Sulfate Ion in Brackish Water, Seawater, and Brines: Standard Test Method For
Sulfate Ion in Brackish Water, Seawater, and Brines: Standard Test Method For
Sulfate Ion in Brackish Water, Seawater, and Brines: Standard Test Method For
Designation: D 4130 – 03
1
D 4130 – 03
7. Apparatus 100 mL with water. Adjust the temperature of these solutions to
7.1 Photometer—A filter photometer or a spectrophotom- 25 6 2°C. These solutions will contain 1.0, 2.0, 3.0, 4.0, 5.0,
eter for measurements between 400 to 450 nm, the preferable 6.0, 8.0, and 10.0 mg of sulfate ion, respectively.
wavelength being 425 nm. The cell for the instrument must 10.2 Follow the procedure as given in 11.6-11.8. Prepare a
have a light path of 20 6 2 mm and hold a volume of 25 mL. calibration curve showing sulfate ion content in milligrams on
Filter photometers, spectrophotometers, and photometric prac- the linear axis with the corresponding percent transmittance
tices prescribed in this test method shall conform to Practice (%T) reading of the photometer on the logarithmic axis of a
E 275. one cycle semilogarithmic graph paper (Note 2).
NOTE 2—The plot of concentration versus %T is not linear but shows a
8. Reagents slight s curvature. A separate calibration curve must be prepared for each
8.1 Purity of Reagents—Reagent grade chemicals shall be photometer and a new curve must be prepared if it is necessary to change
used in all tests. Unless otherwise indicated, it is intended that the photo cell, lamp, filter, or if any other alterations of the instrument or
reagents are made. Check the curve with each series of tests by running
all reagents shall conform to the specifications of the commit-
two or more solutions of known sulfate concentrations.
tee on Analytical Reagents of the American Chemical Society,
where such specifications are available. 5 Other grades may be 11. Procedure
used, providing it is first ascertained that the reagent is of
11.1 Filter the sample through a 0.45-µm membrane filter.
sufficiently high purity to permit its use without lessening the
This is necessary to remove nucleating particles.
accuracy of the determination.
11.2 Pipet a volume of filtered sample not to exceed 50 mL
8.2 Purity of Water—Unless otherwise indicated, reference
and 10 mg SO42- into a 100-mL graduated mixing cylinder. The
to water shall be understood to mean reagent water conforming
ionic strength (IS) of the sample when diluted to 100 must not
to Specification D 1193, Type I. Other reagent water types may
exceed 2.00 mol/L.
be used provided it is first ascertained that the water is of
11.3 Add 5 mL of glycerin-acid solution.
sufficiently high purity to permit its use without adversely
11.4 Add by a graduated pipet or a buret a volume of
affecting the precision and bias of the test method. Type III
sodium chloride solution (5 mol) calculated as follows:
water was specified at the time of round robin testing of this
test method. In addition, reagent water used for this test method mL NaCl 5 ~200 2 ~V 3 IS!/5
shall be sulfate-free.
8.3 Barium Chloride—Crystals of barium chloride V = volume of sample, and
(BaCL2·2H2O) screened to 20 to 70 mesh. IS = ionic strength of sample as calculated in Note 1, 4.1.
8.4 Glycerin-Acid Solution—Mix 250 mL of glycerin and 11.5 Dilute with water to 100 mL, mix well, and adjust the
50 mL of hydrochloric acid (HCl, sp gr 1.19) and dilute to 500 temperature to 25 6 2°C.
mL with water.
8.5 Sodium Chloride Solution (5 mol)—Dissolve 584.4 g of NOTE 3—The temperature of the solution in the mixing cylinder during
sodium chloride (NaCl) containing less than 0.001% SO4 in the development and measurement of the turbidity must be within 2°C of
the temperature of the standards when the calibration was performed. A
about 1800 mL of water and dilute to 2 L with water.
higher temperature will result in a positive error, a lower temperature in a
8.6 Sulfate Solution, Standard (1 mL = 1.00 mg SO4 = )— negative error.
Dissolve 1.479 g of anhydrous sodium sulfate, (Na2SO4), in
water and dilute to 1 L in a volumetric flask. 11.6 Pipet a 25-mL aliquot of the sample solution into a
sample cell and place it in the cell compartment. Set the
9. Sampling photometer to 100 % T (transmittance) with the wavelength set
9.1 Collect the sample in accordance with the applicable at 425 nm or blue filter in place.
ASTM standard as follows: Specification D 1192 or Practices 11.7 Add 0.3 6 0.01 g of BaCl2·2H2O crystals to the 75 mL
D 3370. remaining in the mixing cylinder, stopper, set a timer for 5 min,
9.2 Preserve the samples with high purity hydrochloric acid and mix for 30 s by inverting and righting the cylinder 15
to a pH of two or less immediately at the time of collection (2 times.
mL/L). NOTE 4—It is important the mixing be performed at a constant rate and
duplicated in all determinations.
