Solea100 User Manual
Solea100 User Manual
Solea100 User Manual
BIOLABO S.A.S.
02160 MAIZY - FRANCE Version_140709
www.biolabo.fr
Operator‘s manual
Table of content
General Safety Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Warning symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Specifications.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
The Measuring System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Ball Function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Unit overview.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Unpacking the SOLEA 100. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Location. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Preparations for Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Turning on the device. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
The Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Reagent block preparation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
The Cuvette Magazine.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Washing Tank. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Refilling the Washing Tank. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Checking Probe Position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Prime Pumps.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Stand-by Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Turning off. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Taking the analyzer out of operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Routine program. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
The Main Menu Window.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Signs and Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Sample Tray. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Reagent Block. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
STAT. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Run Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Q.C. setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Run Control.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Reagent Database. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Hardware. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Result. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Backup Param. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Backup database. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
2
www.biolabo.fr
Operator‘s manual
Testparameter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Special Functions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Maintenance and Care. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Quick Reference Guide. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Start operation from stand-by. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Place a tray with patient samples in the analyzer .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Adding patient samples - Filling free positions in the sample plate. . . . . . . . . . . . . . . . . . . . . . . . . 56
Measuring Run Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Run controls by means of sample prep. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Displaying a run control. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Generation of a calibration curve. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Checking the calibration (successfully measured) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Printing Calibration Curves (example Fibrinogen)
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Cycle of a measurement (using the example Fibrinogen). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Preparations for the measurement.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Incubating.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Starting the measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Measuring clotting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Cuvette rack ejection and return transportation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Calculating the measurement value. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Pipetting Station and Dilutor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Fluid Sensor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Dilutor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Wash Station. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
List of accessories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
Glossary.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
www.biolabo.fr 3
Operator‘s manual
Wear gloves when handling blood, samples and objects contaminated by blood!
DANGER !
Strictly follow the existing regulations pertaining to the handling and manipulation
of reagents for laboratory use and blood samples!
IMPORTANT!
This instrument may only be operated by trained specialists, who have been
instructed and trained in procedures using In Vitro Diagnostics. They must be familiar
with the instructions and able to work accordingly, to fully utilise the analyzer
capacities.
ATTENTION!
Follow all warnings and instructions affixed to the instrument or stated in the
instructions.
Intervention in and modification of the product, not explicitly approved by the
equipment manufacturer, may result in loss of functional capability. The costs for
necessary repairs are to be borne by the user.
The equipment manufacturer is not liable for any damage resulting from non
compliance of the specifications stated in these instructions, damage caused by
handling of reagents and biological fluids, or other handling of the product which is
not in line with these instructions.
These instructions contain the information necessary for operating the analyzer.
It is strongly recommended that the user reads and understands these instructions thoroughly, in order to
fully utilise the analyzers capacity!
Meaning of the warnings used in these instructions:
4
www.biolabo.fr
Operator‘s manual
Warning symbols
Symbol Explanation
Protected earth
ON (mains switch)
www.biolabo.fr 5
Operator‘s manual
Symbol Explanation
Read and follow this information carefully before using the analyzer.
6
www.biolabo.fr
Operator‘s manual
Specifications
Protection class: 1
Working voltage: 100 to 240 VAC ± 10%
Supply frequency: 50 to 60 Hz
Power Input: 150VA
Fuses: 5 x 20mm, T 2.0A UL / IEC 127
Dimensions
w/o packing with packing
W x H x D: 72.0 x 62.0 x 55.0 cm 120.0 x 80.0 cm x 90.5.0 cm
Weight: 36kg 54kg
Space required
W x H x D: 170cm x 70cm x60cm
Ambient conditions
Operating temperature: +17°C to +28°C
Storage temperature: +10°C to +40°C
Relative humidity: 50% to 80%
Altitude: not for use over 2000 m.
Maximum heat output: 80W
Sound intensity: 65 dB (A)
Overvoltage category: II according to EN 61010 -1:2001
Pollution degree: 2
Usage environment: Indoor use in residential areas, commercial dwellings and light industrial
environments.
Temperature specifications
Incubation: 38.0°C ±0.8°C *
Measuring block: 40.5°C ±0.8°C*
Reagent cooling: 16.0°C to 22.0°C
* Corresponds to a temperature in the cuvette of 37.0oC ±0.8°C after 3 minute waiting period and a filling
volume of 220µl.
Sample volume
(plasma + reagent): minimum 150µl /maximum 260µl
www.biolabo.fr 7
Operator‘s manual
Introduction
The analyzer carries out a majority of sample preps for you and conducts chronometric and chromogenic
measurements fully automatically.
A patented procedure allows for simultaneous incubation of both sample and reagent in one cuvette.
Primary tubes can be placed directly in the sample rack from the centrifuge, after removing their caps.
All necessary analyses for a patient are performed in one pass. The identification (ID) is either entered via
barcode or keyboard. Names, numbers or consecutive numbers starting with any number can be entered.
New IDs can be recorded at any time, even during the measurement, and Stat samples will be performed
with a higher priority.
The analyzers sample throughput is approximately 100 PT’s and/or APTT’s per hour. When the cuvette ma-
gazine has been completely filled.
Due to the micro method a measuring volume of only 150µl is required. The required reagents can be
placed in a refrigerated station with 16 positions for reagents and 4 positions for controls.
You can integrate an emergency analysis into the running sample prep at any time. The analyzer will auto-
matically repeat measurements which are outside the tolerance.
Working with the analyzer is easy and requires only a few simple preparations:
8
www.biolabo.fr
Operator‘s manual
Features
▪▪ Walk-away-system
▪▪ Throughput approx. 100 PT’s per hour
▪▪ Patient oriented processing
▪▪ Automatic predilution
▪▪ Automatic repeat-testing
▪▪ Automatic calibration curve calculation
▪▪ 8 measuring channels
▪▪ Two wave lengths
▪▪ Red 620nm / blue 405nm
▪▪ Derived fibrinogen
▪▪ Automatic level detection
▪▪ Cuvette magazine for 460 tests
▪▪ 1 sample tray for 32 primary receptacles
▪▪ Cooled reagent block for 16 receptacles
▪▪ 4 positions for the quality control
▪▪ Chronometric, chromogenic and immunological tests
▪▪ Automatic reagent change
▪▪ Measuring system with process monitoring
▪▪ Bidirectional interface
▪▪ USB 2.0 interface
▪▪ Continuous operation and emergency analyses
▪▪ Rack return
▪▪ Open system for nearly all reagents
▪▪ Ready for immediate emergency analysis at any time
▪▪ Immediate result display
▪▪ Q.C. programme
▪▪ PC can be used at any time (multitasking)
▪▪ One main menu for the entire sample prep
▪▪ Request of work lists from the host
▪▪ Use of primary tubes
▪▪ Sample and reagent reloading possible
▪▪ Refilling of the cuvettes possible at any time
▪▪ Immediate stop of the analyser when opening the cover
▪▪ Error monitoring during the course of clotting
▪▪ Error criteria output
▪▪ Graphic display of the clotting process
▪▪ Current display of sample status
▪▪ Automatic subsequent testing
▪▪ Intergrated database
www.biolabo.fr 9
Operator‘s manual
incubating
This illustration demonstrates the tilting technology.
