Alanine 55 PDF
Alanine 55 PDF
Alanine 55 PDF
Technique: Spectrophotometry
Application Areas: Clinical
in clinical samples
Introduction
The liver is an important site for this metabolic reaction. Increased levels of
ALT are associated with hepatitis and other liver disorders associated with
hepatic necrosis as in cirrhosis, carcinoma and obstructive jaundice. Elevated
levels occur in circulatory failure and extensive trauma. Expected values are
up to 30U/l.
Principle
Biochrom Ltd
Biochrom Ltd 22 Cambridge Science Park, Milton Road
Certificate No: Cambridge, CB4 0FJ, UK
890333
Tel: +44 (0)1223 423723
Fax: +44 (0)1223 420164
Email: support@biochrom.co.uk
website: www.biochrom.co.uk
Method
The reagent can be prepared ready for use or obtained as a kit (Randox Labs,
UK)
• Insert blank and press green run key. This reference value is used for
subsequent samples until changed
o Read the absorbance rate of the blank over the period 1-4 minutes.
o If there is any change in the blank value over the time period, the
sample value should be appropriately corrected.
o A single blank as above suffices for subsequent analyses in the
same series.
The absorbance rate is shown as the slope of the graph over this time period.
Other information is displayed which, although not required for this calculation,
is a check on the chemical stability of the assay. The assay is shown
graphically as it proceeds, with absorbance difference and the line quality (a
coefficient of determination of > 95 % is expected if the assay was carried out
over a linear section).
The above procedure can be easily used with other instruments in the Libra
range by using the concentration mode and the 1.746 factor.
Ordering Details
Libra S5 80-2115-00
Libra S11 80-2115-15
Libra S12 80-2115-10
Libra S21 80-2115-25
Libra S22 80-2115-20
Libra S32 80-2115-30
The reaction can be accelerated for increased sensitivity if warmed. For this
purpose the Libra S21/S22 have the following accessories:
The Sipper (80-2112-25) enables some automation of the analyses, and can
be used together with a heated (not water heated) or non-heated single cell
holder.