AIP-1200V Infusion Pump Maintenance Manual
AIP-1200V Infusion Pump Maintenance Manual
AIP-1200V Infusion Pump Maintenance Manual
Infusion Pump
(AIP-1200V)
Table of Contents
Chapter 1 Safety Matters ...................................................................................................................................3
1.1 Warnings ....................................................................................................................................3
1.2 Cautions .....................................................................................................................................4
1.3 Dialogue window ...................................................................................................................... 5
1.4 Symbols .....................................................................................................................................5
Chapter 2 Organization of Infusion Pump ........................................................................................................ 6
2.1 Mechanical Structure of Infusion Pump ................................................................................... 6
2.2 Hardware Structure of Infusion Pump ...................................................................................... 7
2.2.1 Function Introductions of Board Card .............................................................................. 7
2.2.2 Main Control Board Interface Details .............................................................................. 8
2.2.3 Driver board interface signal .......................................................................................... 10
2.2.4 Battery adapter board interface signal ............................................................................ 12
2.2.5 External interface signal ................................................................................................. 13
2.3 Software Structure of Syringe Pump ...................................................................................... 14
2.3.1 Interface Flow Chart of Syringe Pump ...........................................................................14
2.3.2 System Interfaces ............................................................................................................ 15
Chapter 3 Common Faults .............................................................................................................................. 16
Chapter 1 ................................................................................................................................................16
Chapter 2 ................................................................................................................................................16
Chapter 3 ................................................................................................................................................16
3.1 Common Faults ....................................................................................................................... 16
3.2 Failure Analysis and Solutions ................................................................................................16
3.2.1 Inaccurate Touch on Display Screen ...............................................................................16
3.2.2 Failure of Motor of Switch Door; ................................................................................... 16
3.2.3 Failure of Fluid Flow Regulator ; ................................................................................... 16
3.2.4 False Alarming of Bubble ............................................................................................... 16
3.2.5 System Failure 2 ..............................................................................................................17
3.2.6 System Failure 4 ..............................................................................................................17
3.2.7 Loudspeaker Failure ........................................................................................................17
3.2.8 Charging Failure ............................................................................................................. 17
Chapter 4 Specific Maintenance Operations ...................................................................................................18
Chapter 4 ................................................................................................................................................18
4.1 Preparations Before Maintenance ........................................................................................... 18
4.1.1 Split Diagram of Infusion Pump Assembly ....................................................................18
4.1.2 Maintenance Tool ............................................................................................................19
4.1.3 Notices .............................................................................................................................19
4.2 Components Disassembly .......................................................................................................19
4.2.1 Battery Removal ............................................................................................................. 19
4.2.2 Split of Front and Back Shells ........................................................................................ 21
4.2.3 Remove Pump Body ....................................................................................................... 22
4.2.4 Removal of Fluid Flow Regulator .................................................................................. 23
4.3 Maintenance Operations ......................................................................................................... 24
4.3.1 Replacement of Display Screen ....................................................................................24
4.3.2 Replacement of Hall Sensor Components ...................................................................... 25
1 / 43
4.3.3 Replacement of Door Motor ........................................................................................... 26
4.3.4 Replacement of Ultrasonic Module ................................................................................ 27
4.3.5 Replacement of Optocoupler PCBA ...............................................................................28
4.3.6 Replacement of Down Pressure Sensor Assembly ......................................................... 29
4.3.7 Loudspeaker Replacement .............................................................................................. 30
4.3.8 Replacement of Battery Adapter Plate ............................................................................31
4.3.9 Replacement of Power Module .......................................................................................32
Chapter 5 Calibration and Testing ...................................................................................................................33
5.1 Sensor Calibration ...................................................................................................................33
5.1.1 Upper Blocking Sensor ................................................................................................... 34
5.1.2 Lower Blocking Sensor ...................................................................................................34