measuring
10
www.biolabo.fr
Operator‘s manual
Ball Function
with ball
2. In case of abnormal samples, the ball
2. Abnormal plasma concentrates the blood clot in the optical path.
The dynamics of the clouding difference between
the fluid and clotted sample are
without ball very high. This leads to positive detection of the
beginning of clotting.
www.biolabo.fr 11
Operator‘s manual
Unit overview
6
12
1 11
2 3
19
18
13 16
9
14 17 10
15
12
www.biolabo.fr
Operator‘s manual
1. Key field
STOP The sampler stops immediately after this button is pressed.
ALARM OFF Messages are acknowledged by pressing the ALARM OFF button.
This also removes the message from the message box. ALARM OFF has the same function as <F4> on the
keyboard.
2. Sample trays
Hold the primary cups to place the patient samples in the SOLEA 100.
3. Reagent block
Reagent blocks are loaded with reagents according to the schematic diagram on the Reagent Prep. menu.
By placing the samples in the SOLEA 100 they are automatically cooled to approx. 16-22°C depending on
room temperature. When the block is in the SOLEA 100, specific positions can be mixed. The control plas-
ma‘ positions are also in the reagent block.
Note: As an option, it is possible to work with several reagent blocks.
4. Wash Station
To avoid contamination, the needle is cleaned on the inside and outside. The washing cycle depends on
which test is selected.
5. Dilutor
The dilutor controls fluid movement together with the sample distributor.
6. Cuvette magazine
The magazine holds 60 cuvette bars. It is easily exchanged.
7. Sample distributor
Distributes plasma and reagent according to the programmed test volume.
8. Pipette probe
Distributes samples and fluids depending on the selected program.
14. Fuse
The fuses protect the instrument from damage.
www.biolabo.fr 13
Operator‘s manual
18. Host
The RS-232 interface is used to establish a connection to the LIS.
19. EXT
This interface is used to connect to the following SOLEA 100 using the same communication protocol.
14
www.biolabo.fr
Operator‘s manual
ATTENTION!
If the packaging or contents are damaged, make a complaint with the forwarding company and notify BIO-
LABO or your local agent. Assembly and installation should only be conducted by trained specialists, your
instrument supplier or a service engineer!
Direct or indirect damage to the instrument, caused by shipment in improper packaging, is excluded from
liability or warranty.
Location
Choose a location where the instrument is not subjected to direct sunlight, excess heat, humidity, dust and
vibrations.
The room temperature should be between 17°C and 28°C.
Place the instrument in a position which allows unhindered access to the mains outlet at all times.
Place the instrument on a firm, level table which has a depth of at least 60 cm and is up to 1.50 m wide.
The stability of the basis (table) should be adequate to support the instrument (50 kg).
The instrument can be lifted by the base plate at any position. 2 people should lift the instrument.
Choose the instrument location so that the instrument is protected from large temperature differences.
Direct sunlight, extremely bright lights, and radiators should be avoided.
The instrument must be located in a position where no water or fluids can enter the instrument.
The mains voltage must coincide with the technical specifications of the instrument.
The mains circuit must have adequate fuse protection.
The instrument must be connected to a properly grounded outlet.
If in doubt about mains voltage or the circuit in general, contact a qualified electrician.
ATTENTION!
Avoid connection to circuits where other appliances consume large amounts of current (for example centrifu-
ges) or which turn on and off frequently (for example refrigerator, water bath, etc.).
Avoid the immediate vicinity of radiators or devices which produce large amounts of heat, etc.
Avoid direct draughts.
Do not connect other electrical appliances which may cause interference with the circuit.
Do not set up the instrument near electrical appliances causing electric interference
(appliances bearing no CE label).
Ensure that other persons can not step on or trip over the power cord.
All connections to the instrument should be made with the instrument turned OFF.
www.biolabo.fr 15
Operator‘s manual
16
www.biolabo.fr
Operator‘s manual
The Screen
After login as user, the main screen appears, which is divided into six areas.
1.Menu window
Main menu for choosing the status windows required for the sample prep.
2. Test
This column is faded out if it is not relevant for the current programme.
3.Status window
The status window is changed by the menu items in the main menu. The selection is determined by the
cursor’s position in the menu. Depending on the selected menu item, information is displayed or data can be
edited.
4.Message box
System status display area (system and error messages).
5.Status display
Status of the function keys (time, date and software version).
On the left of the indicator, the samples which are already entered are indicated; on the right, the selected
reagent block (e.g. rb1) is displayed.
www.biolabo.fr 17
Operator‘s manual
In order to load the reagent block, select the reagent block in the main menu. A
diagram of the reagent block can now be viewed on the desktop. The identifica-
tion codes of the tests to be used, the reagent identification code and the name
of the reagent are displayed in the individual positions. Load the reagent block
according to the diagram on the screen. There are two fixed positions outside of
the reagent block for standard liquids. These two liquids are “Clean” and “Diluti-
on buffer”. The term “Clean” represents a decontamination liquid for the probe.
The following preset identification codes are used for the reagent block:
Place the reagent block into the intended position. In order to cool down the reagent to approx. 18°C, push
the module insert in as far as it goes (sensor controlled). “Mixed” indicates the positions which can be mixed.
18
www.biolabo.fr
Operator‘s manual
Replacement
The cuvette magazine holds 60 cuvette racks.
The cuvette magazine can be replaced or reloaded at any given time,
even if only half-filled.
When the instrument is working, it must be stopped with n before
continuing. As soon as the sampler has stopped, the magazine can
be reloaded or replaced with a full one. Please make sure the bar
cover is returned to its original position. Restart the instrument by
pressing o/ALARM OFF.
