S1-554 2 UG en V06
S1-554 2 UG en V06
S1-554 2 UG en V06
USER MANUAL
G
Safety Warning must be observed during use.
n n n n
SAFETY WARNINGS
The circuit under test must be switched off, de-energised, isolated and checked to be safe before insulation test connections are made. Make sure the circuit is not re-energised whilst the instrument is connected. Circuit connections must not be touched during an insulation test. After completing a test, capacitive circuits must be completely discharged before disconnecting the test leads. Capacitive charges can be lethal. Tested items should be firmly shorted out with a shorting link, after discharge, until required for use. This is to guard against any stored dielectric absorption charge subsequently being released thereby raising the voltage to potentially dangerous levels. The voltage indicator and automatic discharge features should be regarded as additional safety features and not a substitute for normal safe working practice. It is rare, but in certain circumstances, breakdown of the circuit under test may cause the instrument to terminate the test in an uncontrolled manner, possibly causing a loss of display while the circuit remains energised. In this event, the unit must be turned off and the circuit discharged manually. Test leads, including crocodile clips, must be in good order, clean and with no broken or cracked insulation. The instrument should not be used if any part of it is damaged. Insulation testing in wet weather conditions might be hazardous. It is recommended that this instrument is not used in these circumstances. If this is unavoidable, the user must take all necessary precautions. This instrument is not intrinsically safe and must not be used in hazardous atmospheres.
n n n
NOTE THE INSTRUMENT MUST ONLY BE USED BY SUITABLY TRAINED AND COMPETENT PERSONS. Users of this equipment and/or their employers are reminded that National Health and Safety Legislation requires them to carry out valid risk assessments of all electrical work so as to identify potential sources of electrical danger and risk of electrical injury such as inadvertent short circuits. Where the assessments show that the risk is significant then the use of fused test leads may be appropriate.
CONTENTS
Safety warnings 2 Introduction 4 General description 4 Features 4 Cleaning 5 Power lead and battery charging 5 Instrument controls and indicators 6 Power On/Off button 7 Test voltage s and t buttons 7 Test start / stop button 7 /I button 7 B button 10 Fn button 10 Mode button 10 Timer s and t buttons 11 Record button 11 Recording to on-board memory 11 Downloading results 11 Deleting results 11 Recording to a PC 12 High voltage warning LED 12 Line input present LED 12 Test terminals 12 Guard terminal 12 RS232 / USB connections 13 Battery bar graph 13 Voltage at terminals 13 Timer indicator 13 Digital display 13 Analogue display 14 Secondary display 14 Pre-Test / during test key action table 14 Breakdown detection 15 Noise detection 15 Test mode summary 16 Hardware filter 16 Insulation resistance IR test 16 Firmware filtering 16 Breakdown mode / burn mode 17 Alarm limit mode 17 Step voltage SV test 17 Polarisation index PI test, and dielectric absorption ratio DAR test 17 Dielectric discharge DD test 18 Measurements above 100 G 18 Circuit block diagram 19 Specifications 20 Accessories 21 Repair and warranty 22 Battery replacement Symbols used on the instrument are:
F Caution: risk of electric shock G Caution: refer to accompanying notes t Equipment protected throughout by Double
Insulation (Class II)
INTRODUCTION
General Description The S1-554/2 is a microprocessor controlled 5 kV insulation tester offering measurement capability of up to 15 T. The instrument performs automatic tests and has data storage and data retrieval facilities. Polarization index and dielectric discharge tests are performed automatically, and test duration and voltages can be adjusted according to user preference for these tests. A step voltage test can be performed automatically, with a default voltage of 1 kV, and test duration of 5 minutes. It is possible for the user to set different voltages and test durations for the step voltage test. The S1-554/2 incorporates a hardware filter designed to tolerate at least 4 mA rms of noise current at 50Hz and above. This filter is enabled by default, but may be switched off in order to speed up the settling time when there is little noise current. The Insulation Resistance mode offers additional firmware filtering to average out slow variations during testing. The S1-554/2 can be powered from the mains supply or by its own internal rechargeable battery, which provides for at least 6 hours of continuous testing with a 100 M load. A battery level indicator on the LCD display indicates battery capacity. Connecting power to the mains supply connector will automatically charge the battery whether the instrument is switched on or off, except during testing. A high level of internal isolation allows the instrument to be used while the unit is supplied from the mains. An internal battery management system switches the instrument off after ten minutes of inactivity. If the battery approaches a very low state of charge the instrument turns itself off, and mains power must be applied before the instrument can be used again. Recorded test results and settings will not be lost when the instrument is switched off. A comprehensive LCD display shows resistance, current, filter setting, 4 capacitance, time constant, voltage, timer measurements, and figure of merit measurements such as polarisation index. Features