5.1.3 Calibration of Display Screen .........................................................................................34
5.2 Function Test of Infusion Pump ..............................................................................................35
5.2.1 Anti-bolus test ................................................................................................................. 35
5.2.2 Pump stop fluid test ........................................................................................................ 35
5.2.3 Calibration method of the Down Occlusion pressure sensor ......................................... 35
5.2.4 Blocking Pressure Test ....................................................................................................36
5.3 Accuracy Testing .....................................................................................................................36
5.3.1 25mL Accuracy ............................................................................................................... 36
Chapter 6 Care and Maintenance .................................................................................................................... 38
Chapter 2 ................................................................................................................................................38
Chapter 3 ................................................................................................................................................38
Chapter 4 ................................................................................................................................................38
Chapter 5 ................................................................................................................................................38
Chapter 6 ................................................................................................................................................38
6.1 Cleaning, disinfecting and sterilizing ..................................................................................... 38
6.1.1 Cleaning .......................................................................................................................... 38
6.1.2 Disinfecting ..................................................................................................................... 38
6.2 Periodical maintenance ........................................................................................................... 38
6.2.1 Check the Appearance .....................................................................................................39
6.2.2 Performance Check ......................................................................................................... 39
6.2.3 Maintenance Plan ............................................................................................................39
6.3 Add new brand and Calibration .............................................................................................. 39
6.4 Repair ...................................................................................................................................... 41
6.4.1 Normal Repair Process ................................................................................................... 41
6.4.2 Maintenance for Long Term Store ..................................................................................41
6.5 Equipment Components/Accessories ..................................................................................... 41
6.6 Production Date .......................................................................................................................41
6.7 Recycling .................................................................................................................................41
Chapter 7 After-Sales Service ......................................................................................................................... 42
Chapter 8 Appendix .........................................................................................................................................43
Appendix A Material list ......................................................................................................................... 43
2 / 43
Chapter 1 Safety Matters
1.1 Warnings
Before using, please check the equipment, connecting wire and accessories to ensure that it can work
normally and safely. If there’s anything abnormal, immediately stop working and contact our after sale
service department. Additionally, the adhesion or intrusion of fluid/drug may possibly cause the equipment
fault and malfunction. Therefore, please clean the equipment after use, and store it correctly.
This equipment must be operated by trained professional medical care personnel.
It is not allowed to put and use the equipment in the environment with anesthetic and other inflammable or
explosive articles to avoid fire or explosion.
It is not allowed to store or use the equipment in the environment with active chemical gas (including gas
for disinfecting) and moist environment since it may influence the inside components of the infusion
pump and may possibly cause performance drop or damage of the inside components.
The operator shall guarantee that the set infusion parameters of this equipment are the same as the medical
advice before starting infusion.
Please correctly install the infusion apparatus according to the infusion indication direction of this
equipment, ensure that infusion tube smoothly and straightly cross the creep device. Otherwise, it may
possibly suck blood from the patient or fails to reach the expected performance.
Please do not only depend on information prompt during use, please periodically check it to avoid
accident.
Tightly fix this equipment on the infusion stand and ensure the stability of the infusion stand. Be careful
when moving the infusion stand and this equipment to avoid the equipment dropping and infusion stand
falling or knocking the surrounding objects.
If the infusion tube is twisted, or the filter or needle is obstructed, or blood in the needle which may
obstruct the infusion, the pressure in the infusion tube will rise. When removing such occlusion, it may
possibly cause “bolus injection” (temporary excess infusion) to the patient. The correct method is to
tightly hold or clamp the infusion tube near the puncturing position, then open the door to drop the
pressure in the infusion tube. Then loosen the infusion tube, solve the reason of occlusion, and restart
infusion. If infusion is restarted when the occlusion reason exists, then it may cause occlusion alarm
persistently, and the pressure in the infusion tube may keep rising, and may break or cut off the infusion
tube, or hurt the patient.
This equipment injects fluid/drug through extruding the infusion tube, but it can’t detect the leakage if the
infusion line is cut off or broken. Therefore, please periodically check it to avoid above fault during the
working period.
During infusion, please periodically check the dripping state of the fluid and the fluid/drug in the
intravenous infusion bag/container, so as to ensure the correct working during infusion. This equipment
doesn’t directly measure the quantity of infusion fluid, therefore, it is possible that this equipment can’t
detect the free infusion flow under the extremely special condition. Even the drop sensor is adopted, it is
possible that this equipment can’t detect the free infusion flow which is less than the specific value for the
demands of tolerance.
