e
e
e
August 2018
1.14878.0001 7. Preparation
1.14879.0001 • Analyze immediately after sampling.
• The pH must be within the range 2 - 12
Adjust, if necessary, with sulfuric acid.
Spectroquant®
TOC Cell Test TOC 8. Procedure
114878 114879
Pretreated sample 25 ml 1.0 ml Place into a suitable glass
(10 - 30 °C) vessel.
1. Definition Distilled water 1) - 9.0 ml Add and mix.
The term TOC (Total Organic Carbon) is used to describe the total content of Reagent TOC-1K 3 drops 2) 2 drops 2) Add and mix.
organically bound carbon in dissolved and undissolved compounds. The pH must be below 2.5.
Check with pH-indicator
The TOC content is expressed in mg/l. strips.
Adjust the pH, if necessary,
2. Method with sulfuric acid.
By digestion with sulfuric acid and peroxodisulfate, carbon-containing com- Stir for 10 min at medium speed.
pounds are transformed into carbon dioxide. This reacts with an indicator solu-
tion, the color of which is determined photometrically. Stirred sample 3.0 ml 3.0 ml Pipette into a reaction cell.
Inorganically bound carbon (dissolved carbon dioxide and anions of carbon
dioxide) is expelled in gaseous form beforehand by acidification. Reagent TOC-2K 1 level grey 1 level grey Add.
The digestion is analogous to APHA 5310 C (2014). microspoon microspoon
(in the cap of the (in the cap of the
TOC-2K bottle) TOC-2K bottle)
3. Measuring range and number of determinations
Immediately close the cell tightly with an aluminium cap.
Heat the cell, standing on its head, at 120 °C in the preheated thermoreactor for 120
Measuring range Number of
Cat. No. + 5 min.
mg/l TOC determinations
Leave the closed cell, standig on its head, to cool in a test-tube rack for 60 min (cooling
114878 5.0 - 80.0 time).
25 Do not cool with cold water!
114879 50 - 800 After cooling, turn the cell upright and measure in the photometer within 10 min.
For programming data for selected photometers / spectrophotometers see It is recommended to use water for chromatography LiChrosolv®, Cat. No. 115333.
1)