Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, VOL. 61, NO. 3, MARCH 2012 729 A Low Power CMOS Voltage Regulator for a Wireless Blood Pressure Biosensor Paulo Cesar Crepaldi, Tales Cleber Pimenta, Robson Luiz Moreno, and Edgar Charry Rodriguez Abstract—This paper describes a CMOS implementation of a linear voltage regulator (LVR) used to power up implanted physiological signal systems, as it is the case of a wireless blood pressure biosensor. The topology is based on a classical structure of a linear low-dropout regulator. The circuit is powered up from an RF link, thus characterizing a passive radio frequency identification (RFID) tag. The LVR was designed to meet important features such as low power consumption and small silicon area, without the need for any external discrete components. The low power operation represents an essential condition to avoid a high-energy RF link, thus minimizing the transmitted power and therefore minimizing the thermal effects on the patient’s tissues. The project was implemented in a 0.35-µm CMOS process, and the prototypes were tested to validate the overall performance. The LVR output is regulated at 1 V and supplies a maximum load current of 0.5 mA at 37 ◦ C. The load regulation is 13 mV/mA, and the line regulation is 39 mV/V. The LVR total power consumption is 1.2 mW. Index Terms—Implanted device, linear voltage regulator (LVR), low power, pressure biosensor, wireless biomedical device. I. I NTRODUCTION ATIENT monitoring systems can be found in a wide range of application in hospitals, including intensive care units. Once the equipment is connected to a communication network, it forms a telemedicine system in which the patients can be monitored remotely, even over the Internet, thus indicating the portability of these instruments [1]–[3]. The interaction between medicine and technology, as it is the case of microelectronics and biosensor materials, allows the development of diagnosing devices capable of monitoring pathogens and diseases. Once the whole circuitry can be placed directly on the patient or even implanted, it becomes a lab-on-chip and pointof-care device [4]. Since the implanted device becomes part of a biological data acquisition system (biotelemetry), it must meet important constraints, such as reduced size, low power consumption, and the possibility of being powered by an RF link, thus operating as a passive RFID tag [5]. A typical CMOS front-end architecture for an in vivo biomedical implanted device (BID) is shown in Fig. 1. The system consists, basically, of the sensitive biological element, the transducer or detector element, the associate electronics and signal P Manuscript received July 19, 2011; revised September 18, 2011; accepted September 19, 2011. Date of publication November 16, 2011; date of current version February 8, 2012. This work was supported in part by CAPES, by MOSIS, by CNPq, and by FAPEMIG. The Associate Editor coordinating the review process for this paper was Dr. Daryl Beetner. P. C. Crepaldi, T. C. Pimenta, and R. L. Moreno are with the Federal University of Itajuba, Itajuba-MG 37500-903, Brazil. E. C. Rodriguez is with the Universidade de São Paulo, São Paulo-SP 05508-010, Brazil. Digital Object Identifier 10.1109/TIM.2011.2172121 processors, and the RF link to establish a communication with the manager unit. The combination of the implanted device, the local wireless link, and a communication network results in a wireless biosensor network [6]. Among physical parameters, the pressure in various organs and body parts, such blood vessels, heart, brain, eyes, etc., is an important indication of some diseases like glaucoma, hypertension, and neurogenic bladder dysfunction. An implantable telemetric pressure sensor is a good solution for short-term application to avoid percutaneous connections through the patient’s skin (risk of infections) and to give him free movement for other clinical settings. This paper presents the development and implementation of a linear voltage regulator (LVR), as part of the associate electronics of a blood pressure BID [7]. The LVR must provide the power supply voltage for the system, as illustrated in Fig. 2. Variations on the input line voltage, load current fluctuation, and temperature variations may cause the circuit to deviate from its optimum operation bias point and may even loose its linearity. Therefore, the power supply system must assure minimum impacts on the linearity due to those