This equipment has the occlusion detection function for detecting and alarming when the infusion needle
deviates the position in the vein or the needle is not correctly punctured in the vein. However, it only
alarms when the occlusion pressure has reached certain numerical value, and the puncturing part may
possibly have become reddish, swelling or bleeding, additionally, it is possible that the device doesn’t
alarm for a long period if the actual occlusion pressure is lower than the alarm threshold value, therefore,
please periodically check the puncturing part. If there’s any abnormal phenomenon for the puncturing part,
3 / 43
please timely take suitable measures, such as puncturing again.
Only those infusion apparatus, line, infusion needle and other medical components that meet the local
laws and regulations and the requirements covered in and this User Manual can be adopted, it is suggested
to adopt the infusion apparatus with same brand as this equipment. It can’t ensure the infusion accuracy if
the unsuitable infusion line is adopted.
It is not allowed to disassemble or refit this equipment or use it for other purposes except normal infusion.
No one is allowed to repair this equipment except our company or the authorized repair technician of our
company.
To avoid risk of electric shock,this equipment must only be connected to AC with Ground Protection.
1.2 Cautions
Before its first use after purchase, or this equipment is not used for a long period, please charge the
equipment with AC power supply. If it is not fully charged, under power failure, the equipment can’t
continue working with built-in battery power supply.
This equipment can not be used in the places with radiological installation or magnetic resonance
equipment as well as the places with high pressure oxygen therapy.
Other devices near this equipment must meet corresponding EMC requirements, otherwise, it may
influence the performance of this equipment.
Under general conditions, please use AC power supply as much as possible since it can prolong the
service life of the battery at a certain degree. When using AC power supply, ensure that the grounding
wire is reliably connected with the ground, and only the AC power wire attached with this equipment
shall be adopted. The built-in battery can only be used as the assistant power supply when the AC
power supply can’t reliably connected with the ground and is not under normal conditions (power
failure or moving infusion).
Before connecting this equipment with power supply, please keep the power socket and plug dry, and
the power voltage and frequency meet the requirements listed in the equipment label or this User
Manual.
The equipment is equipped with the audible and visual alarm system, and the red and yellow alarm
indicators will light on by turn to check if the alarm system can work normally, and the speaker makes
the “beep” sound.
Please keep the equipment away from the AC power socket for a certain distance to avoid fluid/drug
splashing or dropping in the socket, otherwise, it may possibly cause short circuit.
Please use the fluid/drug after it has reached or nearly reached room temperature. When the fluid/drug
is used at low temperature, the air which is dissolved in the fluid/drug may cause more air bubbles and
result in frequent air bubble alarm.
It is not allowed to press and operate the button with sharp object (such as pencil tip and nail),
otherwise, it may possibly cause early damage to button or surface film.
Please do not use the infusion tube for 8h at the same pumping position. Infusion tube may distort after
using for a long time and cause flow rate error. It is suggested to replace the pumping position or
directly replace the infusion tube every 8h.
Please tightly close the flow rate adjuster of the infusion apparatus before taking out the infusion
apparatus to avoid liquid leakage.
Under the condition of low flow rate infusion, please pay special attention on occlusion. The lower the
infusion flow rate, the longer the time of detecting occlusion, and it in turn may possibly cause a long
time infusion stop during this period.
If the equipment suffered from dropping or impacting, please immediately stop using it, and contact our
4 / 43
after sale service department, because the inside components of the equipment may be possibly
damaged even the appearance is not damaged and abnormality is not occurred when working.
1.4 Symbols
Not all of the below symbols are existed in the equipment you have purchased.
