Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

EJX530A

Download as pdf or txt
Download as pdf or txt
You are on page 1of 10

<<Contents>> <<Index>>

General EJX510A and EJX530A


Specifications Absolute and Gauge Pressure
Transmitter
GS 01C25F01-01E

The high performance absolute and gauge pressure


transmitter EJX510A and EJX530A feature single
crystal silicon resonant sensor and are suitable to
measure liquid, gas, or steam pressure. EJX510A and
EJX530A output a 4 to 20 mA DC signal correspond-
ing to the measured pressure. It also features quick
response, remote setup and monitoring via BRAIN or
HART communications, self-diagnostics, and optional
status output for pressure high/low alarm. FOUNDA-
TION Fieldbus protocol type is also available.
All EJX series models in their standard configuration,
with the exception of the Fieldbus type, are certified by
TÜV as complying with SIL 2 for safety requirement.

 STANDARD SPECIFICATIONS
Refer to GS 01C25T02-01E for Fieldbus communica-
tion type marked with “.”
Ambient Temperature Effects per 28˚C (50˚F)
 SPAN AND RANGE LIMITS Change
(For EJX510A, values are in absolute and lower ⫾(0.15% of Span + 0.15% of URL)
range limits are 0.) Stability (All normal operating condition)
Measurement ⫾0.1% of URL per 1 year
Span/Range MPa psi (/D1) bar (/D3) kgf/cm2 (/D4)
Power Supply Effects
Span 8 to 200 kPa 1.16 to 29 0.08 to 2 0.08 to 2 ⫾0.005% per Volt (from 21.6 to 32 V DC, 350 ⍀)
A
Range –100 to 200 kPa –14.5 to 29 –1 to 2 –1 to 2 Vibration Effects
Span 0.04 to 2 5.8 to 290 0.4 to 20 0.4 to 20 Amplifier housing code 1:
B
Range –0.1 to 2 –14.5 to 290 –1 to 20 –1 to 20 Less than 0.1% of URL when tested per the require-
Span 0.2 to 10 29 to 1450 2 to 100 2 to 100 ments of IEC60770-1 field or pipeline with high
C vibration level (10-60 Hz, 0.21 mm peak to peak
Range –0.1 to 10 –14.5 to 1450 –1 to 100 –1 to 100
Span 1 to 50 145 to 7200 10 to 500 10 to 500 displacement/60-2000 Hz 3 g)
D Amplifier housing code 2:
Range –0.1 to 50 –14.5 to 7200 –1 to 500 –1 to 500
T01E.EPS
Less than ±0.1% of URL when tested per the
requirements of IEC60770-1 field with general
 PERFORMANCE SPECIFICATIONS application or pipeline with low vibration level (10-60
Zero-based calibrated span, linear output, wetted Hz 0.15mm peak to peak displacement /60-500 Hz
parts material code ‘S’ and silicone oil, unless 2g)
otherwise mentioned. Mounting Position Effects
For Fieldbus communication type, use calibrated Rotation in diaphragm plane has no effect. Tilting up
range instead of span in the following specifications. to 90 degree will cause zero shift up to 0.21 kPa
Specification Conformance (0.84 inH2O) which can be corrected by the zero
EJX series ensures specification conformance to at adjustment.
least ⫾3␴. Response Time (All capsules) “”
Reference Accuracy of Calibrated Span 90 msec
(includes the effects of terminal-based linearity, When software damping is set to zero and including
hysteresis, and repeatability) dead time of 45 msec (nominal)
Measurement span A B C D
Reference Span ⭌ X ⫾0.1% of Span
accuracy Span ⬍ X ⫾(0.01+0.009 URL/Span) % of Span
20 kPa 0.2 MPa 1 MPa 5 MPa
X
(2.9 psi) (29 psi) (145 psi) (720 psi)
URL 200 kPa 2 MPa 10 MPa 50 MPa
(Upper range limit) (29 psi) (290 psi) (1450 psi) (7200 psi)
T02E.EPS