n n
Continuous resistance or current measurement on main display. Optional resistance, current, or figure of merit measurements on secondary display. Standard test voltages 250 V, 500 V, 1 kV, 2.5 kV, 5 kV Non-standard test voltages selectable between 50 V and 1 kV in 10 V steps, and selectable between 1 kV and 5 kV in 25 V steps. Test modes include insulation resistance, polarisation index, step voltage and dielectric discharge. 4 mA rms hardware noise current filter (selectable) Firmware filter 10, 30, and 100 seconds (optional) The dielectric absorption ratio is automatically calculated if the corresponding timers are set. Either burn or breakdown selectable in insulation resistance mode. Insulation Alarm Limit available in insulation resistance mode. Programmable timers include a main test duration timer, plus T1 and T2 timers for time resistance method type tests. Load capacitance and time constant measurements displayed at end of test. LCD backlight. Data storage, data retrieval and real time data output. USB or RS232 communications. Battery level / charge level indicator.
n n
n n n
n n n
n n n n
If using a fused plug, ensure that it is fitted with a 3 Amp fuse. The instrument can be powered from 85 - 265 Vrms a.c. at 50/60 Hz. The battery will charge as long as the mains supply is connected, except when a test is in progress. The power On/Off button has a green LED, which illuminates when mains power is present. For optimum battery life, charge the battery after each use. A completely exhausted battery will take 14 hours to recharge.
Line input
USB
RS232
Test modes
Burn indicator
Alarm
Data recording Timers Main, T1, T2 Analogue display High voltage warning indicator
Voltage s t selectors
S1-554
Breakdown indicator
Backlight on/off
Function
Mode
Test start/stop
Digital display
Power On/Off button The instrument will only turn on if this button is pressed, held and then released when the display responds. The instrument will not turn on if the button is released before the display responds, or if the button is held down for too long. This is a safety feature to prevent the instrument being inadvertently turned on. The instrument is turned off either by pressing the button again, or if the instrument is running on the battery, by timing out after 10 minutes of inactivity. Upon switching the instrument on the display will first show Ini while it undergoes a self-checking routine. When Ini disappears, the instrument is ready for use. Test voltage s and t buttons Using these buttons one of five test voltages can be selected: 250 V, 500 V, 1 kV, 2.5 kV, and 5 kV. The selected voltage is shown on the display. A non standard voltage between 50 V and 5 kV can be selected by holding down the Fn function button whilst operating the buttons. The selectable voltage is adjustable in 10 V steps between 50 V and 1 kV, and adjustable in 25 V steps between 1 kV and 5 kV. An auto repeat facility is enabled when the button is held down, allowing faster travel through the range. If there is an external voltage greater than 50 V on the test leads, the high voltage warning indicators are flashed, and the display shows this voltage instead. The instrument will not perform a test if this voltage is greater than 80 V. During a test the display shows the actual voltage on the test leads. If the
test voltage is changed during a test, the new test voltage will be displayed briefly. When the test has stopped, the display continues to show the voltage present on the test leads. Pressing either test voltage s or t button will then display the test voltage immediately before the end of the test. Test start / stop button A test will only start if this button is pressed, held and then released as soon as the red high voltage warning indicator LED lights. The LCD and red LED high voltage warning indicators flash when the test starts. A test will not start if the button is released before the red LED shows, or pressed continuously for longer than 5 seconds. This is a safety feature to prevent a test being started inadvertently. The presence of a voltage greater than 50 V on the test leads is indicated with flashing high voltage warning indicators. Testing is disabled if the external voltage exceeds 80 V. Testing will stop if the test start / stop button is pressed again, the preset test time is reached, or, if the unit is not in burn mode, insulation breakdown is detected. When a test has finished, the instrument will discharge the load, which may take some time. The operator must always check the load has been discharged before touching the test leads. /I button Pressing this button toggles the digital and secondary displays. The details available when scrolling the secondary display will depend upon the test mode selected, whether timers T1 and T2 have been set, and the duration 7
of the test. The digital display toggles between insulation resistance and current. In the IR, PI, and DD modes the secondary display initially shows the PI (polarity index), DAR (dielectric absorption ratio), and, on completion of the test, the TC (time constant) and capacitance measurements. Further presses of the /I button cycles the display through insulation resistances, currents, and the filter setting. In the SV (step voltage) mode, the secondary display cycles through insulation resistances, currents, and the filter setting. Figures 1 to 4 illustrate the display status for sequential /I key presses in the different modes.
