Measurement Procedure - Molspin Magnetometer: North Component Name/No of The Specimen
Measurement Procedure - Molspin Magnetometer: North Component Name/No of The Specimen
Measurement Procedure - Molspin Magnetometer: North Component Name/No of The Specimen
Spin Position
2. Select SER 1 (Serial Port 1) and SHORT (Short Spin). Also keep the SET
GAIN (in the instrument) in maximum position.
3. Calibrate the instrument using the following steps
3.1. Click spin to run an initial spin
3.2. Then enter the value shown on the calibration specimen in the space
given in the program CAL= 1113 and click on CAL=
(For e.g.1113)
1
3.3. Then click spin again and check the north component in the table
(first column in the table). It should show ±10% of 1113.
3.4. If the north component is close to 1113, then the instrument is
Calibrated and is ready for measurements.
Note: While calibrating the instrument, make sure that the attenuator
position is in ‘1’position. Also make sure that the instrument is in RS
232 mode
4. Measurements:
4.1. Click on output File in the Program Window. The following window
will appear. This will help you to save the results automatically in a text file.
2
4.3. Keep the specimen in First Position in the specimen holder, keeping the
v- Notch or slit towards the observer and then click spin
Instrument will take few seconds time for spinning and the result will be
displayed in the first row of the table in the Program window.
4.4. Now click the next button in the program window and observe whether
the spin position is changed or not in the upper right hand side corner in the
program window.
4.5. Now keep the Specimen in Second Position and click spin
4.6. Follow the same for third and fourth positions also by clicking next every
time
4.7. After four spins, click on Sp4= to get the result. The Result will be
displayed in the empty space provided for the result in the Program window.
4.8. Now Click on save to save the result in the output file created in the
beginning (see Step 4.1)
4.9. To measure new specimen click new and follow the steps 4.3 to 4.8
(Note: To be on safer side copy this result in another text file separately
and save it in another location)
4.10. After completing all the measurements, close the program by clicking QUIT
But, before quitting the program close the output file by clicking close file
3
Specimen holder
Set Gain
Attenuator
4
Procedure for IRM measurement
using Molspin magnetometer & Pulsed field magnetometer
Power
A (T)
A (mT)
B
Sample holder Result display Set Field
5
3. Set the required field by pressing “set field” button. (Up to 150 mT, field can
be set by using ‘A (mT) ‘. For higher fields you have to select “A (T)”. That
means if you want to set 300 mT field, you have select ‘A(T), and set field as 0.3
T)
4. Forward field: Keep the specimen in the specimen holder (Coil A) with arrow
on the specimen away from the observer.
6. Once the set field is applied, take out the specimen and measure in the
Molspin magnetometer in Third Position of NRM measurement.
7. Note down the value of the north component shown in the Molspin
Program.
9. Reveres field: Repeat the steps described above for the reversed fields, i.e.
keeping the specimen in the opposite direction to the step 4 in the pulsed field
magnetizer.
Reversed field
10. Continue the experiment till the value of the north component becomes
negative
11. Then stop the experiment and proceed with the new specimen.
6
AMS measurement procedure using KLY-2 instrument
Calibration
2. When it comes to Ready position (i.e. green light on), press Start button and
keep the Calibration sample in the sample holder. Instrument will take few
moments and the result will be displayed on the instrument monitor. It should
show 1167. Note that while calibrating, the Range should be 5.
If this value of 1167 (or nearest value) is seen on the instrument, that
means the instrument is calibrated
7
5. Type the following on the computer monitor
12. Enter Holder Susceptibility (Check it and enter; it will be between 0 and -3)
15. Now start measuring in 15 positions (See instrument manual for 15 directions)
8
AMS measurement procedure using MFK1-FA instrument
F1 F2 F3 F4 F5 F6 F7 F8 F9 F10
BSus Acmd Cal Hol Field AvtVol Help Freq Kill Main
F1 F2 F3 F4 F5 F6 F7 F8 F9 F10
Ax1 Ax2 Ax3 T Sus3 Field AvtVol Help Freq Kill Main
14. Measurement:
15. Press F1 for First position. Instrument will take some time and display the
reading in the following way
9
Ax Range Cosine Sine Error Error %
1
2
3