Golander Pump: Operation Manual

Download as pdf or txt
Download as pdf or txt
You are on page 1of 27

GOLANDER PUMP

Operation Manual
For Basic Variable-Speed Peristaltic Pump

BT100S, BT300S, BT600S

info@golanderpump.com

http://golanderpump.com

1-678-587-8806
Contents
Safety Cautions ..................................................................................... 1
1 Description ......................................................................................... 1
2 Functions and Features...................................................................... 2
3 Components and Connectors ............................................................. 3
4 Display Panel and Operating Keypads ............................................... 4
4.1 Keypad ..................................................................................... 4
4.2 LCD Screen Display ................................................................. 5
5 Parameter Settings ............................................................................ 7
6 External Control Interface ................................................................... 9
7 Operation Instructions ...................................................................... 10
7.1 Before Operation .................................................................... 10
7.2 Power Connection .................................................................. 10
7.3 Mode Change......................................................................... 10
7.4 Internal Control Mode .............................................................. 11
7.5 External Control Mode............................................................ 12
7.6 Time Dispense Mode.............................................................. 15
7.7 Logic Level 1 Control Mode (footswitch) ................................. 16
7.8 Logic Level 2 Control Mode (footswitch and direction switch) . 17
7.9 Communication Mode ............................................................ 18
7.10 Speed Setting....................................................................... 19
8 Maintenance .................................................................................... 20
8.1 Warranty................................................................................. 20
8.2 Regular Maintenance ............................................................. 20
8.3 Malfunction Solutions ............................................................. 20
9 Dimensions ...................................................................................... 22
10 Naming Rule .................................................................................. 22
11 Specifications ................................................................................. 23
Golander Pump golanderpump.com

Safety Cautions
Danger: Please use correct AC power voltage source shown on
the sticker attached on the equipment to avoid any damage.
Please do not open the case. High voltages exist and are
accessible. Use extreme caution when servicing internal components.
For maintenance, please contact the manufacturer or distributor directly.
Danger: Turn drive off before removing or installing tubing. Fingers or
loose clothing could get caught in drive mechanism.

Warning: Tubing breakage may result in fluid being sprayed


from pump. Use appropriate measures to protect operator and
equipment.
Warning: Remove power from pump before attempting any maintenance
or any cleaning operation is started.
Warning: Remove power from pump before connecting or disconnecting
the external control device or communication interface.
Warning: Pump is provided with a grounded plug, it must be well-
grounded at all times.
Warning: This product is not designed for, nor intended for use in patient
connected applications; including, but not limited to, medical and dental
use.

1 Description
BT-S basic variable-speed peristaltic pump provides not only the basic
functions such as reversible direction, start/stop and adjustable speed,
but also Time Dispense Mode and Anti-Drip function. With MODBUS
RS485 interface, pump is easy to communicate with external device, such
as PC, HMI or PLC. This pump series includes:
• BT100S, flow rate 0.00011-720 mL/min, speed 0.1-150 rpm
• BT300S, flow rate 0.006-1600 mL/min, speed 0.1-350 rpm
• BT600S, flow rate 0.006-2900 mL/min, speed 0.1-600 rpm

1
Golander Pump golanderpump.com

2 Functions and Features


Advantage of peristaltic pump: Peristaltic pump can handle extremely
viscous fluids, abrasive slurries and corrosive fluids. There is no seals in
contact with the medium pumped and no valves to clog. The inner
surfaces are smooth and easy to clean; fluid contacts only the tubing or
tube material. Suction lift and priming can be up to 8m water column at
sea level. It can handle the most shear sensitive of fluids like latex or
firefighting foam with low shearing. It is capable of running dry and
pumping fluids with high quantities of entrained air, such as black liquor
soap. The high volumetric efficiency allows operation in metering or
dosing applications where high accuracy is required. Tubing and tube
materials are available for food and pharmaceutical use.

• LCD displays rotating speed and working mode.