Press /I
Press /I
Press /I
Press /I
Figure 1 - The result of an insulation resistance test (IR) Test settings: T1 and T2 times set in order to measure the DAR Test conditions: test runs for longer than 10 minutes as this is required for a PI reading. 8
Press /I
Press /I
Press /I
Press /I
Press /I Press /I
Press /I Press /I
Figure 2 - The result of a polarity index test (PI) Test settings: T1 and T2 times set in order to measure the DAR Test conditions: timer defaults to 10 minutes, as this is required for a PI reading
Figure 3 - The result of a dielectric discharge test (DD) Test settings: T1 and T2 times set in order to measure the DAR Test condition: test runs for longer than 10 minutes as this is required for a PI reading. 9
test voltage buttons. Ts and Tt represent the timer buttons. Button Press Function Comment Fn + Vs Increment in 10 V steps Voltage range between 50 and 1000 V Fn + Vt Decrement in 10 V steps Voltage range between 50 and 1000 V Fn + Vs Increment in 25 V steps Voltage range between 1000 and 5000 V Fn + Vt Decrement in 25 V steps Voltage range between 1000 and 5000 V Fn + Ts or Tt Cycle through timers Select main timer, T1 or T2 Fn + Mode Cycle through IR modes Select breakdown, burn or alarm limit for IR test Fn + /I Cycle through filter options Additional filtering options are available in IR mode Fn + Figure 4 - The result of a step voltage (SV) test Test condition: timer defaults to 5 minutes and test voltage defaults to 1000 V. Fn + Download mode Clear stored data See Downloading results The record button is pressed twice
Press /I Press /I
Press /I
B button
Pressing this button toggles the display backlight on and off. Fn button The function button when pressed allows other keys to perform another function. This is summarized in the table below. Vs and Vt represent the 10
Mode button Press down on the MODE button to cycle through and select the test mode. Modes of test to be chosen from include an insulation resistance IR test, a polarization index PI test, a step voltage SV test, and a dielectric discharge DD test. The insulation resistance IR test operates in either burn or breakdown mode, with an option of setting a resistance threshold alarm limit. These
modes can be cycled through and selected by holding down the Fn button and pressing the Mode button. Timer s and t buttons The main timer can be set up to 99mins 59secs. The Ts button increments the time, and the Tt button decrements the time in ten seconds steps. An auto repeat facility allows the time to be set more quickly. Setting a time of 00:00 disables the timer. With the timer disabled a test has to be manually stopped. To select timer Tmain, T1 or T2, hold down the Fn button whilst repeatedly pressing the Ts or Ttbuttons. To set the selected timer, release the Fn button, and the use the Ts or Ttbuttons. The minimum timer setting is 15 seconds for test voltages of 1000 V or more, and 30 seconds below this. Note: T2 cannot exceed the time on main timer unless it is disabled (00:00). T1 cannot exceed the time on T2. Record button This button is used to start and stop recording. Recording can only be activated before testing. When data recording is enabled the record symbol flashes. Data is stored in solid-state memory and under normal circumstances will maintain its integrity for in excess of ten years, but may rarely be corrupted or lost by external influences such as transients and static discharge. Megger Limited cannot accept responsibility for any losses of data. Regular downloading to a PC using software such as Download Manager will substantially reduce any such risk.
