Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

Pilot A Ce en

Download as pdf or txt
Download as pdf or txt
You are on page 1of 122

TECHNICAL

MANUAL
TECHNICAL MANUAL

PILOT A, CE

Serial N° 15789806 to...


1 Overview ........................................................................................................... 5
1.1 General .................................................................................................... 5
1.2 Overview diagram ................................................................................. 6
1.3 Precautions to be taken before use...................................................... 7
1.4 Internal safety features .......................................................................... 7
1.5 Technical characteristics ...................................................................... 7
1.5.1 Electrical..................................................................................................................... 7
1.5.2 Electronics.................................................................................................................. 7
1.5.3 Mechanical characteristics ......................................................................................... 7
1.5.4 Conformity, standards ................................................................................................ 8

2 Description and operation ............................................................................... 9


2.1 Physical description .............................................................................. 9
2.1.1 The display board and the front panel...................................................................... 10
2.1.2 CPU board ............................................................................................................... 12
2.1.3 Power supply board and battery .............................................................................. 15
2.1.4 Mechanical base unit ............................................................................................... 18
2.1.5 Mechanical plunger unit ........................................................................................... 18
2.2 Functional description......................................................................... 19
2.2.1 Syringe control and maintenance sub-assembly ..................................................... 19
2.2.2 Motorisation sub-assembly ...................................................................................... 19
2.2.3 External connection sub-assembly .......................................................................... 19

3 Description of the menus .............................................................................. 21


3.1 Configuration menu of the basic operation parameters .................. 21
3.1.1 Access to menus ...................................................................................................... 22
3.1.2 PAr1, configuration of the memorisation type ...................................................... 22
3.1.3 Par2, configuration of the syringe selection type ................................................. 23
3.1.4 PAr3, configuration of the maximum flow rate that can
be selected by the keyboard. ................................................................................... 23
3.1.5 PAr4, configuration of the list of syringes that can be selected............................ 24
3.1.6 PAr5, configuration of the compulsory priming .................................................... 25
3.1.7 PAr8, configuration of Master mode..................................................................... 25
3.1.8 PAr9, configuration of the RS232 communication speed..................................... 26
3.1.9 PArA, configuration of the empty syringe mode ................................................... 26
3.1.10 Syringe/name match table ....................................................................................... 27
3.2 Calibration menu .................................................................................. 29
3.3 Service test menu................................................................................. 31

4 Preventive maintenance ................................................................................ 33


4.1 Recommendations ............................................................................... 33
4.2 Maintenance schedule ......................................................................... 33
4.2.1 Use beyond the framework of the departmental order ............................................. 33

PiloteCE_A_enTDM.fm 1
4.2.2 Use within the framework of the departmental order ............................................... 33
4.3 Checks .................................................................................................. 35
4.3.1 Test access .............................................................................................................. 35
4.3.2 Visual check ............................................................................................................. 35
4.3.3 Running time and servicing inspection date............................................................. 36
4.3.4 Indicator lights check................................................................................................ 36
4.3.5 Keyboard check ....................................................................................................... 37
4.3.6 Checking the battery voltage.................................................................................... 38
4.3.7 Testing the last 10 alarms ........................................................................................ 38
4.3.8 Total running time check .......................................................................................... 39
4.3.9 TTL serial link test .................................................................................................... 39
4.3.10 RS 232 serial link check ........................................................................................... 39
4.3.11 Checking the software version. ................................................................................ 40
4.3.12 Checking the ADC.................................................................................................... 40
4.3.13 Checking the position sensor ................................................................................... 40
4.3.14 Buzzer test ............................................................................................................... 41
4.3.15 Displaying the calibration values .............................................................................. 41
4.3.16 Checking the syringe clamp ..................................................................................... 41
4.3.17 Checking the syringe group number ........................................................................ 42
4.3.18 Checking the disengagement................................................................................... 42
4.3.19 Checking the anti-siphon arm .................................................................................. 42
4.3.20 Checking backpressure............................................................................................ 42
4.3.21 Checking the end of infusion pre-alarm ................................................................... 43
4.3.22 Checking the linearity ............................................................................................... 44
4.3.23 Checking mains/battery operation............................................................................ 44
4.3.24 Battery autonomy test .............................................................................................. 45
4.3.25 Continuity test .......................................................................................................... 45
4.3.26 Regular inspection sheet.......................................................................................... 47
4.4 Flow rate control .................................................................................. 49
4.4.1 Measurement with a computer ................................................................................. 49
4.4.2 Measurement with scales......................................................................................... 51
4.4.3 Measurement using a test tube................................................................................ 53
4.5 Cleaning and disinfection ................................................................... 55
4.6 Storage.................................................................................................. 56

5 Diagnosis......................................................................................................... 57
5.1 Troubleshooting................................................................................... 57
5.2 Error messages.................................................................................... 59

6 Operation sheets............................................................................................. 61
N°1, Procedure: Display and central unit boards .................................... 63
N°2, Procedure: Syringe clamp ................................................................. 65
N°3, Procedure: Syringe detection system .............................................. 67
N°4, Procedure: Motor + Opto + Disk........................................................ 69
N°5, Procedure: Dynamometer sensor ..................................................... 71
N°6, Procedure: Plunger advance control potentiometer....................... 77

22 PiloteCE_A_enTDM.fm
N°7, Procedure: Plunger cover and/or disengagement lever
+ anti-siphon arm ........................................................................................ 79
N°8, Procedure: Power supply board........................................................ 81
N°9, Procedure: Battery holder and battery ............................................. 83
N°10, Procedure: Rear plug support ......................................................... 85
N°11, Procedure: Ribbon cable winding kit.............................................. 87
N°12, Procedure: Centering ring kit .......................................................... 91
N°13, Procedure: Flex circuit and tube kit ................................................ 95
N°14, Procedure: Upper and lower cases ............................................... 101

7 Calibration..................................................................................................... 105
7.1 Calibration procedure ........................................................................ 105
7.1.1 Calibration access .................................................................................................. 105
7.1.2 EtA.4 Calibration of the 3 battery voltage levels................................................ 106
7.1.3 EtA.6 Calibration of the position sensor. ........................................................... 106

8 Spare parts catalogue .................................................................................. 107


8.1 Upper case .......................................................................................... 107
8.2 Lower case.......................................................................................... 109
8.3 Plunger unit ........................................................................................ 111
8.4 Mechanical gear box .......................................................................... 115
8.5 Labels .................................................................................................. 117

PiloteCE_A_enTDM.fm 3
44 PiloteCE_A_enTDM.fm
1 Overview
1.1 General
The Pilot A is a syringe pump intended for the infusion of intravenous agents at a accurate,
low flow rate. The wide choice of syringes, the use of the universally recognised control
symbols and the synoptic display of the alarms contribute to making the Pilot A easy to use.
The adjustable occlusion detection, the correct positioning detection and the overall syringe
protection system guarantee optimum safety.
Its technical characteristics, the flow range from 0.1 to 120 ml/hr and its excellent accuracy
(+1% on the equipment) contribute to making the Pilot A, the ideal instrument for medical
services.

Overview

01_007a_en.fm 5
1.2 Overview diagram

Ext
12/15 V
15 W

Step by
DC-DC Motor
230 V Power supply ON / OFF step
converter driver motor

Battery

EPROM
EPROM Motor speed
128K x 8 sensor

RAM
RAM
8K x8 Syringue
Motor body
speed
sensor

EEPROM
EEPROM
512
Syringue
Antisiphonbody
sensor

BUS Antisiphon
sensor
SPI
CPU
CPU
CPU
Watch
dog Displacement
transducer

Interface Opto
UART
bus interface
Switchcall
Nurse
occlusion
(option)
ADC

Switch
Displacement
occlusion
transducer
Disengagement
switch
Nurse call
Disengagement
(option)
switch
LCD LCD
RS 232 driver display
Keyboard Disengagement
switch

LED LED
Master
Buzzer driver display
Overview

6 01_007a_en.fm
1.3 Precautions to be taken before use

The symbol ! in the Quick Start Guide recommends that the manual should be read in full
in accordance with standard EN 60601-1.
Fresenius Vial may in no case be held responsible for medical problems or any other
problems resulting from inadequate use of the equipment.
Refer to the User’s Instructions for further details.

1.4 Internal safety features


As soon as it is switched ON, the equipment activates a continuous function inspection
system. Any internal failure or any problem in the usage procedure is detected immediately.
Nevertheless, abnormal operation of the equipment with no obvious cause must always be
reported to the qualified technicians in your organization or our After Sales service.
If a fault occurs, an alarm is activated for any flow rate deviation greater than 5% in
comparison with the normal flow rate.
A second check activates an alarm in the event of deviation of 1 ml in comparison with the
anticipated infused volume, or if a flow rate greater than 20% is identified. The alarm is
triggered by the deviation detected first.
The Pilot is fitted with an internal battery to continue operation in the event of a power cut.
Furthermore, a safety fuse protects the mains from further disturbance.

1.5 Technical characteristics

1.5.1 Electrical
n Power supply: 230 V - 50-60 Hz.
n Max. consumption : 23 VAC.
n Fuse F2: 100 mAT 250 V IEC 127.
n Battery: 6 V - 1.1 Ah.
n External power supply: 12 - 15 V DC - 15 W.

1.5.2 Electronics
The Pilot syringe pump is fitted with 3 circuit boards:
n Motor power supply and control board.
n CPU board.
n Keyboard display board.

1.5.3 Mechanical characteristics


n Overall dimensions H x W x D: 120 x 330 x 155 mm.
n Weight: 2.2 kg approx.
Overview

01_007a_en.fm 7
1.5.4 Conformity, standards
n In compliance with the European Directive 93/42 EEC related to Medical Equipment:
CE0459.
n In compliance with the European Directive 89/336 EEC: Electromagnetic compatibility.
n Compliant with the standards EN 60601.2 and PrEN60601-2-24.
n Protection against leakage currents: CF type.
n Protection against electric shock: Class II.
n IP34 : Protection against ingress of liquid
Overview

8 01_007a_en.fm
2 Description and operation
2.1 Physical description

Fin positioning groove Anti-siphon arm

Plunger disengagement
control
Syringe clamp

Upper case

Lower case Front control panel

The Pilot A is composed of an upper case and a lower case.


n The upper case holds the syringe clamp and contains:
o A display board associated with the front control panel.
o A CPU board.
n The lower case contains:
o A power supply board and a storage battery.
o A mechanical base unit.
o A plunger unit.
Description and operation

02_007a_en.fm 9
2.1.1 The display board and the front panel
The display board is mounted under the front control panel and is fitted with all the items
required for man-machine interaction.
n Input keyboard.
n Control lamps and overview diagrams.
n 7-segment display units.

