GRP - PM.QA.04A - Prosedur Ultra Sonic Test (AWS) - Update
GRP - PM.QA.04A - Prosedur Ultra Sonic Test (AWS) - Update
GRP - PM.QA.04A - Prosedur Ultra Sonic Test (AWS) - Update
REVISION NOTE
Prepared By Approved By
TABLE OF CONTENT
1. SCOPE
1.1. This testing procedure provide the technical information and detailed steps necessary to ensure
a complete and repeatable ultrasonic testing of welds.
1.2. This Procedure is comply with the recommendation of AWS D1.1 Ultrasonic Testing Only
Non Tubular Conection & Tubular Joint with diameter 24” and greater.
2. REFERENCE
2.1. AWS D1.1, 2020 Edition
2.2. ASME BPV Section V Article 4 and 5 – 2019 (As Complimentary)
2.3. ASTM A 435 Standard Specification for Straight Beam Ultrasonic Examination of Steel Plate
4. EQUIPMENT
4.1. Portable pulse echo 'A' scan (rectified) presentation units with test frequency range of 1.0 MHz
to 6.0 MHz will be used. It shall be equipped with a stepped gain control calibrated in units of
2.0 dB or less.
4.2. Probes, either single or twin crystals, with polarized ceramic crystals shall be used. No
dimension of the crystals shall exceed 25.0 mm (1.0 inch), The angle of angle beam probe shall
be between 45 to 70 within the test material.
4.3. Reference block used for equipment’s calibration and standardization shall be one or more of
the following :
i IIW VI Block, Distance Sensitivity Calibration Block
ii IIW V2 (DIN 54122) Block, Distance Sensitivity Calibration Block
The proper functioning of the these block shall be of similar material as the test components
and meet the requirement of the governing code specification.
4.4. Viewing (light) hood over the CRT screen shall be used if ultrasonic testing is to be carried out
in bright condition.
4.5. Couplants like grease, oil, wall paper paste (cellulose) and other water based couplant may be
use as coupling medium. Same couplant shall be used in calibration and production testing.
High temperature grease shall be used for couplant for examination on surface with
temperature of more than 50o C.
4.6. Temperature of calibration block shall be within range actual on the test specimen or calculate
due to temperature deviation to reference level.
PLATE SERVICE CENTER
Document No. GRP.PM.QA.04A
QUALITY PROCEDURE Revision 01
Issued May 14, 2020
Title : Ultrasonic Testing (In accordance with AWS D1.1/D1.1M- Page 4 of 16
2020) Department QA/QC
6. SURFACE CONDITION
6.1. The surface where scanning is to be carried out shall be free of scale, weld spatters, pitting,
thick coating and other irregularities which may interfere in the transmission of sound wave.
8.5. The length of discontinuity shall be determined by 6dB (1:2) drop technique.
8.6. The cross section of discontinuity shall be sized with 20 dB (1:10) drop method or maximum
amplitude technique or a combination of both if required.
8.7. Reporting of defective areas shall be referred to a zero datum and the weld center line for
constant configuration and defective area physically marked adjacent to the defective weld.
11. REPORTING
11.1. The Client shall receive copies of approved reports generated in accordance with this
procedure. The report shall be counter signed by the Client.
11.2. Reports shall contain full details of test restriction and procedure deviation wherever
applicable.
11.3. Copies of the above reports shall be retained and filed by PT. GUNUNG RAJA PAKSI
APPENDIX I
TRANSFER CORRECTION
1.0 GENERAL.
1.1 There is a difference in instrument sensitivity between the reference block and test
component due to test roughness, contact area, component temperature and
acoustical characteristic. A correction in instrument sensitivity is required to
compensate for these difference (transfer correction) and is described below.
75%
Line R
50 %
25 % F
2.0. PROCEDURE
2.1 Use two angle beam probes of the same type. One acting as transmitter and the
other as receiver with the test unit set to dual mode.
2.2 Place the probe in tandem directed at each other on the reference block at one skip
distance and maximized the signal to 75 % . Note down the gain (dB)
2.3 Mark on the screen the peak of the maximized signal.
2.4 Reposition the probe without altering the test unit sensitivity to achieve peak signal
at two skip distance repeat step 2.3 and join the two marks to obtain line R.
2.5 On the surface of the test component, carry out step 2.2 at the same gain (dB)
setting.
Increase or decrease the maximize signal peak (line F as shown above) as to bring
it to line R.
Note down the gain (dB) difference.
2.6 The difference in gain (dB) in step 2.2 and step 2.5 is the transfer correction and
has to be added or subtracted to the primary reference sensitivity level prior to
scanning.