Table:1.4-1
Defibrillation-proof type CF
Handle with harmless method
applied Part
Unlock Lock
CE-mark/Notified Bod
5 / 43
Chapter 2 Organization of Infusion Pump
6 / 43
2.2 Hardware Structure of Infusion Pump
8 / 43
13 VBUS_IN Power bus Power Output 8~16V
4 POWER_ON Switch button signal Level output 3.3V±0.5V
15 EN_SPEAK Speaker enable port Output 0~3.6V
16 MOTO_CONTRAL Motor control signal Level output 3.3V±0.5V
17 NURSE_CALL_O Nurse call signal Level output OC
18 MAIN_MCU_RXI External infrared receiving signal Pulse output 0~3.6V
Sending signals to the external
19 MAIN_MCU_TX1 Pulse output 0~3.6V
infrared
20 GND Function field / /
9 / 43
4 VDD 3.3V power supply Output 3.3V±0.3V
5~12 R0~R7 LCD interface data cable Two-way 2~3.6V
13~20 G0~G7 LCD interface data cable Two-way 2~3.6V
21~28 B0~B7 LCD interface data cable Two-way 2~3.6V
29 GND Function field / /
30 PLCK Pixel clock Pulse output 0~3.6V
31 DISP / Two-way /
32 HSYNC Horizontal sync signal Two-way /
33 VSYNC Vertical sync signal Two-way /
34 DE Enable port Two-way 3.3V±0.3V
35 NC Unused / /
36 GND Function field / /
37 XR Touch signal Two-way 0~3.3V
38 YD Touch signal Two-way 0~3.3V
39 XL Touch signal Two-way 0~3.3V
40 YU Touch signal Two-way 0~3.3V
10 / 43
24 WAVE Speaker signal input Pulse output 0~3.6V
2.2.3.2 Driver board P6 and motor code disc
Table 2.2.3.2-1 Driver Board Interface P6 Signal Definition
NO. Network label Signal name Input/output Signal range
1 NetR18_1 Transmitter input Input 2V±1V
2 MOTO_SENSOR1 Pulse output Output 0~3.3V
3 GND Function field / /
2.2.3.3 Driver Board P9
Table 2.2.3.3-1 Driver Board Interface P9 Signal Definition
NO. Network label Signal name Input/output Signal range
1 GND Function field / /
2 SWCLK Clock signal Two-way 0~3.3V
3 SWDIO data transmission Two-way 0~3.3V
4 VDD 3.3V power supply Output 3.3V±0.3V
2.2.3.4 Driver Board P3
Table 2.2.3.4-1 Driver Board Motor Interface P3 Signal Definition
NO. Network label Signal name Input/output Signal range
1 2A Motor power supply Output 0~24
2 2B Motor power supply Output 0~24
3 1A Motor power supply Output 0~24
4 1B Motor power supply Output 0~24
2.2.3.5 Driver Board P1
Table 2.2.3.5-1 Driver Board Motor Interface P3 Signal Definition
NO. Network label Signal name Input/output Signal range
1 NETC4-2 Motor power supply Output 0~5V
2 NETC26-1 Motor power supply Output 0~5V
2.2.3.6 Driver Board P2
Table 2.2.3.6-1 Driver Board Motor Interface P2 Signal Definition
NO. Network label Signal name Input/output Signal range
1 NETC5-2 Motor power supply Output 0~5V
2 NETC31-1 Motor power supply Output 0~5V
11 / 43
2.2.3.9 Driver Board U6
Table 2.2.3.9-1 Driver Board U6 Signal Definition
NO. Network label Signal name Input/output Signal range
1 VCC 5V power supply Power Output 5V±0.5V
2 DOOR_ON_COM Door control signal Output 0~3.6V
GND Function field / /
2.2.3.10 Driver Board P7
Table 2.2.3.10-1 Driver Board P7 Signal Definition
NO. Network label Signal name Input/output Signal range
1 GND Function field / /
2 TUBL_OUT- Ultrasonic signal - Pulse signal output 0~5.5V
3 TUBL_OUT+ Ultrasonic signal + Pulse signal output 0~5.5V
12 / 43
2.2.4.2 Battery adapter plate P2
Table 2.2.4.2-1 Battery Adapter Board Speaker Interface P2 Signal Definition
NO. Network label Signal name Input/output Signal range
1 GND Function field / /
2 SPEAK+ Speaker power supply Output 0~5V
3 SPEAK- Speaker power supply Output 0~5V
2.2.4.3 Battery Adapter Plate P3
Table 2.2.4.3-1 Battery Adapter Board Infrared Communication Interface P3 Signal Definition
NO. Network label Signal name Input/output Signal range
1 VCC 5V power supply Power Output 5V±0.5V
