ES 642 9800 Manual Rev K
ES 642 9800 Manual Rev K
ES 642 9800 Manual Rev K
Operation Manual
ES-642 Operation Manual - © Copyright 2013 Met One Instruments, Inc. All Rights Reserved worldwide. No part of this
publication may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any other language in any
form without the express written permission of Met One Instruments, Inc.
Page 2 ES-642-9800 Rev K
Table of Contents
1 INTRODUCTION 6
1.1 About This Manual........................................................................................................... 6
1.2 Technical Service and Warranty ...................................................................................... 6
1.3 About the ES-642 ............................................................................................................ 6
1.4 Laser Radiation Safety and Conformity ........................................................................... 6
1.5 ES-642 Specifications ..................................................................................................... 8
3 SITE SELECTION 16
3.1 Site Selection Requirements ......................................................................................... 16
5 FIELD CALIBRATIONS 18
5.1 Ambient Temperature Calibration .................................................................................. 18
5.2 Ambient Pressure Calibration ........................................................................................ 19
5.3 Sample RH Calibration .................................................................................................. 19
5.4 Flow Sensor Calibration ................................................................................................ 20
5.5 K Factor ......................................................................................................................... 20
9 APPENDIX B - COMMUNICATION 33
9.1 Comet Software ............................................................................................................. 33
9.2 Serial Communication ................................................................................................... 33
9.3 Computer Mode ............................................................................................................. 34
9.4 User Mode ..................................................................................................................... 34
9.5 Output Formats.............................................................................................................. 35
9.6 MetRecord Format......................................................................................................... 35
9.7 Legacy Format............................................................................................................... 35
9.8 Status Field.................................................................................................................... 36
9.9 Example output.............................................................................................................. 36
9.10 Basic Command Summary ............................................................................................ 37
9.11 Report Model and Firmware Version [RV] ..................................................................... 37
9.12 Set Unit ID [ID x] ............................................................................................................ 37
9.13 Set Sample Time [ST x] ................................................................................................. 38
9.14 Set Operation Control [S x] ............................................................................................ 38
9.15 Exit User Mode [Q,X] ..................................................................................................... 39
9.16 Enter Calibration Mode [CAL] ........................................................................................ 39
9.17 Exit Calibration Mode [CAX] .......................................................................................... 39
9.18 Default Sensor Calibrations [DFLT] ............................................................................... 39
9.19 Reference AT [RA] ........................................................................................................ 40
9.20 Reference Flow [RF]...................................................................................................... 40
9.21 Reference Pressure [RP] ............................................................................................... 40
9.22 Reference Humidity [RR] ............................................................................................... 41
9.23 RH Set Point [SPR] ....................................................................................................... 41
9.24 Output Interval [OI] ........................................................................................................ 41
9.25 Advanced Command Summary ..................................................................................... 42
9.26 Modbus Address [MA] ................................................................................................... 42
9.27 Measurement Request [ME] .......................................................................................... 42
9.28 Request Query [RQ] ...................................................................................................... 42
9.29 Set K-Factor Multiplier [SK] ........................................................................................... 43
9.30 Network Mode [NW] ...................................................................................................... 43
9.31 Operation Mode [OP]..................................................................................................... 43
11 APPENDIX D - MODBUS 47
11.1 Modbus 3X Registers .................................................................................................... 47
11.2 4X Registers .................................................................................................................. 48
Table of Figures
Figure 2-1 Sharp cut cyclone attached to intake.............................................................................. 9
Figure 2-2 Basic Mounting Installation ........................................................................................... 10
Figure 2-3 Typical Wall Mounting .................................................................................................. 10
Figure 2-4 Typical Pole Mounting .................................................................................................. 10
Figure 2-5 Connector Layout ......................................................................................................... 12
Figure 2-6 ES-642 Power Connector ............................................................................................. 13
Figure 6-1 Pump Filter and Purge Filter Replacement .................................................................. 25
Figure 8-1 MDF Forward Light Scatter Laser Optical Engine ........................................................ 29
Figure 8-2 MDF Clean Air and Particulate Scatter Conditions in the MD Engine .......................... 30
Figure 8-3 The Effects of RH on Sample Concentration ................................................................ 30
Figure 8-4 ES-642 Flow System Diagram - Simplified for Clarity .................................................. 32
Figure 2-3 Typical Wall Mounting Figure 2-4 Typical Pole Mounting
Pin 1
CONC: 0.000 mg
BP 978 mbar
When the zero process has completed the unit will show the following screens:
The concentration is always displayed on the top line while the second line cycles through the
various measurements.
