VT-350
OPERATOR’S MANUAL
0120
Contents
1.Introduction.......................................................................................5
1 .1 Product features and using range............................................................................................. 5
1.2 Classify.......................................................................................................................................5
1.3 Main performance index...........................................................................................................6
2. Security considerations.............................................................. 11
2 .1 Security identity...................................................................................................................... 11
2 .2 Safety precautions before use................................................................................................ 11
2.3 During Use.............................................................................................................................. 13
2.4 After Use..................................................................................................................................13
2.5 Maintenance and Check..........................................................................................................14
2.6 Disposal....................................................................................................................................15
3.Configuration.................................................................................. 15
3 .1 Refractor head configuration................................................................................................. 15
3.2 Accessories.............................................................................................................................. 17
3 .3 Identification........................................................................................................................... 18
4.Ready use.......................................................................................... 19
4 .1 Setup connection of the device.............................................................................................. 19
4.2 Install near vision chart...........................................................................................................20
4.3 Install face shield.................................................................................................................... 21
4 .4 Install and remove the forehead assemble.............................................................................22
4.5 Adjust the level........................................................................................................................ 22
4.6 Alignment cornea.................................................................................................................... 23
4 .7 Using near vision chart...........................................................................................................23
2
5 .3.4 Vision chart state and vision chart working mode switching area...........................40
5 .3.5 User-defined vision atlas editing interface................................................................ 40
5.4 The system state and menu area............................................................................................ 43
5 .4 .1 The use of customer information management module........................................... 44
5 .4.2 The use of custom program module.......................................................................... 48
5 .4 .3 External connection equipment management............................................................54
5.5 Fast reset and basic Settings.................................................................................................. 59
5.5.1 Change the system Settings........................................................................................ 59
6.Instrument operation.................................................................... 60
6.1 Checking with fog................................................................................................................... 60
6 .2 Precise inspection astigmatism axis and degree................................................................... 62
6 .3 Accurately measure sphere degree if there is a deviation.....................................................63
6.4 Binocular balance test.............................................................................................................64
6 .5 Heterophoria measurement.................................................................................................... 65
7. Printer using.................................................................................. 67
7 .1 Basic function of the printer.................................................................................................. 67
7.1.1 Turn on......................................................................................................................... 67
7 .1.2 Shutdown..................................................................................................................... 67
7.1.3 Self-inspection.............................................................................................................67
7.1.4 Battery Charging..........................................................................................................67
7.2 Printing mode.......................................................................................................................... 68
7 .3 Printer indicator and buzzer signal introduction................................................................... 68
3
Please read this manual before use.
This device is manufactured accord to GB9706. 1 Medical electrical equipment
general requirements for safety, have to connect the power plug to a grounded outlet.
In order to ensure use the product safely, and prevent damage to operator and other
people, or cause a lose to other devices, warning labels are indicate on the machine as
well as in the manual, please pay attention and read the accompanying documents
carefully.
ISO 9001/ 13485 certification— R&D, Manufacture and service of the products by Our
company have passed ISO 9001/ 13485 certification . IEC standards applied in this manual.
The dioptre are indicated with a reference wavelength of 5 4 6 . 0 7 nm . Materials in direct
contact with the patient passed the ISO109 9 3 assess, not made up unacceptable risk.
The safety precautions and operating procedures must be thoroughly understood before
using the device. And please safekeeping this manual. If you encounter any problems or
have questions about the device, please contact Ningbo Ming Sing’ s local authorized
distributor or contact with our customer service department directly.
Information contained in this manual has been confirmed when publish. Product
specifications are subject to change without prior notice. The rights of change the product
which contains in this manual is reserved by our company, and without prior notice. Sold
products does not involve in this change.
Do not copy, retrieve, distribute any chapters of this book by electronic, mechanical,
recording or any other means if not written approval by our company.
4
1 Introduction
Computerized vision tester ( hereinafter referred to as vision tester) is used to
subjective examination of visual acuity and refractive error of subject’ s eyes.
1.2 Classify
Figure 1. 1
5
Figure 1. 1
Storage and transport environment conditions:
Temperature: - 1 0 ~ 5 5 C
Humidity: 10 ~ 85 %
Air pressure: 70 0 hPa~ 106 0 hPa( transport), 50 0 hPa~ 10 60 hPa( storage)
a) Temperature : 10 C - 40 C
b) Humidity ≤80%
c) Air pressure: 800hPa - 1060hPa
1.3.2 Step
Sphere: 0.25D/0.5D/3.0D
Cylinder: 0.25D/ 1.0D
Axis: 1º/5º
Prism: 0. 1△/0.5△/2.0△
Prism base minimum measure step: 1º
PD: 0.5mm
1.3.3 Tolerance
6
Sphereф (D) 0.00≤ ︱ ф ︱ ≤3.00 ±0.06 Remnant
3.00< ︱ ф ︱ ≤6.00 ±0.09 astigmatic
6.00< ︱ ф ︱ ≤9.00 ±0. 12 power≤0.03
9.00< ︱ ф ︱ ≤ 12.00 ±0. 15
12.00< ︱ ф ︱ ≤ 15.00 ±0. 18
15.00< ︱ ф ︱ ±0.25
b) The tolerance of astigmatic power should accord with the requirement table 1. 2
diopter Tolerance D
>5.00- 10.00
c) The optical center tolerance of mechanical system of sphere and cylinder combinative
0.00 0. 12
>0.00-6.00 0.25
7
≤5.00 ±0.25 cm/m
>5.00 ±5%
e) Cylinder axis and prism base tolerance should accord with the requirement of table 1.5
f) PD and the reference range scale should accord with the requirement of table 1. 6
Item Tolerance mm
VD ±0.5
PD ±0.5
1.3.4.1 The minimum aperture of all the lens of vision tester should be not less than 16mm,
for the prism equal to or more than 6 cm/ m, the aperture can be reduced to
11mm.
