Ek205 Manual
Ek205 Manual
Ek205 Manual
Contents
1 General .............................................................................................................................. 7
1.1 Information in these instructions ...................................................................................... 7
1.2 Warranty provisions ......................................................................................................... 7
1.3 Customer service ............................................................................................................ 7
1.3.1 Customer services and repairs................................................................................. 7
1.3.2 Electronics hotline .................................................................................................... 7
1.4 Explanation of symbols ................................................................................................... 8
1.4.1 Safety instructions .................................................................................................... 8
1.4.2 Tips and recommendations ...................................................................................... 8
1.5 Limitation of liability ......................................................................................................... 9
1.6 Copyright ......................................................................................................................... 9
1.7 Scope of delivery ............................................................................................................. 9
1.8 Replacement and accessory parts ................................................................................ 10
1.9 Storage .......................................................................................................................... 11
2 Safety............................................................................................................................... 12
2.1 General ......................................................................................................................... 12
2.2 Intended use.................................................................................................................. 13
2.3 Staff ............................................................................................................................... 14
2.4 Personal protective equipment ...................................................................................... 15
2.5 Special hazards ............................................................................................................. 15
2.6 Environmental protection ............................................................................................... 16
2.7 Operator liability ............................................................................................................ 16
3 Technical data................................................................................................................. 17
3.1 General data.................................................................................................................. 17
3.2 Power supply for EK205 without integrated mains adapter ........................................... 17
3.2.1 Battery supply ........................................................................................................ 17
3.2.2 External power supply for the basic device ............................................................ 17
3.3 Pressure sensor ............................................................................................................ 18
3.3.1 CT30 Type Pressure Sensor .................................................................................. 18
3.4 Temperature sensor ...................................................................................................... 18
3.5 Digital inputs .................................................................................................................. 19
3.5.1 Pulse and reporting inputs NF ................................................................................ 19
3.6 Digital outputs................................................................................................................ 19
3.7 Optical serial interface ................................................................................................... 20
3.8 Electrical serial interface ............................................................................................... 20
3.8.1 RS485 type ............................................................................................................ 20
3.9 Operating conditions ..................................................................................................... 20
3.9.1 Environment ........................................................................................................... 20
3.10 Identification ................................................................................................................ 21
3.10.1 Type plate and EX identification ........................................................................... 21
3.10.2 Device software identification ............................................................................... 21
4 Construction and function ............................................................................................. 22
4.1 External view ................................................................................................................. 22
4 General
1 General
The data and material properties that are presented below are guidelines. They
must be reviewed for each individual case and corrected if necessary.
You can also contact the electronics hotline in the event of faults.
DANGER!
…indicates an immediately hazardous situation that leads to death or severe
injury if it is not avoided.
WARNING!
…indicates an immediately hazardous situation that may lead to death or severe
injury if it is not avoided.
CAUTION!
…indicates an immediately hazardous situation that may lead to minor or slight
injury if it is not avoided.
CAUTION!
…indicates an immediately hazardous situation that may lead to material
damage if it is not avoided.
Read through these operating instructions carefully before beginning any work
to and with the device, especially before commissioning the device!
The manufacturer assumes no liability for loss and malfunctions that result from
non-compliance with these instructions.
We reserve the right to make technical changes within the scope of improving performance
characteristics and continuous development of the device.
1.6 Copyright
These instructions are copyright-protected and intended for internal purposes only. Handover
to third parties, copies of any form, including extracts, as well as utilization and/or notification
of the content without written approval of the manufacturer is not permitted, except for
internal purposes. Violations give rise to compensation. Further claims remain reserved.
WARNING!
Incorrect replacement parts and accessories are a safety risk!
Incorrect or defective replacement parts and accessories may detract from
safety and lead to damage, malfunction, or total device failure.
Therefore:
– Use only original replacement parts and accessories from the
manufacturer.
– Always contact the manufacturer if you are in doubt.
The replacement parts and accessories list can be found in the appendix. Replacement parts
and accessories can be ordered from a contract dealership or directly from our customer
service.
The accessories of the EK205 also include the free enSuite program, available at
www.elster-instromet.com. With them, you can program the Volume Conversion Device
EK205 via its data interfaces in order to carry out further applications.
The EK205 is available in numerous equipment variants.
Details can be found at www.elster-instromet.com as well as in the chapter Installation,
connection and commissioning.
General 11
1.9 Storage
CAUTION!
Reduced performance after exceeding or falling below the valid
temperature range of the batteries.
Exceeding or falling below the valid temperature range can reduce the
performance of the batteries when storing the device.
Therefore:
– For longer-term storage, ensure that the valid temperature range of the
built-in batteries between -25 °C and +55 °C is not exceeded or fallen
below.
CAUTION!
Material damage from formation of condensation!
