Infcs and Infcs-Xxxb: Infinity C Strain Meter/Controller
Infcs and Infcs-Xxxb: Infinity C Strain Meter/Controller
Infcs and Infcs-Xxxb: Infinity C Strain Meter/Controller
Counters
Frequency Meters
PID Controllers
Clock/Timers
Printers
Process Meters
On/Off
Controllers
Recorders
Relative
Humidity
Transmitters
Thermocouples
Thermistors
Wire
It is the policy of NEWPORT to comply with all worldwide safety and EMC/EMI regulations that apply. NEWPORT is constantly pursuing
certification of its products to the European New Approach Directives. NEWPORT will add the CE mark to every appropriate device upon
certification.
The information contained in this document is believed to be correct but NEWPORT Electronics, Inc. accepts no liability for any errors it
contains, and reserves the right to alter specifications without notice.
WARNING: These products are not designed for use in, and should not be used for, patient connected applications.
TRADEMARK NOTICE: a ,
NEWPORT , NEWPORT , newportUS.com,
Case Bezel Design are trademarks of NEWPORT Electronics, Inc.
PATENT NOTICE: This product is covered by one or more of the following patents: U.S. Pat. No. Des. 336,895; 5,274,577; 6,243,021 /
Canada 2052599; 2052600 / Italy 1249456; 1250938 / France Brevet No. 91 12756 / Spain 2039150; 2048066 / UK Patent No. GB2 249
837; GB2 248 954 / Germany DE 41 34398 C2. Other International Patents Pending.
This device is marked with the international caution symbol. It is important to read the Setup Guide before installing or commissioning
this device as it contains important information relating to safety and EMC.
PREFACE
Manual Objectives
This manual shows you how to set up and use the Programmable Digital Meter.
Standard Procedures:
*
*
*
*
*
*
*
*
*
*
*
*
*
*
Optional Procedures:
*
*
*
*
*
ii
Refer to section
Introduction
Getting Started
Display messages
Display Messages
Menu Configuration
Setpoint Configuration
Displays
Specifications
Specifications
Factory Default
Setup as Shipped
Table of Contents
Table of Contents
Section
Page
SEC 1
1.1
1.2
INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
Safety Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
SEC 2
2.1
2.2
2.3
2.4
2.5
2.6
SEC 3
3.1
3.2
3.3
3.4
3.5
3.6
3.7
SEC 4
4.1
4.2
4.3
4.3.1
4.3.2
4.4
4.4.1
4.4.2
4.4.3
4.4.4
Table of Contents
Table of Contents
Section
4.5
4.6
4.6.1
4.6.2
4.6.3
4.7
4.7.1
4.7.2
4.7.3
4.8
4.9
4.10
4.10.1
4.10.2
4.10.3
4.11
4.11.1
4.12
4.12.1
4.12.2
4.12.3
4.13
4.13.1
iv
Page
Selecting a Display Color . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30
Using Setpoint 1 Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30
Setting Setpoint 1's Active Band . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .31
Selecting if Setpoint 1 is Latched or Unlatched . . . . . . . . . . . . . . . . . . . .31
Assigning Setpoint 1 Values to Net or Gross Readings . . . . . . . . . . . . . .32
Using Setpoint 2 Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .32
Setting the Setpoint 2s Active Band . . . . . . . . . . . . . . . . . . . . . . . . . . . .32
Selecting if Setpoint 2 is Latched or Unlatched . . . . . . . . . . . . . . . . . . . .33
Assigning Setpoint 2 Values to Net or Gross Readings . . . . . . . . . . . . . .33
Setting the Setpoint 1 Deadband . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .34
Setting the Setpoint 2 Deadband . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35
Using Output Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37
Enabling or Disabling the Analog Output . . . . . . . . . . . . . . . . . . . . . . . . .37
Selecting Analog Output as Current or Voltage . . . . . . . . . . . . . . . . . . . .38
Selecting Analog Output or Proportional Control . . . . . . . . . . . . . . . . . . .38
Using Output Scale and Offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39
Examples of Output Scale and Offset . . . . . . . . . . . . . . . . . . . . . . . . . . .41
Using Lock Out Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .42
Enabling or Disabling the RESET Button in the Run Mode . . . . . . . . . . .42
Enabling or Disabling Setpoint Changes . . . . . . . . . . . . . . . . . . . . . . . . .42
Setpoint Display Function: Firmware Version or Setpoint Value . . . . . . .43
Using Display Brightness Configuration . . . . . . . . . . . . . . . . . . . . . . . . .43
Changing Brightness Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43
SEC 5
SEC 6
SEC 7
SEC 8
SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50
SEC 9
SEC 10
Table of Contents
List of Figures
Figure
2-1
2-2
2-3
3-1
3-2
3-3
3-4
3-5
3-6
3-7
3-8
3-9
3-10
3-11
3-12
3-13
3-14
3-15
3-16
4-1
8-1
Page
Front-Panel with Big LED Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Front Panel with Standard LED Display . . . . . . . . . . . . . . . . . . . . . . . . . .5
Connectors (AC-Powered and DC-Powered) . . . . . . . . . . . . . . . . . . . . . .8
Main Board Power Jumpers (W1, W2, W3) . . . . . . . . . . . . . . . . . . . . . .11
Main Board Jumper Positions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
Upper Isolated Analog Output Option Board Installation . . . . . . . . . . . . .12
Meter - Exploded View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
Panel Cut-Out . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
Meter-powered Bridge Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
Externally-powered Bridge Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
4-Wire DC Input Connections with External Excitation . . . . . . . . . . . . . .16
Current Input Connections with Internal Excitation . . . . . . . . . . . . . . . . .16
Current Input Connections with External Excitation . . . . . . . . . . . . . . . . .17
Main Power Connections - AC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
Main Power Connections - DC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18
External Tare Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