1 .3.4.2 Manufacturer should indicate the reference plant of vision tester, both sides should
have the identify of reference range.
1.3 .4 .3 The vision tester should have at least one set of mask and dichoptic device, each
system should mount a Jackson cross cylinder.
1.3.4.6 Structure of lens cavity should be not influence the patient examine.
8
1.3.4.7 The device should be alignment and centered when the lanes and accessory fixed
in front of the sightglass.
1.3.4.8 The device should be equipped with auxiliary lens, every auxiliary should be
identify corresponding.
1 .3.4.9 The device using LCD display, the displayed character, data and curve should be
clear.
1 .3.4.10 The PD should be able to continuous adjustment, the range is no less than
50mm—75mm.
1.3.4.11 All the lens disks should be installation corresponding lenses as fixed data, the
center of lenses and holes should be alignment.
1 .3.4.12 All the buttons of the device are identification clearly, function and status should
be convert normally when press the keys.
1.3.5 Appearance
1 .3 . 5 . 1 The surface of vision tester should be smooth, bright and clear, cannot have etch
pit, dirt, scratch and front, rags and etc.
1 . 3 . 5 . 2 Effective clear aperture of the vision tester’ s optical element cannot have bubble,
defect, scratch and any macroscopically observable irregular surface defects.
The material which contact with the patient ( forehead rest and face shield) should be
a) No cytotoxicity.
1.3.7.1 The part and nearby part where contact with the patient or operatives should be
liable to clean. No disinfections dead angles should be on the parts which need to
disinfect.
1 .3.7.2 Please cleaning and disinfect the device according to this user manual, should not
9
cause device damage or material deteriorate, or affect to the security protection
performance.
1.3.8 The environment test of computerized vision tester should be accord with the
requirement of climate environment Ⅱ mechanical environment test group Ⅱ in
GB/T14710- 1993
1.3.9 Safety requirements should accord with stipulates in GB9706. 1
10
2. Security considerations
In this manual, words used to indicate the security alert degree or level are defined as
follows:
Warning: indicates a potentially dangerous situation, such as can' t
warning be avoided, could result in death or serious injury.
Attention: indicates a potentially dangerous situation, such as can' t
attention be avoided, may result in slight or moderate injury or property
damage.
Under certain conditions, even if the conditions for attention are also likely to lead to
serious damage. So safety precautions must be strictly abided by the moment.
warning
• Be sure to secure the refractor head to the refractor arm .
If the refractor head is dropped or falls, injury or device failure may result.
CAUTION
• The safety precautions and operating procedures must be thoroughly understood before
using the device.
Unintended use may cause unexpected malfunction or adverse experience.
• Do not store the device in an area that is exposed to rain or water, do not put any
container which with gas or liquid.
• Do not store the device in dusty, hot, humid places or in direct sunlight.
11
The following conditions must be maintained during use.
Temperature: 10 to 40 ºC (50 to 104ºF)
Humidity: 30 to 85% (No condensation)
Pressure: 760 to 1060 hPa
Installation place: Interior
No hazardous dust or smoke
A dust- free place
A place free from vibration and shock
• Be sure to use a wall outlet which meets the power specification requirements.
If the line voltage is too high or too low, the device may not operate properly. Malfunction
or fire may result.
• Do not use any power cord other than that equipped. Do not use the equipped power
cord for purpose other than as intended.
Or else malfunction or fire may result.
• When installing and operating the device, observe the following instructions about EMC
( electromagnetic compatibility) :
- Do not use the device simultaneously with other electronic equipment to avoid
electromagnetic interference with the operation of the device.
- Do not use the device in the same room with other equipment such as life- support
equipment, other equipment that has major affects on the life of the patient and results of
treatment, or other measurement or treatment equipment that involves small electric
current.
- Do not use the device simultaneously with portable and mobile radio frequency
communication systems because it may have an adverse effect on operation of the
device.
- Do not use cables and accessories that are not specified for the device because that
may increase the emission of electromagnetic waves from the device or the system and
decrease the resistance of the device to electromagnetic disturbance.
12
shielding measures should be taken or change the placement of this equipment, in order
to reduce the possibility of interference.
WARNING
• When moving the refractor head, make sure that there are no obstacles in the path.
If the arm is swung carelessly, it may bump against someone or something and injury or
malfunction may occur.
CAUTION
• Keep the measuring windows free of fingerprints and smudges.
Failure to do so could affect the data accuracy.
• In the event of smoke or strange odors, immediately turn off the device and disconnect
the power plug from the outlet. After you are positive that the smoke has stopped, contact
Ningbo Ming Sing or your authorized distributor.
Continued use of the device under such abnormal conditions may cause fire or electric
shock.
• When the device is not in use, please turn it off and put the dust cover over it.
If the device is not covered for an extended period, the accumulation of dust may affect
the data accuracy.
• Perform the visual and operation checks before using the device. If any abnormality is
found, do not use the device.