2.7 The first skip distance has to be greater than four inches or other wise the
successive skip distance should be employed
PLATE SERVICE CENTER
Document No. GRP.PM.QA.04A
QUALITY PROCEDURE Revision 01
Issued May 14, 2020
Title : Ultrasonic Testing (In accordance with AWS D1.1/D1.1M- Page 8 of 16
2020) Department QA/QC
APPENDIX II
UT ACCEPTANCE-REJECTION CRITERIA AS PER AWS D1.1/D1.1M-2020
PLATE SERVICE CENTER
Document No. GRP.PM.QA.04A
QUALITY PROCEDURE Revision 01
Issued May 14, 2020
Title : Ultrasonic Testing (In accordance with AWS D1.1/D1.1M- Page 9 of 16
2020) Department QA/QC
PLATE SERVICE CENTER
Document No. GRP.PM.QA.04A
QUALITY PROCEDURE Revision 01
Issued May 14, 2020
Title : Ultrasonic Testing (In accordance with AWS D1.1/D1.1M- Page 10 of 16
2020) Department QA/QC
APPENDIX III
TECHNIQUE SHEET: LAMINATION CHECK
0
t
Pate
PLATE SERVICE CENTER
Document No. GRP.PM.QA.04A
QUALITY PROCEDURE Revision 01
Issued May 14, 2020
Title : Ultrasonic Testing (In accordance with AWS D1.1/D1.1M- Page 11 of 16
2020) Department QA/QC
INSPECTION SEQUENCE:
1. Test surface shall be cleaned and smooth as practically allowed. Filling or light grinding
(if permitted) should be carried out on rusted and pitted surface. Paint work shall be
removed for thickness gauging.
2. Apply thin layer of couplant on the test surface and place the probe over the couplant.
Press the probe lightly and evenly into the test surface.
3. Maximized the signal to get a stable reading (digital meter) twist the probe lightly if stable
signal could not be achieved.
4. Record down the reading with sketch. For lamination check at least two signals which
indicate the wall thickness should be seen.
5. Special couplant, probe housing and technique will be detailed if testing to be carried out
on hot surface above 50 C.
6. Defect sizing shall be carried out at primary reference level plus transfer correction not be
achieved.
PLATE SERVICE CENTER
Document No. GRP.PM.QA.04A
QUALITY PROCEDURE Revision 01
Issued May 14, 2020
Title : Ultrasonic Testing (In accordance with AWS D1.1/D1.1M- Page 12 of 16
2020) Department QA/QC
APPENDIX IV
TECHNIQUE SHEET: STRUCTURAL BUTT JOINT, T JOINT & CORNER JOINT
TIME BASE CALIBRATION : Zero degree probe - At least twice the material thick.
: Angle probe - 100 mm across screen or 1-1/2 skip range
whichever is greater.
SENSITIVITY SETTING : Zero degree probe - 1st backwall echo set to 50% to 75%
FSH.
: Angle probe - Echo of 1.5 mm diameter side drill hole in
IIW V1 Block set to 40% to 60% FSH + dB Transfer
Correction
dB SCANNING of Angle Beam : dB Sensitivity Setting 1.5 mm diameter side drill hole IIW
V1 Block + dB Transfer Correction + dB Scanning level
(Non Tubular Joints)
Face A
0 0 Angle probe
Face B
Face A Face B
Face C
Face A Face B
Face C
Face A
Angle Probe 0
Face C
0
Face B
Face A
Angle Probe 0
Face C
0
Face B
Backing Plate
INSPECTION’S SEQUENCE.
1. Zero degree probe scan of the parent material at least 2 skip distance from the weld center
line.Scan both sides of the weld parent material. this will determine the material thickness
and will detect internal reflector which might affect interpretation of angle beam results.
Location and areas of such reflector should be reported.
2. Root scanning - as per section 7.3
3. With appropriate angle probes, scan the whole volume of the welds. See scanning sketch
on this technique sheet.
4. Minimum scanning level - primary reference level plus transfer correction plus required
gain setting where 2 mm high grass signals from the opposite face are observed.
5. Defect sizing shall be carried out at primary reference level plus transfer correction.
Acceptance criteria should be referred back to client's requirements. It should be specific.
6. Defective areas shall be marked and recorded in a simple sketch of cross-section of the
weld indicating defect depth, position and orientation reckoned from reference datum
point.
7. Written report should be generated on weld tested. Rejectable discontinuity shall be shown
in full cross-sectional profile. All scanning limitations and restrictions shall be reported