On-Line Gas Chromatographs
On-Line Gas Chromatographs
On-Line Gas Chromatographs
Assures “out-of-the-box”
reliability / Assures
performance in outdoor
installations
DMC / GC.ppt slide 7 July 01
Product Overview
Model 2350 GC Controller
The Danalyzer
Energy (BTU/CV) Measurement for
Natural Gas Pipelines
REFERENCE
DETECTOR
SAMPLE
LOOP
6 PORT
VALVE
Step 1
Prior to sample Injection
Sample
Column 3
shut-off Valve
Column 1
Vent Detector
Analyser Oven
Step 2
5 - 10 Seconds
Sample
Sample Column 3
shut-off Valve
Sample Back Flush N2
Valve Valve C1
Sample Column 2
Loop Vent
Analyser Oven
Step 3
10 - 42 Seconds
Sample
Sample Column 3
shut-off Valve C2 CO2 C1 N2
Sample Back Flush
Valve Valve Column 2 Column 1
Vent
Analyser Oven
DMC / GC.ppt slide 27 July 01
Gas Chromatographs- Principle of Operation
Step 4
42 - 165 Seconds
Sample
Sample Column 3
shut-off Valve C2 CO 2 C 1 N 2
Sample Back Flush
Valve Valve Column 2 Column 1
Vent
Vent C5,C4
C3 Detector
Dual Column Valve
Analyser Oven
DMC / GC.ppt slide 28 July 01
Gas Chromatographs- Principle of Operation
Step 5
165 - 220 Seconds
Sample
Sample Column 3
shut-off Valve C2 CO2 C1 N 2
Sample Back Flush
Valve Valve Column 2 Column 1
Vent
Detector
Dual Column Valve
Vent
Analyser Oven
DMC / GC.ppt slide 29 July 01
Gas Chromatographs- Principle of Operation
Pressure Balancing
stream to stream.
Minimizes normalization.
Minimizes cross-stream
contamination if solenoids do
not seal completely.
Calibration Standards
may require heated
insulation blankets when
installed in High BTU
applications in cooler
climates.
RS485 from
Ultrasonic Standard Volume Flow Rate
Energy Flow Rate
Total Volume & Energy
Hydrocarbon Processing