Recording to on-board memory Press the record button to start and stop recording. When data recording is enabled the record symbol will flash on and off repeatedly. Recording can only be activated before testing commences. Results are stored at 15, 30, 45 and 60 seconds. After 60 seconds, at minute intervals up to 10 minutes. After 10 minutes, results are recorded at 5 minute intervals until the test terminates. At each interval the recorded data will contain selected voltage, test time elapsed, voltage applied, leakage current, and insulation resistance. Downloading results Connect the instrument to the RS232 / USB port of a PC running Download Manager. Refer to section RS232 / USB connection for set up details. Start Download Manager on your PC, select the S1-554 driver and right click the icon. Select Download. Switch the instrument on and wait until initialisation is complete. Press the function key along with the record key. The instrument now displays dld to indicate download mode. Press and hold down test button until download begins, shown by analogue display lighting clockwise. Results will not be erased during this operation and so may be downloaded repeatedly. Deleting test results Switch the instrument on and wait until initialisation is complete. Hold down the function key and press the record key twice. The instrument now displays clr to indicate clear mode. Press and hold down test button until clear process begins, shown by analogue display lighting anti clockwise. Press the MODE button to exit without deleting the results. 11
Recording to a PC While carrying out a test, the instrument will output the test voltage, test current and resistance every second. Refer to section RS232 / USB connection for setup details. Connect the instrument to the RS232 / USB port of the PC. The data may be captured with Microsoft HyperTerminal or another suitable programme. High voltage warning LED This is a red LED next to the TEST button on the front panel. The LED flashes when the voltage on the test inputs exceeds 50 V. Line input present LED This is a green LED next to the power On/Off button on the front panel. It is illuminated whenever the mains power is connected. Test terminals There are three test terminals marked +, - and G. These terminals are designed to accept only the test leads supplied. Shutters across the terminals prevent accidental ingress of dirt and other objects. Test lead plugs interlock with the shutters and are released by rotating the test lead plug a quarter turn. The Guard terminal is explained below and is only used in cases where surface leakage currents need to be eliminated. Most measurements use just the + and terminals. The instruments internal voltage generator drives the + terminal with respect to the terminal, current being measured in the terminal.
Guard terminal For basic insulation tests and where there is little possibility of surface leakage affecting the measurement, it is unnecessary to use the guard terminal i.e. if the insulator is clean and there are unlikely to be any adverse current paths. However in cable testing for example, there may be surface leakage paths across the insulation between the bare cable and the external sheathing due to the presence of moisture or dirt. Where it is required to remove the effect of this leakage, particularly at high testing voltages, a bare wire may be bound tightly around the insulation and connected via the third test lead to the guard terminal G.
The guard terminal is at the same potential as the negative terminal. Since the leakage resistance is effectively in parallel with the resistance to be measured, the use of the guard causes the current flowing through surface leakage to be diverted from the measuring circuit. The instrument therefore reads the leakage of the insulator, ignoring leakage across its surface. The display will show FUS if the internal guard terminal fuse is found to have blown. The instrument must be switched off to clear the message before further testing is permitted. The fuse should be replaced by an authorised repairer. The instrument may be used in the mean time if the guard terminal not used. Refer to notes regarding measurements above 100 G on page 17.
12
RS232 / USB connection Data can be transferred to a PC via an RS232 or USB port. If using the RS232 port, use the null modem cable supplied. The RS232 settings are 38400 Baud, 8 data bits, 0 parity, 1 stop bit, no flow control (handshake). If using the USB port, ensure that the USB driver supplied on the accompanying product CD has been installed BEFORE connecting the instrument. Installation instructions are also to be found on the CD. Programmes such as Megger Download Manager may be used to download the results stored in memory. Programmes such as Microsoft HyperTerminal may be used to record real time data. Battery bar graph This is a battery symbol on the LCD display comprising 4 pairs of segments. The battery is monitored continuously when the instrument is turned on. The charge remaining in the battery is shown in the table below. Fully charged battery
When mains power is present the indicator shows the battery is being charged by animating the segments of the bar graph. Voltage at terminals The test voltage will default to that selected in the previous test. If the instrument has been switched off since the previous test, the test voltage defaults to 250 V. If there is an external voltage greater than 80 V present, this will be displayed regardless of changes made to the test voltage. In this case the instrument will not perform a test. The high voltage warning indicators flash, and the beeper sounds, to warn of the hazard until the external voltage becomes less than 50 V. When testing, the voltage displayed is the voltage present at the test terminals of the instrument. If the test voltage is changed, by pressing either of the test voltage s and t buttons during a test, the new test voltage is displayed momentarily. After testing, the voltage displayed is the voltage present on the terminals. To see what the voltage was immediately prior to the end on the test, press either test voltage s or t buttons. Timer indicator The timer shows minutes and seconds. At the start of a test the timer will start from zero and at the end of a test it will stop. The duration of the last test remains on the display until another test is started. Digital display The digital display shows the resistance or current being measured during a test. The /I button toggles between the two. After a test the display 13
Tests cannot be started, and the battery may fail at any time
Symbol flashes when there is not enough charge for a test. The instrument then turns itself off
shows the last measurement made until the timer or voltage test settings are changed, or the test start/stop button is pressed. Analogue display This simulates an analogue meter movement to give the user a better feel for how a measurement is progressing. The analogue display shows resistance only. The display is also used to indicate how result download and deletion of results is progressing. Secondary display This part of the display shows the results of time resistance method tests.
