The document discusses different types of magnetometry techniques used to measure magnetic properties of materials, including SQUID magnetometry, inductive pickup coils, vibrating sample magnetometry, pulsed field extraction magnetometry, torque magnetometry, Faraday force magnetometry, optical magnetometry, and survey magnetometry. Each technique is described in one or more paragraphs.

Uploaded by

jafasoh293
Copyright
© © All Rights Reserved
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
7 views

CMP53

The document discusses different types of magnetometry techniques used to measure magnetic properties of materials, including SQUID magnetometry, inductive pickup coils, vibrating sample magnetometry, pulsed field extraction magnetometry, torque magnetometry, Faraday force magnetometry, optical magnetometry, and survey magnetometry. Each technique is described in one or more paragraphs.

Uploaded by

jafasoh293
Copyright
© © All Rights Reserved
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
You are on page 1/ 2

tic moments are primarily classified as diamagnetic, paramagnetic, ferromagnetic, or antiferromagnetic (although the zoology of magnetic ordering also

includes ferrimagnetic, helimagnetic, toroidal, spin glass, etc.). Measuring the magnetization as a function of temperature and magnetic field can give
clues as to the type of magnetic ordering, as well as any phase transitions between different types of magnetic orders that occur at critical
temperatures or magnetic fields. This type of magnetometry measurement is very important to understand the magnetic properties of materials in
physics, chemistry, geophysics and geology, as well as sometimes biology.

SQUID (superconducting quantum interference device)[edit]


Main article: SQUID

SQUIDs are a type of magnetometer used both as survey and as laboratory magnetometers. SQUID magnetometry is an extremely sensitive absolute
magnetometry technique. However SQUIDs are noise sensitive, making them impractical as laboratory magnetometers in high DC magnetic fields, and
in pulsed magnets. Commercial SQUID magnetometers are available for temperatures between 300 mK and 400 K, and magnetic fields up to 7 tesla.

Inductive pickup coils[edit]


Main article: Inductive sensor

Inductive pickup coils (also referred as inductive sensor) measure the magnetic dipole moment of a material by detecting the current induced in a coil
due to the changing magnetic moment of the sample. The sample's magnetization can be changed by applying a small ac magnetic field (or a rapidly
changing dc field), as occurs in capacitor-driven pulsed magnets. These measurements require differentiating between the magnetic field produced by
the sample and that from the external applied field. Often a special arrangement of cancellation coils is used. For example, half of the pickup coil is
wound in one direction, and the other half in the other direction, and the sample is placed in only one half. The external uniform magnetic field is
detected by both halves of the coil, and since they are counter-wound, the external magnetic field produces no net signal.

VSM (vibrating-sample magnetometer)[edit]


Vibrating-sample magnetometers (VSMs) detect the dipole moment of a sample by mechanically vibrating the sample inside of an inductive pickup coil
or inside of a SQUID coil. Induced current or changing flux in the coil is measured. The vibration is typically created by a motor or a piezoelectric
actuator. Typically the VSM technique is about an order of magnitude less sensitive than SQUID magnetometry. VSMs can be combined with SQUIDs
to create a system that is more sensitive than either one alone. Heat due to the sample vibration can limit the base temperature of a VSM, typically to 2
kelvin. VSM is also impractical for measuring a fragile sample that is sensitive to rapid acceleration.

Pulsed-field extraction magnetometry[edit]


Pulsed-field extraction magnetometry is another method making use of pickup coils to measure magnetization. Unlike VSMs where the sample is
physically vibrated, in pulsed-field extraction magnetometry, the sample is secured and the external magnetic field is changed rapidly, for example in a
capacitor-driven magnet. One of multiple techniques must then be used to cancel out the external field from the field produced by the sample. These
include counterwound coils that cancel the external uniform field and background measurements with the sample removed from the coil.

