MINT1400A: Medical - Universal Input Power Supply

Download as pdf or txt
Download as pdf or txt
You are on page 1of 3

MINT1400A

400W Single Output Series


Medical – Universal Input Power Supply 2 Year Warranty
•Compact 3.94” x 7” x 1.58”, 1U compatible
•9.5 W/in³
•Up to 250W Convection Rating
•Efficiency up to 91%
•12Vdc to 48Vdc outputs, up to 400W
•Remote ON/OFF
•Safety approved to Medical and ITE Standards
• EN/IEC/UL60601-1 and CSA 601.1
•CSA/EN/IEC/UL60950-1, 2nd Edition
•Compliant to EN55011/22 Conducted Noise, Class B
•RoHS Compliant – Category 6

Specifications All Specifications are typical at nominal input, full load at 25°C unless otherwise stated.
AC Input 85-270VAC, 47-63 Hz single phase Turn On Time 500mSec maximum
Input Current 120VAC: 4.2A, 230VAC: 2.1A Hold-up Time 20 ms at 400W
Inrush Current Active circuit, 37A max Overvoltage Protection See chart
Input Fuses F1, F2: 6.3A Overload Protection Cycling type, 470W typical
Power Factor 0.99 typical, 0.96 minimum Overtemperature Protection Provided on all models, auto-
Efficiency Typical 88% @ 120VAC, 91% @ 240VAC recovery. See page 2 for cooling information
Remote Sense Compensates for up to 500mV Switching Frequency Output: 170kHz ± 8kHz;
drop on output leads PFC: 90kHz, ± 8kHz; Bias: 100kHz, ± 6 kHz
Output Power 400W with fan cooling. See derating chart for other Safety Standards Medical: EN/IEC/UL 60601-1, CSA 601.1
conditions. Peak: 450W for 1 min., 10% duty cycle ITE: CSA/EN/IEC/UL 60950-1, 2nd Edition
Ripple and Noise 1% pk-pk max., 20MHz BW, differential mode Isolation Input-Output: 4000VAC,
Input-Ground: 1500VAC, Output-Ground: 500VAC
Output Voltages See chart. Voltages adjustable +/- 5% Earth Leakage Current 100µA at 120VAC, 60Hz Normal.
400µA at 264Vac, 60Hz SFC.
Total Regulation See chart Operating Temperature 0° to 70°C, refer to derating
curve for ratings >40°C
Minimum Load Not required Storage Temperature -40 to +85°C
Transient Response 500µs to 1%, 0.2A/µs, 3.5% Deviation Relative Humidity 5% to 95%, non-condensing
Shock and Vibration Shock: MIL-STD-810E, 516.4 part IV Altitude Operating: -500 to 10,000 ft.
Vibration: 5Hz – 500Hz, 2.5 Grms Non-Operating: -500 to 40,000 ft.

Model Number Key


MINT 1 400 A 12 10 L 01
01 = Standard Model, 02 and higher indicates a m odified m odel.
"L" = Input Connector - EBY P/N 5002-03-N-12
"10" = Output Connections - Bus Bars w ith #10 screw s
Output Voltage: "12" = 12V output, "24" = 24V output, etc.
Model Configuration: "A" = Original Configuration; "C" = Optional fan cover included
Output Pow er (Watts) - "400" = 400W
# of Outputs
Product Fam ily: "M" = Medical, "I" = Internal, "NT" = New Technology

______________________________________________________________________________________________________________________________ 1 of 3
SL Power Electronics Corp • 6050 King Drive Ventura, CA 93003 • Phone 805.486.4565 • Fax 858.712.2040 • Email: info@slpower.com • www.slpower.com
MINT1400A
400W Single Output Series
Medical – Universal Input Power Supply 2 Year Warranty

Output Parameters
Model Output Output Current Ripple Total
Number Volts (V) Fan Cooled Convection & Noise Regulation

MINT1400A1210L01 12V 33.0A 21.0A 1% 2%


MINT1400A2410L01 24V 16.7A 10.4A 1% 2%
MINT1400A2810L01 28V 14.3A 8.9A 1% 2%
MINT1400A4810L01 48V 8.30A 5.2A 1% 2%

Signals
Inhibit: DC OK:
TTL LOW to inhibit HIGH when output is >90% of nominal
PS Off: Fan Output:
TTL HIGH to inhibit 12Vdc @ 250 mA.
Pwr OK 5V Standby Voltage:
TTL HIGH when AC is OK, 100-200 mS delay. Always on, up to 100mA.
Pwr OK
PF LOW 4 mS warning before loss of output regulation.

EMI/EMC Compliance
Conducted Emissions EN55011/22 Class B
Radiated Emissions EN55011/22 Class A
Voltage Fluctuations EN61000-3-3, complies
Static Discharge Immunity EN61000-4-2, 6kV Contact Discharge, 8kV air discharge
Radiated RF Immunity EN61000-4-3, 3V/m.
EFT/Burst Immunity EN61000-4-4, 2kV/5kHz..
Line Surge Immunity EN61000-4-5, 1kV differential, 2kV common-mode
Conducted RF Immunity EN61000-4-6, 3Vrms
Power Frequency Magnetic Field Immunity EN61000-4-8, 3A/m
Voltage Dip Immunity EN61000-4-11 criteria A @ 250 Watts or less.
Line Frequency Harmonics EN61000-3-2, Class A

Output Power Derating Curve

MINT1400 Output Power Derating Chart


450
Output Power (Watts)

400
350 Note: For convection cooling, the
300 power supply is required to be
250 mounted on a heatsink or system
200 chassis with a minimum of 150 in² of
Fan Cooled surface area.
150
Convection
100
50
0
0 10 20 30 40 50 60 70
Ambient Temperature (deg C)

______________________________________________________________________________________________________________________________ 2 of 3
SL Power Electronics Corp • 6050 King Drive Ventura, CA 93003 • Phone 805.486.4565 • Fax 858.712.2040 • Email: info@slpower.com • www.slpower.com
MINT1400A
400W Single Output Series
Medical – Universal Input Power Supply 2 Year Warranty

Outline Drawing & Connector Information

Connector Information

Data Sheet © 2011 SL Power Electronics Corp. The information and specifications contained herein are believed to be correct at the time of publication. Rev. 6-7-2013
However, SL Power accepts no responsibility for consequences arising from reproduction errors or inaccuracies. Specifications are subject to change without notice.

______________________________________________________________________________________________________________________________ 3 of 3
SL Power Electronics Corp • 6050 King Drive Ventura, CA 93003 • Phone 805.486.4565 • Fax 858.712.2040 • Email: info@slpower.com • www.slpower.com

You might also like