Overview
MobileDaRt Evolution is a mobile digital X-ray system which performs direct radiography on patient's
various body parts.
This operation guide briefly describes basic operation of MobileDaRt Evolution.
For detailed operation, refer to the relevant instruction manuals.
Instruction Manuals
l
This product includes the following manuals and documents.
Standard
1- Operation Manual (M503-E052)
- Operation Guide (M503-E056)
3- FUJIFILM DR-ID 800 Operation Manual
4- FUJIFILM DR-ID 800CL Operation Manual
5- FUJIFILM DR-ID 800CL Reference Guide
6- FUJIFILM DR-ID 800CL Reference Guide (Image Processing Parameters)
Option
A- FUJIFILM DR-ID 800CL Reference Guide (option)
lSymbols
The following symbols are used in this Operation Guide.
Symbols Meaning
Indicates a potentially hazardous situation which, if not avoided, may result in
! CAUTION minor to moderate injury or equipment damage.
Overview.......................................... 2
Contents ● Instruction Manuals ...............................................
● Symbols ................................................................
Preparation....................................... 7
● Turning the Power ON .......................................... 7
● Driving the Instrument ........................................... 7
Positioning........................................ 8
DR Radiography............................. 10
● Starting up the DR System ................................. 10
● Selecting/Creating Study Order ...........................11
● Confirming Radiography Mode ........................... 16
● Performing X-ray Exposure................................. 17
● Shutting Down the DR System ........................... 1
Cassette Radiography.................... 20
● Selecting Radiograph Method............................. 0
● Performing X-ray Exposure................................. 1
Image Output.................................. 22
● Automation output ...............................................
● Manual output .....................................................
Troubleshooting.............................. 26
● Emergency Stop ................................................. 6
● Releasing Emergency Stop ................................ 6
● If Unable to Travel Electrically............................. 7
● Replacement Time Display ................................. 8
● Warning Indicator Lamp ......................................
● Malfunction Indicator Lamp ................................. 30
System Components
Arm
Grip Bar
Storage Case
lCollimator
Status Indicator Lamp (^Page 6)
Irradiation Field
Adjustment Knob Irradiation Field
Adjustment Knob
Tape Measure
Battery Indicator
Status Indicator Lamp
(^Page 6)
Emergency Stop
Button
DR Button
USB Connector
Charging Lamp
Touch Panel
Monitor Power Line Connection
Indicator Lamp
lX-ray Control Panel FPD/Cassette
Radiography Ready Warning Radiography
Indicator Lamp Indicator Lamp Mode Button
Light OFF Blue light Green light blinking Green light Yellow light Blue light blinking
(arm is released) (returns to blue light)
! CAUTION
If the status indicator lamp is red, the code is displayed on the mAs display on the X-ray control
panel. If the status indicator is red, carefully read "6 Troubleshooting" in the Operation Manual and
take the appropriate measures.
Preparation
lTurning the Power ON
! CAUTION
! CAUTION
If electric travel is disabled due to an instrument malfunction, press the emergency brake release
button and manually move the instrument to a safe location.
For detailed operation, See "lIf Unable to Travel Electrically" (^ Page 7).
@ NOTE
• Maximum Travel Speed:Approx. 5 km/h (Maximum instantaneous travel speed: 6 km/h, may vary
depending on condition)
• When the lock is released, the driving speed becomes about half of the normal speed.
Positioning
radiography position.
Press the arm lock release lever or button View From A
to simultaneously enable column rotation,
arm extension/contraction, and vertical
movement of the X-ray tube unit.
Collimator Knob
Rotate the X-ray tube unit Angle
laterally. Markings
Pointer
and backward.
Angle
Markings
Pointer
Knob
DR Radiography
lStarting up the DR System
2 Press DR Button.
The DR system will start up.
DR Button
10
lSelecting/Creating Study Order
nTo select a study order (Worklist)
2 Press .
[Study List screen] is displayed.
3 Press .
The study order received from RIS is
displayed in the study list.
@ NOTE
When the wireless LAN option is not installed, be sure to connect the LAN cable to the LAN
connector for external connection.
11
4 Select the received study
order, and press .
[Exposure Menu Selection screen] is
displayed.
5 Press .
@ NOTE
If no exposure menu is registered in the Study screen when the system receives patient information
from the Worklist, it is necessary to select an exposure menu in [Exposure Menu Selection screen].
12
6 Press and make
sure that the study order
is in the waiting status.
Although the console (DR-ID 800CL)
is restarted, the waiting status will be
maintained.
@ NOTE
Please make sure to receive the study list in the "Waiting Study list", when moving the FDR Go after
receiving a study list from RIS.
Otherwise, the study list will be lost when the console (DR-ID 800CL) is restarted.
13
8 Select the flat panel
sensor to be used.
14
nTo create a study order (Manual)
15
4 Select the flat panel
sensor to be used.
switch holder.
X-ray Control Panel
The display of kV value and mAs value
starts blinking.
Status Indicator
@ NOTE
To perform radiography consecutively, wait until the status indicator lamp has been changed into
blue color from blinking to lighting.
17
4 The acquired image
is displayed. After
checking the image,
select to to
complete the study.
@ NOTE
Input causes, and then select to close the box. The relevant image will thus be handled as
mis-exposure image.
18
lShutting Down the DR System
2 Press .
The DR system will be shut down.
19
Cassette Radiography
lSelecting Radiograph Method
20
lPerforming X-ray Exposure
Status Indicator
@ NOTE
To perform radiography consecutively, wait until the status indicator lamp has been changed into
blue color from blinking to lighting.
21
Image Output
lAutomatic output
When the system is connected to the network, pressing on the Study screen will
automatically output images.
lManual output
For a study output previously or one suspended for QA application, output the images manually.
22
3 Press .
[Distribution Setting dialog box] is
displayed.
23
6 Press .
24
Shutting OFF and Charging
^ For procedures to shut down DR system, see Page 19.
lCharging Battery
@ NOTE
The (Charging lamp) blinks during charge and will be illuminated after the battery is completely
charged.
When turning ON the key switch, the (Charging lamp) will turn OFF.
25
Troubleshooting
lEmergency Stop
lReleasing Emergency
Stop
the arrow.
26
lIf Unable to Travel Electrically
If the brake release function is used, the brake is released to allow moving the unit by hand.
27
lReplacement Time Display
The system is equipped with a function to notify you of the replacement time for some parts.
kV Display
When any one of the following codes is indicated, replace the relevant part.
^ For detailed operation, refer to the instruction manual.
28
lWarning Indicator Lamp
If the warning indicator lamp illuminates, the warning code appears on the mAs display.
mAs Display
When any one of the following codes is indicated, take the prescribed action.
^ For detailed operation, refer to the instruction manual.
29
lMalfunction Indicator Lamp
If the self-diagnostic function detects any malfunction in the system, malfunction indicator lamp illuminates,
and radiography is disabled.
The malfunction code appears on the mAs display.
If the malfunction indicator lamp illuminates, contact your Shimadzu service representative.
Malfunction Indicator Lamp
mAs Display
However, when the following codes appear, take an appropriate action. Then the normal operation may be
restored.
^ For detailed operation, refer to the instruction manual.
30