Golander Pump: Operation Manual
Golander Pump: Operation Manual
Golander Pump: Operation Manual
Operation Manual
For Basic Variable-Speed Peristaltic Pump
info@golanderpump.com
http://golanderpump.com
1-678-587-8806
Contents
Safety Cautions ..................................................................................... 1
1 Description ......................................................................................... 1
2 Functions and Features...................................................................... 2
3 Components and Connectors ............................................................. 3
4 Display Panel and Operating Keypads ............................................... 4
4.1 Keypad ..................................................................................... 4
4.2 LCD Screen Display ................................................................. 5
5 Parameter Settings ............................................................................ 7
6 External Control Interface ................................................................... 9
7 Operation Instructions ...................................................................... 10
7.1 Before Operation .................................................................... 10
7.2 Power Connection .................................................................. 10
7.3 Mode Change......................................................................... 10
7.4 Internal Control Mode .............................................................. 11
7.5 External Control Mode............................................................ 12
7.6 Time Dispense Mode.............................................................. 15
7.7 Logic Level 1 Control Mode (footswitch) ................................. 16
7.8 Logic Level 2 Control Mode (footswitch and direction switch) . 17
7.9 Communication Mode ............................................................ 18
7.10 Speed Setting....................................................................... 19
8 Maintenance .................................................................................... 20
8.1 Warranty................................................................................. 20
8.2 Regular Maintenance ............................................................. 20
8.3 Malfunction Solutions ............................................................. 20
9 Dimensions ...................................................................................... 22
10 Naming Rule .................................................................................. 22
11 Specifications ................................................................................. 23
Golander Pump golanderpump.com
Safety Cautions
Danger: Please use correct AC power voltage source shown on
the sticker attached on the equipment to avoid any damage.
Please do not open the case. High voltages exist and are
accessible. Use extreme caution when servicing internal components.
For maintenance, please contact the manufacturer or distributor directly.
Danger: Turn drive off before removing or installing tubing. Fingers or
loose clothing could get caught in drive mechanism.
1 Description
BT-S basic variable-speed peristaltic pump provides not only the basic
functions such as reversible direction, start/stop and adjustable speed,
but also Time Dispense Mode and Anti-Drip function. With MODBUS
RS485 interface, pump is easy to communicate with external device, such
as PC, HMI or PLC. This pump series includes:
• BT100S, flow rate 0.00011-720 mL/min, speed 0.1-150 rpm
• BT300S, flow rate 0.006-1600 mL/min, speed 0.1-350 rpm
• BT600S, flow rate 0.006-2900 mL/min, speed 0.1-600 rpm
1
Golander Pump golanderpump.com
LCD Display
Keypad
Coupler
Pump Head
Holder
DB15 External
Control Interface
Power Connector
Fuse
3
Golander Pump golanderpump.com
LCD DISPLAY
UP
Int DOWN
32.5rpm
MENU MODE
golanderpump.com
4.1 Keypad
UP Key. When press it shortly, the last digit of the value will
increase 1. Hold the key to increase the value quickly.
DOWN Key. When press it shortly, the last digit of the value will
decrease 1. Hold the key to decrease the value quickly.
MENU Key. When on main screen, press the MENU key to enter
MENU
the setting menu. When on the setting menu, press the MENU
key to switch between the different setting menus. Press and hold it to
return to main screen. When the drive is running, this key is disabled.
DIRECTION Key. Press to key to change the drive rotating
direction, clockwise or counterclockwise.
MODE key. When the drive is not running, use the MODE key to
MODE
change the working mode: Internal Control mode, External
Control mode, Time Dispense mode, Logic Level 1 control mode
or Logic Level 2 Control mode.
4
Golander Pump golanderpump.com
A B C D E F
G
Int
32.5rpm
Figure 3. Display screen
5
Golander Pump golanderpump.com
Clockwise Counterclockwise
Figure 5. Direction
6
Golander Pump golanderpump.com
#1
Communication Connected, pump
disconnected number is set to 1
Figure 8. Communication state
5 Parameter Settings
On the main screen, press the MENU key to enter the setting menu, and
press UP or DOWN key change the value, then press MENU key again
to confirm the setting. The screen will change to the setting for next
parameter. To return the main screen, press and hold the MENU key or
press the MENU key a few times until it goes back to the main screen
(Figure 9). Please note for Time Dispense mode, there are 5
parameters; for the other modes, there are 6 parameters.