Analog Output Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
Relay Output Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
Isolated Analog Output Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Alarm Example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36
Meter Dimensions/Panel Cutout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .54
Table of Contents
List of Tables
Table
A-1
2-1
2-2
2-3
3-1
3-2
4-1
4-2
4-3
4-4
5-1
6-1
6-2
7-1
8-1
9-1
vi
Page
Sections of the Manuals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ii
Accessories and Add-Ons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
Connector Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
DIP Switch Positions/Input Range & Excitation . . . . . . . . . . . . . . . . . . . .10
S3 Jumper Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
AC-Power Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18
Range Selection Dip Switch Positions for Regular Voltage Input . . . . . .23
Range Selection Dip Switch Positions for Millivolt/Milliamp Input . . . . . .23
Natural Gain . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26
Input Resolution Multiplier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26
Display Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .44
Menu Configuration Displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .45
Run Mode Displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .48
Setpoint Configuration Displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .49
Color Chart for DC Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .53
Factory Preset Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .55
Note
NOTE
WARNING
CAUTION
IMPORTANT
vii
Notes
viii
Introduction
SECTION 1. INTRODUCTION
1.1 UNPACKING
Remove the Packing List and verify that all equipment has been received. If there are
any questions about the shipment, use the phone numbers listed on the back cover to
contact the Customer Service Department nearest you.
Upon receipt of shipment, inspect the container and equipment for any signs of damage.
Take particular note of any evidence of rough handling in transit. Immediately report any
damage to the shipping agent.
Note
The carrier will not honor any claims unless all shipping material is saved for
their examination. After examining and removing contents, save packing
material and carton in the event reshipment is necessary.
Verify that you receive the following items in the shipping box:
QTY
DESCRIPTION
Owner's Manual
Note
If you ordered any of the available options (except the "BL" Blank Lens
option), they will be shipped in a separate container to avoid any damage to
your indicator/controller.
Introduction
1.2
SAFETY CONSIDERATIONS
This device is marked with the international caution symbol. It is important to
read this manual before installing or commissioning this device as it contains
important information relating to Safety and EMC (Electromagnetic Compatibility).
This instrument is a panel mount device protected in accordance with EN 610101:2001, electrical safety requirements for electrical equipment for measurement, control
and laboratory. Installation of this instrument should be done by qualified personnel. In
order to ensure safe operation, the following instructions should be followed.
This instrument has no power-on switch. An external switch or circuit-breaker shall
be included in the building installation as a disconnecting device. It shall be marked to
indicate this function, and it shall be in close proximity to the equipment within easy
reach of the operator. The switch or circuit-breaker shall not interrupt the Protective
Conductor (Earth wire), and it shall meet the relevant requirements of IEC 9471 and
IEC 947-3 (International Electrotechnical Commission). The switch shall not be
incorporated in the main supply cord.
Furthermore, to provide protection against excessive energy being drawn from the
main supply in case of a fault in the equipment, an overcurrent protection device shall
be installed.
Note
Do not exceed voltage rating on the label located on the top of the instrument
housing.
Always disconnect power before changing signal and power connections.
Do not use this instrument on a work bench without its case for safety reasons.
Do not operate this instrument in flammable or explosive atmospheres.
Do not expose this instrument to rain or moisture.
Unit mounting should allow for adequate ventilation to ensure instrument does not
exceed operating temperature rating.
Use electrical wires with adequate size to handle mechanical strain and power
requirements. Install without exposing bare wire outside the connector to minimize
electrical shock hazards.
EMC Considerations
Install Ferrite Bead(s) on signal wires close to the instrument if EMC problems
persist.
Failure to follow all instructions and warnings may result in injury!!
2
2.2 FEATURES
The following is a list of standard features:
*
*
*
*
*
*
*
*
*
Special Calibration/Configuration
NEMA-4 Splash Proof Cover
NEMA-4 Splash Proof Cover, NEW
Accessories
TP1A
RP18
RP28
RP38
Figure 2-2 shows each part of the front of the standard LED display meter.
SETPTS
/NT/GRS
/TARE
MENU
RESET
Setpoint 1 status
Setpoint 2 status
Note
If the dual relay option is not installed or if L.3=1 is displayed on the LK.CF
menu, pressing the SETPTS button will display the meter's firmware version.
/NT/GRS Button
In the Run Mode, this button will toggle between net/gross readings or peak readings
depending upon setup.
In the Configuration Mode, press this button to change the value of the flashing digit
shown on the display and/or toggle between menu choices, such as R.1=R or R.1=N on
RD.CF menu. When configuring your setpoint values, press the /NT/GRS button to
advance the flashing digit's value from 0 to 9 by 1.
/TARE Button
In the Run Mode press the /TARE button to tare your reading (zeroing).
