IC 211 Lab Manual 2018-2019
IC 211 Lab Manual 2018-2019
IC 211 Lab Manual 2018-2019
List of Experiments
M.1 Measurement of Pressure using U-tube manometer, inclined manometer and Dead
weight pressure tester.
a. Comparison of the response of the RTD, thermocouple sensor, Gas pressure
thermometer, bimetallic thermometer and Mercury thermometer with the phase
transition temperatures of water.
b. Calibration of NTC temperature measuring instrument with Pt100 temperature
measuring instrument.
c. Calibration of thermocouple using RTD.
d. Comparison of the Response time for Pt100 and Thermocouple.
Introduction
Post Office Box Trainer is the combination of resistances which works on the principle of
Wheatstone bridge. Post Office Box is helpful to determine the unknown resistance of wire like
manganin, constantan, and Nichrome etc. With the help of this trainer we can determine the
specific resistance of different types of wire. Post Office Box can also verify combination of
resistance in series and parallel.
Post Office Box refers to a box containing a combination of ten resistors of 1 Ω, 10 Ω, 100 Ω and
1000 Ω. They are so connected that when Post Office Box is connected in series in a circuit, it is
possible to introduce any resistance of integer value from 1 Ω to 10000 Ω by pulling out
appropriate keys of the resistors. Post Office Box arranged in the form of a Wheatstone bridge
which is used to find the value of an unknown resistance.
Post office box is a Wheatstone bridge, firstly designed to locate a short circuit in a telephone or
telegraph line by measuring the resistance of the wire to the short circuit. It is called a “Post office
bridge” because this design was adopted by British Post Office, which operated telegraph and
telephone services as well as delivered mail.
Wheatstone was responsible for popularizing the arrangement of four resistors, a battery and a
galvanometer, and gave Christie full credit in his 1843 Bakerian Lecture. Wheatstone called the
circuit a "Differential Resistance Measurer." This arrangement was not invented by Sir Charles
Wheatstone - although it bears his name and is commonly attributed to him, and was employed by
him in some of his electrical researches - but by S. H. Christie, in 1833.
Rx R3
A G D
R4
R2
Figure 1
When the Galvanometer will indicate a current flowing between terminals B and C, it is called
an unbalanced condition. By adjusting R2, R3 and R4, the current flowing through galvanometer
can be set to zero. This is called a balanced condition. By applying circuit laws, one can find an
unknown resistance Rx as
R2
RX R3
R4
One very common application in industry today is to monitor sensor devices such as strain
gauges. Such devices change their internal resistance according to the specific level of strain (or
pressure, temperature, etc.), and serve as the unknown resistor R. Second application is in the area
of electrical power distribution to accurately locate breaks in a power line. The method is fast and
accurate, and does not require a large number of field technicians. A number of resistance
measuring devices have been devised on the principle of Wheatstone bridge.
Experiment 1
Procedure:
1. Connect the -ve terminal of DC Supply to I and +ve terminal to J terminal of the bridge.
2. Connect Galvanometer's +ve and -ve terminals to M and N terminals respectively.
3. Connect K & L terminals of variable resistance to E & G (R3) terminals of bridge.
4. Connect O & P terminals of unknown resistance to H & F (Rx) terminals of bridge.
5. Set all rotaries to 0 Ω before starting the experiment.
6. Keep the key K1 in 'Off' position.
7. Switch 'On' the trainer.
8. Set the resistance Rx at any unknown resistance by rotating its dial knob.
9. Set resistance R1 and R2 at 10 Ω.
10. Switch 'On' the key K1 and observe the deflection on galvanometer.
11. Then adjust the valve of R3 in step of 1Ω, 10 Ω,100 Ω, 1000 Ω, as per the requirement
beginning from zero , till the null point is not obtained in the galvanometer For R 3
resistance of X1, X10, X100, X1000 Ω is provided. For Example, Suppose galvanometer
shows deflection in left direction at 1000 Ω and at200 Ω in right direction then the null
point lies between these two resistance then adjusting resistances of 100 Ω,10 Ω, and 1
Ω, to get the null point. Then add the value of all resistances which you have selected,
this will give R3.
12. Note the value of R3 in the given Observation table.
13. Repeat the same procedure for 1/10 ratio by taking out R1=10 Ω and R2= 100 Ω, and
adjust the resistance R3 till the null point is not obtained. If the null point is not obtained
then repeat the same step no. 10. Note the value of R3 in the given Observation table.
14. Repeat the same procedure for 1\100 ratio by taking out R1= 10 Ω and R2= 1000 Ω.
15. Unknown resistance Rx can be determined by using Wheatstone Bridge relation
R1/R2= R3/Rx Where R1 & R2 are the resistance of arm AB and BC respectively, R3 is
Variable resistance.
16. Tabulate all the readings of Rx in the given Observation table.
17. Take the mean of Rx.
6
18. We get the value of unknown resistance Rx (which is the experimental value) Note : The
actual value of unknown resistance can be determined with the help of dial knob. Here
the dial knob can rotates 10 times corresponding to 0 to 10 numbers, it is said to be Main
Scale reading and its each rotation has 50 divisions, it is said to be Circular Scale reading
(least count =20 Ω). For example , If there are 5 rotations by main Scale it means the
resistance on Main Scale is 5000 Ω and Circular Scale is at 30, it means no. of divisions
are 15 then the actual of resistances becomes 5000 Ω +(15×20 Ω)= 5300 Ω.
1. 10 10
2. 10 100
3. 10 1000
7
Experiment 2
1. Connect the -ve terminal of DC Supply to I and +ve terminal to J terminal of the bridge.
2. Connect Galvanometer's +ve and -ve terminals to M and N terminals respectively.
3. Connect K & L terminals of variable resistance to E & G (R3) terminals of bridge.
4. Take Experimental wire and connect it between the terminals H & F (Rx) whose
resistivity we have to determine (Wire material like manganin, Nichrome etc). Note: Set
all rotaries all 0 Ω before starting the experiment.
5. Keep the key K1 in 'Off' Position.
6. Switch 'On' the trainer.
7. Then set resistance R1 and R2 at 10 Ω.
8. Switch 'On' the key K1 and observe the deflection on galvanometer
9. Switch 'On' the key K1 and observe the deflection on galvanometer
10. Note the value of R3 in the given Observation table
11. Repeat the same procedure for 10/1 ratio by taking out R1=100 Ω and R2= 10 Ω, and
adjust the resistance R3 till the null point is not obtained. If the null point is not obtained
then repeat the same step no. 9. Now note the value of R3 in the given Observation table.
12. Repeat the same procedure for 100\1 ratio by taking out R1= 1000 Ω and R2= 10 Ω.
13. Resistance Rx can be determined by using Wheatstone Bridge relation
R1/R2= R3/Rx Where R1 & R2 are the resistance of arm AB and BC respectively, R3 is
Variable resistance at which point we determined null point.
14. Tabulate all the readings of R= in the given Observation table.
15. Take the mean of Rx. We get the Value of Resistance Rx of the material of wire.
16. Specific resistance of the material of wire can be determine by using formula
r2
RX
l
17. The resistivity of the wire is ρ = ----- Ω-m
8
Observation Table:
2. 10 100
3. 10 1000
9
Experiment 3
Procedure:
1. Connect the -ve terminal of DC Supply to I and +ve terminal to J terminal of the
bridge
2. Connect Galvanometer's +ve and -ve terminals to M and N terminals respectively.
3. Connect K & L terminals of variable resistance to E & G (R3) terminals of bridge
4. Take two resistance of 100 Ω as r1and r2 and check with multimeter.
5. Connect K & L terminals of variable resistance to E & G (R3) terminals of bridge
Note: Set all rotaries all 0 Ω before starting the experiment.
6. Then set resistance R1 and R2 at 10 Ω
7. Keep the key K1 in 'Off' Position.
8. Switch 'On' the trainer.
9. Switch 'On' the key K1 and observe the deflection on galvanometer
10. Then adjust the valve of R3 in step of 1Ω, 10 Ω,100 Ω, 1000 Ω, as per the
requirement beginning from zero , till the null point is not obtained in the
galvanometer For R3 resistance of X1, X10, X100, X1000 Ω is provided. For
Example, Suppose galvanometer shows deflection in left direction at 1000 Ω and
at200 Ω in right direction then the null point lies between these two resistance
then adjusting resistances of 100 Ω,10 Ω, and 1 Ω, to get the null point. Then add
the value of all resistances which you have selected, this will give R3.
11. Note the value of R3 in the given Observation table
12. Repeat the same procedure for 1/100 ratio by taking out R1=10 Ω and R2= 100 Ω,
and adjust the resistance R3 till the null point is not obtained. If the null point is
not obtained then repeat the same step no.10. Now note the value of R3 in the
given Observation table
13. Repeat the same procedure for 1\100 ratio by taking out R1= 10 Ω and R2= 1000
Ω.
14. Resistance Rx of the combination of Resistances either in series or in parallel, can
be determined by using Wheatstone Bridge relation:
10
Observation Table
2. 10 100
3. 10 1000
11
Introduction
Kelvin’s bridge is useful for measuring very small values of resistance. Kelvin’s double bridge or
Kelvin’s bridge (as it is commonly known as) is a variation of Wheatstone bridge and is based on
the same principle. Consider the value of resistance is in the magnitude of contact leads. For low
resistance measurement, the resistance of lead and contacts becomes significant and can
introduce an error; this can be eliminated using Kelvin’s bridge. This bridge is a modification
over other DC bridges and provides greatly increased accuracy in measurement of low
resistance. Figure 1 illustrates the basic principle of Kelvin’s bridge.