Continued use of the device under such abnormal conditions may affect the data
accuracy. Unexpected malfunction or faulty diagnosis may induce unexpected health
hazards.
• Before each patient, always clean the patient' s contact area ( forehead rest and face
shields) using disinfectant alcohol.
• This instrument must be operated by the trained or qualified people, or after their
instruction.
CAUTION
• If the device will not be used for the extended period, disconnect the power cable from
13
the wall outlet.
• Occasionally clean the prongs of the power plug with a dry cloth.
If dust settles between the prongs, it may collect moisture, and short circuit or fire may
result.
CAUTION
• Any repair or service to the instrument must be done by Ningbo Ming Sing Optical R & D
Co. , Ltd or people trained and can be operate correctly with experience or contact with
the authorized distributor to do the maintenance.
• Screw out the safe screw may cause the machine separate from the support arm and
cause the serious damage.
• Do not open the body cover or try to repair the interior components. Any repair or service
to the instrument must be done by experienced staff from Ningbo Ming Sing Optical R &
D Co., Ltd or the authorized distributor.
• The adjustment must be done by Ningbo Ming Sing Optical R & D Co., Ltd technician or
other authorized staff.
• The refractor head operation must be strictly according to the user manual. If not follow
the user manual, the patient safety and the normal work of instrument can not be
guaranteed.
CAUTION
• Never use organic solvents or high washing power detergent to clean the exterior of the
device.
It may ruin the surface of the device.
• Never put the refractor head into the liquid, otherwise it will broke the machine.
• Avoid touch the optical components to prevent leaving fingerprint or oil stain to decrease
the performance.
14
2.6 Disposal
CAUTION
Follow the local ordinances and recycling regulations regarding disposal or recycling of
the components.
It is recommended to commission the disposal to a designated industrial waste disposal
contractor.
When disposing of packing material, follow the local ordinances and recycling regulations.
3 Configuration
15
Figure 3. 1
1) Examination window … … … .The patient’s eyes are observed through the window, and
patient observe charts through the window.
2) Near vision illumination … … Light up when near vision testing, light off when far vision
testing.
3 ) Corneal aligning windows … …The position of patient’s cornea can be observed through
the window.
4 ) Forehead rest adjust knob … … Moves the forehead rest forward and backward in order
to adjust the VD.
5) Near-point rod holder … … … … .The near-point rod is inserted and attached here.
16
6) Near-point rod clamp screw … ..Fix the near-point rod.
7) Spirit level … … … … … … … … … … … Used to confirm that the refractor head is level.
8) Leveling knob … … … … … … … .Adjust the level of the refractor head.
9) Forehead rest magnet … … … … .Fix the face shield.
10) Forehead rest … … … … … … …The patient’s forehead rests here when test.
11) Cornea alignment scale … … … Measures the VD.
3.2 Accessories
Figure 3.2
1 2 ) Face shield (2 sets.) … … … ..Attached to the instrument to put the patient’s face in
place, use one while the other is being sterilized.
13) Dust cover … … … … … … … Covers and protects the instrument body from dust and dirt
during storage.
14) Forehead rest (2 sets.) … … . The patient’s forehead rests here. Detachable by a touch,
use one while the other is being sterilized.
1 5) Near-point rod … … … … …The card holder is attached at the near point examination
position.
16) Card holder … … … … … … …The near-point card is attached.
17) Operator’s manual … … … … Describes instructions about handling the instrument
18) Near-point card … … … … … .. Contains the near-point charts, for test the near vision.
17
Warning
The accessory in this device supplied by our company, other supplier’ s accessory
could be use only after verify the safety by our company
3.3 Identification
Moisture proof
Manufacturer
EU authorized representative
Production date
Series number
AC
Fuse indicate
18
Figure 3.3
There is warning mark in front of device: Do not close to the near point rod, prevent of
crushing
4 Ready use
Profession staff of our company or authorized personnel who have been trained by
our company are required on the installing process .
It use with a secure connector if the refractor head install on the stand or ophthalmic
unit which from other companies, as showing in figure 4. 1. Adjust the leveling knob which
is showing in figure 3 . 1 after connect the refractor head, to make the refractor head at
level condition.
19
Figure 4. 1
If the refractor head install on our ophthalmic unit, it is no need for connector, as show
in figure 4 . 2 , it could be install on our ophthalmic unit directly. Power supply unit and
cables are fixed in the ophthalmic unit.
Figure 4.2
Install the near point rod, card holder and near point card before using the device.
Insert the card holder into the top end of the near point rod. The card holder is
20
designed to slide along the rod. As shown in the Figure 4.3.
Figure 4.3
Insert the near point card into the spring which is attached in the card holder.
Next, insert the near point rod into the near point rod, fasten with the set screw. When not
in use, hold it upright as shown in the next illustration.
Figure 4.4
Warning
To avoid injury due to contact, do not bring the face close to the near point rod.
21
Figure 4.5
Remove the forehead rest as Figure 4 . 6 . To mount the forehead rest, reversely follow the
remove procedure.
Figure 4.6
Before using, make sure the refractor head is level. If the bubble is not between two lines,
make it level by operating the leveling knob.
22
4.6 Alignment cornea
Make the forehead rest move back and forth by turning the forehead rest knob.
With the patient’ s forehead set to the forehead rest, look through the cornea aligning
window from about 25 cm apart.
