IR400 Manual
IR400 Manual
IR400 Manual
The information and technical data disclosed in this document may be used and disseminated only for the purposes and to the extent specifically authorized in writing by General Monitors. Instruction Manual 04-11 General Monitors reserves the right to change published specifications and designs without prior notice. MANIR400 Part No. Revision MANIR400 D/04-11
Model IR400
ii
Model IR400
Table of Contents
TABLE OF FIGURES ..................................................................................................................IV TABLE OF TABLES....................................................................................................................IV QUICK-START GUIDE................................................................................................................. 2 1.0 INTRODUCTION .................................................................................................................... 4
1.1 1.2 1.3 Protection for Life .......................................................................................................................4 Special Cautions and Warnings.................................................................................................4 Installation, Operation, and Maintenance ..................................................................................4
6.0 TROUBLESHOOTING ......................................................................................................... 25 7.0 MODBUS INTERFACE ........................................................................................................ 27 8.0 CUSTOMER SUPPORT ....................................................................................................... 28 9.0 APPENDIX............................................................................................................................ 29
9.1 9.2 9.3 Warranty...................................................................................................................................29 Principle of Operation...............................................................................................................30 Specifications ...........................................................................................................................31 iii
Model IR400
Table of Figures
Figure 1: IR400 with Junction Box.......................................................................................................................... 2 Figure 2: Wiring Diagram from IR400 to Control Equipment.................................................................................. 3 Figure 3: Model IR400 ............................................................................................................................................ 5 Figure 4: Model IR400 with a Junction Box............................................................................................................ 8 Figure 5. IR400 Mounting Dimensions .................................................................................................................. 9 Figure 6: IR400 with Junction Box........................................................................................................................ 13 Figure 7: Wiring Diagram from IR400 to Control Equipment................................................................................ 14 Figure 8. IR400 Zero Switch / LED During Zeroing and Calibration .................................................................... 19 Figure 9. IR400 Optical Window Locations ......................................................................................................... 24 Figure 10: Outline and Dimensional Drawing, IR400, Inch Measurements ......................................................... 35
Table of Tables
Table 1: Wiring Chart.............................................................................................................................................. 2 Table 2: Wiring Chart............................................................................................................................................ 11 Table 3: IR400 without HART Option ................................................................................................................... 17 Table 4: IR400 with HART Option ........................................................................................................................ 17 Table 5: GM Locations.......................................................................................................................................... 28 Table 6: Analog Current Output............................................................................................................................ 33 Table 7: Maximum Distance between IR400 and Power Source......................................................................... 33 Table 8: Maximum Distance between IR400 and 500-Ohm Input Impedance..................................................... 33
iv
Model IR400
Quick-Start Guide
Mounting and Orientation
The Model IR400 should be mounted horizontally (Figure 1) to reduce the likelihood of dirt and dust build-up on the windows. For optimum performance, the splashguards should be located on the top and bottom as shown in Figure 1. Apply the supplied thread lubricant/sealant to all conduit entries before use to prevent binding.
Wiring Connections
TERMINAL 1 2 3 4 5 6 7 8 9 10 WIRE COLOR BLACK GREEN RED WHITE BLUE WHT/BLU BROWN SIGNAL COM FIELD GROUND (FG) +24 V 4-20 mA MOD1+ MOD1 CAL MOD2+ (no connection for IR400) MOD2 - (no connection for IR400) No Connection
Table 1: Wiring Chart NOTE: Power should remain disconnected until all other wiring connections are made.
Model IR400
Power Connections
To supply power to the IR400, connect the red lead from the IR400 to the +24 VDC terminal on the power supply. Connect the black lead from the IR400 to the power supply Common. Refer to the manual of the power source being used, for more detailed instructions. NOTE: If the Model IR400 is being used with a +24 VDC power supply and an industrial analog to digital (A/D) converter, then the negative supply (Common) of all three must be connected.
Applying Power
Before applying power to the system for the first time, all wiring connections should be checked for correctness. Upon initial power-up or after a fault condition has been corrected, the unit will enter a start-up mode for 2 minutes before normal operation. IR400 Ethylene start-up can take up to 10 minutes. As a unit warms up, output may briefly be negative. After power is applied, the IR400 should be allowed to stabilize for approximately 60 minutes while the unit attains the proper operating temperature. After stabilization, it is recommended that the IR400 be zeroed, per the procedure in section 4.1 (step 1 only). A gas check should then be performed to ensure that the unit is operating properly. Use the General Monitors Gas Check Kit (P/N 32548) to perform this check.