J1
J3

Solder side display board.

J2

Component side display board.


This board is connected to the different parts of equipment by means of connectors.

J1 connector to CPU board

Pin Description
1 SEG1 display matrix Line 1
2 SEG2 display matrix Line 2
3 SEG3 display matrix Line 3
4 SEG4 display matrix Line 4
5 SEG5 display matrix Line 5
6 SEG6 display matrix Line 6
7 SEG7 display matrix Line 7
8 SEG8 display matrix Line 8
9 COL1 display matrix Column 1
10 COL2 display matrix Column 2
11 COL3 display matrix Column 3
12 diode FAIL control Fail
13 COL/DIG 9 LED type control "
14 LIG1 keyboard interface Line 1
15 LIG2 keyboard interface Line 2
16 LIG3 keyboard interface Line 3
17 LDSECT lighting control Mains LED
Description and operation

18 +5V power supply


19 VBAT power supply
20 GND power supply

10 02_007a_en.fm
J2 connector to keyboard

Pin Description
1 Column 1
2 Column 2
3 Column 3
4 Column 4
5 Column 5
6 Column 6
7 Line 1
8 Line 2
9 Line 3
10 TON
11 TOFF
12 Gnd power supply

J3 connector to CPU board

Pin Description
1 TON on key
2 TOFF off key
3 SPI SI bus
4 SPI CLK bus
5 SPI CSLCD bus
6 BUZZER buzz control
7 VBAT power supply
8 Gnd power supply

Description and operation

02_007a_en.fm 11
2.1.2 CPU board
The CPU board holds an 80C32 microcomputer. It is mounted and connected to the display
board by connectors J4 and J5.
A ribbon cable connects this to a power supply board by means of a connector J3.

J2

J4

J3 J5

CPU board
J2 connector to fin detection switch and syringe detection opto-electronic sensor

Pin Description
1 Opto anode diode +5V
2 Common points between cathode diode and opto 1 and opto 2 transistor
emitters.
3 Opto 1 transistor collector
4 Opto 2 transistor collector

J3 connector to power supply board

Pin Description
1 +5V controlled power supply
2 Gnd power supply
3 +VBAT power supply
4 Gnd power supply
5 Phase A motor control
6 Phase B motor control
7 Phase C motor control
8 Phase D motor control
9 I signal motor control
10 BOOST signal motor control
11 Sopt1 opto rotation motor output
12 Sopt2 opto anti-siphon
13 n.u.
Description and operation

14 Cdopt1 opto rotation motor control output


15 Cdopt2 opto anti-siphon module control
16 OFF signal off key pressed
17 SECT mains supply presence signal
18 CDALIM power cut signal
19 LDSECT mains LED control
20 CTS clear to send Line 2

12 02_007a_en.fm
J3 connector to power supply board

Pin Description
21 DEB/OFF disengagement signal
22 RTS request to send Line 2
23 OCC/OFF stoppage signal
24 BUZ nurse call relay control
25 EOC end of adc conversion
26 CSADC spi adc bus select
27 CLK spi adc bus clock
28 SI data in spi adc bus
29 SO data out spi adc bus
30 CDANA analogue sensor power supply control
31 Rx2 receive data ttl Line 2
32 Tx2 transmit data ttl Line 2
33 Txd1 transmit data ttl Line 1
34 Rxd1 receive data ttl Line 1
35 TON on key
36 TOFF off key
37 +VBAT power supply
38 Gnd
39 +5V
40 Gnd

J4 connector to keyboard

Pin Description
1 SEG1 display matrix Line 1
2 SEG2 display matrix Line 2
3 SEG3 display matrix Line 3
4 SEG4 display matrix Line 4
5 SEG5 display matrix Line 5
6 SEG6 display matrix Line 6
7 SEG7 display matrix Line 7
8 SEG8 display matrix Line 8
9 COL1 display and keyboard matrix Column 1
10 COL2 display and keyboard matrix Column 2
11 COL3 display and keyboard matrix Column 3
12 FAIL LED control fail
13 RSCRT display power reduction control
14 LIG1 keyboard interface Line 1
15 LIG2 keyboard interface Line 2
16 LIG3 keyboard interface Line 3
Description and operation

17 LDSECT mains LED control


18 +5V power supply
19 VBAT power supply
20 Gnd power supply

02_007a_en.fm 13
J5 connector to display board

Pin Description
1 TON on key
2 TOFF off key
3 SI bus spi
4 CLK bus spi
5 CSLCD bus spi
6 BUZZ buzzer control
7 VBAT power supply
8 Gnd power supply
Description and operation

14 02_007a_en.fm
2.1.3 Power supply board and battery
The power supply board is mounted on the bottom of the lower case. It supplies all electronic
parts using the mains 230 V AC or the external 12 V DC. It also provides electrical supply to
the 1.1 or 1.2 Ah battery.

J6
J7
J8

J9
J5 F2
J3
J4 J1
J2 F1

Power supply board.


This board is connected to the different items of equipment by means of connectors.

J1 connector to mains power supply board

Pin Description
1 Neutral
2 Phase

J2 connector to CPU board

Pin Description
1 +5V controlled power supply
2 Gnd power supply
3 +VBAT power supply
4 Gnd power supply
5 Phase A motor control
6 Phase B motor control
7 Phase C motor control
8 Phase D motor control
9 I signal motor control
10 BOOST signal
11 Sopt1 opto rotation output module
12 Sopt2 opto anti-siphon output module
13 N.U
14 Cdopt1 opto rotation control module
Description and operation

15 Cdopt2 opto anti-siphon control module


16 OFF off signal key pressed (ON/OFF)
17 SECT mains supply presence signal
18 CDALIM power cut signal
19 LDSECT mains LED control
20 CTS clear to send
21 DEB/OFF disengagement signal active at 0

02_007a_en.fm 15
J2 connector to CPU board

Pin Description
22 RTS request to send
23 OCC/OFF occlusion signal active at 0
24 BUZ nurse call relay control
25 EOC end of ADC conversion
26 CSADC SPI ADC bus select
27 CLK SPI ADC bus clock
28 SI data IN SPI ADC bus
29 SO data out SPI ADC bus
30 CDANA analogue sensors power supply control
31 RX2 TTL receive data Line 2
32 TX2 TTL transmit data Line 2
33 TXD1 TTL transmit data Line 1
34 RXD1 TTL receive data Line 1
35 TOFF OFF key
36 TON ON key
37 +VBAT power supply
38 Gnd
39 +5V
40 Gnd

J3 connector to potentiometer

Pin Description
1 VREF
2 Centre point
3 Gnd

J4 connector to internal battery

Pin Description
1 + battery
2 - battery

J5 connector to motor

Pin Description
1 +VBAT
2 +VBAT
3 Phase D
4 Phase C
5 Phase B
Description and operation

6 Phase A
7 Opto rotation anode diode /+5V
8 Opto rotation cathode diode
9 Opto rotation transistor collector
10 GND/ opto rotation transistor emitter

16 02_007a_en.fm
J6 connector to RS232 and Master plugs

Pin Description
1 TX1 transmit data Line 1
2 +5V
3 RX1 receive data Line 1
4 Gnd
5 Interface validation
6 Nurse call relay common point
7 Nurse call relay normally open
8 Nurse call relay normally closed
9 CD ON external on
10 CD OFF external off
11 I-OPTON motor output control
12 I-SECT mains LED
13 +VBAT external power supply plug
14 RX2 receive data Line 2
15 TX2 receive data Line 2
16 Gnd
17 CTS
18 RTS

J7 connector to external DC power supply

Pin Description
1 ± external power
supply
2 ± external power
supply

J8 connector to disengagement micro-switch, force sensor and anti-siphon switch

Pin Description
1 +VREF + internal gauge bridge
2 E1 Internal gauge bridge input / occlusion ON
3 E2 Internal gauge bridge input / occlusion ON
4 Gnd – internal gauge bridge
5 Opto anti-siphon LED cathode
6 Opto anti-siphon LED anode / +5V
7 Opto anti-siphon transistor collector
8 Disengagement micro-switch ON
9 Disengagement micro-switch OFF
Description and operation

10 Gnd

Remember to dismount the ribbon cable holder on the power supply board before
extracting the mechanical assembly from the case (the ribbon cable may break).

02_007a_en.fm 17
J9 connector, test points

Pin Description
1 gnd
2 Position sensor output
3 Battery discharge control output
4 Amplified force sensor output
5 n.u.
6 Motor control optical switch output
7 Force and position sensor reference voltage
8 Piston head detection optical switch output

2.1.4 Mechanical base unit


The mechanical base unit is composed of a motor-reducer block driving a screw-and-nut unit.
At the shaft end, the motor receives a control panel associated with an opto-electronic switch.
The mechanical base unit also accommodates a potentiometer fitted with a rack pinion
system.

2.1.5 Mechanical plunger unit


The mechanical plunger unit is mounted onto the mechanical base unit. The latter carries out
the long travel guiding and displacement of the plunger by means of the screw-and-nut
system.
The plunger is fitted with a disengagement control allowing to separate it from the screw-and-
nut system.
Description and operation

18 02_007a_en.fm
2.2 Functional description
From a functional point of view, the Pilot A is composed of three sub-assemblies :
n A syringe position control and maintenance sub-assembly.
n A motorisation sub-assembly.
n An external connection sub-assembly.

2.2.1 Syringe control and maintenance sub-assembly


The syringe if fitted into the upper case and held in position by means of a syringe clamp.
Accurate positioning of fins is carried out by the syringe clamping groove.
Detection of the syringe size (60 cc or 20 cc) is carried out by two opto-electronic sensors
mounted onto the syringe clamp.
Associated with an opto-electronic sensor, the anti-siphon arm controls the piston position.
Composed of a micro-switch fitted to the plunger, an anti-occlusion system triggers an alarm
whenever force on the piston is excessive.

2.2.2 Motorisation sub-assembly


This sub-assembly moves the piston in the syringe.
It is put into motion by means of a motor-reducer unit associated with a screw-and-nut
system.
A motor rotation disk mounted on the shaft end of the motor and associated with an opto-
electronic sensor controls the rotation.
A potentiometer controls the plunger movement by means of a rack pinion system.
A micro-switch allows for control of the disengagement device.

