You are on page 1of 4

Multileaf Collimator AL 01 II

Multileaf Collimator AL 02 II
Multileaf Collimator AFL 01 II
Multileaf Collimator AF 01 II

OEM
Multileaf Collimator AL 01 II, AL 02 II, AF 01 II, AFL 01 II

Product description Options and accessories

The purpose of the multileaf collimator is to Set of three-field templates for


collimate exactly the radiation beam emerg- SID 90 – 110 cm / 36 – 42 ‘’
ing from the X-ray tube assembly onto the SID 110 – 130 cm / 42 – 50 ‘’
size of the object to be displayed. SID 130 – 175 cm / 50 – 70 ‘’
SID 175 – 220 cm / 70 – 90 ‘’
Collimation is manual with the aid of the field
light localizer or motor-driven by automatic The collimator is prepared for installation of a
format collimation system. DAP chamber as well as a CANopen DAP
chamber.
The current source-image distance (SID) and
the size of the collimated radiation field are
displayed on a LCD display where available.

A line light localizer is available for centering


Customizing
the bucky except for type AF 01 II.
In order to fulfill your particular requirements,
For optimizing the image and radiation quality we also provide you with customized ver-
three different Cu-filter can be selected. sions of this product.

Features

• Lamellae close to the source to shield off


extrafocal radiation
• Additional prefilters integrated in the col-
limator for manual (AL 01 II) or motorized
operation (AL 02 II, AF 01 II, AFL 01 II)
• Bucky light generated by laser guarantees
a sharp light line for all source-image dis-
tances (not AF 01 II)
• Display of field sizes optionally in cm (de-
fault) or inch
• Rotation by ±45° with a stop in 0° position
by swiveling flange (not AF 01 II)
• Tape measure integrated in a user
friendly way in the collimator (not AF 01
II)
• Two foam protected accessory rails facili-
tate attachment of additional devices and
further filters
• 16 bit µController

2 of 4
Multileaf Collimator AL 01 II, AL 02 II, AF 01 II, AFL 01 II

Technical data

Type AL 01 II AL 02 II AFL 01 II AF 01 II
Part number 10092602 10092604 10092608 10092603
Collimation
Minimum field size 3 cm x 3 cm with SID 100 cm
Maximum field size 43 cm x 43 cm with SID 90 cm
Radiation field scale tolerance < 1,2% of the SID
Iris
(data for SID = 100 cm)
Shape none none Octagonal Octagonal
Minimum diameter 7 cm 7 cm
Maximum diameter 50 cm 50 cm
Additional prefilter
Type 0, 0.1, 0.2, 0.3 mm Cu
Operation manual motor-driven motor-driven motor-driven
Light localizer no light localizers
(data for SID = 100 cm)
Projection lamp 24 Vac / 150 W
Bucky light line laser
Brightness > 200 lx
Contrast > 4.5
Radiation field – light field tol. < 1.2% of the SID
Time switch 0...255 sec via CAN-Bus, 10, 30, 60 or 90 sec fix per jumper
Radiological properties
Inherent filtration 1.0 mm Al at 70 kV
Radiation shielding 150 kV
Leakage radiation < 0.5 mGy/h (at 150 kV / 3mA, at 1 m distance from tube assembly)
Electrical interface
Collimator power supply 24 Vdc (± 10%), – 30 Vdc (± 10%), max. 2.5 A
Light localizer power supply 28 – 40 Vac (± 10%) max. 10 A
> 30 Vac max. 8 A
CAN interface power supply 13 – 27 Vdc (± 10%) 150 mA
Data interface CAN-BUS
Mechanical interfaces
Flange to the tube assembly Diameter: 136 mm, thickness: 2 mm
Source flange distance 80 mm
Accessory rails 2 rails
Width: 178 mm; Length: 178 mm; Height of the guide: 2 mm
Quality standards CE, UL, CSA, DHHS
Weight 10.5 ± 1.0 kg
Ambient conditions
Operating temperature 10 ≤ T ≤. 40° C
Storage temperature – 40 ≤ T ≤. 70° C
Relative humidity 20 – 75%
Barometric pressure 70 – 106 kPa
Service maintenance-free

3 of 4
Multileaf Collimator AL 01 II, AL 02 II, AF 01 II, AFL 01 II

Dimensions
All dimensions in mm
Models AL01 II, AL02 II and AFL01 II

Model AF01 II

The Siemens Components Division is ISO 9001 certified, manufactures in accordance with the Quality System Regulations (QSR) as
defined by the Food and Drug Administration (FDA) and endeavors to comply with legal requirements concerning the environmental
compatibility of its products. In addition, the performance within the scope set by the guidelines of European Foundation of Quality
Management (EFQM) for the European Quality Award (EQA) is continuously monitored.
The reproduction, transmission or use of this document or its content is not permitted without expressly written consent. Offenders will
be liable for damages. This document is not considered to be a contractual specification. Kindly contact Siemens AG prior to using this
information for equipment design. Siemens reserves the right to modify the design and specifications contained herein without prior
notice. All rights reserved, particularly in connection with patent applications or registrations of utility model or design.

OEM
Siemens AG, Medical Solutions File: data_al01_al02_af01_afl01_oem_versb.doc
Henkestrasse 127, D-91052 Erlangen, Germany Printed in Germany
Telephone: +49-9131-84 2519 © 2007 Siemens AG
Fax: +49-9131-84 2501
http://oem.med.siemens.com E-mail: med.oem@siemens.com
Corporate headquarters:
Siemens AG, Wittelsbacher Platz 2,
D-80333 München, Germany
4 of 4