P Q
p q
R S
m n
Figure 2
Kelvin’s bridge consists of purely resistive elements; its first and third arms have standard
resistance values, while the second arm consists of a variable resistor and the fourth arm has an
unknown resistance to achieve balance condition.
P, Q, p, q = standard resistance
R = variable resistance
S = unknown resistance
The ratio p/q is made equal to ratio P/Q. Under the balance conditions there is no current through
the galvanometer. By using circuit laws,
Since P/Q=p/q
S=R Q/P
This is the practical balance equation for Kelvin’s bridge. It indicates that the resistance of
connecting lead, r has no effect on the measurement; provided that two set of ratio arms
have equal ratio. However, an error will be introduced in case the ratios are not exactly same.
13
Experiment 1
Equipments:
1. Kelvin’s Bridge
2. 2 mm patch cords
3. Digital multimeter
Circuit diagram:
Mimic design used to illustrate the Kelvin's bridge method to measure the value of unknown
resistance is shown in figure 2 below
14
Procedure:
1. Connect a patch cords between sockets '1' and '2' , and in between sockets '9' and '6' to
maintain the 10:1 ratio of bridge.
2. Connect a patch cord between sockets '10' and '11' to measure the unknown resistance
RX1.
3. Connect positive terminal or socket '19' of galvanometer with 2mm patch cord to
socket '17' of the Kelvin's bridge.
4. Connect negative terminal or socket '20' of galvanometer with 2mm patch cord to
socket '18' of the Kelvin's bridge.
6. Rotate the potentiometer R2 till the galvanometer gives the null deflection.
7. Switch 'OFF' the power supply and disconnect the patch cords between sockets '10' and
'11'
8. Take reading of potentiometer R2 between test -points '16' and '15' using digital
multimeter.
R2
RX R3
R1
Where Rx = Rx1.
10. One can even verify the calculated value by measuring the value of Rx1 in between
socket '11' and test-point '14' using digital multimeter.
15
Observation Table
Introduction
Maxwell's bridge is useful device for measuring inductance. By setting the null point we can
evaluate the unknown inductance value. For this purpose null detector with amplifier circuit is
implemented on the trainer board.
Principle of AC Bridge
The AC Bridge is a Wheatstone bridge except that resistance is replaced by impedance Z and DC
source is replaced by AC source as shown in figure 1.
Z1 Z2
E
D
Z3 Z4
Figure 3
When the Detector detects a current between terminals B and C, the bridge is in an unbalanced
condition. By adjusting Z1, Z2, Z3 and Z4 the owing through galvanometer can be set to zero.
This is called a balanced condition. By applying circuit laws, one can derive that
Z1 Z 4 Z 2 Z 3
17
There are two conditions that should get satisfy for balanced bridge that both magnitude as well
as phase should be equal on both right and left sides.
This bridge is the simplest method of comparing two inductance values and to determine the
values of unknown inductance. Figure 2 shows the basic circuit for Maxwell’s inductance bridge
circuit configuration.
A
Rx R3
Lx
E
D
R2
L2 R4
B
Figure 4
By applying the AC bridge balance equations one gets
R3
L1 L2
R4
R3
R1 R2
R4
Rx R3
Lx
E
D
R4
R2
C4
B
Figure 5
By applying the AC bridge balance equation one gets
R3
L1 R2 R3C4 and R1 R2
R4
L1
Q C4 R4
R1
Advantages:
2. Useful for measurement of wide range of inductance at power and audio frequencies.
The limitation of this bridge is that it can measure inductance of only medium Q coils (1<Q<10).
19
Experiment 1
Equipments:
2. Digital Multimeter
Circuit diagram:
Mimic illustrate with connections for Maxwell's inductance bridge method for measurement of
unknown inductance is shown in the figure 4 below
Figure 6
20
Procedure:
1. Connect a patch cord between socket '13' of vin terminals of Maxwell's inductance
bridge and socket '29' of Vout terminals of the 1 KHz sine wave generator.
2. Connect a patch cord between socket '14' of vin terminals of Maxwell's inductance
bridge and socket '30' of Vout terminals of the 1 KHz sine wave generator.
3. Connect a patch cord between socket '1' and '2' and connect another patch cord between
Sockets '8' and '11' to determine the value of Lx1 and Rx1.
4. Connect a patch cord between socket '15' and '17' and sockets '16' and '18' for the
purpose of null detection.
5. Switch 'ON' the power supply and the Null Detector.
6. Set the amplitude or loudness of the audio detector as per your requirement by rotating
amplitude control knob of 1 KHz sine wave generator.
7. Adjust the pot in the minimum position so that the resistance R2 between the arms is
minimum by measuring it with help of multimeter. Rotate the potentiometer R2 to find
a condition where null (or a minimum sound) is generated.
8. Switch 'Off' the power supply and the Null Detector
9. Remove the patch cords between sockets '1' and '2'.
10. Take the reading of resistance R2 between test-points '5' and '6' using a digital
multimeter.
11. Calculate the value of inductance Lx1 and resistance Rx using the formula Lx =
L1.R2/R4 where Lx = Lx1.
Note: Value of R4 and L1 are indicated on the front of the board.
12. Calculate the value of unknown resistance using the formula.
Rx = R2.R3/R4.
16. Repeat the above steps from 5 to 12 and calculate the value of inductors Lx3 and
resistor Rx.
Observation Table
S. No R2 R4 R3 L1 Lx = L1.R2/R4 Rx = R3/R4
Ω Ω Ω µH µH Ω
Calculation:
Lx = L1.R2/R4
Rx = R2.R3/R4.
Note: value of R3, R4 and L1 are indicated on the front of the board.
22
Result:
Experiment 2
Equipment:
2. Multimeter.
Circuit diagram:
Mimic illustrate with connections for Maxwell's inductance capacitance bridge method for
measurement of unknown inductance is shown in the figure 5 below.
Figure 7
24
Procedure:
1. Adjust the pot in the minimum position so that the resistance R2 between the arms is
minimum by measuring it with help of multimeter
2. Connect a patch cord between socket '26' of vin terminals of Maxwell's inductance
bridge and socket '29' of Vout terminals of the 1 KHz sine wave generator.
3. Connect a patch cord between socket '27' of vin terminals of Maxwell's inductance
bridge and socket '30' of Vout terminals of the 1 KHz sine wave generator output.
4. Connect a patch cord between socket '19' and '17' and sockets '20' and '18' for the
purpose of null detection.
5. Connect the unknown inductor Lx4 with internal resistance Rx1 from socket '22' to
arm consisting resistance R5 at socket '21'.
6. Switch 'ON' the power supply and the Null Detector
7. Now vary the resistance R7 with the help of pot till the null position (or the first
minimum sound position) is achieved.
8. Now Switch 'Off' the power supply and remove the patch cord from socket; 22; and
measure the resistance R7 with the help of multimeter.
9. Calculate the value of unknown inductor using the following equation
Where, Lx = Lx4.
Note: Value of R5 and C1 are indicated on the front of the board.
10. Calculate unknown value of internal resistance by using the following equation
Rx = R5.R7/R6 where Rx= Rx1.
Note: Value of R5 is indicated on the front of the board and the value of R6 is
1.122 kΩ.
11. Calculate the value for Q factor by using the following equation.
Q = ωLx/Rx = ω.C1.R6
Note: ω = 2π f, where f = frequency = 1 KHz and π= 3.14
12. Verify the result for calculation of Q factor using both the formulae in the above
step.
13. Connect the unknown inductor Lx5 from socket '23' to arm consisting resistance R5
at socket '21'.
14. Repeat steps from 6 to 12 and calculate the values of Lx5, Rx2, and their
corresponding Q factor.
15. Connect the unknown inductor Lx6 from socket '24' to arm consisting resistance R5
at socket '21'.
16. Repeat steps from 6 to 12 and calculate the values of Lx6, Rx3, and their
corresponding Q factor.
25
Observation Table
Calculation:
Measured value of R7 is............................................ Ω
Now measure the value of Lx by the formula:
Lx = R5.R7.C1
Now measure the value of Rx by the formula:
Note: Value of R5 is indicated on the front of the board and the value of R6 is 1.122 K Ω.
Result:
Introduction
Desauty`s and Schering Bridge is useful for measuring very small value of capacitances. These
are based on the principle of Wheatstone bridge.
DeSauty’s Bridge
The Desauty’s bridge is a direct carry over of the Wheatstone bridge with the DC source replaced
by an AC source. The null detector we will be using also has an amplifier where the gain can be
adjusted. This is connected to Null detector which is used for getting the null point. The bridge is
as shown in the figure 1.
Cx R1
E
D
C2 R2
Figure 8
R2
C X C2 F
R1
27
Schering Bridge
A Schering bridge is a bridge circuit used for measuring an unknown electrical capacitance and its
dissipation factor. The dissipation factor of a capacitor is the ratio of its resistance to its capacitive
reactance. The Schering Bridge is shown in figure 2.