2.2.3 External connection sub-assembly


The Pilot A has three connectors located at the rear end of the lower case:
n A 12-15 V DC, 15 W type external power supply connector.
n An optional RS 232 connector.

Description and operation

02_007a_en.fm 19
Description and operation

20
02_007a_en.fm
3 Description of the menus
3.1 Configuration menu of the basic operation parameters
The configuration menu enables users to adapt the Pilot to specific needs. It provides access
to the menus allowing for customisation of the parameters associated with basic operation
modes.
Fresenius Vial recommends users to implement the selected configuration procedures in the
presence of qualified personnel or an employee of the technical department.

It is possible to exit the configuration mode at any time by pressing the OFF key.

This menu enables users to:


n PAr1: Select the type of flow rate memorisation.
n PAr2: Select the syringe selection mode.
n PAr3: Modify the maximum flow rates which can be selected using the keyboard.
n PAr4: Configure the list of syringes that can be selected.
n PAr5: Select compulsory draining.
n Par8: Select the Master mode.
n PAr9: Select the RS232 communication speed.
n PArA: Select the empty syringe mode.

Description of the menus

03.2_007a_en.fm 21
3.1.1 Access to menus

Keys used
Key Function
ON, is used to switch the machine on.
OFF, is used to switch the machine off and, when pressed for over three
seconds, to exit the configuration mode.
SILENCE ALARME, is used to access the configuration mode of the
current operation parameters.
The selection keys allow to scroll the figures and letters on the tens,
units, tens segments etc.

VALIDATION, is used to validate a choice.

STOP, is used to cancel the current configuration.

Switch to configuration mode.


n Press "SILENCE ALARME" and "DIZAINE" (TENS)
simultaneously.
n Maintain this position while pressing "ON".

n When PAr is shown on the display unit, release the


selection of "SILENCE ALARME" and "DIZAINE" (TENS)
Validate within three seconds by pressing "VALIDATION".
n PAr1 is shown by default.
n Switching from PAr1 to PArA is carried out
using the "tens" keys.

3.1.2 PAr1, configuration of the memorisation type


This configuration enables users to choose whether or not to memorise the infusion flow rate
when the Pilot is shut down.
n PAr1, press "VALIDATION"
o If MEM, the flow rate is memorised when the Pilot is
shut down, this value will be displayed by default when the
machine is next switched on.
o If noME, the flow rate is not memorised, the default
value is 00.0 each time the machine is switched on.
Description of the menus

s Select the memorisation type using the selection


keys.

o By validating once again, the type is memorised and it is


possible to select another configuration .

22 03.2_007a_en.fm
The modification can be cancelled by pressing the "STOP" key.

It is possible to exit the configuration mode at any time by pressing the "OFF"
key.

3.1.3 Par2, configuration of the syringe selection type


This configuration enables users to choose the type of syringe selection.
n PAr2, press "VALIDATION"
o If SEL1, automatic scrolling of the various types of

syringues available, every 5 seconds.


o If SEL3, automatic validation of the only syringe that
can be selected.
o If SEL4, when the Pilot is switched on, the user needs
to select the type of syringe installed.

o Choose the selection type using the selection keys.

o By validating once again, the type is memorised and it is


possible to select another configuration .

When mode SEL3 is selected, and if there is a choice of more than one syringe, the
Pilot automatically moves onto the configuration of the list of syringes that can be
selected PAr4 when the machine is next switched on.

The modification can be cancelled by pressing the "STOP" key.

It is possible to exit the configuration mode at any time by pressing the "OFF"
key.

3.1.4 PAr3, configuration of the maximum flow rate that can be


selected by the keyboard.
This configuration enables users to choose the maximum flow rate that can be selected using
the keyboard for each type of syringe.

Syringe type Min. flow rate (ml/hr) Max. flow rate (ml/hr)
50/60 cc 0,1 120
20 cc 0,1 60
Description of the menus

n Par3, press "VALIDATION"

o Select the syringe type using the keys.


s 20c, 20 ml.
s 50c, 50 ml.

03.2_007a_en.fm 23
o Press "VALIDATION"

s Select the maximum flow rate using the keys.

o By validating once again, the type is memorised and it is


possible to select another configuration .

The modification can be cancelled by pressing the "STOP" key.

It is possible to exit the configuration mode at any time by pressing the "OFF"
key.

3.1.5 PAr4, configuration of the list of syringes that can be selected


This configuration enables users to choose whether or not it can be selected for each type of
active syringe.
n PAr4, press "VALIDATION"
o The LED of the syringe to be configured (50ml ou 20ml)
flashes.
s If SEL, this type of syringe can be selected, in the
configuration of the Pilot list.
s If noSEL, this type of syringe cannot be selected,
in the configuration of the Pilot list.

o Make your choice (whether it can be selected or not)


using the keys.

n Press "VALIDATION" to memorise the modification.


o The LED of the configured syringe is:
s On, if it can be selected.
s Off, if it cannot be selected.

Pressing the "tens" key once displays the name of the


syringe (see chapter 3 “Syringe/name match
table”).
Description of the menus

The modification can be cancelled by pressing the "STOP" key.

It is possible to exit the configuration mode at any time by pressing the "OFF"
key.

24 03.2_007a_en.fm
3.1.6 PAr5, configuration of the compulsory priming
This configuration enables users to choose whether or not draining is compulsory after the
syringe is selected.
n PAr4, press "VALIDATION"
o If PurG, compulsory draining, pressing "BOLUS"
during start-up is compulsory to be able to select the flow
rate.
o If noPu, draining is not compulsory, the flow rate may
be selected upon start-up straight after the syringe is
validated.

o Make your choice using the keys.

o By validating once again, the type is memorised and it is


possible to select another configuration.

The modification can be cancelled by pressing the "STOP" key.

It is possible to exit the configuration mode at any time by pressing the "OFF"
key.

3.1.7 PAr8, configuration of Master mode


This configuration allows to select between normal and Master modes.

In order to control a Pilot with a PC, users need to transmit the request using the
appropriate command (see the RS232 link protocol for the Pilote computer interface).

n Par8, press "VALIDATION"


o if -SP-, the pilot operates in a normal way.

o if MAST, the pilot operates in Master mode.


s At start-up, the MAST message flashes. The pilot is
expecting a command to switch to Master mode.
s The pilot keyboard is locked and an alarm is heard
whenever it is used.
Description of the menus

o By validating once again, the type is memorised and it is


possible to select another configuration.

The modification can be cancelled by pressing the "STOP" key.

It is possible to exit the configuration mode at any time by pressing the "OFF"
key.

03.2_007a_en.fm 25
3.1.8 PAr9, configuration of the RS232 communication speed
This configuration enables the user to select the communication speed of the RS232 link.
n PAr9, press "VALIDATION"
o If 19K2, speed at 19,200 Bauds.

o If 19K2, speed at 9,600 Bauds.

o If 4800, speed at 4,800 Bauds.

o Make your choice using the keys.

o By validating once again, the type is memorised and it is


possible to select another configuration.

The modification can be cancelled by pressing the "STOP" key.

It is possible to exit the configuration mode at any time by pressing the "OFF"
key.

3.1.9 PArA, configuration of the empty syringe mode


This configuration enables users to select a type of operation for the instrument using the
empty syringe mode.
n PArA, press "VALIDATION"
o If SVId, empty syringe mode activated.

o If SVId, empty syringe mode desactivated.

o Make your choice using the keys.

o By validating once again, the type is memorised and it is


possible to select another configuration.

The modification can be cancelled by pressing the "STOP" key.

It is possible to exit the configuration mode at any time by pressing the "OFF"
key.
Description of the menus

26 03.2_007a_en.fm
3.1.10 Syringe/name match table

Brand Capacity (ml) Name Number


BD Perfusion 50 BPf 1
BD Plastipak 60 BDL 2
BD Plastipak 20 BDL 3
Braun Omnifix 50 BrO 4
Braun Perfusor 50 BrP 5
Braun Omnifix 20 BrO 6
Didactic Line 60 DiL 7
Fresenius Injectomat 50 FrI 8
Map Gliss 50 MGL 11
Monoject 60 SMJ 12
Terumo 50 Trm 13
Terumo 20 Trm 14
Monoject 20 SMJ 15
Didactic Perfusion 60 DiP 16
Dipomed Spritze 50 Dis 17
Fresenius P Spritze 50 FrP 18
Tutoject type T 50 DsP 19
Ivac 50 IVa 20
Braun Perfusor 20 BrP 24

Description of the menus

03.2_007a_en.fm 27
Description of the menus

28
03.2_007a_en.fm
Configuration, calibration and inspection

3.2 Calibration menu

The calibration menu is reserved for authorised personnel only.


To determine the operation mode of the different calibrations, refer to the
“Calibrations” chapter

This menu allows for calibration:


n ETA4 : of the three levels of battery voltage alarms.
n ETA6 : of the displacement potentiometer.

Configuration, calibration and inspection

03.3_002a_en.fm 35
Configuration, calibration and inspection

36
03.3_002a_en.fm
Configuration, calibration and inspection

3.3 Service test menu


The service test menu is reserved for authorised personnel only. It enables users to perform a
series of Pilot inspections to validate its operation (see “Checks”chapter). This must be
carried out each time parts are replaced.

The Service tests may also be performed more easily and more quickly using a PC
with installed maintenance software (consult our After Sales Service).

The service test menu enables users to perform a series of 15 tests or checks:
n tSt1: Displays the running time and the maintenance date.
n tSt2: Indicator lights test.
n tSt3: Keyboard test.
n tSt4: Displays the battery voltage.
n tSt5: Displays the codes of the last 10 alarms.
n tSt6: Displays the total running time.
n tSt7: TTL serial link test.
n tSt8: RS 232 serial link test.
n tStA: Displays the software version.
n tStb: Displays the analogue input.
n tStC: Displays the plunger position.
n tStd: Buzzer test.
n tStE: Displays the calibration values.
n tStf: Displays the syringe type.
n tStG: Displays the syringe group.

Configuration, calibration and inspection

03.4_007a_en.fm 31
Configuration, calibration and inspection

32
03.4_007a_en.fm
4 Preventive maintenance
4.1 Recommendations
The Pilot syringe pump can only be inspected, serviced or repaired by Fresenius Vial or by
an authorised and appointed service. The qualified technicians in your establishment and our
After Sales Service should be notified of any abnormal operation of the device.
If a repair is necessary, send the instrument in its original packaging if possible with a precise
description of the observed fault, to the official dealer for Fresenius Vial.
For further information concerning troubleshooting or the usage procedure, please contact
our After Sales Service or our Sales Department.
Fresenius Vial is not liable for loss or damage to the equipment during transport to our After
Sales Service.