Figure 4.7
Observe the anterior pole of cornea and check it on which line as show in figure 13 . If
the eye comes on the long line, it means the distance from the eyeglass to the cornea
surface is 12 mm, short lines are given at 2 mm intervals, the dotted line is 13 . 7 5 mm from
the cornea surface. For example, if the cornea surface is 4 mm away ( on the second short
line) from the long line, it means the glasses wearing distance is 12+ 4=1 6 mm. If the
eyeglass wearing distance and the measuring distance is different, correction the
When the patient wears multi- focal lenses, detection the diopter when look at the near
distance. It needs to install the near point rod, card holder and near point card
Bring down the near point rod ,and screw the retaining screw.
The scale of the near point rod is graduated in cm and inch unit.
Value matching the both end lines of the card holder indicates the distance between
the eye and the card. (Figure 4.8)
23
Figure 4.8
Turn the rotary part at the bottom part of the card to select the required chart. The chart
will appears in the sight window( Figure 4.9)
24
Figure 4.9
Remark: The number and name of the charts which doctor seeing is corresponding with
1 :Letters
2 :Letters&Numbers
3 :Numbers(V)
4 :Astigmatism
5 :Letters
6 :Cross grid
7 :Numbers(H)
8 :Landolt’s rings
25
5 Tablet application interface and function
introduction
The main interface of operation program is divides into three parts: refractor head
operate and data display area, Monitor chart operate area, system condition and menu
area. Main interface layout and the above three areas layout are showing in figure 5. 1.
26
5.2 Refractor head operate and data display area
Refractor head operate and data display areas are consist of 7 parts, as showing in
figure 5.2.
This area realizes three main functions: 1. Auto display the data of the lens of refractor
head. 2. Click to select the optometry items. 3. To switch the status and parameter of the
The detailed function block of this area is showing as figure 5.3. If the item changes to
27
white, for example , it means it’s not selected.
1) Right eye spherical power display area and right eye spherical power test mode
select button.
2) Binoculus spherical power test mode select button, display the current step- length of
sphere.
3) Left eye spherical power display area and right eye spherical power test mode select
button.
*Long press any button of the 1,2,3 for at least one second, the sphere step-length
select menu will appear, click to select the needed sphere step- length. Sphere
step- length menu is showing in figure 5.4, button functions are as below:
0.25D 5D
4) Right eye cylinder power display area and right eye cylinder power test mode select
button.
28
5) Binoculus cylinder power test mode select button, display the current -/+ mode of
cylinder.
6) Left eye cylinder power display area and left eye cylinder power test mode select
button.
*Long press any key of the 4,5,6 for at least one second, the cylinder mode select
menu will appear, click to switch the cylinder mode. Cylinder mode menu is showing in
Button display - +
7) Right eye axis display area and right eye axis test mode select button.
8) Binoculus axis test mode select button, display the current axis step-length.
9) Left eye axis display area and left eye axis test mode select button.
* Long press any key of the 7,8,9 for at least one second, the axial step and power
select menu will appear, click to select the axial step(1 °or 5°) or specified axial power
(180°/ 135°/90°/ 45°) . Axial setting menu is showing in figure 5.6, button functions are as
below:
Button 1 5
display
function
29
display
10, 11, 12, 13, 14, 15 is corresponds to prism item, the display mode is different according to
rectangular coordinate mode or Polar Coordinates mode as showing in figure 5.7 and 5.8.
1 0) Right eye horizontal prism power display area and horizontal prism test mode
select/ right eye data display area and test mode select under polar Coordinates mode.
1 1) Binoculus horizontal prism power test mode select/ test mode select under binoculus
polar Coordinates mode, display prism step in corresponding mode.
1 2) Left eye horizontal prism power display area and horizontal prism test mode select/
left eye data display area and test mode select under polar Coordinates mode.
1 3) Right eye vertical prism power display area and vertical prism test mode select/ right
eye data display area and test mode select under Rectangular coordinate mode.
14 ) Binoculus vertical prism power test mode select/ test mode select under binoculus
rectangular coordinate mode, display the prism step in corresponding mode.
1 5) Left eye vertical prism power display area and vertical prism test mode select/ left eye
data display area and test mode select under Rectangular coordinate mode.
*Long press any key of the 10, 11, 12, 13, 14, 15 for at least one second, the prism step
and mode select menu will appear, click to select the needed prism step (axial 1°or 5°,
30
0. 1△ or 0.5△) . Prism menu is showing in figure 5.9 under Rectangular coordinate mode,
Button function Prism step Prism step 0.5△ Switch to polar Clear
0. 1△ Coordinates
mode
Axial test mode setting menu is showing in figure 5. 10 under polar Coordinates mode,
Button 1 5 X-Y 0
display
function 1° 5° Rectangular
coordinate mode
Test mode setting menu is showing in figure 5. 11 under polar Coordinates mode,
display
31
function Rectangular
coordinate mode
16) Right eye ADD power display area and ADD test mode select
18) Left eye ADD power display area and ADD test mode select
This button realizes two functions: switch the near/ far test mode and display current near,
Select the far or near point test mode. Far point test mode will be selected automatic
* If current is far point test mode, when switch to near point test mode, if have not
process near point test, it will automatic copy the S+ADD,C,A power of far point test
32
to the S,C,A power of near point test mode. It will not copy if near point test have been
taken.