Storing the device can lead to the formation of condensation resulting from
variations in temperature. This may result in the device malfunctioning at a
later time.
Therefore:
– After the device has been stored or transported in cold weather or if it has
been subject to extreme variations in temperature, bring it slowly to room
temperature before it is commissioned.
– The device must undergo a waiting period of at least 12 hours before it is
put into operation if condensation formed during storage.
If the power supply of the device is interrupted during storage from clamping
the batteries, the time and date need to be reset.
2 Safety
This section provides an overview of all key safety aspects for optimal protection of the staff
as well as for safe and fault-free operation.
Non-observance of these operating and safety instructions may result in considerable
hazards.
2.1 General
The EK205 is an intrinsically safe device and is suitable for operation within the zones at risk
of gas explosion 0, 1 or 2 for gases of temperature class T4.
Check for use in zones 0, 1 or 2 in accordance with the currently applicable rules of the art:
see chapter 9.3: EC type examination certificate no. LCIE 16 ATEX 3007 X and no. IECEx
LCIE 16.0003X1
DANGER!
Risk of explosion by connection to non-intrinsically-safe and associated
equipment.
When operating the EK205 in an area at risk of explosion (zones 0, 1 or 2) and
connecting devices without certification as an "associated piece of equipment",
there is a risk of explosion.
Therefore:
– When used in an area at risk of explosion, connect the EK205 to certified,
associated pieces of equipment only in accordance with the valid rules of the
art stated in the design test certificates.
– Connect the EK205 only to intrinsically safe pieces of equipment whose
electric data correspond to the requirements stated in the design test
certificates of the EK205 (see chapter 9.3).
DANGER!
Risk of explosion from use of wrong batteries.
Use only batteries stated in the design test certificates of the EK205 (see
chapter 9.3). Order number, see section 9.1.4.
During connection and operation of the EK205 in areas at risk of gas explosion,
observe the associated norms:
DIN EN 60079-0
The device may only be used in an area at risk of gas explosion if the
installation is carried out in accordance with the general requirements of DIN EB
60079-0 and the operating conditions (see chapter Technical data) and
connection conditions (see chapter Installation, connection and commissioning)
are observed.
The device can be hazardous if it is used by staff not trained correctly in an improper or
unintended manner.
Anyone appointed to perform work to or with the device must read and understand
these operating instructions before beginning work on the device. This also applies if
the individuals concerned have already worked with such a device or a similar one, or
were instructed by the manufacturer.
Knowledge of the content is one of the requirements of protecting staff from hazards
and avoiding faults and therefore operating the device in a safe and fault-free manner.
In order to avoid risks and to ensure that the device performs in an optimal manner, no
changes or modifications that were not expressly authorized by the manufacturer may
be performed to the device.
Always keep all operating instructions on the device in easily legible condition. Renew
damaged or illegible instructions.
Observe settings values or value ranges stated in the instructions.
The Volume Conversion Device EK205 serves to convert a gas volume of a gas line in
basic condition measured by a meter in measurement conditions as well as to assign the
volumes measured to tariffs. In addition, using the device, further parameters can be
measured, recorded and monitored depending on the configuration set by the user.
Compliance with all the specifications in these operating instructions also falls under the
device's intended use. Any use of the device that goes beyond or deviates from its intended
use is considered a misuse of the device, and may lead to hazardous situations. Claims of
any kind due to loss resulting from non-intended use of the device are excluded.
WARNING!
Hazard from incorrect use!
Incorrect use of the device may result in hazardous situations.
Therefore:
– Use the device only as intended.
– Do not use the device to control the gas flow or other parameters influencing
the gas volume within the framework of the overall system.
14 Safety
2.3 Staff
WARNING!
Risk of injury in the event of insufficient qualification!
Improper use may result in considerable personal injury and material damage.
Therefore:
– Have all activities carried out by qualified staff only.
The following qualifications for different areas of activity are listed below in the instructions:
Instructed staff
was instructed by the plant operator in an informational session on the tasks assigned
to him or her, and on possible hazards in case of improper behaviour.
Specialist staff
has the ability, because of his or her technical training, knowledge and experience, as
well as his or her knowledge of the relevant regulations, to carry out the work to the
device assigned to him or her, and to recognize and avoid possible hazards on his or
her own.
Gas specialist
has the ability, because of his or her technical training, knowledge and experience, as
well as his or her knowledge of the relevant standards and regulations, to carry out
work to gas systems, and to recognize possible hazards on his or her own. A gas
specialist receives training for the specific location in which he or she works, and is
acquainted with the relevant standards and regulations.
Calibration officer
has the ability, because of his or her technical training, knowledge and experience, as
well as his or her knowledge of the relevant standards and regulations, to carry out
calibration-protected work to gas systems. The calibration officer is trained to work
with calibration-protected devices and systems, and is acquainted with the relevant
standards and regulations that apply.
