On-Line Gas Chromatographs

DMC / GC.ppt slide 1 July 01


Daniel On-Line Gas Chromatographs

•World leader in Gas Chromatography


•Over 5000 GC’s Installed Worldwide
•Two Manufacturing/Application Centers:
Houston, TX, and Larbert, Scotland

DMC / GC.ppt slide 2 July 01


Core Technology
Chromatograph Valves
Sample inject and
Column Switching Valves
6 port, Piston-operated diaphragm
Valve
Unique air-to-air diaphragm actuation
based on original NASA design
No springs, washers, or lubricants
Sample does not come in contact with
internal moving parts
>5 Million operations per valve
Lifetime warranty

DMC / GC.ppt slide 3 July 01


Core Technology
Micro-Packed Columns
Faster analysis times with extremely
low carrier consumption
. (10-15
cc/min. typical)
Superior component separation
enhances baseline separation and
sharpens peak resolution
Combines durability of packed
columns with advantages of capillary
columns
.043” ID
Designed together with valve and
detector to minimize internal dead
volume
2 year warranty under normal
operation

DMC / GC.ppt slide 4 July 01


Core Technology
Micro-Thermal Conductivity Detector
Uses (TCD) thermistors to
measure the concentration of
each component.

Improved sensitivity for low


level measurements (eliminates
FID requirement for some
applications)

Rugged -- Immune to carrier


gas failure and vibration.

Detector block and column


blocks are heated. A thermal
cutout prevents heating beyond
85° C.

DMC / GC.ppt slide 5 July 01


Core Technology
Heat Sink Oven
Integral Heat Sink Oven contains
all columns, valves and TCD
detectors

Approved for Class I Div I Groups


C&D

Minimal Utilities -- No Air or


Hydrogen Required

+/- 0.1° C over ambient


temperature (-18° to 55° C)
Ensures maximum stability and
repeatability in outdoor
environments.
DMC / GC.ppt slide 6 July 01
Core Technology
Environmental Test Chamber
Analyzers and electronic
components are operated over
the specified temperature
range (-18° to 55° C) on
actual calibration standards,
for 24-72 hours

Eliminates weaker component


parts prior to shipment

Assures “out-of-the-box”
reliability / Assures
performance in outdoor
installations
DMC / GC.ppt slide 7 July 01
Product Overview
Model 2350 GC Controller

NEC/CEC Standards Explosion-proof,


Class 1 Div. 1
IEC Standards-EEx d IIC T4 SUITABLE
FOR ZONE 1
19” Rack-mount -- with and without
Integral Field mount with Keypad and
Keypad/Display
Display
Complies with EMC Directive
Complies with EMC Directive

DMC / GC.ppt slide 8 July 01


Model 2350
Features
Controls up to 12 streams - 4 methods.
Four (4) serial ports for convenient communication to multiple hosts or
flow computers (232, 422, 485)
Up to 32 units on a Modbus “network” -- without additional peripherals
Dual Detector Processing capability for faster applications
Electronic data storage eliminates paper from printers and strip chart
recorders
Diagnostics - hardware and software checks
All data exported in comma delimited ASCII
Available as a retrofit to all 2251 controllers
Full remote diagnostics including chromatograms and raw data, with built
in optional modem

DMC / GC.ppt slide 9 July 01


MON 2000
Windows-Based GC Diagnostic Software

DMC / GC.ppt slide 10 July 01


MON 2000
Website / Internet access
•Website registration
directly from MON
2000.
•Notification of
engineering updates by
e-mail to registered
users
•Registered users
download software
revisions from website

DMC / GC.ppt slide 11 July 01


MON 2000
Windows-Based Polling

DMC / GC.ppt slide 12 July 01


MON 2000 Features
Easy to use PC software included with each GC at no
extra cost.
Runs previous MON versions -- DOS executable
files/applications (*.bin &*.cgm files).
Field Configurable Modbus Communications.
4 independent serial ports (RS, 232, 422, 485)..
Archives - Following information always in memory
Most recent 400 calibrations.
Most recent 1200 runs.
Most recent 64 days of 24 hr. averages.
Most recent chromatogram and calibration for each
stream.

