SyncScan 2 Manual en
SyncScan 2 Manual en
SyncScan 2 Manual en
Contents
Preface............................................................................................................................. Ⅷ
I
Contents
8.6 Cannot Auto Identify the Phased Array Probe ................................................... 8-3
DCY2.781.SYNCSCAN 32PTSS/V1.0/A-E
VII
preface
Preface
The SyncScan ultrasonic flaw detector, with compact size and powerful functions, has
imaging features of ultrasonic flaw detection, Time of Flight Diffraction (TOFD), ultrasound
phased array (PA). It is suitable for ultrasonic testing on surface and internal defects of
Smart wizard, which can guide users to finish imaging setup easily and improve inspection
speed.
Multi-peripheral connectors: encoder input/output port, Ethernet port, USB port and video
Tips:
The system screenshots in this manual are generated by effective software version, which
Some of the features described in the manual are optional, and your system may not have
these features. If you have any questions, please contact the distributor.
To know system features and operation, as well as how to perform ultrasonic testing with
the system, please read this Operation Manual carefully before using the system.
Contact Information:
Website: http://www.siui.com
E-mail: siui@siui.com
VIII
Operation Cautions and Safety Guide
hazards any time. The Company shall not be liable for any damage or loss
Operation Manual.
b) The user should receive proper training on ultrasonic testing before using the
manual properly, and always operate the system by following the instructions
c) The user shall be responsible for the testing process, The Company has no
d) The Company shall not be liable for any loss of data stored in the system due
e) After purchasing the system, the user shall have full responsibility for the
f) Do not try to dismantle and repair the system unless specifically described in
g) Do not place the system in the place of moisture or near radiator, strong
h) When the system is not in use for a long time, it should be powered up at
system shall be certified in accordance with relevant IEC standards (e.g. IEC
60950 for data processing equipment and IEC 60601-1 Safety requirements
n) The Company shall not be responsible for the safety, reliability and
1) Damage or loss caused by the operation not following the Operation Manual
lightning.
3) Damage or loss caused by failure to meet the conditions of use for the
national standards.
X
Basic Information
Chapter 1
Basic Information
1.1 Appearance
1-1
Basic Information
(reserved, unavailable)
(reserved, unavailable)
1-2
Basic Information
1-3
Basic Information
1.2 Keys
For the key function description, please see Table 1-5.
key function and use with the knob to quickly adjust gain
step.
1-4
Basic Information
The system can be connected to one smart lithium-ion battery only, supporting hot swap.
a) When installing or removing the battery, open the rubber cover at the left of the
system, as shown in Fig. 1-5. Remove the screw fixing the battery compartment, and
If the battery is connected to the system, in power-on state, the upper left corner of the
【Note 1】: When battery power is very low, to avoid battery over-discharge, the
【Note 2】: When the battery is connected with the system, even if the system is
powered off, there is also current consumption. It is suggested to take out the
Charging the battery should be in a dry and ventilated place under normal temperature.
Operation:
1-5
Basic Information
2) Connect the adapter to the power supply. Now the adapter indicator (referred
normally.
3) Insert the adapter plug into the DC power socket on the system. The battery
charging indicator at the lower right corner of the system indicates the current
charging state, red for charging, and no indicator (power-off state) or green for
【Note】: Battery recharge can be done either when the system is powered on or off.
a) The capacity and service life of the rechargeable battery depend on whether the
battery is used properly. Please recharge the battery with the configured adapter.
1) When charging the battery through the adapter, use stable and reliable mains
2) The system battery can only be recharged with the configured adapter and the
3) Please recharge the system battery in a cool and ventilated room, otherwise the
4) If the system is not in use for a long time, the battery should be discharged and
5) During storage or transportation, the battery should maintain about half of its
capacity.
6) When system battery comes to the end of its service life, in order to avoid
1-6
Basic Information
environmental pollution, do not discard, break or crush the battery at will. Please
standards.
toxic chemicals.
8) Do not charge the battery near fire source or in extremely hot environment.
10) Keep the battery from moisture and never throw it into water.
11) Do not store the battery at the temperature higher than 60℃ or lower than -20℃.
12) Do not use the battery if it has peculiar smelling, heat, distortion, changed color
The adapter may supply power for the system or recharge the battery. After connecting
the adapter to AC power supply, if the adapter indicator is green, this means normal.
Connect the adapter output plug to the system adapter input socket for proper use.
【Note 1】:Connect the adapter by aligning the red mark on the adapter output plug
【Note 2】:Do not use the adapter on a plug-loosen socket. Be noted that the total
load of adapter plugging into the socket shall not exceed the rated load.
【Note 3】:Do not try to open the adapter if the adapter makes abnormal sound, or
In power up state, the Power Supply mark is displayed at the upper left of the screen. For
1-7
Basic Information
power.
1.4.1 Power On
Operation:
2) Press Power to boot up the system. The power/charging indicator at the lower
right of the system panel turns on, and the system enters work mode after
self-test.
【Note】: If any problem occurs during system power up, see Chapter 8.
Operation:
Long press Power for about 1 second, and the system will enter the shutdown state and
turn off.
【Note】: If any problem occurs during system power down, see Chapter 8.
1-8
Basic Information
are basically the same. The screen layouts are shown as Fig. 1-5 and Fig. 1-6. For
1-9
Basic Information
Menu: Main menu and the 1st level submenu are shown on the left of the
1
screen, below which are the 2nd and 3rd level submenus.
System info: For displaying info such as battery, adapter status or system
2
time.
Image area and scale: For displaying ultrasound image data and the
4
corresponding scales.
1.7 Scale
Subject to the specific system display view, the scales on the left and/or below the view
a) The scales for phased-array (PA) B-scan and A-scan waveforms are as shown in
1-10
Basic Information
Fig. 1-8 PA A-scan and B-scan image scales, with the definition in Table 1-8.
1 3 4
2 5
b) The scales for phased-array (PA) C-scan and D-scan in multiple views are as
1-11
Basic Information
c) The scales for UT A-scan are as shown in Fig. 1-10, with the definition in Table
1-10.
d) The scale for TOFD scan is as shown in Fig. 1-11, with the definition in Table
1-11.
1-12
Basic Information
Long touch the non-image area to pop up a “Full Screen” selection menu. Select it and the
image area (including the above data information bar and the below menu column) will be
1.8.2 Gate
In A-mode image, click the gate to move it to the desired area. Drag both sides of the gate
to change the width. When the start point and the end point overlap, click the gate and a
menu pops up for selecting “gate width” or “gate start”. The user can choose a menu to
adjust.
In B-mode image, click the gate to correspondingly adjust the gate start and the width of
A-mode image.
In phased-array mode, when the C-mode and D-mode images displayed, there is a
rectangle box appeared in the B-mode image. Click the four sides to change the box size.
Click the rectangle center, the cursor changes to cross cursor, then to move the box
1-13
Basic Information
position.
In B-mode image, click the scan line to move the A scan line position. When there are
1.8.4 Cautions
1) The touch screen is made of glass. Please do not exert great impact. Avoid touching
2) They system should be placed in the packing case with face up.
3) When the system is placed in the packing case, please DO NOT lay any other objects
4) Please use soft cloth with petroleum ethers to clean the touch screen surface. DO
NOT use any corrosive organic solvent to clean, e.g. industrial alcohol.
The typical appearance of phased array probes is as shown in Fig. 1-12. See the
1-14
Basic Information
1. Probe head
WARNING: When inserting or removing the probe, the system should be turned off
See Fig. 1-13 for the connection status between the phased array probe and the system.
As shown in Fig. 1-13, connect the phased array probe securely to the system. Then push
the locking mechanism to the right, and the phased array probe can be locked firmly on
the system.
WARNING Before pushing the locking mechanism to fix the probe, make sure that
the phased array probe is connected properly and securely to the system,
otherwise the phased array probe connector or the phased array probe socket on
1-15
Basic Information
Wedges are used with phased array probes, so as to achieve functions such as probe
protection, beam delay or beam conversion. See Fig. 1-14 for wedge types.
1 2
1. 0-degree wedge
2. Angle wedge
2) Use the hexagonal screwdriver and the screw configured with the system to
Make sure the installation direction of the probe and the wedge is correct. As indicated by
1) Ensure the mark face the bottom of the wedge. The wrong installation may
3) Use the hexagonal screwdriver and the screw configured with the system to
1-16
Basic Information
1-17
Ultrasonic Testing(UT)
Chapter 2
Ultrasonic Testing(UT)
【Note】: For the menus with the mark * in the menu structure below, tap the menu
-- -- Reject -- --
2-1
Ultrasonic Testing(UT)
ClearEnv. -- -- aStart* --
Clear -- -- aWidth* --
-- -- -- aThresh --
Chn.Sel BEA -- -- --
Frequency StartPt* -- -- --
Range* EndPt* -- -- --
MeasFreq -- -- -- --
WaveWidth -- -- -- --
PPSens -- -- -- --
Rectify -- -- -- --
2-2
Ultrasonic Testing(UT)
Display Display -- -- --
CurveType LineSel -- -- --
CurveType Curve1 -- -- --
LineSel Curve2 -- -- --
LockPara. Curve3 -- -- --
ClearCur Unit -- -- --
Adjust ClearCur -- -- --
Criterion -- -- -- --
Level -- -- -- --
2nd Submenu
CurveNum -- -- -- --
Curve1 -- -- -- --
Curve2 -- -- -- --
Curve3 -- -- -- --
Curve4 -- -- -- --
Curve5 -- -- -- --
Curve6 -- -- -- --
ThickBlock -- -- -- --
FlawBlock -- -- -- --
-- -- PRF* -- HCursor1
2-3
Ultrasonic Testing(UT)
-- -- -- -- Vcursor1
-- -- -- -- Hcursor2
-- -- -- -- Vcursor2
2-4
Ultrasonic Testing(UT)
Open Rename -- -- --
FileInfo -- -- -- --
Order -- -- -- --
Format -- -- -- --
VGA AutoGain -- -- --
LastFName AutoMode -- -- --
OutsideMode ScaleValue -- -- --
FanSw -- -- -- --
2-5
Ultrasonic Testing(UT)
The main menus are displayed in desktop menus at the center of the screen. To select
one main menu, tap the menu on the screen directly, or turn the knob to select it and press
Enter. Exit the desktop menu if the main menu is activated, with its submenu displayed
on the left and at the bottom of the screen.
There are three levels of submenus: 1st submenu, 2nd submenu, and 3rd submenu which
are shown as Fig. 2-1. All the menus can be selected by directly tapping the touch screen,
or activated by turning the knob to select and press Enter. The value of the 2nd submenu
【Note】: All menus can be selected by turning the knob and press Enter to activate,
2-6
Ultrasonic Testing(UT)
2.3.1 Basic
a) WorkPiece
b) Transmit
PulseWd: For setting transmit pulse width, the value of pulse width related to
TxMode.
