1. General notes for users of the Service Manual VAP 20/30/40
This manual was written exclusive for the service staff of C. Gerhardt GmbH & Co. KG and for the
service staffs of our international agencies.
This service manual shall help to get a deeper knowledge of how the instruments are constructed and
how they work. It is an important tool to service and repair the instruments in a safe, competent and
specific way. The transparent structure and the appendix of modification information will enable to
order and to store the most useful spare parts. This manual will make servicing easier, in particular for
our international agencies.
To provide yourself against risks during operation or service, the following safety instructions should be
obeyed:
Make sure that no liquid gets into contact with cable connections or the interior of the electrical
parts of the unit! Danger of electrical shock!
Always switch off the apparatus at the mains and pull the plug before opening! Danger of
electrical shock!
The apparatus must be used according to the instruction manual. It is not allowed to make
changes to the apparatus in order to modify its application.
2 0 0 2 9 0 3 0 0 2 4 1 2 4 0 2 8 0 3 8 0 2 6 1 2 6 0 5 8 0 1 1 0 0 5 0 3 1 1 1 0 4 4 0
2 5 0
2 5 1
2 1 0
2 7 0
5 1
3 3 0
3 4 0
3 2 0 3 1 0 2 0 1 0 2 2 0 1 1 2 0 2 3 0 2 0 8 0 4 5 0
9 9 5
3 5 0
F r o n t V ie w
V A P 2 0 /3 0 /4 0
Date: Key to Front view VAP 20/30/40
Drawing-No. Part-No. Description Modif. 1 Modif. 2 Modif. 3 Modif. 4
Page 1
1 6 3 1 6 21 1 7 01 6 0 1 6 2 1 1 6 0 4 0 0 3 9 0 3 6 0 4 1 0 1 1 1 0 4 9 0 1 0 3 7 0
S 3
3 8 5
9 0 4 5 6
4 5 5
1 1 1 0
S 2
S 3
S 2
1 1 7 0
4 3 0
7 0
1 1 5 0
5
S 1
8
S 1
to c o n d e n s e r
3 0 9 9 0 4 8 0 8 1 8 0 2 9 0 9 9 0 4 0 9 9 8 5 2 0
R e a r V ie w
4 8 6 6 0
V A P 2 0 /3 0 /4 0
Date: Key to Rear View VAP 20/30/40
Drawing-No. Part-No. Description Modif. 1 Modif. 2 Modif. 3 Modif. 4
to s te a m g e n e ra to r
5 to 5 0 0 2 4
(re a r)
E 1
2
4
to
g r o u n d c o n n e c t io n to
1 0 0 R e c e iv e r
6 9 0 5
6 6 0 1 0 0
2
1 0 0
E 1
E 2
to 5 0 0 2 4
to 8 0 0
D is t r ib u tio n H e a d
9 8 1
4 6
t o c o n n e c t io n p ie c e
2 9 0
5 1 x 1 0 x 5
E 2
to
S a m p le T u b e C la m p in g D e v ic e
5 7 0
6
9 9 2
6 8 0 1 1 0 9 9 2 3 0 9 9 0 4 8 0
1 0 0
4 1 0 7 0 0 4 0 0
P u m p H o ld in g M o d u le
V A P 2 0 /3 0 /4 0
Date: Key to pump-holding module VAP 20/30/40
Drawing-No. Part-No. Description Modif. 1 Modif. 2 Modif. 3 Modif. 4
to P o w e r P C B
8 1 5 0
to c o n d e n s e r
4 2 0 to p e r is ta lt ic -
p u m p (p re s s u re )
4 8 5
4 3 0
to c o n d e n s e r
1 1 0 0 9 9 0 4 0 5 1 0 5 0 0 1 1 5 0
T u b in g C o n n e c tio n M o d u le
V A P 2 0 /3 0 /4 0
Date: Key to tubing connection module VAP 20/30/40