Washing Tank
The washing solution is stored in a receptacle which is situated
outside the analyzer. For the washing solution, de-ionized water or
distilled water may be used.
The water level is monitored by a sensor. If the water level is insuffici-
ent, an alarm is sounded and the following message is displayed in
the message window:
“Distilled reservoir low level. Refill”
www.biolabo.fr 19
Operator‘s manual
1. Test position
Note:
The arm moves to the “Home Position”.
2. Probe check
Select the menu item “Hardware“ by pressing e.
Confirm “Probe Check” with e.
Follow the instructions in the message window: “place the test receptacle in the wash station”, then e.
Note:
The probe positions itself over the washing station and dispenses approximate 3ml into the test receptacle.
Should the dosage be under the 2ml marker, please contact the service department.
Follow the instructions in the message window: “please take the test receptacle from the washing station,
then press o“. Select the item “Washstation“ e. Exit the work field “Hardware“ with ^.
Prime Pumps
Select the menu item “Hardware“ / “Prime Pumps“. Confirm with e. The tubing system of the analyzer
fills with washing solution. The next message box reads:
Note:
“Prime Pumps“ is only necessary if the system was switched off for several hours. In Stand-by mode, an
automatic short rinsing takes place every 30 minutes.
Stand-by Mode
If no patients’ specimens are being processed, the analyzer automatically goes into stand-by mode. In the
stand-by mode, a short washing of the probe is carried out every 30 minutes.
20
www.biolabo.fr
Operator‘s manual
Turning off
Select “Exit“ in the menu using yw. Confirm by pressing etwice.
Note:
The arm moves and a final washing process is conducted.
www.biolabo.fr 21
Operator‘s manual
Routine program
- Sample Prep.
Here the patients’ IDs and the desired tests can be entered.
- Reagent preparation
Display of the reagent block with reagent positions. Replacement of the reagent block when several blocks
are available.
- STAT
For inserting STAT’s into the running routine. STAT’s are given priority in processing.
- Run Display
Status display of the cuvette racks in the incubation / measuring block, as well as the results of the last three
racks.
- Calibration
Display and input of calibration curves and standard values.
- QC
Entry of the controlling plasmas for the “Q.C” selection and start of the “Q.C.”
- Q.C. Setup
Display, input and control of the “Q.C.” limit values for every test.
- Reagent Setup
Data display and data input of the reagents used.
- Calibration Setup
Data display and data input of the calibrators used.
- Hardware
Maintenance and control menu for “Probe”, “Syringe”, “Racks”, “LED”, “Water level and Temperatures”.
- Result
For searching the result database for measuring results, quality controls and system messages for viewing,
printing and transmitting this information to the host.
- Test Parameter
Selection and control of the selected test parameters. Changes only via the maintenance menu.
- Exit
The main menu must be exited before the PC is turned off.
22
www.biolabo.fr
Operator‘s manual
► “STAT”
Manual input
Sample Tray
Load the sample tray in numbering order.
Enter the patient data in the “ID No” column and the identification codes of the tests to be performed in the
“test” column. “Prep” indicates the sample tray number (1 or 2) and the position of the belonging patient plas-
ma in the sample tray (01 to 32).
Data Entry
The data can either be entered manually or with the optional internal scanner.
www.biolabo.fr 23
Operator‘s manual
If the data is entered manually, and the instrument is connected to an EDP system (Electronic Data Proces-
sing), then the test must be actively queried from the host by pressing p (for each patient) or jp (for all
patients entered).
If the data is entered with the internal scanner, and the instrument is connected to an EDP system, then the
corresponding tests are received from the host for each patient entered.
Once all the tests for a tray are completed, the status display below on the right in the routine screen chan-
ges from red to green. Now you can replace this tray with a new tray, scan or enter data and start.
You can add specimens at any time during the running sample prep by 2 ways :
- opening the cover. In this case, the arm stops immediately working, and goes to home position. From that
time, the analyzer gives you 1 minute to act. Move the cursor to a free position and enter the sample ID ma-
nually or with a scanner. Place the sample at the appropriate position. Subsequently, push the tray back into
the device.
=> The cover must be closed again within a minute to allow the analyzer continuing the work.
If the cover remains open for more than a minute, the current cuvettes rack is transferred into incubation
position.
- pressing n. Message “stopped (arm is still working) wait for ‘ready’ – restart F4“ appears. The analyzer
finishes pipetting the current cuvette rack and interrupts its current routine. Wait until the message “stopped
with F3 (arm ready: restart F4)“ appears. Open the cover, and remove the sample tray from the instrument.
Move the cursor to a free position and enter the sample ID manually or with a scanner. Place the sample in
the appropriate position. Subsequently, push the tray back into the device, press the m button in order to
read the specimen and reactivate the interrupted routine with o. In case of connected EDP and already
running routine, you need not press m, because the readjusted specimens are automatically started.
24
www.biolabo.fr
Operator‘s manual
Reagent Block
When the cursor is on “Reagent block” in the menu window, a diagram of the reagent block’s rack appears in
the screen’s work field.
In the reagent block, the reagent positions 1 to 12 are available for larger reagent bottles. Positions 13-16 are
available for smaller reagent bottles. Positions 1 and 3 are stirred.
Reagent monitoring
During the pipetting process, the XRC monitors the reagent volume. If there is insufficient reagent in the
reagent block, this is reported in the message box “not enough reagent ... for further testing”.
If the reagent is used multiple times, then the positions are used one after another. If all reagent positions
are empty, then the message “Liquid XY not found” appears. Display for the automatic liquid level control.
The graphic level display shows the current liquid level for each reagent.
Refilling Reagents
When pressing n the message “stopped (arm is still working) wait for ‘ready’ – restart o/ALARM OFF
(sample prep.)” is displayed. The analyzer finishes pipetting the previously started cuvette rack and interrupts
its current routine. Wait until the message “stopped with n(arm ready: restart o/Alarm Off.) sample prep.“ ap-
pears. Remove the reagent block from the device, and replenish the corresponding reagent. Put the reagent
block back into the device and start “Sample prep” with o/ALARM OFF.
After a reagent has been changed or refilled, the liquid level is automatically checked. This is done for safety
reasons.
www.biolabo.fr 25
Operator‘s manual
Never place foreign objects, e.g. coins, under the reagent receptacles.
There is also the alternative of extracting the reagent block without using n in order to replace re-
agents. Simply open the cover. In this case, the pipetting probe stops immediately.
ATTENTION !