10. Calibration
11.8 Just before 5 min has expired, check the blank setting.
10.1 Prepare standards by adding 1.0, 2.0, 3.0, 4.0, 5.0, 6.0,
Adjust to 100 %T if drifting has occurred. Replace the blank
8.0, and 10.0 mL of sulfate standard solution (1 mL = 1.00 mg
with the sample cell and measure turbidity at 5 min. If the % T
SO4 = ) to separate 100 mL graduated mixing cylinders. Add
is greater than 80 % or less than 30 % T, the determination
5.0 mL of glycerin-acid solution and 40.0 mL of sodium
with a smaller or larger sample volume providing the restric-
chloride solution (5 mol) to each of the cylinders and dilute to
tions in step 11.2 are not violated.
NOTE 5—The most reproducible section of the calibration curve is from
5
Reagent Chemicals, American Chemical Society Specifications, Am. Chemical 80 to 30 % T. Very low concentrations of sulfate ion can be determined by
Soc., Washington, DC. For suggestions on the testing of reagents not listed by the adding 3 mL of sulfate standard (1 mL = 1.00 mg SO42-) before diluting
American Chemical Society, see Analar Standards for Laboratory Chemicals, BDH to 100 mL in step 11.5 and then subtracting the 3 mg SO42- from the final
Ltd., Poole, Dorset, U.K., and the United States Pharmacopeia. results.
2
D 4130 – 03
TABLE 1 Determination of Precision and Bias TABLE 2 Composition of Synthetic Brine Samples
Amount Amount Statistically g/L
Added, Found, ST SO % Bias Significant (95 % Sample No.
1 2 3
mg/L mg/L confidence level)
NaCl 47.74 61.04 95.33
60.3 61.7 9.35 2.47 +2.32 no
CaCl2 27.22 40.85 54.47
86.3 83.9 6.98 2.45 −2.78 no
MgCl2 3.84 7.68 7.71
128.9 126.1 6.15 2.67 −2.17 no
SO4 = 0.0863 0.1289 0.0603
3
D 4130 – 03
14.6.1 To check for interferences in the specific matrix results must be qualified with an indication that they do not fall
being tested, perform a MS on at least one sample from each within the performance criteria of the test method.
batch by spiking an aliquot of the sample with a known NOTE 7—Acceptable spike recoveries are dependent on the concentra-
concentration of sulfate and taking it through the analytical tion of the component of interest. See Test Method D 5810 for additional
method. information.
14.6.2 The spike concentration plus the background concen-
tration of sulfate must not exceed the high calibration standard. 14.7 Duplicate
The spike must produce a concentration in the spiked sample 14.7.1 To check the precision of sample analyses, analyze a
that is 2 to 5 times the analyte concentration in the unspiked sample in duplicate with each batch. If the concentration of the
sample, or 10 to 50 times the detection limit of the test method, analyte is less than five times the detection limit for the analyte,
whichever is greater. a matrix spike duplicate (MSD) should be used.
14.6.3 Calculate the percent recovery of the spike (P) using 14.7.2 Calculate the standard deviation of the duplicate
the following formula: values and compare to the precision in the collaborative study
using an F test. Refer to 6.4.4 of Practice D 5847 for
P 5 100 [A~Vs 1 V! – BVs# / CV information on applying the F test.
where 14.7.3 If the result exceeds the precision limit, the batch
must be reanalyzed or the results must be qualified with an
indication that they do not fall within the performance criteria
A = analyte concentration (mg/L) in spiked sample, of the test method.
B = analyte concentration (mg/L) in unspiked sample,
C = concentration (mg/L) of analyte in spiking solution, 14.8 Independent Reference Material (IRM)
Vs = volume (mL) of sample used, and 14.8.1 In order to verify the quantitative value produced by
V = volume (mL) added with spike. the test method, analyze an Independent Reference Material
14.6.4 The percent recovery of the spike shall fall within the (IRM) submitted as a regular sample (if practical) to the
limits, based on the analyte concentration, listed in Test laboratory at least once per quarter. The concentration of the
Method D 5810, Table 1. If the percent recovery is not within IRM should be in the concentration mid-range for the method
these limits, a matrix interference may be present in the sample chosen. The value obtained must fall within the control limits
selected for spiking. Under these circumstances, one of the established by the laboratory.
following remedies must be employed: the matrix interference
must be removed, all samples in the batch must be analyzed by 15. Keywords
a test method not affected by the matrix interference, or the 15.1 brackish; brine; seawater; sulfate
SUMMARY OF CHANGES
Committee D19 has identified the location of selected changes to this standard since the last issue
(D 4130 – 99) that may impact the use of this standard.
(1) Section 8.2 was modified. (3) Section 14 (Quality Control) was added to the test method.
(2) Section 13.4 was added.
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org).