Drawing-No. Part-No. Description Remark Modif. 1 Modif. 2 Modif. 3
to p re s s u re s e n s o r
( c o o lin g w a te r ) 4 0
1
2
to m ic r o s w itc h
3
4
5
6
( tu b . c la m p . d e v ic e )
7
8
f1 to S te a m -
g 1
g e n e ra to r (fro n t)
f2
g 2
rd t
rh a
G e
5 1 b 1
d
b 2
d b 3
b 4
c 1 h 1
e 1
a
3 0
e 2
s t
d e
b 3
p o
f2
V a
2
5 0
3
1
f1
h 1
h 2
b 4
c 2
8 0 to
re a r c o v e r
c 1 to
g r o u n d c o n n e c tio n
c 2
h 2
7 0 , 7 1 b 1
b 2
e 1
e 2
P C B - m o d u le
6 0 7 2
g 2
g 1
9 0
V A P 2 0 /3 0
Date: Key to PCB-module VAP 20/30/40
Drawing-No. Part-No. Description Modif. 1 Modif. 2 Modif. 3 Modif. 4
a 1 1 1 0 1 2
to p re s s u re s e n s o r
rd t
rh a
( c o o lin g w a t e r )
G e
1
2
to m ic r o s w itc h
3
4
5
b 1
( t u b e c la m p . d e v ic e )
6
7
8
d
f1 to S te a m - b 2
g 1 b 3
g e n e ra to r (fro n t) b 4
f2 h 1
g 2
a
4 0
s t
d e
p o
V a
5 1
d
c 1
e 1
e 2
b 3
f2
2
5 0
3
1
f1
h 1
h 2
b 4
c 2
8 0 to
re a r c o v e r
c 1 to
g r o u n d c o n n e c t io n
c 2
h 2
7 0 , 7 1 b 1
P C B - m o d u le
e 1
e 2
b 2
g 2
6 0 7 2
g 1
9 0
V A P 4 0
3 2 0 2 1 0 2 0 0 3 3 0 1 4 0
3 1 0
5 0 2
5 3 5
5 1 5
3 0 0
E 2 5 0 0
5 0 5
3 5 0
E 1
5 0 0
5 3 0
8 9 0
E 2
3 4 0 4 0 0
E 1
S te a m G e n e ra to r
V A P 2 0 /3 0 /3 0
3 2 0
n e w s e n s o r , P a r t- N o . 4 0 8 5 7 s in c e ( 1 2 /9 5 )
3 3 0
5 1 5
3 0 0
E 2 5 0 0
5 0 5
3 5 0
E 1
5 0 0
5 3 0
8 9 0
E 2
M o d ific a tio n 1 1 .0 8 .1 9 9 5 , P a g e 2 3 4 0 4 0 0
E 1
T o g e th e r w ith th e c h a n g e m e n t o f th e le v e l s e n s o r , a ls o th e s te a m g e n e r a to r h a s b e e n c h a n g e d .
U n d e r th e s a m e P a r t- N o . 4 0 8 5 0 n o w o n ly th e n e w s te a m g e n e r a to r w ill b e d e lie v e r e d .
Date: Key to steam generator VAP 20/30/40
Drawing-No. Part-No. Description Modif. 1 Modif. 2 Modif. 3 Modif. 4
F Y D L /N M 2 M 1 B Z
D ia p h r a g m P u m p
E C K 1 0 3 1 0
Spare parts for diaphragm pump 10310
Attention !
In instruments Vap 20/30/40 there are different angle-connections for the pump
NaOH.
Angle-connections Part-No. 17911 and 17910 are replaced against the following
connections.
2) Other spare parts are not available. In case of damage customers only can
order
the complete pump 10310 .
Date: Spare parts and accessories VAP 20/30/40
Part-No. Description Modif. 1 Modif. 2 Modif. 3
Modification to Büchi
Modification to micro
Modification to acid-restistant
After switching-on the Vapodest the electronics measures the level of water inside the steam
generator via the level sensor. In case of a level below minimum, the pump for H2O steam
generator is switched on until the level sensor detects sufficient water inside the steam
generator. This will take some seconds.
