PC3400-750 Operations Manual
PC3400-750 Operations Manual
PC3400-750 Operations Manual
PC3400
Contents
SAFETY PRECAUTIONS ............................................................................................................................. 4
WARRANTY INFORMATION........................................................................................................................ 5
1.0 GENERAL INFORMATION .............................................................................................................. 6
1.1 BEFORE GETTING STARTED.................................................................................................... 6
1.2 UNPACKING UNIT ...................................................................................................................... 6
1.3 UNIT LABELING .......................................................................................................................... 7
1.4 FEATURES AND BENEFITS ....................................................................................................... 7
1.5 SPECIFICATIONS ....................................................................................................................... 8
1.5.1 WIRING SPECIFICATIONS .................................................................................................. 9
1.6 BASIC THEORY OF OPERATION ............................................................................................ 10
2.0 INSTALLATION.............................................................................................................................. 13
2.1 WEIR ASSEMBLY ..................................................................................................................... 13
2.2 MOUNTING ............................................................................................................................... 14
2.3 WIRING...................................................................................................................................... 15
2.3.1 WIRING OF MAINS POWER .............................................................................................. 16
2.3.2 WIRING OF RS-485 COMMUNICATIONS.......................................................................... 17
2.3.3 WIRING OF ETHERNET COMMUNICATIONS .................................................................. 18
2.3.4 WIRING OF ANALOG I/O .................................................................................................... 18
2.3.5 ANALOG INPUT CONFIGURATION................................................................................... 19
2.4 SAMPLE SETUP ....................................................................................................................... 20
2.4.1 SAMPLE CONSIDERATIONS ............................................................................................ 20
2.4.2 SAMPLE TAPS ....................................................................................................................... 20
2.4.3 FLOW CONTROL .................................................................................................................... 21
2.4.4 FLOW CONTROL W EIR ASSEMBLY .......................................................................................... 21
2.4.5 PROBLEMS WITH AIR BUBBLES ............................................................................................... 22
2.4.6 INITIAL STARTUP .................................................................................................................... 22
3.0 PARTICLE COUNTER OPERATION ............................................................................................. 23
3.1 USER INTERFACE (KEY PAD) ................................................................................................... 23
3.2 READOUT / DISPLAY SCREENS ............................................................................................. 24
3.2.1 MAIN DISPLAY SCREEN .......................................................................................................... 24
3.2.2 TOTAL COUNTS SCREEN ........................................................................................................ 24
3.2.3 TREND GRAPH ...................................................................................................................... 24
3.2.4 STATUS SCREEN ................................................................................................................... 25
3.3 USER MENU / SETUP SCREENS ............................................................................................ 26
3.3.1 USER LOGIN .......................................................................................................................... 26
3.3.2 SAMPLE SETUP ..................................................................................................................... 27
3.3.3 GENERAL SETUP ................................................................................................................... 28
3.3.4 SERVICE SETUP .................................................................................................................... 30
3.4 RS-485/RS-232 COMMUNICATION AND POWER INDICATORS ........................................... 31
3.5 CLEANING PROCEDURE......................................................................................................... 32
3.5.1 Non-brush Cleaning ............................................................................................................ 33
3.5.2 Brush Cleaning .................................................................................................................... 33
3.5.3 Replacing Tubing ................................................................................................................ 33
4.0 DOWNLOADING DATA ................................................................................................................ 34
4.1 INSTALLING AND SETTING UP TRACDATA SOFTWARE .................................................... 34
4.2 CONNECTING RS-232 DOWNLOAD CABLE .......................................................................... 35
4.3 DOWNLOADING DATA............................................................................................................. 36
4.1 EXPORTING DATA ................................................................................................................... 37
5.0 RECOMMENDED SPARE PARTS LIST....................................................................................... 38
5.1 ORDERING SPARE PARTS...................................................................................................... 38
APPENDIX A .............................................................................................................................................. 39
USER MENU MAP .................................................................................................................................... 39
MODEL PC3400 2
LIST OF FIGURES
Figure 1 Light Obscuration Measurement (Diagram 1) .............................................................................. 10
Figure 2 Particle Size Comparison ............................................................................................................. 11
Figure 3 System Components.................................................................................................................... 12
Figure 4 PC3400 and Flow Control Weir Plumbing .................................................................................... 12
Figure 5 Weir Assembly .............................................................................................................................. 13
Figure 6 System Dimensions ...................................................................................................................... 14
Figure 7 115/230V and Power Switches ..................................................................................................... 15
Figure 8 Warning Label ............................................................................................................................... 15
Figure 9 Power Terminal Block ................................................................................................................... 16
Figure 10 RS-485 Communications ............................................................................................................ 17
Figure 11 Optional MODBUS TCP Communications ................................................................................. 18
Figure 12 Analog I/O Wiring Terminals ....................................................................................................... 18
Figure 13 Analog Input Type Jumper Selection .......................................................................................... 19
Figure 14 Optional Analog Input Card ........................................................................................................ 19
Figure 15 Sample Tap................................................................................................................................. 20
Figure 16 Sensor Flow Rate Adjustment .................................................................................................... 21
Figure 17 Optional Analog Output Board Adjustment ................................................................................. 30
Figure 18 Sensor Cleaning Methods .......................................................................................................... 33
MODEL PC3400 3
SAFETY
In order to provide maximum user safety the PC3400 particle counter was designed with the following
issues in mind.