• Membrane keypad.
• Reversible direction, start/stop control and adjustable speed.
• Time dispense mode: automatically dispense by setting the
duration for each dose, lag time between doses and number of
cycles.
• 0.5% high precision rotating speed control with 0.1 rpm speed
resolution when speed is between 0-100 rpm, 1 rpm speed
resolution when speed is between 100-600 rpm.
• External logic level signal can control start/stop, direction and
dispense functions; external analog signal can adjust the rotating
speed. Signal is optically isolated.
• With RS485 MODBUS interface, easy to communicate with
external device.
• Internal double-layer isolation structure; circuit board with
conformal coating makes the pump dust-proof and moisture
proof.
• Anti-electromagnetic interference feature, wide input voltage
range for complex power environment.
• Stainless steel enclosure, easy to clean, resistant to the
corrosion of the acid, alkali, sodium and organic solvents.
2
Golander Pump golanderpump.com

3 Components and Connectors


Handle

LCD Display

Keypad
Coupler

Pump Head
Holder

DB15 External
Control Interface

Power Switch Cooling Fan

Power Connector

Fuse

Figure 1. Components and Connectors

3
Golander Pump golanderpump.com

4 Display Panel and Operating Keypads

LCD DISPLAY
UP

Int DOWN

32.5rpm

MENU MODE

golanderpump.com

MENU MODE PRIME DIRECTION START/STOP

Figure 2. Display Panel

4.1 Keypad
UP Key. When press it shortly, the last digit of the value will
increase 1. Hold the key to increase the value quickly.
DOWN Key. When press it shortly, the last digit of the value will
decrease 1. Hold the key to decrease the value quickly.
MENU Key. When on main screen, press the MENU key to enter
MENU
the setting menu. When on the setting menu, press the MENU
key to switch between the different setting menus. Press and hold it to
return to main screen. When the drive is running, this key is disabled.
DIRECTION Key. Press to key to change the drive rotating
direction, clockwise or counterclockwise.
MODE key. When the drive is not running, use the MODE key to
MODE
change the working mode: Internal Control mode, External
Control mode, Time Dispense mode, Logic Level 1 control mode
or Logic Level 2 Control mode.

PRIME key. Press the key to run pump at maximum allowed


speed in the direction shown on the display. Press again to
return to the previous state.

4
Golander Pump golanderpump.com

START/STOP key. Press to start or stop the drive.

4.2 LCD Screen Display

A B C D E F

G
Int
32.5rpm
Figure 3. Display screen

4.2.1 A. Control Mode


It displays current control mode. Press MODE key to switch the control
mode: Internal Control mode, External Control mode, Time Dispense
mode, Logic Level 1 control mode or Logic Level 2 Control mode.
• Internal Control Mode: Use the keypad to operate pump. Use
optional external pulse signal to control start and stop.
• External Control Mode: Use external analog signal to control
rotation speed. Use external logic level signal to control direction,
start and stop. The keypad is disabled.
• Time Dispense Mode: Dispense fluid automatically by setting
duration for each dose, time between doses and number of cycles.
• Logic Level 1 Control Mode (footswitch): Use external logic level
signal to control start and stop. Use the keypad to control direction
and speed.
• Logic Level 2 Control Mode (footswitch and direction switch): Use
external logic level signal to control start, stop and direction. Use
the keypad to adjust speed.
4.2.2 B. Running State
It shows current drive running state.

5
Golander Pump golanderpump.com

Stopped Running Paused


Figure 4. Running State

4.2.3 C. Direction State


It shows current drive direction setting.

Clockwise Counterclockwise
Figure 5. Direction

4.2.4 D. Key Tone State


It shows the state of the key tone, on or off.

Tone on Tone off


Figure 6. Key Tone

4.2.5 E. Keypad Lock State


It shows the state of the keypad lock. When the keypad is locked, only
the START/STOP key will work. On the main screen, press and hold the
DIRECTION key to lock the keypad; press and hold the START/STOP
key to unlock the keypad.