4.2 Maintenance schedule

4.2.1 Use beyond the framework of the departmental order

Frequency Name
12 months Carry out a servicing inspection.
3 years Replace the battery (see "Battery-holder and battery" operation sheet).

4.2.2 Use within the framework of the departmental order


When the equipment is used within the framework of the departmental order of October 3
1995, inspections are performed on a less frequent basis. This is due to the fact that the
equipment is inspected before each use.

Frequency Name
1st inspection in Perform the first servicing inspection.
the 3rd year
Then every 2 Perform a servicing inspection.
years
3 years Replace the battery (see "Battery-holder and battery" operation sheet).
Preventive maintenance

04.1&2_001a_en.fm 41
Preventive maintenance

42
04.1&2_001a_en.fm
Preventive maintenance

4.3 Checks
In order to perform equipment follow-up by means of preventive maintenance, a regular
servicing inspection is recommended every 12 months (see “Regular servicing sheet”).

To ensure the check procedure is carried out efficiently, recharge the battery
beforehand (16 hours).

4.3.1 Test access

Keyboard description.
Key Function
ON, is used to switch the machine on.
OFF, is used to switch the machine off when pressed for over three
seconds.
SILENCE ALARME, is used to access the test mode.

STOP, is used to cancel the test in progress.

VALIDATION, is used to validate a choice.

"PRESSURE LIMIT", is used for the backpressure test.

The selection keys allow to scroll the figures and letters on the tens,
units, tens segments etc.

Activate the service test.


n Press "SILENCE ALARME" and "UNITS" keys
simultaneously.

n Maintain this position while pressing "ON".

n When tSt. is displayed on screen, release selection of


the "SILENCE ALARME" and "UNITS" keys, then validate
within three seconds by pressing the "VALIDATION" key.
n By default, the equipment starts with test n°1 tSt.1
n By using the selection keys "+ or -", scroll the different tests
on the display unit.
Preventive maintenance

4.3.2 Visual check


Check the general appearance of the case and labels, and check for any traces of shock.

04.3_007a_en.fm 35
4.3.3 Running time and servicing inspection date
This test allows for n tSt.1, press the "VALIDATION" key.
display of the Pilot o If xxx.H: number of hours of use, 999 hours max.
running time since
o If xxx.J: number of days of use, 999 hours max.
its last serving
inspection. It also o If xxx.M : number of months, 999 months max.
allows for display (average duration of one month considered as 30 days).
and modification of
the “last” servicing n Press "VALIDATION" again to display the date of the last
inspection date. servicing inspection.
When the servicing o xx.d, day of the servicing inspection date.
inspection date is o Validate once again,
modified, the
s xx.m, month of the servicing inspection date.
running time is
reset. o By validating once again,
s xxx.x, year of the servicing inspection date.

Each time this information is read, the month, day and year of the
servicing inspection date may be modified by using the tens and
units "- and +" keys. This date will be stored in the EEPROM and
the running time will be reset.

o By validating once again, a different test may be selected.

The test may be stopped at any time by pressing the "STOP" key, and a
different test may be selected.

4.3.4 Indicator lights check


This test checks n tSt.2, press "VALIDATION".
the efficiency of o All LEDs, 7-segment display units are ON.
the indicator o By validating once again,
lamps, the display s The LEDs and display units are scrolled one by one
units and the front from left to right. (display of the LEDs, 7-segment
panel LCD screen. display unit by segment and then by sets of 8.
The text is OK if all indicator lamps are lit up.

The test may be stopped at any time by pressing the "STOP" key, and a
different test may be selected.
Preventive maintenance

36 04.3_007a_en.fm
4.3.5 Keyboard check
This test allows to n tSt.3, press "VALIDATION".
check that all o The message CLAV is displayed.
keyboard keys
o Keep each key pressed down, one by one,
function correctly.
s Check the name of the key displayed on the display
unit.

The name of each key is displayed as follows:

Display Selected key


SIA.L SILENCE ALARME

StoP STOP

VAL VALIDATION

boLu BOLUS

S-.. + tens

S.- + units

S..- + tens

S_.. - tens

S._. - unit

S_.. - tens

If two or more keys are pressed simultaneously, the display unit shows Err and three
beeps are heard, indicating an error.
The "OFF" key is not included in the keyboard test.

The test may be stopped at any time by pressing the "VALIDATION" key
for over 3 seconds. Another test can then be selected.

If the display is faulty, replace the display board (see “Display and CPU boards”).
Preventive maintenance

04.3_007a_en.fm 37
4.3.6 Checking the battery voltage
This test enables n tSt.4, press the "VALIDATION" key.
users to display the o The voltage is displayed in Volts.
battery voltage in
Volts and tenths of
o By validating once again, a different test may be selected.
a Volt.

4.3.7 Testing the last 10 alarms


This test enables n tSt.5, press the "VALIDATION" key.
users to display the o The code of the most recent alarm is displayed xxx.0.
last ten Pilot
alarms.
o Press the keys to display the following codes for xxx.1
to xxx.9.

o By validating once again, a different test may be selected.

Meaning of the codes:

Alarm Description Error Description


A10 Battery alarm E50 ADC access self-test error.
A11 Syringe clamp E60 Error on: verification of syringe parameter consistency (inconsistency
alarm of the syringe diameter in relation to the motor pitch for 0.1 ml
calculated at the time of syringe validation).
A12 End of infusion E70 Error on: motor frequency fault (motor pitch period calculated in
alarm relation to the syringe diameter and the flow rate selected, either too
low or too high).
A13 Volume limit alarm E80 Keyboard error or high electromagnetic interference (with V7.OC
memory only).
A14 Disengagement E01 Rotation control error.
alarm
A15 Piston head alarm E32 Error on: segment advance control
A16 Occlusion alarm E52 Error on: advance during compensation for play.
E72 Error on: advance over the whole length.
E03 Communication error

n Errors 10, 20, 30 and 40 cannot be stored in EEPROM.


n For a standard shutdown, the OF message is displayed.
n For a standard shutdown, Off is displayed with a flashing F.
Preventive maintenance

38 04.3_007a_en.fm
4.3.8 Total running time check
This test enables n tSt.6, press the "VALIDATION" key.
users to display the o If xxx.H: number of hours of use, 999 hours max.
total running time
o If xxx.J: number of days of use, 999 days max.
of the Pilot. It is
not possible to o If xxx.M : number of months, 999 months max.
modify this time (average duration of one month considered as 30 days).
manually. s By keeping the "tens" keys pressed down, the number
of times the equipment has been switched on is
displayed.

o By validating once again, a different test may be selected.

4.3.9 TTL serial link test

This check only applies to CE2 Pilots.

4.3.10 RS 232 serial link check


This test enables To perform this test, use a plug on which lines Rx and Tx, RTS and CTS are short-circuited (2
users to check the and 3, 7 and 8) on the RS232.
efficiency of the
RS232 (2691) n tSt.8, press the "VALIDATION" key.
serial link. o NoUA is displayed if the UART is absent, the test
cannot be performed.
o Operation test of the RTS and CTS lines.
s If the test is unsuccessful NoRC is displayed, the
test cannot go on.
o Lr is displayed.
s If LrEK is displayed, the test is successful.
s If LrEr is displayed, the test is unsuccessful.

o By validating once again, a different test may be selected.

It is impossible to carry out this test when the PC is in communication with the Pilot. In this
case, “OPEN” is displayed.
Preventive maintenance

04.3_007a_en.fm 39
4.3.11 Checking the software version.
Run this test to n tSt.A, press the "VALIDATION" key.
display the o ex. V07.0 is displayed:
software version
and revision s 07, software version number.
numbers. s .0, revision number
o Press on the tens keys to display the Eprom checksum.
Ex.: 738.C.
o Press on the other keys to display the software version
index. Ex.: b.
o By validating once again, a different test may be selected.

4.3.12 Checking the ADC


Run this test to n tSt.b, press the "VALIDATION" key.
read the result of o The result displayed is the first of the five analogue inputs
the conversion of
(from 0 to 4). Ex.: 318.0.
the five analogue
o The second type of result displayed corresponds to the
inputs and three
test inputs of the converter test based on the channel number, xxx.Lis
converter. test input 0, xxx.M is the mid-scale test and xxx.H
is the full-scale test for the ADC.

Use the "-" and "+" selection keys to move from one input to another.

4.3.13 Checking the position sensor


This test enables n tSt.C, using spacers ref. T300940E and T300775G,
users to display the o Position the spacer ref. T300940E and press
plunger position in "VALIDATION".
mm and tenths of a s In the Up position the LCD screen shows
mm. 115.0 + 0.5 mm.
o Position the spacer ref. T300940E and press
"VALIDATION".
s In the Down position the LCD screen shows
20.0 + 0.5 mm.
o By validating once again, a different test may be selected.

If the value is out of limits, recalibrate the position sensor (see “EtA.6 Position sensor
calibration.”).
Preventive maintenance

40 04.3_007a_en.fm
4.3.14 Buzzer test
Run this test to n tSt.d, press the "VALIDATION" key.
check that the
o The buzzer rings non stop and bEEp is displayed.
buzzer is working.
o By validating once again, a different test may be selected.

4.3.15 Displaying the calibration values


Run this test to n tSt.E, press the "VALIDATION" key.
display the
o bAT.1 is displayed alternately with its calibration
calibration values
stored in the value.
EEPROM. s Press a Tenths key to display the calibration number
of that value

o Press a Unit or Tens key to move onto other values.

Display Name
bat.1 Alarm and pre-alarm battery voltage 6.3 V

bat.2 Pre-alarm battery voltage 5.9 V

bat.3 Alarm battery voltage 5.7 V

HIG.H Displacement potentiometer with large 115.0 mm spacer

Lou Displacement potentiometer with small 20.0 mm spacer

0.G Force meter with 0 kg

St.G Force meter with 5 kg

4.3.16 Checking the syringe clamp


This test displays n tSt.F, press the "VALIDATION" key
the type of syringe using a 50 cc and 20 cc capacity syringe or the T301521 spacer.
fitted to the Pilot. o Place the syringe clamp in the higher position.
The display shows ---.-
o Position the 50cc syringe or the T301521 spacer, 50cc
mini/maxi. The display shows 50c.c
o Position the 20cc syringe or the T301521 spacer, 20cc
mini/maxi. The display shows 20c.c
o Place the syringe clamp in the lower position.
The display shows ---.-
Preventive maintenance

Capacities which are non-existent or non-configured in the EEPROM are displayed in the
form of ---.-.

o By validating once again, a different test may be selected.