All electrical circuitry is enclosed within a protective non-conductive housing.
No user serviceable parts.
No user replaceable parts.
No access to laser.
SAFETY PRECAUTIONS
It is important to review this list of precautions prior to installation.
Do not attempt to disassemble the unit.
Water must not be allowed to enter the housing of the unit.
Close and fasten shut the door of the unit prior to any external cleaning to prevent water ingress.
Do not drop or jar the unit.
Use a mild non-abrasive cleanser when cleaning the outer cover of the unit.
Do not modify any internal electrical wiring or electronics.
The PC3400 sensor is equipped with a class III laser and should not be tampered with in any
way. The following warning label is affixed to the laser assembly housing inside of the unit.
In order to provide maximum user safety this instrument was designed with all electrical circuitry
enclosed within a protective non-conductive housing. The label below will be visible at any
location where high voltage is present.
MODEL PC3400 4
WARRANTY INFORMATION
Chemtrac, Inc. warrants its equipment to be free from defects in material and workmanship for a period of
one (1) year from date of shipment to the original purchaser. Upon receipt of written notice from
purchaser, seller shall repair or replace the equipment (at option of Chemtrac, Inc.).
Chemtrac, Inc. assumes no responsibility for equipment damage or failure caused by:
This warranty represents the exclusive remedy of damage or failure of equipment. In no event shall
Chemtrac, Inc. be liable for any special, incidental, or consequential damage such as loss of production
or profits.
Chemtrac, Inc
6991 Peachtree Industrial Blvd., Building 600
Norcross, GA 30092
MODEL PC3400 5
1.0 GENERAL INFORMATION
1.1 BEFORE GETTING STARTED
NOTE: Do not plug in or power up this unit without first verifying the 115/230 VAC Select Switch
is in the proper position (see figure 7). In order to insure proper operation of the PC3400, it is
important to follow the installation guidelines outlined in this manual. Individual plant requirements and
layouts will vary, resulting in alterations to the procedures.
The particle counter is one of the most sensitive and precise on-line instruments in use in water treatment
today. In order to achieve and maintain optimal performance, it is imperative that proper installation and
operational procedures be followed.
The sensitivity of the unit to small concentrations of particles well below the range of visibility requires that
the sample be delivered to the sensor without contamination or alteration. Data produced is based on the
sample volume, so flow rate must be maintained at a constant value. The flow path through the sensor is
small (750 um x 750 um) in comparison to other types of on-line instruments. Strainers should be used
on raw water samples, and a proper maintenance schedule must be followed. Since these instruments
count and size particles, each counter produces several channels of data. This data is most useful when
trended, so provision should be made for recording and storage.
The sensor is designed to detect particles in the 2 to 750 µm range (upper limit defined by flow cell size),
and size particles in the 2 to 125 µm range (any particles larger than 125 micron will be sized as >125
µm). The PC3400 uses a light blockage method to determine particle size and count. The light source is
an infrared laser diode.
The analyzer features an entirely external sampling and measurement system which prevents sample
leaks from damaging sensitive electronic components. The 750 µm x 750 µm flow cell is easy to access
and clean. The constant head, flow control weir ensures stable, reproducible flow and accurate particle
counts.
The PC3400 has a LCD backlit display which allows the user to see the individual size ranges and
particle counts in each range. The display also provides the cell condition readout, which is an indication
of the sensor’s cleanliness and allows the user to know when cleaning is necessary. It is recommended to
clean the sensor when the cell condition drops by more than 10% from the reading obtained after the
previous cleaning.
Integration to the customer’s SCADA / DCS can be accomplished either via analog 4-20 mA outputs (4
output channels), or by using the digital communication features. RS/485 and RS/232 communication
ports are standard on the PC3400, and an Ethernet connection is optional. Using these communication
ports, the customer can communicate to the PC3400 via Modbus (RTU, or TCP). Option for Profibus is
also available (contact factory for more details).
The PC3400 has a built in data logging feature that allows the unit to record over 60,000 sample strings.
A single sample string includes size bins, counts in each bin, cell condition, laser condition, and analog
inputs (optional). The data is downloaded to a computer using Chemtrac’s data download utility software.
Once downloaded, the data can be exported to a spreadsheet program such as Excel. For instant trend
viewing, a graphical trend of the last 60 recorded readings is also provided on the unit’s display.
Another option for viewing and collecting the data from a one or several PC3400’s is to use Chemtrac’s
TracWare software. TracWare also provides the user with the ability to customize particle size ranges,
change sampling duration and frequency, plot trends, and prepare reports. The user must provide the
personal computer for running TracWare.
Another unique feature of the PC3400 is its ability to perform Automatic Sizing Calibration via the user
menu and front panel keypads. No software or computer is required to calibrate the PC3400.
Note
Ninety-six particle counters can be multi-dropped on a RS485 serial communications loop. The
maximum network length is 4000 ft. To avoid problems with communications, be sure to contact
the factory to discuss how the RS/485 network will be laid out in the plant. A poorly designed
network can result in excessive communication timeouts.