variations. The impact of temperature variations in implantable devices is minimized since the body temperature is kept stable at approximately 37 ◦ C [8]. In the next sections, a more detailed description of the LVR circuit topology will be presented, as well as some considerations about the safety regarding the RF power link [7]. II. S AFETY C ONSIDERATIONS The user of an implantable blood pressure monitoring system is exposed to a source of RF radiation near the skin, and therefore, few safety considerations must be taken into account. The main biohazards and risks due to the RF exposure are mainly the heating from the electromagnetic field distribution on biological tissues [9]. In our work, the frequency operation of the RF link was fixed into 10 MHz since it provides a good compromise between power levels and human tissue penetration [10]. The specific absorption rate (SAR) represents a direct measurement of both the electric field and induced current density over the human tissue. The temperature variation over the time indicates the local heating factor. Both relations are given by [11] 0018-9456/$26.00 © 2011 IEEE SAR = σ|E|2 ρ dT SAR = dt c [W/Kg] [◦ C/s] (1) (2) 730 IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, VOL. 61, NO. 3, MARCH 2012 Fig. 1. Simplified diagram of an RFID BID. Fig. 2. Block diagram of the blood pressure monitoring system. where E is the incident electric field intensity (rms) and σ, ρ, and c represent the conductivity, the human tissue mass density, and the specific heat capacity, respectively, at the implant location. Based on expressions (1) and (2), a safe value for the power transferred by the RF link is 10 mW/cm2 [12]. The LVR power dissipation should be taken as just 10% of it, corresponding to 1 mW, which represents twice as much as required by the subsystems (0.5 mW). Reported voltage regulators for implanted devices list a power dissipation range that can be as high as tens of milliwatts [13]. The circuit topology and its components will be detailed next, as well its frequency stability, which reveals an unconditional stable system. III. LVR C IRCUIT T OPOLOGY The classic topologies designed to provide stable power supply voltage are the linear and the switched regulators. Switched regulators present a complex topology, mainly due to its control systems, and generally require more power consumption and larger silicon area. Additionally, it provides more noise at the regulated output due to its switched operation [14]. The low-dropout (LDO) voltage regulator is one of the most popular power converter used in power management [15], [16] and is more suitable for implanted systems. In the linear regulator, the output voltage is monitored and compared to a voltage reference through an error amplifier [usually an operational transconductance amplifier (OTA)]. The amplifier output acts directly on the pass element in such a way that, during its normal operation, there should be no fluctuations on the output voltage. The pass element can be implemented using bipolar or MOS transistors. Since a MOS transistor is controlled by its gate voltage, it offers the advantage of smaller power consumption and, consequently, higher efficiency for the voltage regulator. The MOS transistor can be either N or P type. The NMOS transistor requires a gate voltage higher than the source voltage, and therefore, a charge pump may be necessary to increase the voltage level. The proper choice for low-voltage systems, such as implantable devices, is the use of a PMOS LDO [17], [18] since its gate voltage is smaller than the source voltage. An NMOS LDO without charge pump is reported in [19] using native transistors (zero threshold) and an internal capacitor to improve the stability, at the expense of two external capacitors. Fig. 3 illustrates the contribution presented in this work, by the introduction of a source follower stage (transistor MNFOL ) and the alternative implementation of the resistive sampler (R1 and R2 ). The resistive sampler was replaced by a single MOS transistor along with a grounded MOS resistor (biasing from transistor MNAUX ). In the proposed circuit, the load capacitance (CL ) represents the internal capacitance of the subsystems that are powered up by the regulator (blood pressure microsensor, signal conditioner, processing unit, and transmitter). The input voltage comes from the RF link after rectification and filtering. Finally, the current source IL represents