Keypad unlocked Keypad locked


Figure 7. Keypad Lock

4.2.6 F. Communication State


It shows the current RS485 communication state.

6
Golander Pump golanderpump.com

#1
Communication Connected, pump
disconnected number is set to 1
Figure 8. Communication state

4.2.7 G. Rotating Speed State


It shows the current rotating speed, rpm. When the drive is running at full
speed, the display will be >>>>>>>.

5 Parameter Settings
On the main screen, press the MENU key to enter the setting menu, and
press UP or DOWN key change the value, then press MENU key again
to confirm the setting. The screen will change to the setting for next
parameter. To return the main screen, press and hold the MENU key or
press the MENU key a few times until it goes back to the main screen
(Figure 9). Please note for Time Dispense mode, there are 5
parameters; for the other modes, there are 6 parameters.

7
Golander Pump golanderpump.com

Time Dispense Mode Int, Ext, Lev1, Lev2


Disp Int
32.5rpm 32.5rpm

1.Time Unit: 5.Reverse Angle:


Second 360

2.Disp Time: 6.Max Speed:


5.1Second 150rpm

3.Pause Time: 7.Pump No:


2.0Second #1

4.Cycles: 8.Beep Setting:


3 Enable

5.Reverse Angle: 9.Language:


360 ENGLISH

10.Contrast:
10

Figure 9. Parameter Setting Flow Chart

Note 1. Time Unit: Time unit for dispense mode. The time can be in
hours, minutes or seconds.
Note 2. Disp Time: Dispense time for Time Dispense mode. It is the
dispense duration for each dose. The range is 0.1-999
seconds/minutes/hours.

8
Golander Pump golanderpump.com
Note 3. Pause Time: Pause time for dispense mode. It is the lag time
between doses when the number of cycles setting is more than 1. The
range is 0.1-999 seconds/minutes/hours.
Note 4. Cycles: Number of cycles for dispense mode. The range is 0-
999 cycles. When it is set to 0, the dispense process will keep running
until START/STOP is pressed. When it is set to the other value, pump
will stop when pump finishes dispensing the set number of cycles
(Figure 20).
Note 5. Reverse Angle: To minimize the drip after a dispense, the
drive can reverse direction to draw the fluid back at the end of the
tubing. The reverse angle range is 0-720 degrees. When it set to 0,
the anti-drip function is disabled.
Note 6. Max Speed: Maximum speed for External Control mode. It is
the maximum speed that the external analog signal can control.
Note 7. Pump No: It is the pump communication address for
communication mode. When the number is changed, please restart
pump to apply the setting.
Note 8. Beep Setting: Set the key tone on or off.
Note 9. Language: System language setting, English or Chinese.
Note 10. Contrast: LCD backlight contrast setting.

6 External Control Interface

1 2 3 4 5 6 7 8

9 10 11 12 13 14 15

DB15 Mark Note


1 ADC_W Positive of external analog input
2 B Communication interface, B pole of RS485
3 A Communication interface, A pole of RS485
4 VCC_W External DC power input

9
Golander Pump golanderpump.com
5
6 CW_W External input signal to control direction
7
8 COM Ground of external power
9 AGND Negative of analog signal input
10 +12V Positive of internal +12V power source
11 GND Ground of Internal power source
12
13 RS_W External start/stop signal input terminal
14
15

7 Operation Instructions
7.1 Before Operation
1) Please check the packing slip to make sure nothing is wrong or
damaged in the package. If there is problem, please contact the
manufacturer or distributor.
2) Read through the instruction.
3) There should be more than 200 mm space for the back of the
pump when it is running.

7.2 Power Connection


The voltage of the power supply should be marked on the sticker of the
pump. Please make sure to use the right power source for the pump.
Please plug the power cord into the IEC Power Connector on the rear of
the pump and plug the opposite end of the power cord into an electrical
outlet. Flip the power switch located on the rear of the pump.

7.3 Mode Change


Turn on the power switch. The display will show welcome message then
show the main screen. Press the MODE key to change the working mode
(Figure 10).