MODEL PC3400 7
1.5 SPECIFICATIONS
GENERAL
MODEL PC3400 8
SAMPLE REQUIREMENTS
CALIBRATION
Conductor AWG: 24
Conductor Stranding: (7x32)
Conductor DCR: 78.7 ohms/km
Conductor Type: TC
Description:
Tinned copper, polyethylene insulated, twisted pairs. Overall Beldfoil® aluminum-polyester
shield. 24 AWG stranded tinned copper drain wire. Overall tinned copper braid shield (90%
coverage). Chrome PVC jacket (100% coverage). 30V 80° C
MODEL PC3400 9
Power Cable (not supplied with units shipped to EU)
3-Core Round
CMA Ref. No. 2182Y
Bar Code No. H03VV-F
BASEC approved
CSA: 0.5mm″
Current Rating: 3A
Voltage Rating 300/300/V
Nominal Dia.: 5.8mm
The particle counter consists of an optical sensor and counting electronics. The sensor consists of a 780
nm infrared laser diode, a light detector, and two transparent "windows". The detector converts light
energy into electrical voltage. The laser light is directed through both windows and on to the detector.
Sample flow passes between the windows, so that any particles in the sample will pass through the laser
beam. Each particle will block a small percentage of the light that hits the detector, changing the
electrical output of the detector. Since the particles are moving rapidly, each particle will produce a short
electrical pulse at the output of the detector. The amplitude of the pulse correlates to the size of the
particle. The output of the sensor is a stream of pulses of varying amplitude, each corresponding to a
particle. This type of device is known as a light blocking or light extinction sensor, see Figures 1.
MODEL PC3400 10
The amount of light blocked from the detector is the sum of the light absorbed by the particle and the light
scattered or reflected by the particle. The size and composition of each particle will determine how much
light is scattered and how much is absorbed. Carbon particles will absorb most of the light and scatter
very little of it. Organic particles have an index of refraction close to the value of water, and tend to
refract more light. The result of this is that an organic particle 5 microns in size will block less light than
an inorganic particle of the same size, and will appear smaller to the particle counter. For this reason,
Giardia and Cryptosporidium will be counted in size ranges several microns below their actual size. The
orientation of the particle as it passes through the light beam will also affect how much light is blocked.
These factors make it necessary to count particles over a range of sizes as opposed to an exact size.
The sensor output is fed into the counter electronics, which sorts the pulses according to amplitude and
counts them. In a typical application, three or four ranges might be used. The first one might be set to
count particles in the range from 2 to 5 microns, the second 5 to 10, the third 10 to 15, and so on.
The Model PC3400 typically operates at a flow rate of 75 ml/min. The Flow Control Weir, which is a
constant head overflow device, is used to maintain this flow. Flow adjustment is accomplished by moving
the drain cup up or down the weir assembly, which decreases or increases flow (See Section 4.1.3).
Other flow rates can be used and are set via the user menu. However, the instrument is calibrated at the
factory for a flow rate of 75 ml/min and the sizing accuracy will not be accurate if that flow rate is changed
by more than 10% and the unit is not recalibrated at the new flow rate. If your application requires the
flow the sensor to be something different than 75 ml/min, then contact the factory for assistance.
MODEL PC3400 11
Laser Diode
Analog Board Main Processor
(Optional) Board
and / or
MODBUS TCP
Board (Optional)
I/O Connectors
Sensor
Flow Cell
MODEL PC3400 13
2.2 MOUNTING
See Figure 6 below for mounting details. The total space needed to mount the PC3400 is an area 23
inches wide and 43 inches tall. The PC3400 should be permanently mounted in an upright position using
the mounting holes on the back of the counter. The mounting hole dimensions are 6.0 inches horizontally
and 10.75 inches vertically.
The weir should be mounted using the supplied hardware (weir mounting arms) and the weir height
adjusted so that he top of the weir is approximately 10 to 12 inches above the top of the PC3400.
Once the PC3400 and Flow Control Weir are mounted, the tubing can be attached to the Weir and
Sensor as detailed in Figure 4. Make the sensor tubing going between the sensor and drain cup long
enough to reach the drain cup with it slid all the way down on the weir. It is also recommended to leave a
little extra tubing on the bottom of the drain cup. Final adjustment to the drain cup position and tubing
lengths will be made in section 4.1.3.
MODEL PC3400 14
2.3 WIRING
WIRING SAFETY - Please observe the following safety precautions prior to wiring the unit.
Power must be locally fused or switched prior to entering the unit.
Power should be hardwired to the unit. Flexible power cords are not recommended around
water.
Disable power circuit prior to hardwiring the unit.
Observe polarity when connecting the unit. The unit will not work and can be damaged if polarity
is crossed.
All wire connections must be tinned prior to connecting to ensure proper contact and to prevent
corrosion.
The unit must be connected to an Earth Ground. Note: Poor grounding can produce electrical
noise in the instrument which could affect the instrument’s accuracy.
If using the original power cord supplied with your unit, make sure the proper AC voltage is set on
the 115/230 VAC Select Switch. Once checked for proper AC power requirements, remove
WARNING! Label to gain access to the power ON/OFF switch (see Figures 7 and 8).
