699 Pressure Sensor PDF
699 Pressure Sensor PDF
50 mbar
Huba Control type 699 - Technical data subject to change - Edition 02/2015
1/5
Technical overview
Pressure range
Relative and differential
Operating conditions
Medium
Medium / ambient
Storage
No condensation
< 3 mbar
> 3 mbar
< 3 mbar
> 3 mbar
ambient temperature
70 C
Temperature
Application at over pressure range
Tolerable overload on one side
Application at under pressure range
Rupture pressure
P1 = 50 mbar
P1 = 100 mbar
P1 = -4 mbar
P1 = -4 mbar
2 x overload
1.5 x overload
50 mbar
P2 =
4 mbar
P2 =
4 mbar
P2 = -50 mbar
P2 = -100 mbar
Load
Current consumption 2)
voltage - 8 V
[Ohm]
< 20 mA
< supply0.02
A
> 10 kOhm
< 10 mA
< 500 Ohm
< 30 mA
< 500 Ohm
< 30 mA
> 10 kOhm
< 10 mA
Response time switchable by
off / 0.2s / 1s / 5s / 20s
Short circuit proof and protected against polarity reversal. Each connection is protected against crossover up to max. supply voltage.
Dynamic response
Response time
Load cycle
< 20 ms
< 10 Hz
Protection standard
Without cover
IP 00
IP 54
IP 65
With cover
Display
Double spaced
per 8 digit alphanumeric
At additional backlight LCD-Display 30 mA current consumption
RTU RS-485
LCD Display
Module MODBUS
Ranges of adjustment
The zero piont is adjustable by reset button.
The Full scale is adjustable by DIP-Switch and additional by the turbopoti.
Adjustability
Optional version with self configurable parameters (see order code selection table)
Electrical connection
Screw terminals for wire and stranded conductors up to 1.5 mm2
Cable gland with built-in strain relief PG11
Pressure connection
Connection pipe
6.2 mm
Mounting instructions
Installation arrangement
Mounting
Recommendation:
Tests / Admissions
UL
Electromagnetic compatibility
Weight
Without display
With display
~ 90 g
~ 100 g
Packaging
Single packaging in cardboard
Multiple packaging
20 / 40 / 120
Accuracy
Parameter
Tolerance zero point
Tolerance zero full scale
Resolution
Total of linearity,
hysteresis and repeatability
Long therm stability acc. to DIN EN 60770
TC zero point 4)
TC zero point 4)
TC sensitivity 4)
TC sensitivity 4)
Unit
0 ... 5 mbar
% fs
% fs
% fs
1.0
1.0
0.2
1.0
1.0
0.2
0.7
0.7
0.1
0.7
0.7
0.1
0.7
0.7
0.1
max.
% fs
% fs
% fs/10K
% fs/10K
% fs/10K
% fs/10K
1.0
1.0
0.2
1.0
0.3
0.6
1.0
1.0
0.2
1.0
0.3
0.6
1.0
1.0
0.2
0.5
0.2
0.5
1.0
1.0
0.1
0.4
0.1
0.5
0.6
1.0
0.1
0.4
0.1
0.2
typ.
max.
typ.
max.
0.5 mbar
2/5
Adjustable by DIP-Switch
TC = Temperature coefficient
2)
At nominal pressure
0 ... 1 mbar
0 ... 10 - 50 mbar
0 ... 3 mbar
max.
max.