Do not extract the reagent block if the pipetting probe is currently pipetting a reagent.
The software will allow this procedure within 60 seconds. In case the cover is not closed again within the 60
second time limit, the rack currently being pipetted is discarded.
26
www.biolabo.fr
Operator‘s manual
STAT
Analyses for STAT patients can be performed at any time. STATS, including result printing, are prioritised.
The sequence is the same as in the menu option “Sample prep.“, only that you can access the ID field using
menu item “STAT“ e.
Now the specimen IDs read manually entered are STAT specimens and are processed with priority. These
specimen are marked with ►.
You can place STAT specimens in any free place in the tray.
End the data input ^ and start with m for manual input. The STAT specimens are inserted according to
the process described in chapter “adjusting specimen“. Processing and measuring of STATs in “sample prep”
processing have the highest priority.
www.biolabo.fr 27
Operator‘s manual
Run Display
Status Display of being processed Cuvette racks.
The menu item “Run Display” displays the status of eight cuvette racks. The display scrolls from bottom to top.
28
www.biolabo.fr
Operator‘s manual
Calibration
In the “Calibration” menu values and times for your calibration are registered. There are three alternatives of
calculating a calibration curve: “Manual”, “Automatic” and “Fully automatic”.
1 3
2
4
6
7
8
9
10
1) Test
The name of the previously selected test is automatically shown here.
2) Reagent
The reagent used is assumed from the “Reagent Database” entries.
3) Date
This date and time is the validation date of the current calibation chart.It is automatically entered for each
value change in the chart if it is confirmed with „yes“ at the option „valid“.
4) Lot-No
Lot number of the corresponding reagent in the left side. It is transferred from the entries in the “Reagent
Database“.
6) Automatic
The analyzer determines the values from calibration plasmas with different concentrations.
www.biolabo.fr 29
Operator‘s manual
7) Fully Automatic
The analyzer calculates calibration curves using fully automatic plasma predilution.
8) Manual
Entering individual calibration curve values using the keyboard.
9) Curve
When the cursor is in the “Curve” box and eis pressed, the graphics screen opens and displays the
calibration curve.
10) Normal/ISI
Optional when calculating the INR value.
ATTENTION !
The following parameters of the calibration curve display can only be altered by the system administrator:
30
www.biolabo.fr
Operator‘s manual
- LOG/LOG - LOG/LIN
- LIN/LOG - LIN/LIN
The form of scaling is set according to the reagents you use and the parameters before or during the installa-
tion of the device. The same is applicable for the allocation types:
Values of the curve are given e.g. in %, mg/dl, g/l, IU/ml on the X-axis of the curve. The Y-axis shows the
corresponding times. All calibration points must be within the “min. cal.“ and “max. cal. values“. No values
outside this range are calculated. The curve can be printed using i.
With e you can enter the first value for the desired dilution series. If you press e after entering the
value, the cursor goes to the next field of this column. All the calibration curve points are entered ^. The
cursor switches to the field “Start“. Diluted plasmas are placed in the corresponding positions in the speci-
men stand and started with e. Once the measured values are automatically entered in the chart, the
device shows the message “System has completed calibration (see curve)“. Press e to move the cursor
to the field “Curve“. Now you can view the curve e. After leaving the chart with ^ the cursor is in the
field “Accept“, which has changed from “no“ to “yes“. If you want to accept the new curve, press ^. If you
want to view all old curves, change the field “Accept“ with k from “yes“ to “no“.
Note:
If you do not wish to alter the dilution series values of your calibration curve, go directly to “start” to begin
your measurements.
www.biolabo.fr 31
Operator‘s manual
Dilution values are displayed in the right column. The message box specifies the positions in which the diluti-
on buffer, reference plasma and empty receptacles for dilutions must be placed.
Once you have brought calibrator, empty receptacles, buffer and all the reagents in the correct positions,
confirm with e, when the cursor is in the field “Start“. The analyzer starts pipetting the dilutions and
shows the message “System is working (calibration)“. Once the calibration curve is successfully measured,
the message “System has finished calibration (see curve)“ appears in “Message“. Confirm from menu item
“Calibration curves“ with e, the cursor switches to the field “Curve“. Confirm again with e and you
can view the graphic display of the calibration curve. Once you leave the curve with eor ^, the cur-
sor is in the field “Accept“. If you want to use the new calibration curve, press e. If you want to use your
old calibration curve, switch from “yes“ to “no“ with kand leave the work area with ^.
32
www.biolabo.fr
Operator‘s manual
In case a calibration curve was measured incorrectly, move the cursor in the menu to “Calibration curves“
e. The cursor is in the work area “Calibration curves“ directly in the field “Values“. If you want to supple-
ment the missing value, press e. Go to the position of the missing value and enter the value. Once you
have entered the value, you will see the calibration curve and you can accept or reject it. If you do not want to
enter the missing value of the calibration curve, exit the work area “Calibration“ using ^. Your old calibration
curve remains unaltered.
www.biolabo.fr 33
Operator‘s manual
Q.C. setup
On the “Q.C. setup” menu, the confidence values of the control plasmas are entered. A chart is available for
every control plasma measured so far.
Plasma with Lot No.: Plasma is adopted with name and lot number from the menu item
„QC“. When using a same plasma for several parameters, make sure to enter the same name and lot num-
ber for all of them. If not, an error message will be displayed.
Date : enter the expiry date stated on the label of the control plasma.
Valid (h) : validity of the control result. After this time a QC must be run again to be valid.
Range (Low and High) : validity range. These limits will be checked, and used for the chart.
34
www.biolabo.fr
Operator‘s manual
Use the k to toggle between the following displays: all, unflagged values, or values with error codes
(flags).
To return to the main menu, press ^ 2-3 times, depending on the position.
SD
Specifies the calculation of standard deviation of the chart.
CV
Specifies the calculated coefficients of variation of the chart.
i
The chart can be sent to the printer for printing by pressing .
By pressing ^ three times you return to the main menu field.
www.biolabo.fr 35
Operator‘s manual
Run Control
Select “QC” on the Menu yw.
Status window
In the control plasma status window, you can specify the plasma to be used. The measurements for the run
control can also be started from this window.
For each parameter, you can launch up to 3 controls. Here is the description of the screen :
Test :
Name and abreviation of each test
Plasma :
Name of each plasma set in the „QC Setup“.
Position :
The position of each plasma will be automatically defined after requesting QC.
Start :
For each parameter, select the controls to perform by switching from „No“ to „Yes“ this item.