2-7
Ultrasonic Testing(UT)
c) Receive
DisplayDly: For setting the image display of ultrasound pulse delay number.
【Note】: The system gain and gain step are displayed in the info bar at the
d) Probe
angle).
e) WaveSetup
FullScrn: For zooming the image area to full screen. Press Cancel to exit full
screen.
2-8
Ultrasonic Testing(UT)
2.3.2 Meas
a) GateA
b) GateB
c) GateC
d) Reading: For setting displayed content in the data info bar. For the meaning of
2-9
Ultrasonic Testing(UT)
e) Indn
2.3.3 Assistant
a) API 5UE
Dr: For setting the reference indicator for calibration (the depth of a groove
K: Correction factor.
b) AWS
CSC Type: For selecting the CSC type (outer diameter/ inner diameter)
2-10
Ultrasonic Testing(UT)
angle).
d) WeldSimu
DistanceE: For setting the distance from probe tip to the weld center
e) CrackHeight
system.
Height2: The echo height reflected by the second endpoint recorded by the
system.
Height2.
f) Spect.Anal
DspDly: For setting the screen display of ultrasound pulse delay number.
a) BEA
2.3.4 CurveCtrl
a) DAC
LineSel: Select one of the DAC curves, and the selected one is displayed in
white.
LockPara: If enabled, all the system setup parameters that affect DAC
setup.
ThickBlock: When the “GB/T 11345”, “GB/T 29712” or “NB/T 47013” DAC is
selected, the menu is adjustable, and the thickness step for the tested work
FlawBlock: When the "GB/T 11345" or "GB/T 29712" DAC is selected, the
menu is adjustable, and the work piece thickness step for the tested flaw
can be selected.
b) AVG
2-12
Ultrasonic Testing(UT)
LineSel: Select one of the DAC curves, and the selected one is displayed in
white.
2.3.5 BScan
a) Control
ImageDir: For setting the display direction of BScan images (DirHor /DirVer).
b) Setup
ScanStep: For setting the trigger time (time base mode) or trigger distance
ScanTime/ ScanDis: For setting the longest scan time and the distance.
c) Parameter
DspDly: For setting the pulse delay value of AScan echoes on screen
display.
Gain: For setting receiving amplifier gain. The higher the gain, the more
2-13
Ultrasonic Testing(UT)
d) Chn.Pos
e) Meas
2.3.6 AutoCalib
2.3.7 AngleCalib
2.3.8 DAC
2.3.9 AVG
a) Storage
1) Parameter file: for saving the setting parameters, and the UT parameter
2-14
Ultrasonic Testing(UT)
3) Record file: for saving the cine clips of BScan files, with the format
*.urec.
4) Report file: for saving system reports, with the format *.pdf.
Save: For saving the files according to the file type setting.
Order: For setting the file arrangement method of the current activated folder.
Format: For formatting the storage device inserted into the system.
Note: When inserting a new storage device for the first time, the
b) Encoder
c) WorkPiece
Edit: For re-editing the work piece parameters created by the user.
2-15
Ultrasonic Testing(UT)
Delete: For deleting a saved work piece name created by the user.
d) Prode
2.3.12 Setup
a) System
BRT: For setting screen brightness, which is displayed in a value. The higher
Default: For setting system parameters as default, and all the current
LastFName: For turning on naming the current file based on the last bit of
Note: if the last bit of the last file is z, the current file name is moved
forward and incremented. (e.g. the last file name is az, then the file
b) Reference
Unit: For setting the units for measurement and display values.
ColorEdit: For setting the image and line display color of the system.
2-16
Ultrasonic Testing(UT)
c) Info
ExportLog: If any error occurs in the system, select ExportLog to save the
relevant information and provide it to the engineer for analysis and trouble
shooting.
About: Display info such as system serial number and software version.
2.3.13 Mode
To switch to other imaging mode and enter the corresponding mode menu, with PA, UT,
The information bar of UT is as shown in 1 of Fig. 2-2, which is for parameter setting
2-17
Ultrasonic Testing(UT)
Symbol Meaning
ALA ALA: P means the alarm type is WaveIn alarm (positive); ALA:N means the
2.5 Measurement
The submenu Reading of the main menu Meas is for setting up the display of the data
information bar. Touch the data information bar to enter the Reading submenu, shown as
in Fig. 2-3. Tick in front of Reading1 ~ Reading 8, and select the desired displayed
readings in the Name list. The right column displayed the meaning of each selected
symbol.
2-18
Ultrasonic Testing(UT)
For the meaning of each symbol in the submenu Reading, see Table 2-4.
【Note】: "x" in Table 2-4 stands for Gate A/B/C, indicating the measurements in
Gate A/B/C.
Symbol Meaning
Horizontal distance from the probe flank to the echo measurement point
Rx
within the gate.
A Ind
D Rating
2-19
Ultrasonic Testing(UT)
With gate measurement, information such as amplitude, sound path and depth within the
Operation:
In Meas menu, adjust the submenus Start, Width and Thresh of Gate A, B or C to record
the echo, and the measurement results are shown in the data information bar real time.
【Note】: Track: When the gate is set “Track”, the echo of this tracked gate will be
the measurement reference point for the other gate measurement. Only one of the
There are 5 methods for gate MeasPoint: Peak, JFlank, Flank, GPeak and GFlank. The
values for Sound Path and Amplitude are different subject to the methods.
1) Peak: To calculate sound path and amplitude based on the peak within the gate
range.
2) JFlank: To calculate sound path from the first intersection with the gate within
the gate range, based on which calculate the amplitude by searching for the
3) Flank: To calculate sound path from the first intersection with the gate, and to
calculate the amplitude based on the peak within the gate range.
4) GPeak: To calculate sound path and amplitude based on the peak exceeding
5) GFlank: To calculate sound path from the first intersection with the gate
exceeding the gate thresh within the gate range, and calculate amplitude
If Peak is selected for MeasPoint, Sound Path is S3, and Amplitude is H2.
2-20
Ultrasonic Testing(UT)
2.5.4 Alarm
Alarm Includes visual and audible alarm. Two alarm methods are available: WaveIn
(positive) and WaveLost (negative). The alarm indicators (from left to right)at the upper
right of the system panel correspond to Gate A, B and C. If the gate alarm is valid, its
indicator is red. If Buzzer is enabled, the system will give the sound alarm at the interval of
0.5 seconds.
a) WaveIn Alarm (positive): Alarm when the echo amplitude within the gate is higher
b) WaveLost Alarm (negative): Alarm when the echo amplitude within the gate is
A quantitative method for defect depth according to API (American Petroleum Institute).
Operation:
2) Select the submenu Dr, and input the reference indicator for calibration (the
3) Place the probe on the standard test block and ensure the coupling is good.
Moving the probe to bring on screen a dynamic echo envelope required by API
2-21
Ultrasonic Testing(UT)
4) Select the menu aStart, adjust gate start to record the echo envelope peak value.
Tap the submenu K, and the system will calculate the factor K.
5) Place the probe on the standard test block and ensure the coupling is good.
Moving the probe to bring on screen a dynamic echo envelope required by API
5UE.
6) Select the menu of aStart, adjust gate start to record the echo envelope peak
value. Tap the submenu Di, and the system will calculate the factor Di.
2.6.2 AWS
Welding level is calculated as per AWS (American Welding Society) structural welding
code (D1.1) and bridge welding code (D1.5). Before operation, select the appropriate
Operation:
2) Place the probe on the test point of the standard test block, and ensure good
coupling. Move the probe back and forth to get the highest echo, and adjust the
gain to keep the echo amplitude within 20%~100%, select aStart menu, adjust
2-22
Ultrasonic Testing(UT)
3) Set the menu Operation to Record, and the submenu B_Ref will record the
4) In AWS screen, Place the probe on the tested work piece and ensure good
coupling. Follow Step 2) to operate and acquire the max. defect echo of the work
piece. The system will automatically record the current echo amplitude that
Indication Level, and the record result will be displayed in the extended submenu
formula. And the results are displayed in the extended submenu CAtt and
2-23
Ultrasonic Testing(UT)
【Note 1】: All calculations are based on AWS D1.1-2006 and AWS D1.5-2002.
【Note 2】: When recording B Reference Level in step 3), the values of A, C, D are
reset to 0.
For depth and horizontal distance correction when testing circumference with an angle
Operation:
2) Select the submenu OutDiam, and input the outer diameter of the tested
workpiece.
3) Select the submenu Thickness, and input the thickness of the tested
workpiece.
4) Select the submenu CSC Type, and set the correction type of circumference
testing:
workpiece
workpiece
2-24
Ultrasonic Testing(UT)
5) Select the submenu Angle or K_Value, and input the angle or KValue of the
probe in use.
6) Select the submenu X_Value extended submenu, and input the X Value of the
probe in use.
2.6.4 WeldSimu
WeldSimu is to simulate the real weld shape and use the dynamic beam tracking function
to assist the use to quickly determine the location of defects in the weld.
Operation:
2) If a waveform with very high amplitude in the effective AScan image area is
detected, now fix the probe still, and use a ruler to measure DistanceE from
4) After the parameters are set, tap the touch screen to move the gate (or use the
menu setup for gate Start, Width and Thresh) to record the echo, now the red
dot on the weld simulation figure refers to the defect position in the workpiece.
2-25
Ultrasonic Testing(UT)
2.6.5 CrackH
To measure crack height with angle probe endpoint peak echo method.
When incident waves enter the crack tip, some of them are reflected along their original
path, which is called endpoint peak echoes. The endpoint peak echo method is to
calculate crack height by measuring the sound path of endpoint peak echo produced
Operation:
3) Tap the touch screen to shift the gate (or select gate Start) to record the first
endpoint reflection echo. Press Height1 menu to record the depth of the echo.
4) Tap the touch screen to shift the gate (or select gate Start) to record the
second endpoint reflection echo. Press Height2 menu to record the depth of
the echo. Now the system will automatically calculate the recorded depth
difference between H1 and H2, with the result displayed in Result menu.
2.6.6 Spect.Anal
signals, and the frequency related measurements can be further obtained by measuring
the spectrogram.
2-26
Ultrasonic Testing(UT)
Operation:
1) Record the echo in Basic screen and enter Spect.Anal. The system shall
expand according to the peak of the amplitude to obtain the desired Range and
2.6.7 BEA
The BEA (Backwall Echo Attenuator) function is for image reject and attenuation of local
echoes.
Operation:
When the BEA function is enabled, adjust StartPt and EndPt to determine the image area
2.7 CurveCtrl
2.7.1 DAC
DAC control menu is for adjustment and setup of the created DAC curves, including curve
display On/Off, curve manual adjustment, curve type selection, and conversion between
2-27
Ultrasonic Testing(UT)
1) Click Adjust menu, and a green circle appears on the DAC curve points.
