Exp Fluids (2010) 49:257–266
DOI 10.1007/s00348-010-0853-9
RESEARCH ARTICLE
Dual-tracer fluorescence thermometry measurements
in a heated channel
Myeongsub Kim • Minami Yoda
Received: 13 July 2009 / Revised: 19 February 2010 / Accepted: 22 February 2010 / Published online: 13 March 2010
Ó Springer-Verlag 2010
Abstract The exponential growth of component density
in microelectronics has renewed interest in compact and
high heat flux thermal management technologies that can
handle local heat fluxes exceeding 1 kW/cm2. Accurate
and spatially resolved thermometry techniques that can
measure liquid-phase temperatures without disturbing the
coolant flow are important in developing new heat
exchangers employing forced-liquid and evaporative
cooling. This paper describes water temperature measurements using dual-tracer fluorescence thermometry (DFT)
with fluorescein and sulforhodamine B in laminar Poiseuille flow through polydimethyl siloxane-glass channels
heated on one side. The major advantage of using the ratio
of the signals from these two fluorophores is their temperature sensitivity of 4.0–12% per °C—a significant
improvement over previous DFT studies at these spatial
resolutions. For an in-plane spatial resolution of 30 lm, the
average experimental uncertainties in the temperature data
are estimated to be 0.3°C.
1 Introduction
In 1965, Gordon Moore of Intel predicted that the density
of transistors on a computer chip would double every two
years. Since then, the microelectronics industry has managed to follow Moore’s Law, with a doubling in component
density every 18–24 months. This exponential growth in
component density has however created huge thermal
M. Kim M. Yoda (&)
George W. Woodruff School of Mechanical Engineering,
Georgia Institute of Technology, Atlanta, GA 30332-0405, USA
e-mail: minami@gatech.edu
M. Kim
e-mail: mskim@gatech.edu
management challenges as more and more transistors (the
Intel dual-core Itanium 2, released in 2006, contains 1.7
billion transistors, for example) dissipate more and more
heat. Chip average heat fluxes are projected to approach
100 W/cm2 within a decade (ITRS 2006), with local heat
fluxes over hot spots with areas of O(103 lm2) up to an
order of magnitude greater than this value. Even with the
latest advances in heat sinks, forced-air cooling cannot
dissipate heat fluxes of this magnitude and still meet size
and space restrictions.
Future microprocessor designs will therefore require
new single-phase (liquid) and two-phase (evaporative)
thermal management technologies with micron-scale
addressability such as microchannel networks, jet and
spray arrays, heat spreaders and thermosyphons (Asheghi
and Yang 2005). A single chip cooled by a microchannel
heat sink with a footprint of a few cm2 will need a network
of up to several hundred microchannels connected by
numerous manifolds for adequate cooling at such high heat
flux values. Although thermal transport (due, presumably,
to laminar flows) in such microchannel networks can be
numerically simulated, these simulations require significant
computation time because of the geometric complexity of
these networks. These simulations are made even more
complex by the spatial variations in heat flux due to chip
‘‘power maps,’’ i.e., the power dissipation profile.
Such simulations are therefore too time-consuming at
present for design optimization purposes. ‘‘Reduced order’’
models, which predict thermal performance using heat
transfer correlations, are a more efficient way to design and
optimize the cooling performance of such microchannel
networks. It is unlikely, however, that classic convective
heat transfer correlations apply to microchannel networks,
because thermal coupling between the channels cannot be
neglected due to heat spreading via the high thermal
123
258
Exp Fluids (2010) 49:257–266
conductivity of silicon, the leading material for microelectronic components and MEMS devices.
By Newton’s Law of Cooling, the local heat flux normal
to the surface
q00s ¼ h ðTs Tm Þ
ð1Þ
where h is the local heat transfer coefficient, and Ts and Tm
are the local wall surface and mean (bulk) fluid temperatures, respectively. Developing local heat transfer correlations for various microchannel array geometries therefore
requires accurately measuring the difference between the
wall and bulk temperatures. Although a number of techniques such as infrared thermometry (IRT) can be used to
measure Ts with good spatial resolution in many cases,
there are far fewer techniques that can accurately and nonintrusively measure mean fluid temperatures, especially at
micro(meter-)scale resolution.
Although variations in the emission lifetime of phosphorescent and fluorescent compounds with temperature
have been used for liquid thermometry (Hu et al. 2006;
Mendels et al. 2008), the most commonly used optical
technique is fluorescence thermometry (FT), which
exploits the changes in the emission intensity of various
fluorophores (Nakajima et al. 1990). For a given fluorophore solution at a known molar concentration C, the
emission intensity imaged at a given location
If ðTÞ ¼ AIo gðTÞ
ð2Þ
where A is related to the collection efficiency of the
imaging system, and Io is the intensity of the excitation
(i.e., illumination). The temperature-dependent portion of
If, g(T), depends on the quantum yield and the molar
absorption coefficient of the fluorophore. The majority of
FT studies use the temperature-sensitive species rhodamine
B (RhB), usually excited at a wavelength of 532 nm, to
measure water temperatures ranging from *20°C to
*60°C (Coolen et al. 1999; Seuntiëns et al. 2001).
