Andro Dia
Andro Dia
Andro Dia
About AndroDiags
AndroDiags Installation
2.1 2.2 2.3 2.4 2.5 What Is in this Chapter Installing AndroDiags on your PC Configuring COM ports Setting up a Field Modem Configuring a system
abc
AndroDiags
About AndroDiags
This section describes the functionality of Andromeda Diagnostics.
In some cases toolbar buttons and menu entries are disabled. E.g., if you make a local connection, the remote connection button is disabled.
What is it?
AndroDiags is an easy-to-use servicing tool. To clarify button functionality, dragging the mouse over a button will cause a short button description to pop up.
If you are not able to solve a problem yourself, or if you need more information, please feel free to contact your local AGFA representative.
abc
AndroDiags
2
2.1
AndroDiags Installation
What Is in this Chapter
This chapter provides information on how to install AndroDiags for servicing. The contents of this chapter is essential during the installation process of equipment.
Introduction
Overview
This chapter covers the following topics: Topic What Is in this Chapter Installing AndroDiags on your PC Configuring COM ports Setting up a Field Modem Configuring the equipment
For more information, refer to: Troubleshooting on page 131 in the Instruction Manual.
abc
AndroDiags
2.2
Four disks
Proceed as follows to install AndroDiags on your PC: Step 1 2 Action Insert the disk labeled disk 1 into the floppy drive of your PC. Click Add/Remove Programs in the Control Panel dialog box. or Select Run from the Start Menu. 3 On the Install/Uninstall tab, click Install. or Type a:\setup and click OK. Windows now scans the floppy drive, searching for the setup command. 4 Follow the instructions on the screen.
Installing updates
Agfa will issue new releases of the AndroDiags software at regular intervals. Ask your local Agfa representative for any new releases. Proceed as follows to install updates: Step 1 2 3 Action Insert the disk into the floppy drive of your PC. Select Run from the Start Menu. Type a:\update and click OK.
abc
AndroDiags
2.3
COM ports
X20
X23 X16 X21 X15 X17 X27 X26 X33 X32 X31 X34 X29 X28 X22
CO
X18
CO
M1
or
X30
X14
Proceed as follows to configure COM ports: Step 1 Action Select COM-Port Settings from the Tools > Options menu. The following window pops up:
2 3
Select a COM port for local connections, i.e. the serial cable port. Select the baud rate at which the connection operates, normally 9600.
Tick the AutoBaud Up/Download box to automatically increase the baud rate while downloading software. 4 5 Select a COM port for remote connections, i.e. the modem port. Click OK to confirm the settings.
abc
AndroDiags
2.4
Modem setup
Proceed as follows to set up a field modem: Step 1 2 3 4 Action Disconnect the modem from the electrical box. Connect the modem to a local port (COM1) of your PC using an RS232 cable. Launch AndroDiags. Select Program Field Modem from the Tools menu. The following window pops up:
If ... the modem at the customer site does not occur in the list the modem at the customer site occurs in the list
Then ... click Add Modem and proceed with step 6. proceed with step 8.
abc
AndroDiags
Step 6 7
Action Enter the name of the modem. Enter all the other data required in the remaining text boxes.
Refer to the documentation shipped with the modem and the table below. 8 Click Program Modem to initiate the modem setup. The following window pops up:
9 10
Connect the modem to COM1 of the electrical box. Click Yes to confirm.
abc
AndroDiags
Remaining data
The table below describes the remaining settings to be entered. You can find these settings in the documentation shipped with the modem. Box Reset to Factory Settings No Handshake Ignore DTR Auto Answer Baudrate to 9600 Disable Status Messages No Echo Save Factory Settings Description The modem must ignore all settings and return to the original default settings. The modem may not support software or hardware data flow control (XON/XOFF - RTS/CTS). The modem must ignore the Data Terminal Ready (DTR) signal. The modem automatically answers incoming calls via the telephone line. The baud rate of the modem must be set to 9600. The modem may not send messages of his own, such as RING or CONNECT messages. The modem may not echo received messages. The settings above become the standard parameters after switching the modem off.
abc
AndroDiags
2.5
Configuring a System
During the installation process, the configuration of the system (e.g. LP150) must be specified in the System Configuration dialog box of the AndroDiags software.