the overall current consumption of the powered subsystems. The nominal values adopted for the design are based on the actual subsystem (blood pressure microsensor, signal conditioner, processing unit, and transmitter) requirements. The supply voltage goal is 1 V for a maximum current load of 0.5 mA. Consequently, the total power required CREPALDI et al.: LOW POWER CMOS VOLTAGE REGULATOR FOR A WIRELESS BLOOD PRESSURE BIOSENSOR Fig. 3. Simplified diagram of the proposed LVR. by the remaining subsystems is 0.5 mW. The source follower stage (transistor MNFOL ) was added in order to provide stability to the LVR system. The rectifier, voltage references, OTA, and the sampler will be discussed next. IV. F RONT-E ND I NTERFACE The front-end interface is presented is Fig. 4. The rectifier was implemented in the chip. The use of regular diodes is prohibitive due to the high frequency circuit operation. The NMOS rectifier shows a good performance and has the advantage of being totally compatible with the CMOS digital technology [20]. The cross gate rectifier structure suppresses the threshold voltage drop of the switching transistors (MN1 and MN2 ). Additionally, the drain–source voltage VDS(ON) can be minimized by proper W/L aspect ratio. Transistors MN3 and MN4 cause the larger voltage drop since they operate as diodes. Nevertheless, since they are NMOS, their voltage drop is smaller than that of PMOS. In the technology used in this project, the threshold voltage is approximately 520 mV. This smaller voltage represents an improvement on the rectifier efficiency [21]. The average voltage at the rectifier output (VRET ), which is the LVR input voltage (VIN ), is fixed in 2.2 V in order to operate the voltage references circuit, to be presented in Section VI. The capacitive filter (CFILT ) was implemented using a poly–insulator–poly capacitor and was designed to provide a ripple voltage of ≈ ±10% (2 V a 2.4 V) at a maximum load current condition of 0.5 mA. Fig. 5 shows the simulation results for the rectifier, and Table I compares with measured values. V. OTA The OTA must present few optimized features to improve the overall LVR performance. It must present a low offset 731 voltage (ViO ) since it affects directly the output voltage through the control loop, as it can be modeled as a voltage source in series with VREF . Additionally, ViO must present a low thermal drift, even though it can be discarded in this project due to the fairly stable human temperature. The power supply rejection ratio (PSRR) must be maximized to avoid propagating unregulated power supply ripple to the output. The open-loop gain (AOL ) must be maximized to assure proper closed-loop operation. In this design, the OTA load is purely capacitive, represented by the input gate capacitance of the pass transistor MPPASS . The OTA circuit is presented in Fig. 6, and it is implemented as a self-biased folded cascode [22]–[25]. The cascode topology has the advantages of improved PSRR and presents a dominant pole that is determined only by the load capacitance. The self-biased cascode also represents a good solution to minimize silicon area since it does not need any additional biasing circuitry. The power consumption of the OTA is kept as low as possible to avoid degrading the efficiency of the LVR. The OTA layout was carefully developed using a centroid configuration to minimize any offset voltage. Table II summarizes the main OTA-measured parameters. The load capacitance (CL ≈ 30 pF) represents the sum of all capacitances at node OUT(OTA) . The rail-to-rail topology needs an extra circuitry to compensate the gm variations due the PMOS and NMOS differential input pairs. The frequency response of the OTA is directly proportional to gm. However, as will be seen later on item 8, evaluating the frequency response of the LVR, these variations are not so significant provided that a pole splinting locates the dominant pole and unit frequency gain at least three decades before the other poles of the system. The advantage of the absence of the gm compensation circuitry is less silicon area and additional power dissipation. VI. LVR VOLTAGE R EFERENCES The LVR requires two voltages references, VG and VREF . The VG reference is necessary to bias the source follower stages represented by MNFOL and MNAUX . The VREF reference is used, on the closed-loop system, to obtain the LVR output   R1 VOUT = 1 + (3) VREF . R2 Bandgap references are generally used in voltage