front of the probe shoe
51