Electrical expert
has the ability, because of his or her technical training, knowledge and experience, as
well as his or her knowledge of the relevant standards and regulations, to carry out
work to electrical systems, and to recognize and avoid possible hazards on his or her
own. A qualified electrician receives training for the specific location in which he/she
works, and is acquainted with the relevant standards and regulations.
WARNING!
Risk for unauthorized persons!
Unauthorized persons who do not fulfil the requirements described do not know the
hazards in the work area.
Therefore:
– Keep unauthorized persons away.
– In the event of doubt, speak to people and direct them out of the working area.
– Interrupt the work as long as unauthorized persons are in the working area.
Safety 15
Only persons are permitted as staff from whom it can be expected that they will perform their
work reliably. Persons whose reactivity is influenced, e.g. from drugs, alcohol or medication,
are not permitted.
When selecting personnel to operate the device, make sure you comply with the
specific regulations of the overall gas system that concern age and occupation.
WARNING!
Risk of injury when handling batteries incorrectly!
WARNING!
Fire hazard from easily flammable substances!
Easily flammable substances, liquids or gases can ignite and cause severe to fatal injuries.
Therefore:
– Do not smoke inside the hazardous area and in the near proximity. Do not handle open
flames or sources of ignition.
– Keep a fire extinguisher ready.
– Report suspicious substances, liquids or gases to the person in charge immediately.
– Suspend work immediately in the event of fire. Leave the hazardous area until the all-
clear has been given.
16 Safety
CAUTION!
Substances harmful to the environment!
When handling substances harmful to the environment, in particular in the event of
incorrect disposal, severe damage may arise for the environment.
Therefore:
– Observe the below notes at all times.
– Take suitable measures immediately if substances harmful to the environment enter the
environment by mistake. In the event of doubt, inform the municipal authorities in charge
of the damage.
3 Technical data
3.9.1 Environment
Information Value Unit
Temperature range -25 – +55 °C
Relative humidity, maximum 93 %
Humidity conditions acc. EN12405-1 condensation
Point of use acc. EN12405-1 open
Mechanical ambient conditions acc. EN12405-1 M2
Electromagnetic ambient conditions acc. EN12405-1 E2
2
Unit load: Standard RS485 receiver with an input resistance = 12kOhm
3
Details on the connection of the RS485 interface, see application manual
Technical data 21
3.10 Identification
The EK205 is approved as a volume corrector as per the Measuring Instruments Directive
(MID). The identification of the EK205 is carried out on the front side (see chapter Structure
and function).
1
10
2
9
3
8
4
5 6 7
Ill.: 1
4 Depending on the device design or country of destination, the type plate may contain other information.
5 Not available for all device types.
22 Construction and function
7 6 5 10
Ill.: 2
Ill.: 3
4.4 Connections
The Volume Conversion Device EK205 has connection options for:
Two batteries
To monitor and register the calculated data and to pass on the data and program its
functions, the following are available:
2 meter and reporting inputs DE1, DE2
2 pulse and reporting outputs DA1, DA2
Serial data interface
Optical data interface
Details on the connection options of the EK205 and the deliverable equipment
variants can be found under chapter Technical data and under chapter
Installation, connection and commissioning.
24 Installation, connection and commissioning
5.1 Installation
The EK205 can be installed either on the gas meter, on the pipeline or on a wall.
In the event of issues during installation, e.g. when selecting the suitable
installation accessories, please contact our customer service (see chapter
General).
Ill.: 6
3 1
of the Volume Conversion Device
3 : Test connection: Possibility to take test
2 pressure or to gibe external pressure on
the pressure sensor of the Volume
Conversion Device.
Ill.: 7
When installing the three-way cock, it must be noted that the position of the
operating lever is checked with the corresponding throughputs, as the lever
can be removed and may be installed in a twisted manner.
The pipeline from the pressure sensor to the meter must be laid in a falling
manner so that, where applicable, water cannot damage the pressure sensor
or influence the measurement precision.
5.2 Connection
The EK205 is available in various equipment variants. Information about this can
be found at www.elster-instromet.com.
DANGER!
Risk of explosion by connection to non-intrinsically-safe and associated
equipment.
When operating the EK205 in an area at risk of explosion (zones 0, 1 or 2) and
connecting devices without certification as an "associated piece of equipment",
there is a risk of explosion.
Therefore:
– When used in an area at risk of explosion, connect the EK205 to certified,
associated pieces of equipment only in accordance with the valid rules of the
art stated in the design test certificates.
– Connect the EK205 only to intrinsically safe pieces of equipment whose
electric data correspond to the requirements stated in the design test
certificates of the EK205 (see chapter 9.3).
WARNING!