Rx R3
Cx
E
D
R4
C2
C4
Figure 9
At balanced condition, unknown capacitance Cx can be calculated as
R4
C X C2 F
R3
28
Experiment 1
Equipments:
1. Deasuty’s bridge
2. 2 mm patch cords
Procedure:
18. Report above procedure for different value of frequency and different value of unknown
capacitors (i.e. Cx2 and Cx3)
19. Tabulate all the retried data in observation table below.
Observation Table
Capacitor
Unknown Frequency Resistance Resistance C2
S.No
capacitor Frequency R1 ohm R2 ohm µF
f1
1. Cx1 f2
f3
f1
2. Cx2 f2
f3
f1
3. Cx3 f2
f3
Calculation:
Experiment 2
Equipments Needed:
1 Schering bridge
2 2 mm patch cords
Procedure:
1. Connect mains cord to the mains and trainer.
2. Connect terminal 13 to 16 ( for evaluating unknown capacitance Cx4)
3. Turn variable resistance R3 towards anticlockwise direction
4. Connect null detector (i.e. terminal 10 to 12 and 11 to 19)
5. Keep toggle of null detector towards ‘off’ condition
6. Select Frequency Selector for any desired range of frequency.
500 Hz to 1 kHz
1 kHz to 10 kHz
10 Hz to 60 kHz
7. For example 8 kHz frequency, select frequency selector between the range 1 Hz to
-10 kHz
8. Chose any ambient frequency (let it be 10 kHz)
9. use frequency variable knob to set 10 kHz frequency on display screen
10. connect terminal 20 to 17 and 21 to 18
11. Now “on” the mains and trainer switch
12. set toggle of null detector towards ‘on’ condition
13. Vary amplitude variable for enough sound of speaker.
14. Vary resistance R3 towards clockwise direction slowly. ( sound diminishes)
15. Keep varying R3 until you get very low sound or null sound ( Null Condition
further varying R3 in same direction speaker starts sounding
16. finally adjust the value of R3 to get null point.(where sound completely diminishes)
17. Take multimeter and record the value of R3 in the observation table.
31
18. Report above procedure for different value of frequency and different value of
unknown capacitors (i.e. Cx5 and Cx6).
Tabulate all the retried data in observation table below
Observation Table
For Capacitor
Frequency Resistance Resistance C3µF
S.No Unknown
R3 ohm R4ohm
capacitor
f1
1. Cx4 f2
f3
f1
2. Cx5 f2
f3
f1
3. Cx6 f2
f3
Calculation:
Linear variable differential transformers (LVDT) are used to measure displacement. LVDTs
operate on the principle of a transformer. As shown in figure 4, an LVDT consists of a coil
assembly and a core. The coil assembly is typically mounted to a stationary form, while the core
is secured to the object whose position is being measured. The coil assembly consists of three
coils of wire wound on the hollow form. A core of permeable material can slide freely through the
center of the form. The inner coil is the primary, which is excited by an AC source as shown.
Magnetic flux produced by the primary is coupled to the two secondary coils, inducing an AC
voltage in each coil.
Figure 10
The main advantage of the LVDT transducer over other types of displacement transducer is the
high degree of robustness. Because there is no physical contact across the sensing element, there
is no wear and tear in the sensing element. Because the device relies on the coupling of magnetic
flux, an LVDT can have infinite resolution. Therefore the smallest fraction of movement can be
detected by suitable signal conditioning hardware, and the resolution of the transducer is solely
determined by the resolution of the data acquisition system.
LVDT Measurement:
LVDT measures displacement by associating a specific signal value for any given position of the
core. This association of a signal value to a position occurs through electromagnetic coupling of
an AC excitation signal on the primary winding to the core and back to the secondary windings.
The position of the core determines how tightly the signal of the primary coil is coupled to each of
the secondary coils. The two secondary coils are series-opposed, which means wound in series but
in opposite directions. This results in the two signals on each secondary being 180 deg out of
phase. Therefore phase of the output signal determines direction and its amplitude, distance.
Figure 2 depicts a cross-sectional view of an LVDT. The core causes the magnetic field generated
by the primary winding to be coupled to the secondary. When the core is centered perfectly
between both secondary and the primary, as shown, the voltage induced in each secondary is
equal in amplitude and 180 deg out of phase. Thus the LVDT output (for the series-opposed
33
connection shown in his case) is zero because the voltages cancel each other.
Displacing the core to the left (figure 3) causes the first secondary to be more strongly
coupled to the primary than the second secondary. The resulting higher voltage of
the first secondary in relation to the second secondary causes an output voltage
that is in phase with the primary voltage.
Figure 2
Likewise, displacing the core to the right causes the second secondary to be more strongly
coupled to the primary than the first secondary as shown in figure below. The greater voltage of
the second secondary causes an output voltage to be out of phase with the primary voltage.
Figure 11
Likewise, displacing the core to the right causes the second secondary to be more strongly
coupled to the primary than the first secondary as shown in figure below. The greater voltage of
the second secondary causes an output voltage to be out of phase with the primary voltage.
34
Figure 12
To summarize, "The LVDT closely models an ideal zero-order displacement sensor structure at
low frequency, where the output is a direct and linear function of the input. It is a variable-
reluctance device, where a primary center coil establishes a magnetic flux that is coupled through
a center core (mobile armature) to a symmetrically wound secondary coil on either side of the
primary. Thus, by measurement of the voltage amplitude and phase, one can determine the extent
of the core motion and the direction, that is, the displacement." Figure below shows the linearity
of the device within a range of core displacement. Note that the output is not linear as the core
travels near the boundaries of its range. This is because less magnetic flux is coupled to the core
from the primary. However, because LVDTs have excellent repeatability, nonlinearity near the
boundaries of the range of the device can be predicted by a table or polynomial curve-fitting
function, thus extending the range of the device.
Figure 13
Construction:
The physical construction of a typical LVDT consists of a movable core of magnetic material and
three coils comprising the static transformer. One of the three coils is the primary coil and the
other two are secondary coils.
35
Figure 14
LVDTs have certain significant features and benefits, most of which derive from its fundamental
physical principles of operation or from the materials and techniques used in its construction.
1. Friction-Free Operation :
One of the most important features of an LVDT is its friction-free operation. In normal
use, there is no mechanical contact between the LVDT's core and coil assembly, so there
is no rubbing, dragging or other source of friction. This feature is particularly useful in
materials testing, vibration displacement measurements, and high resolution dimensional
gauging systems.
2. Infinite Resolution :
Because there is normally no contact between the LVDTs core and coil structure, no parts
can rub together or wear out. This means that an LVDT features unlimited mechanical
life. This factor is especially important in high reliability applications such as aircraft,
satellites and space vehicles, and nuclear installations. It is also highly desirable in. many
industrial process control and factory automation systems.
36
The internal bore of most LVDTs is open at both ends. In the event of unanticipated over
travel, the core is able to pass completely through the sensor coil assembly without
causing damage. This invulnerability to position input overload makes an LVDT the ideal
sensor for applications like extensometers that are attached to tensile test samples in
destructive materials testing apparatus.
An LVDT responds to motion of the core along the coil's axis, but is generally insensitive
to cross-axis motion of the core or to its radial position. Thus, an LVDT can usually
function without adverse effect in applications involving misaligned or floating
moving members, and in cases where the core doesn't travel in a precisely straight line.
Because the only interaction between an LVDT's core and coil is magnetic coupling, the
coil assembly can be isolated from the core by inserting a non- magnetic tube between the
core and the bore. By doing so, a pressurized fluid can be contained within the tube, in
which the core is free to move, while the coil assembly is unpressurized. This feature is
often utilized in LVDTs used for spool position feedback in hydraulic proportional and/or
servo valves.
7. Environmentally Robust :
The materials and construction techniques used in assembling an LVDT result in a rugged,
durable sensor that is robust to a variety of environmental conditions. Bonding of the
windings is followed by epoxy encapsulation into the case, resulting in superior moisture
and humidity resistance, as well as the capability to take substantial shock loads and high
vibration levels in all axes. And the internal high-permeability magnetic shield minimizes
the effects of external AC fields. Both the case and core are made of corrosion resistant
metals, with the case also acting as a supplemental magnetic shield. And for those
applications where the sensor must withstand exposure to flammable or corrosive vapors
and liquids, or operate in pressurized fluid, the case and coil assembly can be hermetically
sealed using a variety of welding processes. Ordinary LVDTs can operate over a very
wide temperature range, but, if required, they can be produced to operate down to
cryogenic temperatures, or, using special materials, operate at the elevated temperatures
and radiation levels found in many nuclear reactors.
37
The location of an LVDT's intrinsic null point is extremely stable and repeatable, even
over its very wide operating temperature range. This makes an LVDT perform well as a
null position sensor in closed-loop control systems and high performance servo balance
instruments.
The absence of friction during ordinary operation permits an LVDT to respond very fast to
changes in core position. The dynamic response of an LVDT sensor itself is limited only
by the inertial effects of the core's slight mass. More often, the response of an LVDT
sensing system is determined by characteristics of the signal conditioner.
Experiment 1
Procedure:
Sensitivity of real time software is 0.5mm (i.e. on real time software window, readings
will change after every 0.5mm displacement).