regulator designs. There are alternative circuits capable of obtaining low voltage and high accuracy; nevertheless, those approaches may require components not readily available in CMOS technology (additional fabrications steps), or CMOS transistor on subthreshold operation, or floating gate or even need of trimming [26]–[28]. Since the circuit is intended to be used in an implanted device, the temperature range is very small, and therefore, it is not taken into account. The reference voltage sensitivity assumes the major factor quality. Fig. 7 presents the proposed topology that is simple, requires small silicon area (0.003 mm2 ), and requires just a few microwatts. The core of this reference circuit is the self-biased current mirror composed by MN1 , MP1 , MP2 , and Q1 . The use of 732 IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, VOL. 61, NO. 3, MARCH 2012 Fig. 4. LVR front-end interface. TABLE II OTA-MEASURED PARAMETERS Fig. 5. Rectifier simulation results for average value and ripple. TABLE I RECTIFIER : S IMULATED AND M EASURED VALUES Fig. 7. Proposed voltage reference circuit. the parasitic PNP bipolar transistor Q1 in a CMOS digital technology is justified since it offers a well-known VBE voltage (approximately 0.676 V for this circuit) and its corresponding temperature behavior. To obtain higher voltage VG , other current mirrors and successive cascade of bipolar transistor are used. A 2-V value is needed to bias the source follower stage (MNFOL ) and the auxiliary voltage follower MNAUX . This voltage reference is obtained by VG = VEB (Q2 )+VEB (Q3 )+VEB (Q4 ) = 3VEB (Q1 ) ≈ 2[V]. (4) Fig. 6. OTA circuit. Voltage VREF is derived from VBE of transistor Q2 and uses a composite structure of two MOS transistors, MN2 and CREPALDI et al.: LOW POWER CMOS VOLTAGE REGULATOR FOR A WIRELESS BLOOD PRESSURE BIOSENSOR 733 The self-biased current mirror was designed to operate with a 4-µA base current, which is the MN1 transistor drain current. Since the current mirrors are implemented by transistors MP3 , MP4 , MP5 , and MP6 (current gain of 1), the total current of the reference generator circuit is 24 µA. Table III presents a comparison of this work with others presented in the literature. Since the circuit is intended to be used in an implanted device, the temperature variation is negligible, and therefore, it is not taken into account. The reference voltage sensitivity assumes the major quality factor. The reference voltage sensitivities (considering the quiescent point Q) of VG and VREF are given respectively by REF SV VIN = VIN VREF Q ∂VREF ∂VIN Q G and SV VIN = VIN VG Q ∂VG ∂VIN . Q (7) The evaluated derivatives for VG and VREF are given as Fig. 8. UT λN IDQ λN ∂VREF G = and SV VIN =3 2UT 2UT ∂VIN 1− 1− VEB−VTHO(N) (VEB−VTH0(N) ) Composite structure of two NMOS transistors to generate VREF . (8) Fig. 9. Adjustment of the aspect ratio of MNREF1 by simulation. MN3 . The composite MOS transistor improves the PSRR of the voltage reference [29]. This topology is shown in Fig. 8. Although the NMOS transistors, MNREF1 and MNREF2 , carry the same current ID , they are operated in weak inversion and strong inversion, respectively. This condition leads to ID (MNREF1 ) = ID (MNREF2 )     VEB − VREF − VTH0(N) W IX exp L 1 nUT  2 = βN VEB − VTH0(N) (1 + λN VREF ) . where UT is the temperature equivalent voltage (kT/q), λN is the channel modulation factor, and VTHO(N) is the threshold voltage of the MOS transistor without the body effect factor. It can be observed that VG voltage sensitivity is multiplied by three due to the three voltages VBE added up along the way of transistors Q2, Q3, and Q4. Table IV provides a comparison of the sensitivity obtained by using expressions (5) and (6), and the measured values. The self-biased current mirror requires a start-up circuitry, that it is implemented through capacitor CSTART and transistor MSTART , as shown in Fig. 10. Simulation results (see Fig. 11) show that the current ID (MSTART ) has a peak value of 160 µA and vanishes to almost zero in less than 10 ns. Therefore, the reference circuit is led to its operating point and avoids the undesirable zero-state condition. Figs. 12 and 13 show respectively the step response of the two voltage references. As can be seen, their final values are 201.2 mV and 2.047 V that represent a small deviation from the target values. VREF has an overshot with a duration of approximately 1 µs, and VG takes 7 µs to reach its final