Risk from faulty connection of the device!
The connection of the device may be carried out by a gas expert only (see
chapter Safety). Faults in this process may result in fatal situations or
considerable material damage.
Therefore:
– Have the connection of the calibrated device be carried out by a gas expert
only.
– Also involve this person for subsequent relocations.
– Do not perform any connection changes or relocations of the device
independently.
During connection and commissioning of the EK205, observe the requirements of the
associated norms DIN EN 60079-0 and DIN EN 60079-14.
Have the wiring of the connections carried out correctly by a gas expert or a calibration
officer.
Do not switch active outputs against one another.
Close unused cable screw connections in accordance with DIN EN 60079-14 using a
plug or a suitable screw lid.
To program the device and carry out additional applications, you can also connect in addition
to the components stated in this chapter the serial and optical interfaces of the EK205 (see
chapter Construction and function).
Installation, connection and commissioning 27
Ill.: 8 1
Lubricate the temperature sensor before connecting with heat transfer paste to
increase its functionality.
28 Installation, connection and commissioning
6 1
1. Insert the temperature sensor Pt 500 4 in the
temperature sensor pocket 5 (see appendix).
2
2. Fasten the temperature sensor through the
5
capstan screw 2 and the planned screw
connection 6 .
3. Have the temperature sensor sealed using a
lead-sealing socket 1 and the wire seal 3 in
3 accordance with Ill. 9 by a calibration officer.
Ill.: 9
7 8 1
1. Insert the temperature sensor Pt 500 5 in the
temperature sensor pocket 6 (see appendix).
2
2. Use the adapter for lead-sealing the
connection 2 (see appendix).
6
3. Fasten the temperature sensor through the
3 capstan screw 3 and the planned screw
connection 8.
4 4. Have the temperature senor sealed using a
5
lead-sealing socket 1 and the wire seal 4, 7 by
a calibration officer.
Ill.: 10
1 2 3
1. Connect the pressure connection 1 using
the union nut and the seal insert to the
connection of the internal pressure
sensor 4 .
2. Seal the connection with the sealing
socket 2 and a wire seal 3 .
Ill.: 11 4
Various instruments can be connected to the digital output of the EK205 To this
end, the outputs are preset (see chapter 5.3.1.10)
5.3 Commissioning
Ill.: 13
5.3.1.2 Setting parameters for the NF pulse transducer of the gas meter
If a low-frequency pulse transducer is connected in accordance with 5.2.1.1, you can set the
input mode and cp value as follows:
1. Setting the input mode:
Move the cursor to the "Serv" tab and via the following path to the value "Md.I1" input
mode:
Serv. Inputs Input 1 Md.I1
Press the key combination ENTER + The value set flashes.
Repeatedly press one of the or arrow buttons until the text "Pulse input" flashes.
Press the ENTER + button to confirm the set value. It is possible to abort the
entry by pressing the key combination ESC + .
2. Setting the cp value (pulse constant):
Move the cursor to cp value "cp.I1" in the same path
Press the key combination ENTER + One digit of the value flashes.
Move the cursor with the arrow buttons or to the digits and change these with the
arrow buttons or .
Press the ENTER + button to confirm the set value. It is possible to abort the
entry by pressing the key combination ESC + .
5.3.1.3 Setting the volume at measurement conditions
In order to control the recording of volumes within measurement conditions, the EK205's
counter can be set to the same value as the gas counter once when the administrator lock is
open. Setting the volume is always possible when the calibration lock is open.
Open the administrator lock or the calibration lock
Move the cursor to the "Serv" tab and via the following path to the value "VmA"
(adjustable volume at measurement conditions) input mode:
Serv. Volume Actual volume Synchronization Vm VmA
Installation, connection and commissioning 31
Press the key combination ENTER + one digit of the value flashes.
Move the cursor with the arrow buttons or to the digits and change these with the
arrow buttons or .
After all digits have been changed, press the key combination ENTER + in order
to confirm the input. It is possible to abort the entry by pressing the key combination
ESC + .
To adopt the value set, move the cursor to the Store value.
Press the ENTER key combination + "0" flashes.
Change the value using the arrow buttons or to "1".
After the value has been changed, press the key combination ENTER + in order
to confirm the input. It is possible to abort the entry by pressing the key combination
ESC + . For Vm, VmA is adopted.
If the administrator lock is open and the transfer of VmA for Vm is rejected with
the –13– message then this process has already been carried out.
Further changes are possible when either the calibration lock or the
administration lock is open, after deleting the change information regarding
setting Vm at the same value as the gas counter (see chapter 5.3.1.4).