2) Select the adjust point, and press Enter. The selected DAC curve point
indication circle turns from green to red, which means for adjustment. Turn
2.7.2 AVG
AVG control menu is for adjustment and setup of the created AVG curves, including curve
2-28
Ultrasonic Testing(UT)
2.8 BScan
BScan is to display type A echo in imaging mode, so as to achieve more intuitive test
2.8.1 Control
After entering the BScan interface, set up the scan parameters in the Control menu, and
tap Start to begin scanning. When the scan is finished, use Browse to view the results,
2-29
Ultrasonic Testing(UT)
Save: to save the scan result. Tap it to open the save screen, and the scan result
Browse: to browse the scan result, with Manual, AutoForward and AutoBack for
selection.
ImageDir: For setting the encoder moving direction, with DirVer or DirHor for
2.8.2 Setup
The user sets the scan mode and parameters according to the actual scanning
2.8.3 Parameter
The user may set up parameters such as Range, DisplayDly and Gain as required. As
2-30
Ultrasonic Testing(UT)
2.8.4 Chn.Pos
With dual-channel scan, and when the trigger mode is Encoder, the Chn1_Pos and
Chn2_Pos menus are activated for setting the positions of Channel 1 and Channel 2.
2.8.5 Measurement
Cursor1 and Cursor2 are for setting the start and end points of measurement.
Operation steps:
1) Adjust the position of Cursor1 or Cursor2 (or through the touch screen) to find the
2-31
Ultrasonic Testing(UT)
2) Adjust the parameters of Gate A (or through the touch screen) to record the echo
with the gate, then record the echo Sa, Da, Ra values.
3) Then, in the B image area, adjust Cursor1 position (or through the touch screen)
to bring it at the start of the defect; adjust Cursor2 position (or through the touch
4) Read the horizontal and vertical value difference of Cursor1 and Cursor2, and
2.8.6 Reading
The image measurement result can be found from the reading. The 8 data sets at the top
right of the screen are for the current measurement data, from left to right and top to
bottom, matching Reading 1~8 at the screen bottom. Tap Reading menu, or the data
select either material velocity calibration or probe zero calibration, or both at the same
time.
b) Zero calibration: Calculate probe zero by using the difference of the echo propagation
time difference of one reflector with known sound path, and the echo time measured
by the system.
See Fig. 2-17 for the calibration block as an example. The sound path of the first reflector
is 50mm, and the sound path of the second reflector is 100mm. The user can also use a
2-32
Ultrasonic Testing(UT)
Take the angle probe (frequency 2.5MHz, angle 68.2°) as an example to introduce how to
Operation:
1) Select AutoCalib in the UT imaging screen, and enter the Auto Calibration
2) Set the sound path of Ref 1 and Ref 2 of calibration block according to the test
block parameters (set 50mm and 100mm here) and adjust Option to Vel & Zero.
3) Set up each parameter of the system according to the connected probe, as shown in
Fig. 2-19.
2-33
Ultrasonic Testing(UT)
4) Place the probe on the calibration block as shown in Fig. 2-17 and get the echo
of Ref1 as shown in Fig. 2-20. Adjust aStart and aWidth to bring the gate record
the echo, move the probe back and forth to get the highest echo of the reference
5) Move the probe to get the echo of Ref2 as shown in Fig. 2-21. Refer to the
operation in step 4) to get the highest echo of Ref2 and press Record key to
2-34
Ultrasonic Testing(UT)
6) Now the values of MtVel and ProbeDly are updated. Press Finish and the
screen pops up an input box to input the exact X_Value, and finish the auto
calibration.
2-35
Ultrasonic Testing(UT)
AngleCalib is for measuring the angle of an angle probe with a reflector (e.g. side drilled
【Note】: Before angle calibration, probe zero should be calibrated first, otherwise
A test block with a side drilled hole depth 30mm, diameter 3mm is used in the example,
see Fig. 2-23. Here we use an angle probe (frequency 2.5MHz) as an example to
Operation:
1) Select AngleCalib in the UT imaging screen, and enter the Auto Calibration
screen as shown in Fig. 2-24. Adjust ReflDep and ReflApet according to the test
2-36
Ultrasonic Testing(UT)
3) Place the probe on the test block as shown in Fig. 2-23. If velocity calibration is
not done, tap1/2 menu and set MtType and Mtvel according to the probe and the
test block in use. Otherwise, adjust aStart and aWidth to bring the echo go
through the center of the gate. Move the probe back and forth till the position of
the red circle on screen does not update. Now the angle information of the probe
is shown at the upper right corner of the screen as shown in Fig. 2-26.
2-37
Ultrasonic Testing(UT)
【Note】: The red circle on the echo means the highest echo position.
Otherwise, press Reset to repeat step 3) to have calibration till the result is good.
DAC wizard is for guiding the user to make DAC curves based on the detection
requirement.
Operation:
1) Select DAC in the in the UT screen, and enter DAC wizard as shown in Fig. 2-27.
2-38
Ultrasonic Testing(UT)
2) Adjust the parameters aStart and aWidth, bring the echo go through the gate A
center. Set the gain to make the echo amplitude bigger than 20% and smaller
than 100%.
3) Press Record menu to finish setting a DAC curve record point. 1 is shown in the
4) Repeat steps 2)~3) to continuously increase DAC curve record points to meet
the requirement. In the process, if any wrong record point is selected, tap
DeletePt, rotate the knob to select the point to be deleted and press Enter key
2-39
Ultrasonic Testing(UT)
to delete a record point. Tap ClearCur to delete the whole DAC curve.
5) Set up evaluation reference lines. The user can choose whether to add reference
lines depending on the testing technique. The CurveNum menu is for setting the
number of reference lines, with 6 curves at maximum for setup, as shown in Fig.
2-29.
6) Curve repair: This step is to correct the reference lines set up by the user.
AVG wizard is for guiding the user to make AVG curves based on the detection
requirement.
【Note】: Before using AVG Wizard, Velocity and Zero should be calibrated first.
Operation:
【Note】: A transverse angle probe, with 5MHz, 10×10mm, angle 70°is used in this
example.
1) Select AVG in the UT screen, and enter AVG wizard as shown in Fig. 2-30.
2-40
Ultrasonic Testing(UT)
the length of both probe sides in AVGLenA and AVGLenB, as well as input
2) Input parameters according to the type of the test block in use, as shown in Fig.
2-31.
Select the submenu AVGBaseType, adjust it to the type of the reference test
selected, the system automatically ignores the value entered in the sub-menu
FBH equivalent aperture of the 3 created AVG curves by selecting the sub-menu
2-41
Ultrasonic Testing(UT)
aStart, and adjust gate start to record the gate, adjust the gain to bring the echo
amplitude to about 80%. Press Record to record the echo. Now the curves are
created automatically.
【Note】: If the echo amplitude of the reference test block or the workpiece
does not comply with the theoretical value, Sur_Comp (Surface Compensation)
echo serves as the reference. Since BW echo amplitude usually cannot reach
2-42
Ultrasonic Testing(UT)
the theoretical difference with the BW hole echo amplitude of the same
2-43
Ultrasonic TOFD
Chapter 3
Ultrasonic TOFD
the mark * in the menu structure below, tap the menu repeatedly to toggle between
ImageDir EncodeSel -- -- --
Browse ScanStep* -- -- --
Chn.Sel StartPos -- -- --
-- ScanSpeed -- -- --
-- PRF -- -- --
-- Average -- -- --
3-1
Ultrasonic TOFD
-- LW Filter -- --
-- Meas Recover -- --
-- MeasRes -- -- --
-- CurveLen -- -- --
3-2
Ultrasonic TOFD
Open Rename -- -- --
FileInfo -- -- -- --
Order -- -- -- --
Format -- -- -- --
VGA AutoGain -- -- --
LastFName AutoMode -- -- --
OutsideMode ScaleValue -- -- --
FanSw -- -- -- --
3-3
Ultrasonic TOFD
2.2.
3.3.1 Scan
a) Control
ImageDir: For setting the display direction of TOFD images (DirHior /DirVer).
scanned images.
b) Setup
screen.
scanning.
Average: During the average scan number, the scan lines will finish
3-4
Ultrasonic TOFD
scanning and send to the system to form echoes. The higher the scan
number is, the lower speed that the echoes are formed.
c) Chn.Pos
3.3.2 Analyse
a) Basic
ImageDir: For setting the image display direction, with DirHor and DirVer for
selection.
b) ImageMeas
cursor.
c) Strn.Filtr.
3-5
Ultrasonic TOFD
d) ImageProc
scanned images.
3.3.4 CalcPCS
3.3.5 CalibZero
3.3.6 UTPara
3.3.7 TimeWnd
3-6
Ultrasonic TOFD
3.3.8 Mng
3.3.9 Setup
3.3.10 Mode
tested workpiece. There are eight steps: parameter and workpiece selection, setting the
number of channels, calculating PCS, zero calibration, ultrasonic parameter setting, time
Operation:
1) Select ScanWiz in the UT imaging desktop menu, and enter the wizard, as shown in
Fig. 3-1. This step is for ParamOpen and editing workpiece. When the setup is
3-7
Ultrasonic TOFD
2) Set up channel number, as shown in Fig. 3-2. The number of scan channels can be
set to 1 or 2, and the channel type can be selected as TOFD, Angle or Creeping.
3) “CalcPCS”, see Fig. 3-3Fig. 3-3 Calculating PCS values. Set up probe and wedge
rule. The system will simulate the beam coverage, and calculate the theoretic test
blind area according to the setup to assist making testing process. When the setup is
3-8
Ultrasonic TOFD
Note: Parameter "PulseVol" cannot exceed 300V while the probe frequency is
above 7MHz.
4) Zero Calibration.
i. After installing the TOFD probe on the wedge, contact the two probes as shown
in Fig. 3-4. Adjust the parameters such as Gain and Range to enable the probe
ii. Adjust GateStart and GateWidth to enable the main peak of the pulse echo go
through the gate center. Move the probe wedge back and forth till the highest
3-9
Ultrasonic TOFD
iii. Keep the probe position unchanged, and press CalibZero to complete the probe
zero calibration.
iv. Use a ruler to measure the d value shown in Fig. 3 4, divide this value by 2 and
v. Subtract the theoretical PCS value by twice of the calibrated X_Value, which is
the actual required probe spacing, Adjust the probe spacing again. After setup,
Set up UT parameters, shown as in Fig. 3-6 Set up the parameter of Voltage, Pulse Wd
and PRF, etc. Tap Next to go to step 5) when the setup is finished.