NOTE
The particle counter is one of the most sensitive and precise on-line instruments in use in water
treatment today. In order to achieve and maintain optimal performance, it is imperative that proper
installation and operational procedures be followed.
MODEL PC3400 15
2.3.1 WIRING OF MAINS POWER
Following the installation, guidelines outlined in this manual should result in proper operation with minimal
problems. Individual plant requirements and layouts will vary, resulting in alterations to the procedures.
1. Open the enclosure by first unscrewing all 4 screws on the front cover.
2. Locate the 115/230 VAC select switch and verify it is in the proper position for your
mains power.
3. Remove the 4 screws that hold the aluminum protective cover and unplug the 3
position green connector to gain better access to the connector screws.
4. During installation, a local isolation point must be supplied.
5. Disable the power circuit that will be used to supply power to the unit.
6. With the power off, feed the electrical supply line through the compression fitting on
the bottom right side of the unit.
7. Electrical power should be connected as shown in Figure 9. On the PC board, under
the green connector, it is labeled from left to right, “NEU”, “GND”, and “HOT”
NOTE
When feeding wire into unit it is always important to feed through the compression fitting that is
closest to that wire connection.
• Insert the neutral wire into the left terminal labeled “NEU”.
• Insert the ground (earth) wire into the middle terminal labeled “GND”.
• Insert the hot wire into the right terminal labeled “HOT”.
HOT
NEUTRAL GROUND
After wiring of mains power is complete, and after verifying the 115/230 VAC select switch is in the right
position, enable the external power circuit and flip the “On/Off” switch to the “On” position and verify the
PC3400 display comes on.
MODEL PC3400 16
2.3.2 WIRING OF RS-485 COMMUNICATIONS
If utilizing RS-485 Communications, connect the RS-485 Cable to the RS-485 terminals shown on the
following page. Ensure shielding is tied together and connected to ground at one end.
TX – RX - GND TX + RX +
MODEL PC3400 17
2.3.3 WIRING OF ETHERNET COMMUNICATIONS
The optional MODBUS TCP card with RJ45 connector (PN: 34791) is required to utilize the optional
Ethernet Communication capability. The location of the MODBUS board and RJ45 connector are shown
below. Route the CAT5 cable with RJ45 male connector through the strain relief fitting located on the
bottom of the PC3400 enclosure and connect it to the female RJ45 connector as shown.
RJ45 Connector
MODEL PC3400 18
2.3.5 ANALOG INPUT CONFIGURATION
The PC3400 provides for up to eight analog inputs with optional Analog Input cards. (PN: 34761) Each
card provides four inputs. The analog inputs can be configured for either 0 - 5 VDC, 0 - 10 VDC or 0 - 20
mA (4 – 20 mA). The input type is selected via the General Setup -> Outputs & Alarms menu. (See
3.3.3.4) Once the input type is determined, jumper settings on the analog input card must be configured
to match the input type. See Figure 13.
Information from the analog inputs is available over any of the digital communications options (RS232,
RS485, Modbus TCP). In addition, analog input data, for the first 4 input channels, is logged in the
instruments memory. The data is logged based on the interval selected in the General Setup -> Log
Frequency menu. It is also possible to select one of the analog input signals for trending on the Trend
Graph. (See 3.3.3.1)
MODEL PC3400 19
2.4 SAMPLE SETUP
2.4.1 SAMPLE CONSIDERATIONS
The high sensitivity of the particle counter makes sample handling and delivery critical for proper
operation. Poor sample point selection and/or improper sample line installation can result in a substantial
loss of accuracy.
It is not difficult to achieve proper sample delivery in virtually any treatment plant. The important thing is
to make sure the sample is properly representative of the process stream. There are three ways in which
the sample can be distorted:
1. Adding particles to the sample stream. Choosing an improper location for the sample tap, such
as the bottom of a pipe where sediment can accumulate, or an open sample point where particles
can be introduced from outside the process usually causes this.
2. Losing particles from the sample stream. Long sample lines can cause particle dropout,
especially at low flow rates. If the sample line must be long (more than 10 or 15 feet) then it will
be important to keep the sample flowing swiftly (i.e. 4 ft per second) to prevent particulate from
settling out in the tubing. The higher flow rate will also help keep the tubing cleaner and thereby
help minimize particle shedding which can be caused by fluctuations in flow.
3. Altering the particle distribution. Sample pumps can chop up larger particles creating more small
particles. Valves and other obstructions can cause "shedding". Particles collect on cracks and
edges and then break loose later. "T" fittings can cause larger particles to miss the sample line
because they can't make the sharp right angle "turn".
Of course, it is seldom possible to avoid all of these pitfalls. Frequently valves are needed to control
sample flow rates, or pumps are used to get the sample to a usable point. The important thing is to
minimize these problems. Filtered water can be pumped with fewer problems than raw or settled, since
only very small particles are usually present, and they are less likely to be broken up. Ball valves or other
types with minimal edges to trap particles can be used. "Y" fittings can be used instead of "T's".
Make sure that the sample point has the minimum pressure available at all times to provide enough flow.