Then the semiconductor relay switches on the tubular heating element with maximum power
to reach the necessary steam pressure for the distillation. When this pressure is reached, the
maximum heating power is switched off and the instrument reverts to stand-by with a
regulated and controlled heating. The instrument now is ready for distillation.
When starting a distillation the instrument will heat with a selected power from 30% - 100%
(VAP 20/30) or 40% - 100% (VAP 40). The steam will reach the sample tube via pinch-
solenoid valve(s), ventilation glass, PP-distributor and steam inlet tube. During this procedure
the level inside the steam generator will decrease. This is recognized by the level sensor and
the pump H2O for the filling of the steam generator will be switched on. If there is no sample
tube present or if the tap for cooling water is not open, a distillation is not possible.
Remark:
Especially if the water for the steam generator is not very clean or if there any kind of dirt
within the reservoir, the following problem may occur:
- After some distillations the instrument shows the error message „E3“, no water for steam
generator and the program stops.
- After pressing „RUN“ the display shows „H“, wait for steam.
Explanation:
If the steam generator is contaminated with dirt, the boiling and bubbling water will build up
foam above the water level. The level sensor detects the foam as water and prevents the
pump from running even though the water level in the steam generator is not sufficient. After
some time the liquid level or the foam will have diminished to such an extent that no more
liquid is detected, and the system switches on the pump. Within 15 s a sufficient water level
should now be detected. As this is not possible in such a short time the error „E3“ will appear
on the display.
If the „RUN“ key is now pressed, the system will get into the filling mode and the steam
generator will be filled to the height stipulated. However, at the same time, due to the cold
water, the steam pressure collapses, and the unit heats up again until stand-by is reached.
Only now further distillations can be run. So please note that a reaction as described is not a
malfunction, but a sign of unwanted foam in the steam generator. If this happens, clean the
steam generator and the reservoir.
8.2 Safety devices for steam generation
The level sensor constantly controlls and regulates the level of water inside the steam
generator.The sensor itself is constantly checked by the electronics; in case of a malfunction,
the instrument will give an error message. If the level detection system fails totally, the tubular
heating element will be switched off in case of a level decreasing under the allowed minimum
range.
The level sensor can only compare between insufficient or sufficient water inside the steam
generator. This is done by changing the frequency it sends to the Control-PCB.
In case that due to a total failure of the level sensor and the control electronics lack of water
in the steam generator cannot be detected, the tubular heating element must be protected
against overheating. In such a case the excess-temperature device will switch off the heating
circuit at about 150°C. After trouble shooting and cooling off of the steam generator the pin of
the excess-temperature device has to be pressed back into position. Now the next distillation
can be started.
The pressure sensor is placed directly on the Control PCB and continuously checks the
pressure inside the steam generator. If this sensor detects a higher pressure than 0,45 bar,
the tubular heating element is switched off, and the instrument gives an error message.
This switch ensures that a distillation is not run without a sample tube. In case you try to start
the system without a sample tube, the start is prevented by the switch. Thus there is no
danger of scalding.
8.3 Testing the safety devices for steam generation
To measure the frequency the gauge has to be connected to collection plug P4 at pins 2 + 3.
P4 is to be found on the Power PCB, above those orange coloured plugs.The occupation of
P4 is to be found in the circuit diagram (see 4.2).
P4 KHz
8
7
At instruments of the type Vapodest 40 the frequency can be checked with the inbuilt test-
.
program without opening the rear wall. To enter the test-program the keys „ “ and „ 9 “ have
to be pressed at the same time. The instrument shows the menu display for the I/O Test and
the frequency can be read:
The adjustment of the excess-temperature device should not be changed. It has been
adjusted here at C. Gerhardt so that an excess-temperature of about 150° C will open the
heating circuit. This device is directly switched into the the 230 V circuit. A damaged excess-
temperature device should not be bridged but replaced immediately.
The way to measure the pressure inside the steam generator is as follows:
VAP 20/30
When the Vapodest is still cold, a gauge for pressure measurement 0 - 1 bar has to be
connected into the line from steam generator to pressure sensor as the schematics shows.