DMC / GC.ppt slide 13 July 01


MON 2000
Benefits
Easy to use diagnostics simplify troubleshooting.
E-mail chromatograms with site specific comments for
peer review.
Eliminates PC and modem compatibility issues
associated with DOS.
Integral polling software can be easily set up for
intermittent or continuous polling.
Polling software saves data in industry standard file
formats for easy transfer to spreadsheets (.xls files)
word processors (.txt files ) or web browsers (html
files).

DMC / GC.ppt slide 14 July 01


Product Overview
Danalyzer Gas Chromatographs

The Danalyzer
Energy (BTU/CV) Measurement for
Natural Gas Pipelines

DMC / GC.ppt slide 15 July 01


Danalyzer BTU/CV Chromatographs
Model 570 & 565

Model 570 Model 565


with integral controller with rack-mount controller
3 valves 2 valves
(+/- .50 BTU,
. 4 min. analysis time) (+/- 1 BTU, 12 min. analysis)

DMC / GC.ppt slide 16 July 01


Danalyzer BTU/CV Chromatograph
Model 590

C9+ (BTU) alone or C6+(BTU) and H2S (0-30ppm)


Model 590 (+/- 0.50 BTU, 5 min. analysis time)
with integral controller
Dual Oven -- Natural Gas Applications

DMC / GC.ppt slide 17 July 01


Danalyzer CV Chromatograph
Model 571 (Cenelec Version)

•EEX d IIC T4 suitable for use in


Zone 1 hazardous areas

•CE approved -- complies with


EMC directive

•Approved for custody transfer in


all European Countries

•Integral x-proof controller or


general purpose 19” rack
mounted remotely in non-
hazardous areas.
.
DMC / GC.ppt slide 18 July 01
Danalyzer Gas Chromatograph
Features

+/-0.5 BTU (+/- 0.05%) repeatability over complete temperature range.


+/- .25 BTU (+/- .025%) repeatability in temperature controlled buildings.
Approved by CSA and CENELEC (Zone 1) for Class 1 Div 1 (Groups C & D).
Meets NEC, CEC & IEC electrical standards.
No instrument air required -- Solenoid actuation via Helium or Nitrogen.
Thermal Conductivity Detectors (TCD) with high resolution 16 bit A/D
conversion.
Each unit individually tested to ensure operation in ambient temperatures from
0-130° F (-18 to +55° C).
Multiple stream operation - up to 12 streams max.
Full Chromatogram Diagnostics for troubleshooting on-site or remote via
modem.
MON windows-based diagnostic software included.

DMC / GC.ppt slide 19 July 01


Danalyzer Gas Chromatographs
Benefits
“Legendary Performance” - “Lowest cost of Ownership”
Increased performance
Maximum custody transfer accuracy and assurance
Reduced installation costs
Building not required; minimal temperature control in cold climates.
Reduced utility costs
No instrument air required
Reduced maintenance costs
Sample valves warranted to operate 5+ years between service
QC/Environmental Chambering reduces probability of component failure
Reduced training costs
Use with 2350 and easy to use MON (Windows) software for local and full
remote diagnostics.

DMC / GC.ppt slide 20 July 01


Gas Chromatographs- Principle of Operation

The GC oven contains 3 main components mounted


together in an electrically heated heat-sink-oven:
GC Valves
Micro-packed Columns
TCD Detectors

DMC / GC.ppt slide 21 July 01


HEATSINK OVEN DETECTOR
AND VALVES HOUSING

DMC / GC.ppt slide 22 July 01


MEASUREMENT
DETECTOR

REFERENCE
DETECTOR

DMC / GC.ppt slide 23 July 01


COLUMNS
HEATSINK
OVEN

SAMPLE
LOOP
6 PORT
VALVE

DMC / GC.ppt slide 24 July 01


Gas Chromatographs- Principle of Operation

Step 1
Prior to sample Injection
Sample
Column 3
shut-off Valve

Sample Back Flush


Valve Valve Column 2
Sample Vent
Loop

Column 1
Vent Detector

Dual Column Valve

Helium Carrier gas

Analyser Oven

DMC / GC.ppt slide 25 July 01


Gas Chromatographs- Principle of Operation

Step 2
5 - 10 Seconds
Sample

Sample Column 3
shut-off Valve
Sample Back Flush N2
Valve Valve C1
Sample Column 2
Loop Vent