Yokogawa Electric Corporation GS 01C25F01-01E


2-9-32 Nakacho, Musashino-shi, Tokyo, 180-8750 Japan ©Copyright Jun. 2004
Phone: 81-422-52-5690 Fax.: 81-422-52-2018 7th Edition Feb. 2006
<<Contents>> <<Index>> 2
 FUNCTIONAL SPECIFICATIONS SIL Certification
EJX series transmitters except Fieldbus communica-
Output tion type are certified by RWTÜV Systems GmbH in
Two wire 4 to 20 mA DC output with digital communi- compliance with the following standards;
cations, linear or square root programmable. BRAIN IEC 61508: 2000; Part1 to Part 7
or HART FSK protocol are superimposed on the 4 to Functional Safety of Electrical/electronic/program-
20 mA signal. mable electronic related systems; SIL 2 capability for
Output range: 3.6 mA to 21.6 mA single transmitter use, SIL 3 capability for dual
Output limits conforming to NAMUR NE43 can be transmitter use.
pre-set by option code C2 or C3.
Failure Alarm (Output signal code D and E)  NORMAL OPERATING CONDITION
Output status at CPU failure and hardware error;
(Optional features or approval codes may affect
Up-scale: 110%, 21.6 mA DC or more (standard)
Down-scale: ⫺5%, 3.2 mA DC or less limits.)
Damping Time Constant (1st order) Ambient Temperature Limits
Amplifier’s damping time constant is adjustable from ⫺40 to 85˚C (⫺40 to 185˚F)
0.00 to 100.00 sec by software and added to re- ⫺30 to 80˚C (⫺22 to 176˚F) with LCD display
sponse time. Process Temperature Limits
⫺40 to 120˚C (⫺40 to 248˚F)
Note: For BRAIN protocol type, when the software
damping is set to less than 0.5 sec, communica- Ambient Humidity Limits
tion may occasionally be unavailble during the 0 to 100% RH
operation, especially while output changes Maximum Over Pressure
dynamically. The default setting of damping
ensures stable communication. Pressure
Update Period “” Capsule EJX510A EJX530A
Pressure: 45 msec A and B 4 MPa abs (580 psia) 4 MPa (580 psig)
Zero Adjustment Limits C 20 MPa abs (2900 psia) 20 MPa (2900 psig)
Zero can be fully elevated or suppressed, within the D 75 MPa abs (10800 psia) 75 MPa (10800 psig)
lower and upper range limits of the capsule.
External zero Adjustment Working Pressure Limits (Silicone oil)
External zero is continuously adjustable with 0.01% Maximum Pressure Limits
incremental resolution of span. Re-range can be
done locally using the digital indicator with range- Pressure
setting switch. Capsule EJX510A EJX530A
Integral Indicator (LCD display, optional) “” A 200 kPa abs (29 psia) 200 kPa (29 psig)
5-digit numerical display, 6-digit unit display and bar B 2 MPa abs (290 psia) 2 MPa (290 psig)
graph.
C 10 MPa abs (1450 psia) 10 MPa (1450 psig)
The indicator is configurable to display one or up to
three of the following variables periodically; D 50 MPa abs (7200 psia) 50 MPa (7200 psig)
pressure in %, scaled pressure, measured pressure.
See also “Factory Settings.”
Burst Pressure
A, B and C capsule : 30 MPa
D capsule : 132 MPa
Self Diagnostics
CPU failure, hardware failure, configuration error,
process alarm for pressure or capsule temperature.
User-configurable process high/low alarm for
pressure is also available, and its status can be
output when optional status output is specified.
Signal Characterizer (Output signal code D and E)
User-configurable 10-segment signal characterizer
for 4 to 20 mA output.
Status Output (optional, output signal code D and E)
One transistor contact output (sink type) to output the
status of user configurable high/low alarm for
pressure.
Rating: 10.5 to 30 V DC, 120 mA DC max.
Note: A check meter cannot be connected when status
output option (/AL) is specified.
Refer to ‘Wiring Example for Analog Output and
Status Output.’

