Calibration of Micro - Pipette
Calibration of Micro - Pipette
Calibration of Micro - Pipette
Programme
Calibration of Micro-pipette
By
Gravimetric Method
Nihal Gunasekara
Sri Lanka
Bangladesh BEST
Programme
Types of Micro-pipettes
Single Channel
Multi- Channel
Bangladesh BEST
Programme
Requirements
Unit of volume
l
Ref. temperature
To deliver
calibrated to
transfer the stated
another container.
Cubic millimeter :
20 C
Apparatus is
accurately
volume to
Bangladesh BEST
Programme
Apparatus Needed
Balance
Thermometer
Hygrometer
Barometer
Bangladesh BEST
Programme
1 l V
Resolution Std.
of
Deviation
Balance
(mg)
(mg)
10 l
0.002
0.001
10 l < V
100
0.01
0.02
100 l < V
1000
0.1
0.2
1 ml < V
10 ml
0.1
0.2
Bangladeshi BEST
Programme
Recommended
Apparatus/Liquids
Apparatus/Liqui
d
Thermometer
0.2 C
Hygrometer
10%
Barometer
0.5 kPa
Calibration
Liquid
Uncertainty
Distilled /de-ionized
water
Bangladesh BEST
Programme
Less than 60 s
Bangladesh BEST
Programme
Test Volume
Fixed Volume M
Variable Volume
volumes
Preferably
nominal volume
Measurements
Nominal Volume
At least three
10%, 50 % and
10 at each volume
Bangladesh BEST
Programme
Bangladesh BEST
Programme
Bangladesh BEST
Programme
Calibration Procedure
Replace the disposable tip of the piston pipette
Fill the piston pipette with test liquid, immersing its
delivery orifice 2 to 3 mm below the surface of water
Release the operating button slowly and draw the
pipette vertically against the side wall of container .
Expel the water to waste to pre-wet the tip and refill
the piston pipette as above.
Record the mass m0 of the weighing vessel or tare the
balance to zero
Deliver the contents of the pipette into the weighing
vessel by touching pipette tip against the container
wall just above liquid surface at an angle 30 to 45
Record the mass m1 of the weighing vessel and
determine the quantity delivered
Repeat the above cycle for ten measurements
Bangladesh BEST
Programme
Bangladesh BEST
Programme
vi mi Z
The Z correction factor is a function
of test temperature and air
pressure
13
Bangladesh BEST
Programme
Correction
Factors for Distilled
Water
80
85
90
95
100
101.3
105
17
17.5
18.0
18.5
19.0
19.5
20.0
20.5
21.0
21.5
22.0
22.5
Bangladesh BEST
Programme
Mean Volume
1 10
V Vi
10 i 1
15
Bangladesh BEST
Programme
Data Recording
Use spread sheet for data recording
Record Temperature, Humidity,
Barometric pressure
Record final data in a data recording
sheet
Review from a third party
Prepare the final calibration
certificate
Bangladesh BEST
Programme
Thank you