Column 1 C5,C4 Detector


Vent C8,C7,C 6 C3 ,C2, CO 2

Dual Column Valve


Helium Carrier gas

Analyser Oven

DMC / GC.ppt slide 26 July 01


Gas Chromatographs- Principle of Operation

Step 3
10 - 42 Seconds

Sample
Sample Column 3
shut-off Valve C2 CO2 C1 N2
Sample Back Flush
Valve Valve Column 2 Column 1
Vent

Vent C5,C4 C6 ,C7 ,C8


C3 Detector
Dual Column Valve

Helium Carrier gas

Analyser Oven
DMC / GC.ppt slide 27 July 01
Gas Chromatographs- Principle of Operation

Step 4
42 - 165 Seconds

Sample

Sample Column 3
shut-off Valve C2 CO 2 C 1 N 2
Sample Back Flush
Valve Valve Column 2 Column 1
Vent

Vent C5,C4
C3 Detector
Dual Column Valve

Helium Carrier gas

Analyser Oven
DMC / GC.ppt slide 28 July 01
Gas Chromatographs- Principle of Operation

Step 5
165 - 220 Seconds

Sample
Sample Column 3
shut-off Valve C2 CO2 C1 N 2
Sample Back Flush
Valve Valve Column 2 Column 1
Vent

Detector
Dual Column Valve
Vent

Helium Carrier gas

Analyser Oven
DMC / GC.ppt slide 29 July 01
Gas Chromatographs- Principle of Operation

DMC / GC.ppt slide 30 July 01


Gas Chromatographs- Principle of Operation

DMC / GC.ppt slide 31 July 01


Sample Conditioning and Sample Transport

The sample conditioning system is a very important


component in any analytical system. This is a
frequently overlooked component that must deliver a
representative sample from the pipeline to the GC
without changing the composition.

This is not a difficult task when low BTU gas is


measured in warm climates. When high BTU gas is
measured the potential exists to lose the heavier
components as they drop into the liquid phase due to
sample cooling.

DMC / GC.ppt slide 32 July 01


Sample Conditioning and Sample Transport

The Joules-Thompson (JT)


cooling effect causes heavier
components to drop into liquid
phase.

Will bias energy readings low


because the high BTU contributors
drop out of the vapor-phase
sample before the GC even has a
chance to measure it.

DMC / GC.ppt slide 33 July 01


Sample Conditioning and Sample Transport

Pressure Balancing
stream to stream.

Minimizes normalization.

Minimizes cross-stream
contamination if solenoids do
not seal completely.

DMC / GC.ppt slide 34 July 01


Sample Conditioning and Sample Transport

Calibration Standards
may require heated
insulation blankets when
installed in High BTU
applications in cooler
climates.

Sample transport lines


require heat tracing to a
heated area or all the way
to the sample valve.

DMC / GC.ppt slide 35 July 01


Sample Conditioning and Sample Transport

DMC / GC.ppt slide 36 July 01


Sample Conditioning and Sample Transport

DMC / GC.ppt slide 37 July 01


Sample Conditioning and Sample Transport

DMC / GC.ppt slide 38 July 01


Sample Conditioning and Sample Transport

DMC / GC.ppt slide 39 July 01


Key Products for Natural Gas Applications
Differentiators - Danalyzer Gas Chromatographs

Daniel 6-port GC valve guaranteed for 5 million operations


between rebuilds (lifetime warranty on valve)

Micro-packed columns improve baseline separation

Single or dual TCD detectors for applications up to C9+

Advanced Windows based software for extensive chromatogram storage

Environmental testing guarantees performance over -17ºC to 55ºC

Controller stores 1200 analyses and 400 calibrations

DMC / GC.ppt slide 40 July 01


Key Products for Natural Gas Applications
Gas Chromatographs

DMC / GC.ppt slide 41 July 01


RS232 from
chromatograph
Pressure
&
Temp Digital Connection to SCADA
including alarms and diagnostics

