MAR COR Competent, Consistent, & Compliant

700 Series Portable Medical Device


REVERSE OSMOSIS SYSTEM
650 to 2600 ml/min

Standard Features
• Specially designed for quiet operation.
• Advanced performance monitoring includes percent
rejection and a digital display of product water total
dissolved solids as mg/L. For added safety, an audible
alarm with adjustable set points is incorporated
into the TDS monitor. The monitoring and
instrumentation meets the standards recommended
by the Association for the Advancement of Medical
Instrumentation (AAMI).
• Panel-mounted pressure gauges continuously display
important prefilter, pump, and product pressures.
• Space-efficient design for easy portability.
• Chemically compatible with 1% Minncare® Cold
Sterilant disinfectant solution.
• Stainless steel RO pump and pressure vessels are
used for maximum strength, reliability, and longevity.
• Thin film RO membranes for high quality & long life. • Listed as certified by CSA to Canadian and U.S. medical
• Incoming water solenoid allows for easy changing electrical standards. The 700 Series RO has been
of the pre-filter element and system cleaning/ tested to and demonstrates compliance with CAN/
disinfecting without disconnecting the incoming CSA C22.2 No. 601.1-M90 (Safety of Medical Electrical
water supply. Equipment, Part I; General Requirements for Safety)
• Product water recirculation returns the excess and UL STD 2601-1 (Safety of Medical Electrical
product water into the feed stream, which reduces Equipment Part I; General Requirements for Safety).
stress on the RO membranes, conserves water,
inhibits bacterial growth, and lowers operating costs.
Optional Features
• Easy to expand design allows for additional
membranes to be added, increasing the capacity Mar Cor Purification offers a complete line of system
up to 2600 ml/min at 77°F (25°C); (see operating components and accessories including:
parameters). Mar Cor Purification strongly suggests • Replacement Prefilters
operating only one dialysis machine per 700 Series • 36” (91 cm) Tall Cabinet
RO because the life of the pump and membranes will • Pretreatment Cart
be significantly decreased. • AAMI Water Analysis
• FDA registered manufacturer meets stringent • Disinfectant Tank
medical device requirements. • Hardness & Total Chlorine Test Kits
• Carbon Tanks, Sediment Filters & Water Softeners
• High and Low pH Cleaners
700 Series RO Specifications
Model 701 702 703 704 704 (Cold)1
Production Rate
ml/min 656 1312 1968 2624 2060
L/hr 39 79 118 157 124
gallons/hr 10.4 20.8 31.2 41.6 32.7

Recovery Rate 17% 29% 38% 45% 45%


Dry Weight
lbs 55 73 79 85 90
kg 25 33 36 39 41
Power Required 115/230 V, 60/50 Hz, 8/5 amps

(Short Cabinet) Depth - 14” (36 cm) l Height - 26” (66 cm) l Width - 18” (46 cm)
Dimensions
(Tall Cabinet) Depth - 14” (36 cm) l Height - 36” (92 cm) l Width - 18” (46 cm)

NOTE: Initial product flow rate based upon membrane manufacturers’ specifications of pretreated feed water, 200 PSI (14 bars) applied pressure, 77°F
(25°C) incoming water temperature, 1,000 ppm (mg/L) NaCl, 10% recovery and 1 PSI (0.07 bars) product pressure. Membrane flux may vary ± 10%. Product
water flow rates specified @ 40oF (5oC). Operation of the 704 (Cold) Series RO unit at temperatures above 55oF (13oC) will result in shortened membrane
life. 1Cold water option available in tall cabinet only.

Feedwater Requirements

pH Operating: 5.5-10.5 Cleaning: 2-11


pH with Chloramines Operating: 5.5-8.5 Cleaning: 2-11
Temperature 35-90°F (1.7-32°C)
Water production decreases approximately 1.5% per 1°F (3% per 1°C) incoming water temperature drop.
Inlet Pressure (Dynamic) Minimum: 30 PSI (2 bars) Maximum: 100 PSI (7 bars)
Inlet Water Flow Minimum: 2 gpm (7.6 L/m)
Total Hardness 100 ppm (mg/L) (pH dependent)
Silt Density Index (SDI) <5

Note: Detailed analysis of the feed water will dictate the components, if any, needed in pre- or post-treatment of the 700 Series RO unit. This will
help assure a constant supply of water, meeting AAMI Dialysis Water Quality Standards.

Mar Cor Purification Mar Cor Purification Mar Cor Purification Mar Cor Purification Mar Cor Purification
4450 Township Line Road 14550 28th Avenue North 3250 Harvester Road - Unit 6 Sourethweg 11 1A International

MAR COR Skippack, PA 19474-1429


Tel: (484) 991-0220
Plymouth, MN 55447
Tel: (484) 991-0220
Burlington, ON L7N 3W9
Tel: (905) 639-7025
6422 PC Heerlen
The Netherlands
Business Park, #05-01
Singapore 609933
Toll Free: (800) 346-0365 Toll Free: (800) 633-3080 Toll Free: (800) 268-5035 Tel: (+31) 45 5471 471 Tel: (+65) 6227 9698
Fax: (484) 991-0230 Fax: (763) 210-3868 Fax: (905) 639-0425 Fax: (+31) 45 5429 695 Fax: (+65) 6225 6848

Visit www.mcpur.com for more information or call 1-800-633-3080 ©2014 Mar Cor Purification. All rights reserved. P/N:1238642 Rev.G

You might also like