Figure 2: Wiring Diagram from IR400 to Control Equipment The instrument is now ready to operate. Please further consult this manual for more information on the instruments many features. 3
Model IR400
1.0 Introduction
1.1 Protection for Life
General Monitors mission is to benefit society by providing solutions through industry-leading safety products, services and systems that save lives and protect capital resources from the dangers of hazardous flames, gases and vapors. The safety products you have purchased should be handled carefully and installed, calibrated and maintained in accordance with this instruction manual. Remember, these products are for your safety.
Model IR400
Model IR400
Model IR400
3.0 Installation
Receipt of Equipment
All equipment shipped by General Monitors is packaged in shock absorbing containers, which provide considerable protection against physical damage. The contents should be carefully removed and checked against the packing list. If any damage has occurred or there is any discrepancy in the order, please notify General Monitors as soon as possible. All subsequent correspondence with General Monitors must specify the equipment part number and the serial number.
NOTE: The Model IR400 is factory calibrated and needs no routine calibration. However, if the IR400 is to be installed at altitudes greater than 1000 ft (300m), it must be re-calibrated on-site (Section 4.1).
Model IR400
Figure 4: Model IR400 with a Junction Box Fasten the IR400 wires to a wiring connector inside the junction box, IR4000M or IR4000S enclosure. Once the two units are attached, you may replace the cover on the enclosure attached to the IR400 using the four captive screws, or leave it off until additional cabling from the enclosure is completed.
Model IR400
MODEL IR4000M
P/N 32580
0518 0539
CALIBRATION PROCEDURE 1. HAVE DETECTOR IN CLEAN AIR 2. APPLY MAGNET, AND WAIT FOR "AC" ON DISPLAY 3. REMOVE MAGNET, WHEN d# IS DISPLAYED, APPLY MAGNET TO SELECT DETECTOR # 4. REMOVE MAGNET, FLASHING "AC" IS DISPLAYED 5. WHEN "AC" IS STEADY, APPLY GAS 6. WHEN GAS IS DETECTED, "CP" IS DISPLAYED 7. WHEN CALIBRATION IS COMPLETE, "CC" IS DISPLAYED 8. REMOVE GAS
ALARM LED
WARN LED
S/N CONFIG
YEAR OF CONSTRUCTION
MAN
Model IR400
To Mount the IR400 The IR400 must rest horizontally to reduce the possibility of dirt and dust building up on the lens. The open slots of the gas passage must be straight up and down for the gas to rise up and through the unit. If the detector is installed with the metal section blocking the gas flow, it will slow down the detectors response.
Mount the attached junction box or IR4000M(S) enclosure using the two bolt holes. NOTE: There is also a duct mounting kit available from General Monitors with separate instructions.
10
Model IR400
Table 2: Wiring Chart The IR400 operates on nominal power of +24 VDC. The customer must provide primary DC voltage power, unless a General Monitors readout/relay display module with an internal power supply is used. Since the IR400 is designed to continuously monitor for leaks of hydrocarbon gas, a power switch is not included to prevent accidental system shut down. NOTE: Power should remain disconnected until all other wiring connections are made. The maximum distance between the IR400 and the power source is specified in Section 9.3.3.
11
Model IR400
3.3.2 4- 20 mA Output
A 4 to 20 mA output signal is provided by the Model IR400 and can be sent to a General Monitors readout/relay display module or any industrial device that can accept a 4 to 20 mA signal for computer based multi-point monitoring. The Analog Output connection provides a signal for use in displaying current LEL readings, special operation or fault conditions. The maximum distance between the IR400 and the device connected to the Analog Output signal is specified in Section 9.3.5. To access the 4-20 mA signal, connect the white lead from the IR400 to the signal-in terminal of the input unit. Refer to the manual of the display or other device being used for detailed instructions. Connect the black lead from the IR400 to the device Common. The Common connection serves both the analog signal and the power connections.
12
Model IR400
13
Model IR400
14
Model IR400
NOTE: This chapter describes how to use the junction box magnetic switch for stand-alone IR400 operation; the IR4000 manual describes how to use the IR4000 menus for IR400 system operation and configuration.