In the Configuration Mode, press the this button to scroll to the next digit.
MENU Button
In the Run Mode, press the MENU button to terminate the current measuring strain and
enter you into the Configuration Mode.
Note
Only if you have not installed the lock out jumpers on the main board.
In the Configuration Mode, press the MENU button to store changes in the nonvolatile
memory and then advance you to the next menu item.
RESET Button
If you hard reset (press the MENU button followed by the RESET button) or power off/on
the meter, it shows RST , followed by STRN.
In the Run Mode, press the RESET button to reset the latched setpoints. The meter
shows SP.RS and returns to the Run Mode.
In the Configuration Mode, press the RESET button once to review the previous menu.
Press the RESET button twice to perform a hard reset and return to the Run Mode.
In the Peak Mode, press the RESET button to reset peak values. The meter shows
PK.RS and returns to the Run Mode.
In the Setpoint Mode, press the RESET button to reset the latched setpoint. The meter
shows SP.RS and enters the Run Mode.
In the Tare Mode, press the RESET button to reset. The meter shows T.RST.
Note
Description
Setpoint 1: Normally open (N.O.1) connection
Setpoint 1: Normally closed (N.C.1) connection
Setpoint 1: Common (COM1) connection
Setpoint 2: Normally open (N.O.2) connection
Setpoint 2: Normally closed (N.C.2) connection
Setpoint 2: Common (COM2) connection
AC line connection (no connections on DC-powered units)
AC neutral connection (+ Input on DC-powered units)
AC earth ground (DC-power return on DC-powered units)
Analog voltage output
Analog current output
Analog return
-E: Negative excitation connection from meter (5, 10, 12 V)
+E: Positive excitation connection from meter (5, 10, 12 V)
+20 mA connection for analog input
+R (Not used)
+24 V output connection
+S: Positive signal input
-S: Negative signal input and return for +20 mA or +24 V
-R (Not used)
Isolated Analog Voltage Output
Isolated Analog Current Output
Isolated Analog Output Return
Remote tare connection with a momentary switch
O
C
C
O
O
O
O
C
C
O
C
Settings for Input Ranges
O
C
O
O
C
O
C
O
O
C
O
O
O
C
C
The display must also be configured to the selected input type after setting
the DIP switches (see Section 4.1, Selecting the Input Type)
2.6 DISASSEMBLY
You may need to open up the meter for one of the following reasons:
Note
10
Getting Started
1.
Remove the main board from the case. Refer to Section 2.6.
2.
Locate the solder jumpers W1, W2, and W3 (located near the edge of the main
board alongside the transformer).
3.
T1A
W2
W1
DISPLAY BOARD
W3
Getting Started
3.2 MAIN BOARD POWER JUMPERS (Continued)
Figure 3-2 shows the location jumper positions on the main board.
DIS
PL
AY
BO
AR
D
T1
S3
MAIN BOARD
TP1
TP2
TP3
TP4
TP5
TP6
TP7
TP8
TP9
TP10
S4
S2
TB1
TB2
S1
J1
A B
E D CB A
S3
Refer to
table 3-1
S2
C D
TP11
Attach cable
to P1.
TB5
P1
Getting Started
Test pins TP1 - TP11 are for testing purposes. Do not use as reading errors may result.
S4-A Factory default jumper is removed.
Description
Install to enable front panel push-buttons.
Remove to disable all front panel push-buttons.
Removed. For factory calibration only.
Removed. Not used.
Installed for external ratiometric.
If installed without S3-B, the MENU button locks out.
If you press the MENU button, the meter shows LOCK .
13
Getting Started
3.3 PANEL MOUNTING
RE
A
(R R C
EM OV
OV ER
ED
CONNECTOR
)
PANEL
CUT-OUT
"NEW" STYLE
MOUNTING
BRACKET
LABEL
PRODUCT
LABEL
CASE
GASKET
"OLDER" STYLE
MOUNTING
BRACKET 2 PCS
FRONT BEZEL
1,5
R(.06)
4 PLCS
45,00 + 0,61/-0,00
(1.772 + .024/.000)
92,00 + 0,81/0,00
(3.622 + .032/.000)
NOTE: Dimensions in Millimeters (Inches)
Getting Started
+E +20mA +R
2
3
4
+24
5
+S
6
-S
7
-R
8
TB2
+
-
-E
+E +20mA +R
+24 +S
5
-S
-R
TB2
+EX
-EX
15
Getting Started
3.4 CONNECTING SENSOR INPUTS (Continued)
16
Getting Started
Getting Started
3.5 CONNECTING MAIN POWER (Continued)
Table 3-2 shows the wire color and respective terminal connections for both USA and
Europe.
AC POWER
WIRE COLORS
EUROPE
USA
~ AC Line
~ AC Neutral
~ AC Earth
Brown
Blue
Green/Yellow
Black
White
Green
TB1
TB2
1 2 3 4 5 6 1 2 3 4
7 8 9 10 11 12 5 6 7 8
+ DC
- DC
DC POWER
When using DC power, refer to the Table 8-1 Color Chart in the Specifications
Section for Display Color, Intensity, Excitation Voltage and Current, and Analog
Output Isolated Option. Failure to use proper ratings may result in damaging
the unit.