All Rights Reserved. Copyright © 2004, Yokogawa Electric Corporation GS 01C25F01-01E Aug. 31, 2005-00
<<Contents>> <<Index>> 3
Minimum Pressure Limit
See graph below
100(750)
600

E-10.5
R=
0.0244 Digital
External Communication
load range
10(75) BRAIN and HART
resistance
250
2.7(20)
Working R (⍀)
pressure Applicable range
kPa abs 1(7.5)
(mmHg abs)

10.5 16.6 25.2 42


0.13(1) Power supply voltage E (V DC)
0.1(0.75) F03E.EPS

Figure 2. Relationship Between Power Supply Voltage


and External Load Resistance

0.013(0.1) Supply Voltage “”


0.01(0.075) -40 0 40 80 120 (248) 10.5 to 42 V DC for general use and flameproof type.
(-40) (32) (104) (176) 85 10.5 to 32 V DC for lightning protector (option code A).
Process temperature ⬚C(⬚F)
(185) 10.5 to 30 V DC for intrinsically safe, type n, non-
incendive or non-sparking type.
F01E.EPS
Minimum voltage limited at 16.6 V DC for digital
Figure 1-1. Working Pressure and Process Temperature communications, BRAIN and HART
[For EJX510A] Load (Output signal code D and E)
0 to 1290 ⍀ for operation
250 to 600 ⍀ for digital communication
100(14.5) Atmospheric
pressure Communication Requirements “”
Working
pressure (Approval codes may affect electrical require-
kPa abs ments.)
(psia)
BRAIN
Applicable range
Communication Distance
10(1.4)
Up to 2 km (1.25 miles) when using CEV polyethyl-
ene-insulated PVC-sheathed cables. Communication
distance varies depending on type of cable used.
2.7(0.38) Load Capacitance
0.22 ␮F or less
1(0.14)
Load Inductance
-40 0 40 80 120 3.3 mH or less
(-40) (32) (104) (176) (248) Input Impedance of communicating device
Process temperature ⬚C (⬚F) 10 k⍀ or more at 2.4 kHz.
F02E.EPS

Figure 1-2. Working Pressure and Process Temperature


[For EJX530A]

Supply & Load Requirements


(Output signal code D and E. Optional features
or approval codes may affect electrical require-
ments.)
With 24 V DC supply, up to a 550 ⍀ load can be
used. See graph below.

All Rights Reserved. Copyright © 2004, Yokogawa Electric Corporation GS 01C25F01-01E Aug. 31, 2005-00
<<Contents>> <<Index>> 4
HART < Related Instruments>
Communication Distance Power Distributor: Refer to GS 01B04T01-02E or
Up to 1.5 km (1 mile) when using multiple twisted pair GS 01B04T02-02E
cables. Communication distance varies depending on BRAIN TERMINAL: Refer to GS 01C00A11-00E
type of cable used.
< Reference >
Use the following formula to determine cable length 1. Hastelloy; Trademark of Haynes International Inc.
for specific applications: 2. HART; Trademark of the HART Communication
Foundation.
65 ⫻ 106 (Cf + 10,000) 3. FOUNDATION Fieldbus; Tradmark of Fieldbus
L= -
(R ⫻ C) C Foundation.
Other company names and product names used in this
Where: material are registered trademarks or trademarks of
L = length in meters or feet their respective owners.
R = resistance in ⍀ (including barrier resistance)
C = cable capacitance in pF/m or pF/ft
Cf = maximum shunt capacitance of receiving devices
in pF/m or pF/ft
EMC Conformity Standards ,
EN 61326, AS/NZS CISPR11
European Pressure Equipment Directive 97/23/EC
Sound Engineering Practice (for all capsules)

With option code /PE3 (for D capsule)

Category III, Module H, Type of Equipment: Pressure


Accessory-Vessel, Type of Fluid: Liquid and Gas,
Group of Fluid: 1 and 2

 PHYSICAL SPECIFICATIONS
Wetted Parts Materials
Diaphragm, Process Connector
Refer to “MODEL AND SUFFIX CODE.”
Non-wetted Parts Materials
Housing
Low copper cast aluminum alloy with polyurethane,
mint-green paint (Munsell 5.6BG 3.3/2.9 or its
equivalent) or ASTM CF-8M Stainless Steel
Degrees of Protection
IP67, NEMA4X, JIS C0920
Pipe
Polypropylene
Cover O-rings
Buna-N
Name plate and tag
304 SST
Fill Fluid
Silicone, fluorinated oil (option)
Weight
Capsule A, B and C: 1.1 kg (2.4 lb)*
Capsule D: 1.3 kg (2.9 lb)*
*: Without integral indicator and mounting bracket.
Add 1.5 kg (3.3lb) for Amplifier housing code 2.
Connections
Refer to “MODEL AND SUFFIX CODE.”