5.3.1.4 Deleting change information for alignment of Vm with the gas meter
To allow for resetting of the volume within measurement conditions when the administrator
lock is open, the change information regarding setting Vm at the same value as the gas
counter must be deleted:
Open the calibration lock
Move the cursor to the "Serv." tab and via the following path to the value "Clear"
(adjustable volume at measurement conditions) input mode:
Serv. Volume Actual volume Synchronization Vm Clear
Press the key combination ENTER + "0" flashes.
Change the value using the arrow buttons or to "1".
After the value has been changed, press the key combination ENTER + in order
to confirm the input. It is possible to abort the entry by pressing the key combination
ESC + . "TimeX" (time stamp of the alignment), "Vm_o" (Vm old value),
"Vm_n" (Vm new value) are set to the default values.
5.3.1.5 Setting volume at base conditions
It is possible to set up the meter when first operating the EK205 provided that the
administrator lock is open. Setting the volume is always possible when the calibration lock is
open.
Open the administrator lock or the calibration lock.
Move the cursor to the "Serv" tab and via the following path to the value "VbA"
(adjustable standard volume meter) input mode:
Serv. Volume Standard volume Synchronization Vb VbA
Press the key combination ENTER + One digit of the value flashes.
Move the cursor with the arrow buttons or to the digits and change these with the
arrow buttons or .
After all digits have been changed, press the key combination ENTER + in order
to confirm the input. It is possible to abort the entry by pressing the key combination
ESC + .
32 Installation, connection and commissioning
To adopt the value set, move the cursor to the Store value.
Press the key combination ENTER + "0" flashes.
Change the value using the arrow buttons or to "1".
After the value has been changed, press the key combination ENTER + in order
to confirm the input. It is possible to abort the entry by pressing the key combination
ESC + . For Vb, VbA is adopted.
If the administrator lock is open and the transfer of VbA for Vb is rejected with the
–13– message then this process has already been carried out.
Further changes are possible when either the calibration lock or the
administration lock is open, after deleting the change information regarding
setting Vb (see chapter 0).
Press the ENTER + button to confirm the set value. It is possible to abort the entry
by pressing the key combination ESC + .
Move the cursor with the arrow keys or further to the gas analysis values (Ho.b,
CO2, etc.).
Change the values in accordance with the gas used by enabling with the ENTER +
key combination the entry, going with the arrow keys or to the figures and changing
them with the keys or .
Press the ENTER + button to confirm the set value. It is possible to abort the entry
by pressing the key combination ESC + .
5.3.1.8 Setting the alarm limits for gas pressure and temperature measurement
Ex works, the alarm thresholds are set to sensible values. If a change is necessary, proceed
as follows:
Move the cursor to the "Serv" tab and via the following paths to the alarm thresholds:
Serv. Measured values Pressure pMin and pMax9
Serv. Measured values Temperature TMin and TMax
Press the key combination ENTER + to enable the entry.
Move the cursor with the arrow buttons or to the digits and change these with the
arrow buttons or .
Press the ENTER + button to confirm the set value. It is possible to abort the entry
by pressing the key combination ESC + .
5.3.1.9 Setting replacement values for gas pressure and temperature
If the gas pressure or the gas temperature cannot be measured due to a fault or if pressure
and temperature are outside the alarm limits defined in chapter 5.3.1.8, an adjustable
replacement value (fixed value) is used to calculate the standard volume. The volume
thereby calculated is counted in the separate fault volume meter VbD (Base Volume
Disturbed).
The total base volume counter VbT is keep on counting although there is a disturbance. The
total base volume calculated under these faulty conditions has no measuring relevance. It
must not be used for billing purposes.
The replacement values can be found at the following paths:
Serv. Measured values Pressure p.F10
Serv. Measured values Temperature T.F
The values are changed as a rule in the same way as described in chapter 5.3.1.8.
The following functions are preset for the outputs as factory settings:
9 These values are not required in the performance as Temperature Conversion device.
10 These values are not required in the performance as Temperature Conversion device.
34 Installation, connection and commissioning
Output Function
DA1 Pulse output for the total standard volume "VbT"11 (CP.O1= 0.1 Imp/m3)
DA2 Pulse output for the total standard volume "VmT" (CP.O2= 0.1 Imp/m3)
The cp value of an output states how many pulses were emitted per cubic meter. A cp value
of 0.1/m3 (0.1 pulses per m3) means that e.g. a pulse is emitted per 10m3.
To change the cp values for output 1 or output 2, move the cursor to the "Serv." tab and via
the following path to "CP.O1" (output 1) or "CP.O2" (output 2):
Serv. Outputs Output 1 CP.O1
or Serv. Outputs Output 2 CP.O2
Press the key combination ENTER + One digit of the cp value flashes.
Move the cursor with the arrow buttons or to the digits and change these with the
arrow buttons or .
Press the key combination ENTER + to confirm the set value. It is possible to abort
the entry by pressing the key combination ESC + .