* If current is far point test mode, when switch to near point test mode, if have not
adjust the PD value of near point test mode, the PD value of near point will calculate
according the far point PD value. If near point PD value have been adjusted, then it
* The light will turn on default if switch the far point condition to near point condition,
the light could be turn off by long press for two seconds if no need near
point light,
There are three buttons corresponding with NC, CV and Final switching. The
NC: Uncorrected visual acuity test mode, in this mode, all lens will withdraw, input VA
value.
Final: The final prescription. Cope the data of refractor head to Final, could adjust the
The rightmost two buttons are program step switch buttons after load the custom
program or standard program . There are two display conditions:
This area realizes the PD test and input NC data operate. It’s showing in Figure 5. 13
33
Figure 5. 13 PD and NC test area
* Cross lens will put in as well as cornea light turn on automatic when click 1 , 2 , 3 . To
*It will exit PD test mode automatic as well as withdrwa corss lens when click
4. Right eye VA display and right eye VA input and select button
6. Left eye VA display and left eye VA input and select button
* When click 4,5,6, it will pop up the select box according to the sight of current vision
chart, when click the target sight, the data will copy to corresponding VA data area of this
eye. The choice box will not pop up if the currenty vision chart is not belong to general
34
5.2.4 Right/ Left eye auxiliary lens operate buttons
This area realize the display of current put in auxiliary lens, and auxiliary lens select
Click the left and right auxiliary lens operate button, it will pop up the left auxiliary lens
select box (Figure 5. 14) and right auxiliary lens select box (Figure 5. 15) respectively.
When click the corresponding auxiliary, the lens will be put in automatic, and will back to
* Click anywhere except the auxiliary lens, the select box will disappeared automatic,
35
or ratate the plate counterclockwise: Minus the data of current test item .
: In the screen of main testing interface, if not insert cross cylinder and under
binoculus or right eye test mode, it will cover the left eye automatically, and then insert
right cross cylinder, set the dial of right eye testing, and set cross cylinder as side [ 1] .
In the screen of main testing interface, if not insert cross cylinder and under left eye test
mode, it will cover the right eye automatically, and then insert left cross cylinder, set the
dial of left eye testing, and set cross cylinder as side [ 1] .
In the screen of main testing interface, if the cross cylinder insert already, then the cross
cylinder will be at side [ 1] .
: In the screen of main testing interface, if not insert cross cylinder and under
binoculus or left eye test mode, it will cover the right eye automatically, and then insert left
cross cylinder, set the dial of left eye testing, and set cross cylinder as side [ 1] .
In the screen of main testing interface, if not insert cross cylinder and under right eye test
mode, it will cover the left eye automatically, and then insert right cross cylinder, set the
dial of right eye testing, and set cross cylinder as side [ 1] .
In the screen of main testing interface, if the cross cylinder insert already, then the cross
cylinder will be at side [2] .
36
5.3 Monitor chart operate area
This area realizes the management, callout, display of the built- in vision charts and
37
5.3.1 Vision chart type switching area
The current visual chart has three types: visual functional examination type, general
letter chart type and color blindness type. Six types of key functions in Program is as
follows,:
Button Function
Choose general letter type chart, through the short press can switch
the content of the visual figure in the vision chart selection area.
Choose general visual function type chart, through the short press can
switch the content of the visual figure in the vision chart selection area.
Choose color blindness type chart, through the short press can switch
the content of the visual figure in the vision chart selection area.
User- defined vision charts 1, through the long press can turn out the
user- defined vision atlas 1 editor interface, through the short press
User- defined vision charts 2, through the long press can turn out the
user- defined vision atlas 2 editor interface, through the short press
User- defined vision charts 3, through the long press can turn out the
user- defined vision atlas 3 editor interface, through the short press
* The icon style shows the currently selected type The icon style show
that the type is not selected
Vision chart choose area, a page at most shows 24 thumbnail images of vision chart,
38
by clicking on the thumbnail, automatically control LCD chart display the specified vision
chart, and control the refractor head switch to the corresponding state .
one page. Color- blind figure, general letter and visual function check types need more
pages to display, such as 5.3. 1 described, through the short press can realize content
switch, also can realize the flip over switch by horizontal stroke screen operation in vision
* Long press the binocular visual function chart , will show the vision chart simple
Vision chart display area has two types, corresponding the general letter vision chart
and general visual function test vision chart respectively. For example, figure 5. 17 is a
general letter vision chart, figure 5. 18 is a general visual function test figure.
* General letter type, according to the chart 5. 17, for example, with the blue button 1, 2, 3,
by clicking on, can realize single block or cancel block. Button marked with 0. 1, 0. 16,
shows the sight values of this line, by clicking on, can realize single block or cancel a
single block.
* Directly click on the specified chart, can realize single letter block or cancel a single letter
block.
* When vision chart under block condition, swiping the screen on the vision chart display
area for horizontal and vertical operation, can realize the block up, down, left, right four
39
directions of movement.
* General visual function check vision chart display ,see 5. 18 for example.
switching area
Realize genera letter enter into/ exit red and green contrast
mode
Through long press user- defined atlas, will enter to the user- defined instance atlas editing
40
Figure 5. 19 user-defined instance atlas editing interface
The definition of the button type is consistent in 5.3. 1. Click on the button, will display
the corresponding all vision chart in the built- in vision chart area.
Show all the built- in vision charts of the specified type. There are three types: the
current general letter vision chart, visual function test vision chart and color blindness
user- defined atlas, will cover the previous user- defined atlas.