10
Golander Pump golanderpump.com

Int lev2
32.5rpm 32.5rpm

Ext V lev1
32.5rpm 32.5rpm

Disp
32.5rpm
Figure 10. Change Working Mode

7.4 Internal Control Mode


Use the keypad to operate pump. Use optional external pulse signal to
control start and stop.
1) Turn on the power switch. Pump will display the main screen.
2) Press MODE key to change the mode to Internal Control mode
(Int shown on the screen).
3) Press UP or DOWN key to adjust the speed.
4) Press DIRECTION key to change the rotating direction.
5) Press START/STOP key to start or stop pump.
6) Press PRIME key, pump will run at maximum allowed speed.

Int Int
32.5rpm >>>>>>
Normal Speed Full Speed
Figure 11. Normal vs Full Speed

To use external pulse signal to control start and stop, wire the DB15
11
Golander Pump golanderpump.com
connector as shown on Figure 12 or Figure 13, and connect it to the DB15
port on the rear of the pump.

1 2 3 4 5 6 7 8

9 10 11 12 13 14 15

RS_W

Figure 12. Control Start/Stop with internal 12V Power Source

1 2 3 4 5 6 7 8

9 10 11 12 13 14 15

RS_W

VCC_W COM

12VDC

Figure 13. Control Start/Stop with External 12V Power Source

The RS_W would act like a momentary switch. When the switch is pushed
then released, pump will start or stop.

7.5 External Control Mode


On this mode, use external logic level signal to control direction, start and
stop; use external analog signal to control rotation speed. The keypad is
disabled. The analog signal could be 0-5V, 0-10V or 4-20mA. By default,
the signal is 0-5V. For 0-10V or 4-20mA, the jump setting on the analog
signal control board has to be changed (Figure 14).

12
Golander Pump golanderpump.com

0-5V (default) 0-10V 4-20mA


Figure 14. Analog Signal Control Board Setting

ExtV ExtI
0.0rpm 0.0rpm
For External Voltage Signal For External Current Signal
0-5V/0-10V 4-20mA
Figure 15. External Control Mode

To control pump by external signal:


1) Switch the power off. Wire the DB15 connector as shown on
Figure 16 or Figure 17. Connect it to the DB15 port on the rear of
the pump.

1 2 3 4 5 6 7 8

9 10 11 12 13 14 15

CW_W
ADC_W AGND
RS_W

0-5V/0-10V VCC_W COM


/4-20mA
12VDC

Figure 16. DB15 Wiring with External 12VDC Power Source

13
Golander Pump golanderpump.com

1 2 3 4 5 6 7 8

9 10 11 12 13 14 15

CW_W
ADC_W AGND
RS_W

0-5V/0-10V
/4-20mA

Figure 17. DB15 Wiring with Internal 12VDC Power Source

2) Turn on the power switch. Pump will display the main screen.
3) Press MODE key to change the mode to External Control mode
(ExtV or ExtI shown on the screen).
4) Close the external RS_W switch, and turn on the external analog
signal power source. The speed will change according to the
intensity of the input signal. Open the RS_W to stop the drive.
5) Open CW_W switch, then pump will run in clockwise direction;
close the CW_W switch, then pump will run in counterclockwise
direction.
Note: The external DC power source can be 5V or 12V. If it is 24V,
1.5K resistor is needed to protect internal circuit.

1 2 3 4 5 6 7 8

9 10 11 12 13 14 15

1.5K CW_W

ADC_W AGND RS_W


1.5K

0-5V/0-10V VCC_W COM


/4-20mA
24VDC

Figure 18. DB15 Wiring with External 24VDC Power Source

14
Golander Pump golanderpump.com
7.6 Time Dispense Mode
Pump will dispense fluid automatically by setting duration for each dose,
pause time between doses and number of cycles. When dispensing, the
display will show the dispensing time or lag time that has passed (1.2s on
the figure below), and the total cycles (4 on the figure) that has dispensed.