Vs/Vt Increase / decrease the test voltage in major steps Vs AND Vt Ts/ Tt Sets voltage to 500V
Increase / Increase / decrease the decrease the test voltage in test voltage in major steps minor steps (IR only) Show set voltage
Increase / Display set time Select main decrease the on main timer timer T1 or T2 time of the briefly selected timer Display set Reset selected time on main timer to zero timer briefly Cycle through IR modes
Display time set for main timer, T1 or T2 briefly Display set time on main timer briefly
/I Cycle through Cycle through Select filter Select firmware results display results display option filtering level of last test of current (IR only, completed active test hardware filter ON) Record Turn recording Select On/Off download or clear stored data
14
Vs and Vt represent the test voltage buttons. Ts and Tt represent the timer buttons.
On screen warning codes Breakdown detection If insulation breakdown is detected, the unit aborts the test and flashes brd and the final test result alternately in the main digital display. This display is disabled when in burn mode as the unit will continue the test without aborting. The instrument detects breakdown in terms of a rapid drop in test voltage commensurate with that caused by an arc across or though the insulation under test. Noise detection Excessive noise during test In environments such as HV switch yards or substations electrical noise levels can be high enough to effect insulation test measurements. This noise takes the form of induced currents in the measurement circuit. The noise current is either picked up by test leads or induced directly in the item under test. The MIT510/2, MIT520/2, MIT1020/2, S1-552/2 and S1-1052/2 are all capable of accurate measurements with up to 2mA of induced noise current. The S1-554/2 and S1-1054/4 are design for very high noise environments and can provide accurate measurements with up to 4mA of noise current present. A level of noise exceeding the instruments specified capability will be indicated by an alternating sine wave on the display. The test continues while displaying the noise. If the noise drops back down the normal display is resumed.
The effects of noise can be reduces significantly with the use of high quality Megger screened test leads. The picture displayed on the right shows the expected screen result of an IR test on with a 5000V test voltage after 2 minutes 9 seconds. The black and grey waveforms are displayed alternately while the noise is present. When the noise is removed the wave form is replaced by the correct reading.
Test aborted due to extremely high noise Should an extremely high level of noise be present, particularly at the start of a test, the instrument will indicate nSE (Noise). The instrument will require switching off and on again to reset. In certain rare circumstances breakdown of the circuit under test may interfere with the sensitive current measuring circuit. In this event the unit stops and also displays nSE. The instrument will again require switching off and on again to reset.
15
Press and hold the Fn button while repeatedly pressing the /I button to scroll through the filter options. Breakdown mode / burn mode The insulation resistance IR test operates in either the Burn or Breakdown mode of operation. The default mode is breakdown. To change mode press and hold the function button, then press and release the mode button repeatedly until the desired mode is indicated by the flashing symbol on the display. Release the function button to select it. In the breakdown mode the Breakdown Indicator will be illuminated refer to page 6. In this mode the test will automatically terminate should the insulation break down to prevent damage to the insulation under test. In the burn mode the Burn Indicator is illuminated refer page 6. The burn mode disables the normal breakdown detection and enables the insulation test voltage to continue even after breakdown of the insulation. This will enable the location of the failure to be seen and heard. Due to the potential damage that could occur, the unit produces two long beeps when starting a test with burn mode activated. Alarm limit mode The insulation resistance IR test has an option of setting an alarm limit. If this mode is selected, the instrument will beep should the resistance reading exceed a user selectable threshold. To set the limit, hold down the function button, then press and release the mode button repeatedly until the A symbol is flashing on the display. Use the timer s and t buttons to set the resistance threshold between the limits of 10 k and 15 T. Release the function button to save the current limit. Adjusting the alarm threshold level automatically activates the alarm limit mode. This is indicated by the A symbol turning solidly on.