The accuracy of FT is improved by using a ratiometric
approach where the variations in emission intensity
reflecting the temperature distribution are decoupled from
those due to nonuniformities in the excitation intensity.
Dual-tracer FT (DFT) takes the ratio of the emissions from
two different fluorescent species, A and B, excited by the
same illumination (so IoA ¼ IoB ):
I 0 ðTÞ
IfA ðTÞ gA ðTÞ
/
:
IfB ðTÞ gB ðTÞ
ð3Þ
Coppeta and Rogers (1998), who summarized the
properties of a number of temperature-sensitive species,
demonstrated in their ‘‘dual-emission’’ technique that the
ratio of the signals from fluorescein (Fl) and RhB gave a
2D mean steady-state temperature fields in a thermal plume
with an accuracy of 1.8°C. Sakakibara and Adrian (2004)
123
used the ratio of the emissions from RhB and Rhodamine
110 (Rh110) instead to measure water temperatures with an
accuracy of 0.17°C.
Two-color and three-color FT takes instead the ratio of
the emissions from a single species over two and three
distinct wavelength bands, respectively, from a single
excited fluorophore. Lavieille et al. (2001, 2004) used RhB
to measure temperatures in a combusting and evaporating
ethanol spray and reported accuracies of 1°C for two-color
FT and 0.5–1°C for three-color FT.
More recently, fluorescence thermometry has also been
used to measure temperature fields in micron-scale flows
with a spatial resolution of O(10–100 lm). Ross et al.
(2001) used RhB to measure bulk temperatures in
electrokinetically driven flows in a 1-cm-long microchannel
with a trapezoidal cross-section of 30 9 20–75 lm at an
estimated precision of 2.4–3.5°C for fluid temperatures
ranging from 15 to 90°C. Kim et al. (2003) used RhB and
Rh110 in their ‘‘ratiometric’’ technique to measure steadystate temperature fields for the buoyancy-driven flow inside
a 1 9 10 9 45 mm test cell and reported uncertainties
(based on 95% confidence intervals) of 2–0.4°C for spatial
resolutions of 150–1,200 lm, respectively. Natrajan and
Christensen (2009) used RhB and sulforhodamine 101
(Sr101) to measure the steady-state temperature gradient
maintained across an array of parallel microchannels by hot
and cold thermal reservoirs and reported standard deviations as great as 0.6 and 0.5°C for ethanol and water
temperatures, respectively, at a spatial resolution of about
22 lm.
The majority of DFT studies use a temperature-sensitive
species, such as RhB, and a temperature-insensitive species,
such as Rh110, where the signal from the latter should be
directly proportional to the excitation intensity. More
recently, however, it has been shown that the accuracy of
DFT can be further improved by using two temperaturesensitive species whose temperature sensitivities are
‘‘inverted,’’ where ogA =oT [ 0 and ogB =oT\0. Shafii et al.
(2009) reported temperature sensitivities of 4% per °C in
their macroscale studies of solidification using Fl and sulforhodamine B (SrB), while Sutton et al. (2008) found
temperature sensitivities ranging from 4.5 to 6.5% per °C
using fluorescein 27 and RhB. These sensitivities are a major
improvement over the values of -1.7 and -2.7% per °C
reported by Sakakibara and Adrian (2004) and Natrajan and
Christensen (2009), respectively, in their DFT studies.
In the experiments described here, water temperatures
were measured using Fl and SrB in the laminar (Reynolds
numbers of 3 and 8) Poiseuille flow over a *23-lm-thick
‘‘slice’’ of the flow near the wall of a square channel heated
along one sidewall. Given that this flow is steady, the
emissions from the two species were imaged by the same
camera separated by a few seconds using different filters.
Exp Fluids (2010) 49:257–266
The setup used here therefore cannot capture time-resolved
(i.e., instantaneous) temperature field data, but, as demonstrated by Shafii et al. (2009) for Fl and SrB, as well as
by other researchers for other fluorophore pairs (Sakakibara
and Adrian 2004; Natrajan and Christensen 2009), the DFT
technique can measure transient temperature fields using
two cameras with appropriate filters.
The results presented here are, to our knowledge, the
first time that DFT using two species with inverted temperature sensitivity has been used and validated at micronscale spatial resolutions. Section 2 describes the aqueous
solution of the two temperature-sensitive fluorescent dyes,
fluorescein (Fl) and sulforhodamine B (SrB) and then
details the heated channel and optical system used in these
studies. Results from in situ calibration studies of the
temperature sensitivity of Fl and SrB illuminated volumetrically at 514 nm, as well as the temperature sensitivity
of the ratio of the Fl and SrB signals, are also presented.
Section 3 details numerical simulations performed with the
computational fluid dynamics (CFD) software package
FLUENT to validate these DFT data, which were obtained
in the presence of a significant temperature gradient within
the fluid. Section 4 presents and discusses the initial DFT
temperature results and compares them with the numerical
predictions, while Sect. 5 summarizes this work.