All Rights Reserved. Copyright © 2004, Yokogawa Electric Corporation GS 01C25F01-01E Aug. 31, 2005-00
<<Contents>> <<Index>> 5
 MODEL AND SUFFIX CODES

Model Suffix Codes Description


EJX510A ························ Absolute pressure transmitter
EJX530A ························ Gauge pressure transmitter
Output signal -D · · · · · · · · · · · · · · · · · · · · · · 4 to 20 mA DC Output with digital communication (BRAIN protocol)
-E · · · · · · · · · · · · · · · · · · · · · · 4 to 20 mA DC Output with digital communication (HART protocol)
-F · · · · · · · · · · · · · · · · · · · · · · Digital communication (FOUNDATION Fieldbus protocol, refer to GS 01C25T02-01E)
Measurement A····················· 8 to 200 kPa (1.16 to 29 psi)
span (Capsule) B····················· 0.04 to 2 MPa (5.8 to 290 psi)
C····················· 0.2 to 10 MPa (29 to 1450 psi)
D····················· 1 to 50 MPa (145 to 7200 psi)
Wetted parts Process connection Diaphragm
material *4 S ·················· 316L SST# Hastelloy C-276 *1#
H ·················· Hastelloy C-276 * 1# Hastelloy C-276 *1#
Process connections 4················· 1/2 NPT female
7················· 1/2 NPT male
8················· G1/2 DIN 16 288 male *2
9················· M20⫻1.5 DIN 16 288 male *2
— N··············· Always N
— -0 · · · · · · · · · · · · · Always 0
Amplifier housing  1············ Cast alluminum alloy
2············ ASTM CF-8M Stainless Steel *5
Electrical connection 0 ········· G1/2 female, one electrical connection without blind plugs
 2 ········· 1/2 NPT female, two electrical connections without blind plugs
4 ········· M20 female, two electrical connections without blind plugs
5 ········· G1/2 female, two electrical connections with a blind plug
7 ········· 1/2 NPT female, two electrical connections with a blind plug
9 ········· M20 female, two electrical connections with a blind plug
Integral indicator D······· Digital indicator
E······· Digital indicator with the range setting switch *3
 N······· None
Mounting bracket F······ 304 SST 2-inch pipe mounting
 N······ None
Optional code /  Optional specification
T03.EPS
The “” marks indicates the most typical selection for each specification. Example: EJX530A-DAS4N-012NN/.
*1: Hastelloy C-276 or ASTM N10276.
*2: Not applicable for combination of capsule code D and wetted parts material code H.
*3: Not applicable for output signal code F.
*4: Users must consider the characteristics of selected wetted parts material and the influence of process fluids. The use of
inappropriate materials can result in the leakage of corrosive process fluids and cause injury to personnel and/or
damage to plant facilities. It is also possible that the diaphragm itself can be damaged and that material from the broken
diaphragm and the fill fluid can contaminate the user’s process fluids.
Be very careful with highly corrosive process fluids such as hydrochloric acid, sulfuric acid, hydrogen sulfide, sodium
hypochlorite, and high-temperature steam (150°C [302°F] or above). Contact Yokogawa for detailed information of the
wetted parts material.
*5: Not applicable for electrical connection code 0.

# The ‘#’marks indicate the construction materials conform to NACE material recommendations per MR01-75. For the use
of 316 SST material, there may be certain limitations for pressure and temperature. Please refer to NACE standards for
details.

