Pod Controller 1008s
Pod Controller 1008s
Pod Controller 1008s
CONTENTS
1. INTRODUCTION................................................................................................. 3
3. SPECIFICATIONS................................................................................................ 7
3.1 Measured input signal ........................................................................................ 7
3.2 Loop Power Supply Specification ..................................................................... 7
3.3 Retransmission Output ....................................................................................... 7
3.4 Contact Input (Digital input) .............................................................................. 7
3.5 Contact Output ................................................................................................... 8
3.6 Display Unit Specification ................................................................................. 8
3.7 Electrical Specification ...................................................................................... 8
3.8 Isolation specification ........................................................................................ 8
3.9 Environmental Specification .............................................................................. 9
3.10 Alarm Specification .......................................................................................... 9
3.11 Display Specification ........................................................................................ 9
3.12 Communication ................................................................................................. 9
3.13 Other Specification ........................................................................................... 9
7. WIRING DIAGRAM.......................................................................................... 14
7.1 Wiring Diagram Of 1008S Standard.................................................................. 14
7.2 Wiring Diagram Of 1008S With Mass Flow ..................................................... 15
User Manual 1
Flow Indicator Totaliser 1008S
REF NO: m18A/om/101
Issue No: 17
9. CALIBRATION PROCEDURE........................................................................ 31
9.1 Input Calibration .............................................................................................. 31
9.2 Output Calibration ........................................................................................... 31
9.2.1 Out zero and Out span................................................................................. 31
11. Miscellaneous..................................................................................................... 37
User Manual 2
Flow Indicator Totaliser 1008S
REF NO: m18A/om/101
Issue No: 17
1. INTRODUCTION
Forew ord
Thank you for purchasing FLOW INDICATOR TOTALISER 1008S. This
manual describes the basic functions and operation methods of 1008S.
Please read through this user‟s manual carefully before using the product.
Overview
This is a microcontroller based Indicator - Totaliser unit, with very high
performance to price ratio. It is highly versatile, accurate and different from the
conventional indicators.
Notice
The contents of this manual are subject to change without notice as a result of
continuing improvements to the instrument‟s performance and functions.
Every effort has been made to ensure accuracy in the preparation of
this manual. Should any errors or omissions come to your attention, however,
please inform MASIBUS Sales office or sales representative. Under no
circumstances may the contents of this manual, in part or in whole, be
transcribed or copied without our permission.
Trademarks
Our product names or brand names mentioned in this manual are the
trademarks or registered trademarks of Masibus Automation and
Instrumentation (P) Ltd. (herein after referred to as MASIBUS).
User Manual 3
Flow Indicator Totaliser 1008S
REF NO: m18A/om/101
Issue No: 17
ORDERING CODE
Digital Communic Retransmis Mass Flow
Model Input Types Power Supply Relay Mounting
Input ation sion O/P Measurement
1008S X X XX X X X XX X
85-265V
C 4-20mA N None A1 N None N NONE N None P0 Panel N None
AC
PT
D 0-20mA Y Yes A3 24V DC 2 RS485 2 2 Relays Y Yes W1 Wall-IP55 PT Compensati
on
Saturated
E 1-5V DC 4 4 Relays FP Wall-FLP SS
Steam Flow
Super
F 0-5V DC SH Heated
Steam
G 0-10V DC
N 0-10 KHz
S SPECIAL*
User Manual 4
Flow Indicator Totaliser 1008S
REF NO: m18A/om/101
Issue No: 17
Note:
o Information in this manual is subject to change without prior notice or
permission due to continuous improvement.
User Manual 5
Flow Indicator Totaliser 1008S
REF NO: m18A/om/101
Issue No: 17
CAUTION:
High voltage transients may occur when switching inductive loads such as
some contactors or solenoid valves. Through the internal contacts, these
transients may introduce disturbances which could affect the performance of
the instrument.
For this type of load it is highly recommended that a “snubber” is connected
across the normally open contact of the relay switching though load. The
snubber recommended consists of a series connected resistor/capacitor
(typically 15nF/100 Ohms). A snubber will also prolong the life of the relay
contacts. A snubber should also be connected across the output of a trip
output to prevent false triggering under line transient conditions.
User Manual 6
Flow Indicator Totaliser 1008S
REF NO: m18A/om/101
Issue No: 17
3. SPECIFICATIONS
Note:
Input type is factory set.
