Standard Test Methods in Water': LLS%L
Standard Test Methods in Water': LLS%L
Standard Test Methods in Water': LLS%L
O - a d - 7s
4CrI~ Designation: D 1339 - 84
NOTICE: This standard has either been superseded and replaced by a new version or discontinued.
Contact ASTM International (www.astm.org) for the latest information.
1. Scope 3. Definitions
1.1 These test methods cover the volumetric 3.1 For definitions of terms used in these
determination of sulfite ion in water. The follow- methods refer to Definitions D 1129.
ing three methods are included:
4. Significance and Use
Range Section
Method A 0.1 to 6 mg/L 7 to 14 4.1 The determination of sulfite is usually
Na2S03 made on process waters, boiler waters, and other
Method B 6 mg/L Na2S03 15 to 22 waters that have been treated with sodium sulfite.
or higher Generally, in boiler applications, a cataIyzed so-
Method C 3 mg/L Na2S03 23 to 29
or higher lution of sodium sulfite is injected continuously
into the feedwater system for the purpose of
1.2 Method B may be used for samples con- removing dissolvedoxygen to prevent pitting and
taining less than 6 mg/L Na2S03with some loss excessive oxidation of metal components. In
of precision; in such applications the end point most boiler operations, the sulfite concentration
may be determined with starch indicator. is maintained above 20 mg/L in the boiler water.
1.3 This standard may involve hazardoiis ma- Sulfite injection is restricted to Iower-pressure
terials, operations, and equipment. This standard boilers (<lo00 psi, 6900 kPa) because of the
does not purport to address all of the safety prob- tendency for decomposition at higher tempera-
lems associated with its tise. It is the responsibil- tures. When exposed to the environment, dis-
ity of whoever uses this standard to constilt and charges containing sulfites wilI react with the
esrablish appropriate safety and health practices oxygen to form sulfates. Generally speaking, sul-
and determine the applicabilityof regulatorylimi- fites are not found in natural waters.
tations prior to use.
5. Purity of Reagents
5.1 Purity of Reagents-Reagent grade chem-
2. Applicable Documents icals shall be used in all tests. Unless otherwise
2.1 ASTM Standards: . indicated, it is intended that all reagents shall
D 1129 Definitions of Terms Relating to conform to the specifications of the Committee
Wate8 on Analytical Reagents of the American Chemi-
D 1192 Specifcation for Equipment for Sam-
piing Water and Steam'
D 1193 Specification for Reagent Wate8 'These test methods are under the jurisdiction of ASTM
Committee D-19 on Water and are the direct responsibility of
D 2777 Practice for Determination of Preci- SubcommitfeeD19.05 on Inorganic Constituents in Water.
sion and Bias of Methods of Committee Current edition approved Jan. 27, 1984. Published March
1984. Originally published as D 1339 - 54 T. Last previous
D- 19 on Wate? edition D 1339 - 78.
D 3370 Practices for Sampling Wate? Anniial B w k of ASTM Siandards, Vol 11.O 1.
treated with 5 drops of 1 N sulfuric acid (H2S04) lution prepared in this manner will remain chem-
and 1 mL of freshly prepared starch solution. i d l y stable for I year.
1 1.5 Potassium Iodide Solution (50 g/L)- NOTE2-If the indicatoris not prepared as specified
Dissolve 50 g of iodate-free Ki and 0.5 g of or a proprietarystarch indicator preparation is used the
sodium bicarbonate (NaHC03) in freshly boiled report of analysis shaii speii out the deviation.
and cooled water and dilute to 1 L.