18
Getting Started
TB1
TB2
1 2 3 4 5 6 1 2 3 4
S1
OPEN
7 8 9 10 11 12 5 6 7 8
J1
1 2
CLOSED
3 4
TB2
1 2 3 4 5 6 1 2 3 4
7 8 9 10 11 12 5 6 7 8
+
ANALOG
VOLTAGE
0-10V dc
+
-
ANALOG
CURRENT
0-20mA
or
4-20mA
ANALOG
RTN
19
Getting Started
3.7 CONNECTING ANALOG AND RELAY OUTPUTS (Continued)
Relay 2
COM2
N.C.2
N.O.2
COM1
N.C.1
N.O.1
}
}
Relay 1
NO2
EXTERNAL LOAD
COM2
TB1
N
L
FUSE
TB2
1 2 3 4 5 6 1 2 3 4
7 8 9 10 11 12 5 6 7 8
010VDC
020mA
OR
420mA
1 2 3
TB1
TB2
TB5
1 2 3 4 5 6 1 2 3 4
7 8 9 10 11 12 5 6 7 8
Before proceeding, set the input DIP switch settings at the back of your
meter. (Refer to Table 2-3).
1.
2.
Press the TARE button. The meter flashes one of the following:
5V (for 5 Vdc)
3.
4.
Press the MENU button to store your choice. The meter momentarily shows
STRD , followed by DEC.P (Decimal Point).
21
2.
Press the TARE button. The meter shows one of the following:
FFF.F
FF.FF
F.FFF
FFFF (Default)
3.
4.
Press the MENU button to store your choice. The meter momentarily shows STRD ,
followed by the next menu RD.S.O (Reading Scale and Offset). Or you can press
the RESET button to abort and go back to the DEC.P menu.
To scale the meter to show readings in engineering units. There are two methods. One
method is to scale with known inputs. Another method is to scale without known inputs:
you calculate input values based on the transducer specifications and manually enter
them through the keyboard.
22
For maximum resolution, find the maximum signal that will be applied to the
meter input.
For regular voltage input, refer to the main body of Table 4-1.
For millivolt or milliamp input, refer to the main body of Table 4-2.
Set the DIP switch positions as indicated at the top of either Table 4-1 or 4-2. The
numbers 1 through 8 in the top row of either table represent dip switches 1 through 8,
and the O, C or X directly below the number indicates the correct position of each switch.
Once Dip switches have been positioned correctly, apply power. Proceed to the RD.CF
(Reading Configuration) and set R2 equal to the value in the right hand column of the
chart.
Table 4-1. Range Selection Dip Switch Positions For Regular Voltage Input
12345678
XCOOOXXC
0 - 10 V
0-5V
0-3V
0-2V
0-1V
12345678
XCOOCXXC
5 V
5 V
3 V
2 V
1 V
RD.CF*
R.2=
4
3
2
1
0
Table 4-2. Range Selection Dip Switch Positions For Millivolt/ Milliamp Input
12345678
12345678
12345678
RD.CF*
Table 4-1. Range
Selection DipXOCOCXX0
Switch Positions
For Regular Voltage
Input
XOCOOXX0
XOCCOXX0
R.2=
0 - 100 mV
50 mV
0 - 20 mA
4
0 - 50 mV
50 mV
0 - 10 mA
3
0 - 30 mV
30 mV
0 - 6 mA
2
Table
4-2.
Range
Selection
Dip
Switch
Positions
0 - 20 mV
20 mV
0 - 4 mA
1
For Millivolt
and Milliamp
Input
0
10
mV
10
mV
0
2
mA
0
p
* Reading Configuration
23
2.
3.
Press the TARE button. The meter shows IN1 (Input 1).
Note
4.
Press the TARE button again. The meter shows last stored value for Input 1.
5.
Press the TARE button once more. The meter shows the actual signal being
received.
6.
Press the MENU button to store this value as IN1 (Input 1). The meter shows
RD1 (Read 1).
Note
7.
Press the TARE button. The meter shows the last stored value for Read 1.
8.
9.
10.
Press the MENU button to store value as RD1 . The meter shows IN!2 (Input 2).
Note
24
12.
Press the TARE button again. The meter shows the last stored value for Input 2.
13.
Press the TARE button once more. The meter shows the actual signal being
received.
14.
Press the MENU button to store Input 2 value. The meter shows RD!2 (Read 2).
Note
15.
Press the TARE button. The meter shows the last stored value for Read 2.
16.
17.
18.
Press the MENU button to store value as RD!2 (Read 2). The meter momentarily
shows STRD , followed by RD.CF . Meter scaling is now complete.
25
100.0 lbs
3.1 mV/V
10 Vdc
31mV = (3.1 mV/V) x (10 V)
Determine the correct values for IN1 and IN!2 based on the load cell specifications.
In most cases, RD1 and RD!2 are equal to the minimum and maximum of the
transducer output span. The example assumes RD1 and RD!2 are equal to the
range of the load (RD1 = 0 and RD!2 = 100.0). Calculate IN1 and IN!2 using the
loadcell output span and the following equation:
IN = (Sensor Output) x (Natural Gain) x (Multiplier).
Span Units
Millivolts
Millivolts
Volts
Volts
Milliamps
Natural Gain
100 cts/mV
40 cts/mV
1000 cts/V
400 cts/V
500 cts/mA
Determine the multiplier by the Input Resolution setting (R.2 in the RD.CF menu)
and the input range selected. Typically R.2=4 is suitable for most applications.