15
Model IR400
The portable purge calibrator is a compact, accurate and safe system containing a nonexplosive gas concentration. The lecture bottle is filled with a standard 50% LEL mixture of gas/air. Using a known gas/air mixture reduces the likelihood of error in field calibration. The hose and cup adapter that are included allow for quick calibrations and gas checks. Pre-mixed calibration gases at approximately 50% LEL are available in lecture bottles. ButaneC4H10 EthyleneC2H4 MethaneCH4 PropaneC3H8 EthaneC2H6 HexaneC6H14 PentaneC5H12
Spare bottles containing these gases may be ordered. Methane lecture bottles may be returned to General Monitors for refilling with the standard 2.5% by volume methane (50% LEL per ISO 10156 and NFPA 325).
16
Model IR400
General Monitors configures the IR400 with three calibration input options: magnetic switch (default), manual solenoid, and automatic remote gas calibration device (ARGC). To use the manual solenoid or ARGC, purchase a factory configured IR400 or change the calibration input via Modbus or HART.
Meaning Startup mode and critical fault for non HART unit Startup mode and critical fault if current range is low Zero, Calibration and Gas Check Mode Non critical fault All faults and startup if unit is configured with current range set to high 0 100% LEL or % by volume Over range Table 4: IR400 with HART Option
17
Model IR400
NOTE: This table is for Zero Switch / LED calibration input configurations only.
18
Model IR400
2 Seconds
Zeroing
Step 1
Zero Complete
Step 2
Gas Present
No Gas, Error
Step 3
Fault
Re-Calibration Complete
Figure 8. IR400 Zero Switch / LED During Zeroing and Calibration Review the general guidelines in Section 4.1 on Zeroing, Gas Check Tests, and Calibration. Make sure the windows are clean and there is nothing blocking the optical beam.
Zero Switch
19
Model IR400
Zero the Unit. Apply the General Monitors magnet that was included with the unit to the Zero Switch / LED for approximately three seconds. The LED in the switch will light to show proper placement. Remove the magnet and the LED will flash on for one second and off for one second to indicate that the unit is attaining a zero value (Zeroing in the figure). When the unit has finished zeroing, the LED will turn on and flash off quickly once per second for 30 seconds (Zero Complete in Figure 8).
NOTE: If an error occurs during the zeroing/calibration sequence, the LED will flash on and off rapidly (Fault).
NOTE: If an error occurs during the zeroing/calibration sequence, the LED will flash on and off rapidly (Fault). 2. Return to Normal Operation or Start Calibration. If the magnet is not applied again, the unit will return to normal operation. To continue on and calibrate, apply the magnet again and the unit will enter the calibration mode. The LED will flash off quickly once every half-second while the unit is waiting for gas to be applied (Waiting for Gas). 3. Apply Gas. Apply a 50% LEL gas using a Gas Check Kit with portable purge calibrator equipment (or 50% by volume of the gas being detected in nitrogen for units designed for monitoring 0 to 100% by volume). When the unit detects the gas, the LED will flash on for a half second every one and one-half seconds (Gas Present). If the unit does not detect the gas, the LED will flash off quickly once every half-second while the unit is still waiting for gas to be applied (No Gas Error). Once calibration has been completed, the LED will turn on and flash off once every second (Calibration Complete).
4. Return to Normal Operation. Remove the gas and the unit will return to normal operation once the gas has fallen below 5% of full scale.
20
Model IR400
NOTE: 3.
Once the detector is placed in gas check mode and the gas is applied, monitor the % LEL reading for the detector to see if it is functioning properly using the FMD or IR4000S display or Modbus/HART commands. When the reading is stabilized, it should be 50% LEL if the gas from the check kit is applied. Remove the gas. If the manual solenoid calibration input is configured, send the Modbus or HART command to turn off the solenoid.
4.
NOTE:
The unit will return to normal operation when the concentration drops below 5% full-scale. Since the IR400 zeros before calibrating, you must remove the gas completely before going to calibration mode.
Model IR400
T response times: According to CSA C22.2, T50 equals time to 50% of full-scale meaning 50% LEL and T90 which is 90% of the final reading. CSA C22.2 Section 6.9 states: Beginning with the gas sensing element in clean air it shall be suddenly exposed to a prepared mixture of gas in air having a concentration corresponding to 100% of the full scale gas concentration. From the instant of exposure to this gas mixture the instrument shall respond to provide an indication within the time specified as follows: 50% of full scale gas concentration in 10 sec, and 90% of maximum indicated gas concentration in 30 sec. The products offered by GM are not for the purpose of testing T response time, but as a method to allow a user to check that the unit is responding to gas and that final response is within tolerance. The products offered by GM are not for the purpose of testing T response time, but as a method to allow a user to check that the unit is responding to gas and/or that final response is within tolerance. If its required to demonstrate a reading of 50% LEL at site this can be achieved using the calibration cup, however it should be noted that you need to apply the test gas for approximately three minutes to get a reading of 50% LEL. This time is due to the ambient air located in the optical path of the detector having to be replaced progressively by the test gas. This replacement of ambient air is quicker at the beginning but longer for the last percentage because it is linked to the slow replacement of air dilution by the test gas. This test is only to indicate a gas level of 50% LEL within the calibration cup and is not a reflection of the response time of the detector.