12. Procedure
i 1.6 Sodiiim Thiosulfate Solution, Standard
(0.10 N)-Using a 1000-mL volumetric flask, €2.1 sampling- Arrange the sample tube in
dissolve 24.82 g of sodium thiosulfate (NazSzO3. a support so that it is vertical, with its upper
5Hz0) in approximately 800 mL of water that outIet free of hose connections and at a level
has just been boiled and cooled, by inverting the higher than the valve for adjustment of sample
flask at reguIar short intervals. Stabilize the so- flow. Connect the lower end to the sampling line
lution by dissolving in it, in the same manner, l by means of rubber tubing. The sampling line
g of sodium carbonate (Nazco3), and make up shall contain a suitable cooling coil if the water
to 1 L with boiled water. Standardize against being sampled is above temperature, in which
KzCrz07,as follows: case the sample shall be cooled to ambient-air
11.6.1 Dissolve 2 g of iodate-free Ki and 2 g temperature. If a cooling coil is used, the valve
of NaHC03 in 300 mL of water in a 500-mL for dooling-water adjustment shaü be at the inlet
Erlenmeyer flask and add HCl (sp gr I. 19) slowly, to the cooling coil, and the overkìow shall be to
while swirling the flask, until carbon dioxide a point of lower elevation. The valve for adjusting
(COZ)gas evolution ceases. Add 10 mL excess of the flow of sample shall be at the outlet from the
HCI, mix, and thendissolve 0.098 g of dried cooling coil. Adjust the sample to a rate that will
K2Crz07(see 1 1.2). Wash down the inside of the fill the tube in 40 to 60 s and continue this flow
flask with a small amount of water without agi- long enough to provide at least ten changes of
tating the flask, and allow to stand for 10 min. water in the sampling tube. if the sampling line
11.6.2 Titrate with the Na&O3 solution, us- is used intermittently, alíow a suitably longer
time for the first sample to ensure adequate
ing starch solution as the indicator, until the color
just changes from blue to the green color of the flushing of the sampling line and cooling coil.
chromic salt. Close the upper stopcock of the tube, and im-
1 1.6.3 Calculate the normality of the Na2S203 mediately close the lower sfopcock and remove
solution, as follows: the tubing connection. If the line being sampled
is under such high pressure that the sampling
N = (A149.94B) X loo0 tubes or connecting tubing may burst with the
where: water hammer when the stopcocks are closed,
N = normality of the NazSz03solution, throttle the sample flow with the control valve
A = grams of KzCr207used, and just before the sample is removed, but do not
B = millilitres of NazSz03solution required for shut off the sample flow completely.
the titration. 12.2 Fixing-Fill the three burets with HCI
1 I .I Sodium Thiosulfate Solution. Standard (If. I), KI solution, and 0.025 N K I 0 3 solution,
(0.005 N)-Transfer 25 mL of the nominally respectively. Hick the water from the upper nip-
O. 10 NNaZSzO3solution to a 500-mLvolumetric ple of the sample tube and fill the nipple to the
flask. Dilute to the mark with water and mix upper calibration mark with HCl. If a bubble is
completely. This solution shall be prepared not entrapped in the nipple within or below the
more than 12 to 15 h before using. reagent, remove it by probing with a clean plati-
i 1.8 Starch Indicator-Mate a paste of 6 g of num wire until the bubble rises to the surface.
arrowroot starch or soluble iodometric starch Open the lower stopcock and admit the reagent
with cold water. Pour the paste into 1 L of boiling by control with the upper stopcock until the
water. Then add 20 g of potassium hydroxide meniscus in the nipple coincides with the lower
(KOH), mix thoroughly, and allow to stand for calibration mark. If the sample contains more
2 h. Add 6 mL of glacial acetic acid (99.5 %). than 800 mg/L of NaOH, add 2 mL more of
Mix thoroughly and then add sufficient HCl (sp acid for each IO00 mg/L of alkalinity or fraction
gr 1.19) to adjust the pH value of the solution to thereof in excess of 800 m a . Close both stop-
4.0. Store in a glass-stoppered bottle. Starch so- cocks and ense both nipples with a fine stream
ofwater. Mix the contents by shaking or rotation, 14. Precision and Bias'
and flick out the excess water from each nipple. 14.1 The precision of this method may be
Invert the tube and fill what is now the upper expressed as follows:
nipple with KI solution to the 2-mL calibration
mark, Introduce the solution into the sample,
s, = 0.2
repeating the stopcock control, the rinsing of where:
both nipples, the shaking and rotating, and the S,= overall precision expressed in mg/L of
flicking of water from the nipples. Again invert NazSO3.