R.2=4
1.000
1.000
1.000
1.000
1.000
R.2=3
2.000
2.000
2.000
1.000
1.000
R.2=2
3.333
3.333
3.333
1.667
1.667
R.2=1
5.000
5.000
5.000
2.500
2.500
R.2=0
10.00
10.00
10.00
5.000
5.000
Determine IN1 and IN!2 input range and resolution. The example selects the
0 to 100 mV range and 10 uV resolution (R.2=4 ).
Example: IN1 = (0 mV) x (100 cts/mV) x (1.000) = 0
IN!2 = (31 mV) x (100 cts/mV) x (1.000) = 3100
RD1 = 0000
RD!2 = 100.0
4.
5.
6.
Press the TARE button again, the meter shows the last Input 1 value, with the
fourth digit flashing.
7.
8.
9.
Press the MENU button to store this value. The meter shows RD1 .
10.
Press the TARE button. The meter shows the last value for Read 1.
Repeat steps 7, 8 and 9 until RD1 , IN!2 and RD!2 have been displayed, verified,
changed (if necessary) and stored.
You may use Reading Configuration RD.CF to configure your meter for the following:
27
2.
Press the TARE button. The meter shows one of the following:
3.
Press the NT/GRS button to view last stored selection. Press the NT/GRS
button to toggle between selections.
4.
Press the TARE button to select input resolution or press the MENU button to
store your selection and shows COLR menu.
Press the MENU button until RD.CF displays, then press the TARE button twice.
or
Press the TARE button from R.1 .
One of the following displays (default is R.2=4 ):
R.2=4 =
R.2=0 =
R.2=1 =
R.2=2 =
R.2=3 =
Example: 3 V resolution means that if you input 0-30 mV, at 30 mV the display shows
9999 .
28
2.
3.
Press the TARE button to display the filtered/unfiltered signal input or press the
MENU button to store your selection and shows COLR menu.
Press the MENU button until RD.CF displays, then press the TARE button three
times.
or
Press the TARE button from R.2 .
One of the following displays:
2.
3.
Press the MENU button to store your selections. STRD momentarily displays,
followed by COLR Menu.
Press the MENU button until RD.CF displays, then press the TARE button three
times.
or
Press the TARE button from R.3.
One of the following displays:
2.
3.
Press the MENU button to store your selections. STRD momentarily displays,
followed by COLR Menu.
29
2.
Press the TARE button. The meter shows one of the following:
GRN
RED
AMBR
3.
4.
Press the MENU button to store your choice. The meter momentarily shows
STRD , followed by the next menu S1.CF (Setpoint 1 Configuration). Or you can
press the RESET button to abort and go back to the RD.CF menu.
Setpoint 1 Configuration S1.CF is not active unless your meter has dual relay output
capabilities. The LED's will display whether the S1.CF is active or not. You may use
Setpoint 1 Configuration S1.CF for the following:
30
To set the setpoint's active band above or below your chosen value
2.
Press the TARE button. The meter shows one of the following:
3.
4.
Press the MENU button until S1.CF displays, then press the TARE button twice.
or
Press the TARE button from S.1 .
The meter shows one of the following:
2.
3.
Press the TARE button to assign Setpoint 1 values to net or gross reading or
press the MENU button to enter S2.CF (Setpoint 2 Configuration).
31
Press the MENU button until S1.CF displays, then press the TARE button twice.
or
Press the TARE button from S.2.
The meter shows one of the following:
2.
3.
Press the MENU button to store your selection(s). The meter momentarily shows
STRD, followed by S2.CF (Setpoint 2 Configuration).
Setpoint 2 Configuration S2.CF is not active unless your meter has dual relay output
capabilities. The LED's will display whether the S2.CF is active or not. You may use
Setpoint 2 Configuration S2.CF for the following:
To set the setpoint's active band above or below your chosen value
To select whether the setpoint operation is latched or unlatched
To assign setpoint values to the net or gross reading
32
3.
4.
Press the MENU button until S2.CF displays, then press the TARE button twice.
or
Press the TARE button from S.1.
The meter shows one of the following:
2.
3.
Press the MENU button to store your selection(s). The meter momentarily shows
STRD , followed by S1.DB (Setpoint 1 Deadband).
Press the MENU button until S2.CF displays, then press the TARE button twice.
or
Press the TARE button from S.2.
The meter shows one of the following:
2.
3.
Press the MENU button to store your selection(s). The meter momentarily shows
STRD, followed by S1.DB (Setpoint 1 Deadband).
33
Setpoint 1 Deadband S1.DB is not active unless your meter has dual relay output
capabilities. The LED's will display whether the S1.DB is active or not. The Setpoint 1
Default deadband is 0003. To change the deadband (hysteresis) of Setpoint 1, follow
these steps:
34
1.
2.
Press the TARE button. The meter shows the last previously stored 4-digit
number (0000 through 9999) with flashing 4th digit.
3.
Press the NT/GRS button to change the value of the flashing digit. If you
continue to press the NT/GRS button, the flashing digit's value continues to
change.
4.
5.
Press the MENU button to store your selection. The meter momentarily shows
STRD , followed by S2.DB (Setpoint 2 Deadband).
