MOVING IRON (MI) INSTRUMENTS

Where ruggedness is more important than high degree of accuracy,


moving iron instruments are used.
Used for measurement of AC and DC
• Types of moving iron instruments:
1. Attraction type
2. Repulsion type

https://www.youtube.com/watch?v=7QDMrj0r
8Mc
PRINCIPLE OF OPERATION
• Construction of the attraction type moving iron
instrument is as shown in the given figure.
• The moving iron, i.e. the disc of soft iron, is
eccentrically mounted. Coil is situated around the disc.
When the coil is excited it produces magnetic field.
• Due to magnetic field the moving iron moves from the
weaker field outside the coil to the stronger field inside
the coil. Thus moving iron gets attracted inwards and
thus the name attraction type.
Continued…
• The controlling torque is provided by the balance weights attached to
the moving iron. Spring also can be used to provide controlling torque.
• Damping is provided by air friction in which aluminum piston is
attached to the moving system and moves in a closed air damping
chamber.
ATTRACTION TYPE MOVING IRON INSTRUMENT

Control Mechanism
PRINCIPLE OF OPERATION
PRINCIPLE OF OPERATION
PRINCIPLE OF OPERATION
REPULSION TYPE MOVING IRON INSTRUMENT
REPULSION TYPE MOVING IRON INSTRUMENT

•In these type of instruments there two vanes present inside


the coil in which one is fixed while other is movable. These
both vanes gets similarly magnetized when coil is excited.
Thus there is a force of repulsion between both vanes causing
movement in movable vane.
•The controlling torque is provided by the springs or
balancing weights (in vertically mounted instrument). The
damping is provided by air friction damping same as in
attraction type moving iron instruments.
ADVANTAGES

1. Suitable of a.c. as well as d.c. measurements.


2. Good accuracy.
3. Cheaper in cost as compared to permanents magnet moving coil
instruments.
4. The instrument has high torque to weight ratio.
5. The instrument can be used for low frequency measurement also.
DISADVANTAGES

1. Power consumed by the instrument is high as compared to that of


the permanent magnet moving coil instrument.
2. The scale is non-uniform.
3. Due to the presence of iron part in the operating system, error due to
the hysteresis effect is introduced. To reduce this effect nickel iron
alloys are used.