* Max. Pulse level<=24V , Min. pulse level > 1.0V
** ±(0.45% of Reading + 1 Digit)
User Manual 7
Flow Indicator Totaliser 1008S
REF NO: m18A/om/101
Issue No: 17
INPUT TYPE Non- voltage contact input or transistor open collector input
INPUT CONTACT 12VDC,10mA or more (for non – voltage contact input)
CAPACITY
ON/ OFF DETERMINATION For non-voltage contact input
ON = contact resistance of 1KΩ or less,
OFF = contact resistance of 20KΩ or more
For transistor contact input
ON = 2V or less
OFF = leak current of 100µA or less
MINIMUM RETENTION About 1 Second
TIME FOR STATUS
DETECTION
User Manual 8
Flow Indicator Totaliser 1008S
REF NO: m18A/om/101
Issue No: 17
3.12 Communication
PROTOCOL Modbus RTU serial
STANDARD EIA RS-485
MAX. COMMUNICATION DISTANCE 1200 mtrs. (For 9600 bps RS 485)
COMMUNICATION METHOD 2 wire half duplex (RS 485)
DATA FRAME N, 8, 1
COMMUNICATION RATE 9600, 19200 bps
MAX. CONNECTABLE 32
CONTROLLERS/ INDICATOR
ADDRESS RANGE 1 to 99
User Manual 9
Flow Indicator Totaliser 1008S
REF NO: m18A/om/101
Issue No: 17
User Manual 10
Flow Indicator Totaliser 1008S
REF NO: m18A/om/101
Issue No: 17
User Manual 11
Flow Indicator Totaliser 1008S
REF NO: m18A/om/101
Issue No: 17
5. MECHANICAL GUIDELINES
User Manual 12
Flow Indicator Totaliser 1008S
REF NO: m18A/om/101
Issue No: 17
User Manual 13
Flow Indicator Totaliser 1008S
REF NO: m18A/om/101
Issue No: 17
7. WIRING DIAGRAM
7.1 Wiring Diagram Of 1008S Standard
User Manual 14
Flow Indicator Totaliser 1008S
REF NO: m18A/om/101
Issue No: 17
User Manual 15
Flow Indicator Totaliser 1008S
REF NO: m18A/om/101
Issue No: 17
8. OPERATING DETAILS
The following paragraphs give detailed description of how to operate the unit.
Before using the instrument, make sure to study and understand this section.
Parameter‟s Value
Escape sequence
When Esc key is pressed in Menu, the instrument will come in
RUN Mode. If user wants to go in EDIT mode, he will have to
enter the correct password again.
2. When Esc key is pressed in RUN Mode, it will directly enter
in to the set point menu. This function is only applicable when
type of instrument is totaliser.
User Manual 16
Flow Indicator Totaliser 1008S
REF NO: m18A/om/101
Issue No: 17
Batch total:
This is an eight digit totalized value, displayed as Batch total. As per the selected time
base, Zero and Full-scale settings, this total is updated continuously, proportional to
input. When New Batch Starts or Integration total is reset this value also gets
initialized to 0.
Integration total:
This is an eight digit totalized value, displayed as integrated total. As per the selected
time base, Zero and Full-scale settings, this total is updated continuously,
proportional to input.
Integrated total, batch total and roll count are cleared by special password.
On power fail detection, current value of Batch total, integrated total and rollover
count are stored in NVRAM.
User Manual 17
Flow Indicator Totaliser 1008S
REF NO: m18A/om/101
Issue No: 17
START key:
If this key pressed batching relays (WP/EP) gets ON & BATCH TOTAL = 0.It
means a new batch is started. If START key is pressed again and again, it
will not start the batch until WP/EP relays gets OFF.
Relays get OFF by two ways:
1. STOP key
2. BATCH TOTAL crosses the value of SETPOINT.
STOP key:
If user presses STOP key in running batch, batching relays (WP/EP) gets off.
If again STOP key is pressed again and again only batching relays gets
ON/OFF (only if BATCH TOTAL < (WP/EP) values. No new batch will be
started by STOP key.
When batch total value crosses the WP/EP value, relays gets OFF and batch
is over.
Again if START key is pressed then a new batch will start making BATCH
TOTAL = 0. Batching relays gets ON.