7
Golander Pump golanderpump.com
10.Contrast:
10
Note 1. Time Unit: Time unit for dispense mode. The time can be in
hours, minutes or seconds.
Note 2. Disp Time: Dispense time for Time Dispense mode. It is the
dispense duration for each dose. The range is 0.1-999
seconds/minutes/hours.
8
Golander Pump golanderpump.com
Note 3. Pause Time: Pause time for dispense mode. It is the lag time
between doses when the number of cycles setting is more than 1. The
range is 0.1-999 seconds/minutes/hours.
Note 4. Cycles: Number of cycles for dispense mode. The range is 0-
999 cycles. When it is set to 0, the dispense process will keep running
until START/STOP is pressed. When it is set to the other value, pump
will stop when pump finishes dispensing the set number of cycles
(Figure 20).
Note 5. Reverse Angle: To minimize the drip after a dispense, the
drive can reverse direction to draw the fluid back at the end of the
tubing. The reverse angle range is 0-720 degrees. When it set to 0,
the anti-drip function is disabled.
Note 6. Max Speed: Maximum speed for External Control mode. It is
the maximum speed that the external analog signal can control.
Note 7. Pump No: It is the pump communication address for
communication mode. When the number is changed, please restart
pump to apply the setting.
Note 8. Beep Setting: Set the key tone on or off.
Note 9. Language: System language setting, English or Chinese.
Note 10. Contrast: LCD backlight contrast setting.
1 2 3 4 5 6 7 8
9 10 11 12 13 14 15
9
Golander Pump golanderpump.com
5
6 CW_W External input signal to control direction
7
8 COM Ground of external power
9 AGND Negative of analog signal input
10 +12V Positive of internal +12V power source
11 GND Ground of Internal power source
12
13 RS_W External start/stop signal input terminal
14
15
7 Operation Instructions
7.1 Before Operation
1) Please check the packing slip to make sure nothing is wrong or
damaged in the package. If there is problem, please contact the
manufacturer or distributor.
2) Read through the instruction.
3) There should be more than 200 mm space for the back of the
pump when it is running.
10
Golander Pump golanderpump.com
Int lev2
32.5rpm 32.5rpm
Ext V lev1
32.5rpm 32.5rpm
Disp
32.5rpm
Figure 10. Change Working Mode
Int Int
32.5rpm >>>>>>
Normal Speed Full Speed
Figure 11. Normal vs Full Speed
To use external pulse signal to control start and stop, wire the DB15
11
Golander Pump golanderpump.com
connector as shown on Figure 12 or Figure 13, and connect it to the DB15
port on the rear of the pump.
1 2 3 4 5 6 7 8
9 10 11 12 13 14 15
RS_W
1 2 3 4 5 6 7 8
9 10 11 12 13 14 15
RS_W
VCC_W COM
12VDC
The RS_W would act like a momentary switch. When the switch is pushed
then released, pump will start or stop.
12
Golander Pump golanderpump.com
ExtV ExtI
0.0rpm 0.0rpm
For External Voltage Signal For External Current Signal
0-5V/0-10V 4-20mA
Figure 15. External Control Mode
1 2 3 4 5 6 7 8
9 10 11 12 13 14 15
CW_W
ADC_W AGND
RS_W
13
Golander Pump golanderpump.com
1 2 3 4 5 6 7 8
9 10 11 12 13 14 15
CW_W
ADC_W AGND
RS_W
0-5V/0-10V
/4-20mA
2) Turn on the power switch. Pump will display the main screen.
3) Press MODE key to change the mode to External Control mode
(ExtV or ExtI shown on the screen).
4) Close the external RS_W switch, and turn on the external analog
signal power source. The speed will change according to the
intensity of the input signal. Open the RS_W to stop the drive.
5) Open CW_W switch, then pump will run in clockwise direction;
close the CW_W switch, then pump will run in counterclockwise
direction.
Note: The external DC power source can be 5V or 12V. If it is 24V,
1.5K resistor is needed to protect internal circuit.
