Improving Automatic Viscosity Monitoring in The Lab
Improving Automatic Viscosity Monitoring in The Lab
Improving Automatic Viscosity Monitoring in The Lab
Lab
The viscosity of a lubricating oil is perhaps its most important physical property. Among other
factors, it determines the dynamic load that may be supported, what the lubrication film thickness
will be, and what the flow rate will be under a given pressure. Designers of machinery specify a
viscosity or “viscosity grade” which must be used in their equipment. The specification comes from
a combination of design parameters such as bearing loads, machining tolerances and the required
rate of heat removal.
The vast majority of automatic viscometers in use today measure each sample only once. The
main reason is that it is not cost effective for the laboratory to run every sample twice to
crosscheck results, because this doubles analysis times. By making only one measurement, the
lab is in effect relying on the accuracy of the instrument. While these instruments are accurate
when conditions are ideal (for example with new oils) when measuring used oil, particles can
collect on the viscometer tube walls, shifting the calibration. When a viscometer tube is out of
calibration, it will deliver consistent but inaccurate results. To counteract this, most labs will run
periodic quality control (QC) samples to verify calibration. When this QC check fails, all samples
back to the previous QC check should be remeasured. However, what happens when a temporary
condition presents itself: one that corrects after a few more samples have been run?
A large particle or filter fiber can partially plug a tube for the next several samples. However, if that
particle is flushed prior to the next QC check, how can the lab know there was a problem with
some of the measurements since the last QC check? Likewise, a sample may contain water or
other contaminants that are not easily flushed by the cleaning solvent. In this case, it may take
several cycles to properly clean out that viscometer tube. The samples run in the meantime can
have inaccurate measurements, which again would never be picked up by QC samples unless a
quality check is run every couple of samples. If any other conditions occur, a situation may arise
where a highly accurate laboratory viscometer can generate a viscosity measurement that is off by
several percentage points.
To counteract these problems, a new type of automatic viscometer has been designed by
Edmonton, Alberta-based Dare Instruments. The Trivisc, recently introduced by Dare, is one of
only a few automatic viscometers specifically designed for used oil analysis. The instrument
utilizes multiple sensors to derive three unique measurements on each oil sample. The sensor’s
unique, two-piece design allows the viscometer to detect slight inconsistencies caused by
calibration shift, partial blockages, water or entrained gases. Comparative analysis allows a
confidence factor to be generated. When this result is greater than a set limit, the test is
considered valid and the viscosity accepted as accurate. In the case of a low confidence factor,
the sample may be retested or the measurement parameters may be adjusted. This built-in quality
control check happens automatically on every sample tested.
To cost-effectively measure viscosity at two temperatures and calculate the VI, commercial labs
need an accurate, high-speed instrument. The Trivisc instrument can provide rapid turnaround,
allowing more than 90 viscosity measurements, at two different temperatures. Adaptive, automatic
cleaning algorithms ensure that minimum time and solvent is required between samples. Wide-
range viscometer tubes are utilized so that tubes do not need to be changed. On new oil, relative
standard deviation (RSD) repeatability ranges from 0.24 percent to 0.50 percent above normal
viscosity ranges with an accuracy of 0.5 percent.
At this level of accuracy, commercial oil analysis labs now can have the rapid throughput of the
highest quality automatic viscometer with comparable repeatability and reproducibility of the
stringent ASTM D445 procedure.