10. Compare above results with the observation table.
11. Plot the graph between displacement (mm) indicated by micrometer and Display reading
(mm).The graph will be linear as shown in figure 7.
Figure 7
Observation Table
Experiment 2
Procedure:
2. Plot the graph for the display reading with the proportional change with the displacement
(as done in the previous Experiment). Find the point from which it disobeys the linear
relation.
3. Note the displacement reading of micrometer for the above point and subtract 10 mm from
it. The difference is the linear range of LVDT.
4. The trainer is linear over full scale that is 10 mm, so linear range is 10 mm.
41
Experiment 3
DC output/Displacement: Vdc/mm
Procedure:
4. Measure the differential voltage between Test Point TP6 and TP7 with multi-meter in mV
range.
6. Repeat step 4.
7. Calculate S
42
Experiment 4
Procedure:
9. Adjust vertical position of signals displayed so as negative peak of one touches positive
peak of other as shown in figure below
Figure 8
11. Make micrometer to read 5 mm and observe waveforms at Test Point TP4, TP5.
13. Make micrometer to read 15 mm. observe waveforms at Test Point TP4, TP5.
1. Introduction
Pressure is usually expressed with reference to either absolute zero pressure (a complete
vacuum) or local atmospheric pressure. The absolute pressure is the difference between the value
of pressure and absolute zero pressure. Gauge pressure is the difference between the value of the
absolute pressure and the local atmospheric pressure. When the gauge pressure becomes negative
is called as Vacuum Pressure. The schematic is shown in Fig.1
2. Measurement of Pressure
The direct proportional relation between gauge pressure and the differential height “h”
for a fluid of constant density enables the pressure to be simply visualized in terms of the vertical
height, h = P/ρg. The height “h” is termed as pressure head corresponding to pressure “P”. For a
liquid without a free surface in a closed pipe, the pressure head P/ρg at a point corresponds to the
vertical height above the point to which a free surface would rise, if a small tube of sufficient
44
length and open to atmosphere is connected to the pipe. Such a tube is called a piezometer tube,
and the height h is the measure of the gauge pressure of the fluid in the pipe. If such a piezometer
tube of sufficient length were closed at the top and the space above the liquid surface were a
perfect vacuum, the height of the column would then correspond to the absolute pressure of the
liquid at the base. This principle is used in the well-known mercury barometer to determine the
local atmospheric pressure.
Even if the air is completely absent, a perfect vacuum at the top of the tube is never possible. The
space would be occupied by the mercury vapour and the pressure would equal to the vapour
pressure of mercury at its existing temperature. This almost vacuum condition above the mercury
in the barometer is known as Torricellian vacuum. The pressure at A equal to that at B (Fig. 3)
which is the atmospheric pressure patm since A and B lie on the same horizontal plane. Therefore,
we can write
PB = Patm=Pv+ρgh (1)
The vapour pressure of mercury Pv, can normally be neglected in comparison to Patm. At 200C, Pv
is only 0.16 Patm, where Patm =1.0132 X105 Pa at sea level. Then we get from Eq. (1)
h= Patm/ρg= (1.0132×105)/(13560×9.81)=0.752 m of Hg.
For accuracy, small corrections are necessary to allow for the variation of density with
temperature, the thermal expansion of the scale (usually made of brass), and surface tension
effects. If water was used instead of mercury, the corresponding height of the column would be
about 10.4 m provided that a perfect vacuum could be achieved above the water. However, the
vapour pressure of water at ordinary temperature is appreciable and so the actual height at say,
15°C would be about 180 mm less than this value. Moreover with a tube smaller in diameter than
about 15 mm, surface tension effects become significant.
Manometers are devices in which columns of a suitable liquid are used to measure the
difference in pressure between two points or between a certain point and the atmosphere.
Manometers are generally used for measuring large gauge pressures. They are basically the
modified form of the piezometric tube. A common type manometer is like a transparent "U-tube"
as shown in Figs. 4.
46
Fig. 4: A simple manometer to measure gauge Fig.5:A simple manometer to measure vacuum
pressure pressure
One of the ends is connected to a pipe or a container having a fluid (A) whose pressure is to be
measured while the other end is open to atmosphere. The lower part of the U-tube contains a
liquid immiscible with the fluid A and is of greater density than that of A. This fluid is called the
manometric fluid. The pressures at two points P and Q (Fig. 4) in a horizontal plane within the
continuous expanse of same fluid (the liquid B in this case) must be equal. Then equating the
pressures at P and Q in terms of the heights of the fluids above those points, with the aid of the
fundamental equation of hydrostatics, we have:
Manometers are also frequently used to measure the pressure difference, in course of flow,
across a restriction in a horizontal pipe. The axis of each connecting tube at A and B should be
perpendicular to the direction of flow and also for the edges of the connections to be smooth.
Applying the principle of hydrostatics at P and Q we have:
Where, ρ m is the density of manometric fluid and ρw is the density of the working fluid flowing
through the pipe. We can express the difference of pressure in terms of the difference of heads
(height of the working fluid at equilibrium).
As the pressure difference {P1-P2} reduces, the corresponding value of ‘X’ {Eq. 6} (Difference
of differential height of u-tube manometer) reduces and it becomes difficult to read it. This
situation can be solved by two methods.
1) By using mano-metric fluid of less density in order to reduce (ρm-ρw) (No constraint is there
on use of manometric fluid.)
2) For few working fluids i.e. air, the difference of (ρm-ρw) is always very high even if we
change the manometric fluid from mercury to water. In such cases, it is better to utilize inclined
manometer, which increases the value of Δx by inclined one of the limb of manometer. If the
transparent tube of a manometer, instead of being vertical, is set at an angle θ to the horizontal
(Fig.7), then a pressure difference corresponding to a vertical difference of levels x gives a
movement of the meniscus s = x/sin along the slope.
The present experimental unit consists of U-tube manometer, inclined manometer, manometer
panel and pneumatic quick action and dial gauges shown in the Fig. 8. The tubes are filled with
water to which ink is added as a dye to improve readability. The panel has two dial gauges for
measurement of gauge and vacuum pressure. Both the inclined and U tube manometer can be
connected and combined with each other by using connecting hoses.
Fig. 2
A.U-Tube manometer
1. Check the levels in the manometer and be sure that water is properly filled in the
manometer.
2. Make the required connection to measure pressure through U- tube manometer.
3. Plunge the syringe in the inlet port of the U-tube manometer (be sure that it contains
sufficient amount of air in it).
4. Push the syringe plunger until it reaches to the next small mark (3mm) into the u –tube.
50
5. Take the difference in water column between the left and right limbs of the
manometer.
6. Repeat the procedure 4 and 5 and take at least eight readings.
7. Calculate the pressure from the manometer readings.
8. Plot a curve between the pressure (calculated) and the distance travelled by the syringe
plunger.
B. Inclined manometer
1. Check the levels in the manometer and be sure that water is properly filled in the
manometer.
2. Make the required connection to measure pressure through inclined manometer.
3. Plunge the syringe in the inlet port of the inclined manometer (be sure that it contains
sufficient amount of air in it).
4. Repeat the procedure from A.4- A.8 for inclined manometer.
C. U- tube manometer and inclined manometer in series
1. Connect the right limb of the U tube manometer with the left limb of inclined
manometer by using a hose pipe.
2. Check the levels in the manometer and be sure that water is properly filled in the
manometer.
2. Make the required connection with the pipe to connect the u tube manometer and
inclined manometer
3. Plunge the syringe in the inlet port of the inclined manometer (be sure that it contains
sufficient amount of air in it).
4. Repeat the procedure from A.4- A.8 for the above test set up.
D. Dial gauge (positive pressure)
1. Plunge the syringe in the inlet of positive pressure gauge.
2. Push the syringe plunger until next mark (3mm).
3. Note three reading of pressure for the particular displacement of plunger.
E. Dial Gauge (vacuum pressure)
1. Plunge the syringe in the inlet of vacuum pressure gauge.
2. Push the syringe plunger until next mark (3mm).
51
These Devices operate on the basis of a change in mechanical dimension with a change in
property. Different temperature measuring techniques based on mechanical effect are elaborated
below.
𝑃
𝑇 = 𝑇𝑟𝑒𝑓 ( )
𝑃𝑟𝑒𝑓
𝑐𝑜𝑛𝑠𝑡 𝑣𝑜𝑙
Mercury Thermometer is one of the most common types of temperature measurement devices. It
has a large bulb at the lower portion to hold Mercury, which rises or falls in the capillary tube
due to temperature fluctuations. The capillary tube is having appropriate scale markings to
measure the temperature of the surrounding space.
At the top of the capillary tube another bulb is placed to provide a safety feature in case the
temperature range of the thermometer is inadvertently exceeded.
During operation the bulb of the mercury in thermometer is exposed to the environment whose
temperature is to be measured. An increase in temperature causes the liquid to expand in the bulb
and rise in the capillary thereby indicating the temperature and vice versa.
Alcohol and mercury are the most commonly liquids used for measurement of temperature.
Alcohol has the advantage that it has a higher coefficient of expansion than mercury, but it is
limited to low temperature measurements because it tends to boil away at higher temperatures.
Mercury cannot be used below its freezing point -38.78°F (-37.8°C). The size of the capillary
depends on the size of the sensing bulb, the liquid, and the desired temperature range for the
thermometer.