2 MAIN_MCU_TX4 Serial port send signal Pulse output 0~3.6V
3 MAIN_MCU_RX4 Serial port accept signal Pulse output 0~3.6V
4 GND Function field / /
2.2.4.4 Battery Adapter Plate P5
Table 2.2.4.4-1 Battery Adapter Board Power Module Interface P3 Signal Definition
NO. Network label Signal name Input/output Signal range
1 ACDC 15V power supply Power Output 15V±0.5V
2 ACDC 15V power supply Power Output 15V±0.5V
3 GND Function field / /
4 GND Function field / /
6 NC NC NC NC
13 / 43
4 DROP_EN Drop rate potential signal Level signal 0~3.6V
14 / 43
2.3.2 System Interfaces
The interfaces of software system and external components are as follows:
System
interface
15 / 43
Chapter 3 Common Faults
17 / 43
Chapter 4 Specific Maintenance Operations
18 / 43
4.1.2 Maintenance Tool
(1)Electric screwdriver
Use different electric screwdriver according to different screw specifications, and adjust torque value
of the electric screwdriver, as follows:
Table 4.1.2-1 Screw specification table
Screw Specifications Torque value
M3 machinery tooth screw 3.0kgf/cm±0.5
M5 machinery tooth screw 4.5kgf/cm±0.5
ST2.9 Cross recessed pan head tapping screw 4.0kgf/cm±0.5
ST2.6 Cross recessed pan head tapping screw 3.5kgf/cm±0.5
(2)Diagonal pliers
4.1.3 Notices
① You should remove battery to disconnect machine before disassembly, in order to prevent damage
to the main and accessory boards of the infusion pump;
② Pay attention to the wire when disassembling the machine, to avoid damage to the wire, and tie the
wires with strapping tape when installed;
③ Be careful not to leave debris in the machine when installed.
(1)Before removing the battery, press the button of door opening 1 to open the door, as shown in
Figure 4.2.1-1.
19 / 43
(2) Turn the infusion pump to the back shell position, the right side on the door of battery store of rear
shell is equipped with button position, open the door of battery store 3 from this position, move the battery
latch 2 to the right, and then pull out the battery 1, as shown in Figure 4.2.1-2 :
1 2 3
Figure 4.2.1-2 Battery removal
Table 4.2.1-1 Material list
NO. Material number Material name Quantity
1 09-000039-XX Battery 1
2 24-000075-XX Battery latch 1
3 24-000036-XX Door of battery store 1
20 / 43
4.2.2 Split of Front and Back Shells
21 / 43
4.2.3 Remove Pump Body
22 / 43
4.2.4 Removal of Fluid Flow Regulator
23 / 43
4.3 Maintenance Operations
24 / 43
4.3.2 Replacement of Hall Sensor Components
25 / 43
4.3.3 Replacement of Door Motor
26 / 43
4.3.4 Replacement of Ultrasonic Module
Steps: ① Disassemble front and back shell, the operations are shown in 4.2.1-4.2.2;
② Remove the connecting wire of ultrasonic sensor, and then remove the locking screw 1 of
ultrasound sensor module, remove the ultrasound sensor module 2 for replacement.
Note: It is recommended to replace the ultrasonic sensor 1 first, and check if the problem is solved after
installation. If not, replace the ultrasonic sensor 2. The method is the same as above. Replacement of
Optocoupler PCBA
27 / 43
4.3.5 Replacement of Optocoupler PCBA
28 / 43
4.3.6 Replacement of Down Pressure Sensor Assembly
29 / 43
4.3.7 Loudspeaker Replacement
30 / 43
4.3.8 Replacement of Battery Adapter Plate
31 / 43
4.3.9 Replacement of Power Module
32 / 43
Chapter 5 Calibration and Testing
5.1 Sensor Calibration
33 / 43
5.1.1 Upper Blocking Sensor
1 2
3 4
5
Figure 5.1.1-1 Calibration operation
Do not install the infusion tube, click [System Information]-> [System Maintenance]-> Enter the system
password "2341", click [√]-> [Sensor Information] and check the [Pressure up] , And record it in the table.