Disp Disp
32.5rpm 32.5rpm 1.2s 4
Dispense Stopped Dispense Running
Figure 19. Time Dispense Mode

To set Time Dispense mode:


1) Turn on the power switch. Pump will display the main screen.
2) Press MODE key to change the mode to Time Dispense mode
(Disp shown on the screen).
3) Press MENU key to enter the Setting menu.
4) Set duration for each dose, pause time between doses and
number of cycles.
5) Return to main screen.
Dispense
1 Stop
Duration

Dispense Press
Start Cycles 0 Lag Time Stop
Duration START/STOP

Dispense
N Lag Time Run N times Stop
Duration

Figure 20. Dispense Cycle Setting

To run Time Dispense:


1) Press DIRECTION key to change the running direction, clockwise
or counterclockwise.
2) Press START/STOP key to start dispensing.
3) When the drive is running, press START/STOP key to stop the
15
Golander Pump golanderpump.com
drive anytime when an accident occurs.
4) A footswitch can be used to start/stop pump.

7.7 Logic Level 1 Control Mode (footswitch)


Use external logic level signal to control start and stop. Use the keypad
to control direction and speed.

lev1
32.5rpm
Figure 21. Logic Level 1 Control Mode

1) Switch the power off. Wire the DB15 connector as shown on


Figure 22 or Figure 23, and connect it to the DB15 port on the rear
of the pump.

1 2 3 4 5 6 7 8

9 10 11 12 13 14 15

RS_W

Figure 22. Logic Level 1 Control with Internal 12V Power Source

1 2 3 4 5 6 7 8

9 10 11 12 13 14 15

RS_W

VCC_W COM

12VDC

Figure 23. Logic Level 1 Control with External 12V Power Source

2) Turn on the power switch. Pump will display the main screen.
16
Golander Pump golanderpump.com
3) Press MODE key to change the mode to Logic Level 1 control
mode (shows lev1 on the screen).
4) Press UP or DOWN key to adjust the speed.
5) Press DIRECTION key to change the rotating direction.
6) When the switch is closed, the drive will start running; when the
switch is open, the drive will stop.
Note: Use this mode to work with a TIME CONTROLLER.

7.8 Logic Level 2 Control Mode (footswitch and


direction switch)
Use external logic level signal to control start, stop and direction. Use the
keypad to control speed.

lev2
32.5rpm
Figure 24. Logic Level 2 Control Mode

1) Switch the power off. Wire the DB15 connector as shown on


Figure 25 or Figure 26, and connect it to the DB15 port on the
rear of the pump.

1 2 3 4 5 6 7 8
9 10 11 12 13 14 15

CW_W

RS_W

Figure 25. Logic Level 2 Control with Internal 12V Power Source

17
Golander Pump golanderpump.com

1 2 3 4 5 6 7 8

9 10 11 12 13 14 15

CW_W

RS_W

VCC_W COM

12VDC

Figure 26. Logic Level 2 Control with External 12V Power Source

2) Turn on the power switch. Pump will display the main screen.
3) Press MODE key to change the mode to Logic Level 2 control
mode (shows lev2 on the screen).
4) Press UP or DOWN key to adjust the speed.
5) When the switch RS_W is closed, the drive will be running at the
set speed; when the switch RS_W is open, the drive will stop.
6) When the switch CW_W is open, the drive will be running in
clockwise direction. When the switch CW_W is closed, the drive
will be running in counterclockwise direction.

7.9 Communication Mode


The RS485 interface supports standard MODBUS protocol. Pump can
communicate with external device via the communication port. Please
refer to the Communication Instruction manual for the parameters and
supported commands.

Int Int #1
32.5rpm 32.5rpm
Communication Disconnected Communication Connected
Pump number is set to 1.
Figure 27. Communication Mode

18
Golander Pump golanderpump.com
1) When the power is off, wire the DB15 connector as shown on
Figure 28, and connect it to the DB15 port on the rear of the pump.
External DC power source is recommend to avoid electrical
interference.