N:
Define here the number of repetition of each control. This value is not accessible in „Routine“ mode.
Result :
When a QC result is found within the allowable limits, the QC is tagged „ok“. When not found within the
limits, or if validity is expired, it is tagged „not ok“.
Valid :
When a new QC is performed, the validity time is reset. Then the remaining time to expiration of QC result is
indicated here. After this time, the QC is tagged „not ok“.
36
www.biolabo.fr
Operator‘s manual
Select “QC“ in the menu by pressing ywe. The cursor switches to “no“ of the first position.
You can switch all or some of control levels for each parameters. The software will allocate the positions for
controls according to you choice.
Check to make sure that all required control plasma and reagents are in their determined positions. Press
m to start the run control.
“System is operating (run control)” is displayed in the message window.
When the analyses are completed the established results are transmitted to the result database and the
corresponding chart.
www.biolabo.fr 37
Operator‘s manual
Reagent Setup
Select “Reagent Database” in the menu window yw. The test screen will appear in the status window.
Using “Reagent Database” e the entries in this menu item can be edited.
To move to a different column, press ywe. For each test displayed the following information can be
specified:
- Reagent name
- Lot number
- Expiration date of the reagent
- Comment
The information regarding reagent name and lot number are automatically entered in the item calibration
curves.
38
www.biolabo.fr
Operator‘s manual
Hardware
The menu item “Hardware” is a control and maintenance programme (also see chapter maintenance and
care).
Status Window
To start the programme, select “Hardware” yw and confirm with e. Use yw to move in the status
window, with e the desired action is conducted.
- Sampler:
Test Position
Probe Check
Select “Hardware“ e.
Select “Sample“ e.
Select “Probe Check“ e.
ATTENTION
The arm moves to the “Home Position”.
www.biolabo.fr 39
Operator‘s manual
ATTENTION:
The probe moves over the washing station and then pipettes the solution into the test receptacle.
The lower section of the test container should be filled. The pipetting stream should be vertical and should
come out of the probe as a single stream. After this has been done, the tip of the probe should be checked
for any development of water drops. Water droplets on the tip of the probe are an indication that the system
might be leaky.
Message: “Please remove test receptacle from the wash station and press o“ .
Change Position
- Push the protective tube back until the probe is exposed on the top.
Attention!
When removing the tube, tightly hold on to the probe’s black guide.
- Press the probe out and away from the notch in the guide.
40
www.biolabo.fr
Operator‘s manual
- Press the silver locking cap upwards and then pull the probe up and out.
- Replace the needle in the opposite order
Attention!
When fitting the probe make sure it is inserted as far as it goes.
right wrong
www.biolabo.fr 41
Operator‘s manual
Reagent Position
Confirm the “Reagent Position“ e. A list of reagent positions is displayed for selection. Enter the position
to be approached e. After a brief washing the arm moves to the corresponding position. This serves the
purpose of controlling individual positions in the reagent block. Exit the “Reagent Position” with ^.
Select the “Wash Position“ by pressing yw, then press e and exit the work area with ^.
Probe Clean
In the menu “Hardware“ select the submenu “Probe Clean“ e. The probe exits the wash position, please
pay attention to the message box.
Pipette 1ml of 5% hypochloride in the washstation e.
The probe moves into the “Wash Position” and draws up the solution, allow this to work for approximate
15 - 30 min. e. Use ^ to exit needle cleaning.
The probe should be cleaned manually if any droplets, on the shaft of the probe after pipetting, are observed
The probe must first reach its final position before any cleaning can be done.
Please be careful and avoid any injuries which could be caused by the tip of the probe.
Always wear protective gloves in order to protect yourself against contamination!
ATTENTION
The probe should only be cleaned from top to bottom. Please pay special attention to the
instructions displayed in the message box!
42
www.biolabo.fr
Operator‘s manual
A cuvette rack is transported to the pipetting position. The pipetting sensor searches for the reference point
on the rack and automatically corrects the probe’s Z - position. Using the cursor keys, both X and Y positions
can be optimised (maximum of 10 steps in one direction).
Press ^ to finish the procedure. The arm then drives to its “Home Position”, and the rack returns to the
wait position.
Wash Position
The “Wash Position” must always be selected in order to be able to exit the work area “Hardware“ (exit with
^).
Syringe
“Syringe“ is a maintenance programme for the extraction of dosing syringes and filling prime
pumps.
Select “Syringe“ yw in the operating area “Change Position“ e. The syringe moves a few
steps down. You can now uninstall the syringe:
www.biolabo.fr 43
Operator‘s manual
Please ensure that the syringe is tightly fitted on the connection as well as at the bottom on the dilutor.
Select “Start Position“ e. Now the syringe moves back to its initial position.
Select “Prime Pumps“ e. “System is not yet ready (Prime Pumps)“ appears in the message box. Wait
until the message “System ready (Hardware)“ appears and exit the work area with ^.
Prime Pumps
Select the menu item “Hardware“ / “Prime Pumps“. Confirm with e. The tubing system of the analyzer
fills with washing solution. The next message box reads:
“System is not yet ready (Prime Pumps)“
After completion the message box reads:
“System ready (Hardware)“
Exit the work field with ^.
Note:
“Prime Pumps“ is only necessary if the system was switched off for several hours. In Stand-by operation, an
automatic short rinsing takes place every 30 minutes.
- LEDs:
LED Test
By confirming with e, the light intensity of the measuring system is checked. At the end of the test, the
message “LED is OK” appears in the message box.
If an error message appears in the message box, please clean the measuring block.
44
www.biolabo.fr
Operator‘s manual
Upon selecting this menu item, a window opens which displays the A/D values of the four measuring chan-
nels. Please clean the measuring block if one or more channels are outside the displayed range.
Water reservoir status and the temperatures are displayed in a chart. If the water level is not shown to be
OK, refill the water reservoir as described in chapter “the washing tank”. If one of the temperatures is not OK,
the analyzer stops the measurement routine.
Result
In the menu option “Result“ you can view and edit all saved messages and measurement values. In case the
cursor is on “Result” in the menu, press e and the screen changes to database selection.
www.biolabo.fr 45
Operator‘s manual
Results: Test results for all patients incl. all reaction curves.
Controls: Contains data from measurements of “Q.C. Set-up“.
Calibration curves: Contains all the executed calibration curves.
Protocol: Contains the work log of the previous measurements.
Messages: List of generated error messages.
Status: Number of specimens, Q.C.’s and error messages.