Mercury-in-glass thermometers are generally applicable up to 600°F (315°C), but their range
may be extended to 1000°F (538°C) by filling the space above the mercury with a gas like
nitrogen, this increases the pressure on the mercury, raises its boiling point and thereby permits
the use of thermometers at higher temperatures.
55
In Bimetallic Thermometer two pieces of metal with different coefficients of thermal expansion
are bonded together to form the device. When the strip is subjected to a temperature higher than
the bonding temperature, it will bend in one direction when it is subjected to lower temperature it
will bend in the other direction. This motion is exploited to sense the temperature variation.
Bimetallic strips are frequently used in simple on-off temperature-control devices (thermostats
).Movement of the strip has sufficient force to trip control switches for various devices .The
bimetallic strip has the advantages of low-cost, negligible maintenance expenses and stable
operation over extended periods of time. Alternate methods of construction can use a coiled strip
to drive a dial indicator (similar to bourdon gauge) for temperatures.
Electrical methods of temperature measurement are very convenient because they furnish a
electrical signal that can be easily detected, amplified. These types of temperature measurement
methods are very useful in online monitoring as well as automatic control of temperature. In
addition, they are usually quite accurate when properly calibrated and compensated.
56
Various methods are employed for construction of resistance thermometers, depending on the
application. In all cases care must be taken to ensure that the resistance wire is free of
mechanical stresses and so mounted that the moisture cannot come in contact with the wire and
influence the measurement.
one construction technique involves winding the platinum on a glass or ceramic bobbin followed
by sealing with molten glass . This technique protects the platinum RTD element but is subject to
stress variations over wide temperature ranges. Stress relief techniques can alleviate the problem
so that the element may be used for temperature measurements within ±0.1°
57
Practical problems which are encountered with RTDs involve lead error and relatively bulky size
which sometimes gives rise to poor transient response and conduction error. These can be used
either as 2 core wire configuration or 4 core wire configuration.
2.2 Thermistor
1 1
𝑅 = 𝑅0 𝑒𝑥𝑝 [𝛽 ( − )]
𝑇 𝑇0
Thermistors are semiconducting ceramics composed of mixtures of several metal oxides. Metal
electrodes or wires are attached to the ceramic material so that the thermistor resistance can be
measured conveniently. The resistance does not change linearly with temperature.
The thermistor is a very sensitive device, and consistent performance 0.01°C may be anticipated
with proper calibration. A rather nice feature of the thermistor is that it may be used for
temperature compensation of electric circuits. This is possible because of the negative
temperature characteristic that it exhibits so that it can be used to counteract the increase in
resistance of a circuit with a temperature increase.
The thermistor is an extremely sensitive device because its resistance changes so rapidly with
temperature. However, it has the disadvantage of highly non-linear behavior. This is not a
particularly severe problem because data acquisition systems can employ computing programs to
provide direct temperature readout from the resistance measurement.
Thermistors can be classified into two types, depending on the change in Resistance with the
Temperature. If the resistance increases with increasing temperature, the device is called as
positive temperature coefficient (PTC) thermistor. If the resistance decreases with increasing
temperature, the device is called as negative temperature coefficient (NTC) thermistor.
Because the resistance of the thermistor is so high the error due to lead resistance is small
compared to that for the RTD.
2.3 Thermocouple
The most common electrical method of temperature measurement uses the thermocouple. when
two dissimilar metals are joined together and the junction is kept at any temperature. An emf will
exist between two free ends which is primarily a function of the junction temperature. This
phenomena is called the Seebeck effect. If the two materials are connected to an external circuit
in such a way that a current is drawn, the emf may be altered slightly owing to phenomena called
Peltier effect. Further if a temperature gradient exists along either or both of the materials, the
junction emf may undergo an additional slight alteration .This is called the Thomson effect. The
Seebeck emf is of prime concern since it is dependent on junction temperature. If the emf
generated at the junction of two dissimilar metals is carefully measured as a function of
temperature, then such a junction may be utilized for the measurement of temperature. The main
problem arises when one attempts to measure the potential. When the two dissimilar materials
are connected to a measuring device, there will be another thermal emf generated at the junction
of the materials and the connecting wires to the voltage measuring instrument. This emf will be
dependent on the temperature of the connection, and provision must be made to take account of
this additional potential.
It may be observed that all thermocouple circuits must involve at least two junctions. If the
temperature of one junction is known, then the temperature of the other junction may be easily
calculated using the thermoelectric properties of the materials. The known temperature is called
the reference temperature.
Thermocouples are used in applications ranging from measurement of room air temperature to
that of a liquid metal bath.
Figure 5: Thermocouple
59
Experiment 1
Comparison of the response of the RTD, thermocouple sensor, Gas pressure Thermometer,
Bimetallic Thermometer and Mercury thermometer with the phase transition temperatures of
water.
Aim:
Studying the response of the different temperature measurement instruments with the phase
transition of water at 0° C with ice water and at 100° C with boiling water
Preparation:
Fill the insulated tank with very small crushed pieces of ice cubes and sufficient amount of water
to form ice water.
Connect the Pt100 and Thermocouple temperature sensors and multimeters to the displays on the
WL202 base unit.
Place a container of water on the laboratory heater. Bring the water (purified RO water) to the
boil and record the measured values.
Procedure:
2) Fill the stainless steel tank with very small crushed pieces of ice cubes and sufficient amount
of water (purified RO water) to form ice water.
3) Using the five instruments record the measured temperature values for 0° C with ice water.
Then measure for the Pt100 (RTD) and Thermocouple the generated voltage through multimeter
(by connecting the multimeter to relevant port and setting it for voltage measurement).
6) Dip the Temperature sensors into the water bath at the same height.
8) Bring the water gradually to the boiling stage and at this stage record the measured values of
boiling point temperature, through sensors display and multimeter simultaneously.
(Similar procedure adopted for ice point temperature measurement through different sensors.
Observation table 1:
0° C ice water
Temperature Quantity Unit Measured Set point Variation
sensor
Pt100 Temperature °C
Voltage mV
Thermocouple Temperature °C
Voltage mV
Gas Pressure Temperature °C
Thermometer
Bimetallic Temperature °C
Thermometer
Mercury Temperature °C
Thermometer
Graph: Draw the graph comparing the response of all five temperature measuring instrument.
Experiment 2
Aim:
Thermistor or any other electrical temperature measuring method, gives the electrical output.
These electrical outputs should be calibrated either through first principle or some other
instrument more accurate than thermocouple.
Assuming RTD to be more accurate, perform the calibration of NTC temperature measuring
instrument.
Preparation:
Connect the Pt100 and NTC temperature sensors and multimeters to the displays on the WL202
base unit.
Ice water and boiling water are required to perform the experiment. The water is added to the
stainless steel tank.
Procedure:
2) Fill the stainless steel tank with suitable amount of water mixed with proper ice to bring down
temperature of water till 100 C.
3) Dip the RTD and NTC both inside the stainless steel tank (tip of both sensors should be
dipped inside the water level at equal height) and start with very slow rate of heating of the
stainless steel tank. Connect multimeter across the thermocouple port to note down generated
voltage.
4) Note down the reading of RTD (0C) and corresponding reading of multimeter (voltage;
connected across NTC) at interval of 50C (refer table 3). Control the heating rate accordingly to
achieve higher temperature.
5) Repeat the experiment for few of the readings in cooling mode by switching of the heater and
allowing the water to cool down.
62
Observation table 2:
Measurement
Pt100 Pt100 NTC NTC Variation Variation
Temp. In °C Voltage In Temp. In °C Voltage In In °C In mV
mV mV
10
15
20
25
30
35
40
45
50
55
60
65
70
75
80
85
90
95
Graph:
1. Study the responses of NTC with temperature variation by drawing the graph between
temperature and corresponding voltage generated.
2. Study the responses of RTD with temperature variation by drawing the graph between
temperature and corresponding voltage generated.
Conclusion:
63
Experiment 3
Thermocouple or any other electrical temperature measuring method, gives the electrical output.
These electrical outputs should be calibrated either through first principle or some other
instrument more accurate than thermocouple.
Preparation:
Connect the Pt100 and Thermocouple temperature sensors and multimeters to the displays on the
WL202 base unit.
Procedure:
2) Fill the stainless steel tank with suitable amount of water mixed with proper ice to bring down
temperature of water till 100 C.
3) Dip the RTD and thermocouple both inside the stainless steel tank (tip of both sensors should
be dipped inside the water level at equal height) and start with very slow rate of heating of the
stainless steel tank. Connect multimeter across the thermocouple port to note down generated
voltage.
4) Note down the reading of RDT (0C) and corresponding reading of multimeter (voltage;
connected across thermocouple) at interval of 50C (refer table 3). Control the heating rate
accordingly to achieve higher temperature.
5) Repeat the experiment for few of the readings in cooling mode by switching of the heater and
allowing the water to cool down.
64
Observation table 3:
Experiment No.4
Aim: Comparing the response time of two different methods measuring same parameter.
Response time of the instrument shows, how early the instrument catches up the steady state
value of the parameter. Scientifically the response time of the instrument is stated as the time for
the system to settle to within ±10% of the steady state value.
Procedure:
This will be open ended experiment. In order to check students understanding, their ability to
perform experiment etc, the procedure of this experiment is kept open ended. Students are
supposed to decide procedure, perform experiment and comment on their experimental result.
