Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                
100% found this document useful (1 vote)
485 views11 pages

Wiring Diagram - spr8-230.844.01.05.02

Download as pdf or txt
Download as pdf or txt
Download as pdf or txt
You are on page 1/ 11

X074J Wiring Diagrams

s
MOBILETT XP
WIRING DIAGRAMS
SPR8-230.844.01.05.02
Siemens - MED PROD & I+D SPR8-230.844.01.05.02 MOBILETT XP
C.T.S. Getafe Madrid, Spain Page 1 of 11
X074J
C.T.S. Getafe Madrid, Spain
Siemens - MED PROD & I+D

D908

L1

C1/R1 D916 Remote

R3-R6

D962 CABLE WINDER Z1


D972

REAR ZOOM
SPR8-230.844.01.05.02

D927 K1
C2

D952

D972

U1

U2

K22

HW Overview
RIGHT SIDE ZOOM
MOBILETT XP

MOBILETT
XP
Page 2 of 11

HW OVERVIEW
X074J
C.T.S. Getafe Madrid, Spain
Siemens - MED PROD & I+D

UNIT KEYBOARD
EXPOSURE
REMOTE CONTROL RELEASE USER CONTROL
SWITCH KV, mAs, Col

H1 SINGLE TANK
X28 X15
D991
D908 DISPLAY K3 DAP
X1 CHAMBER

XRMT X27 X12

D916 CPU D907 L1


XDAP
KV Temp COLIMATOR
&
SPR8-230.844.01.05.02

& mA
Pressure
Measurement
sensors
X8 X1

X20 K2

X2 X10 X11 K1

FIL X-RAY
D927 X7 TRAFO TUBE
POWER BOARD 11
X5 12

0
ANODE HT
I
X6 STATOR TRAFO
II

U1 V1

Functional Blocks
X20
D952 D972 D962
X20 CAPACITOR CAPACITOR L1
CHARGER BANK INVERTER
MOBILETT XP
Page 3 of 11

C1 PE
X074J
C.T.S. Getafe Madrid, Spain
Siemens - MED PROD & I+D

K22

1 2
3 4
7 8
X1A 5 6
1 2 4

X14
D927
X1B
Z1 4 F8 2
3 F7 15A 1
1 15A
X8 0
3 K1
4 I
1
PE A1
RC
2 1
A2
K22
X9
2 8 7
4 4 3
3
1

X4 K2 K1
3 6 U2 1
Switch on/off
SPR8-230.844.01.05.02

F3 F2 5A
Logic 1 1 +24VDC 3
4
1A
H1
Connectors
X13
PE PE
3

K1 Collimator
X7
1 1
1
F6 K2 X15
2
2 15A X15 K1
1 5 D916
3 3 U1 2 4
F2 5A 4 3
1 1 +5VDC 5 6 CPU
±15VDC 6 7
Heating 1
D907 Inverter
X5
1
11
2

3 +V
12
X3
D952 D972 D962
X3
3 7 F13 F14

Anode
9 20A 80A
Rotation 9 F1
20A
X6 ... F12
5 F5 7
II 3
C1
3 1 1
0 6A ... C12
5
1 2
I -V
61 71 3 1
X9

Connections
1 2
XCAP R3 R1 C1
MOBILETT XP

K22 R5
C2
R4 L1
Page 4 of 11

PE
R2
62 72
X074J
C.T.S. Getafe Madrid, Spain
Siemens - MED PROD & I+D

UNIT KEYBOARD

D908
+ - + -
mAs BUZZER kV mAs
kV X15
READY

COL EXP

X12

D916 PREPARATION
REQUEST
+V EXPOSURE
RELEASE
X27 SWITCH
SPR8-230.844.01.05.02

RESET BUZZER 3
SW3 μ CONTROLLER DATA BUS
EXPOSURE +V
REQUEST
RUN C167
+5 2
MODE RAM EPROM
SW2
SERVICE 1
+5 H1
X8 K2 p
60ºC
8 1
X
TEMP OR OIL PRESURE OIL
PRESSURE X>Y Y LIMIT 3 TEMPERATURE
9
ERROR
(ERR 3)
MOBILETT XP
Page 5 of 11

CPU
X074J
C.T.S. Getafe Madrid, Spain
Siemens - MED PROD & I+D

D972 D962 H1
F14 U+

100A
BOLTS AT
BOTTOM SIDE
L1
CAPACITOR
U1
KVC
BANK X-RAY
V5 C1 TUBE V (VDC)
450V DC C1
V1
3,5
! (350A)