Exit the edit button. Don’t save the current state of editing,
42
5.3.5.5 Add new vision chart to user-defined atlas
Add one or more new vision charts to define atlas, or replace with a new vision chart
instead of original vision figure, need to make the following three steps:
1. If add the needed vision charts in the built-in visual figure display area, to drag
into the user- define display area, cannot release in the process . Can see a small
chart move with finger, display content is the new vision chart which need to
added.
2. Drag the vision chart to the specified location of user-defined visual figure shows
area. If moved to the location can be placed, will display a larger vision chart in
the upper left corner. Placed in empty vision chart position, it means add in the
empty position. If placed in the vision chart already exist, then it will replace with
Remove specified visual figure from the original user- defined images, need to make
1. Press on the figure which needs to delete in the user- defined vision chart display
area for 2 seconds, the specified vision chart will become empty state.
Button in this area has realized the basic function of initialize program loading, the
user information loading, equipment connection, printing, the system Settings. Keys
43
Customer information management key. Showing the customer
function.
loading state, and the user- defined program callout function, user- defined
editing functions.
callout function.
Quick reset button . Rapid reset the measurement interface data, fast
Customer information not loaded, click the button, will enter into the
44
registration or the old users searching.
Customer information loaded, if click the button, will directly enter into
Click on the customer information management button, as shown in figure 5.2 3 interface:
Click on the [Add] button. Switch to the new customer information input interface, as
45
Figure 5.2 4 Customer information input interface
Customer name and phone number should be fill in, other content is optional, can be
modified in the future. After confirmed, click on the [ confirm] button, if the construction of
the new customer profiles successfully filed, it will jump directly to the customer optometry
Under this interface, it will display the current client' s name, phone number and
optometry historical information. The history list as below is briefly shows the final
prescription for the power of sphere and cylinder and the corresponding date.
Click on [modify] button in figure 5.25, will callout the interface as shown in figure 5.24.
46
Resume load the customer information need to modify and click on the confirm button, it
Click on the[login out ] button in figure 5.25 ,will exit the current loaded customers and
If the current client loaded, click login out to return to the interface as shown in figure
5 .23. If no current customers loaded, click on the user management key, to enter into the
Input the customer's name (or last name) in the textbox in figure 5.23 and click search.
For example two customers have registered "Lucy" and " Linda", only need to input “L” in
the textbox, When click search, the list will show all the name with "L" + XX (for one or
If we need to load current customer is the "Lucy", directly click on the list of Lucy, it
will automatic loading this customers, and jump to a customer optometry history is shown
47
5.4.1.5 Customer optometry historical data loading
in the list below, It means the historical optometry data is exist, according to the date
display, select the optometry history data and click once, it will jump out optometry data of
the specified user and the specified date interface, as shown in figure 5.27:
The interfaces in detail show the naked eye, Auto ref- keratometer data, lensmeter data,
Corresponding function: read the display data from the file, and loaded into
the current measuring interface
Corresponding function: not load data, returns to the historical records list to
reselect.
Custom procedure not load, click on the button, will enter into custom
program choose interface. Long press button for 2 seconds, will enter
48
the custom program editing interface.
Custom procedure loaded, click on the button, will enter into custom
program choose interface. Long press button for 2 seconds, will enter
There is one default standard procedures and six user- defined procedures in the
program . Selection interface loaded in user-defined program as shown in figure 5.28, click
49
Figure 5. 29 custom program editing interface
Clicking to select target edit program . The standard procedure is protected and
cannot be edited.
It is consistent with the vision chart selection area of 5.3. 1. Do not keep three
user- defined vision atlas, click on the specified vision chart in 3 ( The currently selected
vision chart display area) and insert the specified vision chart, shows the default auxiliary
lenses in 4 ( auxiliary lens operation area), update the contents of 6 (optometry options
50
5.4.2.2.3 The currently selected vision chart display area
Display the current step auxiliary lenses. The same use with 5.2.4 right/left eye
Display the current selected program step list. There is the step number, vision chart
selection and left/ right eye auxiliary lens selection in display content . Currently selected
steps will become dark blue background, steps which not selected is light blue.
Click on the specified steps, will automatically load the specified steps vision chart,
The steps of the current single program better no more than 32 steps. Too long may
The area is used to control the basic condition of refractor head in the process of the
1 Specify measuring Right eye, left eye, left first than right, binoculus
eye
state
51
fog degree
Implement save, delete, insert and steps of switching and exit and other operations in
the process of program editing. Buttons from left to right function is described below:
Pre step Save the current page changes, and return to previous page
Next step Save the current page changes, and display the content of next
save Save the current page modify settings, update step list
insert To save the current page content, as a new step into the next
delete Delete the current step, display the next step, update step list
Close Dot save the current page changes, exit the program editing
interface
The function of the custom optometry program block is introduced in the previous
details. Next we will illustrate the specific steps of last optometry case.
For example, in the process of accurate measurment on cylinder, will use Cross
First choose point cloud icon, click on the specified measuring project button, switch
to the " cylinder", In general, the cloud point figure insert automatically join with crossed
cylindrical lens, if you are not careful operation to cancel it by mistake, you can insert
again cloud point figure, and it will automatically to join. Click to specify additional fog
cloud button to add a certain value, the final screen display as shown in figure 5.30:
52
Figure 5.30 Insert point cloud figure in custom edit program
In the process of measuring heterophoria with Maddox, it can be set like this.