The Chemtrac sensor requires approximately 2 feet of head to maintain a 75-ml/min flow, so ensure there
will be a minimum of 2 feet of head throughout the entire filter run (i.e. during high head loss towards the
end of filter run).
The flow rate through the sensor is adjusted by moving the Drain Cup up or down the vertical pipe until a
75 ml per minute flow is achieved. Measure the flow rate using a graduated cylinder and a stopwatch.
Once the flow is set, the flow will remain stable as long as some amount of overflow is maintained at the
top of the weir. As seen in figure 6, the head required to achieve the 75 ml/min flow rate is approximately
24 to 30 inches (as measured from the clear horizontal pipe down to drain cup’s tubing holder). Changes
in water temperature, as well as the sensor & tubing getting dirty, will affect the flow rate.
As part of the regular maintenance cycle, periodic checks should be made to the flow rate. If the flow rate
is found to be low, try replacing the sensor tubing if it looks dirty and clean the sensor as outlined in
section 3.5. If this maintenance action does not achieve the proper flow rate, then adjust the Drain Cup
up or down the weir to obtain the proper flow rate of 75 ml/min.
The weir assembly and PC3400 should be mounted at an elevation sufficient to achieve overflow at
minimum source pressure. A ball valve can be used to regulate the flow into the weir assembly.
IMPORTANT NOTE ON WEIR INSTALLATION
MODEL PC3400 21
Weir arms are provided to help simplify the mounting of the Weir. However, some users prefer to mount
the Weir directly to a back plate versus using the Weir arms. Not using the weir arms can make it easier to
remove the PC3400 if the equipment needs to be serviced. If the weir arms are not to be used, just
ensure two things:
1. Make sure the clear horizontal pipe at the top of Weir is higher than the top of the PC3400 by 10 to 12
inches. This ensures flow will start reliably through the sensor. If the Weir is mounted too low, the flow
may not re-start reliably through the sensor if flow was ever lost for any reason.
2. Mount the Weir to the right of the unit (as shown in figure 16) and insure there is a minimum of 4 inches
between the Weir and the sensor (the sensor protrudes from the top right side of the enclosure). The cap
on the sensor may need to be removed at some point for maintenance and mounting the Weir too close to
the sensor will make it impossible to remove the cap without first removing the weir.
The flow control weir is designed to aid in the removal of air bubbles. However, air bubbles that have
formed due to degassing are difficult to remove using the flow control weir. In some cases, the sample
may need to be kept pressurized through the sensor to avoid the formation of bubbles. Contact the factory
for assistance if you suspect air bubbles are interfering with the measurement. Air bubbles forming in the
sensor tubing is a possible indication that air bubbles may be a problem.
Once installation is complete and the sample is hooked up to the sensor, the sample lines may still be
somewhat dirty and not truly representative of the actual particle concentration at the sample point.
Typically the sample lines will sufficiently flush and stabilize within 15 to 30 minutes. If the particle counter
output is being trended, it is easy to tell when the lines have been sufficiently flushed, as the data will level
out.
Once the counts have appeared to stabilized, a quick test can be done to verify the PC3400 is not
experiencing any sort of noise issue that can arise to due to issues with poor electrical grounding. Simply
pinch the tubing coming off the top of the sensor (which is the tubing going to the drain cup). Pinching this
tubing will stop flow through the sensor. If the flow is stopped and the tubing held stationary, the counts
should drop to zero within 1 to 2 minutes providing the sample is filter effluent (use DI water if filter effluent
is not available or is higher than 0.3 NTU).
With water flowing through the sensor, a check should be performed to make sure the PC3400 has a laser
and cell condition reading of 90 to 110% as seen on the Status screen. Try cleaning the sensor as detailed
in section 3.5 if the cell condition is below 90%. Contact factory for assistance at 770-449-6233 if the cell
condition does not come up after cleaning, and/or if the laser is outside of the 90 to 100% range. NOTE:
Laser and Cell condition are temperature dependent, so water and/or air temps near freezing or above 100
degrees F can cause these readings to change. The PC3400 is working properly as long as the laser and
cell condition reading are in the 90 to 100% range with water and air temps in the 40 to 90 degree F range.
At this point the unit should be ready for user customized menu setup and operation.
MODEL PC3400 22
3.0 PARTICLE COUNTER OPERATION
3.1 USER INTERFACE (Key Pad)
The PC3400 is equipped with user interface keys, and LED indicators. The interface
keys are up (), down (), left (), right (), and the Menu and Enter function keys.
On power up, you see the screen below. In the Appendix of this manual there is a flow
chart of display and menu screens.
= UP key
= DOWN key
There are three LED lights to the left of the graphical display. These are:
Power –Red LED that illuminates when there is power to the unit.
TX – Yellow LED that illuminates when the unit is transmitting data.
RX – Green LED that illuminates when the unit is receiving data.
MODEL PC3400 23
3.2 READOUT / DISPLAY SCREENS
3.2.1 Main Display Screen
Upon completion of first sample period, the screen will update with the measurement results
broken out by size range.
MODEL PC3400 24
3.2.4 Status Screen
Pressing the Down Arrow key again will access the Status Screen.