To toggle the alarm limit mode on/off, hold down the function button, then press and release the mode button repeatedly until the A symbol is flashing on the display. Release the function key to toggle. Step voltage SV test This is a test based on the principle that an ideal insulator will produce identical readings at all voltages, while an insulator which is being over stressed, will show lower insulation values at higher voltages. The main timer and test voltage settings can be adjusted if desired from their default values of 5 minutes and 1 kV respectively. During the test the applied test voltage incrementally steps by one fifth of the test voltage setting (final value) every minute, for 5 minutes, taking successive measurements until the final voltage is reached. Readings for the first 4 recorded values are displayed under the consecutive segment headings 1m to 4m. The 5 minute reading is displayed by the main display. If the range of measured insulation between readings is too wide for the instrument to display, then those readings too small compared to the final reading will be represented by ---. Polarisation index PI test, and dielectric absorption ratio DAR test The PI test is a particular example of a time/resistance method, which takes the ratio of the insulation measured at 1 minute and at 10 minutes. Good insulation generally shows an increase in resistance over a 10 minute period. Readings for contaminated insulation are fairly constant because any absorption effects are masked by high leakage currents. On completion of the test the polarisation index is displayed under the segment heading PI.
17
Polarisation Index (PI) = R10 min R1 min The Dielectric Absorption Ratio is the term applied to the polarisation index using other time intervals set by T1 and T2. If timers T1 and T2 are both set, then the insulation resistance measured at these times is also recorded. These are displayed together with the calculated dielectric absorption ratio under the T1, T2, and DAR segments respectively. The instrument can be made to display either the insulation resistances, insulation currents or ratios plus capacitance measurements by using the /I toggle button. See section /I button for cycling through results and toggling units. Dielectric Absorption Ratio (DAR) = RT2 min RT1 min
The main timer will default to 30 minutes, which is normally sufficient time for full absorption to take place in an insulation material. The default test voltage is set to 500 V. The DD test requires the instrument to measure the discharge current 1 minute after the removal of the test voltage. At this time the capacitive current should be insignificant compared with the released absorption current. On completion of the test, the instrument uses this measurement along with the test voltage and calculated capacitance to produce a figure of merit indicating the quality of the insulation. Dielectric Discharge (DD) = I1 min VxC where I is the measured current expressed in milliamps (mA), V is the test voltage in Volts (V), and C is the measured capacitance in Farads (F). Measurements above 100 G Measurements up to 100 G can be made without any special precautions, assuming that the test leads are reasonably clean and dry. The guard lead can be used to remove the effects of surface leakage if necessary. When measuring resistances above 100 G, the test leads should not be allowed to touch each other, or any other object since this will introduce leakage paths. Sharp points at the test lead connections should also be avoided since this will encourage corona discharge. The output is isolated, and so will float relative to ground such that the positive terminal is at plus half of the test voltage, and the negative terminal is at minus half of the test voltage with respect to ground. Leakages therefore occur between the positive terminal and ground, between the negative terminal and ground, and directly between the positive and negative terminals. These leakages have a significant effect
Dielectric discharge DD test The DD test is a diagnostic insulation test that allows ageing deterioration, and voids in the insulation to be assessed. The result is dependent on the discharge characteristic, so the internal condition of the insulation is tested, largely independent of any surface contamination. On discharge the capacitive component of the discharge current decays from a high value with a relatively short time constant of a few seconds. The other current component, comprising the released absorption current, decays from a lower value with a relatively long time constant of up to several minutes. If this component of the discharge current is large (>7 @ 500 V test voltage) then the insulation condition is poor. 18
- Test V 2
When making measurements above 100 G therefore, the user should ground the Guard Lead where possible, otherwise parallel leakage paths may occur. Alternatively, screened leads are available as an optional accessory from Megger. The lead to the negative terminal is fully screened. The screen is plugged into the Guard terminal, diverting any stray leakage currents. This considerably improves measurements made with a floating output, where the leads might touch each other or anything other than the test piece. Circuit block diagram
R1 100 M High voltage source Current Limit Voltage control Fuse Cable under test C1 Volts Discharge resistance G R2 +
Ground
If the guard lead is connected to ground, then since the negative terminal is at the same voltage as the guard terminal, the leakage into the negative terminal will be considerably reduced. This will improve accuracy because the current flowing into the negative terminal is measured by the instrument and used to calculate resistance. This technique is only permissible if the item under test is isolated from ground. Isolated in this context means insulated by a resistance of at least 5 M for the positive terminal, or at least 10 k for the negative terminal. Conversely, if the positive terminal is grounded, then the negative terminal will be at a voltage equal to the test voltage relative to ground, which will result in an increase in leakage current, and worsening of measurement accuracy.