Torque magnetometry[edit]
Magnetic torque magnetometry can be even more sensitive than SQUID magnetometry. However, magnetic torque magnetometry doesn't measure
magnetism directly as all the previously mentioned methods do. Magnetic torque magnetometry instead measures the torque τ acting on a sample's
magnetic moment μ as a result of a uniform magnetic field B, τ = μ × B. A torque is thus a measure of the sample's magnetic or shape anisotropy. In
some cases the sample's magnetization can be extracted from the measured torque. In other cases, the magnetic torque measurement is used to
detect magnetic phase transitions or quantum oscillations. The most common way to measure magnetic torque is to mount the sample on
a cantilever and measure the displacement via capacitance measurement between the cantilever and nearby fixed object, or by measuring
the piezoelectricity of the cantilever, or by optical interferometry off the surface of the cantilever.

Faraday force magnetometry[edit]


Faraday force magnetometry uses the fact that a spatial magnetic field gradient produces force that acts on a magnetized object, F = (M⋅∇)B. In
Faraday force magnetometry the force on the sample can be measured by a scale (hanging the sample from a sensitive balance), or by detecting the
displacement against a spring. Commonly a capacitive load cell or cantilever is used because of its sensitivity, size, and lack of mechanical parts.
Faraday force magnetometry is approximately one order of magnitude less sensitive than a SQUID. The biggest drawback to Faraday force
magnetometry is that it requires some means of not only producing a magnetic field, but also producing a magnetic field gradient. While this can be
accomplished by using a set of special pole faces, a much better result can be achieved by using set of gradient coils. A major advantage to Faraday
force magnetometry is that it is small and reasonably tolerant to noise, and thus can be implemented in a wide range of environments, including
a dilution refrigerator. Faraday force magnetometry can also be complicated by the presence of torque (see previous technique). This can be
circumvented by varying the gradient field independently of the applied DC field so the torque and the Faraday force contribution can be separated,
and/or by designing a Faraday force magnetometer that prevents the sample from being rotated.

Optical magnetometry[edit]
Optical magnetometry makes use of various optical techniques to measure magnetization. One such technique, Kerr magnetometry makes use of
the magneto-optic Kerr effect, or MOKE. In this technique, incident light is directed at the sample's surface. Light interacts with a magnetized surface
nonlinearly so the reflected light has an elliptical polarization, which is then measured by a detector. Another method of optical magnetometry
is Faraday rotation magnetometry. Faraday rotation magnetometry utilizes nonlinear magneto-optical rotation to measure a sample's magnetization. In
this method a Faraday modulating thin film is applied to the sample to be measured and a series of images are taken with a camera that senses the
polarization of the reflected light. To reduce noise, multiple pictures are then averaged together. One advantage to this method is that it allows mapping
of the magnetic characteristics over the surface of a sample. This can be especially useful when studying such things as the Meissner effect on
superconductors. Microfabricated optically pumped magnetometers (μOPMs) can be used to detect the origin of brain seizures more precisely and
generate less heat than currently available superconducting quantum interference devices, better known as SQUIDs.[13] The device works by using
polarized light to control the spin of rubidium atoms which can be used to measure and monitor the magnetic field.[14]
Survey magnetometers[edit]
Survey magnetometers can be divided into two basic types:

 Scalar magnetometers measure the total strength of the magnetic field to which they are subjected, but not its direction
 Vector magnetometers have the capability to measure the component of the magnetic field in a particular direction, relative to the spatial
orientation of the device.

A vector is a mathematical entity with both magnitude and direction. The Earth's magnetic field at a given point is a vector. A magnetic compass is
designed to give a horizontal bearing direction, whereas a vector magnetometer measures both the magnitude and direction of the total magnetic field.
Three orthogonal sensors are required to measure the components of the magnetic field in all three dimensions.

They are also rated as "absolute" if the strength of the field can be calibrated from their own known internal constants or "relative" if they need to be
calibrated by reference to a known field.