If available, connect an instrument to measure the current the instrument draws into the
mains circuit.
Switch on the Vapodest. The electronics will measure the level of water inside the steam
generator and, if necessary, the pump for H2O will run. Then the heating element is switched
on.
When the pressure inside the steam generator begins to increase, the ventilation valve will
close, and the pressure of about 0.2 bar will build up. The instruments switches into stand-by
and indicates that it is ready for distillation.
If the instrument goes into stand-by well below or well above than 0.2 bar, the Control-PCB
has to be replaced (see 10.).
When measuring the pressure it is important to make sure that all tubing connections are
fixed properly. The connections should only be loosened when the instrument has cooled
down:
Risk of scalding because of hot steam of more than 100°C!
to steam valve
bar
to pressure sensor
steam generator
H2O steam generator
VAP 40
At instruments of the type Vapodest 40 the pressure can be checked with the inbuilt test-
.
program without opening the rear wall. To enter the test-program the keys „ “ and „ 9 “ have
to be pressed at the same time. The instrument shows the menu display for the I/O test. The
pressure as measured by the pressure sensor can now be read.
.
To quit the test-program the keys „ “ and „ Reset “ have to be pressed at the same time.
The adjustment of the excess pressure valve should not be changed. It has been adjusted
here at C. Gerhardt so that an excess-pressure of about 0.5 bar will activate the valve.
The following instructions will help to adjust the excess-pressure valve if it becomes
necessary nevertheless.
When the Vapodest is still cold, a gauge for pressure measurement 0 - 1 bar has to be
connected to the steam generator as the schematics show. The tubing to the pressure
sensor is left without any connection to prevent the switching off of the heating when reaching
a pressure of 0.2 bar.
If available connect an instrument to measure the current the instrument draws into the
mains circuit.
to steam valve
bar
to pressure sensor
X
steam generator
H2O steam generator
Switch on the Vapodest. The electronics measures the level of water inside the steam
generator and, if necessary, the pump for H2O will run. When a sufficient water level is
reached the heating element is switched on.
When the pressure inside the steam generator begins to increase the ventilation valve will
close, and the pressure is able to increase more and more because of the disconnected
pressure sensor. Observe carefully the increasing pressure, and make a note at which
pressure the excess-pressure valve opens. In case that the valve does not open at a
pressure of about 0.5 bar, immediately switch off the instrument!
Excess-pressure valve has to open earlier (lower pressure):
Remove the brass cap and slightly loosen the big nut. Turn the adjusting screw slightly anti-
clockwise, and tighten the nut very well. Take care that the adjusting screw stays in the
adjusted position. When the adjustment is done, repeat the above described procedure to
test the excess-pressure valve.
When measuring the pressure it is important to make sure that all tubing connections are
fixed properly. Carefully observe the increasing pressure and make sure you do not touch the
hot excess-pressure valve. The connections should only be opened when the instrument has
cooled down:
Risk of scalding because of hot steam of more than 100°C!
This switch prevents the start of a distillation without any sample tube. Without a tube the
switch is activated, which means that the circuit is open. With a sample tube the switch is not
activated, and the circuit is closed.The switch is located inside the instrument directly fixed on
the clamping device for the sample tube.
Steam inlet:
During stand-by steam of about 0.2 bar only reach the steam valve. During stand-by both
valves, steam valve and shut-off valve are dead. After starting a distillation, the steam valve
gets alive which means it will open. The ventilation valve however, still stays dead, so that it
is open for steam. The steam now can pass through the steam valve, ventilation glass, shut-
off valve, PP-distributor and steam inlet tubing into the sample tube. From there the distillate
will pass through the distribution head and the condenser into the receiver. The ventilation
valve of the condenser must be closed. The steam can also pass through the PP-distributor
until it reaches the pump for sample suction. Steam will also pass through the tubing for the
dilution of the sample until it reaches the non-return valve of the pump for H2O sample.