04.3_007a_en.fm 41
4.3.17 Checking the syringe group number
This test allows to n tSt.G, press the "VALIDATION" key.
display the list of o The syringe group number programmed EEPROM is
syringes for which
displayed, e.g. G01.
the equipment has
been configured.
o By validating once again, a different test may be selected.

4.3.18 Checking the disengagement


To carry out this operation, exit the test mode, press “OFF”.

n Press "ON".
o Lift the disengagement lever.
s Check the mechanical disengagement alarm (red
LED at the end of the syringe diagram).
o Fit the 50 cc syringe onto the equipment, ensuring the fin
and plunger are in position.
o Release the disengagement lever.
s Check that there is no mechanical lever release
alarm.
s Check that the plunger is locked.

4.3.19 Checking the anti-siphon arm

This check must be carried out for every syringe provided with the equipment.

n Check the functionality.


o Free travel, with no shaft play, and no dismounting of the
latter.
n Check that there is an alarm by locking only the syringe body
in the clamp.
o Anti-siphon arm in the upper position.
o Anti-siphon arm in the lower position.

Check that no alarm goes off with spacers T301519 and T301518.

4.3.20 Checking backpressure


To carry out this operation, exit the test mode, press “OFF",

n Press "ON".
o Fit the 50 cc syringe on the equipment and lock it into
Preventive maintenance

position using the clamp, ensuring the fin and plunger are
correctly positioned.
o Place the manometer (or any other pressure
measurement equipment) at the syringe outlet.
o Select a 50 ml "B-D PLASTIPAK" syringe by pressing
"VALIDATION".
o Check the position of the backpressure adjustment button .

42 04.3_007a_en.fm
o Select a maximum flow rate of 120ml/h and initiate the
infusion by pressing the “VALIDATION” key.
s Ensure there is no acoustic and visual alarm
(backpressure LED off).
s Check that the infusion LEDs flash correctly.
s Ensure the alarm is triggered for a value of 0.7 bar +
0.1 bar.
o Stop the infusion cycle by pressing the "ARRET
PERFUSION" ("STOP INFUSION") key.

o Run this test again, adjusting the backpressure level to


minimum and maximum levels.
s Min. backpressure = 0.4 bar + 0.1 bar.
s Max. backpressure = 1.2 bar + 0.2 bar.

The backpressure button should not be raised once the backpressure alarm threshold
has been adjusted with the adjustment screw.

4.3.21 Checking the end of infusion pre-alarm


To carry out this operation, exit the test mode, press “OFF",

n Press "ON".
o Fit the syringe on the equipment, ensuring the fin and
plunger are in position.
o Check that the correct syringe is detected, e.g. 50c.c.
o Select a 50 m/l "B-D PLASTIPAK" syringe filled to 20 ml.
o Select a flow rate of 120 ml/h.
s For normal flow rates, the pre-alarm is activated 5
minutes before end of infusion.
s Flow rate example: > 72 ml/h, the pre-alarm is
activated when the remaining volume equals 10% of
the total syringe capacity.
s Ensure the end of infusion pre-alarm is present.
o Press the "SILENCE ALARME" and "TENS"
simultaneously.
s The acoustic alarm is silenced and the visual signal is
maintained.
o Measure the “hard height” at “end of infusion”.
s 18.5 < x < 19.5.

Top view

For accurate checking of the "hard height", do not move the plunger when measuring.
Preventive maintenance

If the "hard height" reading is out of limits, recalibrate the position sensor (voir “EtA.6
Position sensor calibration.”).

04.3_007a_en.fm 43
4.3.22 Checking the linearity
To carry out this operation, exit the test mode, press “OFF",
n Equipment required: Chronometer, calliper square, BD
Plastipak 50 ml. syringe.

n Press "ON".
o Fit the equipment with the "B-D PLASTIPAK" 50 ml
syringe, ensuring the fin and plunger are in starting
position.
o Measure the distance X in mm.

Top view

o Check that the correct syringe is detected, e.g. 50c.c.


o Select a B-D PLASTIPAK" syringe filled to 50 ml.
o Select a flow rate of 50 ml/h.
o Press "VALIDATION" to start infusion and simultaneously
start the chronometer

o At 50 minutes, stop the infusion by pressing "STOP" and


measure the distance X2.
o Ensure X = X1 - X2 lies between
74.96 mm < X < 76.47 mm.

For accurate checking of the linearity do not move the plunger when measuring.

4.3.23 Checking mains/battery operation


To carry out this operation, exit the test mode, press “OFF",
n Connect the equipment to a mains supply.
o Check the operation of the mains presence Led (indicator
in the shape of a plug).
n Connect the equipment to a test power supply.
o Disconnect the equipment from the mains.
o Remove the battery holder.
o Remove the battery.
o Remove the connection lugs.
o Connect the battery lugs to a stabilised supply set at
6.3 V. Mind the polarities.
o Press "ON".
Preventive maintenance

o Select a syringe from the syringe list proposed by the


equipment.

o Press "VALIDATION"
o Select a flow rate and validate.

44 04.3_007a_en.fm
o Reduce the test power supply voltage until the battery
discharge is triggered.
s Ensure this is triggered between 5.8 V and 6 V.

The acoustic alarm can be temporarily silenced by pressing "SILENCE


ALARME" (2 minutes).

o Reduce the test power supply voltage again until the


battery discharge alarm is triggered.
s Ensure this is triggered between 5. 6V and 5.8 V.

If the results obtained during the tests differ from the values indicated, recalibrate the battery
voltage levels (see “EtA.4 Calibration of the 3 battery voltage levels.”).

4.3.24 Battery autonomy test


n Recharge the battery for 15 hours.

(100 % of capacity) Battery life is 5 hours minimum, with a 50 ml BD Plastipak syringe,


with a flow rate of 5ml/h.

n Run a battery life test greater than one hour.


o Select a flow rate of 120 ml/h for a 50 ml/h B-D Plastipak
syringe, then validate.
o The "battery discharge" pre-alarm informs the user on the
remaining time, between 60 minutes to 5 ml/h.
s The "total battery discharge" will sound before the
infusion is stopped.

4.3.25 Continuity test


Using a multimeter.
To carry out this operation, exit the test mode, press “OFF",
n Connect the multimeter to an ohmmeter.
o Check the electrical resistance shown by the ohmmeter is
over 10 MΩ:
s between phase and metal tube.
s between neutral and metal tube.
Preventive maintenance

04.3_007a_en.fm 45
Preventive maintenance

46
04.3_007a_en.fm
Preventive maintenance

4.3.26 Regular inspection sheet


Use this table to note the results of the different tests.

Equipment type: Code: Equipment series N°:

Conformity
N° Procedure Resulting value
Yes No
1 n Check the general condition of the case and its labels.
2 n Display total running time, tSt1
(in hours, days or months):***************************************** ..........................
3 n Display the last servicing inspection date, tSt1
(in days, months or years):***************************************** ..........................
4 n Check all indicator lights, tSt2.
5 n Check the keyboard, tSt3.
6 n Display the total running time, tSt6
(in hours, days or months):***************************************** ..........................
7 n Check the position sensor, tStC:
o High position with T300940 spacer, check that the displayed
value is 115.0 + 0.5 mm: ********************************** ..........................
o Low position with T300775 spacer, check that the displayed
value is 20.0 + 0.5 mm: ************************************ ..........................
8 n Check the syringe clamp, tStF:
o Syringe clamp in high position, check that the displayed
value is ---.- :******************************************** ..........................
o Position the 50cc syringe or the T301521 spacer, 50cc mini/maxi.
Check that he value displayed is 50c.c :**************** ..........................
o Position a 20cc syringe or a T301521 spacer, 20cc mini/maxi.
Check that he value displayed is 20c.c :**************** ..........................
o Syringe clamp in low position, check that the displayed
value is ---.- :******************************************** ..........................
9 n Check the anti-siphon arm:
o Free travel without end play.
o Presence of the alarm in high and low position.
o No alarm in presence of 20 cc and 50 cc syringe piston or T301519 et
T301518 spacers.
10 n Check the backpressure (use a Fresenius Vial dynamometer):
o Position of adjustment knob
Preventive maintenance

0.6 < X bar < 0.7: ****************************************** ..........................


11 n Check the end of infusion pre-alarm:
o Eg.: for a flow rate of 50 ml/hr with a 60 cc BD, check that
the pre-alarm is triggered at 5 mn + 10 s before the end
of infusion: ******************************************************* ..........................

fiche_006a_en.fm 47
Conformity
N° Procedure Resulting value
Yes No
12 n Check the end of infusion alarm:
o With a 60 cc BD, check that the hard height range
is 18.5 < x mm < 19.5: ****************************************** ..........................
13 n Check the linearity (60 cc BD Plastipack):
o Measure the plunger starting position range, XI mm:********* ..........................
o Measure the plunger range after 50 mn pump running, X2 mm: ......................
o Check: 74.96 < X1-X2 mm < 76.47: *************************** ..........................
14 n Check the battery autonomy:
o Recharge the battery for 15 hours.
o Operate the Pilot for 1 hr at a flow rate of 120 ml/hr:********* ..........................
15 n Carry out the electrical tests according to standard EN 60601-1

Name: Date: Signature:

Comments:
Preventive maintenance

48 fiche_006a_en.fm
Diagnosis

4.4 Flow rate control

4.4.1 Measurement with a computer


ISDébit software is required for measuring the flow with a computer. This software is the
property of Fresenius Vial. Please contact our After Sales Service for any further
information.
The operation mode defined by this software follows the project protocol of standard
PrEN-60-601-2-24 for infusion pumps. It is up to the user to adapt this procedure to the
software used.

In order to purchase the ISDébit flow rate control software, contact the Fresenius Vial
customer service.

The test procedure defined below must be carried out using a 50 ml or 20 ml syringe.