1 2 3 4 5 6 7 8
9 10 11 12 13 14 15
1.5K CW_W
14
Golander Pump golanderpump.com
7.6 Time Dispense Mode
Pump will dispense fluid automatically by setting duration for each dose,
pause time between doses and number of cycles. When dispensing, the
display will show the dispensing time or lag time that has passed (1.2s on
the figure below), and the total cycles (4 on the figure) that has dispensed.
Disp Disp
32.5rpm 32.5rpm 1.2s 4
Dispense Stopped Dispense Running
Figure 19. Time Dispense Mode
Dispense Press
Start Cycles 0 Lag Time Stop
Duration START/STOP
Dispense
N Lag Time Run N times Stop
Duration
lev1
32.5rpm
Figure 21. Logic Level 1 Control Mode
1 2 3 4 5 6 7 8
9 10 11 12 13 14 15
RS_W
Figure 22. Logic Level 1 Control with Internal 12V Power Source
1 2 3 4 5 6 7 8
9 10 11 12 13 14 15
RS_W
VCC_W COM
12VDC
Figure 23. Logic Level 1 Control with External 12V Power Source
2) Turn on the power switch. Pump will display the main screen.
16
Golander Pump golanderpump.com
3) Press MODE key to change the mode to Logic Level 1 control
mode (shows lev1 on the screen).
4) Press UP or DOWN key to adjust the speed.
5) Press DIRECTION key to change the rotating direction.
6) When the switch is closed, the drive will start running; when the
switch is open, the drive will stop.
Note: Use this mode to work with a TIME CONTROLLER.
lev2
32.5rpm
Figure 24. Logic Level 2 Control Mode
1 2 3 4 5 6 7 8
9 10 11 12 13 14 15
CW_W
RS_W
Figure 25. Logic Level 2 Control with Internal 12V Power Source
17
Golander Pump golanderpump.com
1 2 3 4 5 6 7 8
9 10 11 12 13 14 15
CW_W
RS_W
VCC_W COM
12VDC
Figure 26. Logic Level 2 Control with External 12V Power Source
2) Turn on the power switch. Pump will display the main screen.
3) Press MODE key to change the mode to Logic Level 2 control
mode (shows lev2 on the screen).
4) Press UP or DOWN key to adjust the speed.
5) When the switch RS_W is closed, the drive will be running at the
set speed; when the switch RS_W is open, the drive will stop.
6) When the switch CW_W is open, the drive will be running in
clockwise direction. When the switch CW_W is closed, the drive
will be running in counterclockwise direction.
Int Int #1
32.5rpm 32.5rpm
Communication Disconnected Communication Connected
Pump number is set to 1.
Figure 27. Communication Mode
18
Golander Pump golanderpump.com
1) When the power is off, wire the DB15 connector as shown on
Figure 28, and connect it to the DB15 port on the rear of the pump.
External DC power source is recommend to avoid electrical
interference.
1 2 3 4 5 6 7 8
9 10 11 12 13 14 15
A VCC_W COM
12VDC
2) Turn on the power switch. Pump will display the main screen.
3) Press MODE key to change the mode to Internal Control mode or
Time Dispense mode.
4) Control pump with communication interface.
5) Press the START/STOP key to stop the drive anytime.
19
Golander Pump golanderpump.com
8 Maintenance
8.1 Warranty
The product comes with one-year labor and parts warranty. The limited
warranty does not cover any damage that is caused by improper usage
and handling.
20
Golander Pump golanderpump.com
runs in one drive board and the main control
direction board.
5 Hardware Keypad 1. Check the wire connection between
does not keypad and the main board.
work 2. Check if the key is broken.
6 Hardware External 1. Check the wiring of the connector.
control does 2. Check if the external control power
not work voltage is provided.
3. Check the connections of the
external control board.
7 Hardware RS485 com 1. Check the wiring of the connector.
does not 2. Check if the external control power
work voltage is provided.
3. Check the connections of the
communication board.
8 Hardware Noisy when Check the screws and level on pump
running head to make sure they are secure.
9 Software External Check if pump is on External Control
control does Mode.
not work
10 Software RS485 does 1. Check if the display shows the
not work communication is ready.
right 2. Reset the address of the pump.