R1

MEASUREMENT
CURRENT
U- 0
R+ S R-
D907
t (µs)
30μs 30μs
KV
MEASURE
RESONANT CURRENT AT MAIN INVERTER
SHORT EXPOSURE 70kV 5mAs
CIRCUIT
DETECTION
* WAVEFORM ONLY VALID FOR FIRST
PULSE TRIGGER
+ - + - + - + -
V (VDC) FC2
SPR8-230.844.01.05.02

GND
+12
GND
+5

3,25

t (µs)
22 21 20 μs
20 19 14 13 18 17 16 15 23 24 5 3,6 4
X20
X1
SET SERVICE PROGRAM 7 AND PRESS
EXPOSURE SWITCH TO PREPARATION

D927 X10 D916 * ADJUSTED BY P4 AT CPU D916

5,7
K4
KVC
SCM
13,14 ZERO
+15V CROSS X8

KV VALUE 1
1
SET BY CPU
KVS X=Y KV 3
1/3

KV REF X Y 1
2

P4
FREQUENCY Y TEST POINT DESCRIPTION SCALE
SETTING
(50kHz)
KVN NEGATIVE BRANCH KV 1V ≡ 10kV
Y>X X KVP MAX REFERENCE
KVP POSITIVE BRANCH KV 1V ≡ 10kV

Main Inverter
Y KV TOTAL KV 1V ≡ 30kV
MOBILETT XP

KVM 1
Y>X X KVN MAX REFERENCE KVS KV SET VALUE 1V ≡ 30kV
Page 6 of 11

SCM SHORT CIRCUIT MAIN INVERTER ACTIVE - HIGH


KVM KV MAXIMUM ACTIVE - HIGH
KVC 1V ≡ 100A
KVS

FC2

MAIN INVERTER CURRENT


XGND

KV
KVP
GND

KVN

SCM

KVM
KVC

FC2 MAX FREQUENCY INVERTER * SEE WAVEFORMS


HIGH: 5V LOW: 0V
X074J
C.T.S. Getafe Madrid, Spain
Siemens - MED PROD & I+D
D952 D927 H1
X3
7 F5

6A

X-RAY
X5 TUBE
1 11
450VDC

! 3 12

PE
2

1
P N
PE D907
SHORT X-RAY TUBE
CIRCUIT CURRENT
DETECTION MEASURE
FILAMENT
CURRENT
MEASUREMENT

12 14 13 16 1 2 2 1
X11 X1
SPR8-230.844.01.05.02

D916
JR X
MAX JR
VALUE Y X>Y 1
X8
IH X 1 1
6
MAX IH X>Y 1
Y 5
VALUE
JIM SCH I
SW
4
DISABLE
HEATING RMS

1
DISABLE PWM CONTROL
GAIN
ADJUSTMENT
ZERO G P5
IHS CROSS
IH JR
FILAMENT SET VALUE
JRS I SET I REAL 1
X-RAY CURRENT SET VALUE X=Y

X Y

Filament Circuit
SWITCH X-RAY/FILAMENT CURRENT TEST
FC1 DESCRIPTION SCALE
POINT
REGULATION
P3 mAs JR X-RAY TUBE CURRENT 1V ≡ 100mA
REACHED I FILAMENT CURRENT 1V ≡ 5A
MOBILETT XP

FREQUENCY
SETTING IH FILAMENT CURRENT (RMS) 1V ≡ 1,2A
SWT SWITCH CURRENT X-RAY/FIL ACTIVE: HIGH
(80kHz) F1
X=Y JRS X-RAY CURRENT SET 1V ≡ 100 mA
1V ≡ 1,2A
Page 7 of 11

VOLTAGE IHS FILAMENT CURRENT SET


mAs Y X COUNTER FREQUENCY JIM MAXIMUM CURRENT ERROR ACTIVE: HIGH
TARGET CONVERTER SCH ACTIVE: HIGH
SHORT CIRCUIT ERROR
SCH
SWT
IH