Installation
Proceed as follows to configure a system: Step 1 Action Connect your PC to the equipment using a serial cable.
Refer to Configuring COM ports on page 11 of AndroDiags Installation. 2 3 Select Local Connect from the Connect menu or click the button on the toolbar to connect to the equipment. Select Configure System from the Software menu. The following window pops up:
abc
AndroDiags
3
3.1
Introduction
Overview
This chapter covers the following topics: Topic What Is in this Chapter AndroDiags Quick Reference Guide The AndroDiags Toolbar
For more information, refer to: Troubleshooting on page 131 in the Instruction Manual.
abc
AndroDiags
11
3.2
Quick start
If ... Before first use (Installation) you wish to set up a modem at the customer site you wish to download software onto the equipment you wish to configure equipment Diagnostics you wish to establish a local connection you wish to establish a remote connection you wish to upload the current logbook of the equipment you are connected to you wish to view the up-tothe-second process diagram of the equipment you do not know the cause of a problem, enter terminal mode
select Program Field Modem from the Tools menu. select Download Software from the Software menu. select Configure System from the Software menu.
Setting up a Field Modem on page 12. Downloading New Software onto Equipment on page 34. Configuring the LITHOSTAR LP150 on page 15.
select Local Connect from the Connect menu or click the button on the toolbar. select Remote Connect from the Connect menu or click the button on the toolbar. select Upload Logbooks from the Diagnostics menu or click the button on the toolbar. select Process Diagram from the Diagnostics menu or click the button on the toolbar. select Terminal Mode from the Diagnostics menu or click the button on the toolbar.
Establishing a Remote Connection on page 23. Uploading and Opening Logbooks on page 25. Retrieving the Process Diagram on page 31. Entering Terminal Mode on page 33.
abc
AndroDiags
12
3.3
Overview
The table below lists all buttons of the toolbar of the AndroDiags main window: button Description Click this button to exit AndroDiags.
Click this button to establish a local connection, i.e. through the COM port. Click this button when you wish to end the local connection with the LITHOSTAR LP150. Click this button to establish a remote connection, i.e. via the modem. Click this button when you wish to end the remote connection.
abc
AndroDiags
13
4
4.1
Introduction
Overview
This chapter covers the following topics: Topic What Is in this Chapter Checking Equipment Information Establishing a Remote Connection Uploading and Opening Logbooks Analysing Logbook Information Retrieving the Process Diagram Entering Terminal Mode Downloading New Software onto Equipment Installing New Equipment Releases on your PC AndroDiags Troubleshooting
abc
AndroDiags
14
4.2
What
Proceed as follows to check equipment information: Step 1 2 Action Make sure you are properly connected to the equipment. Select Equipment Info from the Software menu. The following window pops up:
3 4 5
Check the equipment type or software release installed. Enter the serial number (e.g. SN1012) and the location of the equipment. Click OK to save the settings.
abc
AndroDiags
15
4.3
Remote connection
Proceed as follows to add a telephone number to the list: Step 1 Action Select Remote Connect from the Connect menu or click the corresponding button on the toolbar. The following window pops up:
If ... you wish to add a number to the end of the list you wish to add a number before the selected one in the list
3 4
Enter a site name, its modem number, extra site info and a contact person. Click Save Edit to add it to the list.
abc
AndroDiags
16
Select Tone or Pulse Dial when the telephone line does not support the default dialling method, i.e. Regular Dial. 3 If ... it is a direct modem number it is not a direct modem number 4 5 Then ... click the Modem Dial button. proceed with step 4.
Dial the phone number and wait until you are through-dialled. Click the Manual Dial button, when you hear the remote modem is expecting a dial tone.
Proceed as follows to modify or delete telephone list entries: Action Modify telephone list entries Delete telephone list entries What to do? Click Modify underneath the list. Click Delete underneath the list.
abc
AndroDiags
17
4.4
What is a logbook?
Overview
With respect to logbooks, there are three major activities: For details on ... Selecting a logbook viewer Opening saved logbooks Uploading logbooks Refer to ... How to select a logbook viewer? How to open saved logbooks?. How to upload logbooks? .