Ref
Current
+ Test V
+ -
0V Guard Ground For 5 kV instruments C1 = 47 nF, R1 = 50 k, R2 = 40 k For 10 kV instruments C1 = 15 nF, R1 = 156 k, R2 = 110 k
19
SPECIFICATIONS
Voltage input range 85-265 V rms, 50/60 Hz, 60 VA Battery life Typical capacity is 6 hours continuous testing at 5 kV with a 100 M load Test voltages 250 V, 500 V, 1 kV, 2.5 kV, 5 kV ranges, adjustable in 10 V steps from 50 V to 1 kV, and 25 V steps from 1 kV to 5 kV Accuracy (23 C, 5 kV) 5000 V 2500 V 1000 V 500 V 5% to 1 T 20% to 10 T 250 V Capacitance measurement accuracy (23 C) 5% 5 nF Voltage output accuracy (0 C to 30 C) +4%, -0% 10 V of nominal test voltage at 1 G load Current measurement range 0.01 nA to 5 mA Current measurement accuracy (23 C) 5% 0.2 nA at all voltages Interference rejection 1 mA per 250 V up to a maximum of 4 mA (selectable) Filter 10, 30 and 100 second time constants (selectable) Timer range Counts up to 99 minutes and 59 seconds from start of test 15 second minimum setting for test voltage 1000 V 30 second minimum setting for test voltage <1000 V Test regimes Auto IR, PI, SV, DD DAR is calculated automatically if timers T1 and T2 are set Interface RS232: 38400 Baud, 8 data bits, 0 parity, 1 stop bit, no flow control. USB Data storage Data stored: selected voltage, test time elapsed, voltage applied, leakage current, and insulation resistance. The PI, DAR, capacitance, time constant and DD values are also stored if available at the end of the test. Megger
Guard 2% error guarding 500 k leakage with 100 M load Display range Digital display Analogue display 10 k to 15 T (3 digits) 100 k to 1 T
Short circuit/charge current 5 mA @ 5 kV Capacitor charge time <1.5 seconds per F at 5 mA to 5 kV Capacitor discharge time <120 ms per F to discharge from 5 kV to 50 V Capacitance measurement (500 V minimum test voltage) 10 nF to 50 F (dependent on test voltage) 20
ACCESSORIES
Download Manager may be used to transfer this data to a PC. Data output Real time serial data output once per second of the test voltage, test current, and resistance Lead set Three flexible silicon insulated leads with compact clamp Safety Meets the requirements of IEC61010-1 CAT IV 600 V EMC Meets the requirements of IEC61326-1 Operating uncertainties -20 C Operating temperature Refer to www.megger.com Storage temperature -25 C to 65 C Ingress protection (lid closed) IP65 Humidity 90% RH non-condensing at 40 C Dimensions 305 x 194 x 360 (mm) (12 x 7.6 x 14.2 inches) Weight 7.1 kg (16lb) approx. Included Accessories 3 m lead set, medium size insulated clips User guide on CD-ROM Order Code 6220-820 2000-213
RS232 cable 25955-025 USB cable 25970-041 Optional Accessories HV test lead sets 3 x 3 m with un-insulated small clips 3 x 8 m with un-insulated small clips 3 x 15 m with un-insulated small clips 3 x 10 m with medium insulated clips 3 x 15 m with medium insulated clips 3 x 3 m with large insulated clips 3 x 10 m with large insulated clips 3 x 15 m with large insulated clips Control circuit test lead set 2 x 3 m with small insulated clips 6220-822 8101-181 8101-182 8101-183 1000-441 1000-442 6220-811 1000-443 1000-432
Screened HV test leads 1 x 3 m, wtih 5 kV screened un-insulated small clips 1 x 15 m, with 5 kV screened un-insulated small clips 6220-835 6311-080
21
Megger operate fully traceable calibration and repair facilities, ensuring your instrument continues to provide the high standard of performance and workmanship you expect. These facilities are complemented by a worldwide network of approved repair and calibration companies to offer excellent in-service care for your Megger products. Returning your product to Megger - UK and USA service centres 1. When an instrument requires recalibration, or in the event of a repair being necessary, a Returns Authorisation (RA) number must first be obtained from one of the addresses shown above. You will be asked to provide the following information to enable the Service Department to prepare in advance for receipt of your instrument, and to provide the best possible service to you. n Model, e.g. S1-554/2. n Serial number, to be found on the underside of the case or on the calibration certificate. n Reason for return, e.g. calibration required, or repair. n Details of the fault if the instrument is to be repaired. 22