Setpoint 2 Deadband S2.DB is not active unless your meter has dual relay output
capabilities. The LED's will display whether the S2.DB is active or not. The Setpoint 2
default deadband is 0003. To change the deadband (hysteresis) of Setpoint 2, follow
these steps:
1.
2.
Press the TARE button. The meter shows the last previously stored 4-digit
number (0000 through 9999) with flashing 4th digit.
3.
Press the NT/GRS button to change the value of the flashing digit. If you
continue to press the NT/GRS button, the flashing digit's value continues to
change.
4.
5.
Press the MENU button to store your selection. The meter momentarily shows
STRD , followed by OT.CF (Output Configuration) if you have analog output
capabilities.
35
SETPOINT
SIGNAL
LEVEL
ON
OFF
ON
ON
OFF
ACTIVE BELOW
OFF
ACTIVE ABOVE
3
DEADBAND
SETPOINT
3
SIGNAL
LEVEL
ON
OFF
ON
OFF
ON
OFF
ON
ACTIVE BELOW
WITH DEADBAND 3
ACTIVE ABOVE
WITH DEADBAND 3
SETPOINT
SIGNAL
LEVEL
ON
ON
ON
OFF
ON
36
ON
Output Configuration OT.CF is not active unless your meter has analog output
capabilities. The menu will display whether analog output is present or not. Analog
output must be ordered at the time of purchase.
Use Output Configuration OT.CF to select the following:
2.
Press the TARE button. The meter shows one of the following:
3.
4.
Press the TARE button to select analog output as current or voltage or press the
MENU button to store your selection and enter OT.S.O (Output Scale and Offset).
37
Press the MENU button until it shows OT.CF , then press the TARE button twice.
or
Press the TARE button from 0.1.
The meter shows one of the following:
2.
3.
Press the TARE button to select analog output or proportional control or press the
MENU button to store your selection and enter OT.S.O (Output Scale and Offset).
Press the MENU button until it shows OT.CF , then press the TARE button twice.
or
Press the TARE button from 0.2.
The meter shows one of the following:
38
2.
3.
Press the MENU button to store your selection. The meter momentarily shows
STRD , followed by OT.S.O (Output Scale and Offset).
Output Scale and Offset OT.S.O is not active unless your meter has analog output
capabilities. The menu will display whether analog output is present or not. Output Scale
and Offset OT.S.O scales your analog output to be equal to the meter's display and/or
any engineering units you require. You may scale the output for direct (4-20 mA, 0-10 V,
etc) or reverse acting (20-4 mA, 10-0 V, etc).
1.
2.
3.
Press the TARE button again. The meter shows the last previously stored
4-digit number (-1999 through 9999) with flashing 4th digit.
4.
5.
6.
Press the MENU button to store your selection. OUT.1 (Output 1) displays.
Note
7.
If you select O.2=V for voltage, the maximum signal you may select is
10.00 for an 0-10 Vdc signal output. If you select O.2=C for current, the
maximum signal you may select is 20.00.
8.
Press the NT/GRS button to enter the Output 1 signal selection. If you continue
to press the NT/GRS button, the flashing digit's value continues to change.
9.
10.
Press the MENU button to store your selection. RD!2 (Read 2) displays.
Note
Press the TARE button. The meter shows last previously stored 4-digit number
(-1999 through 9999) displays with flashing 4th digit.
12.
Press the NT/GRS button to change the value of the flashing digit. If you continue
to press the NT/GRS button, the flashing digit's value continues to change.
13.
14.
Press the MENU button to store your selection. The meter shows OUT.2 (Output 2).
Note
15.
Note
If you select O.2=V for voltage, the maximum signal you may select is 10.00
for an 0-10 Vdc signal output. If you select O.2=C for current, the maximum
signal you may select is 20.00 for a 0-20 or 4-20 mA DC signal output.
16.
Press the NT/GRS button to change the value of the flashing digit. If you
continue to press the NT/GRS button, the flashing digit's value continues to
change.
17.
18.
Press the MENU button to store your selection. The meter momentarily shows
STRD , followed by LK.CF (Lockout Configuration).
WARNING: If the meter displays all flashing values on any item, the value
has overflowed. Press the NT/GRS button to start new values.
40
41
Press the MENU button until the meter shows LK.CF (after OT.S.O ).
Press the TARE button. The meter shows one of the following:
RS.=E To enable the RESET button in the Run Mode (Default)
4.
5.
Press the MENU button to store the changes. The meter shows STRD if the new
value is different otherwise the meter shows BRIT and returns to the Run Mode.
42
1.
Press the MENU button until the meter shows LK.CF (after OT.S.O ).
2.
Press the TARE button twice. The meter shows one of the following:
3.
4.
Press the MENU button to store the changes. The meter shows STRD if the new
value is different otherwise the meter shows BRIT and returns to the Run Mode.
Press the MENU button until the meter shows LK.CF (after OT.S.O ).
2.
Press the TARE button three times. The meter shows one of the following:
L.3=0 SETPTS button will display setpoint values.
L.3=1 SETPTS button will display the meter's firmware version.
3.
4.
Press the MENU button to store the changes. The meter shows STRD if the new
value is different otherwise the meter shows BRIT and returns to the Run Mode.