2 Experimental description
2.1 Fluorophore solution
The majority of temperature-sensitive fluorophores have a
quantum yield, and hence an emission intensity, that
decreases with increasing temperature, because the nonradiative dissipation in most cases increases with increasing temperature. Fluorescein, however, has an emission
intensity that, when excited at a wavelength k = 514 nm
259
(significantly off its absorption peak), increases with
increasing temperature, because its absorption at 514 nm
also increases with temperature (Coppeta and Rogers
1998). Although there are several fluorophores whose
emission intensity decreases with increasing temperature,
SrB was chosen among these as the second species
because: (1) it can also be excited at 514 nm due to its
relatively broad absorption spectrum; and (2) its emission
band, with a peak at 591 nm, has little overlap with that of
Fl, with a peak at 518 nm (Shafii et al. 2009).
Figure 1 shows the normalized absorption spectrum
measured by a spectrometer (HR4000CG-UV-NIR, Ocean
Optics, Inc.) (left) and the normalized emission spectrum
measured by a fluorescence spectrometer (QuantaMaster,
PTI, Inc.) (right) as a function of wavelength k for 5 lmol/L
Fl (solid line) and 5 lmol/L SrB (dashed line) both in
the same aqueous solution, at pH 9.2. The absorption
spectra verify that both Fl and SrB can be excited at a
wavelength of 514 nm. The emission spectra show that the
fluorescence from the Fl and the SrB can be isolated from
each other using appropriate filters. Shafii et al. (2009) isolated the SrB emissions from those of the Fl using a longpass
filter that only transmitted light at k C 600 nm and reported
that the leakage of Fl emissions through this filter, or
‘‘cross-talk’’ with the Fl, was no more than 0.1% of the total
signal.
In microscale thermometry, it is especially important to
ensure that the fluorophores have minimal surface
adsorption on channel materials such as silicate glasses,
silicon (oxide) and polydimethyl siloxane (PDMS) to
minimize background noise in geometries where the surface area to volume ratio is significantly greater than that at
the macroscale. Although RhB is often used in FT because
of its marked temperature sensitivity, most of the rhodamine dyes are zwitterions with a positively charged portion
that strongly adsorbs onto negatively charged surfaces such
as glass, silicon and PDMS at moderate pH. Aqueous
Fig. 1 Absorption (left) and
emission (right) spectra for Fl
(solid line) and SrB (dashed
line). The arrows on both
spectra indicate the excitation at
514 nm; the shaded and hatched
regions on the emission
spectrum denote the
wavelengths transmitted by the
filters that isolate the emissions
from Fl and SrB, respectively
123
260
Exp Fluids (2010) 49:257–266
solutions of 10 lmol/L Fl, where the Fl is a divalent anion
in solution, and 10 lmol/L SrB, both at pH 9.2, were found
to have negligible adsorption on fused-silica surfaces at
temperatures of 20–60°C (Suda-Cederquist 2007).
Dual-tracer fluorescence thermometry was used to
measure temperature fields in an aqueous solution of
5 lmol/L fluorescein, 5 lmol/L sulfohodamine B and
7 mmol/L sodium tetraborate (Na2B4O7), which was prepared by dissolving appropriate amounts of the disodium
salt of Fl (FL116, Spectrum Chemical), the sodium salt of
SrB (86183, Fluorescence grade, Fluka), and the decahydrate salt of Na2B4O7 (ACS reagent grade, Acros Organics)
in deionized water. The addition of the tetraborate salt
ensured that the solution pH was 9.2 at a solution temperature of 19°C, as measured by a pH meter (WD-3561420, Oakton), to maximize the quantum yield of Fl. The pH
of the Fl-SrB solution varied by no more than 0.2 pH units
as the solution temperatures varied from 20 to 60°C. The
solution was filtered through 1.5 lm pore size filter paper
(09-804-70A, Grade G6, Fisher Scientific) with a Büchner
funnel and degassed by a vacuum pump for [2 h before
use.
2.2 Heated channel and optical setup
The temperature of the Fl-SrB solution was measured in
Poiseuille flow through a channel with a square crosssection 1 mm on a side and a length of 28 mm heated on
one side. The channel (Fig. 2) was molded in a block of
polydimethyl siloxane (PDMS) of nominal dimensions
35 9 25 9 5 mm, which was cured using standard procedures (Duffy et al. 1998) at 60°C for 12 h and plasma
oxidized to make the PDMS surface hydrophilic. A 50-W
cartridge heater (FIREROD, Watlow), 3 mm in diameter
and 31 mm in length, was embedded in the PDMS at a
minimum edge-to-edge distance of 0.5 mm from the
channel sidewall to create a temperature gradient across the
channel. The 1-mm square trench in the PDMS block was
sealed by a fused-silica ‘‘lid’’ (44.5 9 44.5 9 1 mm) on
the bottom. The PDMS sides of the channel were exposed
to the atmosphere under ambient conditions, while the
fused-silica bottom of the channel sat upon a steel microscope stage.
A syringe pump (NE-500, New Era Pump System, Inc.)
drove the Fl-SrB solution through the PDMS-glass channel
at volume flow rates of Q = 1 and 8 lL/s. The Reynolds
numbers Re for these two Poiseuille flow cases based on the
channel cross-sectional dimension and average speed were 1
and 8, respectively. The pump was connected to the channel
inlet port by 0.8-mm ID silicone tubing (51134K14,
McMaster-Carr), and the solution was discarded after
exiting the channel through more 0.8-mm ID silicone tubing
attached to the outlet port.