All Rights Reserved. Copyright © 2004, Yokogawa Electric Corporation GS 01C25F01-01E Feb. 10, 2006-00
<<Contents>> <<Index>> 6
 OPTIONAL SPECIFICATIONS (For Explosion Protected type) “”
Item Description Code
FM Explosionproof Approval *1
Applicable Standard: FM3600, FM3615, FM3810, ANSI/NEMA 250
Explosionproof for Class I, Division 1, Groups B, C and D, Dust-ignitionproof for Class II/III, Division 1, FF1
Groups E, F and G, in Hazardous locations, indoors and outdoors (NEMA 4X)
Temperature class: T6, Amb. Temp.: –40 to 60⬚C (–40 to 140⬚F)
FM Intrinsically safe Approval *1*2
Factory Applicable Standard: FM3600, FM3610, FM3611, FM3810
Mutual Intrinsically Safe for Class I, Division 1, Groups A, B, C & D, Class II, Division 1,
(FM) Groups E, F & G and Class III, Division 1, Class I, Zone 0, in Hazardous Locations, AEx ia IIC
Nonincendive for Class I, Division 2, Groups A, B, C & D, Class II, Division. 2, FS1
Groups F & G, and Class III, Division 1, Class I, Zone 2, Group IIC, in Hazardous Locations
Enclosure: "NEMA 4X", Temp. Class: T4, Amb. Temp.: –60 to 60⬚C (–75 to 140⬚F)
Intrinsically Safe Apparatus Parameters
[Groups A, B, C, D, E, F and G] Vmax=30 V, Imax=200 mA, Pmax=1 W, Ci=6 nF, Li=0 ␮H
[Groups C, D, E, F and G] Vmax=30 V, Imax=225 mA, Pmax=1 W, Ci=6 nF, Li=0 ␮H
Combined FF1 and FS1 *1*2 FU1
CENELEC ATEX (KEMA) Flameproof Approval *1
Applicable Standard: EN 50014, EN 50018, EN 50281-1-1
Certificate: KEMA 03ATEX2570
II 2G,1D EExd IIC T4, T5, T6 Type of protection : IP66 and IP67
Amb. Temp. (Tamb) for gas-proof : KF2
T4 ; –50 to 75°C (–58 to 167°F), T5; –50 to 80⬚C (–58 to 176⬚F), T6; –50 to 70°C (–58 to 158°F)
Max. process Temp.(Tp): T4; 120⬚C (248⬚F), T5; 100⬚C (212⬚F), T6; 85⬚C (185⬚F)
Max. surface Temp. for dust-proof : T80°C (Tamb: –40 to 40⬚C, Tp:80⬚C),
CENELEC T100°C (Tamb: –40 to 60⬚C, Tp:100⬚C), T120°C (Tamb: –40 to 80⬚C, Tp:120⬚C)
ATEX CENELEC ATEX (KEMA) Intrinsically safe Approval *1*2
Applicable Standard: EN 50014, EN 50020, EN 50284, EN 50281-1-1
Certificate: KEMA 03ATEX1544X
II 1G, 1D EEx ia IIC T4 Type of protection : IP66 and IP67
KS2
Amb. Temp.(Tamb) for gas-proof: –50 to 60⬚C (–58 to 140⬚F)
Maximum Process Temp.(Tp) for gas-proof :120⬚C
Electrical data : Ui=30 V, Ii=200 mA, Pi=0.9 W, Ci=10 nF, Li=0 mH
Max. surface Temp. for dust-proof : T85°C (Tamb: –40 to 60⬚C, Tp:80⬚C),
T100°C (Tamb: –40 to 60⬚C, Tp:100⬚C), T120°C (Tamb: –40 to 60⬚C, Tp:120⬚C)
Combined KF2, KS2 and Type n *1*2
Type n
Applicable Standard: EN 50021, EN 60529 KU2
II 3G EEx nL IIC T4, Amb. Temp.: –50 to 60⬚C (–50 to 140⬚F)
Ui=30 V DC, Ci=10 nF, Li=0 mH
T04E.EPS
*1: Applicable for electrical connection code 2, 4, 7, and 9.
*2: Not applicable for option code /AL.