Test conditions:
Huba Control type 699 - Technical data subject to change - Edition 02/2015
Pressure unit
Option
Pressure connection /
Pressure orifices
Accessories /
Connection Kit
10
699. X X X X X X X X X X
max. range
middle range
min. range
Pressure range of
0 % bis 100% fs
Pressure range of -10 % bis 90% fs
Pressure range of -100 % bis 100% fs
mbar (hPa)
Pa
mmWS
inH2O
0 ...0.3/0.5
30/50
3/5
0.1/0.2
0 ... 0.3/0.5/1
30/50/100
3/5/10
0.1/0.2/0.3
0 ... 0.5/1/3
50/100/300
5/10/30
0.3/0.5/1
0 ... 1/3/5
100/300/500
10/30/50
0.5/1/2
0 ... 3/5/10
300/500/1000
30/50/100
1/2/3
0 ... 5/10/16
500/1000/1600
50/100/160
2/3/5
0 ... 10/16/25
1000/1600/2500
100/160/250
3/5/10
0 ... 16/25/50
1600/2500/5000
160/250/500
5/10/20
mbar
hPa
Pa
kPa
mmWS
inH O
Linear
without Filter
dual DIP-Switch
Linear
with Filter (transposable)
tenfold DIP-Switch
Square root extracted
without Filter
dual DIP-Switch
Square root extracted
with Filter (transposable)
tenfold DIP-Switch
0 ... 10 V
13.5 ... 33 VDC / 24 VAC 15 %
(3 wire)
0 ... 20 mA
13.5 ... 33 VDC / 24 VAC 15 %
(3 wire)
4 ... 20 mA
13.5 ... 33 VDC / 24 VAC 15 %
(3 wire)
4 ... 20 mA
8.0 ... 33 VDC
(2 wire)
Output signal complimentary selectabel, at delivery no pre-adjustment
without display
with display in pressure unit chosen above
with display in % fs
with module MODBUS
Connection pipe 6.2 mm without pressure orifice
Connection pipe 6.2 mm pressure orifice on P1
Connection pipe 6.2 mm pressure orifice on P2
Connection pipe 6.2 mm pressure orifice on P1 and P2
IP 54 without
IP 54 with connection kit (metal), 90 angled including tube 2 m long (Fig. 1)
IP 54 with connection kit (plastic), straight including tube 2 m long (Fig. 2)
IP 65 without
IP 65 with connection kit (metal), 90 angled including tube 2 m long (Fig. 1)
IP 65 with connection kit (plastic), straight including tube 2 m long (Fig. 2)
Indicate W and state range on order (e.g.: W0 ... + 8mbar/OUT1...6V)
9
B
C
1
2
3
0
1
2
3
4
5
6
7
1)
2)
1)
1)
1)
1)
1)
1)
0
4
2
5
3
6
Output signal /
adjustment
1
2
4
3
0
0
1
3
4
5
6
2,3 1
0
1
2
3
1
2
3
4
0
1
2
3
4
5
b)
a)
Out
100 %
Out
100 %
max. low
pressure
-1 mbar
Zero @ 10%
Pressure
100 %
-10 %
Pressure
Out
100 %
90 %
Zero @ 50%
0%
-100%
Pressure
100 %
Variable parameters
Pressure ranges in grades
Tenfold DIP-Switch
Pressure ranges in grades; stepless adjustable with Turbo-Poti / output signals / filter (off / on) / response curve (linear / root extracted)
Pressure ranges in grades; stepless adjustable with Turbo-Poti / pressure units / pressure range character / output signals; additional
0 ... 5 V / filter (off / 0.2s / 1s / 5s / 20s) / response curve (linear / root extracted) / backlight (off / 5min / on)
Accessories 3)
Connection kit for vent duct (metal), 90 angled
Connection kit for vent duct (plastic), straight
DIN-rail mounting adaptor (Fig. 3)
Module MODBUS
Calibration certificate
1)
2)
Huba Control type 699 - Technical data subject to change - Edition 02/2015
Order number
104312
100064
112854
117305
104551
3)
3/5
2 wire
3 wire
Universal
2 and 3 wire
Accessories
Fig. 1
Fig. 2
Fig. 3
92
6 11.6
35.2
28
51
97
EN 60 715-TH 35-7.5
4/5
Huba Control type 699 - Technical data subject to change - Edition 02/2015
Huba Control AG
Headquarters
Industriestrasse 17
5436 Wrenlos
Telefon +41 (0) 56 436 82 00
Telefax +41 (0) 56 436 82 82
info.ch@hubacontrol.com
Huba Control AG
Niederlassung Deutschland
Schlattgrabenstrasse 24
72141 Walddorfhslach
Telefon +49 (0) 7127 23 93 00
Telefax +49 (0) 7127 23 93 20
info.de@hubacontrol.com
Huba Control SA
Succursale France
Rue Lavoisier
Technople Forbach-Sud
57602 Forbach Cedex
Tlphone +33 (0) 387 847 300
Tlcopieur +33 (0) 387 847 301
info.fr@hubacontrol.com
Huba Control AG
Vestiging Nederland
Hamseweg 20A
3828 AD Hoogland
Telefoon +31 (0) 33 433 03 66
Telefax +31 (0) 33 433 03 77
info.nl@hubacontrol.com
Huba Control AG
Branch Office United Kingdom
Unit 13 Berkshire House
County Park Business Centre
Shrivenham Road
Swindon Wiltshire SN1 2NR
Phone +44 (0) 1993 776667
Fax
+44 (0) 1993 776671
info.uk@hubacontrol.com
www.hubacontrol.com
5/5
Huba Control type 699 - Technical data subject to change - Edition 02/2015