3. Check whether on the bus there
are two pumps using the same
address
21
Golander Pump golanderpump.com
9 Dimensions
238
10 Naming Rule
B T 1 0 0 S 1 65
Drive Type Flow Rate Rotation Speed Version Pump type Function Code IP rating
B: Step motor Q: Microflow 05: 50 rpm 0-9 D: Fixed speed 1-9 for special Default:
W: DC brushless T: Laboratory 08: 80 rpm S: Variable speed pumps IP31
motor G: Industrial 10: 150 rpm * L: Flow pump
J: AC motor P: Batch transfer 30: 350 rpm F: Dispensing
F: Explosion Proof 60: 600 rpm pump
Motor
* BT102S: 100 rpm
Z: DC motor
22
Golander Pump golanderpump.com
11 Specifications
Speed resolution 0.1-100 rpm, 0.1 rpm resolution;
100-600 rpm, 1 rpm resolution
Speed accuracy 0.5%
Power supply AC 85-265V, 50/60Hz
Power consumption BT100S: < 40W; BT300S:< 50W;
BT600S: < 60W
External logic level control 5V, 12V (standard), 24V (optional)
signal
External analog control 0-5V (standard); 0-10V, 4-20mA (optional)
signal
Communication interface RS485 MODBUS
Operating condition Temperature 0~40°C
Relative humidity <80%
IP grade IP31
Display 132x32 LCD
Dimensions (LxWxH) 226 x 150 x 238 mm
(8.90 x 5.91 x 9.37 inch)
Weight BT100S: 4.7 kg (10.4 lbs)
BT300S: 4.9 kg (10.8 lbs)
BT600S: 5.2 kg (11.5 lbs)
23
Golander Pump golanderpump.com
BT100S Suitable Pump Heads and Tubing, Flow Parameters
Flow rate
Drive Pump
Ch Tubing size (mm) per channel
type head
(mL/min)
DG6-1
1 Wall:0.8~1, ID:≤2.4 0.00016~26
(6rollers)
DG10-1
1 Wall:0.8~1, ID:≤2.4 0.00011~20
(10rollers)
DG6-2
2 Wall:0.8~1, ID:≤2.4 0.00016~26
(6rollers)
DG10-2
2 Wall:0.8~1, ID:≤2.4 0.00011~20
(10rollers)
DG6-4
4 Wall:0.8~1, ID:≤2.4 0.00016~26
(6rollers)
DG10-4
4 Wall:0.8~1, ID:≤2.4 0.00011~20
(10rollers)
BT100S
DT10-18 1 13# 14# , Wall: 0.8~1, ID:≤3.17 0.0002~82
DT10-28 2 13# 14# , Wall: 0.8~1, ID:≤3.17 0.0002~82
DT10-48 4 13# 14# , Wall: 0.8~1, ID:≤3.17 0.0002~82
YZ15 1 13# 14# 16# 19# 25# 17# 0.006~420
YZ25 1 15# 24# 0.17~420
2 x YZ15 2 13# 14# 16# 19# 25# 17# 0.006~420
2 x YZ25 2 15# 24# 0.17~420
YT15 1 13# 14# 16# 19# 25# 17# 18# 0.006~570
YT25 1 15# 24# 35# 36# 0.17~720
2 x YT15 2 13# 14# 16# 19# 25# 17# 18# 0.006~570
2 x YT25 2 15# 24# 35# 36# 0.006~720
DT15-24 2 19# 16# 25# 17# 0.05~400
24
Golander Pump golanderpump.com
BT300S Suitable Pump Heads and Tubing, Flow Parameters
Flow rate
Drive Pump
Ch Tubing size per channel
type Head
(mL/min)
YZ15 1 13# 14# 16# 19# 25# 17# 0.006~990
YZ25 1 15# 24# 0.16~990
2 x YZ15 2 13# 14# 16# 19# 25# 17# 0.006~990
2 x YZ25 2 15# 24# 0.16~990
BT300S DT15-24 2 19# 16# 25# 17# 0.05~930
DT15-44 4 19# 16# 25# 0.05-610
YT15 1 13# 14# 16# 19# 25# 17# 18# 0.006~1300
YT25 1 15# 24# 35# 36# 0.16~1600
2 x YT15 2 13# 14# 16# 19# 25# 17# 18# 0.006~1300
25