22
Model IR400
5.0 Maintenance
The Model IR400 is calibrated at the factory and is fail-to-safe; once it is correctly installed and calibrated upon start-up, it requires little maintenance other than periodic cleaning, gas checks, zeroing and recalibration to ensure system integrity. Integrity checks can be performed using General Monitors Gas Check Kit (P/N 32548). WARNING: Disconnect or inhibit external devices such as Trip Amplifiers, PLCs, or DCS systems before performing any maintenance. NOTE: If an optical fault still occurs after cleaning and re-zeroing of an IR400 detector is complete, then you must return the unit to the factory for service. The systems full twoyear warranty will be voided if customer personnel or third parties damage the system during repair attempts or maintenance activities. Gassing into the screened splashguard will not provide a stable or accurate reading.
23
Model IR400
Figure 9. IR400 Optical Window Locations NOTE: Do not clean the windows while an IR400 unit is zeroing or in recalibration mode. The unit must be re-zeroed after cleaning.
5.4 Storage
The Model IR400 Gas Detector and IR4000 Monitor System should be stored in a clean, dry area, and within the temperature and humidity ranges noted for environmental specifications in Section 9.3.3 Electrical Specifications for the IR400, and the separate user manual for the IR4000. Insert the red dust caps into any vacant cable entry holes while the unit is stored.
24
Model IR400
6.0 Troubleshooting
The IR400 will alert the operator that there are problems in a number of ways. The table on the following page shows the value for the analog output, the value read through Modbus register 2 (if Modbus is being used), and the fault code, shown as F and a number, on the IR4000 display (if connected). If the IR400 is connected to a zero switch, then the LED in the zero switch will blink at a rate of 5 flashes per second to show that the detector needs attention. HART Troubleshooting 1. Verify that the IR400 is configured for HART 2. Verify that the HART modem or HART handheld device (375 or 475) is working by checking it against another HART field device 3. Verify that the HART software or handheld device has the DD for the IR400. If the DD is not present, download it from the HART Communication Foundation (HCF) website. 4. Verify the IR400 wiring 5. Verify that power is off when removing or attaching wires - Red to power - Black to common - White to a 250 ohm resistor, and the other end of resistor to common. Resistor tolerance should be +/-5% or less. - Verify HART modem or 475 wiring: Leads are across the 250 ohm resistor - After verifying the IR400 wiring, verify that power is on to the IR400 and to the HART modem or handheld device 6. Verify IR400 4-20mA output. When no gas and no faults are present, a voltmeter across the 250 ohm resistor will read 1.0 volt DC.
25
Model IR400
FAULT CODE
DESCRIPTION
POSSIBLE CAUSE
CORRECTIVE ACTION
F0 F1
1)Fouling of the detector windows or beam Gas Concentration is path is excessive and requires attention 1) Clean detector beam path and windows. Excessively Negative Gas Concentration is 1) Detector windows or beam path are Negative beginning to be obscurred. 1) Clean detector beam path and windows.
1) Cal bottle empty and time-out occurred 1) Obtained filled CAL bottle and re-CAL. 2) Failed to remove gas at end of CAL and 2) Remove gas when directed. time-out occurred. 3a) Check CAL cup for proper seating on unit. 3) Leaky CAL cup resulting in unstable signal. 3b) Check CAL cup seal and replace CAL cup if damaged. 4) Calibrate at a much less windy time or shield the unit and CAL cup from the wind.
F2
Bit 6
1.25
4) Attempted to CAL with too much wind resulting in unstable signal. 1) Detector windows or beam path are blocked by dirt, spider webbing or other foreign matter. 2) Detector or source failure.