RS485 from
Ultrasonic Standard Volume Flow Rate
Energy Flow Rate
Total Volume & Energy

DMC / GC.ppt slide 42 July 01


Product Overview
Daniel Process Gas Chromatographs

Hydrocarbon Processing

DMC / GC.ppt slide 43 July 01


Product Overview
Model 500 Process GC

Sample System Oven (Close up) --


Heating up to 85C without instrument air.
Model 500 Ideally suited for outdoor installations close--
with lower sample system oven coupled to the sample point

DMC / GC.ppt slide 44 July 01


Model 500
Process Gas Chromatographs

• Installed cost is lower without the need


for expensive analyzer
buildings/heated sample line
• Ideal for NGL applications in Gas
Processing Plants NGL Pipelines
• Sample transport lines are shorter for
improved analytical speed
• Results:
• Improved Process Control
• Reduced Maintenance Costs

DMC / GC.ppt slide 45 July 01


Model 500 Gas Chromatograph
Features
CSA/CENELEC versions for Class 1 Div 1 (Groups C & D)
Heated (airless) sample oven to 85C
1 or 2 detector/oven combinations
No instrument air required (for operation or safety)
Thermal Conductivity Detectors (TCD)
Tested to ensure operation in ambient temperatures from 0-130° F
(-18 to +55° C)
12 streams max; Dual sample valve capability allows for improved
analysis of liquids and gases in the same sample system by grouping
similar streams.
Ideal for liquid fractionation facilities in Natural Gas Processing
Plants. Demethanizer overheads and bottoms, deethanizer
overheads and bottoms, depropanizer overheads and debutanizer
overheads

DMC / GC.ppt slide 46 July 01


Model 500 Gas Chromatograph
Benefits
Superior stability/accuracy:
+/- 0.05 % Repeatability
Complete component separation
Rugged and Reliable:
Environmental Chambering reduces
field failures, and installation costs
(shelters, sample line, or A/C units,
etc..)
5 Million + valve operations
between service
Versatile and Efficient:
Single unit can measure both gas Field-mounted Model 500
and liquid phase samples (Gas Processing Plant)
Dual detectors/oven applications “Legendary Performance”
for speed or complex analysis “Lowest Cost of Ownership”

DMC / GC.ppt slide 47 July 01


Model 1000
Process Gas Chromatographs

Model 1000 Air bath Oven


Model 1000
with integral controller/keypad/display

DMC / GC.ppt slide 48 July 01


Model 1000 Gas Chromatographs
Detectors

Flame Ionization Detector


for trace hydrocarbon analysis.

Thermal Conductivity Detector


for multi-component analysis.

Flame Photometric detector


for sulfur component analysis.

DMC / GC.ppt slide 49 July 01


Model 1000 Gas Chromatograph
Features

Air bath oven - 150° C max


Integral Controller with keypad/display
Detectors: TCD, FID, FPD
Up to 6 analytical valves max
6 Port diaphragm or liquid injections valve
Environmental Chamber testing standard (35-130° F)
12 streams max
Applications include liquid streams, depropanizer bottoms
and debutanizer bottoms (natural gasolines). Also MTBE,
aromatics, olefin gas or other liquid phase samples
requiring higher GC oven temperatures.

DMC / GC.ppt slide 50 July 01


Model 1000
Benefits
Proven Performance & Reliability
Same high-reliability valve/QC as Danalyzer
Versatility/Application flexibility
Multiple detectors in single or dual configurations
Oven temperature to 150 C
Liquids and Gases in the same GC with integral Sample
Conditioning System.
Reduced training costs - Uses same 2350 controller
SW/Menu/MON as all other Daniel GCs
Single source GC solution
Pipeline/Petrochemical/Chemical Processing

DMC / GC.ppt slide 51 July 01


Product Support

24 Hr. Telephone or PC based service


Field Service
Spare parts delivery
Maintenance contracts
Worldwide/Regional Service
Daniel USA- Houston, Tulsa, New Orleans, Lafayette,
Midland
Daniel Canada- Calgary
Daniel UK- Larbert Scotland (Applications & Service)
Daniel Asia- Singapore
Latin America- Regional service through authorized
representatives

DMC / GC.ppt slide 52 July 01