1 2 3 4 5 6 7 8

9 10 11 12 13 14 15

A VCC_W COM

12VDC

Figure 28. RS485 MODBUS Wiring

2) Turn on the power switch. Pump will display the main screen.
3) Press MODE key to change the mode to Internal Control mode or
Time Dispense mode.
4) Control pump with communication interface.
5) Press the START/STOP key to stop the drive anytime.

7.10 Speed Setting


On the main screen, the speed resolution is 0.1 rpm when the speed is
between 0.1 and 100 rpm; the speed resolution is 1 rpm when the speed
is between 100 and 600 rpm. The speed can be set by pressing the UP
or DOWN key. Press the UP or DOWN key shortly, the last digit of the
value will change 1. Hold the UP or DOWN key to change the value fast.
Hold the PRIME key, then press the UP key to set the speed to maximum
directly; hold the PRIME key, then press the DOWN key to set the speed
to minimum (0.1 rpm) directly.

19
Golander Pump golanderpump.com

8 Maintenance
8.1 Warranty
The product comes with one-year labor and parts warranty. The limited
warranty does not cover any damage that is caused by improper usage
and handling.

8.2 Regular Maintenance


1) Always check the tubing and connections to make sure there is
no leakage.
2) Do not cover the fan on the rear of the pump.
3) Do not use water to wash the pump. Keep pump head dry.
4) Do not use chemical solvents to clean pump and pump head.

8.3 Malfunction Solutions


No. Malfunction Description Solution
1 Hardware No display 1. Check the power cord
2. Check the fuse. If it was blown,
replace it with a 1A slow-blow fuse
3. Check the internal power cord
connection inside the pump.
2 Hardware Motor does 1. Check the indicator of the driver
not work board.
2. Check the wire connection between
the motor and the driver board.
3. Check the wire connection between
the driver and the main board.
4. Check the power voltage for the
pump.
3 Hardware Motor is 1. Check the wire connection between
trembling the motor and the driver board.
2. The motor is overloaded. Check
the mechanical connection.
4 Hardware Motor only Check the connection between the

20
Golander Pump golanderpump.com
runs in one drive board and the main control
direction board.
5 Hardware Keypad 1. Check the wire connection between
does not keypad and the main board.
work 2. Check if the key is broken.
6 Hardware External 1. Check the wiring of the connector.
control does 2. Check if the external control power
not work voltage is provided.
3. Check the connections of the
external control board.
7 Hardware RS485 com 1. Check the wiring of the connector.
does not 2. Check if the external control power
work voltage is provided.
3. Check the connections of the
communication board.
8 Hardware Noisy when Check the screws and level on pump
running head to make sure they are secure.
9 Software External Check if pump is on External Control
control does Mode.
not work
10 Software RS485 does 1. Check if the display shows the
not work communication is ready.
right 2. Reset the address of the pump.
3. Check whether on the bus there
are two pumps using the same
address

If the problem can not be solved, please contact the manufacturer


or distributor.

21
Golander Pump golanderpump.com

9 Dimensions

238

Figure 29. Dimensions (mm)

10 Naming Rule

B T 1 0 0 S 1 65

Drive Type Flow Rate Rotation Speed Version Pump type Function Code IP rating
B: Step motor Q: Microflow 05: 50 rpm 0-9 D: Fixed speed 1-9 for special Default:
W: DC brushless T: Laboratory 08: 80 rpm S: Variable speed pumps IP31
motor G: Industrial 10: 150 rpm * L: Flow pump
J: AC motor P: Batch transfer 30: 350 rpm F: Dispensing
F: Explosion Proof 60: 600 rpm pump
Motor
* BT102S: 100 rpm
Z: DC motor