Password Protection
When INDEX key is pressed, following prompt will be displayed at lower panel 7-segment
display.
Enter the correct password value with the help of START (digit select key) and STOP key
(Incr key) and press INDEX KEY (INDEX key is used as ENTER key)
If correct password is entered & INDEX key is pressed, it will show the first mode PROG
MOD. If wrong password is entered, then “0000” MSG will be displayed again at lower
display window.
Parameter Description
There are total three Menu Items:
1. Program Mode
2. Configuration Mode
3. Calibration Mode
User Manual 18
Flow Indicator Totaliser 1008S
REF NO: m18A/om/101
Issue No: 17
High-Alarm value should be greater than Full-scale value. . If user tries to set value of
High-alarm greater than Full-scale value, error message will be displayed.
If (High Alarm > full-scale value), “error” message appears.
When we press key in RUN Mode we can directly enter in to Set point
and pre-warn menu. This function is only applicable when type of instrument is
totaliser.
User Manual 19
Flow Indicator Totaliser 1008S
REF NO: m18A/om/101
Issue No: 17
User Manual 20
Flow Indicator Totaliser 1008S
REF NO: m18A/om/101
Issue No: 17
User Manual 21
Flow Indicator Totaliser 1008S
REF NO: m18A/om/101
Issue No: 17
32 Rounding 2 1,2,5,10
33 Over Range If Yes then integration total will be - Yes/No
running according to process value
corresponding to 20ma and if No then
integration total will stop after over limit is
reached.
34 No of Channel To convert 3 channel Mass flow unit to 1 - 1-CH , 3-CH
channel unit.
Batch Mode:
If „Batch Mode’ (Batch nod) is selected to „Counter‟, then two parameters “Batch
count”(Bat cnt) and “No of Batches” (No_of_batch) will be displayed in Configuration
mode.
If „Batch Mode’ is selected to „Normal‟, then two parameters will not be displayed.
Input Type:
Based on requirement, user can select input type. It will be either Voltage or Current or
Pulse.
Instrument Type:
If “instrument type” is set as indicator, then Pre warn and Set point parameter will not be
displayed in Program mode.
Square root:
If user selects this mode as „yes‟, flow rate will calculated using square root algorithm.
Alternatively linear calculation will be done.
Digital Input:
If Digital input is selected as „yes‟ than only digital inputs will work. Four digital inputs are
there.
When any of these inputs is connected to DIN-, it will perform its specific function.
Digital Filter:
If Digital filter is selected as „yes‟ than in programming mode user can see „filter no‟
parameter. If selected ‘no‟ then ’filter no‟ will no be displayed.
Time base: Time base is used for calculation & display of flow rate.
User Manual 22
Flow Indicator Totaliser 1008S
REF NO: m18A/om/101
Issue No: 17
Note: To reset batch counter, Enter in batch count (bat cnt) parameter in configuration
mode. Press “Start” key to see its value. Now if user presses “Stop” key Batch counter
value will be cleared to 0.
Relay-mode:
In „Relay-mode’ (relay nod), if set to „normal „mode then alarm relays and LEDs will work
according to alarm values.
i.e. Relays on, LEDs on
Relays off, LEDs off
But if set to „Failsafe’ Mode then alarm relays and LEDs will operate reversibly.
i.e. Relays on, LEDs off
Relays off, LEDs on
Here we are defining five scale factors for the current input 4 mA to 20mA.
Let’s have one example to understand this concept.
For Example:
ZR = 00000 and FS = 10000, Let us assume the Unit is current input mode.
User Manual 23
Flow Indicator Totaliser 1008S
REF NO: m18A/om/101
Issue No: 17
Input
N Flow Param Display
Value Current
o. rate eter Value
mA
1 FL1 10% SF1 2000 5.6
2 FL2 50% SF2 4000 12.0
3 FL3 60% SF3 4500 13.6
4 FL4 80% SF4 7500 16.8
In above example
Flow rate on 5 FL5 100% SF5 10000 20 Display varies
between:
0 – 2000 for input of 0 % – 10%
2000 – 4000 for input of 10% – 50%
4000 – 4500 for input of 50 % – 60%
4500 – 7500for input of 60 % – 80%
7500– 10000 for input of 80 % – 100%
If user wants a flow to be linear throughout the span than configure this parameter as:
SF1 = FS1, FL1 = 100 %.