F3
Bit 2
Beam Block
2) The IR400 must be returned to the factory or authorized service center for repair. 1) Ensure baud rate, data format and address matches at both the IR400 and the bus master. 2) Check and correct wiring. 3) Check and correct wiring. 1) Check and ensure proper connection of wire to board connector 1) Check the supply voltage and replace power supply if necessary 2) Check the supply voltage at the IR400 field and adjust supply to +24VDC at the IR400 or replace wiring with larger gauge. 1) Cycle power, wait 2 minutes, if the fault clears then check all menu settings and recalibrate the IR400. 2) Call your local General Monitors representative for advice. 1) Ensure clean air is available for zeroing or provide zero air to the IR400 during zeroing. 1) Remove the gas
F4
1.25
F5
Bit 4
1.25
Communications Fault 1) Incorrect communications set-up. (Applies to IR4000M display only) 2) Communications wiring open. 3) Communications wiring shorted. CAL Wire (brown) 1) CAL Wire (brown) not in connector Shorted 1) Power supply not outputing greater than 20VDC. Low Supply Voltage 2) Voltage loss due to wiring.
F6
Bit 5
1.25
1) Internal memory glitch F7 Bits 9 - 13 & 15 1.25 0 Electronics Error 2) Internal error with the electronics. F8 Bit 7 1.25 0 Failed to Zero CAL Check Period Exceeded 1) Unstable signal due to gas present. 1) Test gas still present after gas check completed.
F9
Bit 8
1.25
26
Model IR400
27
Model IR400
Phone: +353-91-751175 Fax: +353-91-751317 Email: info@gmil.ie Phone: +65-6-748-3488 Fax: +65-6-748-1911 Email: genmon@gmpacifica.com.sg
28
Model IR400
9.0 Appendix
9.1 Warranty
General Monitors warrants the Model IR400 to be free from defects in workmanship or material under normal use and service within two (2) years from the date of shipment. General Monitors will repair or replace without charge any such defective equipment found to be defective during the warranty period. Full determination of the nature of, and responsibility for, defective equipment will be made by General Monitors personnel. Defective or damaged equipment must be shipped prepaid to General Monitors plant or representative from which shipment was made. In all cases this warranty is limited to the cost of the equipment supplied by General Monitors. The customer will assume all liability for the misuse of this equipment by its employees or other personnel. NOTE: The Model IR400 Infrared Point Detector is easy to install; however, you should read and understand this manual before attempting to install or operate the device. It includes important safety information. All warranties are contingent upon proper use in the application for which the product was intended and do not cover products which have been modified or repaired without General Monitors approval, or which have been subjected to neglect, accident, improper installation or application, or on which the original identification marks have been removed or altered. Except for the express warranty stated above, General Monitors disclaims all warranties with regard to the products sold, including all implied warranties of merchantability and fitness and the express warranty stated herein are in lieu of all obligations or liabilities on the part of General Monitors for damages including, but not limited to, consequential damages arising out of/or in connection with the use or performance of the product.
29
Model IR400
Control Electronics
The Model IR400 operates from an unregulated +24 VDC (nominal) input, which is fed to an onboard power-supply that produces all of the necessary voltages within the unit. The microprocessor constantly monitors the infrared wavelengths and performs mathematical operations on these values in conjunction with values obtained during the factory set-up process. The microprocessor generates output information and feeds it to the digital analog converter to produce a 4 to 20 milliampere (mA) signal that is proportional to the 0 to 100% LEL (or methane % by volume) concentration of gas at the sensor. The microprocessor program also monitors other conditions such as the supply voltage and the optical path integrity. The Model IR400 provides a two-wire RS-485 addressable communications link conforming to the Modbus protocol that is used to monitor the IR400s status and settings in order to simplify installation and maintenance.
30
Model IR400
9.3 Specifications
9.3.1 System Specifications
Detector Type: Detector Life: Measuring Range: Zero Drift: Response Time:
(With 100% LEL methane applied) (With 100% LEL ethylene applied)
Infrared absorption Greater than 5 years 0 to 100% LEL < 2% per year T50 7 seconds, T60 8 seconds T90 10 seconds T50 5 seconds, T60 6 seconds T90 11 seconds 3% FS 50% FS, 5% FS > 50% FS Methane, propane, ethane, ethylene, butane, hexane, pentane at LEL levels defined by ISO 10156/NFPA 325 and at levels defined by IEC 61779-1. Consult factory for other calibrations. DC110: Multi-Channel, Rack Mounted1 DC130: Dual Channel, Rack Mounted2 TA102A: Single Channel, Zero Two Series3 IR4000 display and relay alarms Re-calibration Error Optics Failure/Blockage Low Supply Voltage Reference or Active Lamp Failure Heater Failure Time to Re-zero unit Program Memory Checksum Error (EPROM) Data Non-Volatile Memory Checksum Error (EEPROM) Short Circuit on CAL_IO Wire Two years CSA FM CE Marking ATEX IECEx SIL 3 suitable
Malfunctions Monitored:
Warranty: Approvals:
FM required statement: This Approval does not include or imply Approval of apparatus to which the subject instrumentation may be connected. In order to maintain an FM Approved system, the apparatus to which this instrument is connected, must also be Approved by FM Approvals.