value. VII. S AMPLER C IRCUIT (5) By neglecting the term λN VREF , (5) can be rewritten as   2 βN VEB −VTH0(N)   VREF = VEB −VTH0(N) −nUT ln . IX W L 1 (6) VREF is adopted as 200 mV. The aspect ratio of MNREF1 is adjusted by simulation to achieve that value, as illustrated in Fig. 9. Fig. 14 presents the sampler circuit. Since resistors are avoided, mainly to save silicon area, R1 is implemented as a MOS diode (transistor MN2 ), and R2 is implemented by grounding a MOS resistor [33]. The use of the source follower transistor MNAUX guarantees that the MOS resistors are isolated from VIN , thus avoiding a significant transference of ripple voltage to the output line. MNAUX also imposes a smaller effective voltage to the MOS resistor, thus reducing the sampling current. The relationship R1 /R2 is optimized by the adjustments of the aspect ratio of transistor MN2 . The sampler circuit current IRES is designed to be ≈1% of the maximum current load (≈5 µA). 734 IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, VOL. 61, NO. 3, MARCH 2012 TABLE III COMPARISON W ITH OTHER R EFERENCES TABLE IV REFERENCE VOLTAGE S ENSITIVITY Fig. 12. Voltage reference VREF step response. Fig. 13. Voltage reference VG step response. Fig. 14. Sampler circuit. Fig. 10. Inset of start-up circuit. Fig. 11. Simulation of the start-up current. By using the following relations, it is possible to determine the values of R1 and R2 :   R1 R1 =4 (9) 1V = 1 + 0, 2V ⇒ R2 R2 1V = IRES (R1 + R2 ) ⇒ R1 + R2 = 200 KΩ. (10) The equation system composed of (9) and (10) is solved to R1 = 160 KΩ and R2 = 40 KΩ. CREPALDI et al.: LOW POWER CMOS VOLTAGE REGULATOR FOR A WIRELESS BLOOD PRESSURE BIOSENSOR 735 Fig. 15. Classic PMOS LDO with discrete frequency compensation scheme. The aspect ratio of MN1 was adjusted in order to set IRES as close as to the target value of 5 µA. The PMOS array and transistor MNAUX have also their aspect ratio adjusted by simulation in order to minimize current consumption. The measured current consumption of the PMOS transistor array is 2.94 µA, and the current IRES is 5.2 µA. Thus, the equivalent resistor R2 is approximately 38.5 KΩ, and R1 is approximately 153.8 KΩ. Fig. 16. Frequency response of a PMOS LDO regulator with external compensation capacitor. VIII. S TABILITY A NALYSIS The use of an LDO circuit requires a stability analysis since it forms a closed-loop system. The frequency response is degraded by the presence of two poles in addition to the dominant pole that can lead an unstable system condition. It is necessary to add a zero between these two poles to achieve frequency compensation. The insertion of this zero is normally done by adding a discrete electrolytic capacitor (CCOMP ) at the output node that also contributes with an additional resistance RESR (see Fig. 15). Also, ROTA is the output resistance of the transconductance amplifier, CGPASS is the gate capacitance of the PMOS pass transistor, and RDS is the channel resistance of the PMOS pass transistor. The frequencies of these poles and zero are given by [34] fP0 = −1 −1 ≈ 2π(RDS + RESR ).CCOMP 2πRDS CCOMP (11) fP1 = −1 −1 ≈ 2π(RDS //RESR )CL 2πRESR CL (12) −1 2πRESR CCOMP −1 . = 2πROTA CGPASS fZ0 = (13) fP2 (14) Equation (11) shows that the frequency of the dominant pole depends on the drain–source resistance, which in turn depends on the drain current. As a consequence, the dominant pole can change its position according to the load. To overcome this situation, the zero must follow the pole. It is common to establish not only a single value for RESR but a range of values as a function of load current. Fig. 16 presents the frequency response of a PMOS LDO. The use of an external capacitor, such as an electrolytic capacitor, is prohibitive for an implantable device. The literature provides many contributions to solve the LDO stability Fig. 17. Proposed topology to set a dominant pole in the LVR system. problem. Few approaches maintain the external capacitor and modify the internal feedback loop by using buffers [35] and Miller compensation capacitor [36]. Other approaches insert an internal zero, discarding the compensation capacitor, by using controlled sources and even Miller compensation [37]. The solution proposed in this work is the introduction of a source follower (MNFOL ) stage in between the input voltage and the LDO block, and the removal of the compensation capacitor CCOMP , as shown in Fig. 17. This stage maintains the PMOS pass transistor in the