If the value has a small change, record the intermediate value. The AD value within the range of [-200,
1000] is qualified.
5.1.2 Lower Blocking Sensor
No infusion tube is installed, as shown in Figure 5.1.1-1, click [Menu]-> [System Information]-> [System
Maintenance]-> [Sensor Calibration], check the [Pressure down] value, click [AD0] and record. If there is a
small change in the value, it is sufficient to record the intermediate value. The AD value within the range of
[10,000, 65,000] is qualified.
5.1.3 Calibration of Display Screen
Steps:
① Start the machine, enter into the starting interface, press the power button 1, and enter into the power
off interface;
② Simultaneously press the BOLUS key 2 and OPEN key 3 to enter into the calibration interface of
display screen, and then follow the prompts to calibrate.
34 / 43
1
3
Figure 5.1.4-2 Calibration interface of screen of infusion pump
35 / 43
Figure 5.2.3-1
(Path: [Menu] -> [System] -> [Maintenance] -> Enter the system password "2341", click【√】->【Sensor
Calibration】)
5.2.4 Blocking Pressure Test
Method: As shown in Figure 5.1.1-1, click [Menu]-> [System Information]-> [System Maintenance]->
Page Turn and turn off the Anti-bolus switch.
Install the infusion pipeline according to Figure 5.1-2, and rotate the three-way valve to make the three
channels conductive. Press the exhaust button to evacuate until all the air bubbles in the infusion line are
discharged. Rotate the three-way valve to make the test infusion pump communicate with the pressure
gauge. At this time, the pressure reading should be 0kPa.
Set the infusion speed and blocking level of the machine according to the following table, perform a first
test on each group of data, and record the pressure value displayed by the pressure test tool when the alarm
occurs.
Occlusion Theoretical Real Real Real
No. rate
level pressure range pressure1 pressure2 pressure3
1 200ml/h Level-1 7~33Kpa
2 200ml/h Level-2 48~72Kpa
3 200ml/h Level-4 96~144Kpa
Note: When the pressure value of the first test is within the theoretical pressure range, it is judged as
passing; the second and third tests are not required.
When it fails for the first time, reinstall the tube, and measure the parameter twice in a row, and record the
test data. If it is within the range for two consecutive times, it will pass, otherwise it will fail.
When the first test is failed, then re-install pipe, and make two consecutive measurements for the
parameters, and record test data. It will be qualified if the results in the two consecutive tests are in the
range; otherwise, it is unqualified.
The weight of the empty cup of the measuring cup is measured with an electronic balance, and the weight
of the empty cup is defined as W1, and the value of W1 is filled into the recording table.
Set the infusion rate at 25ml / h, the preset amount is 25ml, and the KVO is 0ml / h. Insert the venous
needle into the measuring cup with lid. Click [Start] button to start the infusion. After completing the
36 / 43
infusion, click [Stop] button to complete the infusion.
Measure the measuring cup weight containing liquid with a balance, the weight is defined as W2, and the
value of W2 is filled into the recording table.
First, use the formula to calculate the reference weight of the infusion Wref = 25 * 0.998 = 24.95g, then use
the formula {(W2-W1) / Wref-1} * 100%Calculate the actual infusion accuracy, accuracy error within ±
5% shall be qualified.
37 / 43
Chapter 6 Care and Maintenance
6.1.1 Cleaning
(1) The daily maintenance is mainly to clean the housing and pump body. It is inevitable that
fluid/drug may flow in the equipment during infusion. Some fluid drug may corrode the pump and
cause working fault. After infusion, please timely clean the equipment, wipe it with moist and clean
soft fabric, and then naturally dry it.