3-10
Ultrasonic TOFD
5) Set up time window, shown as in Fig. 3-7. This step is for setting channel time window
i. Adjust the pulse shift to find the lateral wave, and change the gain to adjust
the amplitude; adjust “StartWnd.” and “EndWnd.” to finish the setting of time
window.
ii. Setup the value of the calibration reference point, adjust the calibration line
to the reference position, and tap CalibPCS to finish depth calibration. If the
3-11
Ultrasonic TOFD
calibrated PCS value is quite different from the theoretical value, select "
iii. After the calibration is finished, adjust the “Line 1” and “Line 2” position,
which correspond to MD1 and MD2, and MT is for the workpiece thickness,
and you may verify if the time window settings and depth calibration is
【Note 1】: Setup the reference point. When there are N channels in the system, set
Table. 3-2.
Channel Reference Point
1 0 means LW
2~N-1 test block aperture depth
N workpiece thickness, i.e. BW
【Note 2】: Refer to test standards for “StartWnd.” and “EndWnd.” setup.
【Note 3】: MD1: depth value of measure line 1; MD2: depth value of measure line 2;
6) Setup “Encoder”. In this step, the user may select, add, calibrate, test, delete, rename
or change scan step, as shown in Fig. 3-8. See for details. Tap Next to go to step 7)
3-12
Ultrasonic TOFD
8) After finishing steps 1) to 8), tap Finish, the screen prompts whether to save the
parameters or not. Choose Yes to bring up the saving setting parameter screen, and
the user can store the setting parameters in the SD card on the system or an external
USB disk. Choose No to exit the wizard and enter the Scan screen to start scanning.
Note 1: If no SD card or USB disk is connected to the system, tap "Finish" and
the screen will not prompt to "save parameters?", and all the measured
Note 2: The parameters acquired from the testing can only be saved in the SD
card or USB disk, but not saved directly in the system, otherwise the image
measurement will be affected. As each image data from testing is not saved,
the data will be lost and not for browse, and the cursor of image measurement
cannot be adjusted.
3-13
Ultrasonic TOFD
3.6 Scan
3.6.1 Control
Set up the scan parameters in Control menu, and tap Start to scan. When the scan is
finished, tap Save to save the result, and review the result through Browse. Adjust
3.6.2 Setup
In TOFD scan screen, the user may set up parameters such as PRF, EncodeSel and
3-14
Ultrasonic TOFD
3.7 Analyse
3.7.1 Basic
3-15
Ultrasonic TOFD
Take a TOFD image with thickness 40mm as an example, here is a demo of how to
perform measurement:
1) Set up CurveLen, change the menu value, and set up the length of the arc fitting
cursor.
2) Select the menu LW, turn the knob to make the LW measured cursor coincide
3) Select HCursor1 and turn the knob to set up the location; Select VCursor1 and
turn the knob to set up the location. Now the setup of measurement start is
finished.
4) Select HCursor2 and turn the knob to set up the location; Select VCursor2 and
turn the knob to set up the location. Now the setup of measurement end is
3-16
Ultrasonic TOFD
6) Tap MeasRes to see the measurement result, as shown in Fig. 3-15 TOFD
Measurement Result.
The straight and filter for TOFD scan images, as shown in Fig. 3-16.
3-17
Ultrasonic TOFD
a) Straight
1) Select HCursor1, and turn the knob to set up its location; Select VCursor1,
and turn the knob to set up its location. Now the horizontal and vertical setup
2) Select HCursor2, and turn the knob to set up its location; Select VCursor2,
and turn the knob to set up its location. Now the horizontal and vertical setup
3) Tap the menu Straight, and a yellow solid line appears on the image. Turn
the knob to bring the yellow solid line in the area set up in step 1) and 2), as
3-18
Ultrasonic TOFD
4) Press Enter on the keypad to finish straighten processing. Tap the menu
b) Filter
Filter is to remove lateral wave or bottom wave on images. Here is an example for
1) Select HCursor1, and turn the knob to set up its location; Select VCursor1,
and turn the knob to set up its location. Now the horizontal and vertical setup
2) Select HCursor2, and turn the knob to set up its location; Select VCursor2,
and turn the knob to set up its location. Now the horizontal and vertical setup
3) Tap the menu Filter, and a yellow solid line appears on the image. Turn the
knob to bring the yellow solid line in the area set up in step 1) and 2), as
3-19
Ultrasonic TOFD
4) Press Enter on the keypad to finish filter processing. Tap the menu
3-20
Ultrasonic TOFD
3.7.4.1 Gain
Gain: To perform gain processing for the scanned images, which enables easy analysis of
3.7.4.2 Contrast
Contrast is the mapping relation of the amplitude of ultrasonic echo images and the
display grayscale. Adjust Contrast menu to process the scanned images for easy
3.7.4.3 Zoom
To perform zoom processing for the scanned 2D images. Here are the operation steps:
1) Tap Zoom menu, and the TOFD scan image area appears a red sample box.
Tap and hold the center of the sample box, move the finger to change the
position of the sample box; Tap and hold the right of the sample box to
adjust the width of the sample box; Tap and hold the bottom of the sample
box to adjust the height of the sample box. Adjust the position and size to
3.7.4.4 SAFT
SAFT: SAFT (Synthetic Aperture Focusing Technique), for post processing of TOFD
appears on TOFD scan images due to probe beam divergence angle. When TOFD scan
is finished, set SAFT menu on, and the SAFT processing can be performed on the
scanned image.
3.7.4.5 ColorBar
Enter the color bar setup screen for TOFD image. Select the desired color bar, to show
the " Current " area. Tap Ok and you can change the color of ColorBar, or Cancel to exit
3-21
Ultrasonic TOFD
3-22
Ultrasonic Phased Array(PA)
Chapter 4
Ultrasonic Phased Array(PA)
mark * in the menu structure below, tap the menu repeatedly to toggle between different
step values.
-- SampleFrec ClearAllGrp -- --
Delay* Focus* -- -- --
MtType FocusMode -- -- --
EchoType -- -- -- --
4-1
Ultrasonic Phased Array(PA)
MtVel * -- -- -- --
-- GateAlarm -- Exchange
-- -- Buzzer -- DefaultPos
-- -- GateMode -- FullScreen
-- -- DisER* -- --
-- -- Coupling -- --
-- -- AutoSearch -- --
GateLogic VCursor1 -- -- --
GateStart* HCursor1 -- -- --
2nd Submenu
GateWidth* VCursor2 -- -- --
GateThres HCursor2 -- -- --
A Line HorZero* -- -- --
4-2
Ultrasonic Phased Array(PA)
A Group Meas -- -- --
Range* MeasRes -- -- --
MeasPoint -- -- -- --
GateAlarm -- -- -- --
Buzzer -- -- -- --
GateMode -- -- -- --
DisER* -- -- -- --
Coupling -- -- -- --
AutoSearch -- -- -- --
DSP6 -- Curve2 -- --
2nd Submenu
Layout -- Curve3 -- --
SrcArea -- Curve4 -- --
DesArea -- Curve5 -- --
Exchange -- Curve6 -- --
DefaultPos -- -- -- --
FullScrn -- -- -- --
4-3
Ultrasonic Phased Array(PA)
Back Back -- -- --
4-4
Ultrasonic Phased Array(PA)
Next Next -- -- --
Goto Finish -- -- --
Finish -- -- -- --
Open Rename -- -- --
FileInfo -- -- -- --
Order -- -- -- --
Format
VGA AutoGain -- -- --
LastFName AutoMode -- -- --
Outside ScaleValue -- -- --
FanSW -- -- -- --
4-5
Ultrasonic Phased Array(PA)
same as 2.2.
4.3.1 Basic
a) Transmit
Voltage: For setting transmit pulse voltage. The higher the voltage, the
TxEnergy: For setting transmit pulse current. With the same voltage, the
larger the energy, the higher the sound pressure emitted by the probe.
AutoMax mode, the system will set PRF to the maximum automatically
b) Receive
Gain: For setting receiving amplifier gain. The higher the gain, the more
4-6
Ultrasonic Phased Array(PA)
【Tip】: The system gain and gain step is defaulted to be shown in the info bar at the
c) Calibrate
d) WorkPiece
4.3.2 Scan
a) Probe
ProbeSel: For selecting the probe connected to the system, with auto and
manual modes. In auto mode, the system will identify the probe connected
manual mode, the user selects the probe type from the probe list manually,
MtType: For setting material type of the tested object. If the material type is
4-7
Ultrasonic Phased Array(PA)
not included in the list or the built-in material velocity does not meet the
user’s requirement, the user may choose user-defined and input the real
b) FocusRule
StartAng: For setting scan start angle. In linear scan, StartAng is the scan
AngSpace: For setting scan interval angle, available in sector scan only.
c) Aperture
EndEle: For setting the end element of scanning, available in linear scan
only.
EleStep: For setting element step between adjacent scan lines in linear
scan.
Aperture: For setting the excited element number at one time for scanning.
【Tip】: The adjustable maximum EleDiam is subject to the probe model, and the
【Note】: Apply the changes made to the parameters of Basic, FocusRule and
4-8
Ultrasonic Phased Array(PA)
4.3.3 AdvScan
a) Control
Browse: To browse the scan record, with manual and auto modes for
selection.
Reset: To clear the value of the current area and restart the scan.
C/DRect: For using the keyboard knob to fine adjust the C/D sample
rectangle.
b) Setup
ScanMeth.: For selecting the scan mode, including scan mode and data
source, etc.
c) Gate
4-9
Ultrasonic Phased Array(PA)
AGroup: For switching to another A-scan select line in group scan “3A+B”
mode.
d) ImageMeas
e) DspMode
window.
adjusted.
adjusted.
Exchange: For exchanging the window location selected for SrcArea and
DesArea.
4-10
Ultrasonic Phased Array(PA)
4.3.4 Meas
a) Gate
AGroup: For switching to another A-scan select line in group scan “3A+B”
mode.
b) ImageMeas
4-11
Ultrasonic Phased Array(PA)
c) Reading: For setting displayed content in the data info bar. For the meaning of
4.3.5 Display
a) DspMode
window.
adjusted.
adjusted.
Exchange: For exchanging the window location selected for SrcArea and
DesArea.
b) indn
FocusInd: For setting the display of focus position indication line on/off.
c) TCG
d) BEA
4-12
Ultrasonic Phased Array(PA)
Enter the ScanWiz screen, and set up the imaging parameters and display mode specific
Operation:
4.3.7 VelCal
Enter the VelCal screen, and complete velocity calibration of work piece.
4.3.8 DelayCal
Enter the DelayCal screen, and complete ultrasonic delay calibration of probe or wedge.