S T A T U S
U n i t : 1 V e r : 1 . 0 7 h
S a m p l e # : 3 1 2
M e m o r y : 4 5 % D a y s : 2 5
S a m p l e P e r i o d : 1 5
S a m p l e F r e q u e n c y : 6 0
L a s e r : 1 0 0 % C e l l : 9 8 %
F l o w R a t e : 7 5 m l / m i n
S a m p l i n g T i m e : 1 6 : 2 0
Unit Address: The units unique address which is used to identify the unit for digital communications
(e.g. MODBUS RTU).
Sample #: Number of measurements taken since the unit was last powered on. This number will rollover
back to 1 when the number reaches 65000. This number will reset to 1 whenever an “Exit & Save” is
performed from the menu. The main purpose if the Sample # is just to provide an indication that the
PC3400 is performing measurements at the proper interval.
Memory / Days: The PC3400 can hold up to 65,000 data strings stored in internal memory and the %
value shown here indicates how full the memory is. The “Days” readout indicates how many days until
the memory will be full.
NOTE: Once the memory is 100% full, the PC3400 will display “Overwrite”. This indicates that
the PC3400 is writing the newest data over top of the oldest data in memory. Data that is written
over will not be available for download. After downloading data, the user can erase the memory
by going to Menu>Service Setup>Resets>Erase Data Memory.
Sample Period: Number of seconds that a measurement lasts. Default value is 15 sec.
Sample Frequency: Number of seconds between the start of measurements. Default value is 60 sec.
Laser: Percentage readout reflecting changes in laser operating current. This value is provided for
diagnostic information only. Laser current will change as a function of laser temperature and age. The
laser percentage will increase over the lifetime of the laser and also when ambient air or water
temperature increases. With yearly calibrations, a laser percentage between 90 to 110% is considered
normal.
Cell: Percentage readout reflecting the cleanliness of the sensor. The sensor should be cleaned after
there has been a 10% decrease following the previous cleaning. Note: Changes in temperature which
affect the laser percentage will also have an affect on the cell percentage. When the laser percentage
drops as a function of cooler temperatures, a corresponding drop in the cell percentage is to be expected.
With yearly calibrations, a cell percentage between 90% to 110% after cleaning is considered normal.
Flow Rate: Flow rate value entered into and being used by the PC3400 electronics to calculate count/ml
or count/100ml data. Note: Standard factory calibration is performed at 75 ml/min. The user can change
the Flow Rate value via the User Menu and set the actual flow rate to match (e.g. 60 ml/min), but particle
sizing may not be accurate if the unit has not been calibrated at that flow rate.
MODEL PC3400 25
3.3 USER MENU / SETUP SCREENS
NOTE:
The PC3400 will not communicate (RS232, RS485 or Ethernet) nor will it perform any measurements
while in the Setup menu. The PC3400 will exit the menu and return to normal operation if the unit is left in
the Setup menu for 5 minutes with no activity.
NOTE:
When exiting the Menu, it is necessary to select “Exit and Save” in order to store any changes to
permanent memory. Alternatively, if the Menu is exited by pressing the Menu button, the changes will be
saved to temporary memory only and changes to settings will be lost if the power is cycled off and on.
If changes were made to the settings that you do not wish to keep, simply exit out of the menu using the
Menu button (do not perform an “Exit and Save”) and then turn the unit off and then back on. This will
erase any and all changes made to the menu since the last time an “Exit and Save” was performed.
NOTE:
User Login is disabled by default which means user has access to all menu functions. To enable User
Login, go to Service Setup > Resets > Tech User Login. A 4 digit login code will be needed to enable the
Login. The unit comes from the factory with a default code of “1234” which can be changed once the
Login is Enabled. Be sure to record this code in a secure place. If the code is forgotten, contact the
factory for assistance.
If the User Login is enabled, then any user who doesn’t have the login code will only be able to perform
two functions:
1. Change Trend Graph mapping .
2. Turn on the Output Hold so cleaning can be performed without disrupting the output signals (4-
20mA, Modbus, Data Log).
MODEL PC3400 26
3.3.2 Sample Setup
From the top Setup menu move the arrow down to “Sample Setup” and press Enter. You should see the
below screen.
MODEL PC3400 27
When entering the size settings, it is not necessary to use all 8 size Bins. Entering a value of 0 will turn
the Bin(s) off. The default setting for Bin 7 and Bin 8 is 0, which is why there are only 6 Size Ranges
active (not counting the ‘greater than’ range) when the unit is received from the factory.
NOTE:
Particle sizing may not be accurate if the unit has not been calibrated at that flow rate being used. It is
recommended to leave the Flow Rate at 75 ml/min unless changes to this setting have been previously
discussed with the factory
MODEL PC3400 28
3.3.3.1 Trend Settings
The Trend Settings menu allows the user to map one of several variables to the Trend Graph. These
include Total Counts, all Bin counts, Cell Condition, Laser Condition, and Analog Inputs. Upon selecting
which variable to map to the graph, the menu prompts the user to select the max scaling for the graph.
The range of the Trend Maximum scaling is from 1 to 65000. The default value is 200 counts.
The Alarm Setup allows the user to Map up to two Alarms and set the Alarm Threshold. Note: The Alarm
settings also control the optional alarm relays. Alarm 1 controls Relay 1 and Alarm 2 controls Relay 2.