(2) When cleaning the equipment interface, please wipe it with dry and soft fabric, confirm the interface is
dry before using.
(3) Please do not soak the equipment in water. Although this equipment has certain waterproof function,
when fluid splashes on the equipment, please check if it works normally, perform insulation and
electric leakage test if needed.
6.1.2 Disinfecting
(1) Disinfecting may possibly cause harm of certain degree to the equipment, it is suggested to disinfect
the equipment if it is needed.
Please disinfect the equipment with common disinfecting agent such as 50% sodium hypochlorite,
cidex 2% glutaraldehyde + activating agent, 70% ethanol, 70% isopropyl alcohol and so on. Please
follow the instructions of the disinfecting agent.
(2) After disinfecting, wet the soft fabric with warm water, dry the fabric and then wipe the equipment
with it.
(3) Do not sterilize the equipment with high pressure steam sterilizer, do not dry the equipment with dryer
or similar product.
Warning: ● Please do not adopt Cidex OPA orthophthalaldehyde, methyl ethyl ketone or similar
solvent, otherwise, it may corrode the equipment.
38 / 43
6.2.1 Check the Appearance
1) The appearance of the equipment shall be clean and under good condition without crack and water
leakage.
2) The buttons are flexible and effective without invalid phenomenon; the sensitivity of the touch
screen is normal,
3) The slider of the syringe pump is flexible in movement, and the clamp is ok.
4) The power wire is under good condition and installed tightly.
5) After connecting with external power supply, check if the AC indicator of the equipment AC
indicator lights on normally.
6) Adopt the accessories designated by our company.
7) The environment meets the requirements.
In the 『System』 submenu, click 『Brand maintenance』 to enter into brand setting interface, create
the consumables brand, delete and calibrate the brand.
Warning: ● It is suggested to contact our company or local dealer, and customize and calibrate it
by professional technician, otherwise, it can’t guarantee the infusion accuracy.
Note: ● If the actual using infusion apparatus brand is not listed in the system built-in brand,
39 / 43
please create the infusion apparatus brand in this interface.
4) Press【Bolus】to remove air bubble in the line, put the needle into the measuring cup for collecting
fluid.
6) After 5mins, the equipment will automatically stop, then record the net weight of fluid by ml;
8) Calibration completed.
(2) Delete
(3)Calibrate
Note
Please calibrate the infusion apparatus when using the built-in brand infusion apparatus for the first
time, or the first user-defined infusion apparatus brand, or after periodical maintenance.
One new and unused infusion apparatus, 20ml measuring cup or 20ml injector.
Calibrating Steps:
3) Press【Bolus】to remove air bubble in the line, put the needle into the measuring cup for collecting
fluid
5) After 5mins, the equipment will automatically stop, then record the net weight of fluid by ml;
7) Calibration completed.
Note:
40 / 43
When the『Volume』is less than 10ml, the infusion rate is ≤1500ml / h.
When the『Volume』is less than 7.5ml, the infusion rate is ≤1000ml / h.
6.4 Repair
6.7 Recycling
The normal service life of this equipment is 10 years, and depends on the use frequency and maintenance.
The equipment must be rejected after reaching the service life, please contact the manufacturer or the dealer
to get more detailed information.
1. The obsolete equipment may be returned to the original dealer or manufacturer.
2. The used lithium-ion polymer battery has the same treatment method, or according to the applicable laws
and regulations.
3. Please handle according to the equipment rejecting flow of your medical mechanism.
41 / 43
Chapter 7 After-Sales Service
One-year free warranty is provided to this product after purchase. The warranty period begins on the date of
installation on the attached Product Warranty Card. "Product Warranty Card" is the only proof of the
warranty period. In order to safeguard your interests, please carefully fill in and keep the "product warranty
card", and hand over the company page to the installer.
In case of damages or faults in the above circumstances, the Company may provide maintenance, but will
charge a fee according to the maintenance cost.
Site included: 5th Floor, Block A, Defengsheng Building, No.41 Dabao Road, Bao'an District 23, Shenzhen,
518101, P.R.China
42 / 43
Chapter 8 Appendix
43 / 43