The temperature of the fluid inside the channel T was
monitored by two T-type miniature thermocouples (TC)
(HYP1-30-1/2-T-G-60-SMPW-M, Omega Engineering,
Inc.) with a bead diameter of 0.3 mm, which were inserted
through the PDMS block and placed symmetrically
upstream and downstream of the imaged region roughly in
the center of the channel with a streamwise spacing of
about 1.2 cm (cf. Fig. 2). The TC were calibrated to an
accuracy of 0.2°C, and the TC readings were recorded by
an A/D data acquisition board (HP34970A, Agilent Technologies) onto a PC HD.
For all experiments, the PDMS-glass channel was
placed on the steel stage of an inverted microscope (DM
IL, Leica), and the entire channel cross-section of the
channel was volumetrically illuminated by a light sheet
formed from the 60 mW beam at k = 514 nm of an argonion laser (543-MA-A03, Melles Griot) passing through an
excitation filter (z514/109, Chroma Tech). The power of
the illumination beam varied by no more than 1% over 5 h,
35 mm
Heater
Flow
25 mm
Channel
Heater
1 mm
TC
1
0.5
Lid
Stage
TCs
Filters
Objective
Top View
Fig. 2 Top (left) and side (right) views of the channel with singlesided heating. The checkered rectangle and line (in the channel)
indicate the region imaged by the microscope objective in the top and
123
Side View
side views, respectively. Flow goes from left to right in the top view
and out of the page in the side view
Exp Fluids (2010) 49:257–266
as measured by a laser power meter (Lasermate Q,
Coherent).
As shown in Fig. 2, the fluorescent emissions from a
‘‘slice’’ of the flow just above the fused-silica wall were
imaged through the bottom of the channel by a 109
magnification, 0.25 numerical aperture microscope objective (506084, N Plan, Leica) and appropriate filters and
recorded by a CCD camera (Pixelfly vga, Cooke Corporation) with an intensifier (VS4-1845, OPELCO, Inc.) as
640 9 480 pixels 8-bit images on a PC HD. The Fl
emissions from a ‘‘slice’’ of the flow just above the fusedsilica wall were imaged through a bandpass filter (NT48083, Edmund Optics) that only transmitted light at
k = 529–555 nm (shaded region, Fig. 1). The SrB emissions were isolated from those of the Fl by a longpass filter
(NT47-618), Edmund Optics) that only transmitted light at
k C 600 nm (hatched region, Fig. 1), similar to that used
by Shafii et al. (2009).
A sequence of 140 images spanning 3.4 s of the emissions from Fl and SrB was recorded at a framing rate of
41.7 Hz and an exposure of 20 ms. Both sequences were
recorded within a few seconds of each other by switching
between the appropriate filters that were next to each other
in a translating filter holder. The thickness, or dimension
along the optical axis, of the imaged region was estimated
to be about 23 lm, based on the depth of field of the
objective. The magnification of the imaging system,
determined from a scale with 20 lm divisions, was measured to be 2.67, corresponding to 3.7 lm/pixel for the
9.9 lm CCD pixels. The physical dimensions of the
640 9 480 images are therefore 2.37 9 1.78 mm.
The relationship between the emission intensity incident
upon the intensified CCD camera and the resultant grayscale recorded on the digital image was found to be
somewhat nonlinear. Camera calibrations were therefore
performed to determine this relationship. Since Fl and SrB
have different emission spectra and different quantum
yields when excited at 514 nm, separate calibrations were
performed for the emissions from these two species for a
range of fluorophore concentrations, illumination intensities and imaging parameters. The fluorescence intensity for
both species was found to be well approximated by secondorder polynomials, with an adjusted coefficient of determination R2 ¼ 99:7% and a root mean squared error
(RMSE) below 1%.
261
passing the fluid through a copper-coil heat exchanger
consisting of 15 m of 1.5 mm ID copper tubing immersed
in a constant-temperature circulating water bath (GH-D8,
Haake) before entering the PDMS-glass channel. The
Reynolds number for the laminar Poiseuille flow Re
based on the channel cross-sectional dimension of 1 mm,
and average speed was 67, corresponding to a volume flow
rate Q = 0.067 mL/s, to minimize thermal losses. Since no
power was supplied to the cartridge heater during these
calibrations, the fluid temperature T was determined by the
set point temperature of the water bath. Data were acquired
at least 20 min. after flow startup when the TC readings
were within 0.2°C (i.e., the measurement accuracy of the
TC) of the set point temperature of the water bath.
As described previously, two sequences, each 140
images long, were acquired of the Fl and SrB emissions.