If an error occurs, the display will change to show the fault on the second line as shown below:
CONC: 0. 010 mg
Flow Error!
Example:
The maximum value of concentration desired is 1000 µg/m3. Type SPA 1000 <ENTER> The
concentration alarm relay set point is now 1000 µg/m3.
CONC: 0. 010 mg
TEMP 25.2 C
Note: The ES-642 ambient temperature sensor is a thermistor bead on the bottom panel of the
ES-642. As such it is not particularly accurate compared to solar shielded or motor aspirated
temperature sensors. An accuracy of ±2 degrees C is adequate for flow control purposes.
To change the temperature calibration using a serial communication program you would collocate
a reference standard near the intake head and then use the RA serial command to change the
temperature calibration.
Example:
The collocated temperature sensor reads 26 degrees C and the temperature sensor on the ES-
642 reads 23 degrees C. Type RA 26 <ENTER> The temperature calibration has now been
corrected to read 26 degrees C on the display.
CONC: 0. 010 mg
BP 978 mbar
The BP parameter is the current reading from the ES-642 pressure sensor. To change the BP
calibration using a serial communication program you would collocate a reference standard and
then use the RP serial command to change the BP calibration.
Example:
The collocated pressure sensor reads 974 mbar and the BP sensor on the ES-642 reads 978
mbar. Type RP 974 <ENTER> The BP calibration has now been corrected to read 974 mbar on
the display.
5.3 Sample RH Calibration
The CALIBRATE RH screen is used for field audits or calibrations of the sample relative humidity
measurement of the ES-642.
CONC: 0. 010 mg
RH 42 %
Note: The ES-642 sample RH sensor is located inside the pulsation dampening chamber. If the
ES-642 has been operating with the inlet heater running, the inside of the unit will likely be hotter
than ambient temperature, resulting in sample RH readings that will be lower than an ambient
standard. If the sample RH sensor is to be calibrated, make sure that the unit has equilibrated to
ambient conditions first, or a large artificial positive offset could be calibrated into the sensor
reading. This will result in the ES-642 over-reading the sample RH and turning the inlet heater on
when the actual RH is lower. Leave the factory default calibration in place unless necessary.
If the sensor fails, it will usually read an impossible value such as 125% or -25%.
The RH parameter is the current reading from the ES-642 humidity sensor. To change the RH
calibration using a serial communication program you would collocate a reference standard and
then use the RR serial command to change the RH calibration.
Example:
The collocated RH sensor reads 55% and the RH sensor on the ES-642 reads 50%. Type RR 55
<ENTER> The RH calibration has now been corrected to read 55% on the display.
CONC: 0. 010 mg
FLOW 2.0 lpm
The FLOW parameter is the current reading from the ES-642 flow sensor, in actual volumetric
liters per minute. You must send the serial command CAL <ENTER> to put the ES-642 into the
flow calibration mode and will automatically regulate to the set point (2.0 LPM). This may take a
moment.
To change the FLOW calibration using a serial communication program you would use the value
from the flow reference standard and then use the RF serial command to change the flow
calibration.
Example:
The flow reference sensor reads 2.2 LPM and the flow sensor on the ES-642 reads 2.0 LPM.
Type RF 2.2 <ENTER>. The flow calibration has now been corrected and will adjust itself back to
2.0 LPM on the display.
Remember to exit the flow calibration mode by typing CAX <ENTER>
5.5 K Factor
The ES-642 is calibrated on latex 0.6 micron micro-spheres. These provide an extremely
consistent calibration, but do not generally match the characteristics of all ambient particulate.