4.3.9 SenCal
4.3.10 TCG_Cal
4.3.11 App
4.3.12 Mng
a) Storage
4-13
Ultrasonic Phased Array(PA)
b) Encoder
c) Probe
d) Wedge
e) WorkPiece
4.3.13 Setup
4.3.14 Mode
Definition
Linear scan, also called electronic scan or E scan, is a scanning method with a fixed
number of elements scanning along the probe arrangement direction in the same focus
rule until the whole probe scanning is finished. That is to add a select window with a fixed
size on the phased-array probe, and the elements in that window are T/R elements. When
interval to select the T/R elements for the next transmission/ reception without changing
the focus rule, and it works in the same way till the whole probe selection is completed,
which is similar to manual straight-line scanning with a conventional probe. The scanning
principle is shown in Fig. 4-1 Linear Scan Diagram, and the image display as in Fig. 4-2.
4-14
Ultrasonic Phased Array(PA)
(Every three elements in one group, and scan at an interval of one element)
Operation
1) Set the transmit pulse width according to the probe frequency, and set the
transmit voltage and transmit energy on the basis of the detection depth.
corresponding wedge model (wedge model is labeled on the wedge). All the
WedgeSel.
3) Select the test material type. Manually enter the material velocity according to the
4) Set EchoType according to the ultrasonic incidence angle. Select LWave when
not using the wedge; when using the wedge for scanning, wave type should be
selected according to the wedge usage. Select LWave when using a 0° wedge,
4-15
Ultrasonic Phased Array(PA)
and select TWave when using a transverse wave wedge (such as 55° wedge).
5) Set the scan type, select FocusRule -> Type, and adjust it to Linear.
6) Set the steer angle, adjust StartAng submenu to set the scan angle.
7) Set the number of elements for scanning. Select the Aperture submenu to set
8) Calculation: Select Calculate of the Scan main menu and the settings take effect.
【Tip】: The values in the menus Depth, Delay take effect immediately after user
setup without pressing Calculate menu. Other scan parameter setup takes
4.4.2.1 Definition
Sector scan, also called S scan, is a scanning method with a fixed number of elements
That is to add a select window with a fixed size on the phased-array probe. The elements
in that window are T/R elements. When transmitting/receiving of elements in the window
is finished, change the focus rule without moving the window, perform the next
transmission/ reception, and it works in the same way until the whole scanning is
completed. The scanning principle is shown in Fig. 4-3, and the image display as in Fig.
4-4.
4-16
Ultrasonic Phased Array(PA)
4.4.2.2 Operation
1) Set PulseWd according to the probe frequency, and set Voltage and TxEnergy
2) When scanning with a wedge, select WedgeSel submenu to select the wedge
model (which is labeled on the wedge). All the default wedge parameters and
3) Select the test material type; manually enter the velocity according to actual
4) Set EchoType according to the ultrasonic incidence angle, select LWave when
not using the wedge; when using the wedge for scanning, wave type should be
selected according to the wedge usage. Select LWave when using a 0° wedge,
and select TWave when using a transverse wave wedge (such as 55° wedge).
5) Set the scan type, select FocusRule -> Type, and adjust it to Sector.
6) Set the scanning angle, adjusting StartAng and EndAng to determine and set
7) Set the number of elements for scanning. Select the Aperture submenu to set
8) Calculation: Select Calculate of the Scan main menu and the settings take effect.
4-17
Ultrasonic Phased Array(PA)
On top of the A-scan waveform area and B-scan image, there is an info bar, which is for
scan imaging information. See Table 4-2 for the specific definition.
Symbol Meaning
1-1 A-1 One current group only, and the image is the first A-scan image in
The number behind “L” refers to the number of the scan line(e.g.
C-X Stands for the display mode of abscissa in A mode image. S stands
ER Stands for the actual distance between wedge flank and the weld
center.
1-1 B One current group only, and the image mode is B-mode.
A Stands for scan angle. In linear scan mode, only display the scan
B-Scan AP Stands for aperture. In linear scan mode, the start element(S), the
Area element step(P), the end element(E), and the element aperture (N)
are displayed.
4-18
Ultrasonic Phased Array(PA)
Symbol Meaning
ER Stands for the actual distance between wedge flank and the weld
center.
internal time base or system external encoder trigger. The user may select multi-view
such as A, B, C, D or their combination for image display and full course recording.
Operation:
1) Select AdvScan from the desktop menu. The screen is as displayed in Fig. 4-5.
2) Set up display mode: Select the menu DspMode to set up the advanced scan
3) Set up gate mode: Select the menu Gate to set up the gate number and mode for
the image.
4) Set up scan mode: Select the menu Setup to set up the trigger mode of scan,
4-19
Ultrasonic Phased Array(PA)
the right), Acquire (if “Single Line” is selected as Scanning Path, only “C/D Rect”
user can select the gate displayed position in Acquire-Gate list ), Data Src
CScan image. Choose “Depth” to form a depth CScan image), Scan Mode
are listed on the right column; if “TimeBase” is selected, the column is not
displayed).
5) If B-scan image display is set in DspMode, the user may set the position of
sample box for the B-scan image: in B-scan view, press and hold the center of the
sample box (the rectangular box in orange lines) till a red cross appears at the
center of the sample box (as shown in Fig. 4-7). Continue to hold on the sample
box center, shift it to a proper location and then release it to finish the adjustment
4-20
Ultrasonic Phased Array(PA)
6) Similar to step 5), set up the size of sample box: In B-scan view, press and hold
the bottom right corner of the sample box (the rectangular box in orange lines) to
have dual arrow indication lines appear on the right and below the sample box (as
shown in Fig. 4-8). Continue to hold on the bottom right of the sample box, change
the sample box to a proper size and then release it to finish the adjustment of
7) Start and stop scanning: After finishing the setup in steps 2)~7), select the menu
Control as shown in Fig. 4-9. Press Start to start scanning. After the scan starts,
4-21
Ultrasonic Phased Array(PA)
8) Browse the scan: Select the menu Browse, then select Manual, AutoForward or
AutoBack to browse the scanned image. If Manual is selected, turn the knob on
9) Refresh [B]: when there is [B] mode in display setup, press Refresh [B] to refresh
the image.
10) NextArea: in grating scan mode, press NextArea to go to the next area
scanning.
11) Dynamic rendering: After finishing one scan, by setting up the position and size
of the sample box again, the C/D scan image can be recreated dynamically
without rescanning. After finishing the scan, adjust the position and size of the
sample box, select the menu Rebuild if the adjustment is finished, and the system
will reconstruct the C or D san image according to the new setup of the position
12) Save the scan file: The result of advanced scan can only be saved to the SD
card or a USB storage. Before scanning, connect the SD card or USB storage
device to the system. As it prompts the identification of the storage (as the symbol
indicated in the information bar at the upper left of the screen). When the scan is
finished, select Save to save the scan file to the external storage.
【Note】: If one scan is finished without being saved, and you have started a
4-22
Ultrasonic Phased Array(PA)
new advanced scan or exited the advanced scan menu, then the scan data will
13) Recall a saved file: The files saved in the external storage can be recalled in
14) Tap FullScreen in DspMode or long press the screen and press FullScreen
menu to full screen display and operate the image. The data information bar is
displayed above of the image for instant check. Press Cancel to exit the full
screen.
4.6 Measurement
The PA image measurement includes: gate measurement and image measurement.
The submenu Reading of the main menu Meas is for setting up the display of the data
information bar. Touch the data information bar to enter the Reading submenu. Tick in
front of Reading1 ~ Reading 8, and select the desired displayed readings in the Name list.
In Table 4-3, x refers to gate A/B/C, which is the measurement values of gate A/B/C.
Symbol Meaning
Horizontal distance from the probe flank to the echo measurement point
Rx
within the gate.
4-23
Ultrasonic Phased Array(PA)
Symbol Meaning
Horizontal distance between the MeaLen1 position and the probe central
PC1
scale line.
RC1 Horizontal distance between the MeaLen1 position and the probe flank.
Horizontal distance between the MeaLen2 position and the probe central
PC2
scale line.
RC2 Horizontal distance between the MeaLen2 position and the probe flank.
AScanAng The selected A-scan line angle displayed in the information bar.
Similar to gate measurement for UT, information such as amplitude, sound path and depth
In Meas menu, adjust the submenus Start, Width and Thresh to record the echo and
acquire the measurements of echo, and the measurement results are shown in Reading.
According to the image display mode, select Depth or Sound Path in Logic submenu.
To turn on the “AutoSrch.” function, the gate will automatically search the highest scan line
4-24
Ultrasonic Phased Array(PA)
Image measurement is for measuring B-mode, C-mode and D-mode image area to
acquire the horizontal distance of the corresponding points in the image and the relative
2) Select HCursor1 and VCursor1 menus, turn the knob or tap the touch screen to
3) The operation of HCursor2 and VCursor2 is the same as that of HCursor1 and
VCursor1.
After finish the measurement, the result is saved in MeasRes menu, and the
measurement menu of each image mode is displayed in different list, shown as in Fig.
4-10.
Tap Report to create a test report, shown as in Fig. 4-11. In the report window, user can
edit the report information, preview the content and save the report in the system internal
disk.
4-25
Ultrasonic Phased Array(PA)
supports maximum six image windows at the same time. The definition of each display
ultrasound signal.
B Side view of workpiece: to display the actual echo position after volume
4-26
Ultrasonic Phased Array(PA)
of beam or coverage.
beam or coverage.
[B] [B] mode is defined as a 2D image created on the time or distance axis
by the same A-mode echo, with the amplitude displayed in color code.
The horizontal axis is depth or sound path, and the vertical axis is
distance or time.
4.8 BEA
BEA(Backwall Echo Attenuator)function is for restrain and attenuate some part of the
image echo.
Operation:
Turn on BEA, and adjust the StartPt and EndPt for the desired image area. Use
4-27
Ultrasonic Phased Array(PA)
an example. Press Next or Back to continue setup after finish every setup. Press Goto to
Operation:
1) Tap Para.Open to open the scan wizard parameter for use, which is saved in the
system.
2) Group and Workpiece, shown as in Fig. 4-13 Workpiece Setup Window. It is for
setting scan group numbers, edit or select the MtType. The wizard supports for
4-28
Ultrasonic Phased Array(PA)
3) Select probe and wedge, as shown in Fig. 4-14. Socket menu is used to select
the probe switcher for the connected probe. which is required when using the
Note:
a) ET: stands for the theoretical distance between the probe/wedge flank
b) ER: stands for the actual distance between the probe/wedge flank
4-29
Ultrasonic Phased Array(PA)
4) Set up scan type. Set up scan type and display depth based on testing needs,
5) Set up focal law. Set up imaging parameters based on testing needs, as shown
in Fig. 4-16. The system will simulate the ultrasound transmission in workpiece
4-30
Ultrasonic Phased Array(PA)
6) Set up DspMode, to setup the display mode of the group, as shown in Fig. 4-17
The position of the image display window can be changed by SrcArea and
DesArea setup.