First select point light source, then click the right eye auxiliary lens operation button,
Click , will automatically return to the program editor interface, set the test
project as " horizontal prism", measured the eye as the "binoculus ". The final screen
53
Figure 5. 3 2 Heterophoria test program editing interface
Similarly, optometrists can be input every optometry steps in the program, if use the
program next time, can call out the preset program selection menu inside the main testing
( click can return to the previous optometry step) set projection icon or select
auxiliary lenses will greatly improve the efficiency.
Refractor no
head
management
54
Monitor chart no
management
Auto
Refractometer
management
Lensmeter
management
Printer no
management
management interface
management interface
management interface
management
The operation way of Optometry head, chart monitor, lensmeter printer and
refractometer is the same. Next make the optometry head device as an example,
Long press the refractor head management keys for 2 seconds, turn out the refractor
55
head connection management interface, as shown in figure 5.33:
Device list is divided into two parts, one is the current established head connected,
the other part is the equipment record which has set up a connection records with control
site before ( No matter whether these devices are real), is referred to as the device history.
The current connection has been established, will show optometry icon to
identify. When the connection has been established, click on the equipment records in any
If the current device not connected, click on the equipment records in any of the
connection work with refractor head will be carried out in the background, when
56
the connection is successful, the refractor head management buttons in the right
After the success connection with auto refractometer by Bluetooth, control site will
monitor message from the auto refractometer. When the auto refractometer has the
optometry data transfer action, the control site will remind the user by changing control
Short click the auto refractometer management button will turn out the auto
57
5.4.3.3 Print data preview and printer operation
Short click the printer management button will turn out the print preview and print
It is divided into five parts according to the optometry item. Each part can be
custom to add t print function or not . Just click the specified print part, will change whether
to join the print. As shown in figure 5.37, the five parts of the content is selected to print, if
do not need CV data, then click the print content of CV can remove it. The results are
Contrast figure 5.37 and figure 5.38, the color of the wire frame on behalf of its
whether to print
58
5.5 Fast reset and basic Settings
Click reset button will pop up prompt box as shown in figure 5.39:
Fast reset function will empty the current naked eye, auto refracotmeter, refractor
head and print data. General suggest printing operations, first to save data to the
When device is in abnormal status, for example, refractor head chuck or dislocation,
Click the system settings button, will turn out the basic attribute setting of the system .
Including related to measuring step set, the relevant content of output printing and the
program default load after reset and so on. As shown in figure 5.4.
For example, to change the default sphere lens step length from 0.25 D to 0.5 D:
Click [ sphere step] button on the right, the menu will pop up as shown in figure
5 .41, directly click [0.5 D] button, the default sphere lens step length is changed to 0.5
D
59
Figure 5.41 sphere step length selection menu
6 Instrument operation
This section will combine a specific case, demonstrates the use of conventional
optometry.
L :- 1.25DS/0-0.50DC X 180°
PD=65mm
Vision correction only about 0 . 7 , may be the original degree is not enough lead to
1 、sphere power of Right eye manual set in higher + 3.00 D than the real power. Because
the right eye of the customer is - 1.00 D, plus + 3.00 D, become + 2.00 D.
2 、At this time, the vision value should be lower than 0. 1, manual cast a vision icon, and
then gradually adjust the degree, to reduce the sphere for - 0.25 D one step, from +2.00D
to + 1.75 D and then to + 1.50 D, until his vision value about 0.5.
60
Figure 6. 1 radiation vision chart
Ask the customer " can you see some of the line particularly clear? ”
If the customer answers " all look the same," it shows this custom have no astigmatism,
measuring end.
If the customer answer “one line seems particularly clear "", then multiply small
Adjust astigmatism axis, and then adjust the astigmatism degree, from 0 . 0 0 to - 0 . 2 5 ,
The customer has myopia, the sphere degree of - 1. 7 5 D, after inspection, his degree
61
Figure 6. 2 Customer vision inspection situation
1 . Manual option dot chart, crossed cylinder lens automatically applied in the right eye. Or
from the auxiliary lens take crossed cylinder lens to be supplemented single chart, or
both sides, ask customer which side is more clear, for example, if is more clear, it
should to click the upward button , if is more clear, it should to click the
upward button , device will automatically change the astigmatism axis. After then
press and repeat the aboved steps, until the both sides is similar clear then
62
lens status shall be maintained) . Click or to change the two sides of
crossed cylinder lens, ask customer which side is more clear. if is more clear,
it should to click the upward button ,if is more clear, it should to click
the upward button , the device will automatically change the astigmatism
degree. After then click and repeat the above steps, until the both
sides is similar clear then definite the precise astigmatism degree. under the condition
of changes astigmatism degree, device can automatically change sphere and cylinder
linkage. It means : If cynlinder change ±0.50DC , sphere will automatically
change±0.25DS.
deviation
( Test with red and green filter, if a customer shall not apply to the red and green test can
Project red and green chart , ask the customer which chart is the more clear,
if the customer says green is more clear, it means that myopia is overcorrection, should
add sphere degree+0 . 2 5 D 。 or directly enter the green button , device can
automatically add the degree.