Equipment required
n Scales coupled to a microcomputer:
Flow rate value Scales sensitivity
x < 5 ml/hr 1/10000th
5 ml/hr < x < 30ml/hr 1/1000th
x > 30 ml/hr 1/100th
n Multi scales acquisition programme.
n Test tube or beaker with 1 ml graduating
n Liquid: distilled water and oil
n Luer lock type plastic syringe (50 or 20 ml).
n Catheter extension with Luer Lock (length 100 cm, inside diameter 2,5 mm).
n Needle:
Flow rate value Needle type
x < 30ml/hr G26
x > 30 ml/hr G18 or G21

Installation
n According to the installation drawings shown below.

1 cm

Cord
Anti vibration measurement table
RS232

Make sure the horizontal installation plane is respected.


Diagnosis

04.4_001a_en.fm 59
n Fill the syringe with 50 ml of distilled water.
o Prime to eliminate any air bubbles.
n Secure the female Luer Lock end piece of the catheter
extension onto the syringe and the male Luer lock end piece
onto the needle.
n Fit the syringe onto the device.
n Fill the test tube ensuring the needle is dipped in the liquid
(> 1 cm).
o Add several drops of oil to create a greasy film on the
surface of the liquid. This way the user will avoid any
measurement error due to evaporation of the liquid.
n Place the test tube in the centre of the scales platform.
n Place the needle inside the test tube.

The infusion line (needle/catheter extension) must not rest on the scales/test tube
assembly.

n Press "ON" (device in mains supply mode).


o Prime the infusion line using the "BOLUS" key.
o Check that there are no air bubbles.

Operating mode

The software works following the operating mode described in the PrEN-60-601-2-24
standard project for infusion pumps.

n Start the acquisition programme for the scales.


n Enter the necessary data to launch the programme without
validating the flow rate.
n Adjust the scales to the specified flow rate.
n Confirm the flow rate on the microcomputer so that the
automatic setting of the scales can take place.
n Start the infusion by pressing the "CONFIRM" key, when
00.00 appears on the scales display screen.
n When the specified time is over, note the error percentage
displayed on the screen.
Diagnosis

60 04.4_001a_en.fm
4.4.2 Measurement with scales

Equipment required
n Stop clock
n Scales
Flow rate value Scales sensitivity
x < 5 ml/hr 1/10000th
5 ml/hr < x < 30ml/hr 1/1000th
x > 30 ml/hr 1/100th
n Test tube or beaker with 1 ml graduating
n Liquid: distilled water and oil
n Luer lock type plastic syringe (50 or 20 ml)
n Catheter extension with Luer Lock (length 100 cm, inside diameter 2,5 mm).
n Needle:
Flow rate value Needle type
x < 30ml/hr G26
x > 30 ml/hr G18 or G21

Installation
n According to the installation drawings shown below.

1 cm

Anti vibration measurement table

Make sure the horizontal installation plane is respected.

n Fill the syringe with 50 ml of distilled water.


n Prime to eliminate any air bubbles.
n Secure the female Luer Lock end piece of the catheter
extension onto the syringe and the male Luer lock end piece
onto the needle.
n Fit the syringe onto the device.
n Fill the test tube ensuring the needle is dipped in the liquid
(> 1 cm).
n Add several drops of oil to create a greasy film on the
surface of the liquid. This way the user will avoid any
measurement error due to evaporation of the liquid.
n Place the test tube in the centre of the scales platform.
n Place the needle inside the test tube.
Diagnosis

04.4_001a_en.fm 61
The infusion line (needle/catheter extension) must not rest on the scales/test tube
assembly.

n Press "ON" (device in mains supply mode).


o Prime the infusion line using the "BOLUS" key.
o Check that there are no air bubbles.

Operating mode
n Select a flow rate.

For low flow rates (< 5 ml/hr), validate and wait for the infusion to stabilise for 1 hour. For
higher flow rates, wait for 10 to 30 minutes after infusion.

n Set the scales at 00.00 g.


n Start infusion by pressing the "CONFIRM" key, and start the
stop clock at the same time, (if necessary make a note of the
stop clock start value).
n Press the "STOP INFUSION" key to stop the test one hour
later.
n Note the value in grams of the “infused” liquid.
n Calculate the difference between the design value and the
real value.

1 gram = 1 ml.

n The error percentage can be calculated from this difference :


( Real value – Design value )
--------------------------------------------------------------------------- × 100 = Error percentage
Design value
Diagnosis

62 04.4_001a_en.fm
4.4.3 Measurement using a test tube

Equipment required
n Stop clock.
n Test tube or beaker with 1 ml graduating
n Liquid: distilled water and oil
n Luer lock type plastic syringe (50 or 20 ml)
n Catheter extension with Luer Lock (length 100 cm, inside diameter 2,5 cm).
n Needle:
Flow rate value Needle type
x < 30ml/hr G26
x > 30 ml/hr G18 or G21

Installation
n According to the installation drawings shown below.

1 cm

Anti vibration measurement table

Make sure the horizontal installation plane is respected.

n Fill the syringe with 50 ml of distilled water.


o Prime to eliminate any air bubbles.
n Secure the female Luer Lock end piece of the catheter
extension onto the syringe and the male Luer lock end piece
onto the needle.
n Fit the syringe onto the device.
n Fill the test tube ensuring the needle is dipped in the liquid
(> 1 cm).
o Add several drops of oil to create a greasy film on the
surface of the liquid. This way the user will avoid any
measurement error due to evaporation of the liquid.
o Place the needle inside the test tube.
Diagnosis

04.4_001a_en.fm 63
Operating mode
n Select a flow rate.

For low flow rates (< 5 ml/hr), validate and wait for the infusion to stabilise for 1 hour. For
higher flow rates, wait for 10 to 30 minutes after infusion.

n Start infusion by pressing the "CONFIRM" key, and start the


stop clock at the same time, (if necessary make a note of the
stop clock start value).
n Once the whole infused syringe is in the test tube, calculate
the difference between the design value and the real value:
Real flow rate = 50 ml/time in hours
n The error percentage can be calculated from this difference :
( Real value – Design value )
--------------------------------------------------------------------------- × 100 = Error percentage
Design value
Diagnosis

64 04.4_001a_en.fm
Diagnosis

4.5 Cleaning and disinfection


The syringe pump is part of the patient's immediate environment. It is advisable to clean and
disinfect the external surfaces of the device on a daily basis in order to protect both patient
and personnel from any risks of contamination.
n Disconnect the power cable from the wall socket before cleaning.
n Do not place in an AUTOCLAVE or IMMERSE the device, and do not allow liquid to
penetrate inside the equipment case or power supply cover.
o Use a cloth soaked in DETERGENT-DISINFECTANT, diluted in water if necessary, to
eliminate micro organisms.
o Avoid excessively abrasive brushing that could scratch the case.
o Do not rinse or wipe the surfaces.
n If the equipment is used in a department with severe contamination risks, after disinfecting
by wiping with a damp cloth, equipment should be left in the room during aerial disinfection.

Do not use TRICHLOROETHYLENE-DICHLOROETHYLENE.

n ¨TRICHLOROETHYLENE-DICHLOROETHYLENE:
o AMMONIA.
o AMMONIUM CHLORIDE
o CHLORINE AND AROMATIC HYDROCARBON.
o ETHYLENE DICHLORIDE-METHYLENE CHLORIDE
o CETONE based cleaning products.
These aggressive agents could damage the plastic parts and lead to apparatus malfunctions.

Also beware of ALCOHOL SPRAYS (20% -40% alcohol) that tarnish and crack the
plastic and fail to provide the cleaning action required prior to disinfection.

For further information, please contact the competent department in your hospital for supply
of the appropriate cleaning and disinfecting products.

Diagnosis

04.5&6_001a_en.fm 65
4.6 Storage
In case of prolonged storage, it is advisable to disconnect the battery using the battery access
door on the bottom of the device. This operation should be done by an experienced
technician.
The equipment must be stored in a dry and cool place.
n The recommended environmental temperature conditions for storage of the equipment are
between 0°C and 40°C.
n Relative humidity tolerated: max. 85%, no condensation.

Fully recharge the battery before using the equipment to avoid any risks caused by micro
power cuts in the mains supply and to ensure maximum autonomy.
Diagnosis

66 04.5&6_001a_en.fm
5 Diagnosis
5.1 Troubleshooting

Problem Cause Solution


End of infusion detected too n The syringe used does not n Change the syringe.
early (at approximately 10 ml). correspond to that selected.
No pre-alarm or alarm at end of
infusion.
Too much flow rate or
displacement control drift.
Occlusion alarm upon start-up n Inappropriate calibration of the force n Recalibrate the force sensor
sensor. (see “EtA.9 Force sensor
calibration”).

n Force sensor out of order. n Check the force sensor (see


“EtA.9 Force sensor
calibration”).

n Ribbon cable cut. n Replace the ribbon cable (see


“Ribbon cable winding kit”
operation sheet).
Occlusion alarm during n Pressure limit selected is too low. n Select a medium pressure
operation. limit.

n Inappropriate calibration of the force n Recalibrate the force sensor


sensor. (see “EtA.9 Force sensor
calibration”).

n Ribbon cable cut. n Replace the ribbon cable (see


“Ribbon cable winding kit”
operation sheet).
Disengagement alarm upon n Faulty disengagement micro-switch. n Replace the micro-switch
start-up (see “Syringe detection system”
operation sheet)
Syringe piston detection alarm n Faulty photoelectric cell and/or n Check the syringe clamp (see
not justified. syringe piston obturator. “Checking the syringe clamp”).
Syringe body clamp alarm not
justified.
Display fault: segments, LEDs n Control transistors and/or display n Check the display (see
board connections. “Checking the indicator lights”).
n Check the connectors.
After a fall. n Damaged mechanical elements. n Check that the input bearing
and mechanical assembly are
intact.
Diagnosis

05.1_003a_en.fm 57
Diagnosis

58 05.1_003a_en.fm
Diagnosis

5.2 Error messages

Error code Description Recommended action


Electronic control and software* errors
Er10 n Internal RAM error. n Reconfigure the Pilot (see “Basic
operation parameter configuration menu”).

Er20 n Enternal RAM error.

Er30 n EEPROM error.

Er40 n EEPROM error.

Er50 n ADC error. n Check the ADC (see “Checking the


ADC”).

Er60 n Syringe parameter error. n Reconfigure the Pilot (see “Current


operation parameter configuration menu”)

Er70 n Motor frequency error.

Er80 n Faulty keyboard. n Check the keyboard.