User is not required to initialize SF2, SF3, SF4, SF5, FL2, FL3, FL4 and FL5.
Pout Selection:
Pout value decides the pulse Output.
There will be a pulse at pulse output pins (PIN NO.9, PIN NO 10), whenever a integrated
total value is greater than by POUT by previous IT value [where last pulse is registered], it
will be added to the integration total value.
i.e.
Current IT value ≥ (Pout + previous IT value).
Example:
1. If Flow rate = 1000, Pout = 1000, Time base = second,
Every second, 1000 is added in integration total. Since pout is 1000, 1 times of pout value
is added to integration-total every sec. Therefore we will have 1 pulse/sec.
Default Display: This parameter will select the parameter to be displayed in run mode.
If Int tot / Bat tot is selected then in run mode „Integration total „/ „Batch total „will be
displayed on lower panel accordingly.
If “instrument type” parameter is set as indicator, then this parameter will not be displayed
in Configuration mode and Integration total will be displayed in run mode.
User Manual 24
Flow Indicator Totaliser 1008S
REF NO: m18A/om/101
Issue No: 17
For example, if process value is 100 liters/minute & conversion factor is 1 then it will
simply integrate process value as it is.
If we change conversion factor = 2, then it will divide process value by that factor i.e.
100/2 = 50 & it integrates that value (in both integral and batch total) but on the display it
shows original process value.
Time base = 2 = 60
Pulse count = 10000
Then FLOW = [(10000 * 60)/2400]
FLOW = 250 per minute.
Clear total:
Parameter to reset IT, BT, and RC.
If user enters the correct password, IT, BT and Roll over count gets cleared (initialized to
Zero). And „cleared‟ message is displayed on the lower window of display. By pressing
ESC key, menu goes back to the clear total. If wrong password is entered, „cleared‟
message will not appear.
Rounding:
This parameter will round the figure of flow rate. It will increase/decrease the flow rate in
multiple of rounding value. E.g. if rounding = 5 selected then flow rate will
increase/decrease by 5, 10, 15….
User Manual 25
Flow Indicator Totaliser 1008S
REF NO: m18A/om/101
Issue No: 17
Over Range:
If this parameter is yes then integration total will be adding according to process value
corresponding to 20ma and if this parameter is no then integration total will stop adding
after over limit is reached.
Pulse Time:
This parameter is used for averaging of input pulse in low frequency mode (Frequency <
300 Hz).
Frequency Selection:
For frequency below 300Hz select frequency type low.
Pulse Width:
This parameter is used for set the input pulse width in low frequency mode (Frequency <
300 Hz).
No of Channel:
This parameter is used to convert mass flow unit to single channel unit. When we convert
mass flow unit to single channel value of P2 = 0 and P3 = 20 set, and P1 = single channel
flow. Value of P2 and P3 are not changeable.
8.4.3.1 Calibration Mode Parameters Details This menu allows user to perform
calibration of analog input & output.
No Setting Description No of Lo limit Hi limit
digit
1 Cal Zero Zero Cal. Count 4
2 Cal Span Span Cal. Count 5
3 Out Zero Zero Cal. Count for 4 0000 4095
Output
4 Out Span Span Cal Count for 5 0000 4095
Output
5 Default If user selects „yes‟, Yes/No
Out then Out zero =
800 and Out span
(4000) is selected.
INDICATION:
For conversion in engineering value, the input is scaled between Zero and Full scale
set values as per following formula (for linear mode):
For any type of input:
If input signal is outside the set Zero and Full-scale limit, all the four digits of the
Process variable starts flashing. Only when input signal comes back into the allowed
range, display becomes steady.
NOTE:
User Manual 26
Flow Indicator Totaliser 1008S
REF NO: m18A/om/101
Issue No: 17
1). When no input is connected to the unit, it will display “open” message in pulse mode,
current mode and voltage mode. (Pulse input 0 to 10KHz, 4-20mA, 1-5V)
2). When input is out of the range the display will blink.