Then ask the customer again, if he says red is more clear, it means that myopia is
undercorection ,add-0.25D ,or directly enter the red button ,device can
63
automatically decrease the degree. It indicates that the degree of customers should be
between- 1.50D and - 1.75D, according to the requirement of the customer to choose high
or low power. Screen will display like this:
Next ,check left eye. Cover the right eye, left eye cancel to cover. With the samestep as
In order to make the customers obtain the better eyesight condition, we need to do
( Test with polarization separation chart, also can according to need to use single chart
64
Figure 6.5 Binocular balance test chart
2. Customer look at the upper of the picture with right eye, see the lower of the picture
with the left eye. If it is same clear or fuzzy with the two lines, it means that binocular
vision is balance. if not ,add sphere +0.25D on the eye which can see more clear ,
3. Next, select the binocular mode, do binocular fog, add + 1.00D sphere for both eyes on
the basis of the data measured before.
4. Add -0.25D for both eye every time ,Until reach to the desire vision value.
First test horizontal heterophoria, select binocular mode, use the right eye lens option
65
Projector dot chart, the customer see one red vertical line (42a) with right eye,see a
lighting point with left eye (42b)。Blending image of both eyes shown as the picture 42c or
42d. If the picture is shown like 42d ,There is no or low strabismus with melt image by
itself , no need to adjust prism degree. If it shown as in picture 42 c, use left prism degree
button BI or BO, move left and right lighting point, remind the customer, if see the picture
Next to test the vertical heterophoria. Select binocular mode, use the left eye lens
option buttons, add red Maddox on the left eye, cancel the cover with right eye.
the customer see one red horizontal line with left eye ,see a lighting point with right
eye. ,use right prism degree button BU or BD, move left and right lighting point, remind the
customer, when see horizontal line and lighting point then stop measure, repeatedly
If the heterophoria degree is bigger, it needs to make correction when fitting glasses.
in this case the customer is not very serious, so no need to correct, the final result is:
66
7. Printer using
7.1.1 Turn on
Under printer shutdown state, press [ POWER] for one second, the printer will start to
work, the power light (red) turn on.
Attention: If the paper indicator light ( blue) of printer is flashing but paper is enough, it
means the power of the printer will use up, please change the battery or charging.
7.1.2 Shutdown
Under printer work state, press [ POWER] for one second, printer turn off and all lights
turn off.
7.1.3 Self-inspection
When customer found the printer abnormal working, can perform self- test of th
e printer and checking the current settings and state.
Shutdown and keep pressing [FEED], then press [POWER] to turn on, when h
ear the sound drops ”DI” , release the [POWER], the printer will print the self-test p
aper, including the current situation of the printer, printer setup, the printer examp
le, such as electronic quantities detection, Bluetooth ID etc.
After the printer used for a period, the battery will run down , when the battery close to
use up , the sending paper indicator light ( blue) will flash and with a alarm sound at the
same time, the customer need to prepare charging . If the printer auto shut off during the
printing, customer need to charge before continue using.
Insert the charge plug to 2 2 0 V, 5 0 HZ power socket first , than put charge plug to
printer DC socket, the printer will in charge state . During the charging, if printer working,
the paper indicator light ( blue) is flashing , till the battery power full, charge indicator light
is on for a long time. In the process of charging, can normal use the printer.
67
7.2 Printing mode
After Printer startup, search the printer by program on the tablet, select the
corresponding printer to connect, if it is the first time to connection, you need to enter the
password "00 00" , generally within 10 seconds, the connection is successfully established.
If the connection failed to establish within 1 0 seconds, check that the Bluetooth on the
tablet is on or not, try to connect it again.
8.1 Troubleshooting
When the device is working abnormal, please try to troubleshoot according to the
following form.
symptom treatment
When the switch is in the on position, · Confirm the power plug in the socket
the head without any response · Confirm the power switch is open on the
table
The tablet could not be opened · Confirm whether the tablet battery running
68
down of battery
The printer could not be opened · Confirm the printer battery running down of
battery
If there is no troubleshooting way according to the inside of the form, please contact
the dealer.
Cleaning:
1 . Keep products clean ordinary, cannot use strong volatile solvent, diluent and benzene
to clean.
2 . With a soft cloth dipped in soapy water which after dry to wipe the product part.
3 . Wipe the lenses and mirrors, blowing off foreign object on the surface, then with a soft
dry cloth to wipe.
Window cleaning: When fingerprint, oil of eyelashes or dust is sticky on the inspection
window, accuracy of result will be influenced. Make sure that the inspection window is
clean before inspection. If there has stain, please using soft and clean cloth to clean it.
Before each test, must make the Windows clean. If Windows have besmirch, with a clean
soft cloth to clean.
If there is water vapor condensation, please unload the mask, then unload the
protection glass, clean it carefully. When unloading the protecting glass you have to use
special screwdriver to unfix four screws.
69
Figure 8. 1
Disinfect:
After the completion of each patient, please disinfect face shields and forehead.
Methods: with a soft cloth dipped in 75% concentration of medical alcohol to wipe,
8.3 Scrapping
Discarded products, accessories and spare parts, in accordance with the local
government about accessories and spare parts discarded laws and regulations and
recycle relevant laws and regulations, especially lithium battery can cause pollution to the
environment.
When discarded packaging materials, should also be in accordance with the local
government of relevant laws and regulations about packing material waste or recycling
· Appearance specifications
70
power 125mm×50mm×31.5mm net weight 0.32kg
· Contraindications:none
5 years
( 2 ) Operation and maintenance accord with operator’ s manual and service manual.
71