22
Golander Pump golanderpump.com

11 Specifications
Speed resolution 0.1-100 rpm, 0.1 rpm resolution;
100-600 rpm, 1 rpm resolution
Speed accuracy 0.5%
Power supply AC 85-265V, 50/60Hz
Power consumption BT100S: < 40W; BT300S:< 50W;
BT600S: < 60W
External logic level control 5V, 12V (standard), 24V (optional)
signal
External analog control 0-5V (standard); 0-10V, 4-20mA (optional)
signal
Communication interface RS485 MODBUS
Operating condition Temperature 0~40°C
Relative humidity <80%
IP grade IP31
Display 132x32 LCD
Dimensions (LxWxH) 226 x 150 x 238 mm
(8.90 x 5.91 x 9.37 inch)
Weight BT100S: 4.7 kg (10.4 lbs)
BT300S: 4.9 kg (10.8 lbs)
BT600S: 5.2 kg (11.5 lbs)

23
Golander Pump golanderpump.com
BT100S Suitable Pump Heads and Tubing, Flow Parameters
Flow rate
Drive Pump
Ch Tubing size (mm) per channel
type head
(mL/min)
DG6-1
1 Wall:0.8~1, ID:≤2.4 0.00016~26
(6rollers)
DG10-1
1 Wall:0.8~1, ID:≤2.4 0.00011~20
(10rollers)
DG6-2
2 Wall:0.8~1, ID:≤2.4 0.00016~26
(6rollers)
DG10-2
2 Wall:0.8~1, ID:≤2.4 0.00011~20
(10rollers)
DG6-4
4 Wall:0.8~1, ID:≤2.4 0.00016~26
(6rollers)
DG10-4
4 Wall:0.8~1, ID:≤2.4 0.00011~20
(10rollers)
BT100S
DT10-18 1 13# 14# , Wall: 0.8~1, ID:≤3.17 0.0002~82
DT10-28 2 13# 14# , Wall: 0.8~1, ID:≤3.17 0.0002~82
DT10-48 4 13# 14# , Wall: 0.8~1, ID:≤3.17 0.0002~82
YZ15 1 13# 14# 16# 19# 25# 17# 0.006~420
YZ25 1 15# 24# 0.17~420
2 x YZ15 2 13# 14# 16# 19# 25# 17# 0.006~420
2 x YZ25 2 15# 24# 0.17~420
YT15 1 13# 14# 16# 19# 25# 17# 18# 0.006~570
YT25 1 15# 24# 35# 36# 0.17~720
2 x YT15 2 13# 14# 16# 19# 25# 17# 18# 0.006~570
2 x YT25 2 15# 24# 35# 36# 0.006~720
DT15-24 2 19# 16# 25# 17# 0.05~400

24
Golander Pump golanderpump.com
BT300S Suitable Pump Heads and Tubing, Flow Parameters
Flow rate
Drive Pump
Ch Tubing size per channel
type Head
(mL/min)
YZ15 1 13# 14# 16# 19# 25# 17# 0.006~990
YZ25 1 15# 24# 0.16~990
2 x YZ15 2 13# 14# 16# 19# 25# 17# 0.006~990
2 x YZ25 2 15# 24# 0.16~990
BT300S DT15-24 2 19# 16# 25# 17# 0.05~930
DT15-44 4 19# 16# 25# 0.05-610
YT15 1 13# 14# 16# 19# 25# 17# 18# 0.006~1300
YT25 1 15# 24# 35# 36# 0.16~1600
2 x YT15 2 13# 14# 16# 19# 25# 17# 18# 0.006~1300

BT600S Suitable Pump Heads and Tubing, Flow Parameters


Flow rate
Drive Pump
Ch Tubing size per channel
type head
(mL/min)
YZ15 1 13# 14# 16# 19# 25# 17# 0.006~1700
YZ25 1 15# 24# 0.16~1700
2 x YZ15 2 13# 14# 16# 19# 25# 17# 0.006~1700
2 x YZ25 2 15# 24# 0.16~1700
BT600S
DT15-24 2 19# 16# 25# 17# 0.047~1600
YT15 1 13# 14# 16# 19# 25# 17# 18# 0.006~2300
YT25 1 15# 24# 35# 36# 0.16~2900
2 x YT15 2 13# 14# 16# 19# 25# 17# 18# 0.006~2300

25