Except for the settings described here, many additional functions, such as time
synchronous pulses, can be set for the outputs.
The complete description can be found in the application manual EK205, which
you can download from www.elster-instromet.com.
Press the ENTER + button to confirm the set value. It is possible to abort the entry
by pressing the key combination ESC + .
5.3.1.12 Delete measurement archives
Move the cursor to the "Serv." tab and via the following path to the value "Clr.A" (delete
measurement archives):
Serv. Edit and delete Clr.A
All measurement archives (no logbooks) are deleted.
To avoid deleting the archives by mistake, when the calibration lock is open the
serial number (on the type plate of the device) of the EK205 must be stated.
Press the key combination ENTER + One digit of the value flashes.
Move the cursor with the arrow buttons or to the digits and enter with the arrow
buttons or the serial number of the device.
After stating the serial number of the device, press the key combination ENTER + in
order to confirm the input. It is possible to abort the entry by pressing the key combination
ESC + .
5.3.1.13 Deleting the certification data log
Deleting the log is only possible when the calibration lock is open.
Move the cursor to the "Serv." tab and via the following path to the value "ClCDL" (delete
certification data log):
Serv. Edit and delete ClCDL
Press the key combination ENTER + "0" flashes.
Change the value using the arrow buttons or to "1".
After the value has been entered, press the key combination ENTER + in order to
confirm the input. It is possible to abort the entry by pressing the key combination
ESC + .
5.3.1.14 Close and secure calibration key
After making all settings subject to calibration requirements, close the calibration lock. Press
the key described in chapter 5.3.1.1 again.
The calibration lock is calibrated with an adhesive label as described in chapter 5.3.2.2.
36 Installation, connection and commissioning
5.3.2 Sealing
5.3.2.1 External view
1 1 Optional user protection: Lid sealing
using wiring seals through sealing
plugs.
1
Ill.: 14
Ill.: 16
The sealing of the sealing points Ill.: 15/3 for securing the pulse inputs and pulse
outputs is subject to national law (see WELMEC 11.1, chapter 2.7.1).
Depending on the legal situation in the country of use, seals from a group of
people set out in law or from meter operators must be used.
If devices are delivered with connected input and/or output cables then seals
with the manufacturer's symbol are applied at the plant. If necessary, these can
be substituted on site as described above.
Installation, connection and commissioning 37
5.3.2.3 Sensors
The sealing of the connected temperature and pressure sensors is illustrated in
chapters 5.2.3 and 5.2.4 as an example.
CAUTION!
Damage to property caused by improper device closure!
If the device is not properly closed cables can get pinched and cause damage.
Therefore:
– Ensure correct positioning of the cable guide when closing.
Ill.: 17
5.3.4 Check installation and connection
WARNING!
Risk from improper installation and connection!
Improper installation and connection of the EK205 may result in fatal situations.
Therefore:
– Carry out the installation and connection of the EK205 correctly.
– Observe the requirements of the standards and guidelines stated in the
design test certificates.
6 Operation
Using the enSuite software and the data interfaces of the EK205, further
applications than those described below are possible. Instructions are available
at www.elster-instromet.com.
6.1 Safety
6.2.2 Specialist
The specialist on operating the EK205
has the ability, because of their technical training, knowledge and experience, as well
as their knowledge of the relevant regulations, to carry out the work to the device
assigned to them, and to recognize and avoid possible hazards on their own.
is authorized to read and note values and parameters of the operating elements of the
EK205 and to make changes not subject to calibration requirements.
6.3 Basics
As already explained in the chapter Construction and function, you can operate and program
the EK205 using the operating elements on the front plate of the device:
1 Alphanumerical display
1
2 Arrow key◄
2 3 4 5
3 Arrow key▲
4 Arrow key▼
6 8 5 Arrow key►
6 HOME/CLR key combination
7
7 ENTER key combination
On plate in the housing floor, the calibration lock can be found in the form of a
key that can be sealed with an adhesive label. This needs to be operated to
change calibration-protected values and parameters. In the event of a
calibration-protected device, this may be carried out by a calibration officer only.
6.3.1 Display
The display is divided ex works into the five tabs of Main, Cust., Admin, Serv. and Ctrl., in
which the measurement values, settings and other data are displayed.
3 4 5 6
+ By pressing the two keys at the same time, you can execute the
HOME/CLR function and thereby
display the Main tab.
initiate a value in the input mode (reset to its start value).
+ By pressing the two keys at the same time, via the context menu displayed,
you can
display the Main tab.
freeze the current display.
delete the status tab.
Symbol Meaning
At the top left edge of the display, individual letters are displayed as symbols
for the following messages:
ok. No special messages
A Alarm – collective message "any alarm"
p Pressure alarm – the gas pressure cannot be measured
T Temperature alarm – the gas temperature cannot be measured
W Warning – collective message "any warning"
B Battery warning – the battery is nearly empty
L12 Certification data log - the certification data log is full.