After compensating for camera nonlinearities, the grayscale values were spatially averaged over a 150 9 250
pixels (0.56 9 0.93 mm) region in the center of the
channel (Fig. 3) and then temporally averaged over all 140
(sub-)images in the sequence. Figure 4 shows this average
normalized by that at 20°C, I20 , as a function of
intensity, I,
solution temperature T measured by the TC at T = 20–
60°C for Fl (red triangles) and SrB (blue circles). The error
bars denote the standard deviation of the data. The fluorescence from Fl increases by about 2.44% per °C, based
on a linear curve-fit, whereas the signal from SrB decreases
by about 1.54% per °C. Both values are in good agreement
with the 2.43% per °C and the 1.55% per °C reported by
Coppeta and Rogers (1998) for Fl and SrB, respectively.
The calibration results for Fl and SrB were consistent over
multiple independent experiments.
The ratio of these two normalized signals, I 0 (gray
squares), clearly has much greater variation with temperature than either the normalized Fl or SrB emissions,
increasing from 1 at T = 20°C to about 4.0 at T = 60°C.
A fifth-order polynomial curve-fit (R2 = 99.7%, RMSE =
0.06%) to I 0 (solid line, Fig. 4) gave:
2.3 Fluorescence thermometry calibrations
Calibrations of the temperature response of the Fl-SrB
solution at temperatures T = 20–60°C were performed in
the same channel with the same optical system as those
used in the DFT experiments. The temperature of the
working fluid in the channel was kept at a constant value by
Fig. 3 A typical 640 (columns) 9 480 pixels (rows) calibration
image of SrB at T = 20°C showing the 150 9 250 pixels
(0.56 9 0.93 mm) region in the center of the channel used for the
temperature calibrations. The edge of the cartridge heater is just
visible on the right side
123
262
Exp Fluids (2010) 49:257–266
I20 as a function
Fig. 4 Average normalized fluorescence intensity I=
of solution temperature T for Fl (triangles) and SrB (circles). The
ratio of these two normalized intensities I0 (squares) is also given as a
function of T, along with a fifth-order polynomial curve-fit to these
data (solid line), given by Eq. 4
I 0 ¼ 1:64 107 T 5 3:21 105 T 4 þ 2:41 103 T 2
8:49 102 T þ 1:44 T 8:52
ð4Þ
where the temperature T is measured in °C. The slope of
Eq. 4 gives temperature sensitivities dI 0 =dT ranging from
4.0% per °C at T = 20°C to 12% per °C at T = 60°C,
suggesting that this technique has greater sensitivity at
higher temperatures. Nevertheless, these sensitivities are
significantly higher than the -2.7% per °C reported by
Natrajan and Christensen (2009) in their microscale DFT
studies. A fifth-order polynomial curve-fit to the
temperature as a function of I 0 (R2 ¼ 99:6%;
RMSE = 0.6°C) gave:
T ½ C ¼ 47:83 ðI 0 Þ5 þ 250:8 ðI 0 Þ4 476 ðI 0 Þ3
þ 384:7 ðI 0 Þ2 85:41 I 0 þ 14:64
ð5Þ
Equation 5 was used to determine liquid-phase temperatures for the results presented in Sect. 4.
The absorption and emission spectra shown in Fig. 1
suggest that there may be some ‘‘cross-talk’’ between the Fl
and SrB emissions and that a small amount of the Fl
emissions may be imaged through the longpass
(k C 600 nm) filter. Moreover, the emission band of Fl and
the absorption band of the SrB have some overlap, suggesting that some of the emissions from Fl may excite the
SrB, resulting in ‘‘spectral conflict.’’ Both of these phenomena are issues for DFT with a single illumination
source.
To evaluate the effects of cross-talk and spectral conflict, three 140-image sequences were acquired of the flow
123
of three different solutions: the Fl-SrB-tetraborate solution,
5 lmol/L Fl and 7 mmol/L Na2B4O7, and 5 lmol/L SrB
and 7 mmol/L Na2B4O7 at T = 55°C. This temperature,
near the upper end of the temperature range studied here,
was chosen to maximize cross-talk and spectral conflict,
since the Fl emissions will be stronger, and the SrB
emissions weaker, at higher T. All three sequences were
acquired in the same setup used for the FT calibrations
under identical experimental conditions within a single
experimental run over *90 min by switching between
syringes in the syringe pump. In all cases, the emission
intensities recorded for all three solutions were identical
within the standard deviations of the images, suggesting
that these effects are negligible. These results are in
agreement with those reported by Shafii et al. (2009) who
reported that cross-talk and spectral conflict were negligible for 0.5 lmol/L Fl and 0.5 lmol/L SrB over optical
pathlengths of at least 1 cm.
3 Numerical simulations
The DFT results were validated by 3D finite-volume simulations of steady laminar flow through the channel using
the commercial CFD software package FLUENT v6.2 run
on a 2.4 GHz PC. Most of the channel, namely the entire
cartridge heater, the PDMS block, the fused-silica lid
and the steel microscope stage (cf. Fig. 3 (right)), was
simulated in the numerical model, which consisted of
3 9 106 cells meshed using a hybrid tetrahedral scheme
with a spatial resolution of 50 lm inside the channel.
Computational limitations (specifically, the RAM on the
PC was limited to 3.5 GB) precluded using a finer mesh to
improve the spatial resolution of these simulations. Given
that the average difference between the temperatures
obtained at spatial resolutions of 50 and 100 lm was
0.09°C, however, linear interpolations of these numerical
predictions should be a good estimate of the temperatures
in the flow at finer spatial resolution.