7) Finish setup of scan wizard. Tap Finish to complete setup of scan wizard, and
4-31
Ultrasonic Phased Array(PA)
4.10 Calibration
As shown in Fig. 4-18 and Fig. 4-19, there are four reference lines for various calibration
areas.
a) Yellow dotted line: The reference error line, indicating reference error of
calibration.
b) Red dotted line: The sound path or depth location envelope line of echo
c) Green dotted line: The amplitude envelope line of scan line echo maximum
amplitude.
d) White dotted line: The sound path or depth location envelope line of echo live
Generally there are two methods to perform velocity calibration: sound path calibration
and depth calibration. The probe placement and test block appearance for these two
calibration methods are shown in Fig. 4-20, Fig. 4-21 and Fig. 4-22.
The sound path calibration and depth calibration method 1 are mostly used in angle
4-32
Ultrasonic Phased Array(PA)
phased-array probe calibration, and the depth calibration method 2 is usually for 0-degree
Generally, sound path calibration uses two test blocks with known curvature radius to
4-33
Ultrasonic Phased Array(PA)
Operation:
2) Select VelCal from the desktop menu and enter the velocity calibration
screen, as shown in Fig. 4-24. Input the radius values 1 and 2 (Fig. 4-23
Sound Path Calibration) in the menus RefPoint1 and RefPoint2 (50mm and
3) Adjust ALine to select the AScan line (or angle) for calibration. Adjust
the center of the gate and adjust GateThres to the required thresh, and then
4-34
Ultrasonic Phased Array(PA)
adjust Gain to make the echo amplitude to about 80%. Move the probe back
4) Tap the menu Record and change the record value to 1, which means the
5) Repeat step 3)~4) to bring the A-scan echo of RefPoint2 go through the
center of the gate, and acquire the RefPoint2 maximum echo. See Fig.
4-26.
4-35
Ultrasonic Phased Array(PA)
6) Tap the menu Record, which means the echo location of RefPoint2 is
MtVel menu. C:V is shown in the B-scan image window info, which means
calibration, and the screen prompts “Do you want to enter the next
and return to the desktop menu. If the calibrated value of MtVel is not
images.
Operation:
4-36
Ultrasonic Phased Array(PA)
2) Select VelCal from the desktop menu and enter the velocity calibration
screen, as shown in Fig. 4-29, input the depth values 1 and 2 (Fig. 4-28) in
the menus RefPoint1 and RefPoint2 (20mm and 30mm in this example)
3) Adjust ALine to select the AScan line (or angle) for calibration. Adjust
the center of the gate and adjust GateThres to the required thresh, and then
adjust Gain to make the echo amplitude to about 80%. Move the probe back
and forth to enable the echo of RefPoint1 coincide with A scan seletion line
4-37
Ultrasonic Phased Array(PA)
4) Tap the menu Record and change the record value to 1, which means the
5) Repeat steps 3)~4) to bring the A-scan echo of RefPoint2 go through the
center of the gate, and then enable RefPoint2 achieve the maximum echo
4-38
Ultrasonic Phased Array(PA)
6) Tap Record menu, which means the echo position of RefPoint2 is recorded.
When the calculation is finished, and the calibrated velocity is shown in the
MtVel menu. C:V is shown in the B-scan image window info, which means
calibration, and the screen prompts “Do you want to enter the next
and return to the desktop menu. If the calibrated value of MtVel is not
images.
Operation:
4-39
Ultrasonic Phased Array(PA)
2) Select VelCal from the desktop menu and enter the velocity calibration
screen, as shown in Fig. 4-34, input the depth values 1 and 2 (Fig. 4-33) in
the menus RefPoint1 and RefPoint2 (25mm and 40mm in this example)
through the center of the gate, and then adjust GateThres to the required
4) Tap the menu Record and change the record value to 1, which means the
4-40
Ultrasonic Phased Array(PA)
RefPoint2 display in the screen. Repeat steps 3)~4) to enable the gate to
bring the A-scan echo of RefPoint2 go through the center of the gate and to
4-36.
6) Tap Record menu, which means the echo position of RefPoint2 is recorded.
When the calculation is finished, and the calibrated velocity is shown in the
MtVel menu. C:V is shown in the B-scan image window info, which means
4-41
Ultrasonic Phased Array(PA)
calibration, and the screen prompts “Do you want to enter the next
and return to the desktop menu. If the calibrated value of MtVel is not
In multiple group mode, when one group of velocity calibration is finished, tap Next to go
Delay calibration consists of probe delay and wedge delay. Generally, there are two
methods: sound path calibration and depth calibration, which can be found in Fig. 4-38,
Fig. 4-39 and Fig. 4-40. Sound path calibration and depth calibration method 1 are usually
used to calibrate angle phased-array probe, and depth calibration method 2 is mostly
4-42
Ultrasonic Phased Array(PA)
Generally, the sound path calibration uses a test block with known curvature radius to
Operation:
4-43
Ultrasonic Phased Array(PA)
2) Select DelayCal from the desktop menu and enter the delay calibration
screen, as shown in Fig. 4-42, input the radius value of the reference point
(Fig. 4-41) in the menu RefPoint, and set GateMode to Sound Path.
RefError is the visual delay error for each scan line, and the default value is
RefPoint, adjust GateThres to the required thresh, and then adjust Gain to
make the echo amplitude to about 80%. Move the probe back and forth to
enable each scan line to achieve the maximum echo amplitude, that is the
echo amplitude envelope (green dotted line) is not updated, and get the
highest echo wave position envelope curve (red dotted line), as shown in Fig.
4-44
Ultrasonic Phased Array(PA)
Press StartAng and EndAng to decide the start line and end line for
4) Tap CAL to finish delay calibration. Move the probe back and forth to
calibration result is correct, the red position envelope line should be among
the yellow dotted lines, as shown in Fig. 4-44. C:D is shown in the B-scan
4-45
Ultrasonic Phased Array(PA)
calibration, and the screen prompts “Do you want to enter the next
and return to the desktop menu. If the delay calibration is not accepted,
In multiple group mode, when one group of delay calibration is finished, tap Next to go to
Operation:
2) Select DelayCal from the desktop menu and enter the delay calibration
screen, as shown in Fig. 4-46, input the depth value of the reference point
(Fig. 4-45) in the menu RefPoint, and set GateMode to Depth. RefError is
the visual error for each scan line delay, and the default value is 5%. When
adjust GateThresh to the required thresh, and then adjust Gain to make the
echo amplitude to about 80%. Move the probe back and forth to enable each
scan line to achieve the maximum echo amplitude, that is the echo
amplitude envelope (green dotted line) is not updated, and get the highest
echo wave position envelope curve (red dotted line), as shown in Fig. 4-47.
reproduce.
Press StartAng and EndAng to decide the start line and end line for
4-47
Ultrasonic Phased Array(PA)
4) Tap CAL to finish delay calibration. Move the probe back and forth to
calibration result is correct, the red position envelope line should be among
the yellow dotted lines, as shown in Fig. 4-48. C:D is shown in the B-scan
calibration, and the screen prompts “Do you want to enter the next
and return to the desktop menu. If the delay calibration is not accepted,
In multiple group mode, when one group of delay calibration is finished, tap Next to go to
Operation:
2) Select DelayCal from the desktop menu and enter the delay calibration
screen, as shown in Fig. 4-50, input the depth value of the reference point
(Fig. 4-49) in the menu RefPoint, and set GateMode to Depth. RefError is
the visual error for each scan line delay, and the default value is 5%. When
adjust GateThresh to the required thresh, and then adjust Gain to make the
echo amplitude to about 80%. Move the probe back and forth to enable each
scan line to achieve the maximum echo amplitude, that is the echo
amplitude envelope (green dotted line) is not updated, and get the highest
echo wave position envelope curve (red dotted line), as shown in Fig. 4-51.
reproduce. Press StartAng and EndAng to decide the start line and end
4) Tap CAL to finish delay calibration. Move the probe back and forth to
reproduce position envelope of each scan line echo. if the calibration result
is correct, the red position envelope line should be among the yellow dotted
lines, as shown in Fig. 4-52. C:D is shown in the B-scan image window info,
4-50
Ultrasonic Phased Array(PA)
calibration, and the screen prompts “Do you want to enter the next
and return to the desktop menu. If the delay calibration is not accepted,
In multiple group mode, when one group of delay calibration is finished, tap Next to go to
Generally, the sensitivity calibration uses three kinds of test blocks, which are shown in
Fig. 4-53, Fig. 4-54 and Fig. 4-55. The test blocks in Fig. 4-53 and Fig. 4-54 are usually for
sensitivity calibration on an angled phased-array probe, and the test block in Fig. 4-55 is
Operation:
2) Select SenCal from the desktop menu and enter the sensitivity calibration screen, as
shown in Fig. 4-57, RefAmp is the echo amplitude after the calibration is completed,
and the default value is 80%. RefError is the visual error of the echo amplitude after
the calibration is completed, the default value is 5%. Adjust Depth to make the echo
of reference arc appear at the center of the B-scan image, and then set GateMode to
Sound Path. When the setup is finished, select Next to go to step 3).
4-52
Ultrasonic Phased Array(PA)
3) Adjust GateStart and GateWidth to record B-scan echo. Adjust GateThres to the
required thresh, and then adjust Gain to make the echo amplitude to about 80%.
Move the probe back and forth to make each scan line reach the maximum echo
amplitude. The maximum echo amplitude envelope curve (green dotted line) is
shown in Fig. 4-58. In calibration, the white dotted line is the live envelope line of
echo amplitude for each scan line. If the echo amplitude envelope is not acceptable,
press ClearEnv to reproduce it. Press StartAng and EndAng to decide the start line
4) Tap CAL to finish sensitivity calibration. Move the probe back and forth to reproduce
each scan line echo amplitude envelope. If the calibration result is correct, the green
amplitude envelope should among the yellow dotted lines as shown in Fig. 4-59. C:S
is shown in the B-scan image window info, which means sensitivity calibration is
finished.
5) If the sensitivity calibration result is accepted, select Finish to finish the calibration,
and the screen prompts “Do you want to enter the next calibration TCG_Cal?”. Select
Yes to go to TCG_Cal or No to exit calibration and return to the desktop menu. If the
TCG calibration is not accepted, tap ClearCal and repeat steps 3)~4) to re-calibrate.