Firstly, you have to select a logbook viewer, i.e. a text editor, you wish to use to open logbooks. Proceed as follows to select a logbook viewer: Step 1 Action Select Logbook Viewer from the Tools > Options menu. The following window pops up:
Use the pull-down menu to select a text editor program or type the full path of a text editor program in the File name box.
abc
AndroDiags
18
3 4
Type a distinctive name for the logbook in the File name box. Click Save to save the logbook.
Refer to Analysing Logbook Information on page 28 for details on the dialog box displayed.
AndroDiags will save the logbook in the AndroDiags\logs directory. Of course, you can save the file in another directory.
abc
AndroDiags
19
2 3
Select a logbook from the list. Click Open to view the Logbook Analysis dialog box.
Refer to Analysing Logbook Information on page 28 for details on the dialog box displayed.
abc
AndroDiags
20
4.5
Errors histogram
abc
AndroDiags
21
The screen shots above represent an example of the LITHOSTAR LP150. What you can see: amount of developer, finisher and wash water used (in litres) m of plate processed and rewashed.
abc
AndroDiags
22
Click this button to select another logbook, of which then the Errors histogram will be displayed.
Click this button to view the underlying logbook of the histogram displayed.
abc
AndroDiags
23
4.6
Overview
Proceed as follows to open the process diagram: Step 1 2 Action Make sure you are properly connected to the equipment. Select Process Diagram from the Diagnostics menu or click the button on the toolbar. The uploading process will take a few moments, after which the process diagram dialog box is displayed: (example Lithostar LP150)
abc
AndroDiags
24
The table below briefly explains what you can do in the process diagram area: What to do 1 2 Place the cursor on the element. Click the right mouse button and move the cursor to Lock. Select either 0 or 1 and click the left mouse button. Place the cursor on the element. Click the right mouse button. Example
obtain direct info on elements symbol name board location (wiring diagram)
1 2
Check status bar box at bottom of the dialog box. Check message box at bottom of the dialog box. Use X to ignore and V to confirm.
abc
AndroDiags
25
4.7
Proceed as follows to enter terminal mode: Step 1 2 Action Make sure you are properly connected to the equipment. Select Terminal Mode from the Diagnostics menu. A window similar to the one below will pop up:
3 4
Type the number of the menu to be opened and press enter to confirm. To quit terminal mode, click Done.
abc
AndroDiags
26
4.8
Different versions
Proceed as follows to install new software onto equipment: Step 1 2 Action Make sure you are properly connected to the equipment. Select Download Software from the Software menu. The following window pops up:
abc
AndroDiags
27
Step 4
Action Select the equipment type you are dealing with and click OK. The following window pops up:
The crossed-out software version is the last one installed on your PC. 5 Select the equipment software version you wish to download and click OK.
abc
AndroDiags
28
4.9
Updates
Proceed as follows to install new equipment release on your PC: Step 1 Action Select Install New Equipment Release onto PC from the Tools menu. The following window pops up:
Browse to the directory of the new equipment release if a:\ is not its home directory (default).
When equipment software has been found, the three boxes at the bottom of the dialog box indicate the equipment type and its name as well as the equipment software version. 3 Click Install to start the new equipment release installation.
If the version you are about to install appears to be older than the version that is installed on your PC, you will be notified and you will be able to cancel the installation process, if necessary.
abc
AndroDiags
29
If you selected a software version that is older than the one already installed on your PC, this message will ask you for additional confirmation.
In both cases, the already installed software will move to a new sub-directory and it will be crossed-out in the Select Software Version dialog box (See Downloading New Software onto Equipment on page 34). An example of the directory structure: AndroDiags 5891 A0_10_02 B0_11_03
abc
AndroDiags
30
The table below lists possible causes and solutions: Cause The serial cable is connected to COM2 at the electrical box. The COM port settings are not correct. The Baud Rate is not 9600. The modem cable is not connected. The equipment is switched off. Solution Plug it into COM1. Check settings in Tools > Options > COM port settings. Set Baud Rate to 9600 in Tools > Options > COM port settings. Connect the modem cable. Switch on the equipment.
Dialling problems
The table below lists possible causes and solutions: Cause The modem is not connected properly. The modem number is not correct. The modem number is not a direct one. Solution Check modem connection. Check modem number. First dial central phone number.
abc