triode region (small values for VDS ). To investigate the stability of the regulator, an open-loop analysis is necessary to determine the loop gain Aβ. Fig. 18 shows the regulator with open feedback loop point, and in Fig. 19, the OTA and the MPPASS are replaced by their smallsignal counterparts. The loop gain can be stated by Aβ = − vr . vx (15) The total load resistance is minimized by the low value of rds , so the drain–gate voltage gain of MPPASS is K=− vout = −gmPASS rDS . vgs (16) 736 IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, VOL. 61, NO. 3, MARCH 2012 Fig. 20. Frequency response of the proposed LVR. Fig. 21. Microphotography of the chip. Fig. 18. Proposed voltage regulator highlighting the open-loop point. Fig. 19. Small signal circuit to determine the loop gain. The output voltage is vout = − gmPASS .rds gmOTA ro 1+ S p1 S p2 1+ vx . (17) Moreover, vr , considering that rid is much larger than R2 , is vr = vout R2 . R1 + R2 (18) Combining (17) and (18), the loop gain is Aβ = gmPASS .rds gmOTA ro 1+ S p1 1+ S p2 R2 . R1 + R2 (19) IX. LVR M EASUREMENTS By properly adjusting the transistor aspect ratio, rds can be designed to be smaller than resistors R1 and R2 of the sampling circuit. rds was chosen to be less than 100 Ω for our LVR. The system poles, under the new approach, are p1 = 1 and (Cgd + CL )rds p2 = 1 . [Co + Cgs + Cgd (1 + gmPASS rds )] ro The derived values for fP0 and fP2 are 318 MHZ and 133 HZ , respectively, resulting in a unit frequency gain fUG of 620 KHZ . Fig. 20 presents the frequency response of the proposed LVR. The source follower stage is disadvantageous for the system since it represents additional power consumption. Nevertheless, the system does not require the external components to become stable, thus representing an important advantage for an implantable system. (20) The pole p2 is the dominant one since ro , in the range of megohms, can be at least 104 times greater than rds that is in the range of a tenth of an ohm. Therefore, the frequency stability of the regulator is a function of the OTA design, the geometric aspect ratio of MPPASS , and the load. As an application specific integrated circuit (ASIC) application, the load current (IL ), resistance (RL ), and capacitance (CL ) can be stated as constants. The LVR was implemented in a Taiwan Semiconductor Manufacturing Company 0.35-µm CMOS technology through MOSIS. Fig. 21 shows the microphotography of the prototype, and Fig. 22 shows the basic circuit used to characterize the most important features. The antennas are artisanal and are kept 5 cm away from each other. The transmitter is implemented by an RF generator, and the chip input voltage is monitored to avoid exceeding 3.6 V (maximum value allowed by the technology). Table V summarizes the main parameter measurements on the LVR. The dynamic behavior of the LVR is depicted in Fig. 23 where an input step signal is applied at a load current of 0.5 mA and a capacitive load of 30 pF. It can be seen that the LVR output voltage is stable, without any overshoot, with a setup time of just 16 µs. The transient response of the LVR circuit indicates a bounded-inputbounded-output system. Table VI shows a comparison of our work with previous reported regulators. CREPALDI et al.: LOW POWER CMOS VOLTAGE REGULATOR FOR A WIRELESS BLOOD PRESSURE BIOSENSOR 737 Fig. 22. Basic circuit for the LVR measurements. TABLE V LVR-MEASURED PARAMETERS TABLE VI COMPARISON W ITH P REVIOUS R EPORTED R EGULATORS Fig. 23. Step function response at IL = 0,5 mA and T = 37 ◦ C. A set of 5000 runs of Monte Carlo simulation was conducted to evaluate the impact of process variation on the LVR output voltage, and the result is presented in Fig. 24. The VOUT mean value is 0.992 V, and the standard deviation is 80.6 mV. X. C ONCLUSION This paper has presented an LVR for physiological monitoring systems that is powered by an RF link. The regulator was Fig. 24. Monte Carlo analysis of the LVR output voltage. designed specifically to meet important requirement conditions, such as low power consumption, low silicon area, and the absence of external discrete components. The output voltage is regulated to 1 V at a maximum load current of 0.5 mA, and the dissipated power is 1.2 mW. The load regulation is 13 mV/mA, and the line regulation is 39 mV/V. The dynamic behavior shows an unconditionally stable response. 