Datcom K3
Datcom K3
Datcom K3
TYPE
DATCOM-K3
1
Technical description
DATCOM-K3 is designed as accessories for the correctors miniELCOR, maxiELCOR,
midiELCOR, ELCOR-2, microELCOR-2 and electronic data logger miniDATCOM, maxiDATCOM, midi-
DATCOM and DATCOM-2. The device is designed as associated apparatus from the point of intrinsic
safety and the device has to be installed outside of hazardous area during operation. Safety standards
and safety parameters of intrinsic safety have to be respected with regards to using the device with other
devices (see chapter Technical data).
Communication module DATCOM-K3 is power supplied with direct tension 12 V.
Communication module DATCOM-K3 is manufactured into plastic box and is assembled on DIN
bar 35 mm into distributor. Clamps with maximum cross-section 1,5 mm2 are determined for connection of
the cables.
2
Technical data
Mechanical parameters
plastic box for assembly on DIN bar
dimensions (width x height x length) 93 x 42 x 96 mm (height without holder)
weight 0,16 kg
clamps for connection of cables 1,5 mm2
ambient temperature -25 C to +60 C
storing temperature -40 C to +85 C
protection IP20 (according to EN 60529)
Non-hazardous variant
classification II (2)G [Ex ia] IIC
ATEX approval FTZ 05 ATEX 0364
classification of environment - hazardous area
- normal environment
(according to EN 33 2000-3)
galvanic separation (only for DATCOM-K3) 1 500 V
Power supply range (clamps 12V)
power supply range 12 VDC -10% / +25%
max. value of voltage Um 250 V (only for DATCOM-K3)
60 V (only for DATCOM-K3/A)
idle (own) current consumption *) typically 28 mA at 14 V (DATCOM-K3)
typically 2 mA (DATCOM-K3/A)
max. current consumption **) 120 mA (DATCOM-K3)
100 mA (DATCOM-K3/A)
max. length of cable 30 m
*) to the clamps, resp. connector is not connected to any outer circuit
**) to the connector, D-Sub is connected to the device (PC, modem) output clamps of intrinsically safe power supply
(6 V OUT) are short-circuited
Port P0 (intrinsically safe port)
communication interface RS-485
communication speed 4 800 Bd to 38 400 Bd
max. length of cable 100 m 1)
Port P2
communication speed 4800 Bd to 38 400 Bd
- communication interface RS232 (connector D-Sub9)
max. length of cable 30 m
- communication interface RS485 (terminals D-, D+)
max. length of cable 1 200 m
number of line wire 2
Digital inputs DI1 to DI4 (intrinsically safe)
number of inputs 4
max. length of cable 30 m 1)
Digital outputs DO1 to DO4
number of outputs 4
type of output open collectro
max. length of cable 30 m
max. voltage 30 V
max. current 100 Ma
max. resistance at fastened state 10 R
Intrinsically safe power supply (clamps 6V OUT)
off-load voltage typically 6,2 V
current limitation typically 74 mA
max. length of cable 30 m 1)
1)
Inductivity and capacity of cable (depends on used length and type of cable) has to be in accordance
with parameters of system non explosiveness.
3
Basic variant DATCOM-K3
Basic variant is determined for systems which are powered from network voltage (Um = 250V).
DATCOM-K3 may be powered from not back-up mains source or from the source which is back-up with
accumulator, charged with mains charger, etc.
Variant DATCOM-K3/A
Manufacturing variant DATCOM-K3/A has got decreased consumption with comparison to basic
variant. Using of DATCOM-K3/A is determined for systems which are powered from accumulator only.
Accumulator may also be charged with solar panel (Um = 60V).
This manufacturing variant does not have galvanic separation between intrinsically safe input cir-
cuit (blue clamps) and other circuit (orange clamps). All GND clamps are interconnected mutually.
The company is a holder of certified quality system ISO 9001 with international validity
and it is authorised to manufacture equipment for hazardous areas.
L060EN_201201 4