Version n Short circuit in the
7.OC only keyboard.

n Electromagnetic n Check the operation environment of the


disturbance exceeding Pilot.
standard limits.
* : When rewriting the EEPROM, when the equipment is switched off, the checksum is
rewritten in the memory to save the parameters.
If the Hardware cut-off delay is shorter than the software cut-off delay, the equipment is
switched off before the EEPROM is fully rewritten: Checksum not compliant.
Err(-)0 or CFPc: When the equipment is in CFPc, reconfiguration is compulsory: Faulty
WATCH DOG.
Motor errors
Er01 n Motor control failure. n Check the motor power supply.
n Motor fault. n Replace the motor.
Plunger advance errors
Er32 n Error over a short n Check the connectors.
distance. n Check that the potentiometer is
tightened.
Er52 n Error during
compensation for play. n Check the ADC (see “Checking the
ADC”).
Er72 n Error over the whole n Check the position sensor (see
length. “Checking the position sensor.”).
Er82 n Error in relation to the
flow rate.
Diagnosis

05.2_004a_en.fm 59
Error code Description Recommended action
Calculation parameter errors (motor and flow rate)
Er14 n Motor period calculation n Check the ADC (see “Checking the
error. ADC”).
n Check the position sensor (see
Er24 n Motor rotation direction
“Checking the position sensor.”).
error.

Er34 n Flow rate/period


calculation error.

Er44 n UART and micro-


controller crystal frequency
error.
Configuration errors
CFPc n The configuration self- n Reconfigure the Pilot using the IS
test upon start-up was not Control software.
satisfactory.
Diagnosis

60 05.2_004a_en.fm
6 Operation sheets
This chapter lists the set of dismounting/remounting sheets.

Operation sheets

06.0_001a_en.fm 71
Operation sheets

72 06.0_001a_en.fm
Operation sheets

N°1, Procedure: Display and central unit boards

Safety:
For safety reasons, the technician should not carry out any maintenance when the device is
connected to the 230 V mains supply voltage.
Disconnect the mains power supply cable.

Equipment required:
n 1 Posidriv Z1 screwdriver.
n 1 antistatic wriststrap.

Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).

Procedure:

Access
n Turn the Pilot over onto the upper case.
n Unscrew the 6 Phillips head screws (ref. 1) located at the bottom of the lower case, which
link this to the upper case.
n Keep the Pilot assembled and turn it over onto the lower case.
n Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.
n Disconnect the CPU board flat jumper.

Hands must not come into contact with the CPU boards.

1 1
1

1
Operation sheets

06.0_0010a_en.fm 73
Dismounting

When electronic components are handled, it is recommended to wear an antistatic


wriststrap linked to ground and to work on an antistatic foam mat.

n Unscrew the 5 Phillips head screws (ref. 2) located at the display board, which link this to
the upper case.
n Remove the board insulator (ref. 3) located on the left.
n Lift the display board slightly and remove the display unit flat jumper (ref. 4).
n Disconnect the black connector (ref. 5).
2
2

Remounting

A specific type of board corresponds to each Pilot "CPU and display board" ; It is
important to avoid reversing the references between each Pilot and order the part
number corresponding to your device.
Carry out the same procedures in reverse to reassemble the parts.

When mounting the display board, it is important to reduce the torque in the plastic inserts
so as to avoid causing damage to these.
Operation sheets

When fitting the upper case, ensure the joint is perfectly positioned in its slot after
reassembly.
Recalibrate the Pilot (see "Calibrations") then carry out the regular servicing tests (see
"Regular servicing sheet").

74 06.0_0010a_en.fm
Operation sheets

N°2, Procedure: Syringe clamp

Safety:
For safety reasons, the technician should not carry out any maintenance when the device is
connected to the 230 V mains supply voltage.
Disconnect the mains power supply cable.

Equipment required:
n 1 Posidriv Z1 screwdriver.
n 1 Posidriv Z1 flat screwdriver.

Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).

Procedure:

Access
n Turn the Pilot over onto the upper case.
n Unscrew the 6 Phillips head screws (ref. 1) located at the bottom of the lower case, which
link this to the upper case.
n Keep the Pilot assembled and turn it over onto the lower case.
n Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.
n Disconnect the CPU board flat jumper.

Hands must not come into contact with the CPU boards.

1 1
1

1
Operation sheets

06.0_0020a_en.fm 75
Dismounting
n Remove the spring retaining ring (ref. 2) located on the syringe clamp shaft, which holds
this to the CPU support.
n Remove the obturator and the spring (ref. 3).
n Remove the syringe clamp.

Reassembly
Carry out the same procedures in reverse to remount the unit.

When fitting the upper case, ensure the joint is perfectly positioned in its slot after
reassembly.

Perform the regular servicing tests (see "Regular servicing sheet").


Operation sheets

76 06.0_0020a_en.fm
Intervention sheets

N°3, Procedure: Syringe detection system

Safety:
For safety reasons, the technician should not carry out any maintenance when the device is
connected to the 230 V mains supply voltage.
Disconnect the mains power supply cable.

Tools required:
n 1 Posidriv Z1 screwdriver.
n 1 antistatic wriststrap.
n 1 soldering iron
n "RADIEL Sn60Pb RI 1" welding wire (cleaning not required for rewelding) or equivalent.
n Silicone ref. 161249.

Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).

Opto replacement procedure:

Access
n Rotate the Pilot on the upper case.
n Unscrew the 6 Phillips head screws (ref. 1) Located at the bottom of the lower case, which
tighten the latter to the upper case.
n Keep the Pilot assembled and turn it over onto the lower case.
n Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.
n Disconnect the CPU board flat jumper.

Hands must not come into contact with the CPU boards.

1 1
1

1
Intervention sheets

06.0_0031a_en.fm 67
Dismounting
n Disconnect the connector (ref. 2).
n Unscrew the Phillips head screw (ref. 3) which holds the conductor to the upper case.
n Unscrew the 2 Phillips head screws (ref. 5) which hold the opto PCB in position.
n Remove the opto assembly.

5
Green
Black
2 White
Red

Reassembly
n Place the new opto assembly on the PCB support.
n Screw the 2 Phillips head screws (ref. 5) which hold the “PCB and opto” assembly in
position.

When mounting the "PCB and opto", it is important to reduce the torque in the plastic
inserts so as to avoid causing damage to these.

n Connect the conductor to the CPU board.


n Tighten the Phillips head screw holding the conductor (ref. 2).
Carry out the same procedures in reverse to remount the unit.

When fitting the upper case, ensure the joint is perfectly positioned in its slot after
reassembly.

Perform the regular servicing tests (see “Regular servicing sheet”).


Intervention sheets

68 06.0_0031a_en.fm
Operation sheets

N°4, Procedure: Motor + Opto + Disk

Safety:
For safety reasons, the technician should not carry out any maintenance when the device is
connected to the 230 V mains supply voltage.
Disconnect the mains power supply cable.

Equipment required:
n 1 Posidriv Z1 screwdriver.
n 1 antistatic wriststrap.

Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).

Procedure:

Access
n Turn the Pilot over onto the upper case.
n Unscrew the 6 Phillips head screws (ref. 1) located at the bottom of the lower case, which
link this to the upper case.
n Keep the Pilot assembled and turn it over onto the lower case.
n Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.
n Disconnect the CPU board flat jumper.

Hands must not come into contact with the CPU boards.

1 1
1

1
Operation sheets

06.0_0041a_en.fm 83
Dismounting
n Press the syringe pump disengagement lever towards the back of the Pilot as far as
possible.
n Maintain this position and slide the whole unit to the back left.
n Remove the plunger guide.
n Disconnect the black connectors (ref. 2).
n Unscrew the flat jumper mounting lug (ref. 3) without removing it.
n Remove the mechanical plunger unit from its slot.
n Disconnect the blue connector from the flat jumper (ref. 4).
n Remove the mechanical plunger unit.

n Unscrew the 2 Phillips head screws (ref. 5).


n Remove the motor, opto and disk assembly.
5

Reassembly
Carry out the same procedures in reverse to remount the unit.
Operation sheets

When fitting the upper case, ensure the joint is perfectly positioned in its slot after
reassembly.

Perform the regular servicing tests (see “Regular servicing sheet”).

84 06.0_0041a_en.fm
Operation sheets

N°5, Procedure: Dynamometer sensor

Safety:
For safety reasons, the technician should not carry out any maintenance when the device is
connected to the 230 V mains supply voltage.
Disconnect the mains power supply cable.

Equipment required:
n 1 Posidriv Z1 screwdriver.
n 1 hexagon socket key (2.5).
n 1 soldering iron.
n "RADIEL Sn60Pb RI 1" welding wire (cleaning not required for re-welding) or equivalent.

Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).

Procedure:

Access
n Turn the Pilot over onto the upper case.
n Unscrew the 6 Phillips head screws (ref. 1) located at the bottom of the lower case, which
link this to the upper case.
n Keep the Pilot assembled and turn it over onto the lower case.
n Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.

Hands must not come into contact with the CPU boards.

1 1
1

1
Operation sheets

06.0_0051a_en.fm 85
Dismounting
n Press the syringe pump disengagement lever towards the back of the Pilot as far as
possible.
n Maintain this position and slide the whole unit to the back right.
n Remove the plunger guide.
n Unscrew the 3 Phillips head screws (ref. 2) located at the bottom of the plunger end shield
which links this to the plunger cover.
n Remove the plunger end shield.

2
2

n Remove the backpressure adjustment button (ref. 3).


n Remove the disengagement lever and its spring (ref. 4) as well as the protective plunger
film (ref. 5).
n Remove the plunger cover clip (ref. 6).

5
4

6
Operation sheets

86 06.0_0051a_en.fm
n Unscrew the 2 Phillips head screws (ref. 7 and 8) which link the plunger support to the
plunger cover.
n Unscrew and remove the Phillips head screw and the washer (ref. 9) which attach the anti-
siphon cam to the anti-siphon arm.
n Remove the cam/arm assembly.
n Remove the plunger cover fitted with the bonding pad.
8

7 9

n Remove the retaining ring that holds the anti-siphon opto in place (ref. 10).
n Unscrew the 6 hexagon socket screws (ref. 11).

11
10

Operation sheets

06.0_0051a_en.fm 87
n Remove the backpressure micro support/spring leaf (ref. 12).
n Pull off the ribbon cable from the double-sided board

12

ribbon cable stuck


to the double-sided
board

Welding and unwelding operations are carried out using a soldering iron fitted with a tip in
good condition, constantly tapered and kept clean. The temperature of the iron should be
between 315°C and 340°C.

n Unweld the sensor wires on the flex circuit.

