VW Multi Level Piezometer
VW Multi Level Piezometer
VW Multi Level Piezometer
VW Piezometer
52611199
This equipment should be installed, maintained, and operated by technically qualified personnel. Any errors
or omissions in data, or the interpretation of data, are not the responsibility of Durham Geo Slope Indicator
Company. The information herein is subject to change without notification.
This document contains information that is proprietary to DGSI and is subject to return upon request. It is
transmitted for the sole purpose of aiding the transaction of business between DGSI and the recipient. All
information, data, designs, and drawings contained herein are proprietary to and the property of DGSI, and
may not be reproduced or copied in any form, by photocopy or any other means, including disclosure to out-
side parties, directly or indirectly, without permission in writing from DGSI.
SLOPE INDICATOR
12123 Harbour Reach Drive
Mukilteo, Washington, USA, 98275
Tel: 425-493-6200 Fax: 425-493-6250
E-mail: solutions@slope.com
Website: www.slopeindicator.com
Contents
Introduction. . . . . . . . . . . . . . . . . . . . . . . .1
Installation . . . . . . . . . . . . . . . . . . . . . . . . .3
Taking Readings . . . . . . . . . . . . . . . . . . . .7
Data Reduction . . . . . . . . . . . . . . . . . . . . .8
Acceptance Test . . . . . . . . . . . . . . . . . . . .9
Diagnostics. . . . . . . . . . . . . . . . . . . . . . . 10
VW Piezometer in The drawing shows a cutaway of the multi-level Pipe with signal
Multi-Level Housing housing. Couplings at top and bottom accept cables inside.
placement pipe. Signal cable from each sensor runs Coupling at top
upwards through the placement pipe and housings. of housing
Grouting The multi-level system uses fully grouted bore- Cable out
holes. More information on this technique can be
found in technotes on the Slope Indicator website. Grout in
User-Supplied Placement pipe, couplings, and PVC cement. The multi-level housing accepts
Components 1.25 inch, schedule 40 PVC pipe.
Safety cable attached to the deepest piezometer to help retrieve the system if
there is trouble during installation.
Tape to protect ends of signal cables and hold cables together when pushing or
pulling them through the placement pipe.
A tubing cutter to cut off the placement pipe at the required height.
External grout pipe, if needed.
Different instrument, same concept: pipes are arranged for easy installation
2. Connect a safety line to the bottom piezometer. If you are using a flexible
external grout pipe, attach it to the bottom pipe.
3. Position the piezometer and pipe over the borehole. Remove the
filter (knurled ring) from the piezometer, fill the cavity with Filter
Grout Recipes Here are grout recipes. Read the notes below.
Grout Mix for Hard to Medium Soils
Read These Notes Mix cement with water first. Then add the bentonite to produce a grout with the
consistency of heavy cream. If you add bentonite first, the mix will be too thick
when cement is added. To thin the mix, you’ll have to add water. This increases
permeability, which is not desirable in this application. It also increases the risk
of a flash-set. So mix cement with water first, and then add bentonite.
There is no exact quantity of bentonite needed. The amount will vary with the
bentonite itself, the method of mixing, and the pH of the water. If you have a
Marsh funnel, the Marsh number of the liquid grout should be about 55 seconds.
Retired Readouts Manuals for retired readouts are available from the Slope Indicator website.
Data Loggers Instructions the VW MiniLogger can be found in the VW MiniLogger manual,
available on the Slope Indicator web site.
Instructions for Campbell Scientific data loggers can be found in the Tech Notes
section of the Slope Indicator web site.
Serial Number The serial number is found near the top of the page. Use the sensor serial number
to match each sensor with its calibration record.
Calibration Factors ABC Factors: These factors are used to convert Hz readings to units of pressure,
such as kPa or psi. If only small changes in water temperature are expected, ABC
factors, which offer no temperature correction, should work as well as TI factors.
TI Factors: These factors are also used to convert Hz readings to units of pressure,
but use temperature readings from the built-in thermistor to provide temperature
compensation. If the large changes in water temperature are expected, TI factors
should provide better results.
Summary of Results This table of recorded values shows the pressure applied by the calibration device,
the frequency output of the sensor, and the pressure calculated by applying the cal-
ibration factors. It also shows error, the difference between the applied pressure and
the calculated pressure, as a percent of the full range of the sensor.
Using ABC Factors Apply ABC factors to convert Hz readings to units of pressure
Pressure = A x Hz2 + B x Hz + C
Where:
Hz is the frequency reading in Hertz
A, B, and C are ABC factors on the sensor calibration record
Calculating To calculate changes in pressure, subtract the initial reading from the current read-
Changes ing. A positive value indicates increased pressure. A negative reading indicates
decreased pressure.
Pressure = Pressure current - Pressure initial
Operational 1. Connect the end of the signal cable to your readout and take a reading. Check
Check that the reading is stable. You now know that the piezometer is operational.
2. Note the serial number, Hz reading, the temperature, and the barometric
pressure at the time of the reading.
3. The Hz reading should be relatively close to the zero reading on the calibration
record. Higher altitudes or higher temperatures than those at calibration will
cause the reading to be lower than on the calibration record.
Zero Check 1. Let the piezometer adjust to ambient temperature. Do not handle the piezometer
body during the test.
2. Connect the signal cable to readout and obtain a frequency reading. Check that
you have obtained a repeatable reading.
3. Apply calibration factors to convert the frequency to a pressure reading in psi.
4. If your local elevation is above sea level, the pressure reading that you obtain will
most likely be negative because the sensor calibration is referenced to one stan-
dard atmosphere (sea level). To calculate a correction for your elevation, allow-
ing 0.5 psi for every 1,000 feet of elevation above sea level (1.15 mb per 10 m).
Add the correction to your reading. For example, if you are at 5,000 feet, you
would add a correction of 2.5 psi to the pressure reading. (See the chapter on
data reduction for more details). To make a more precise check, you would cor-
rect for the exact altitude, the barometric pressure, and the temperature.
5. The piezometer is working satisfactorily if the difference between the corrected
pressure and zero is within 2 percent of the full scale of the piezometer (2 psi for
a 100 psi piezometer).
No Reading Set your handheld multi meter to a low-ohm range (5k ohm).
Measure the resistance between the two VW wires (orange and white-and-
orange). A normal reading should be about 300 ohms. If the reading is very high
or infinite, the coil is damaged (or the cable is severed). If the reading is very low,
the cable may have been crushed and a short has developed.
Measure the resistance between the temperature sensor wires (blue and white).
Thermistors should read about 3000 ohms. RTDs should read about 2000 ohms.
If the reading is very high or infinite, the temperature device is damaged (or the
cable is severed). If the reading is very low, the cable may have been crushed and
a short has developed.
Unstable Reading Set your handheld multimeter to a high range (10 or 20 M ohm).
Measure the resistance between a VW wire and a Temp wire. The reading should
be infinite or out of range.
Measure the resistance between any of the colored wires and the drain (shield)
wire. The reading should be infinite or out of range.
Measure the resistance between the shield wires of two installed VW sensors.
Wires must be disconnected from data logger or terminal box to make this test.
The reading should be very high or infinite. A lower reading indicates the pres-
ence of a ground loop.
Other sources of unstable readings are electrical noise from nearby power lines,
radio transmitters, or motors. Also, over ranged or shocked instruments can
exhibit this problem.