The mapping options for the Outputs and Alarms includes Total Counts, all Bin counts, Cell Condition,
Laser Condition, and Analog Inputs.
Setup for optional Analog Inputs is also accessed through this menu via the Input Type selection. It is
necessary to identify each input channel as either 0 – 10 VDC, 0 – 5 VDC or 0 – 20 mA. The 0 – 20 mA
option is also used for 4 – 20 mA signals. Jumpers on the analog input board must be configured to
match the selected input type. See Figure 13.
Note: For Serial (RS232 & RS485) Communications, the default address is 1 and the range is from 1 to
255. Communication settings are by default 9600 baud, 1 stop bit, and no parity bit. The PC3400 can
handle a baud rate up to 56K.
Note: The instrument will not communicate when a menu screen is active. Communications will resume
when the menu is exited.
MODEL PC3400 29
3.3.4 Service Setup
From the top Setup menu move the arrow down to “Service Setup” and press Enter. You should see the
below screen.
If the PC3400 is equipped with the optional 4-20mA Output Board, the 4-20 mA Ouputs can be set to 4
or 20 mA by accessing the Output Adjust menu under the Manual Calibration menu. Once the outputs
are set to either 4 or 20 mA, the potentiometers shown below can be adjusted to calibrate the outputs.
MODEL PC3400 30
3.3.4.2 Resets
Under the Resets menu there are three selections.
Restore Factory Cal – This will take all the instrument settings, including the calibration settings, back to
what they were when the unit shipped from the factory. This is provided to allow the user to return back
to factory settings in the event the unit is no longer operating properly and it is suspected that changes to
critical instrument settings may have been made that are affecting the unit’s operation.
Tech User Login – Enabling the Tech User Login will limit normal users to only being able to do two
things:
1. Enable the “Output Hold” so they can service the sensor without disrupting the data being sent to
the SCADA/DCS.
2. Change the scaling for the Trend on the front panel display.
When enabling the Tech User Login, the user will be required to enter the old password. On a new unit
the password is 1234. Once the correct “old” password has been entered, the user is allowed to Enable
the Tech User Login and create a new password. Be sure to record this password in a secure location. If
the password is lost, contact Chemtrac for a password reset.
Erase Data Memory – This will clear the PC3400’s datalog. Be sure to download data as outlined in
section 6 of this manual before erasing data. Once data is erased, it cannot be retrieved.
Note: It does require Level 2 Tech User access to erase memory. It is recommended to enable Tech
User Login in order to prevent unauthorized erasing of the data log
Yellow TX LED
Green RX LED
Status LEDs
MODEL PC3400 31
With TracWare running the LED’s on each unit should flash at least once a minute. If the TX LED does
not flash, there is a break in the communications line prior to this unit or a problem with the computer. If
the TX LED flashes but the RX LED never flashes, the problem may be in the unit. Contact Chemtrac if
the LED’s do not appear to be functioning properly and communication cannot be established with the
host computer.
It is important that a consistent flow rate be maintained since the data produced by the particle counter is
based on the sample volume (particles per milliliter). As the tubing and flow cell become dirty, flow will be
restricted. Therefore it is necessary that a proper maintenance schedule be followed to ensure the flow
rate is not adversely impacted from fouling.
If adequate sample pressure is available, the constant head flow control weir will maintain an accurate
flow rate through the sensor. Flow problems will only be caused by fouling of the flow cell and fouling of
the tubing and barb fittings. Obviously, the more turbid the sample, the more likely clogging / fouling will
occur. If raw water is being sampled, it is recommended that a small 40 mesh strainer be placed in line
before the sensor to catch debris.
Routine cleaning of the flow cell should be performed on a regular basis. How often the flow cell should
be cleaned will vary depending on water quality. The cell condition is monitored by the particle counter
electronics, and can be used as an indication of when cleaning is necessary. A drop in the cell condition
to 85% (or 10 to 15% from previous reading obtained after cleaning) indicates that cleaning should be
performed.
Any standard laboratory cleaner can be used to clean the flow cell. If iron or manganese buildup is a
problem, vinegar, CLR, or another slightly acidic liquid, can be used. The cleaner can be flushed through
the flow cell, or applied using the cleaning brush. The sapphire window material and the external flow cell
design make brush cleaning a safe and effective method for removing tough contaminants.
NOTE
Prior to connecting the sensor, the sample line should be flushed thoroughly. Failure to do this is a
common cause of clogged sensors.
Several methods may be used to clear a clogged flow cell. One of the most effective tools is a small can
of compressed air (available at Radio Shack or most hardware stores). Remove the tubing from both
sides of the flow cell and blow the air through the flow cell from the top of the flow cell (opposite direction
from the flow). If compressed air is not available run the liquid sample in from the top of the flow cell and
increase the flow rate. The flow cell windows are sapphire, which is scratch resistant, but it is not
recommended that pieces of wire or other objects be used to clear clogs. Once the clog is removed, a
light can be shined through the sensor to allow visual inspection. The flow cell opening is large enough to
see through when clear.
NOTE
The particle counter can still count particles with a dirty flow cell, but the accuracy may be
diminished.