JIM
GND

JR
I

F1

JRS
XGND

IHS

FC1

mAs MEASUREMENT
X074J
C.T.S. Getafe Madrid, Spain
Siemens - MED PROD & I+D
R2

C2 7 μ F
400VAC
XCAP
1 2
START PHASE 2,5 seg BRAKE PHASE 10 seg
D952 D927
X3
7 F5

6A
H1 START BRAKE

X6 EXPOSURE
1 I TIME

V40 TF1 5 II M
450VDC
1~
TF2
! 3 0

V (VDC)
V41
1

3,8

CH2
6
SHORT
ACTIVE: CIRCUIT
HIGH DETECTOR 14
CH1
SPR8-230.844.01.05.02

t (ms)
+ - + - CH1: 5V / DIV I1
CH2: 1V / DIV I2

STARTING
ACTIVE: HIGH
+5
+5 +5

7 8 10 19 4 3 5 6
X11
V (VDC)

D916

CH2
ERAR
6

1,5
1 1
ENABLE H PULSES L PULSES ERROR CH1
POWER CONTROL CONTROL t (ms)
CIRCUIT I2 I1 CH1: 1V / DIV I1
I MAIN CH2: 1V / DIV I2

BRAKE
I AUX
PWM GENERATION
MOBILETT XP

TEST DESCRIPTION SCALE


Page 8 of 11

POINT
CURRENT AT MAIN WINDING 1V ≡ 0,25A
GND
XGND

ERAR

I1

Starter
I1
I2

I2 CURRENT AT AVX WINDING 1V ≡ 0,25A


ERAR SHORT CIRCUIT DETECTED HIGH: ERROR
X074J
C.T.S. Getafe Madrid, Spain
Siemens - MED PROD & I+D
H1: SINGLE TANK
L1: COLLIMATOR

D952 D927 H1 L1 tª
X3 THERMAL
X7 PROTECTION
7 F5 +24
1 1
6A

P1 24V DC K1

2 2 -24
450VDC LAMP: 24V 100W

!
1 P2

PWM
CONTROLLER

ENABLING P1

4 2
K2
SPR8-230.844.01.05.02

+15

10
X10
X8
9 7

D916
+5
IR REMOTE
CONTROL
RECEIVER
XRMT
20"
& 3
MAXIMUM 1≤ ACT
COLLIMATOR 14
TIME
DATA BUS

DRIVER

X12

UNIT KEYBOARD X15 D908


4 DRIVER
TO DATA BUS
MOBILETT XP

Collimator
COLIMATOR
USER 1
Page 9 of 11

CONTROL
DISPLAY
X074J
C.T.S. Getafe Madrid, Spain
Siemens - MED PROD & I+D
D927 POWER SUPPLY TO HEATING
& COLIMATOR & STARTER

F5 V40
6A

V41
+V
D952 X3 7 1 D972
X3
3
1

X3 F1 F13 F14
B1 PIN1 TO MAIN
7
(+) INVERTER
10A 20A 100A
D962

! V1 20A F1 ... F12


B2
180÷350VDC
(-)

C1... C12
12x10mF

9
SPR8-230.844.01.05.02

PWM 1/100
CONT
X9 DISCHARGE
3 RESISTORS NIN1
DISABLE
R3 R4 R5
1
1 X -V
X3
OVER VOLTAGE 2 71 72
X>Y Y
VALUE
X TEMP
5
SENSOR
OVER TEMP 1 61 62
+15V

XY Y
-15V
+5V

VALUE K22
SHORT
BE SURE CAPACITORS ARE
! DISCHARGED BEFORE HANDLE
THESE ELEMENTS
PE

3, 4, 6, 8,
1 11 22 10, 12, 21 3 7 9 V (VDC) 200 μs
X20
440V

D916

Capacitor Charger
CPU SW SW VC
POWER ERROR ENABLING
SUPPLY HANDLE
TEST POINT DESCRIPTION SCALE
VC CAPACITORS VOLTAGE 1V ≡ 100V
t (s)
STARTUP 3s
MOBILETT XP
Page 10 of 11

THE 3s PERIOD BETWEEN CHARGES IS


VC
GND
XGND

UNDER STAND-BY CONDITIONS.


IF YOU HAVE SHORTER TIME AT
STAND-BY PLEASE VERIFY D972.
ONE OR MORE CAPACITORS CAN BE
DAMAGED OR EVEN INPUT FUSE F13
CAN BE BLOWN.
X074J DAP
IONIZATION
CHAMBER
DAP

X1
D991
XDAP
L1
H1

K3
I1
DAP

I2
XDAP
2
6
3
+15V
RESET
CPU
D916
Siemens - MED PROD & I+D SPR8-230.844.01.05.02 MOBILETT XP
C.T.S. Getafe Madrid, Spain Page 11 of 11

You might also like