The temperatures were assumed to achieve steady state
after 3 9 103 iterations, based on convergence tests for
temperatures obtained over up to 104 iterations. In all
cases, the numerical simulation results, averaged over the
0.3 mm diameter of the TC bead, were consistent with the
TC readings at the corresponding locations.
The inlet temperature boundary condition was the temperature measured in the FT experiments. Standard
free-convection boundary conditions from empirical correlations for external flows (Incropera and DeWitt 2002)
were imposed on the surfaces exposed to ambient conditions. The Churchill–Chu correlation for the average
Nusselt number NuL for free convection from a vertical wall based on the average surface temperature was
Exp Fluids (2010) 49:257–266
263
used for the sides of the PDMS block at typical Rayleigh
numbers RaL ¼ Oð102 Þ:
1=4
0:67RaL
NuL ¼ 0:68 þ h
9=16 i4=9
1 þ 0:492
Pr
ð6Þ
where Pr is the Prandtl number. An adiabatic boundary
condition was imposed on the (vertical) side surfaces of the
fused-silica lid since the heat transfer through the 1-mmthick lid was, as verified by FLUENT simulations, negligible compared to that through the top and bottom surfaces
of the lid.
For the upper surfaces of the glass lid and the PDMS
block with RaL ¼ Oð105 Þ and RaL ¼ Oð106 Þ, respectively,
the standard power-law correlation for free convection
from the upper surface of a heated horizontal plate was
used:
1=4
NuL ¼ 0:54RaL :
ð7Þ
Finally, for the bottom surface of the glass lid (i.e.,
that contacting the microscope stage) with RaL ¼ Oð106 Þ,
the correlation for a lower surface of a heated plate was
used:
1=4
NuL ¼ 0:27RaL :
ð8Þ
Because heat was only generated over the central
19.3 mm (axial extent) of the cartridge heater, a constant
heat-generation rate boundary condition was imposed at
the cartridge heater over this central section. Freeconvection boundary conditions were imposed over the
Inconel surface of the rest of the heater, where the average
Nusselt number for this heated vertical cylindrical surface
at typical RaL ¼ Oð105 Þ was that given by Eq. 6 scaled by
the transverse curvature parameter n, which was 1.383 for
this geometry (Cebici 1974).
The heat-generation rate was taken to be 90% of the
electrical power input to the heater divided by the volume
of the 19.3 mm long, 3 mm diameter central section. In
this experimental setup, some of the power input to the
heater will be dissipated in the solid portions of the channel
(vs. transferred to the fluid) and the surroundings, since the
upper portion of the cartridge heater is exposed to the
atmosphere at ambient conditions. Given the difficulties in
accurately estimating this thermal dissipation, the percentage of the input power transferred to the fluid was
adjusted to a value that gave numerical predictions that
were consistent with the DFT results in terms of the temperature change across the channel. It was assumed that
only 90% of the input power was transferred to the flow for
all the cases presented here. The major source of error in
the numerical predictions is therefore due to error in estimating the fraction of the input power transferred to the
fluid.
4 Results and discussion
Experiments were then carried out for laminar steady
Poiseuille flow subject to a temperature gradient across the
channel at Reynolds numbers Re = 1 and 8 (corresponding
to Q = 1 and 8 lL/s, respectively). The channel was now
heated by the cartridge heater at an edge-to-edge distance
of 0.5 mm from the sidewall supplied with 1.4 W (20 V at
69 mA) from a DC power supply (E3612A, Agilent
Technologies). The temperature field in this laminar and
steady flow was assumed to have reached steady state once
the temperature reading from the TC downstream of the
imaged region (cf. Fig. 2 (left)) varied by no more than
0.2°C over 2 min.
Two sequences of 140 images, spanning 3.4 s of the
emissions from Fl and SrB, were recorded within a few
seconds of each other. The 640 9 480 pixel images of the
emissions over a 23-lm-thick slice of the flow were truncated to 384 9 480 pixel images, corresponding to a
physical field of view of 1.4 9 1.8 mm, centered about the
channel, which was 277 pixels across. Figure 5 shows
representative images of the Fl (left) and the SrB (right)
emissions. Both images are mapped to the same set of
grayscales, and, as expected, the Fl emissions are for the
most part weaker than those of the SrB when excited at
514 nm. The truncated images were then corrected for
camera nonlinearities, and the average grayscale value and
the standard deviation in the grayscale value were calculated over all 140 images. The ratio of the Fl and SrB
images was used to determine I 0 ; and a standard uncertainty
propagation analysis was used to determine the uncertainty
in the resultant temperature from the standard deviations of
the Fl and SrB images.
At Re = 1, the uncertainties were ±1 and 0.3°C
at spatial resolutions of 3.7 lm (1 pixel) and 30 lm (8
pixels), respectively. At Re = 8, the uncertainties in the
temperatures obtained using DFT were ±1.1 and 0.3°C at
the same spatial resolutions. Given that the spatial resolution of these data normal to the image plane (i.e., along the
optical axis) was *23 lm based on the depth of field of
the microscope objective, the DFT data were averaged over
a region 8 pixels, or 30 lm, square.