P Calibration lock (programming mode) – the calibration lock is open.
o online – data transmission is ongoing
If one of these letters flashes, the condition thereby indicated is still in place (active). If a
letter is displayed consistently, the indicated status is no longer in place, but needs to be
confirmed.
Detailed explanations on the meaning and on any necessary procedure: see chapter 8.1.4
"Conduct in the event of faults"
12
If a calibration logbook is available in the device.
Operation 43
Symbol Meaning
External power supply
If this symbol appears, the EK205 is being supplied power from an external
unit connected to the terminals.
Signal strength of the radio network for the external modem (connected to
the terminals). This display is only shown if an external modem is connected
and active.
Charging status of the device battery
13
If a calibration logbook is available in the device.
Operation 45
16 These values are not required in the performance as Temperature Conversion Device.
48 Operation
17
In the User values submenu, the user can set via the enSuite parameterization software up to 10 and within this list in another submenu
up to 12 freely programmable parameters.
50 Operation
18 Operate the device with high communication reliability (high level security, see application manual) so that this menu item is not
displayed.
Operation 51
7 Maintenance
7.1 Safety
DANGER!
Risk of life from electric current!
When coming into contact with live parts, there is a immediate risk of life.
Damage to the insulation or individual components can be fatal.
Therefore:
– Cover the electric connections and live parts against possible contact.
– Immediately cut the power supply in the event of damage to the insulation
and initiate the repairs.
– Have work on the electric components of the device, such as the connection
of the external power supply, carried out by electrical experts only.
– For all work on the electrical plant, de-energize it and check the freedom from
voltage.
– Before maintenance work, cut the voltage supply and secure it from
reactivation.
– Do not bridge any fuses or deactivate them. When replacing fuses, observe
the correct ampere figure.
– Keep life parts away from humidity. This may result in a short-circuit.
DANGER!
Risk of explosion due to electrostatic charging.
Avoid electrostatic charges.
Therefor:
– Clean the device only with a damp cloth.
– Avoid rubbing with non-conductive materials.
WARNING!
Risk of injury when handling batteries incorrectly!
Batteries need to be treated with special care.
Therefore:
– Do not throw the batteries in the fire or expose them to high temperatures.
There is a risk of explosion.
– Do not charge the batteries. There is a risk of explosion.
– Liquid being released in the event of incorrect use may result in skin
inflammations. Avoid contact with the liquid. Rinse the liquid with lots of water
in the event of contact. If the liquid enters the eyes, rinse the eyes for 10 mins
with water and see a doctor immediately.
Maintenance 53
CAUTION!
Substances harmful to the environment!
When handling substances harmful to the environment, in particular in the event
of incorrect disposal, severe damage may arise for the environment.
Therefore:
– Observe the below notes at all times.
– Take suitable measures immediately if substances harmful to the
environment enter the environment by mistake. In the event of doubt, inform
the municipal authorities in charge of the damage.
7.1.1 Staff
Maintenance work must be carried out correctly.
Have the maintenance work described here carried out by electrical experts only (see
chapter Operation).
WARNING!
Risk of injury from improperly performed maintenance work.
Improper maintenance may result in severe personal injury or material damage.
Therefore:
– Before starting work, ensure sufficient installation freedom.
– Ensure order and cleanliness at the place of installation. Components and
tools lying around and placed on top of one another loosely are sources of
accidents.
– If components were removed, ensure correct installation, reinstall all fastening
elements and observe screw torques.
DANGER!
Risk of explosion when using of wrong batteries.
Use only batteries stated in the design test certificates of the EK205 (see chapter
9.3). Order number see chapter 9.1.4.
WARNING!
Risk of injury when handling batteries incorrectly!
Batteries need to be treated with special care.
Therefore:
– Do not throw batteries in the fire or expose them to high temperatures. There
is a risk of explosion.
– Do not charge batteries. There is a risk of explosion.
– Liquid being released in the event of incorrect use may result in skin
inflammations. Avoid contact with the liquid. Rinse the liquid with lots of water
in the event of contact. If the liquid enters the eyes, rinse the eyes for 10 mins
with water and see a doctor immediately.
The replacement of the batteries can be carried out without damaging lead seals for devices
with a standard structure.
Before removing the plug of the old battery from the plate, ensure that the new
battery has already been connected. As such, the EK205 continues to work
during the battery replacement.
Maintenance 55
Ensure that the new battery is fastened correctly and sits tight inside the EK205.
CAUTION!
Damage to property caused by improper device closure!
If the device is not properly closed cables can get pinched and cause damage.
Therefore:
– During closing, ensure the correct position of the cable ducts.