Results
If the item “Result” of the database is confirmed with e, the display changes to the list of patient data.
The patient measured last is displayed at the bottom. You can scroll with yw through the patient data.
You can select the following functions: “Print“ and “Send“.
Result data
Confirm a new ID number with e, the corresponding index card for that patient is displayed.
Result search
When opening the option “Result search“ under “Result“ by pressing e, an input field is opened. The
system will search for an alpha-numeric combination, which may take place at any place in the “ID.-no.“
(observe upper and lower case letters).
Transmit New
Here only analyses which were not yet sent to the central computer on the current day are displayed. Sen-
ding is started with p.
Control
Move the cursor to the working area “Result“ on “Control“ and confirm with e.
All quality checks available in the database are displayed. After selection of a control with yw and
eall results are displayed for this Q.C. Continue to proceed as in the patient‘s database. With t you
have the opportunity of entering a comment for this control.
Control Search
Control search see “Results Search”.
Calibration
Move the cursor to the work area on “Calibration“ and confirm with e. All calibrations of this system are
displayed with test name, test abbreviation, type of calibration, status, date and time. Select a standard cur-
ve with yw and with e all results are displayed for this calibration. With t you have the opportunity
of entering a comment for this standard curve.
Protocol
Under the item “Protocol“ you can view all measurements from the last 30 days. All individual values and
averages with all possible information are displayed. Additionally after selecting with yw and e,
46
www.biolabo.fr
Operator‘s manual
Errors
Upon pressing ea display appears which gives information regarding the number of patient plasmas
measured up until now, the number of checks conducted and the number of registered messages.
Error search
Status
www.biolabo.fr 47
Operator‘s manual
- “Patients”
- “Controls”
- “Error Messages”
- “Strips”
- “Cuvettes”
You will need to enter exact data outside of these limits in order to recall information from the database.
Results 90 days
Control 750 days
Calibration 750 days
Error Messages 180 days
Protocol 90 days
48
www.biolabo.fr
Operator‘s manual
www.biolabo.fr 49
Operator‘s manual
Select a data record by using yw and then press the ekey. You have a list of all single values.
Press u to display the chart. The chart includes all measuring values contained in the data record.
50
www.biolabo.fr
Operator‘s manual
Mark all data records which are to be deactivated with k. A ”-” symbol will appear in front of each deacti-
vated data record.
Backup Param
These functions are used to make parameter back-ups.
When backup is selected, all system relevant settings and test parameters are saved on a USB stick.
Make sure to always have a current backup of your parameters!
Connect a USB stick (provided USB stick) to the PC and select “Backup Parameter”.
The message “Sucess” appears after finish.
Backup database
These functions are used to make database backup. When selecting this function become secured all pati-
ents/control results without reaction curves and error messages on a USB stick.
Connect a USB stick (USB provided stick) with the PC and select them for “Backup data base”. The mes-
sage “Success” appears after the completion.
www.biolabo.fr 51
Operator‘s manual
Test parameters
The individual parameters for the selected tests are displayed in this working area. It is possible to tell the
system if you would like the selected test done in single or double determination.
All of the other parameters can only be changed if you have the appropriate access rights.
Exit
52
www.biolabo.fr
Operator‘s manual
If you would like to stop working with the analyzer, select “Exit“ in the menu.
Confirm 2 times with e.
ATTENTION!
The arm moves and a final washing cycle is conducted.
The user interface to the login console. Press aS and confirm with e.
The PC switches off automatically.
Now switch off the analyzer.
Special Functions
Altering Data
The following functions are available to alter or delete data in the “Sample Prep.” menu:
a) When the cursor is in the status window the following key combinations are enabled:
Move the cursor to the desired test input with yw. Select the desired line area with xz and edit it. If
you wish to delete the entire line including ID.-no. and tests, press bc.
Attention!
The line will be deleted without any further requirement for confirmation.
After starting sample prep, all data can be viewed with {}, but no longer altered.
Probe
In the menu “Hardware“/”Sampler”:
Select sub menu “Test Position“ and check the probe as described in “Probe Check”. If the probe or their tip
is visibly damaged (bent or deformed), please exchange the probe (please refer to chapter “hardware”).
Visible signs of wear and tear on the coating of the probe are normal and do not negatively influence the
proper functioning of the probe.
Tube system
Select the sub menu “Prime Pumps” in the menu ”Hardware“. Check the general sequence
and check whether air bubbles are present in the system. Check the canister fill level and refill if necessary.
www.biolabo.fr 53
Operator‘s manual
Weekly Care
Probe Cleaning
- In the menu “Hardware“ / “Sampler”, select the sub menu “Probe Clean“ e.
The probe moves out of the wash station. Please pay attention to the message box.
-Pipette 1 ml of 5% hypochloride into the wash station e.
The probe moves into the wash position and draws up the solution,
allow this to take effect for approximately 15 - 30 min. e Use ^ to interrupt/stop the current process
- Press w to switch to “Prime Pumps“ e.
- When done, the sub menu “Probe Check” should be selected.
Please follow the appropriate steps as described in the “Hardware” section.
- Check how firmly the syringe is mounted: both top and bottom.
- Check for condensation beneath the teflon tip.
If there is condensation present, then this is a sign that the syringe may be leaky. The syringe should then be
replaced.
Cleaning
For cleaning the analyzer take a lint - free cloth which is moistened with water.
Incubator (Track)
Wipe with a lint free cloth which is moistened with NaCl 0.9%.
Disposal
The waste water should be disposed according to local regulation.
If the used racks are disposed, strictly follow the existing regulations pertaining to the handling and manipula-
tion of reagents for labotratory use and blood samples.
Maintenance Recommendation
We recommend changing the probe and syringe after 20.000 tests. An alternative to this is a yearly inspec-
tion by an authorised technician.
54
www.biolabo.fr
Operator‘s manual
www.biolabo.fr 55
Operator‘s manual
Menu yw “Hardware” e
“Test Position” ereference point OK.
“Sampler” e Follow the instructions at the message box.
“Probe Check” edosage flow OK.
“Wash Position” e^ basic position
of the probe.
“Prime Pumps” eafter completion exit with ^.
Dissolve the reagents according to the manufacturer’s specifications, and place in the reagent block.
Menu yw “Reagent Block” Display of the reagent
Menu yw “Sample Prep.” Menu yw “Sample Prep.” eManual input of the patient‘s data. Exit
operating area with ^m.
Menu o / ALARM OFF Start the interrupted “Sample Prep.”