Figure 6 compares pseudocolor temperature maps over
a 1.1 9 1 mm region of the flow adjacent to the heater
based on numerical predictions from FLUENT at a spatial
resolution of 50 lm (left column) with those obtained from
the DFT results for I 0 using Eq. 5 at a spatial resolution of
30 lm (right column) with at Re = 1 (top row) and 8
(bottom row). The DFT results for this 2D temperature
field are in reasonable qualitative agreement with the
numerical predictions.
To quantitatively compare the numerical and experimental results, the temperature profile across the channel,
123
264
Exp Fluids (2010) 49:257–266
Fig. 5 Typical 640 9 480 pixels (2.37 9 1.78 mm) grayscale
images at an exposure of 20 ms of Fl (left) and SrB (right) heated
on one side at Re = 1. The white rectangle denotes the 384 9 480
DFT
FLUENT
T [°C]
50
0.5
1.1 mm
Re = 1
Flow
Heater
Fig. 6 Pseudocolor map of
water temperatures over a
1.1 9 1 9 23 lm region near
the bottom of the channel at
Re = 1 (top row) and Re = 8
(bottom row) predicted using
FLUENT at a spatial resolution
of 50 lm (left column)
measured using DFT at an inplane spatial resolution of
30 lm and (right column). The
dashed line represents the
streamwise location where
temperature profiles were
extracted across the channel, or
along the y-direction
pixels (1.4 9 1.8 mm) region processed in these images. The edge of
the heater is again just visible on the right side of both images
40
1 mm
T [°C]
33
Re = 8
Flow
25
or along y, was extracted from both the DFT and FLUENT
temperature maps shown in Fig. 6 at the streamwise
location where the edge of the heater is at its minimum
distance of 0.5 mm from the side wall of the channel
(dashed line in Fig. 6). Figure 7 compares these temperature profiles across the channel (where the coordinate
across the channel, y, is normalized by channel y-dimension L = 1 mm) measured by DFT using Eq. 5 (triangles)
with those predicted by the FLUENT simulations (circles)
at Re = 1 (a) and 8 (b). The error bar denotes the uncertainty in the temperature measured by DFT.
To estimate the discrepancy between the DFT results
and the numerical predictions, which are obtained at different spatial resolutions, fifth-order polynomials were
123
y
curve-fit to both sets of temperature data (with
R2 [ 0:995 in all cases), and the average discrepancy was
determined between these data sampled at a spatial resolution of 30 lm. These discrepancies were 0.1 and 0.3°C
at Re = 1 and Re = 8, respectively. At the higher Re, the
DFT results are mostly higher than the numerical predictions, which may be due to errors in the FLUENT
simulations. As discussed in Sect. 3, it is assumed that
90% of the input power is transferred to the fluid, and
enhanced convection at the higher flow rate should
increase heat transfer to the fluid.
For the Re = 8 case, the discrepancy is greatest, 0.4–
0.5°C, for y=L\0:4, or the region of lowest water temperature of *27°C. This increase in the discrepancy at
Exp Fluids (2010) 49:257–266
265
Channel
y
L
y
Heater
L
Heater
T [°C]
Channel
Re = 8
Re = 1
y /L
y/L
Fig. 7 Comparison of fluid temperatures T as a function of normalized channel coordinate y/L measured by DFT at a spatial resolution
of 30 lm (triangles) and predicted by FLUENT at a resolution of
50 lm (circles) at Re = 1 (left) and 8 (right). These temperature
profiles are obtained at the streamwise location where the heater is
nearest the side wall of the channel
lower T is likely due to the lower sensitivity of the DFT
technique and the relatively low level of Fl emissions at
such temperatures. A temperature discrepancy of 0.5°C
corresponds to a variation of less than 0.02 in I 0
(dI 0 =dT ¼ 0:035 at T = 27°C) or less than half the stan I20 for Fl (cf. Fig. 4) at the same temdard deviation in I=
perature even in the ‘‘ideal’’ case where the temperature is
constant over the entire image.
These experimental results suggest that dual-tracer fluorescence thermometry using Fl and SrB, fluorophores with
inverted temperature sensitivity, can be used to obtain
accurate estimates of liquid-phase temperature fields at a
spatial resolution of 30 lm, even in the presence of temperature gradients of about 4°C/mm. In both cases studied
here, the average discrepancy between the experimental data
and the numerical predictions is within the experimental
uncertainty of the DFT measurements of 0.3°C.
channel with a temperature gradient across the channel of
about 4°C/mm are in agreement, on average, with numerical
predictions.
The largest issue in validating these experimental data is
the uncertainty in estimating how much power is actually
transferred to the fluid. New channels with thin-film heaters
sandwiched between the PDMS block and the glass lid—a
configuration where virtually all the input power remains
within the channel—will therefore be used in subsequent
studies.
In terms of future work, combining fluorescence thermometry with evanescent-wave illumination will give
liquid-phase temperatures averaged over the first *400 nm
next to the wall, which in most cases is effectively the wall
surface temperature. Fluorescence thermometry could then
be used, with further development, to obtain nonintrusive
measurements at the microscale of wall surface and bulk
liquid temperatures in the same flow.