Conclusion:
66
1. Piezo Film
The piezo is a flexible component that includes a piezoelectric polymer film that is laminated to
a polyester substrate. The laminated strip contains an added mass at the end weighing 0.078 kg
Piezo Film produces voltage in proportion to compressive or tensile mechanical stress or strain,
making it an ideal dynamic strain gage. It gives them capability to work as highly reliable low-
cost vibration sensor. Piezo Film is also ideally suited for high fidelity (The degree to which an
electronic system accurately reproduces the sound or image of its input signal.) transducers
operating throughout the high audio (>1kHz) and ultrasonic (up to 100MHz) ranges.
Piezo film is a flexible, lightweight, tough engineering plastic available in a wide variety of
thicknesses and large areas. Piezo film has low density and excellent sensitivity, and is
mechanically tough. When extruded into thin film, piezoelectric polymers can be directly
attached to a structure without disturbing its mechanical motion.
Piezo film is well suited to strain sensing applications requiring very wide bandwidth and high
sensitivity. As an actuator, the polymer's low acoustic impedance permits the efficient transfer of
a broadband of energy into air and other gases.
2. Pressure
The pressure transducer on the Mechatronics sensor trainer has a range of 0-30 PSI, a sensitivity
of 0.133 V/PSI, and outputs a voltage between 0.5-4.5 V. thus it has a zero pressure offset of 0.5
V and a full scale span of 4.5 V.
67
Pressure Sensor
A pressure transducer is a transducer that converts pressure into an analog electrical signal.
A pressure sensor usually acts as a transducer, it generates a signal as a function of the pressure
Although there are various types of pressure transducers, one of the most common is the strain
gauge base transducer. Pressure applied to the pressure transducer produces a deflection of the
diaphragm which introduces strain to the gages. The strain will produce an electrical resistance
change proportional to the pressure.
Pressure sensors can also be used to indirectly measure other variables such as fluid/gas flow,
speed, water level, and altitude. Pressure sensors can alternatively be called pressure transducers,
pressure transmitters, pressure senders, pressure indicators and piezometers, manometers, among
other names. These sensors are commonly manufactured out of piezoelectric materials such as
quartz.
3. Sonar
The sonar range finder device on the trainer has a operation measuring range of 6-254 inches.
Sonar Sensor
The term sonar is also used for the equipment used to generate and receive the sound. Sonar
sensors are commonly used in mobile robotics and variety of other applications for distance
68
measurement. The working of sonar sensor is based on sound propagation. In sonar sensor, a
speaker (Transducer) is used to emit a short burst of sound (Ping). The sound wave travels
through the air and reflects off a target back to the Transducer (Echo). By measuring the Time of
Flight between Ping and Echo detection, one can calculate the distance between the target and
transducer. The sonar sensors are also used in submarines in order to detect the location of other
vessels.
The acoustic frequencies used in sonar systems vary from very low (infrasonic) to extremely
high (ultrasonic).
Active sonar uses a sound transmitter and a receiver. When the two are in the same place it is
mono-static operation. When the transmitter and receiver are separated it is bi-static operation.
When more transmitters (or more receivers) are used, again spatially separated, it is multi-static
operation. Most sonar sensors are used mono-statically with the same array often being used for
transmission and reception.
This pulse of sound is generally created electronically using a sonar Projector consisting of a
signal generator, power amplifier and electro-acoustic transducer/array. A beam former is
usually employed to concentrate the acoustic power into a beam, which may be swept to cover
the required search angles.
4. Infrared
The infrared distance measuring unit uses a triangulation method to detect the distance of an
object and has a distance measuring range of 20-150 cm. it outputs a voltage that correlates to the
distance of the target.
69
Optical triangulation position sensors use reflected waves to pinpoint position and displacement.
The source of these waves may be a light emitting diode (LED), infrared (IR) light, or laser.
Most optical triangulation position sensors house the sensor itself with the emitter. The sensor
detects the beam that is reflected from the surface or objects and provides an output that varies
with the distance. Outputs for optical triangulation position sensors include analog current,
analog voltage, and frequency. Digital, serial, and parallel outputs are also available, along with
alarms and changes in state of switches.
Typically, the sensing component for optical triangulation position sensors is either a charged
coupled device (CCD) or a photodiode (PSD). The sampling rate is important, as this
specification may determine the suitability of optical triangulation position sensors for specific
applications. Beam patterns are designated as point, line, or area. To achieve a line pattern, many
point-style sensors are used together in a line. To create an area scan, several optical
triangulation position sensors are used together in a matrix or other shape.
Optical triangulation position sensors have a working range and a reference or standoff distance.
This standoff distance is the mid-point in the working range. By definition, the working range is
either more or less the reference-distance and usually given as a value. For example, a sensor
with a reference of 9” and a working range of 4” will have an effective total range of 8” over
which it can measure. Since the midpoint of that range is 9” from the sensor, the distance to
measure must be between 5” and 13” from the sensor. Optical triangulation position sensors
may use either visible or invisible light beams. Sometimes, a visible beam can make the sensors
easy to mount and aim at a target. Occasionally, a visible beam can be undesirable. For instance,
in a factory or processing centre personnel are working, a visible beam may be distracting and
potentially lead to accidents. Additional considerations when selecting optical triangulation
position sensors are resolution and accuracy. Resolution is the smallest unit of distinction that
optical triangulation position sensors can detect. Accuracy is a percentage of full scale.
70
5. Optical Position
The optical position sensor on the trainer consists of an infrared emitting disilicon photo-
transistor, both mounted side by side. the range of the optical position sensor is 0.25 inches.
Optical Sensor
These hydraulic cylinders are often quite long, ranging up to several meters in length. Many
71
scientific, industrial, military and aerospace applications require precise and accurate
knowledge of the angular orientation of a shaft or other rotating object or the linear position of
a reciprocating object. Angular and linear position sensors are widely used in automatic
control systems as feedback-sensing devices in one or more control loops of the system.
Various linear sensor assemblies are known, which utilize the eddy current principle to
produce an output signal in accordance with movement of a core element with respect to a
coil. Linear position sensing is generally performed using a variety of electrically based
sensor devices. Present linear position sensors include linear voltage differential transducers
(LDVT), variable inductance, variable capacitance and eddy current killed oscillators. LDVT
and variable inductance sensors determine linear position by means of a rod inserted in a coil
or coils of wire. In contrast, eddy current and capacitance based position sensors measure
frequency. The frequency of the signal depends on the position of the target.
A common method for sensing linear position involves using a magnet connected to a
displaceable member such as a rod or piston and measuring the magnetic field from the
magnet. A contacting position sensor requires physical contact between a signal generator and
a sensing element to produce the electrical signal. Contacting position sensors typically
consist of a potentiometer to produce electrical signals that vary as a function of the
component's position. Contacting position sensors are generally accurate and precise.
A non-contacting position sensor (NPS) does not require physical contact between the signal
generator and the sensing element. For example, a non-contacting position sensor may utilize
magnets to generate magnetic fields that vary as a function of position and devices to detect
varying magnetic fields to measure the position of the component to be monitored.
72
6. Push button
The push button analog line goes to +5 V when the button is pressed down, i.e. when the
switch is closed as shown in figure. Optical switches and generate A and B signals. the index
pulse is generated by a pickup sensor
7. Micro Switch
the analog input line connected to the miniature snap action switch is pulled high to +5V, when
switch is in open position and goes down to low when pressed down. the micro switch circuit
depicted in figure
A miniature snap-action switch, also trademarked and frequently known as a micro switch, is an
electric switch that is actuated by very little physical force, through the use of a tipping-point
mechanism, sometimes called an "over-center" mechanism. Switching happens reliably at
specific and repeatable positions of the actuator, which is not necessarily true of other
mechanisms. They are very common due to their low cost and durability, greater than 1 million
cycles and up to 10 million cycles for heavy duty models. This durability is a natural
consequence of the design.
73
Internally a flat metal spring must be bent to activate the switch. Inside is a small curved spring
(at the top, just right of center in the photo). Before assembly, it is more nearly straight. After
assembly, it presses against a stationary part of the mechanism at one end. At its other end, its
force is in a direction to stretch the flat spring (although far too small a force to cause any
practical change of shape at all). Acting at an angle, it presses the contact end of the flat spring
against one or the other fixed contacts. With the mechanism not actuated, the curved spring's
force aids that from the flat spring, to ensure that the normally-closed contacts have sufficient
force. (When the actuator is released, the flat spring's tension overcomes that from the curved
spring.)
When actuated, the flat spring is displaced to close the normally-open contact. As the "pivoting"
ends of the curved spring move into line with the flat spring, the curved spring's effect decreases.
Continuing movement of the actuator causes "over-center" action, in which the curved spring
now aids motion of the flat spring. As the curved spring straightens out slightly, it progressively
exerts more force on the flat spring. When the normally-open contacts close, the curved spring
again ensures sufficient contact force. This action produces a very distinctive clicking sound and
a very crisp feel. When actuator force is removed the flat spring goes back to its original state.
Common applications of micro switches include the door interlock on a microwave oven,
leveling and safety switches in elevators, vending machines, and to detect paper jams or other
faults in photocopiers. Micro switches are commonly used in tamper switches on gate valves on
fire sprinkler systems and other water pipe systems, where it is necessary to know if a valve has
been opened or shut.
