Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

ERBE Diathermy Setting Poster

Download as pdf or txt
Download as pdf or txt
You are on page 1of 1

•ERBE99791-ORPoster/GastroInt 09/18/2001 9:26 PM Page 1

ICC 200EA Gastrointestional


TM

Settings Chart
Common settings are shown on Units

Polypectomy – •Set AUTOCOAG to Forced


•Set Power: 15-30W
Forced Coag •Use Blue Footswitch for bleeders
Snare Resection •Patient plate on patient

•Set AUTOCOAG to Forced


•Set Power: 25-30W
•Set AUTOCUT to 200W
OPTION
•Effect 3, ENDOCUT ON
Polypectomy – Forced •Precondition with Blue Footswitch –
Coag precondition with Tissue will blanch
•Use Yellow Footswitch
ENDOCUT resection - Hold pedal down for ENDOCUT
resection
•Patient plate on patient

•Set AUTOCOAG to Forced


Hot Biopsy with •Set Power: 15-30W
•Biopsy accessory in Mono port
Forced Coag •Tap Blue Footswitch
•Patient plate on patient

OPTION •Set AUTOCOAG to Soft


Hot Biopsy with •Set Power: 60-70W
•Biopsy tool plugs in Mono port
Soft Coag •Use Blue Footswitch
Reduces Sticking Preserves Pathology •Patient plate on patient

•Set AUTOCOAG to Bipolar


TM •Set Power: 15-30W
BiCap Bipolar/Gold •Accessory into Bipolar Plug
Probe •Use Blue Footswitch
•No patient plate

Over 150 Years of World-Leading Medical Technology 1-800-778-ERBE


LIT/5033/00 10/01

You might also like