Note
If your meter does not have the relay option, setpoint menu items above will
not be available and SETPTS button will always display the meter's firmware
version. These units will have +OL (overload) or +OPN memory indicated by
Alarm 1 & 2 LED displays. LEDs can be reset by pressing MENU then RESET
button or by Power OFF then ON.
Press the MENU button until the meter shows BRIT (after LKCF ).
2.
Press the TARE button from BRIT. The meter shows one of the following:
3.
4.
Press the MENU button to store your selection. The meter momentarily shows
STRD followed by STRD, RST, STRN, then measured value.
43
Display Messages
Table 5-1. Display Messages
MESSAGE
STRN
RST
INPT
DEC.P
RD.S.O
RD.CF
COLR
S1.CF
S2.CF
S1.DB
S2.DB
OT.CF
OT.S.O
LK.CF
BRIT
+OL
-OL
RS.OF
+999
I999
NT.OF
GT.OF
ER1
PEAK
PK.RS
T.RS
SP.RS
NET
GROS
SP1
SP2
R.OV.S
V.-8.8
RUN
44
DESCRIPTION
Strain Meter
Hard (power on) Reset
Input Type
Decimal Point Position
Reading Scale and Offset
Reading Configuration
Display Color
Setpoint 1 Configuration
Setpoint 2 Configuration
Setpoint 1 Deadband
Setpoint 2 Deadband
Output Configuration
Output Scale and Offset
Lock Out Configuration
Display Brightness
+ Overload Signal
- Overload Signal
Resolution Overflow
Value Overflow in Setpoint/Menu Routines
Value Overflow in Setpoint/Menu Routines
Net Value Overflow
Gross Value Overflow
2 Coordinate Format Programming Error
Peak Value
Peak Reset
Tare Reset
Setpoint Reset
Net Value
Gross Value
Setpoint 1 Value
Setpoint 2 Value
Resolution Over Scale
Firmware Version (where 8is 0 ~ 9)
Operating Mode
MENU
100M (Default)
50M
10V
5V
0-20
FFFF (Default)
F.FFF
FF.FF
FFF.F
INPT
DEC.P
(Reading Scale & Offset)
RD.S.O
NT/GRS
Shows IN1
Shows prior value entered and flashing digit.
Scrolls to the next digit.
If TARE is pressed, actual input
45
MENU
TARE
Reading Configuration
R.1=
RD.CF
R.2=
R.3=
R.4=
COLR
46
GRN
RED
(Green)
AMBR
(Amber)
(Red)
Setpoint 1 Configuration
S.1=
S1.CF
S.2=
S.2=U (Unlatched)
S.2=L (Latched)
S.3=
Setpoint 2 Configuration
S.1=
S2.CF
S.2=
S.2=U (Unlatched)
S.2=L (Latched)
S.3=
TARE
NT/GRS
S1.DB
Setpoint 2 Deadband
S2.DB
Output Configuration
OT.CF
0.1 =
0.2 =
0.3 =
OT.S.O
(Shown if O.3=A in Output
Configuration Menu OT.CF )
RD1
1 Shows
Shows prior value entered
47
MENU
TARE
RS=
LK.CF
SP=
L3=
Brightness Configuration
BRIT
Shows
input
choices
TARE
NET / GROS
T.RST
SP.RS
48
Press to
activate.
NT/GRS
Displays NET
or GROSS
reading. Once
reading shows,
respective value
shows.
RESET
Description
Peak Reading
Toggle between
Net and Gross
values.
SP1
SP2
TARE
NT/GRS
Description
SETPOINT 1
Press to scroll to
the next digit to the
right
Press to change
the value of the
flashing digit
Press to scroll to
the next digit to the
right
Press to change
the value of the
flashing digit
49
Specifications
SECTION 8. SPECIFICATIONS
SIGNAL INPUT
Input Ranges:
Isolation:
Noise Rejection:
Resistance:
Big Display:
( -1.9.9.9 ~ 9.9.9.9)
Symbol:
Standard Display:
( -1.9.9.9. ~ 9.9.9.9.)
Symbol:
ANALOG TO DIGITAL
Technique:
Dual slope
Internal resolution:
15 bits
Read Rate:
ACCURACY AT 25C
Max Error Strain:
Span Tempco:
50 ppm/C
Step Response:
1 sec
50
Excitation Voltage:
AC power units
24 V @ 25 mA,
12 V @ 50 mA,
10 V @ 120 mA,
5 V @ 60 mA
Load Regulation:
1.1%
Line Regulation:
DC power units
Refer to Table 8-1
Color chart for DC Output
Excitation
Specifications
Current or voltage
Signal Level:
Function:
Linearity:
0.2%
Current or voltage
Signal Level:
Function:
Linearity:
0.2%
Isolation
Note
Only one analog output is available on each unit and it must be factory
installed.