Menu yw STAT e Arrange the samples in the free positions of the tray. Or manual input of the
patient‘s data. Exit operating area with ^.
Menu yw “Run Display” Run display of the measurement results or protocol of the printer.
56
www.biolabo.fr
Operator‘s manual
Menu yw “Reagent Database” e For Fibrinogen enter the reagent, the new load number and
the expiration date.
Values e Switch to reference using w . Enter the Fibrinogen concentration from the package
insert into the reference, then press e.
www.biolabo.fr 57
Operator‘s manual
Menu yw “Calibration” e The cursor is located on “curve“. If not, press yw to move it the-
re. View the curve by pressing e and confirm it with e.
Accept: “YES“ - if you would like to use the new calibration curve, confirm with e.
If you want to use the old calibration curve, press k to switch from accept: “YES” to
accept “NO”.
58
www.biolabo.fr
Operator‘s manual
Place sample receptacle into the tray (bar code aligned). Enter the ID number manually via the keyboard in the correspon-
ding position.
The instrument status changes from “ready” to “working”. The status display above the message window
changes from green to red, in the message window “system working” is displayed. The sample being pro-
cessed is marked with a * on the left of the ID number. A cuvette rack is moved forward from the holding
position to the pipetting position, the sampler starts the pipetting process with the pipette volume specified in
the volume chart for plasma dilution and reagents.
Reagent pipetting
In the reagent block, the probe takes up 40µl Kaolin, then 20µl Fibrinogen reagent and dispenses this volu-
me in the cuvette rack through the lower opening of the same cuvette (20µl of the fibrinogen reagent taken
up last and 20µl Kaolin taken up first). In the wash position the probe is then cleaned. Since the fibrinogen
reagent is very agressive additional cleaning is conducted with Clean solution, followed by another rinse
cycle. With this there are two separately positioned drops in the cuvette at the pipetting position which are
Incubating
For incubating, the cuvette bar is moved by the cuvette transport system, according to the time setting on “In-
cubation” (normally 180 seconds), through the three incubation stages before reaching the measuring block.
Measuring clotting
After tilting the cuvettes (that is after both reaction liquids have merged together) the electric signal produced
by the optical measuring channel is actively traced. In addition, the “measuring thresholds” are applied above
and below this measuring signal.
www.biolabo.fr 59
Operator‘s manual
If clotting occurs, the optical density changes and the measuring signal will cross one of these two
thresholds, in this moment the measuring time is registered.
With Fibrinogen, it is typical that the optical density decreases when clotting starts (more light reaches the
photoelectric cell). This effect originates from the fact that the Kaolin from the clot produced is pulled out by
the rotating ball.
The measured specimen is marked with ≡ left of the ID number. The device status changes from “operating” to “ready” and
the display status above the message box changes from red to green.
The probe moves into the clean position and takes up 100 µl Clean, water is pumped into the washing posi-
tion, the probe measures 50 µl of Clean into the washing postion, detects the mixture volume, immerges and
takes up 300 µl for the final cleaning process.
60
www.biolabo.fr
Operator‘s manual
Troubleshooting
Probe misses Probe bent during pipetting. Press <Stop> button of the analyzer, select
correct pipetting “exit”, wait 1 min., then turn PC + analyzer
positons. off. Check test position after restarting, adjust
if necessary. With the Analyzer turned off,
manually check left/right/forward/back shaft
guides, with probe in upper position, for smoo-
thness of operation. After another manual
check turn the Analyzer on and check the test
position.
Fluid Sensor
The probe sensor is not The pipette probe is not Check that probe is assembled correctly, check
detecting liquid. secured correctly. The sensor that the cable’s plug connections are secure and
cable is not connected cor- check the cable for damage. The cable must be
rectly or the cable is defec- replaced if it is damaged in any way.
tive.
www.biolabo.fr 61
Operator‘s manual
The pipette probe re- The pipette probe is not Check that probe is assembled correctly, check
mains above the sample secured correctly. The that the cable’s plug connections are secure and
/ reagent and starts to sensor cable is not connec- check the cable for damage. The cable must be
continuously move up ted correctly or the cable is replaced if it is damaged in any way.If the error
and down. defective. Sensor setting is is still present, then undertake sensor setting as
incorrect. described below.
WARNING!
- Press the “Stop” button.
- Press the adjustment button on the pipette
arm.
- Press the “Stop” button.
- Restart the SOLEA 100 using the o button.
The probe position Cause / sources of error:The Check that probe is assembled correctly.
above the cuvette is too pipette probe is not secured
high. correctly.
The probe sensor is not Sensor setting is incorrect. Undertake sensor setting as described below.
detecting liquid or only WARNING!
very large volumes. - Press the “Stop” button.
- Press the adjustment button on the pipette arm
- Press the “Stop” button.
- Restart SOLEA 100 using the o button.
Dilutor
Tube has worked loose at Pull tube from the guide to the compression fit-
the dilutor’s compression ting, then retighten the fitting and run it proper-
fitting. ly. (If only the fitting is tightened, any possible
tension on the tube could cause the fitting to
loosen again. Therefore dismounting and reins-
tallation are necessary after the fitting has been
retightened!
Note:
Following any service to the dilutor / tubing / probe system the following checks must be performed on the
“hardware” menu:
“Test Position” - “Probe Check” - “Prime Pumps”, and strongly recommended:
a series test, e.g. with Fibrinogen, using the same sample 16 times.
62
www.biolabo.fr
Operator‘s manual
Wash Station
General:
The wash station’s rinse cycle is software controlled by the probe sensor.
”Wash station over- Waste water drain tube Check tube path outside of the instrument (pos-
flow”. squeezed. sibly also a blocked drain port, e.g. after transport
of the instrument).
www.biolabo.fr 63
Operator‘s manual
Messages
If you are not able to resolve the errors or problems after consulting the error list and corresponding
descriptions, or if you do not have the necessary access rights, please contact your authorised service
technician.
Note: If a missing reagent is not replaced within 2 min., the sequence is interrupted and the tests,
which do not require this reagent, continue.
64
www.biolabo.fr
Operator‘s manual
EF23 - No clot
There was no coagulation detected within the measuring time 1/measuring time 2. Check the reagents;
check the pipetting system.
www.biolabo.fr 65
Operator‘s manual
66
www.biolabo.fr
Operator‘s manual
EF55 - Noisy
Detection of an irregular (noisy) reaction path during a coagulation test:
The following items could cause this problem:
1) Micro-clot
2) A piece of the rubber (or cap material) is in the cuvette (when working with “cap-piercing”).