2. Make a note of the RA number. A returns label can be emailed or faxed to you if you wish. 3. Pack the instrument carefully to prevent damage in transit. 4. Ensure the returns label is attached, or that the RA number is clearly marked on the outside of the package and on any correspondence, before sending the instrument, freight paid, to Megger. Copies of the original purchase invoice and packing note should be sent simultaneously by airmail to expedite clearance through customs. In the case of instruments requiring repair outside the warranty period, an immediate quotation can be provided when obtaining the RA number. 5. You may track the progress of your return on line at www.megger.com Approved Service Centres A list of Approved Service Centres may be obtained from the UK address above, or from Meggers website at www.megger.com Service note Battery replacement Should the internal lead acid battery require replacement the unit must be sent to a Megger authorised service agent. A replacement battery connection wiring loom must be fitted to ensure future connection integrity. In addition the on screen battery charge level indicator will need to calibrated to the new batterys characteristics. End of life disposal WEEE The crossed out wheeled bin placed on the Megger products is a reminder not to dispose of the product at the end of its product life with general waste. Megger is registered in the UK as a Producer of Electrical and Electronic Equipment. The Registration No is WEE/HE0146QT Batteries The crossed out wheeled bin placed on the batteries is a reminder not to dispose of them with general waste at the end of their life. This product contains a sealed lead acid battery 12V, 4Ah (NP4-12 type) Battery replacement should only be performed by a Megger authorised repair agent, who will correctly dispose of the spent battery. For the purpose of end of life disposal only, the battery can be accessed by removing the four recessed screws on rear of instrument and lifting inner instrument out of outer case. Remove screws and remove the inner front cover to access battery. The battery fitted in this instrument is classified as a Portable Battery and should be disposed of in the UK in accordance with Local Authority requirements. For disposal of batteries in other parts of the EU contact your local distributor. Megger is registered in the UK as a producer of batteries. The registration number is BPRN00142
23
M
Megger Limited Archcliffe Road, Dover Kent CT17 9EN England T +44 (0)1 304 502101 F +44 (0)1 304 207342 E uksales@megger.com Megger 4271 Bronze Way, Dallas, Texas 75237-1019 USA T +1 800 723 2861 (USAONLY) T +1 214 333 3201 F +1 214 331 7399 E ussales@megger.com Megger Z.A. Du Buisson de la Couldre 23 rue Eugne Henaff 78190 TRAPPES France T +33 (0)1 30.16.08.90 F +33 (0)1 34.61.23.77 E infos@megger.com Megger Pty Limited Unit 1, 11-21 Underwood Road Homebush NSW 2140 Australia T +61 (0)2 9397 5900 F +61 (0)2 9397 5911 E ausales@megger.com Megger Limited Unit 106-550 Alden Road Markham ON L3R 6A8 Canada T +1 416 298 9688 (Canada only) T +1 416 298 6770 F +1 416 298 0848 E casales@megger.com
Megger 501 Crystal Paradise Mall Off Veera Desai Road Andheri(w), Mumbai - 400053 Maharashtra India T +91 22 26740468 F +91 22 26740465
Megger AB Eldarvgen 4 Box 2970 SE-187 29 TBY Sweden T +46 8 510 195 00 F +46 8 510 195 95
Megger AG Ob. Haselweg 630 5727 Oberkulm Aargau Switzerland T +41 62 768 20 30 F +41 62 768 20 33
This instrument is manufactured in the United Kingdom. The company reserves the right to change the specification or design without prior notice. Megger is a registered trademark Part No. S1-554_2_UG_en_V06 1112 www.megger.com