All Rights Reserved. Copyright © 2004, Yokogawa Electric Corporation GS 01C25F01-01E Feb. 10, 2006-00
<<Contents>> <<Index>> 7
Item Description Code
CSA Explosionproof Approval *2
Certificate: 1589701
[For CSA C22.2]
Applicable Standard: C22.2 No.0, C22.2 No.0.4, C22.2 No.0.5, C22.2 No.25, C22.2 No.30,
C22.2 No.94
Explosion-proof for Class I, Groups B, C and D. CF1
Dustignition-proof for Class II/III, Groups E, F and G.
When installed in Division 2, “SEAL NOT REQUIRED” Enclosure: TYPE 4X, Temp. Code: T6...T4
[For CSA E60079]
Applicable Standard: CAN/CSA E60079-0, CAN/CSA E60079-1
Flameproof for Zone 1, Ex d IIC T6...T4 Enclosure: IP66 and IP67
Canadian Max.Process Temp.: T4;120⬚C(248⬚F), T5;100⬚C(212 ⬚F), T6; 85⬚C(185⬚F)
Standards Amb.Temp.:–50 to 75⬚C(–58 to 167⬚F) for T4, –50 to 80⬚C(–58 to 176⬚F) for T5,
Association –50 to 70⬚C(–58 to 158⬚F) for T6
(CSA)
CSA Intrinsically safe Approval *2*3
Certificate: 1606623
[For CSA C22.2]
Applicable Standard: C22.2 No.0, C22.2 No.0.4, C22.2 No.25, C22.2 No.94, C22.2 No.154,
C22.2 No.213, C22.2 No.1010.1
Intrinsically Safe for Class I, Division 1, Groups A, B, C & D, Class II, Division 1, Groups E, F & G,
Class III, Division 1, Nonincendive for Class I, Division 2, Groups A, B, C & D, Class II, Division 2,
Groups E, F & G, Class III, Division 1
Enclosure: Type 4X, Temp. Code: T4 Amb. Temp.:–50 to 60⬚C(–58 to 140⬚F)
Electrical Parameters: [Intrinsically Safe] Vmax=30V, Imax=200mA, Pmax=0.9W, Ci=10nF, Li=0 CS1
[Nonincendive] Vmax=30V, Ci=10nF, Li=0
[For CSA E60079]
Applicable Standard: CAN/CSA E60079-0, CAN/CSA E60079-11, CAN/CSA E60079-15,
IEC 60529:2001-02
Ex ia IIC T4, Ex nL IIC T4 Enclosure: IP66 and IP67
Amb. Temp.:–50 to 60⬚C(–58 to 140⬚F), Max. Process Temp.: 120⬚C(248⬚F)
Electrical Parameters: [Ex ia] Ui=30V, Ii=200mA, Pi=0.9W, Ci=10nF, Li=0
[Ex nL] Ui=30V, Ci=10nF, Li=0

Combined CF1 and CS1 *2*3 CU1


IECEx Intrinsically safe, type n and Flameproof Approval *1*3
Intrinsically safe and type n
Applicable Standard: IEC 60079-0:2000, IEC 60079-11:1999, IEC 60079-15:2001
Certificate: IECEx CSA 05.0005
IECEx Ex ia IIC T4, Ex nL IIC T4 Enclosure: IP66 and IP67
Scheme *4 Amb. Temp.:–50 to 60⬚C(–58 to 140⬚F), Max. Process Temp.: 120⬚C(248⬚F) SU2
Electrical Parameters: [Ex ia] Ui=30V, Ii=200mA, Pi=0.9W, Ci=10nF, Li=0
[Ex nL] Ui=30V,Ci=10nF, Li=0
Flameproof
Applicable Standard: IEC 60079-0:2000, IEC60079-1:2001
Certificate: IECEx CSA 05.0002
Flameproof for Zone 1, Ex d IIC T6...T4 Enclosure: IP66 and IP67
Max.Process Temp.: T4;120⬚C(248⬚F), T5;100⬚C(212⬚F), T6; 85⬚C(185⬚F)
Amb.Temp.:–50 to 75⬚C(–58 to 167⬚F) for T4, –50 to 80⬚C(–58 to 176⬚F) for T5,
–50 to 70⬚C(–58 to 158⬚F) for T6
T04Eb.EPS
*1: Applicable for electrical connection code 2, 4, 7, and 9.
*2: Applicable for electrical connection code 2 and 7.
*3: Not applicable for output signal code -F and option code /AL.
*4: Applicable only for Australia and New Zealand area.