In multiple group mode, when one group of delay calibration is finished, tap Next to go to
Operation:
4-54
Ultrasonic Phased Array(PA)
2) Select SenCal from the desktop menu and enter the sensitivity calibration
screen, as shown in Fig. 4-61. RefAmp is the echo amplitude after the
calibration is completed, and the default value is 80%. RefError is the visual
error of the echo amplitude after the calibration is completed, the default
value is 5%, Depth is the scan depth. Adjust Depth to make the reference
echo appear at the center of the B-scan image, and then set GateMode to
to the required thresh, and then adjust Gain to make the echo amplitude to
about 80%. Move the probe back and forth to make each scan line reach the
4-55
Ultrasonic Phased Array(PA)
(green dotted line) is shown in Fig. 4-62. In calibration, the white dotted line
is the live envelope line of echo amplitude for each scan line. If the echo
Press StartAng and EndAng to decide the start line and end line for
4) Tap CAL to finish sensitivity calibration. Move the probe back and forth to
result is correct, the green amplitude envelope line should be among the
yellow dotted lines as shown in Fig. 4-63. C:S is shown in the B-scan image
4-56
Ultrasonic Phased Array(PA)
calibration, and the screen prompts “Do you want to enter the next
and return to the desktop menu. If TCG calibration is not accepted, tap
In multiple group mode, when one group of sensitivity calibration is finished, tap Next to
Operation:
2) Select SenCal from the desktop menu and enter the sensitivity calibration
screen, as shown in Fig. 4-65. RefAmp is the echo amplitude after the
4-57
Ultrasonic Phased Array(PA)
calibration is completed, and the default value is 80%. RefError is the visual
error of the echo amplitude after the calibration is completed, the default
value is 5%. Adjust Depth to make the reference echo appear at the center
of the B-scan image, and then set GateMode to Depth. When the setup is
to the required thresh, and then adjust Gain to make the echo amplitude to
about 80%. Move the probe back and forth to make each scan line reach the
(green dotted line) is shown in Fig. 4-66. In calibration, the white dotted line
is the live envelope line of echo amplitude for each scan line. If the echo
Press StartAng and EndAng to decide the start line and end line for
4-58
Ultrasonic Phased Array(PA)
4) Tap CAL to finish sensitivity calibration. Move the probe back and forth to
result is correct, the green amplitude envelope line should be among the
yellow dotted lines as shown in Fig. 4-67. C:S is shown in the B-scan image
4-59
Ultrasonic Phased Array(PA)
calibration, and the screen prompts “Do you want to enter the next
and return to the desktop menu. If TCG calibration is not accepted, tap
In multiple group mode, when one group of sensitivity calibration is finished, tap Next to go
TCG calibration is for the calibration of the difference in echo amplitude due to the
different distances between the same reflector and the probe in the ultrasound
propagation path. Oblique-hole test block and step test block, shown as Fig. 4-68 and Fig.
4-69, are usually used for TCG calibration. Oblique-hole test block is to calibrate angle
phased array probes and step test block is to calibrate 0-degree phased array probes.
4-60
Ultrasonic Phased Array(PA)
TCG calibration Method 1 uses a test block with a row of oblique holes, usually for angle
Operation:
2) Select TCG_Cal from the desktop menu and enter the TCG calibration
screen, as shown in Fig. 4-71. RefAmp is the echo amplitude after the
calibration is completed, and the default value is 80%. RefError is the visual
error of the echo amplitude after the calibration is completed, the default
value is 5%. Set GateMode to Depth. When the setup is finished, select
4-61
Ultrasonic Phased Array(PA)
3) Adjust GateStart and GateWidth to enable the gate record the first reflector
echo or the B-mode image (or the start depth whole echo desired by the
user). Adjust GateThres to the required thresh. Adjust Depth to the B mode
image display depth, and then adjust Gain to make the echo amplitude to
about 80%. Move the probe back and forth to make each scan line reach the
(green dotted line) is shown in Fig. 4-72. In calibration, the white dotted line
is the live envelope line of echo amplitude for each scan line. If the echo
Press StartAng and EndAng to decide the start line and end line for
4) Tap Record, and the Record menu displays “1-1”, the first number
represents the serial number of current record point, and the second number
represents total record point number. Move the probe back and forth to
result is correct, the green amplitude envelope line should be among the
yellow dotted lines as shown in Fig. 4-73. If the calibration result is not
4-62
Ultrasonic Phased Array(PA)
5) Tap NextPoint , Record menu displays “1-2”, and then finish the calibration
4-63
Ultrasonic Phased Array(PA)
"C: T", indicating TCG calibration is finished. Press Next to go to step 7).
7) Set the curve amplitude and number, as shown in Fig. 4-75. The user can
8) If the TCG calibration result is accepted, select Finish to exit the calibration,
and retu4rn to the desktop menu. If not accepted, tap TCG_Cal to repeat
In multiple group mode, when one group of TCG calibration is finished, tap Next to go to
A step block is used for TCG calibration method 2, mostly for testing 0-degree phased
array probes.
Operation:
4-64
Ultrasonic Phased Array(PA)
2) Select TCG_Cal from the desktop menu and enter the TCG calibration screen,
as shown in Fig. 4-77. RefAmp is the echo amplitude after the calibration is
completed, and the default value is 80%. RefError is the visual error of the echo
amplitude after the calibration is completed, the default value is 5%. Set
GateMode to Depth. When the setup is finished, select Next to go to step 3).
3) Adjust GateStart and GateWidth to record B-scan echo of the RefPoint, adjust
GateThres to the required thresh. Adjust Depth to the B mode image display
depth and then adjust Gain to make the echo amplitude to about 80%. Move the
probe back and forth to enable each scan line to achieve the maximum echo
amplitude, and to acquire the maximum echo amplitude envelope curve (green
dotted line), as shown in Fig. 4-78. In calibration, the white dotted line refers to
the live envelope of each scan line echo amplitude. If the echo amplitude
4-65
Ultrasonic Phased Array(PA)
4) Tap Record to record the echo amplitude of RefPoint 1. Move the probe back
and forth to reproduce echo amplitude envelope of each scan line. If the
calibration result is correct, the green amplitude envelope line should be among
the yellow dotted lines as shown in Fig. 4-79. If the calibration result is not
4-66
Ultrasonic Phased Array(PA)
5) Put the probe in the RecPt 2 position of the step test block, and then repeat step
needed record point, by referring to step 3) to 4). When the calibration is finished,
the B-mode image window Information bar displays "C: T", indicating TCG
7) Set the curve amplitude and number, as shown in Fig. 4-81. The user can set up
4-67
Ultrasonic Phased Array(PA)
8) If the TCG calibration result is accepted, select Finish to exit the calibration, and
retu4rn to the desktop menu. If not accepted, tap TCG_Cal to repeat steps 2)~7)
to re-calibrate.
9) In multiple group mode, when one group of TCG calibration is finished, tap Next
HorZero calibration is for calibrating the distance between the probe central scale or probe
flank and the defect. By calibration, the defect position can be accurately positioned in the
horizontal direction.
Operation:
1) Place the probe on a GS calibration block. Find a side-drilled hole (SDH). Adjust
A Line to the center of B image, and move the probe back and forth to find the
highest echo of the side-drilled hole, as shown in Fig. 4-82 and Fig. 4-83.
4-68
Ultrasonic Phased Array(PA)
2) Adjust the gate to record the SDH echo completely. Then keep the probe position
unchanged and measure the horizontal distance L1 from the probe flank to the
edge of the test block, and the horizontal distance L2 from the SDH to the edge
4-69
Ultrasonic Phased Array(PA)
3) L1 minus L2 to get the horizontal distance D between the probe flank and the
SDH, i.e. D=L1-L2.
4) Now go to HorZero in Gate menu, Adjust HorZero value to enable Ra-1 value in
the reading equal to D.
4-70
Management Operation
Chapter 5
Management Operation
This chapter describes the operation and technique for storage, encoder, probe, wedge
5-1
Management Operation
Before file storage, please set up the storage path. There are two kinds of storage paths:
system internal storage path and external USB storage path. The folder name of system
internal storage path is Store, and the external USB disk storage path is USB storage root
directory.
Operation:
In the Storage management surface, rotate the knob to highlight the storage path. Press
Enter key to enter the storage path, then press main menu Storage to exit storage
management interface. The setting of storage path is completed.
【Note】: If a new folder is created in the storage path, select New Folder, enter the
folder name in the information bar on the right of the screen, and then press Enter
key. Rotate the knob, highlight Back to Previous, press Enter key to return to the
previous directory.
In storage management interface, select FileType submenu, rotate the knob to select to
1) Parameter File: For saving system setup parameters. The formats include: PA
parameter file "* .pdat", UT parameter file "* .udat", and TOFD parameter file "* .tdat".
2) Screenshot File: For saving system screenshots, in the format "* .jpg".
3) Record files are the full record file of the PA advanced scan, BScan or TOFD scan, the
record file format of the phased array advanced scan is “*.prec”, UT ciip record or BScan
4) Report: For saving report screen, with the format "* .pdf".
In storage management interface, use the knob and Enter key to select the target folder,
then press Save onscreen and the virtual keyboard pops up on the screen. Meanwhile
5-2
Management Operation
there will be a file name on the upper right bar. The default file name is “Time & Date”,
After entering the file name, the user could tap the touch screen to switch to the file
After all information is entered, press Enter key and the file storage is complete. To quit
In default, press A2 key, to quickly save screenshots as the picture format to the user
In the storage management interface, the user can create a new folder under the storage
directory. Press New Folder to bring up the new directory dialog box, and then enter the
new folder name and press Enter key or OK on the virtual keyboard to finish creating the
new folder.
In storage interface, rotate the knob to highlight the selected file or folder, then press
Enter key. There will be a √ symbol appearing in the front box of the selected file or
folder, which means the selection is done. To cancel the selection, rotate the knob to
highlight the selected file or folder. The Select key will change to NonSel. Press NonSel
【Note】:Rotate the knob to highlight the Return to Previous icon, press the
onscreen Select key, and you can select all the files and folders on the entire
screen.
Editing the storage files include copying, moving, deleting or renaming the files or folders.
The copy function can copy the selected files or folders to another directory. The moving
Operation:
5-3
Management Operation
2) Press Copy or Move key in the Edit menu, another storage path operation menu will
appear on the right side of screen. Set up the storage path following 5.1.1 to set up the
b) Delete file
File deletion can remove the selected files or folders. The deleted files or folders cannot
be restored. Follow step 5.1.5 to select the files or folders to be deleted, select Delete in
the Edit menu, and the selected files or folders will be deleted forever.
C) Rename
Follow step 5.1.5 to select the file or folder to be renamed, select Rename in the menu
Edit, and the user can input the custom name and press OK on the virtual keyboard to
Select the files following step 5.1.5, press Enter or tap Open menu o recall the files.
After recalling the parameter file (suffix ". pdat ", ". tdat" or ". udat"), the system will reset
Image files (with suffix .jpg) will be displayed in full screen after file recall. Rotate the knob
to highlight the file to recall, and press Enter to open the image file. Turn the knob to
display the last or next image file. Press Cancel key to exit the recall interface.
