VERTICAL LAMINAR FLOW

MODELS INCLUDE:

VLFT1000
VLFT1200
VLFT1500
VLFT1800

WITH VISIONAIRE TOUCH SCREEN CONTROLLER

Thermoline Scientific Equipment Pty. Ltd.


T/A Thermoline Scientific ABN 80 000 859 129
10-12 Ross Place. Wetherill Park. N.S.W. 2164. Australia.
P.O. Box 6862, Wetherill Park Delivery Centre, N.S.W. 1851. Australia.
Phone: (02) 9604 3911. International: 61 2 9604 3911.
Fax (02) 9725 1706. International: 61 2 9725 1706.
Email: sales@thermoline.com.au Web: www.thermoline.com.au
2
CONTENTS
SECTION 1 ....................................................................... 6
DESCRIPTION OF THE CABINET .............................................................. 6
SECTION 2 ....................................................................... 7
INSTALLATION............................................................................................7
GENERAL .......................................................................................................... 7
TESTING / COMMISSIONING............................................................................ 7
SECTION 3 ....................................................................... 8
GENERAL OPERATION ..............................................................................8
CONTROL SYSTEM .......................................................................................... 9
START-UP ......................................................................................................... 9
LAMINAR AIRFLOW........................................................................................ 10
HEPA FILTER PRESSURE.............................................................................. 10
CONTROL BUTTONS...................................................................................... 11
SUPERVISOR SETTINGS................................................................................ 13
UV STERILISATION (if fitted) ...................................................................16
SECTION 4 ..................................................................... 18
MAINTENANCE ......................................................................................... 18
FUSES ............................................................................................................. 18
LED LIGHTING ................................................................................................ 18
PRE-FILTER(S)................................................................................................ 18
MAIN HEPA FILTER ........................................................................................ 18
SECTION 5 ..................................................................... 19
SERVICING.................................................................................................19

3
Warning

This cabinet must be used in compliance with these instructions and


any repairs or maintenance carried out by qualified personnel.

For parts or service information please contact Thermoline Scientific on:


(02) 9604 3911

4
5
SECTION 1

DESCRIPTION OF THE CABINET


The Circulaire range of Vertical Laminar Flow cabinets provide Class 100
clean air at a velocity >0.35m/sec. over the entire working area. The cabinets
are supplied in 4 sizes to suit most applications. All cabinets are fitted with an
airflow monitor which provides visual and audible warning should the airflow
drop below the minimum level. A double electric socket is provided on the
side of the cabinet for operator use.


          

 
 
 
 
External 1000mmWide 1200mmWide 1500mmWide 1800mmWide
740mmDeep 740mmDeep 740mmDeep 740mmDeep
Dimensions 1250mm High 1250mm High 1250mm High 1250mm High
Internal 985mmWide 1185mmWide 1485mmWide 1785mmWide
645mmDeep 645mmDeep 645mmDeep 645mmDeep
Dimensions 700mm High 700mm High 700mm High 700mm High

6
SECTION 2

INSTALLATION

GENERAL

The cabinet is supplied fully assembled and ready for use.


The following guidelines should be observed when installing the
cabinet:
 Site the cabinet in a draught free position with a minimum of 200mm from
the top of the cabinet to the ceiling to prevent obstructing the air inlet and
to provide access to change the pre-filter.
 Connect the cabinet to a 10A socket.

TESTING / COMMISSIONING

A test certificate will be supplied for conformity to CE marking, and electrical


test.
The airflow should be checked using a vane anemometer and the results
recorded.
The main HEPA (U15) filter will have been factory tested before delivery. A
DOP filter challenge test should be carried out to verify filter integrity when the
cabinet is installed.

THE CABINET SHOULD BE TESTED EVERY 12 MONTHS.

7
SECTION 3
GENERAL OPERATION

The main on/off switch is located on the right hand side near the top of the
cabinet. When first turned on the LCD control panel will be displayed
providing operator instructions for cabinet use. The cabinet can be put into
standby mode by pressing the key. When in standby mode the fan, lights
and control system are turned off. Normal operation can be resumed by
touching the screen. An ECO mode can be activated (see page 13) which
turns off the lights if no activity is seen inside the cabinet for a time period set
in the supervisor settings (see page 13). Normal operation is resumed by
touching the screen. If the cabinet is not to be used for several days the
cabinet can be turned off at the main on/off switch if preferred.

8
CONTROL SYSTEM
The cabinet is controlled and monitored by a microprocessor with an LCD
touch screen. The touch screen provides the operator with general
information about the cabinet and displays the current laminar velocity and
differential pressure across the HEPA filter. Control of light level, fan speed
and other operating settings are all accessed via a menu system. The control
system is factory set to maintain the following parameters:
Normal airflow velocity: 0.40m/sec.
Low airflow alarm: 0.30m/sec.
These settings can be changed by a Monmouth engineer. Please contact
Monmouth Scientific for further advice.