1 2
DC110 is not approved for use in ATEX installations DC130 is not approved for use in ATEX installations 3 Rev E S/W and up for EU installations 31
Model IR400
NOTE: Customer supplied PSU must meet requirements IEC 1010-1 limiting current to 8A under Fault conditions, in order to comply with CE Marking requirements. Analog Signal: Range Load (max. resistance) Current Level (mA) 0 1.25* 1.5* 2* 4 20 20.1 21.7 0 - 21.7 mA 600 Meaning Startup mode and critical fault for non HART unit Startup mode and critical fault for HART unit Zero, Calibration and Gas Check Mode Dirty Optics 0 100% LEL Over range
* HART units can be configured to never output current less than 3.5 mA if the host equipment is incapable of working below this level. Area Classification: Class I, Divisions 1 & 2, Groups B, C and D (Ta = -40C to +75C), (Ta = -40C to +60C for ethylene only), IP66, Type 4X Ex d, IIB+H2 T5 Gb, IP66 (Ta = -60C to +75C) Ex t IIIC T100C Db Complies with EN50270, EN61000-6-4
RFI/EMI Protection:
32
Model IR400
Table 6: Analog Current Output * HART units can be configured to never output current less than 3.5 mA if the host equipment is incapable of working below this level.
Table 7: Maximum Distance between IR400 and Power Source Analog Output Signal The maximum total distance between the IR400 and a device with a 500-Ohm input impedance varies according to the wire size. AWG 14 16 18 20 Stranded Cable (/1000 ft) 2.525 4.016 6.385 10.15 Feet 9000 5200 3800 2400 Meters 2740 1585 1160 730
33
Model IR400
-40F to +167F (-40C to +75C) -40F to +140F (-40C to +60C) -76F to +185F (-60C to +85C) 5 to 99% RH non-condensing Accuracy is not affected by humidity as long as no condensation accumulates on the windows
9.5 Communications
9.5.1 RS-485 Interface
The Model IR400 has built-in serial communications in the form of a half duplex RS-485 digital serial interface designed to conform to EIA-485 specifications. The format is in binary data transferred at 9600 baud with 1 start bit, 8 data bits, 1 stop bit and no parity. The bus master sends a command message to the IR400, which is comprised of 5 bytes of data in the following format: The first byte is the address of the slave device (IR400). The second byte is the Command Word. The third byte is the Command Data. The last two bytes are a 16-bit checksum calculated by performing a 16-bit addition of the first three bytes of the message and placing the result in the check sum bytes. A 1" in the most significant bit of the Command Word (byte 2), tells the IR400 to change the settings to those given in the Command Data (byte 3). A 0" in the most significant bit of the Command Word (byte 2), tells the IR400 to return to the current settings. In this case the Command Data (byte 3) will be all 0s. The IR400 then responds by sending back a 5-byte message in the following format: The first byte is the address of the IR400. The second byte is an echo of the Command Word sent by the bus master. The third byte is the data requested by the Command Word. The last two bytes are the 16- bit check sum.
9.5.3 HART
The IR400 HART Field Device Specification provides details on HART commands. specification is available from the General Monitors website. The
34
Model IR400
35
Model IR400
36
Model IR400
9.7.4 Accessories
31305-1 31421-3 31305-2 31421-4 32554-1 32545-1 32545-2 32545-3 32545-4 31306-1 32548-Specify Gas 31420-1 31545-1 Junction Box with magnetic switch, CSA/FM Junction Box with magnetic switch, ATEX Junction Box without magnetic switch, CSA/FM Junction Box without magnetic switch, ATEX Calibration Cup / Flow Block Splash Guard standard Splash Guard for use with remote cal Splash Guard with no screen for areas with occasional moisture Splash Guard with no screen for remote cal Duct Mount Junction Box Gas bottle with regulator and calibration cup Flow Block for gas sampling system Rain Guard Assembly
37
Model IR400
38