The defining feature of micro switches is that a relatively small movement at the actuator button
produces a relatively large movement at the electrical contacts, which occurs at high speed
(regardless of the speed of actuation). Most successful designs also exhibit hysteresis, meaning
that a small reversal of the actuator is insufficient to reverse the contacts; there must be a
significant movement in the opposite direction. Both of these characteristics help to achieve a
clean and reliable interruption to the switched circuit.
74
The first micro switch was invented by Peter McGall in 1932 in Freeport, Illinois. McGall was
an employee of the Burgess Battery Company at the time. In 1937 he started the company
MICRO SWITCH, which still exists as of 2009. The company and the Micro Switch trademark
have been owned by Honeywell Sensing and Control since 1950. The trademark has become a
widely used description for snap-action switches. Companies other than Honeywell now
manufacture miniature snap-action switches.
Micro switches are very widely used; among their applications are appliances, machinery,
industrial controls, vehicles, and many other places for control of electrical circuits. They are
usually rated to carry current in control circuits only, although some switches can be directly
used to control small motors, solenoids, lamps, or other devices. Special low-force versions can
sense coins in vending machines, or with a vane attached, air flow. Micro switches may be
directly operated by a mechanism, or may be packaged as part of a pressure, flow, or temperature
switch, operated by a sensing mechanism such as a Bourdon tube. In these latter applications, the
repeatability of the actuator position when switching happens is essential for long-term accuracy.
A motor driven cam (usually relatively slow-speed) and one or more micro switches form a timer
mechanism. The snap-switch mechanism can be enclosed in a metal housing including actuating
levers, plungers or rollers, forming a limit switch useful for control of machine tools or
electrically-driven machinery.
75
08. LED
Light emitting diode, an electronic device that lights up when electricity is passed through it.
LEDs are usually red. They are good for displaying images because they can be relatively small,
and they do not burn out. However, they require more power than LCDs.
We see LEDs every day. A light-emitting diode (LED) is a semiconductor diode that radiates
light (electroluminescence) when current passes through it in the forward direction. Electrons
move though semiconductor medium and "fall into" other energy levels during their transit of the
p-n junction. When these electrons make a transition to a lower energy level, they give off a
photon of light. This photon may be in the infrared region or just about anywhere across the
visible spectrum up to and into ultraviolet - but we have to pick a color when we make them.
That means we can get an infrared LED to work in our remote controls, and get other ones of
different colors to make indicators or, in large arrays, displays - even color ones.
LEDs usually are made to emit one color of light, though bicolor ones and more can be
fabricated. Intensity is being improved as well. That broadens the applications considerably
(think vehicular tail lights). The device has a pair of leads, and a (usually) plastic body molded
around the leads and the semiconductor crystal itself. That way the light can get out through the
plastic. There is frequently a "round dome" at the top of the LED which serves to focus the light
headed out through it. Remember that it is the material of the semiconductor crystal that
determines the light given off by the device. A normal LED isn't really voltage dependent. When
obtaining the LEDs, a manufacturer or hobbiest specifies the color. Red, green and blue are most
popular, and, though there are other colors, it might be tough(er) to find them. Note that the big
LED color displays use a three-LED pixel, and by driving it (the 3-LED assembly) appropriately,
a wide range of colors can be obtained.
A semiconductor with extra electrons is called N-type material, since it has extra negatively-
charged particles. In N-type material, free electrons move from a negatively-charged area to a
positively charged area.
A semiconductor with extra holes is called P-type material, since it effectively has extra
positively-charged particles. Electrons can jump from hole to hole, moving from a negatively-
charged area to a positively-charged area. As a result, the holes themselves appear to move from
a positively-charged area to a negatively-charged area.
A diode comprises a section of N-type material bonded to a section of P-type material, with
electrodes on each end. This arrangement conducts electricity in only one direction. When no
voltage is applied to the diode, electrons from the N-type material fill holes from the P-type
material along the junction between the layers, forming a depletion zone. In a depletion zone,
the semiconductor material is returned to its original insulating state -- all of the holes are filled,
so there are no free electrons or empty spaces for electrons, and charge can't flow.
The interaction between electrons and holes in this has an interesting side effect -- it generates
light!
The word is used on several levels. In commercial terms (such as "the telecom optical switch
market size") it refers to any piece of circuit switching equipment between fibers. The majority
of installed systems in this category actually use electronic switching between fiber transponders.
Systems that perform this function by physically switching light are often referred to as
"photonic" switches, independent of how the light itself is switched. Away from the world of
telecom systems, an optical switch is the unit that actually switches light between fibers, and a
photonic switch is one that does this by exploiting nonlinear material properties to steer light
(i.e., to switch wavelengths or signals within a given fiber).
An optical switch may operate by mechanical means, such as physically shifting an optical fiber
to drive one or more alternative fibers, or by electro-optic effects, magneto-optic effects, or other
methods. Slow optical switches, such as those using moving fibers, may be used for alternate
routing of an optical transmission path, such as routing around a fault. Fast optical switches, such
as those using electro-optic or magneto-optic effects, may be used to perform logic operations;
also included in this category are the semiconductor optical amplifiers, which are optoelectronic
devices that can be used as optical switches and be integrated with discrete or integrated
microelectronic circuits.
10. Encoder
A rotary encoder, also called a shaft encoder, is an electro-mechanical device that converts the
angular position of a shaft or axle to an analog or digital code, making it an angle transducer.
Rotary encoders are used in many applications that require precise shaft unlimited rotation—
including industrial controls, robotics, special purpose photographic lenses, computer input
devices (such as optomechanical mice and trackballs), and rotating radar platforms. There are
two main types: absolute and incremental (relative).
78
1. MEASURING SYSTEMS
The principle of the incremental measuring system is the scanning of a line pattern on a
glass or plastic disc (see Image 1).
The states of the line pattern transparent or not transparent are converted into electronic
pulses by an opto-electronic unit (e.g. transparent = 5V, not transparent = 0V).
The analysis of the signals is performed in an evaluation unit by counting up or down with
each pulse. The current count is stored in digital form and is instantly available for
evaluation.
However, this method has some serious disadvantages. It is possible that the result is
continuously invalid due to signal glitches, unmeasured impulses or similar problems.
Furthermore, after a loss of the supply voltage it is often necessary to return to a reference
point which can cure complications.
79
For these reasons applications with a high emphasis on precision or applications where it
is complicated or not possible to return to the reference point often use the absolute
measuring system.
The fact that incremental encoders use only two sensors does not compromise their
accuracy. One can find in the market incremental encoders with up to 10,000 counts per
revolution, or more.
There can be an optional third output: reference, which happens once every turn. This is
used when there is the need of an absolute reference, such as positioning systems.
The optical type is used when higher RPMs are encountered or a higher degree of
precision is required.
Incremental encoders are used to track motion and can be used to determine position and
velocity. This can be either linear or rotary motion. Because the direction can be
determined, very accurate measurements can be made.
They employ two outputs called A & B which are called quadrature outputs as they are 90
degrees out of phase.
80
The two output wave forms are 90 degrees out of phase, which is all that the quadrature
term means. These signals are decoded to produce a count up pulse or a countdown pulse.
For decoding in software, the A & B outputs are read by software, either via an interrupt
on any edge or polling, and the above table is used to decode the direction. For example if
the last value was 00 and the current value is 01, the device has moved one half step in the
clockwise direction. The mechanical types would be debounced first by requiring that the
same (valid) value be read a certain number of times before recognizing a state change.
If the encoder is turning too fast, an invalid transition may occur, such as 00->11. There is
no way to know which way the encoder turned; if it was 00->01->11, or 00->10->11.
If the encoder is turning even faster, a backward count may occur. Example: consider the
00->01->11-> 10 transitions (3 steps forward). If the encoder is turning too fast, the
system might read only the 00 and then the 10, which yields a 00-> 10 transitions (1 step
backward).
This same principle is used in ball mice to track whether the mouse is moving to the
81
right/left or forward/backward.
Rotary sensors with a single output are not encoders and cannot sense direction, but can
sense RPM. They are thus called tachometer sensors.
Using this measuring system, every position of the measurement range/angle is identified
by a definite code on a glass or plastic disc. This code is represented on the disc in the
form of light and dark regions within different tracks. This combination relates to an
absolute numerical value. Thus, the position value is always directly available, counters
are not necessary. In addition it is not possible to get continuously invalid values caused
by interferences or loss of the supply voltage. Movements which are done while the
system is turned off are immediately measured after the system is Powered up.
The measuring system consists of a light source, a code disc pivoted in a precision ball
bearing and an opto-electronic scanning device (see Image 3). A LED is used as a light
source which shines through the code disc and onto the screen behind. The tracks on the
code disk are evaluated by an opto-array behind the reticle. With every position another
combination of slashes in the reticle is covered by the dark spots on the code disk and the
light beam on the photo transistor is interrupted. That way the code on the disc is
transformed into electronic signals. Fluctuations in the intensity of the light source are
82
measured by an additional photo transistor and another electronic circuit compensates for
these. After the electronic signals are amplified and converted they are then available for
evaluation.