All Rights Reserved. Copyright © 2004, Yokogawa Electric Corporation GS 01C25F01-01E Aug. 31, 2005-00
<<Contents>> <<Index>> 8
 OPTIONAL SPECIFICATIONS
Item Description Code
Amplifier cover only *14 P
Color change
Painting Amplifier cover and terminal cover, Munsell 7.5 R4/14 PR
Coating change Anti-corrosion coating *1*14 X2
Transmitter power supply voltage: 10.5 to 32 V DC ( 10.5 to 30 V DC for intrinsically
safe type, 9 to 32 V DC for Fieldbus communication type.) A
Lightning protector
Allowable current: Max. 6000 A ( 1⫻40 ␮s ), Repeating 1000 A ( 1⫻40 ␮s ) 100 times
Applicable Standards: IEC 61000-4-4, IEC 61000-4-5
Transistor output (sink type)
Status output *2 AL
Rating: 10.5 to 30 V DC, 120 mA DC (max) Low level: 0 to 2 V DC
Degrease cleansing treatment K1
Oil-prohibited use *13 Degrease cleansing treatment and with fluorinated oilfilled capsule.
K2
Operating temperature –20 to 80°C (–4 to 176°F)
Capsule fill fluid Fluorinated oil filled in capsule K3
P calibration (psi unit) D1
Calibration units *3 (See Table for Span and
bar calibration (bar unit) D3
Range Limits.)
M calibration (kgf/cm2 unit) D4
Failure alarm down-scale: Output status at CPU failure and hardware error is –5%,
C1
3.2mA DC or less.
Output limits NAMUR NE43 Compliant Failure alarm down-scale: Output status at CPU
C2
and failure operation *4 Output signal limits: 3.8mA to failure and hardware error is –5%, 3.2mA DC or less.
20.5 mA Failure alarm up-scale: Output status at CPU
C3
failure and hardware error is 110%, 21.6 mA or more.
Stainless steel tag plate 304 SST stainless steel tag plate wired onto transmitter N4
Data configuration for HART communication type Software damping, Descriptor, Message CA
Data configuration at factory *5
Data configuration for BRAIN communication type Software damping CB
PED 97/23/EC
European Pressure Equipment
Category III, Module H, Type of Equipment: Pressure Accessory-Vessel, PE3
Directive *15
Type of Fluid: Liquid and Gas, Group of Fluid: 1 and 2
Material certificate*6 Process Connector M15
Test Pressure: 200 kPa (29 psi) *7 T05
Pressure test Test Pressure: 2 MPa (290 psi) *8 *11 T06
Nitrogen(N2) Gas or Water
/Leak test certificate *12 Test Pressure: 10 MPa (1450 psi) *9 Retention time: one minute T07
Test Pressure: 50 MPa (7200 psi) *10 T08
T05.EPS
*1: Not applicable with color change option.
*2: Check/External indicator terminals cannot be used when this option code is specified. Not applicable for output signal
code F and amplifier housing code 2.
*3: The unit of MWP (Max. working pressure) on the name plate of a housing is the same unit as specified by option codes
D1, D3, and D4.
*4: Applicable for output signal codes D and E. The hardware error indicates faulty amplifier or capsule.
*5: Also see ‘Ordering Information’.
*6: Material traceability certification, per EN 10204 3.1 B.
*7: Applicable for capsule code A.
*8: Applicable for capsule code B.
*9: Applicable for capsule code C.
*10: Applicable for capsule code D.
*11: Pure nitrogen gas or pure water is used for oil-prohibited use (option codes K1 and K2).
*12: The unit on the certificate is always kPa/MPa regardless of selection of option code D1, D3 and D4.
*13: Applicable for wetted parts material code S.
*14: Not applicable for amplifier housing code 2.
*15: Applicable for measurement span code D. If compliance with category III is needed, specify this option code.

All Rights Reserved. Copyright © 2004, Yokogawa Electric Corporation GS 01C25F01-01E Aug. 31, 2005-00
<<Contents>> <<Index>> 9
 DIMENSIONS
 Model EJX510A and EJX530A
 With process connections code 7 Unit: mm (Approx. inch)

95(3.74)

60(2.36)*2 188(7.4)

External indicator Zero


Conduit connection 110(4.33)
adjustment
(optional) Conduit
connection 12 39
54 (0.47) (1.54)
(2.13)
6

ø77(3.03)

(2.72)
(0.23)

ø69
Integral
129(5.08)

indicator

110(4.33)
163(6.42)

169(6.65)
(optional)

Electrical connection Shrouding bolt


for code 5 and 9. Ground
(for flame-proof type) terminal

Pipe for
open to
atmosphere*1
2-inch pipe
Mounting (O.D. 60.5mm)
bracket 129(5.08)
(optional) 12*3
(0.47)

*1: Only for EJX530A whose measurement span code is A, B, or C.


*2: 58 mm (2.28 inch) for measurement span code D.
*3: 11 mm (0.43 inch) for measurement span code D.