51
Specifications
SECTION 8. SPECIFICATIONS (Continued)
INPUT POWER INFORMATION
AC units
DC units
Fuse
115 V
230 V
63 mA @ 250 (T)
Fuse
115 V
230 V
80 mA @ 250 V Slow-Blow
ENVIRONMENT
Operating temperature:
Storage temperature:
Relative humidity:
MECHANICAL
52
Panel cutout:
Weight:
1.27 lb (575 g)
Case material:
Protection:
Specifications
HIGH BRIGHTNESS
RED
Sensor Excitation:
24 V @ 25 mA,
12 V, 10 V, 5 V @ 35 mA Max
Analog Output:
Non-Isolated option only
GREEN
Warning:
AMBER
Note
Analog Output:
Non-Isolated options or
Isolated Analog option
53
Specifications
SECTION 8. SPECIFICATIONS (Continued)
48,0 (1.89)
96,0 (3.78)
FRONT BEZEL
20,3 (.80)
151,4 (5.96)
RETAINER
CASE
REAR COVER
TOP VIEW
SIDE VIEW
PANEL THICKNESS
6,4 (.25) MAX
0,8 (.03) MIN
1,5
R(.06)
4 PLCS
45,00 + 0,61/-0,00
(1.772 + .024/.000)
92,00 + 0,81/0,00
(3.622 + .032/.000)
NOTE: Dimensions in Millimeters (Inches)
54
COLR
S1.CF
S2.CF
S1.DB
S2.DB
OT.CF
OT.S.O
LK.CF
BRIT
SP1
SP2
Sensor Excitation
CE APPROVALS INFORMATION
This product conforms to the EMC directive 89/336/EEC amended by
93/68/EEC, and with the European Low Voltage Directive 72/23/EEC.
Electrical Safety EN61010-1:2001
Safety requirements for electrical equipment for measurement, control and laboratory.
Double Insulation
Pollution Degree 2
Dielectric withstand Test per 1 min
Power to Input/Output:
2300 Vac (3250 Vdc)
Power to Input/Output:
500 Vac (720 Vdc)
(Low Voltage dc Power Option*)
Isolated Analog:
500 V
Input/Output Signals:
500 V
Note: *Units configured for external low power dc voltage, 10-32 Vdc (Basic Insulation)
EMC EN61326:1997 + and A1:1998 + A2:2001
Immunity and Emissions requirements for electrical equipment for measurement,
control and laboratory.
EMC Emissions Table 4, Class B of EN61326
EMC Immunity** Table 1 of EN61326
Note: **I/O signal and control lines require shielded cables and these cables
must be located on conductive cable trays or in conduits. Furthermore,
the length of these cables should not exceed 30 meters
Refer to the EMC and Safety installation considerations (Guidelines) of this manual
for additional information.
56
Warranty/Disclaimer
NEWPORT Electronics, Inc. warrants this unit to be free of defects in materials and workmanship for a period of one (1)
year from the date of purchase. In addition to NEWPORTs standard warranty period, NEWPORT Electronics will extend
the warranty period for four (4) additional years if the warranty card enclosed with each instrument is returned to
NEWPORT.
If the unit should malfunction, it must be returned to the factory for evaluation. NEWPORTs Customer Service
Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination
by NEWPORT, if the unit is found to be defective it will be repaired or replaced at no charge. NEWPORTs WARRANTY
does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling, improper
interfacing, operation outside of design limits, improper repair, or unauthorized modification. This WARRANTY is VOID if
the unit shows evidence of having been tampered with or shows evidence of being damaged as a result of excessive
corrosion; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating
conditions outside of NEWPORTs control. Components which wear are not warranted, including but not limited to
contact points, fuses, and triacs.
NEWPORT is pleased to offer suggestions on the use of its various products. However, NEWPORT neither
assumes responsibility for any omissions or errors nor assumes liability for any damages that result from the
use of its products in accordance with information provided by NEWPORT, either verbal or written. NEWPORT
warrants only that the parts manufactured by it will be as specified and free of defects. NEWPORT MAKES NO
OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED,
EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF
MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF
LIABILITY: The remedies of purchaser set forth herein are exclusive and the total liability of NEWPORT with
respect to this order, whether based on contract, warranty, negligence, indemnification, strict liability or
otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event shall
NEWPORT be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by NEWPORT is not intended to be used, nor shall it be used: (1) as a Basic
Component under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or
used on humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application, or
used on humans, or misused in any way, NEWPORT assumes no responsibility as set forth in our basic WARRANTY /
DISCLAIMER language, and additionally purchaser will indemnify NEWPORT and hold NEWPORT harmless from any
liability or damage whatsoever arising out of the use of the Product(s) in such a manner.
Return Requests/Inquiries
Direct all warranty and repair requests/inquiries to the NEWPORT Customer Service Department.
BEFORE RETURNING ANY PRODUCT(S) TO NEWPORT, PURCHASER MUST OBTAIN AN
AUTHORIZED RETURN (AR) NUMBER FROM NEWPORTS CUSTOMER SERVICE DEPARTMENT
(IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on
the outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent
breakage in transit.
FOR WARRANTY RETURNS, please have the
following information available BEFORE
contacting NEWPORT:
1. P.O. number under which the product was
PURCHASED,
2. Model and serial number of the product under
warranty, and
3. Repair instructions and/or specific problems
relative to the product.
NEWPORTs policy is to make running changes, not model changes, whenever an improvement is possible. This
affords our customers the latest in technology and engineering.
NEWPORT is a registered trademark of NEWPORT Electronics, Inc.
Copyright 2006 NEWPORT Electronics, Inc. All rights reserved. This document may not be copied,
photocopied, reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in
part, without prior written consent of NEWPORT Electronics, Inc.
M3598/N/0906