738 IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, VOL. 61, NO. 3, MARCH 2012 R EFERENCES [1] W. G. Scanlon, N. E. Evans, G. C. Crumley, and Z. M. McCreesh, “Low power radio telemetry: The potential for remote patient monitoring,” J. Telemed. Telecare, vol. 2, no. 4, pp. 185–191, 1996. [2] R. Puers, in Proc. Implantable Sens. Syst., DISens Symp. Book, 2005, vol. 2.6, pp. 1–14. [3] M. Miyazaki, “The future of e-Health—Wired or not wired,” in Proc. Business Briefing—Hospital Engineering & Facilities Management, 2005, pp. 1–5. [4] J. Colomer-Farrarons, P. Miribel-Catala, I. Rodriguez, and J. Samitier, “CMOS front-end architecture for in-vivo biomedical implantable devices,” in Proc. 35th Annu. Conf. IECON, pp. 4401–4408. [5] J. Landt, “The history of RFID,” IEEE Potentials, vol. 24, no. 4, pp. 8–11, Oct./Nov. 2005. [6] M. Guennoun, M. Zandi, and K. El-Khatib, “On the use of biometrics to secure wireless biosensor networks,” in Proc. 3rd Int. Conf. Digit. Idenfier ICTTA, pp. 1–5. [7] P. C. Crepaldi, T. Pimenta, R. Moreno, and E. Rodriguez, “A CMOS linear power supply for a wireless biomedical sensor,” in Proc. IEEE Int. Workshop MeMeA, 2010, pp. 97–101. [8] P. A. Mackowiak, S. S. Wasserman, and M. M. Levine, “A critical appraisal of 98.6 ◦ F, the upper limit of the normal body temperature, and other legacies of Carl Reinhold,” August Wunderlich. JAMA, vol. 268, no. 12, pp. 1578–1580, Sep. 23–30, 1992. [9] J. M. Osepchuk and R. C. Petersen, “Safety standards for exposure to RF electromagnetic fields,” IEEE Microw. Mag., vol. 2, no. 2, pp. 57–69, Jun. 2001. [10] C. Sauer, M. Stanacevic, G. Cauwenberghs, and N. Thakor, “Power harvesting and telemetry in CMOS for implanted devices,” IEEE Trans. Circuits Syst. I, Reg. Papers, vol. 52, no. 12, pp. 2605–2613, Dec. 2005. [11] A. Pradier, D. Lautru, M. F. Wong, H. V. Fouad, and J. Wiart, “Rigorous evaluation of Specific Absorption Rate (SAR) induced in a multilayer biological structure,” in Proc. Eur. Conf. Wireless Technol., Oct. 3/4, 2005, pp. 197–200. [12] G. Lazzi, “Thermal effects of bioimplants,” IEEE Eng. Med. Biol. Mag., vol. 24, no. 5, pp. 75–81, Sep./Oct. 2005. [13] C. Zheng and D. Ma, “Design of monolithic low dropout regulator for wireless powered brain cortical implants using a line ripple rejection technique,” IEEE Trans. Circuits Syst. II, Exp. Briefs, vol. 57, no. 9, pp. 686–690, Sep. 2010. [14] G. A. Rincon-Mora and P. E. Allen, “Study and design of low dropout regulators,” Sch. Elect. Comput. Eng., Georgia Inst. Technol., Atlanta, GA. [15] G. A. Rincon-Mora, “Active multiplier in Miller-compensated circuits,” IEEE J. Solid-State Circuits, vol. 35, no. 1, pp. 26–32, Jan. 2000. [16] G. A. Rincon-Mora and P. Allen, “A low-voltage, low quiescent current, low drop-out regulator,” IEEE J. Solid-State Circuits, vol. 33, no. 1, pp. 36–44, Jan. 1998. [17] T. Kugelstadt, “Fundamental theory of PMOS low-dropout voltage regulators,” Texas Instrum. Inc., Dallas, TX, Application Note SLVA068, Apr. 1999. [18] C. Simpson, “A user’s guide to compensating low-dropout regulators,” in Proc. Wescon, Nov. 4–6, 1997, pp. 270–275. [19] M. M. Ahmadi and G. A. Jullien, “A wireless implantable microsystem for continuous blood glucose monitoring,” IEEE Trans. Biomed. Circuits Syst., vol. 3, no. 3, pp. 169–180, Jun. 2009. [20] M. Ghovanloo and K. Najafi, “Fully integrated wideband high-current rectifiers for inductively powered devices,” IEEE J. Solid-State Circuits, vol. 39, no. 11, pp. 1976–1984, Nov. 2004. [21] J. Hu and H. Min, “A low power and high performance analog front end for passive RFID transponder,” in Proc. 4th IEEE Workshop Autom. Identification Adv. Technol., Oct. 17/18, 2005, pp. 199–204. [22] B. G. Song, O. J. Kwon, I. K. Chang, H. J. Song, and K. D. Kwack, “A 1.8 V self-biased complementary folded cascode amplifier,” in Proc. 1st IEEE ASICs, AP-ASIC , Aug. 23-25, 1999, pp. 63–65. [23] P. Mandal and V. Visvanathan, “Design of high performance two stage CMOS cascode op-amps with stable biasing,” in Proc. 9th Int. Conf. VLSI Design, Jan. 3–6, 1996, pp. 234–237. [24] P. Mandal and V. Visvanathan, “A self biased high performance folded cascode CMOS op-amp,” in Proc. 10th Int. Conf. VLSI Design, Jan. 4–7, 1997, pp. 429–434. [25] V. Ceperic, Z. Butkovic, and A. Baric, “Design and optimization of self-biased complementary folded cascode,” in Proc. IEEE MELECON, May 16–19, 2006, pp. 145–148. [26] W. Jianping, L. Xinquan, L. Yushan, Z. Jie, and G. Xiaofeng, “A novel low-voltage low-power CMOS voltage reference based on subthreshold MOSFETs,” in Proc. 6th Int. Conf. ASIC, ASICON, Oct. 24–27, 2005, vol. 1, pp. 369–373. [27] B. K. Ahuja, H. Vu, C. A. Laber, and W. H. Owen, “A very high precision 500 nA CMOS floating-gate analog voltage reference,” IEEE J. SolidState Circuit, vol. 41, no. 10, pp. 572–573, May 12, 2005. [28] C. W. Chang, T.