3 2 1
4
5
Steam Generator
6 7
b) Way of steam during distillation
3 2 1
4
5
Steam Generator
6 7
Sample suction:
After the distillation the steam valve will not draw current any longer, and it closes the path to
the ventilation glass. For suctioning now the shut-off valve will get alive and it closes the
tubing. too. Consequently the pump for sample suctioning is able to empty the sample tube.
In this case the ventilation valve of the condenser must open. If not, the pump will also empty
the receiver.
The pump for sample suction also draws against the pump for sample dilution.
3 2 1
4
5
Steam Generator
6 7
8 8 = Pump suctioning
9 = Pump H2O sample
9
sample tube
9.1 Diaphragm pump
To clean the pump it is necessary to disconnect all the tubes and to remove the rear wall.
Loosen the 4 screws from the badly working pump, and take away the cover of the pump
head. Also take away the connection-piece for angle connections and the diaphragm.
Carefully loosen the clamp for the valve plates and the valve plates themselves. Clean all the
parts and re-assemble the pump ensuring that all parts are put back into the right position.
Close the rear wall, and do not forget the ground connection! Install again all the tubings and
let the instrument run to see if everthing works well.
L
D
M1 / M2
( see 6. )
Also the peristaltic pumps are maintenance-free. At least once a year the neoprene tubing
should be replaced:
Description Part-No.
Plastic housing for tubing 10327
Spare tubing Neoprene 10370
Guides for the pump body. 10371
S-shaped locking bar 10373
Replacement of the Control PCB
With the new Vapodest 20/30 it is possible to exchange the Control PCB from a Vap 20 to a
VAP 30.
In case of a service it will now be easier to have the right PCB on stock because now it does
not matter any longer if the PCB is for a VAP 20 or for a VAP 30. The software will check
which keyboard is present, and then the processor will adapt the PCB.
In order to ensure that the software will work after the replacement of the Control PCB it is
necessary to press „PROG“ and „RESET“ at the same time and to switch on the instrument
while still pressing the keys.
Modification 16.03.1995
Draw.-No. 1120
Part.-No. 20903
The tubing with part-no.: 20903 is shortened to a length of 150 mm. When installing the
shortened tubing the angle has to be adjusted to the angle of the condenser outlet.
Draw.-No.: 1150
Part-No.: 20920 only VAP 30/40
The silicone-tubing part-no. 20914 from pump H2O to PP-distributor is replaced against
Verprene-tubing Part-No. 20920.
Additionally there is clamp Part-No. 17907 used. It fixes the new tubing to the PP-distributor.
The PVC-tubing between PP-angle connection (10317) and the PVC non-return valve (10329)
is replaced against Isoversinic-tubing part-no. 20918.
The other diaphragm pumps are not modified.
Additionally there are 2 clamps Part-No. 27101 used to fix the new tubing on both ends.
Due to modifications of the solenoid pinch valves the following kit is to be used for
replacement:
2. After modification ventilation glass part-no. 40269 is replaced by ventilation glass part-no
40272.(Part of modification kit)
As the technology of the new level-sensors has changed, they now have 4 contacts
instead of 3. If you have to install one of the new sensors/boilers into older
instruments, you will have problems with the connections at VAP 20/30. In order to
do so you have to modify the cabling between the power PCB and sensor plug.
Depending on the cabling of the sensor there are two possible ways of modification:
green
brown
1 white
P1, sensor PCB
P4, VAP 20-40
power PCB
ground, boiler
black
brown
1 blue
P1, sensor PCB
P4, VAP 20-40
power PCB
ground, boiler
As adaptor for the outlet of the steam generator a connection piece part-no. 17915
and a silicone-tubing(6x10 mm, lenght 45 mm) part-no. 20903 are used to connect
the Neoprene-tubing 20919.
Of the original Neoprene-tubing you have to cut off 45 mm.
Neoprene
20919
Connector
Neoprene-tubing 17915
Silicone
20903
The valve plates made of stainless steel (10328) are replaced by valve plates made
of PTFE No. 10327.(see 6.)