EF 55 flags the measuring value without overwriting it. With an EDP (Host) connection: The data is not auto-
matically sent to the host. A warning message (EP68) is displayed in the message box. It is recommended to
re-check the result. If any doubts remain, repeat the measurement.
www.biolabo.fr 67
Operator‘s manual
68
www.biolabo.fr
Operator‘s manual
12 - [st] (Calibration)
Status message in (....): This process is started once m is pressed.
14 - [st] (Hardware)
Status message in (....): Always exit the hardware so the needle is in the wash station.
www.biolabo.fr 69
Operator‘s manual
70
www.biolabo.fr
Operator‘s manual
Caution: even blank characters are recognized as names for control plasma!
47 - [me] Down volume for cuvette is too great (max. X, Single / Double)
Wrong volume: maximum X. Check test setting in maintenance.
www.biolabo.fr 71
Operator‘s manual
72
www.biolabo.fr
Operator‘s manual
www.biolabo.fr 73
Operator‘s manual
105 - [me] Wash sequence X in last line of table not possible (must be without ‚C‘)
Check test setting in maintenance.
106 - [me] Please put X mm cup on the wash station and press <ENTER>
Serves as a daily check of the needle (see daily routine).
74
www.biolabo.fr
Operator‘s manual
108 - [me] Dilution X not possible: volume setting of plasma without dilution
Fully automatic calibration: dilutions 4:1, 3:1, 2:1 not possible. Requires plasma dilution in volume setting.
Check plasma pipetting in volume table or use other dilutions.
121 - [me] Menu entrance not permitted, check calibration data via Manual, Curve
Selecting of “Calibration, Curve” not possible. Create Calibration via “Calibration/ Manual / Curve”.
www.biolabo.fr 75
Operator‘s manual
76
www.biolabo.fr
Operator‘s manual
148 - [me] Please put test cup on the washstation and press <ENTER>
Message from “Hardware/Probe Check.”
151 - [me] Please remove test cup from washstation and press <F4>
Now, take the container off the wash station to check the volume and press o.
www.biolabo.fr 77
Operator‘s manual
167 - [me] Error: Thrombolyzer device and host device equal not possible
Check system parameter setting in maintenance. Use different devices for system and host.
168 - [me] Attention: probe drives to manual clean position, don‘t grab into work area! “Hardware/Probe
Clean manually”: be careful!
169 - [me] Attention: wipe probe with alcohol only from top to bottom!
“Hardware/Probe clean manually”.
78
www.biolabo.fr
Operator‘s manual
www.biolabo.fr 79
Operator‘s manual
186- [er] Calibration for test X is not OK, verify via Calibration, Manual, Curve
An attempt was made to start a routine with m . The standard curve for test X is not OK. The current
standard curve for this test must be validated.
80
www.biolabo.fr
Operator‘s manual
201 - [er] Rotor error:waiting of rotor X, answer from rotor X. Please exit!
208 - [er] Communication with SAMPLER not ok (X). Please exit the software
The communication with the specimen distributor was disturbed. The device must be shut down and then
restarted. (errorcode X).
www.biolabo.fr 81
Operator‘s manual
217 - [er] Unexpected ANSWER from X (Y)! Please restart the system
Communication with subcontroller X failed with errorcode Y.
218 - [er] Unexpected EVENT from X (Y)! Please restart the system
Communication with subcontroller X failed with errorcode Y.
220 - [er] File X not found. Please quit the system and call service
Missing parameter file X (e.g. sampler-par.txt): restore it or recreate it.
221 - [er] Error in file X. Please quit the system and call service
Error in parameter file (e.g. sampler-par.txt): restore it or recreate it.
Error in parameter file (e.g. interfaces.txt): check connection, logout, login.
226 - [er] Rotor reflector foil not found (X). Please check, press F4, SCAN
A rotor positioning error occurred during initializing/scanning procedure. Check position of reflector foil.
Press o. The rotor must now move to its “Home Position”. Should this error occur frequently, check via
adjusting rotor.
82
www.biolabo.fr
Operator‘s manual
www.biolabo.fr 83
Operator‘s manual
248 to ER 254:
All the following messages require the routine software to be quit and restarted!
After the restart, ensure that there are no more racks in the transport channel up to the point of rack
ejection on the measuring block. If this is the case, remove them manually if required!! Please check the
transport channel for foreign bodies.
84
www.biolabo.fr
Operator‘s manual
258 to ER 259:
All the following messages require the routine software to be quit and restarted !
After the restart, ensure that there are no more racks in the transport channel up to the point of rack
ejection on the measuring block. If this is the case, remove them manually if required!! Please check the
transport channel for foreign bodies.
www.biolabo.fr 85
Operator‘s manual
86
www.biolabo.fr
Operator‘s manual
www.biolabo.fr 87
Operator‘s manual
88
www.biolabo.fr
Operator‘s manual
List of accessories
Cat.-No. Qty. Description
055-350 1 Wash control cup
660-763 1 Cuvette magazine for 8-fold cuvettes
670-604 1 Protective cap for washing tank sensor
671-604OD 1 Water container 5L
671-603OD 1 Waste container 5L
690-443OD 1 Reagent rack
690-407 1 Sample rack
690-410 1 Reagent rack adaptor 30/22.5mm
690-411 1 Reagent rack adaptor 30/26mm
690-416 1 Reagent rack adaptor 22/12.5mm
691-243 1 Sample tray adaptor - 10 pcs
Consumables
Cat.-No. Description
054-320OD Cuvette racks - 2.320
050-111OD Cuvettes Hitachi - 250 pcs
CO0080 Cuvettes Hitachi - 1000 pcs.
050-610OD Reagent containers 25mm -100 pcs.
050-611OD Reagent containers 30mm -100 pcs.
055-200OD Clean Sticks
050-710OD Lids for 050-610 25mm - 25 pcs.
050-711OD Lids for 050-611 30mm - 25 pcs.
054-522OD Predilution Bars B - 25 for 1000 tests
MS8MM Magnetic Stirrer 8mm x 1.5mm - 10 pcs.
Glossary
Clot: Short for clotting
Coagulation: Latin for clotting
Display: Area displaying information
Incubation: Heating-up period prior to measuring
INR: I nternational Normalized Ratio
ISI: I nternational Sensitive Index
Reconstitute: To prepare, dissolve
www.biolabo.fr 89