-Y. Lo, C.-M. Chen, K.-H. Wu, and C.-C. Hung, “A low power CMOS voltage reference circuit based on subthreshold operation,” in Proc. IEEE ISCAS, May 27–30, 2007, pp. 3844–3847. [29] L. H. C. Ferreira and T. C. Pimenta, “A week inversion composite MOS transistor for ultra-low-voltage and ultra-low-power applications,” in Proc. 13th Int. Conf. Mixed Design Integr. Circuits Syst. Gdynia, Jun. 2006, pp. 1–4. [30] A. Becker-Gomez, T. Lakshmi Viswanathan, and T. R. Viswanathan, “A low-supply-voltage CMOS sub-bandgap reference,” IEEE Trans. Circuits Syst. II, Exp. Briefs, vol. 55, no. 7, pp. 609–613, Jul. 2008. [31] K. N. Leung and P. K. T. Mok, “A sub-1-V 15- ppm/◦ C CMOS bandgap voltage reference without requiring low threshold voltage device,” IEEE J. Solid-State Circuits, vol. 37, no. 4, pp. 526–530, Apr. 2002. [32] G. De Vita and G. Ianaccone, “A sub-1-V 10- ppm/◦ C nanopower voltage reference generator,” IEEE J. Solid-State Circuits, vol. 42, no. 7, pp. 1536–1542, Jul. 2007. [33] K. Dejhan, N. Suwanchatree, P. Prommee, S. Piangprantong, and I. Chaisayun, “A CMOS voltage-controlled grounded resistor using a single power supply,” in Proc. IEEE ISCIT, Oct. 26–29, 2004, vol. 1, pp. 124–127. [34] A. Rogers, “Stability analysis of low-dropout linear regulators with a PMOS pass element,” Analog Applications Journal, pp. 10–12, Dallas, TX, Aug. 1999. [35] C. Stanescu, “Buffer stage for fast response LDO,” in Proc. 8th ICSICT, Sep. 28/Oct. 2, 2003, vol. 2, pp. 357–360. [36] W. Huang, S.-H. Lu, and S.-I. Liu, “CMOS low dropout regulator with single Miller capacitor,” Electron. Lett., vol. 42, no. 4, pp. 216–217, Feb. 2006. [37] W. Huang, S.-H. Lu, and S.-I. Liu, “A capacitor-free CMOS low dropout regulator with slew rate enhancement,” in Proc. Int. Symp. VLSI Design, Autom. Test, Apr. 2006, pp. 1–4. [38] Y. P. Tsividis, Operation and Modeling of the MOS Transistor. New York: McGraw-Hill, 1999. [39] T. Y. Man, P. K. T. Mok, and M. Chan, “A high slew-rate push–pull output amplifier for low-quiescent current low-dropout regulators with transientresponse improvement,” IEEE Trans. Circuits Syst. II, Exp. Briefs, vol. 54, no. 9, pp. 755–759, Sep. 2007. [40] S. Lau, P. K. T. Mok, and K. N. Leung, “A low-dropout regulator for SoC with Q-reduction,” IEEE J. Solid-State Circuits, vol. 42, no. 3, pp. 658– 664, Mar. 2007. [41] R. J. Milliken, J. Silva-Martıacute;nez, and E. Sánchez-Sinencio, “Full onchip CMOS low-dropout voltage regulator,” IEEE Trans. Circuits Syst. I, Reg. Papers, vol. 54, no. 9, pp. 1879–1890, Sep. 2007. Paulo Cesar Crepaldi received the B.Sc., M.Sc., and Ph.D. degrees in electrical engineering from the Universidade Federal de Itajuba, Itajuba, Brazil, in 1984, 1992, and 2009, respectively. Since 1985, he has been a Professor with Universidade Federal de Itajuba. Tales Cleber Pimenta received the B.S. and M.S. degrees in electrical engineering from the Universidade Federal de Itajuba, Itajuba, Brazil, in 1985 and 1988, respectively, and the Ph.D. degree from the Ohio University, Athens, in 1992. Since 1985, he has been a Professor with the Universidade Federal de Itajuba. He was a Visiting Scholar at The Ohio State University, Columbus, and Virginia Polytechnic and State University, Blacksburg, in 1997 and 2005, respectively. CREPALDI et al.: LOW POWER CMOS VOLTAGE REGULATOR FOR A WIRELESS BLOOD PRESSURE BIOSENSOR Robson Luiz Moreno received the B.S. degree in electrical engineering from the Universidade Federal de Itajuba, Itajuba, Brazil, in 1988, the M.Sc. degree from the Universidade de Campinas, Campinas, Brazil, in 1996, and the Ph.D. degree from the Universidade de São Paulo, São Paulo, Brazil, in 2002. Since 2002, he has been a Professor with the Universidade Federal de Itajuba. 739 Edgar Charry Rodriguez received the B.S. degree in electrical engineering from the Universidad Del Valle, Cali, Colombia, in 1962, the M.Sc. degree from Instituto Politécnico Nacional de México, México City, México, in 1970, and the Ph.D. degree from the Universidade de São Paulo (USP), São Paulo, Mexico, in 1974. He is currently a Researcher with USP. He has experience in the area of integrated circuits, mainly in CMOS, and pressure sensors.