START-UP
When first turned on the screen below is displayed

9
LAMINAR AIRFLOW
The microprocessor controls the fan speed to compensate for filter blockage.
The airflow velocity is continuously monitored and displayed. The value will
fluctuate slightly during normal operation, this is normal and is an indication
that the fan is under microprocessor control. If preferred the units displayed
can be changed to ft/min. in the Supervisor Settings (see page 13)

HEPA FILTER PRESSURE


The cabinet is fitted with a pressure sensor which monitors the pressure
across the HEPA filter. As the filter becomes blocked over time the pressure
will slowly increase. When the pressure reaches the recommended maximum
a warning will be displayed. The HEPA filter is protected by a pre-filter and
should last several years unless working in a very dirty environment or the
pre-filters are not changed on a regular basis.

10
CONTROL BUTTONS

Light On / Off

Reduce Brightness

Increase Brightness

Fan On

Fan Off

Settings - access to settings menu

Information - Cabinet and filter information

Standby - Puts cabinet into standby

UV sterilisation settings (where fitted)

11
SETTINGS
Pressing the key on the main screen enters the Set-Up menu access
screen.

The supervisor access code is supplied with the cabinet and allows access to
change operating preferences.
The engineer access code is reserved for service engineers to carry out
maintenance procedures.
The key returns to the normal display screen.

12
SUPERVISOR SETTINGS
When the correct Supervisor code is entered the screen below is displayed.

Laminar Airflow Display - select preferred velocity units.


ECO Mode - turns off the lights if no activity is seen inside the cabinet for a
time period set by using the arrow keys at the bottom of the screen. Normal
operation will be resumed when operator touches the screen.
Audible Alarm - Turns off the audible low airflow alarm.
Keypad Sounds - Turns off the audible key beeps.
Set Time - displays a time / date setup screen

Pressing the key returns to the previous screen.

13
Reset UV - displays a UV reset screen to reset the date a new UV tube is
fitted and the hours run. This key will only be displayed on models fitted with a
UV sterilisation option.

Pressing the key returns to the previous screen without resetting.

14
INFORMATION SCREEN
Pressing the key on the main screen will display the information screen
below.

Service information is reset by a Monmouth engineer during a service visit.


When a service becomes due a warning screen will be displayed momentarily
on start-up during the proceeding month.

Pressing the key will display the screen below providing information on the
filters fitted and UV tube life. The key will return to the previous screen.

15
UV STERILISATION (if fitted)

Pressing the button on the home screen will display the screen below

The safety screen is interlocked with the UV sterilisation cycle and the cycle
will not run if it is not fitted. Use the arrow keys to set the sterilisation time and
cycle start delay if required.

When the button is pressed the screen below is displayed providing a


countdown time to the end of the cycle.

If the safety screen is removed or the button is pressed the cycle will
terminate and the normal running screen will be displayed

16
If the cycle is allowed to run to completion the screen below will be displayed

17
SECTION 4
MAINTENANCE
The cabinet should be isolated from the electricity supply before
carrying out any maintenance procedures.

FUSES
The main fuses are located in the mains inlet socket on the top of the cabinet.
Remove the mains lead and withdraw the fuses using a small screwdriver.
Always replace with the correct type and rating – 5A Type T. The electrical
socket is protected by 5A Type T fuse located adjacent to the socket outlet.

LED LIGHTING
The high efficiency, low voltage LED light tubes are fitted to the inside of the
enclosure. They should provide many years of service without requiring
replacement. Spare tubes are available from Monmouth Scientific and have
the following part numbers:

Circulaire VLFT1000 - GS-01539


Circulaire VLFT1200 - GS-01539
Circulaire VLFT1500 - GS-01540
Circulaire VLFT1800 - GS-01540

Care should be taken to ensure the tubes are fitted in the correct orientation
with the red+ marking on the tube aligning with red mark on the fitting.

PRE-FILTER(S)
This is located on top of the cabinet and can be changed by removing the two
thumbscrews on the securing frame. Monmouth Pre-Filter part No: K-PF0022

MAIN HEPA FILTER


 Open the control panel by removing the 2 screws inside the lower
cabinet.
 Remove the two filter clamp bolts.
 Lift and hold open the head assembly using the prop at the right hand
side.
 The HEPA filter may now be withdrawn from the front of the cabinet
and a replacement fitted.
Part Nos:
VLFT1000 - K-HF0135
VLFT1200 - K-HF0136
VLFT1500 - K-HF0137
VLFT1800 - K-HF0138

 The replacement filter should be DOP tested prior to use and the
airflow and alarm re-calibrated if necessary.

18
SECTION 5

SERVICING
An annual service is recommended to maintain optimum operating conditions
and will include the following points

 Check / replace pre-filter


 DOP test the main HEPA filter
 Check airflow monitor and re-calibrate if necessary
 Check and record downflow velocity readings
 Check general condition of cabinet - glazing, hinges etc.
 Inspect electrical components, lighting, cables etc.
 Issue test report and airflow certificate.

For parts or service information:


Please contact Thermoline Scientific on: (02) 9604 3911