 With process connections code 4


158(6.22)

 With process connections code 8 and 9


20(0.78)

174(6.85)
5(0.20)

ø6(0.23)
F04.EPS

● Terminal Configuration ● Terminal Wiring


Communication + Power supply and output terminal
Check meter SUPPLY –
terminals (BT200 etc.)
connection hook*4*5 + External indicator (ammeter) terminal*4*5
connection hook CHECK –
or or
+ Status contact output terminal*5
ALARM – (when /AL is specified)
Ground terminal
CHECK + or
ALARM + *4: When using an external indicator or check meter, the internal
SUPPLY +
resistance must be 10 ⍀ or less. A check meter or indicator
SUPPLY – CHECK – or ALARM – cannot be connected when /AL option is specified.
*5: Not available for fieldbus communication type.
F05.EPS

All Rights Reserved. Copyright © 2004, Yokogawa Electric Corporation GS 01C25F01-01E Feb. 10, 2006-00
<<Contents>> <<Index>> 10
● Wiring Example for Analog Output and Status Output

Connection Description
Analog output EJX electrical terminal
Distributor
+ 24V DC
SUPPLY +
– 250Ω

CHECK +

Analog and EJX electrical terminal Shielded cable


status output 24V DC
(when /AL is specified) +
SUPPLY + Distributor
– – 250Ω
If shield cable is not
used, communication is ALARM +
not possible.
Magnetic
valve
External power supply
30V DC, 120mA max

Use two-wire separately shielded cables. AC power supply

F06.EPS

< Ordering Information > “” < Factory Setting > “”
Specify the following when ordering
Tag number As specified in order
1. Model, suffix codes, and option codes
2. Calibration range and units Software
damping *1 ‘2.00 sec’ or as specified in order
1) Calibration range can be specified with range
value specifications up to 5 digits Calibration range
(excluding any decimal point) for low or high range lower range value As specified in order
limits within the range of -32000 to 32000. When Calibration range As specified in order
reverse range is designated, specify Lower Range upper range value
Value (LRV) as greater than Upper Range Value [EJX530A]
(URV). Selected from mmH2O, mmH2O(68˚F),
2) Specify only one unit from the table, ‘Factory mmAq*2, mmWG*2, mmHg, Pa, hPa*2,
Settings’ when shipped.’ kPa, MPa, mbar, bar, gf/cm2, kgf/cm2,
inH2O, inH2O(68˚F), inHg, ftH2O,
3. Display scale and units (for transmitters equipped
ftH2O(68˚F) or psi. (Only one unit can be
with integral indicator only) Calibration range specified)
Specify either 0 to 100 % or engineering unit scale units
and ‘Range and Unit’ for engineering units scale: [EJX510A]
Scale range can be specified with range limit Torr, Pa abs, hPa abs*2, kPa abs, MPa
specifications up to 5 digits (excluding any decimal abs, mbar abs, bar abs, kgf/cm2 abs,
mmH2O abs, mmH2O abs(68°F),
point) for low or high range limits within the range of
mmHg abs, inH2O abs, inH2O abs(68°F),
-32000 to 32000. The unit display consists of 6-digit, inHg abs, ftH2O abs, ftH2O abs(68°F),
therefore, if the specified unit is longer than 7 psia, atm.
characters excluding ‘/’, the first 6 characters will be
displayed on the unit display. Designated value specified in order.
Display setting
4. Tag Number (if required) (% or user scaled value.)
For BRAIN communication type, specify upto 16 T06E.EPS

letters. The specified letters will be written in the *1: To specify these items at factroy, /CA or /CB option
is required.
amplifier memory and engraved on the tag plate. *2: Not available for HART protocol type.
For HART communication type, specify software tag
(up to 8 letters) to be written on the amplifier memory
and Tag number(up to 16 letters) to be engraved on
the tag plate seperately.
5. Other factory configurations (if required)
Specifying option code CA or CB will allow further
configuration at factory. Following are configurable
items and setting range.
[/CA: For HART communication type]
1) Descriptor(up to 16 characters)
2) Message (up to 30 characters)
3) Software damping in second (0.00 to 100.00)
[/CB: For BRAIN communication type]
1) Software damping in second (0.00 to 100.00)

All Rights Reserved. Copyright © 2004, Yokogawa Electric Corporation GS 01C25F01-01E Jan. 05, 2005-00
Subject to change without notice.

You might also like