A K-Factor (multiplier) must be established for good accuracy and correlation to collocated
instruments. The base K-Factor multiplier is set to 1.0 at the factory.
The K-Factor is only valid at the same site and for the same particulate type. If the local particulate
source changes, the K-Factor may no longer be valid.
The SK command is used to set the K Factor.
CONC: 0. 010 mg
Counter Fault!
Note: The alarm indication will continue to show until the start of the next sample.
RH Sensor 9278-1
Purge Air
Particulate
Scatter Collection Laser Trap
Lenses
Figure 8-2 MDF Clean Air and Particulate Scatter Conditions in the MD Engine
Inlet
Purge
MD
Laser
Pump
Filter
Purge
Filter
Flow
Sample
RH
Purge
Pulsatio
n
Check
Sampl
e
Purge
Exhaust
OI Output Interval.
Q Exit User Mode.
RA Reference Ambient Temperature for Calibration.
RF Reference Flow for Calibration.
RP Reference BP for Calibration.
RR Reference RH for Calibration.
RV Returns the current software revision.
ST Sets / Reads the Sample Time in seconds.
S Manually Start/Stop a measurement.
SPR Sets / Reads the RH Set Point for the Inlet Heater Control.
9.11 Report Model and Firmware Version [RV]
Command Response Description
RV<CR> This command reports the
model, part number and
firmware version.
RV<CR> ES-642, 80708, R1.1.0 <CR><LF>
9.12 Set Unit ID [ID x]
The SID command is the same and is also supported for backward compatibility.
Command Response Description
ID x<CR> This command will Set or Get the Unit ID
string. This can be anywhere from 1 to 8
characters
ID 001<CR> ID 001<CR><LF> Set Station ID.
Issuing the command without a
parameter will respond with the current
setting value.
ID<CR> ID 01<CR><LF> Station ID “01”.
OP Display Operation.
QH Query Header
RL List Record Formats.
RT Set Record Format Type.
RQ Request Measurement (MetRecord Format).
SK Set K-Factor multiplier
UN Set / Read Units
9.26 Modbus Address [MA]
Command Response Description
MA a<CR> This command will set the Modbus
Address.
a = Is the Modbus Address from 1 to
247
MA<CR> MA 1 Modbus Address is set to 1
MA 3<CR> MA 3 Set the Modbus Address to 3
9.27 Measurement Request [ME]
Command Response Description
ME<CR> Request the Measurement in Legacy
output format.
This command will respond with the current measurement result in Legacy output format
regardless of the current Record Type setting. This command is generally used by a computer or
data logger system to Poll for data and is usually used with the Output Interval set to 0.
9.28 Request Query [RQ]
Command Response Description
RQ<CR> Request the Measurement in
MetRecord output format.
This command will respond with the current measurement result in MetRecord output format
regardless of the current Record Type setting. This command is generally used by a computer or
data logger system to Poll for data and is usually used with the Output Interval set to 0.
Response Description
DS n,id,r The response indicates the general header information
n – Number of field descriptor lines available.
id – Location ID
r – Reserved for future use. 0 is the default.
Example
DS 0<cr>
DS 5,01,0
Response
DS c,FieldName,MeasureType,units,prec,math,max,min
Parameter Description
c Field Number – 1 based.
FieldName Field Name string in printable ASCII.
MeasureType Measurement Type string in printable ASCII.
Units Engineering Units in printable ASCII.
prec Display decimal precision.
math Math type field. Vector (V), Scalar (S), Total (T), Minimum (MIN),
Maximum (MAX), Standard Deviation (STD).
max Maximum measurement value.
min Minimum measurement value.
Example
DS 1<cr>
DS 1,Conc,CONC,mg/m3,3,S,100.000,0.000
9.38 Request Units [UN]
Command Description
UN c Request the list of available channel / field units.
c – Desired channel/field 1 based
Response Description
UN c 1-a,... Returns the available units for the channel/field specified.
c – Specified field number
1-a – Unit enumerator and name.
... – More enumerators and names as the field requires
Example
UN 1<cr>
UN 1 1-mg/m3