Single-Turn
Single turn encoders are encoders that specify the absolute position for one turn of the
shaft i.e. for 360°. After one turn the measuring range is completed and starts again from
the beginning.
Multi-Turn
Linear systems normally need more than one turn of a shaft. A single turn encoder is
unsuitable for this type of application because of the additional requirement of the number
of turns. The principle is relatively simple: Several single turn encoders are connected
using a reduction gear (see Image 4). The first stage supplies the resolution per turn, the
stages behind supply the number of turns.
The total number of steps in this example is 16 x 16 x 16 x 8192 steps = 33,554,432 steps
Binary 111,1111,1111,1111111111111)
9. Rotary potentiometer
The rotary potentiometer outputs a voltage that varies linearly with the angle being measured. the
potentiometer has a mechanical limit of 300 degree.
The pot as we know it was originally known as a rheostat - essentially a variable wire wound
resistor. The array of different types is now quite astonishing, and it can be very difficult for the
beginner (in particular) to work out which type is suitable for a given task. The fact that quite a
few different pot types can all be used for the same task makes the job that much harder -
freedom of choice is at best confusing when you don't know what the choices actually are, or
why you should make them. This article is not about to cover every aspect of pots, but is an
introduction to the subject. For anyone wanting to know more, visit manufacturers' web sites,
and have a look at the specifications and available types.
The very first variable resistors were either a block of carbon (or some other resistive material)
with a sliding contact, or a box full of carbon granules, with a threaded screw to compress the
granules. More compression leads to lower resistance, and vice versa. These are rare in modern
equipment, so we shall limit ourselves to the more common types.
84
The linear magnetic fields transducer on the sensor trainer outputs a voltage that is proportional
to the magnetic field that is applied perpendicularly to the object being measured. The
relationship however, between the output voltage and the target distance is exponential.
A magnetic field transducer outputs a voltage proportional to the magnetic field that is applied to
the target. The magnetic field sensor is the chip located on the bottom of the trainer. It applies a
magnetic field perpendicular to the flat screw head. the position of the screw head is changed by
rotating the knob. This magnetic field transducer has a similar range to the optical position
sensor.
85
86
Experiment 1
Hint: You can use the cursor to take measurements off the graph.
87
V
Sensor Measurement: at -1.0 cm
V
Sensor Measurement: at -1.0 cm
V
Sensor Measurement: at -1.0 cm
V
Sensor Measurement: at -1.0 cm
Conclusion -
88
Experiment 2
Pressure Sensor
This VI can be used to view the pressure sensor measurements as the plunger is moved at
different locations within the syringe on the QNET mechatronic sensors trainer.
Exercise 1: Enter collected results in Table and capture the Sensor Readings scope.
V
Sensor Measurement: 6.0 cm
V
Sensor Measurement: 5.0 cm
V
Sensor Measurement: 4.0 cm
V
Sensor Measurement: 3.0 cm
Sensor Measurement:0.0 cm V
89
Conclusion -
90
Experiment 3
1. Manually perturb the plastic band that is attached to the piezo sensor by flicking it
and examine the response in the piezo (V) Scope.
2 Grab the end of the plastic band and move it slowly up and down. Examine the
response.
3. Exercise 1: From these two tests, what does the Piezo sensor measure? How is this
different then a strain gage measurement? Capture a sample Piezo (V) scope response after it
has been perturbed (by flicking it).
Hint: You can use the cursor to take measurements off the graph.
3.3. Exercise
Conclusion -
91
Experiment 4
4. Fill out table with appropriate amount of data points. Notice that as the measured
potentiometer readings are entered, a curve is automatically generated to fit the data
the slope and intercept of this line is generated as well.
5. Exercise 1: Enter the collected data in Table and capture the Sensor Reading chart.
Conclusion -
93
Experiment 5
i. Get a target, such as a sturdy piece of cardboard, that is at least 10 by 10 cm2 with a
reflective color like white or yellow.
ii. Begin with the target close to the Ir sensor and slowly move it away.
iii. Once its range of operation is found, enter the distance between the target and the IR
sensor in the Target Range (cm) array.
iv. Repeat for different target positions. The IR sensor is quadratic. As the Measurement are
entered, the coefficients for the second-order polynomial are generated and the fitted
curve is automatically plotted.
v. Exercise 1: Record your distance and voltage observation in Table and capture the
corresponding Sensor Reading scope
vi. Exercise 2: What did you notice when the target is close to the IR sensor? That is did the
behavior of the sensor change when the target was in close proximity as opposed to being
further way?
Exercise 1: Collect data
c cm
Conclusion -
95
Experiment 6
Sonar Sensor
Use this VI to view the sonar Measurement as a target is moved at different distances away from
the sensor.
1. Get a target such as a sturdy piece of cardboard that is at least 10 by 10 cm2 with a
reflective color like white or yellow.
2. Begin with the target close to the Ir sensor and slowly move it upwards.
3. Once its range of operation is found, enter the distance between the target and the
Sonar sensor in the Target Range (cm) array
4. Enter the corresponding measured voltage from the sonar sensor in the Sensor
Measurement (V) array,
5. Repeat for different target positions. The Sonar sensor is Linear. The slope and
intercept are generated and the fitted curve is automatically plotted.
6. Exercise 1: Enter your collected target distances and voltages in the Table.
Capture the Sensor reading scope as well.
7. Exercise 2: What is the resolution and operating range of the sonar senor? enter them
in table, below How does the resolution and range compare with the IR sensor?
Range Inch
Resolution inch
Conclusion -
97
Experiment 7
Optical Position
The QNET_ MECHKIT optical VI is used to view the measurements of the optical position
sensors the target is moved at different locations using the knob.
1. Gently turn the knob of the optical position sensor clockwise until the flat metal surface
gently rests on top of the tube. Then, rotate the knob slightly counter-clockwise so the 0
mark on the knob faces up. At this point, the reflective target is very close to the optical
sensor and will be array.
2. Enter the voltage measured by the optical position sensor, when the target is 0 cm away,
in the sensor measurements (V) array.
3. Turn the knob counter-clockwise one rotation to move the target further from the sensor.
The target move 1-inch for every 20 turns. Enter the position the target has moved from
the reference in the Target Range (cm) array.
4. Record the measured sensor voltage in the Sensor Measurement (v) array.
5. Take samples for the entire range of the target (i.e. until the knob cannot be rotated CCW
anymore) Remark that the optical position sensor is exponential as data is being entered,
the exponential parameters are generated and the fitted curve is automatically plotted.
6. Exercise 1: Enter the measured sensor data in table and capture the sensor readings
response.
98
Conclusion -
99
Experiment 8
Using this VI, the magnetic Field measurements can be read as the target is moved at different
locations using this knob on the QNET Mechatronic sensors trainer.
1. Gently turn the knob of the magnetic field sensor clockwise until it is st its limit. Then,
rotate the knob slightly counter-clockwise so the 0 mark on the knob faces up this will be
reference0cm target position. Enter this in the Target Range (cm) array.
2. Enter the voltage measured from the magnetic field position sensor for the reference 0cm
position in the Sensor Measurements (V) array. The array.
3. Turn the knob counter-clockwise one rotation to move the target further from the sensor.
The target moves 1-inch for every 20 turns. Enter the position the target has moved from
the reference in the Target Range (cm) array.
4. Record the measured sensor voltage in the Sensor Measurements (V) array.
5. Take samples for the entire range of the target (i.e. until the knob cannot be rotated CCW
anymore) the magnetic field sensor is exponential. The parameters of the exponential
function are outputted and the fitted curve is automatically plotted as data is entered.
6. Exercise 1: Enter the range and measured sensor voltage in table below, and capture the
Sensor Readings scope.
Conclusion -
101
Experiment 9
Encoder
This VI shows the A, B and Index signals generated by the rotary optical encoder on the QNET
mechatronic sensors trainer as the knob is rotated.
1. Exercise 1: Turn the encoder knob clockwise and examine the response of the A and B
signals Note that the signals are offset by 2.5 V for display purposes. Enter your
observation in Table. Similarly, turn the encoder knob counter-clockwise and enter your
observation.
2. Exercise 2: When is the index pulse triggered? What can this be used for?
Clockwise
Counter-clockwise
1. Using the 16-bit Position (counts) indicator on the VI, rotate the knob and determine how
many counts there are per revolution. Enter your result in the counts per rev box in the VI
Rotate the knob and confirm that the Angle (deg) indicator is displaying an accurate
angle.
2. Turn the knob such that the 0 is in the upward position and reset the counter by clicking
on the Reset button.
3. Enable the index by clicking on the Enable index button.
4. Rotate the knob a full CW turn until the index is triggered. Keep turning the knob until
the 0 mark on the knob is pointing upwards. What do you notice about the 16-bit Position
(counts) and the Angle (deg) indicator values?
102
5. Adjust the Reload Value such that Angle (deg) measures 0 degrees when the 0 mark of
the knob is pointing up. Confirm this by moving the knob CW.
6. Exercise 3: Enter the count per rev and the Reload Value values used for a calibrated
measurement in Table 30.
7. Exercise 4: Position the knob such that its 0 label is pointing upwards again. The counts
per rev and Angle (deg) should both be reading 0. Rotate the knob in the CCW fashion
one full rotation Is Angle (deg) reading 0 degrees? Discuss why or why not.
Conclusion -