5 Summary
Acknowledgments This work was supported by the National Science Foundation and Sandia National Laboratories through NSF grant
CBET-0625825 from the Thermal Transport Processes Program (Dr.
T. L. Bergman, program officer) and the Office of Naval Research
through grant N00014-09-1-0298 from the Thermal Management
Program (Dr. M. Spector, program manager). The authors thank K. D.
Suda-Cederquist and Dr. T. G. Hwang for their help on the Fl and SrB
absorption studies.
This paper presents dual-tracer fluorescence thermometry
results obtained with fluorescein and sulforhodamine B for
liquid-phase temperatures. Although this pair of dyes is not
novel, this is, to our knowledge, the first application of the
Fl/SrB pair to obtain water temperature fields at micronscale spatial resolution for water temperatures ranging from
20 to 60°C. The experimental uncertainties for this method
are ±1.1°C at a spatial resolution of 3.7 lm and ±0.3°C at a
spatial resolution of 30 lm. These values are an improvement over those reported previously for microscale applications of fluorescence thermometry at similar spatial
resolution. The experimental results obtained in steady
Poiseuille flow at Re = 1 and 8 through a 1-mm square
References
Asheghi M, Yang Y (2005) Micro- and nano-scale diagnostic
techniques for thermometry and thermal imaging of microelectronic and data storage devices. Springer, Berlin
Cebici, T (1974) Laminar-free-convective heat transfer from the outer
surface of a vertical slender circular cylinder. In: Proceedings of
123
266
the fifth international heat transfer conference, vol 4. Paper NC1,
pp 15–19
Coolen MCJ, Kieft RN, Rindt CCM, van Steenhoven AA (1999)
Application of 2-D LIF temperature measurements in water
using a Nd:YAG laser. Exp Fluids 27:420–426
Coppeta J, Rogers C (1998) Dual emission laser induced fluorescence for
direct planar scalar behavior measurements. Exp Fluids 25:1–15
Duffy DC, McDonald JC, Schueller OJA, Whitesides GM (1998)
Rapid prototyping of microfluidic systems in poly(dimethylsiloxane). Anal Chem 70:4974–4984
Hu H, Koochesfahani M, Lum C (2006) Molecular tagging
thermometry with adjustable temperature sensitivity. Exp Fluids
40:753–763
Incropera FP, DeWitt DP (2002) Fundamentals of heat and mass
transfer, Chap. 9, 5th ed. Wiley Inc, New York
ITRS (2006) International technology roadmap for semiconductors
(www.itrs.net/Links/2006Update/2006UpdateFinal.htm)
Kim HJ, Kihm KD, Allen JS (2003) Examination of ratiometric laser
induced fluorescence thermometry for microscale spatial measurement resolution. Int J Heat Mass Transfer 46:3967–3974
Lavieille P, Lemoine F, Lavergne G, Lebouche M (2001) Evaporating
and combusting droplet temperature measurements using twocolor laser-induced fluorescence. Exp Fluids 31:45–55
Lavieille P, Delconte A, Blondel D, Lebouche M, Lemoine F (2004)
Non-intrusive temperature measurements using three-color laserinduced fluorescence. Exp Fluids 36:706–716
Mendels DA, Graham EM, Magennis SW, Jones AC, Mendels F
(2008) Quantitative comparison of thermal and solutal transport
123
View publication stats
Exp Fluids (2010) 49:257–266
in a T-mixer by FLIM and CFD. Microfluid Nanofluid 5:603–
617
Nakajima T, Utsunomiya M, Ikeda Y, Matsumoto R (1990) Simultaneous measurement of velocity and temperature of water using
LDV and fluorescence technique. 5th Int Symp on Appl Laser
Tech to Fluid Mech Lisbon, 2.6.1–2.6.6
Natrajan VK, Christensen KE (2009) Two-color laser-induced
fluorescent thermometry for microfluidic systems. Meas Sci
Technol 20:015401
Ross D, Gaitan M, Locascio L (2001) Temperature measurement in
microfluidic systems using a temperature-dependent fluorescent
dye. Anal Chem 73:4117–4123
Sakakibara J, Adrian RJ (2004) Measurement of temperature field of a
Rayleigh-Bénard convection using two-color laser-induced
fluorescence. Exp Fluids 37:331–340
Seuntiëns HJ, Kieft RN, Rindt CCM, van Steenhoven AA (2001) 2D
Temperature measurements in the wake of a heated cylinder
using LIF. Exp Fluids 31:588–595
Shafii MB, Lum CL, Koochesfahani MM (2009) In situ LIF
temperature measurements in aqueous ammonium chloride
solution during uni-directional solidification. Exp Fluids doi:
10.1007/s00348-009-758-7 (online publication)
Suda-Cederquist KD (2007) Near-wall thermometry via total internal
reflection fluorescent micro-thermometry (TIR-FMT). M.S.
Thesis, Georgia Institute of Technology
Sutton JA, Fisher BT, Fleming JW (2008) A laser-induced fluorescence measurement for aqueous fluid flows with improved
temperature sensitivity. Exp Fluids 45:869–881