The recalled full screen image is defaulted in the file path at the top of the screen. Press
After recalling the record file (suffix ".prec" or “.trec”), the user can enter the recall
After recalling the report file (suffix ".pdf"), the user can open the report file.
5-4
Management Operation
With the Wizard mode, the user may select the encoder name for calibration from the
touch screen, select Calibrate, and follow the screen prompt to finish calibration.
The operation of Calibrate Encoder is basically the same as Add Encoder. Refer to 5.2.5.
Edit Encoder is for editing information of new encoder parameters. The user may select
the encoder name from the touch screen, select Edit, and follow the screen prompt to
input the new encoder parameters. The default encoder name cannot be changed.
The user may select the encoder name from the touch screen, select Delete to delete the
5-5
Management Operation
For the encoder connected with the system for the first time, users shall create the
encoder name, so that the system may identify the encoder info such as precision.
Operation:
1) Tap Add menu in Encoder interface to pop up the new encoder calibration distance
window as shown in Fig. 5-3. The user can enter distance to be calibrated (the default
is 500mm) and use the ruler to mark start point and end point of the calibration
distance on the smooth surface. After the calibration distance is confirmed, press Ok
2) The screen display is as shown in Fig. 5-4. Place the encoder at the start position of
calibration according to the wizard and press Step 1 to move the encoder to the end
position of the calibration, and stabilize the encoder, the screen is as shown in Fig. 5-5,
5-6
Management Operation
3) The screen is shown as Fig. 5-6, move the encoder backward to the start of calibration.
If Move Distances equals to the calibration distance entered in the Step 1) or in the
admitted allowance range (like Fig. 5-7), tap OK to enter the next step. Otherwise, the
user can tap Initialize to reposition the encoder at the end of calibration and move it to
5-7
Management Operation
4) The Screen prompts window for inputting encoder name as shown in Fig. 5-8, The
user can input a user-defined encoder name and press OK key in the virtual keyboard
5-8
Management Operation
5) Adding encoder is completed as shown in Fig. 5-9. The properties of the added
encoder are shown in the right side of the screen including precision (mm) of the
Users may select an encoder on the touch screen and select Rename key to pop out the
virtual keyboard. Input the new encoder name. The system default encoder name cannot
be changed.
5-9
Management Operation
Users may manage probe parameters through probe management. Users may also
create custom probe, and perform operation of edit or delete probe parameters.
Users may manage wedge parameters for the available wedges via wedge management.
Users may also create custom wedge, and perform operation of edit or delete wedge
5-10
Management Operation
work piece or edit the created work piece. The user-defined work piece can be recalled in
Scan Wizard.
5-11
System Setup
Chapter 6
System Setup
brightness.
b) Screen brightness setup menu is for setting the system screen brightness. Level
c) Lower screen brightness can ensure a long life of the display screen, and make
the system work longer when the system is charged with the battery. When the
system is in bright environment, brighter screen can improve the visual effect of
the screen.
function.
b) After the touch screen calibration function is activated, please calibrate the touch
6.3 Default
This function may clear all setups in the system, and restore them to factory defaults.
6.4 License
This function is for activating system optional software.
Operation:
1) Connect the SD card or USB disk to the system, and wait till the system
6-1
System Setup
6-1.
4) Tap “Export activation file”, and this file will be stored in the SD card or USB
6) Copy the license key file acquired from the distributor to the root directory of
7) Connect the SD card or USB disk with the system, and wait till the system
9) Select the license key file in the pop-up menu and tap Yes. If the function is
screen operation is disabled, at the same time, the information bar at the top left corner of
the screen will display the disabled touch screen operation symbol as shown in Fig. 6-2.
6-2
System Setup
6.6 VGA
1) Select Setup-> System->VGA to turn on/ off VGA video out.
2) If the user connects the system VGA video out port to an external monitor, turn on
VGA (with the VGA enabled symbol displayed at the upper left of the screen), and
and apply this format in new file naming. For example: the last naming format is “0001”,
then when saving the new file, the file name input box will automatically use “0001” as
default name. The user can decide to use this name or reinput a new name.
Turn off the LastFName function, and the system will not automatically record the last
are two files saved at the same time, the name is repeated.
Operation:
language.
Operation:
Tap the item to modify color, select a suitable color, and then tap Ok to make the setting
take effect.
6-4
System Setup
6.12 AutoGain
The value of AutoGain can be set up as required.
Step: Select Setup -> Reference -> AutoGain, and turn the knob to adjust the gain.
6.13 AutoMode
The mode for AutoGain can be selected, including Single and Continue.
Single: Press the hotkey for AutoGain once, and the echo amplitude within the
Continue: The system will adjust the echo amplitude within the gate
Operation: Select Setup -> Reference -> AutoMode, with Single and Continue for
selection.
6-5
Operator Training and System Maintenance
Chapter 7
Operator Training and System Maintenance
ultrasound testing, especially training on phased-array ultrasound testing, and read this
Operation Manual before operating this system. The lack of relevant knowledge on
ultrasound testing may lead to unpredictable wrong testing results. Therefore, to acquire
information on ultrasonic testing training and qualification, please contact relevant NDT
normal, please contact our customer service. Our service staff will try to provide detailed
Check the functions of system panel buttons and touch screen buttons one by one and
In accordance with Chapter 2 and Chapter 3, adjust the parameters of each mode, and
7-1
Operator Training and System Maintenance
In accordance with Chapter 2 and Chapter 3, check each measurement and calculation
Check the probes for any abnormally like cracks; check if the probe cable sheath has any
peeling.
Check if the power cable sheath has any peeling. An accidental risk might occur in the
Note: As sections 7.2.5 and 7.2.6 above involve safety, always do these inspections for
the 2 items before operating the system every time. Other items may be inspected once
Inspect if the system and the accessories are damaged in mechanics and
functions.
Inspect if all functions of the system comply with the operation instructions.
If the system does not work properly or pass any of the inspections above, the
If the system is not used for a long time, regular power-up maintenance is required.
The system can operate normally the control panel can respond normally for
each operation.
The system display works properly, without any abnormal situations like flashing
or character running.
7-2
Operator Training and System Maintenance
The ultrasonic testing performance of the system complies with relevant industry
To ensure that the fundamental performance of the system is normal, the user shall check
and validate the items above on a regular base or as required according to the industry
requirements. The user may use the standard test block and electronic equipment for
testing and validation, or have the testing and validation performed by a qualified
Except for the standard peripheral interfaces clearly described in the operation manual
(such as USB and network. Please see "Operation Cautions and Safety Guide" for the
requirements to connect the peripheral devices with such interfaces to the system), all the
accessories are dedicated to the system. DO NOT connect any accessory not provided by
SIUI to the system, such as the probe, battery, adapter or encoder, otherwise it might
cause damage to the system, or even accidents like electric shock or fire in some extreme
cases.
7.2.10 Cleaning
DO NOT drop any liquid into the system. The system enclosure, glass panel and other
parts may be cleaned with a piece of cloth soaked with a little water or neutral household
cleanser. DO NOT use any solvent to clean the system. The plastic parts may be
7.3 Service
The system adopts the latest technologies and is made of high quality components. The
process inspection, the in-process testing and the quality management system are all
In case any failure is found in the system, please shut off the system and take out the
7-3
Operator Training and System Maintenance
re-calibrate the system. The contact information is listed in the Preface of the Operation
Manual. The warranty does not cover any unauthorized repair or disassembly.
Under the following situations, the safe operation of the system is not ensured:
b) The system is stored in adverse environment for a long time (e.g. abnormal
Severe fall-down or pressure during transportation. Before transportation, the system shall
7-4
Common Failures and Troubleshooting
Chapter 8
Common Failures and Troubleshooting
1) Copy the software update package to the root directory of the SD card.
port.
update is finished.
4) After system auto shutdown, remove the SD card from the system, and
【Tip】: After the update is finished, the update package in the SD card should be
deleted if the SD card will be used for data storage; otherwise when the SD card is
connected to the system, the auto update runs upon next system power-on.
response:
2) Remove the battery from the system, and wait for about 10 seconds before
installing the battery again in the system. Try to turn on the system.
3) Remove the battery from the system, and try to use the adapter for power
supply.
b) If the power supply is from the adapter, press Power and the system has no
8-1
Common Failures and Troubleshooting
response:
1) Check if the system is installed with battery. If yes, remove the battery, and
use the adapter for power supply separately. Try to turn on the system.
4) Disconnect the adapter from the main unit. Wait for about 10 seconds before
5) If the battery is not used on the system, try to use the system with a full-charge
c) The system can be turned on, but cannot enter the application properly.
1) Turn off the system, and disconnect all the peripherals connected to the
system, including USB device, crawler and network cable. Try to turn on the
system.
2) If step 1) does not work, try to update the system with the update package.
d) If the failure still exists after trying the solution above, please contact the
distributor or SIUI.
to force the system off, or unplug the adapter and/ or remove the battery to cut off power
supply. Restart the system to see whether it is in normal operation. If not, try to contact the
b) Unplug the SD card or USB disk, restart the system and plug in the SD card or
8-2
Common Failures and Troubleshooting
b) Check if the adapter indicates normal operation. If the adapter works normally,
If the failure still exists after trying the solution above, please contact the distributor or
SIUI.
b) Check if the phased array probe is connected to the system properly. See 1.9
Phased Array Probes for the connecting method between the phased array
identified probe model will be shown in the information bar at the top of the
screen.
b) Check if the connected phased array probe matches the current scan mode.
Select Setup->System->Default, and the system will set the default parameter
b) Select the waveform for sector scan subject to the probe and the wedge in use. If
no wedge is used, select LWave. If a wedge is used, and the wedge angle
8-3
Common Failures and Troubleshooting
exceeds the first critical angle, select TWave, otherwise select LWave.
c) If the wedge with an angle bigger than the first critical angle is configured, the
start angle and the end angle for sector scan shall fall in the range specified by
the wedge.
d) Wedge parameter of the phased array probe are described based on common
steel work pieces. If other materials are tested, the types listed may not apply.
8-4
System Technical Specifications
Appendix A
System Technical Specifications
General Technical Specifications
Unit mm/inch
Battery Quantity 1 pc
Ethernet Connector 1 pc
Alarm Connector ——
A-1
System Technical Specifications
transportation
Relative Humidity 10%~95% (without condensation)
conditions
IP Rating ≥IP65
Imaging mode: UT
A-2
System Technical Specifications
A-3
System Technical Specifications
A-4
System Technical Specifications
A-5
System Technical Specifications
A-6
System Technical Specifications
A-7
Description of Symbols and Signs
Appendix B
Description of Symbols and Signs
Table B-1 Description of Symbols and Signs
B-1
Description of Symbols and Signs
ISO 7000-0623 Up
—— USB interface
—— Network Interface
—— Encoder interface
—— SD card socket
—— Video Out
B-2