To open the administrator lock, move the cursor to the Admin tab and via the following
path to the Cod.A value (administrator key entry):
Admin. Device settings Access Cod.A
Press the key combination ENTER + One digit of the value flashes.
Move the cursor with the arrow buttons or to the digits and change these with the
arrow buttons or .
After the value has been entered, press the key combination ENTER + in order to
confirm the input. It is possible to abort the entry by pressing the key combination ESC
+ .
To enter the battery capacity, move the cursor to the Serv. tab and via the following path to
the Bat.C value (battery capacity):
Serv. Batteries Bat.C
Press the key combination ENTER + One digit of the value flashes.
Move the cursor with the arrow buttons or to the digits and change these with the
arrow buttons or .
When using a battery of the D value, set the Bat.C value to 13.0 Ah.
After the value has been entered, press the key combination ENTER + in order to
confirm the input. It is possible to abort the entry by pressing the key combination ESC
+ .
To close the administrator lock, move the cursor to the Admin tab and via the following
path to the St.AL value (administrator lock status):
Admin. Device settings Access St.AL
Press the key combination ENTER + "open" flashes.
Move the cursor with the arrow keys or to change the value to "closed".
After the value has been entered, press the key combination ENTER + in order to
confirm the input. It is possible to abort the entry by pressing the key combination ESC
+ .
After entering a new battery capacity, in the standard operating case (see
chapter 3.2.1) a value of at least 62 months must be displayed for "Bat.R".
Faults 57
8 Faults
The following chapter describes causes of faults of the device and possibilities of rectification.
In the event of faults that cannot be rectified with the following instructions,
please contact our customer service (see chapter General) or our electronics
hotline:
Tel. +49 (0) 6134/605-123
http://www.elster-instromet.com/de
E-mail: support@elster-instromet.com
8.1 Safety
DANGER!
Risk of life from electric current!
When coming into contact with live parts, there is a immediate risk of life. Damage
to the insulation or individual components can be fatal.
Therefore:
– Immediately cut the power supply in the event of damage to the insulation and
initiate the repairs.
– Have work on the electric components of the device, such as the connection of
the external power supply, carried out by electrical experts only.
– For all work on the electrical plant, de-energize it and check the freedom from
voltage.
– Before maintenance work, cut the voltage supply and secure it from reactivation.
– Keep life parts away from humidity. This may result in a short-circuit.
8.1.1 Staff
The work described here for fault rectification, unless indicated otherwise, can be
carried out by the operator.
Some work may only be carried out by specially trained specialist staff or exclusively
by the manufacturer. Special reference to this is made in the description of the
individual faults.
Work on the electrical system may be carried out by electrical specialist staff only as a
rule.
WARNING!
Risk of injury from improper fault rectification!
Improper fault rectification may result in severe personal injury or material
damage.
Therefore:
– Before starting work, ensure sufficient installation freedom.
– Ensure order and cleanliness at the place of installation. Components and
tools lying around and placed on top of one another loosely are sources of
accidents.
– If components were removed, ensure correct installation, reinstall all
fastening elements and observe screw torques.
C-fact. err. The conversion factor could not be calculated, as the measured gas
temperature is outside -100°C to +100°C or no usable compressibility
figure is available.
Data error An error was detected during the cyclic check of the data subject to
calibration regulations. Please contact the Elster customer service (see
chapter 1.3 Customer service).
Faults 59
I2 Pulse cmp During the pulse comparison between, for instance, input 1 and input 2,
too large a deviation occurred.
W Warn Lim. The performance displayed is outside the warning limit set.
The warning limit can be changed with the enSuite software.
60 Faults
Sett. error The programing of the device has generated an unusable combination of
settings.
Please contact the Elster customer service (see chapter 1.3 Customer
service).
I2 Warn.sig. If an input has been configured as a warning input, this message will be
displayed if an active signal appears at the corresponding terminal (e.g.
terminal DE2).
The input configuration can be adjusted using the enSuite software.
Batt. low The remaining battery power is less than six months.
The batteries need to be changed soon.
c) Remarks:
TA Warn Lim. The ambient temperature measured is outside the warning limit set.
The warning limit can be changed with the enSuite software.
Repair mode The device is in revision mode
Clock n. set The internal clock of the Volume Conversion Device has not been
adjusted at the factory.
CDL full The certification data logbook is full. Changing the parameters marked
with "CDL" without a calibration lock is only possible if the content of the
certification data logbook has been deleted.
U.Logb. full The software update logbook is full. An update is only possible if the
content of the software update logbook has been deleted.
Faults 61
9 Appendix
When ordering the universal holding angle with the clamps for pipe installation,
state the pipe diameter.
9.1.5 Documentation
Replacement / spare parts Order no.
German manual 73 023 158
English manual 73 023 159
Russian manual 73 023 160
64 Appendix