the tube and now introduce 2.0 mL of 0.025 N 14.2 The supporting data on this coilaborative
KIO3 solution into the sample foliowing the di- study do not include information on the number
rections given above for adding the reagents and of participating laboratories, water used, or sin-
rinsing the nipples. Once more, shake and rotate gle-operator precision or bias. It is the responsi-
the tube to mix the sample thoroughly. biiity of the analyst to assure the validity of the
12.3 Titration of Sample-Invert the sample method in a particular water.
tube and drain the sample into a clean 800-mL METHOD B
Giffin low-form beaker by opening both stop-
cocks. Do not rinse or blow into the sampling 15. Scope
tube, but shake the last drops from the nipple 15.1 This method covers the determination of
into the beaker. Add several drops of starch sulfite ion in wafer, especially boiler water and
indicator to indicate the approach of the end boiler feedwater, in which the sulfite concentra-
point. Insert the electrodes of the electrometric tion is 6 mg/L or higher.
dead-stop indicating apparatus into the sample, 15.2 The anaIyst shoald be aware that ade-
start the stirrer at low speed, and titrate the excess quate collaborativedata for precision and bias as
iodine chloride with 0.005 N Na2S2O3solution. required by Practice D 2777 is not provided. See
12.4 Determination ofBlank (Note 3)-Make Precision and Bias section for details.
a blank titration of 500 mL of water containing
no sulfife, to which have been added the three 16. Summary of Method
reagents that were added to the sample in 12.2. 16.1 Provision is made for excluding air from
NOTE3-Alternatively, if such tubes are available, the sample while it is being taken. Potassium
one sample of 500 mL and one sample of 250 mL are iodate is then added to form iodine chloride with
collected simultaneously, fixed as described in 12.2, potassium iodide and hydrochloric acid to oxi-
and titrated as described in 12.3. The difference in the dize the sulfite ion. The excess iodine chloride is
net amounts of 0.025 N Ki03solution used by the
sulfite content of the two samples will be the sulfite-ion titrated with sodium thiosulfate solution, using
equivalent of a 250-mL sample. an electrometric indicator to show the end point.
17. inferferences
13. Calculation 17.1 High results may be caused by the pres-
13.1 Calculate the concentrations of sulfite as ence of other reducing materials such as sulfide
milligrams per L of sodium sulfite (Na2S03),as and ferrous iron. Nitrite, if presenf, will oxidize
follows: sulfite when the sample is acidified.
Sulfite as Na2SO3, mg/L = 63000N (A - B)/S 18. Apparatus
where: 18.1 ErlenmeyerFlask, wide-mouth, 500-mL,
N = normality of the Na2S20ssolution, fitted with inlet, outlet, and siphon-leveling con-
A = millilitres of Na2SzO3solution needed to nections, as shown in Fig. 2.
titrate the blank, 18.2 Magnetic StirringApparatus, with swing-
B = millilitres of Na2Sz03 solution needed to ing support if electrometric end point-indicating
titrate the sample, and apparatus is used.
S = sampling-tube volume in millilitres minus 18.3 Dead-Stop End Point-Indicating Appa-
6.
13.2 To convert mg/L of Na2S03to mg/L '
Supporting data for these methods have been fded at ASTM
of sulfite ion, multiply by 0.635. Headquarters as R R D19-155.
-t02~~mm--I-2~+-3mm--lct02~3mm
Theánterican Societyfor Testingand Materials takes noposition respectingthe validity ofany patent rights asserted in connection
with any item mentioned in this standard. Users ofthìs standard are expressly advisedthat determination of the validityof any such
patent rights, and the risk of infringement of such rights, are entirely their own responsibility
This standard is subject to revision aí any time by the responsible technicdcommittee and must be reviewed evmyfive years and
i/ not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional
standards and should be addressed to ASTM Headquarters. Your comments will receive carefir1consideration aL a meeting of the
responsible technical committee, which you may attend. Ifyou feel that your comments have not received a fair hearing yotí should
make your views known to the ASTMîommittee on Standards, 1916 Race SI.. Philadelphia, Pa. 19103.
-.
- . &RS\ULJKWE\$670,QW O DOOULJKWVUHVHUYHG :HG-XQ87&
67 1
'RZQORDGHGSULQWHGE\
8QLYHUVLGDG&HQWURDPHULFDQDSXUVXDQWWR/LFHQVH$JUHHPHQW1RIXUWKHUUHSURGXFWLRQVDXWKRUL]HG
__