8.5 Parameter Flow
8.5.1 Parameter Flow Of 1008S Standard
How to operate menu is shown below in the form of flow diagram
User Manual 27
Flow Indicator Totaliser 1008S
REF NO: m18A/om/101
Issue No: 17
User Manual 28
Flow Indicator Totaliser 1008S
REF NO: m18A/om/101
Issue No: 17
User Manual 29
Flow Indicator Totaliser 1008S
REF NO: m18A/om/101
Issue No: 17
User Manual 30
Flow Indicator Totaliser 1008S
REF NO: m18A/om/101
Issue No: 17
9. CALIBRATION PROCEDURE
1. Switch on the instrument and allow 15 minutes of warm up time before starting
calibration.
2. Take a standard source. Set it‟s output at desired Zero value. Apply output of the
source to the input terminals of connector pin according to input select.
3. In calibration mode, Press shift /start key to calibrate zero. This will show prompt -
“CAL Zr1”. To enter in “CAL Zr1” mode press shift /start key
4. Lower window displays a count. This is raw count proportional to analog input. Wait
till counts are stable, save this count by pressing „Index‟ key twice.
5. Press ESC key to come out from “CAL Zr1”.
6. Now apply input equal to Full scale/Span (CALS):
7. Set output of the source at desired full-scale value. Apply output of the source to the
input terminals
8. Press index key to go to calibration span. This will show prompt -”CAL SPn1”. To
enter in “CAL SPn1” mode press shift /start key.
9. Lower window displays raw counts count proportional to analog input.
10. Allow the reading to settle and Press „Index‟ Key twice to save the count.
11. This reading will be stored as Span Now the instrument is calibrated.
12. Press ESC key to come out from “CAL SPn1”.
1. In calibration mode, go to out zero calibration mode by pressing index key. This will
show “oUT ZEro” prompt. To enter in “oUT ZEro” mode press shift /start key
2. Lower window displays a count. Using shift /start or inc/stop key to set reading (zero
output) of connected ampere meter at output, exactly at 4.000mA and press index
key to store the calibration value.
3. Press ESC key to come out from “oUT Zero”.
4. Press index key to go to out span calibration. This will show prompt- “oUT SPAn”. To
enter in “oUT SPAn” mode press shift /start key
5. Lower window displays a count. Using shift /start or inc/stop key to set reading (span
output) of connected ampere meter at output, exactly at 20.000mA and press index
key to store the calibration value.
6. Press ESC key to come out from “oUT SPan”.
This reading will be stored. Now the instrument is calibrated.
If user enters values greater than 4095 “error” message will appear.
It user enters out zero value > out span value then also “error” message will appear.
User Manual 31
Flow Indicator Totaliser 1008S
REF NO: m18A/om/101
Issue No: 17
10.1 Introduction
The unit can be connected in RS-485 communication data link either in multi drop or repeat
mode. Each unit must have unique Serial Number. Entire range of addresses (1 to 247) may
be used. Before starting any communication, choose a baud rate compatible to the host
computer. The serial protocol used is MODBUS RTU.
The error checking field contains a 16-bit value implemented as two eight-bit
bytes. The error check value is the result of a Cyclical Redundancy Check (CRC)
calculation performed on the message contents.
Note:
For Analog Read:
FLx (FL1, FL2, FL3, FL4, and FL5) consists of 9 selectable values from 10 to 100.
For 10 it will read…. 0
For 20 it will read…. 1
If FL1 =100, Hex address 0x0024 will show 9.
User Manual 32
Flow Indicator Totaliser 1008S
REF NO: m18A/om/101
Issue No: 17
User Manual 33
Flow Indicator Totaliser 1008S
REF NO: m18A/om/101
Issue No: 17
User Manual 34
Flow Indicator Totaliser 1008S
REF NO: m18A/om/101
Issue No: 17
User Manual 35
Flow Indicator Totaliser 1008S
REF NO: m18A/om/101
Issue No: 17
User Manual 36
Flow Indicator Totaliser 1008S
REF NO: m18A/om/101
Issue No: 17
11. MISCELLANEOUS
For 4-20mA:-
NOTE:- For OPEN Sensor Condition Pre-warn and batch total (Set Point) Relays are in OFF
State.
For 0-20mA:-
NOTE:- For OPEN Sensor Condition Pre-warn and batch total (Set Point) Relays are not
affected because during open sensor flow rate will be zero.
User Manual 37
Flow Indicator Totaliser 1008S
REF NO: m18A/om/101
Issue No: 17
12. TROUBLESHOOTING
If the operating display does not appear after turning on the controller‟s power, follow the
measures in the procedure below.
User Manual 38