MODEL PC3400 32
3.5.1 Non-brush Cleaning
Fill a large syringe with cleaner (diluted to manufacturer’s recommendation) and attach it to the top flow
cell barb with a short piece of tubing (See Figure 18). Squeeze the syringe to introduce the cleaner into
the sensor and let it sit for a few minutes and then inject some more and let that stand a few more
minutes. Finally, force the remaining cleaner rapidly through the flow cell. The rapid flow should remove
most buildup that has occurred in the sensor and bring the cell condition back up to an acceptable level.
Non-brush cleaning is not a replacement for occasionally cleaning with a brush, but it useful when you
don’t have a brush on hand.
Non-Brush Brush
Figure 18 Sensor Cleaning Methods
NOTE
The small size of the brush makes it easy to bend or damage. Do not force the brush into the flow
cell. It should fit into the flow cell with light pressure.
Once cleaning has been performed, reattach flow lines and allow a minute or so for the flow cell to flush
out. Verify that the cell condition indicator is back in the 90% range. If it is not, repeat the procedure.
NOTE
The cell condition indicator is not valid if the sensor is not full of water and clear of air bubbles.
MODEL PC3400 33
4.0 DOWNLOADING DATA
The PC3400 records all measurement data in internal flash memory. The internal memory holds over
65,000 sample strings of data. This data can be downloaded over an RS-232 or RS-485 connection
using the TracData software.
Go to the folder where the TracData software was installed and click on TracData.exe. The below screen
will appear.
MODEL PC3400 34
At the top of the TracData screen are the communication settings for Com port and Baud rate. The user
must select the Com port that will be used to communicate with the PC3400. If a USB to Serial converter
is being utilized, the Com port can be identified via the Device Manager (located under windows Control
Panel). As shown in the below example, the USB to Serial converter is connected to COM 5.
Once the Com port has been identified, change the Com port setting to match in TracData. Also, verify
the Baud rate is set for 57600.
At this point, it will be necessary to connect the RS-232 Download Cable between the Com port on the
computer and the RS-232 terminals on the PC3400.
MODEL PC3400 35
Connect the DB9 end of the download cable to a Com port on your PC. Alternatively, if your computer
only has USB ports, you will need to connect the DB9 connector to a USB to Serial converter that gets
plugged into a USB port.
Connect the other end of the download cable, which is the 8 position green terminal, to the PC3400’s RS-
232 Communications connector shown below.
Press the Menu button on the PC3400’s front panel display. If the Login is enabled for Level 2 Tech User
access, you will need to first enter the login code. Then go to the following menu:
Set the Baud Rate to match whatever setting is being used on the TracData software. The default and
recommended setting in TracData is 57600. The default setting for the PC3400 Baud rate is 9600 and so
it will be necessary to change this setting to match TracData’s setting (i.e. 57600) in order to establish
communication. For the new Baud rate settings to take effect, it’s necessary to perform an Exit & Save
from the top menu after changing the Baud Rate.
Note: Data download at 57600 Baud Rate is around 40 records per second. If there are communication
problems, try using a slower Baud Rate setting. Remember that the settings in TracData and the PC3400
must match.
Next, make a selection in the box labeled “Data Retrieval By”. It is recommended to select the “Recent
Records” option which downloads the number of records entered starting with the most recent data.
Alternatively, if you know a date and time that matches a record stored in the PC3400’s memory, you can
select “Date & Time” and tell the software how many records you want to download starting at that point.
Note: When using “Date & Time” option, it is mandatory that the date and time entered matches exactly
with a record stored in memory. If there is no match, data will not download and the PC3400 will freeze in
“Terminal Setup” mode and require the power to be cycled off and back on.
MODEL PC3400 36
Click on the “Retrieve Data” button to begin data download. The recorded measurement data will be read
in and displayed in the TracData software.
Note: The information displayed in the Terminal field is communications back and forth between the
computer and PC3400.
Below is how the data will appear in TracData after it has been successfully downloaded.
Raw, unformatted
measurement data that’s
been downloaded from
PC3400.
Note: If TracData is unable to download data when using the “Date & Time” retrieval option, and the
PC3400 appears to be locked up in Terminal Setup mode, the most likely cause is that an invalid date or
time has been entered. To fix this, simply cycle power to the PC3400 and then download data using the
“# of Recent Records” option versus the “Date & Time” option.
Once it has been verified the data has downloaded and been exported to the desired file type, the
PC3400’s memory can be erased by going to Menu>Service Setup>Resets>Erase Data Memory.
Note: It does require Level 2 Tech User access to erase memory. It is recommended to enable Tech
User Login in order to prevent unauthorized erasing of the data log (see section 5.3.4.2 of this manual for
more info).
MODEL PC3400 37
5.0 RECOMMENDED SPARE PARTS LIST
Part # Description
11420 Sensor Sample Tubing*
11425 Weir Inlet & Drain Cup Tubing*
34400 Sensor Cleaning Brush
*Tubing can also be ordered from Cole Parmer using the below product numbers:
Chemtrac Inc.
6991 Peachtree Industrial Blvd. Building 600
Norcross GA 30092
MODEL PC3400 38
Appendix A
User Menu Map
MODEL PC3400 39