When dismounting the force sensor, ensure the welding pellets are not damaged.

o Add more weld to facilitate the unwelding process.


o Heat and pull on the wires one by one.
Operation sheets

88 06.0_0051a_en.fm
Reassembly
n Weld the ribbon cable to the new pressure switch, ensuring the mounting procedure is
carried out in order:
o Taper the 3 welding points.
o Weld the 3 points.
o Check that the welding has been carried out correctly. It should not form a ball, but
should lie against the length of the component snubbers.
n Repeat the operation for the anti-siphon opto (4 welding points).
n Stick the bottom side of the ribbon cable to the double-sided board.

Anti-siphon opto
w
Backpressure switch

Lower side
(connection specific to Pilot C)
stuck to the double-sided board on the plunger support

n Impregnate with weak loctite and screw the 6 hexagon socket screws (ref. 11), which hold
the spring leaf/backpressure micro support in position.
n Position the opto and its retaining ring (ref. 10).
n Mount the plunger cover onto the plunger support. Impregnate the screw with weak loctite
(rep. 9).
n Position and screw the cam/anti-siphon arm unit using the Phillips head screw and the
washer (ref. 9).
n Screw the 2 Phillips head screws (ref. 8 and 7) which link the plunger support to the
plunger cover.
n Position the plunger cover clip (ref. 6).

Operation sheets

06.0_0051a_en.fm 89
n Cover the plunger with the protective film, mount the disengagement lever, its spring and
the adjustment button. Screw the plunger end shield back on (ref. 5, 4 and 3).
n Position the plunger cover and screw it back on (ref. 2).
Carry out the same procedures in reverse to remount the unit.

When fitting the upper case, ensure the joint is perfectly positioned in its slot after
reassembly.

Recalibrate the pressure sensor (see "Calibrations") then carry out the regular servicing tests
(see "Regular servicing sheet").
Operation sheets

90 06.0_0051a_en.fm
Operation sheets

N°6, Procedure: Plunger advance control


potentiometer

Safety:
For safety reasons, the technician should not carry out any maintenance when the device is
connected to the 230 V mains supply voltage.
Disconnect the mains power supply cable.

Equipment required:
n 1 Posidriv Z1 screwdriver.
n 1 flat screwdriver.
n 1 flat key (12).
n 1 potentiometer dismounting tool ref. T300869.

Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).

Procedure:

Access
n Turn the Pilot over onto the upper case.
n Unscrew the 6 Phillips head screws (ref. 1) located at the bottom of the lower housing
case, which link this to the upper case.
n Keep the Pilot assembled and turn it over onto the lower case.
n Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.
n Disconnect the CPU board flat jumper.

Hands must not come into contact with the CPU boards.

1 1
1

1
Operation sheets

06.0_0060a_en.fm 89
Dismounting
n Press the syringe pump disengagement lever towards the back of the Pilot as far as
possible.
n Maintain this position and slide the whole unit to the back left.
n Remove the plunger guide.
n Disconnect the black connectors (ref. 2).
n Unscrew the flat jumper mounting lug (ref. 3) without removing it.
n Remove the mechanical plunger unit from its slot.
n Disconnect the blue connector from the flat jumper (ref. 4).
n Remove the mechanical plunger unit.
n Unscrew the potentiometer lock nut (ref. 5).
n Remove the tooth lockwasher and the potentiometer.

Reassembly
n Dismount the end shield from the reducer frame.
n Mount the potentiometer onto the end shield (screw the nut onto a thread).
n Position the end shield in equipment T 300 869 and lock it into position using the knurled
screw.
n Position the potentiometer in the device and bring it up against the end shield.
n Tighten the potentiometer.
n Extract the end shield from device T 300 869.
n Mount the pinion on the potentiometer (match up the indexing plane).
o The larger diameter of the pinion must be flattened against the potentiometer.
n With the potentiometer facing you, turn the pinion anticlockwise until it blocks, then turn it
¼ of a turn in the opposite direction.
n Mount the moving mechanical assembly onto the reducer frame.
n Slide the end shield onto the guides and rack.
o Check the position of the input bearing which must be on the plunger side.
o Ensure the flex circuit is not damaged when mounting (folding).
n Secure the end shield using the three M3x3 TC screws.
n Secure the input bearing using the two M3x3 TC screws.
Operation sheets

When fitting the upper case, ensure the joint is perfectly positioned in its slot after
reassembly.

Recalibrate the position sensor (see "Calibrations") then carry out the regular servicing tests
(see "Regular servicing sheet").

90 06.0_0060a_en.fm
Operation sheets

N°7, Procedure: Plunger cover and/or disengagement


lever + anti-siphon arm

Safety:
For safety reasons, the technician should not carry out any maintenance when the device is
connected to the 230 V mains supply voltage.
Disconnect the mains power supply cable.

Equipment required:
n 1 Posidriv Z1 screwdriver.

Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).

Procedure:

Access
n Press the syringe pump disengagement lever towards the back of the Pilot as far as
possible.
n Maintain this position and slide the whole unit to the back right.

Dismounting
n Unscrew the 3 Phillips head screws (ref. 1) located at the bottom of the plunger end shield
which links this to the plunger cover. Remove the plunger end shield (ref. 2).
n Remove the backpressure adjustment button (ref. 3), then the disengagement lever and its
spring (ref. 4) as well as the protective plunger film (ref. 5).
n Unscrew and remove the Phillips head screw and the washer (ref. 6) which attach the anti-
siphon cam to the anti-siphon arm.
n Remove the cam/anti-siphon arm assembly (ref. 7).

4
6-7

3
5 2
1
Operation sheets

1 1

06.0_0071a_en.fm 93
Reassembly
Carry out the same procedures in reverse to remount the unit.
n Impregnate the screw with weak loctite (ref. 6) before mounting the cam and arm
assembly.

When fitting the upper case, ensure the joint is perfectly positioned in its slot after
reassembly.

Perform the regular servicing tests (see “Regular servicing sheet”).


Operation sheets

94 06.0_0071a_en.fm
Intervention sheets

N°8, Procedure: Power supply board

Safety:
For safety reasons, the technician should not carry out any maintenance when the device is
connected to the 230 V mains supply voltage.
Disconnect the mains power supply cable.

Tools required:
n 1 Posidriv Z1 screwdriver.
n 1 small flat screwdriver.
n 1 antistatic wriststrap.

Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).

Procedure:

Access
n Rotate the Pilot on the upper case.
n Unscrew the 6 Phillips head screws (ref. 1) located at the bottom of the lower case, which
tighten the latter to the upper case.
n Keep the Pilot assembled and turn it over onto the lower case.
n Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.
n Disconnect the CPU board flat jumper.

Hands must not come into contact with the CPU boards.

1 1

1
1

1
Intervention sheets

06.0_0082a_en.fm 81
Dismounting
n Press the syringe pump disengagement lever towards the back of the Pilot as far as
possible.
n Maintain this position and slide the whole unit to the back left.
n Remove the syringe pump guide.
n Disconnect the black connectors.

When electronic components are handled, it is recommended to wear an antistatic


wriststrap linked to ground and to work on an antistatic foam mat.

n Unscrew the flat jumper mounting lug (ref 2) without removing it.
n Remove the mechanical plunger unit from its slot.
n Disconnect the blue connector from the flat jumper (ref. 3).
n Remove the mechanical plunger unit.
n Disconnect the connectors (ref 4, ref 5, ref. 6, ref. 7, ref. 8)
n Unscrew the 3 Phillips head screws (ref. 9) located on the power supply board, which
tighten the latter to the lower case.
n Unscrew the 2 Phillips head screws (ref. 10) located on lower case, which tighten the latter
to the power supply board.
n Remove the power supply board.
9
8
3
9

6
5
4
7

10

Reassembly
Carry out the same procedures in reverse to remount the unit.

When fitting the upper case, ensure the joint is perfectly positioned in its slot after
Intervention sheets

reassembly.

Recalibrate the force sensor (see "Calibrations") then carry out the regular servicing tests
(see "Regular servicing sheet").

82 06.0_0082a_en.fm
Operation sheets

N°9, Procedure: Battery holder and battery

Safety:
For safety reasons, the technician should not carry out any maintenance when the device is
connected to the 230 V mains supply voltage.
Disconnect the mains power supply cable.

Equipment required:
n 1 Posidriv Z1 screwdriver.

Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).

Procedure:

Access
n Turn the Pilot over onto the upper case.

Dismounting
n Unscrew the 2 Phillips head screws (ref. 1) to be found in the lower case battery holder kit,
which link this to the lower case.
n Swivel the battery holder and remove the battery.
n Disconnect this.

Reassembly
Carry out the same procedures in reverse to remount the unit.
Operation sheets

It is recommended to perform a full loading and unloading cycle to ensure the battery is in
working condition.

Perform the regular servicing tests (see "Regular servicing sheet").

06.0_0090a_en.fm 95
Operation sheets

96 06.0_0090a_en.fm
Intervention sheets

N°10, Procedure: Rear plug support

Safety:
For safety reasons, the technician should not carry out any maintenance when the device is
connected to the 230 V mains supply voltage.
Disconnect the mains power supply cable.

Tools required:
n 1 Posidriv Z1 screwdriver.

Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).

Procedure:

Access
n Rotate the Pilot on the upper case.
n Unscrew the 6 Phillips head screws (ref. 1) located at the bottom of the lower case, which
tighten the latter to the upper case.
n Keep the Pilot assembled and turn it over onto the lower case.
n Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.
n Disconnect the CPU board flat jumper.

Hands must not come into contact with the CPU boards.

1 1

1
1

1
Intervention sheets

06.0_00103a_en.fm 85
Dismounting
n Disconnect the two black connectors (ref 2).
n Unscrew the 2 Phillips head screws (ref. 3) located on the cabled plug support, which link
this to the lower case.
n Remove the cabled plug support.
3

Reassembly
Carry out the same procedures in reverse to remount the unit.

If the number of rear door connector points is different to the number of power
supply board points, make the connections as described below.

Perform the regular servicing tests (see “Regular servicing sheet”).


Intervention sheets

86 06.0_00103a_en.fm
Operation sheets

N°11, Procedure: Ribbon cable winding kit

Safety:
For safety reasons, the technician should not carry out any maintenance when the device is
connected to the 230 V mains supply voltage.
Disconnect the mains power supply cable.

Equipment required:
n 1 Posidriv Z1 screwdriver.

Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).

Procedure:

Access
n Turn the Pilot over onto the upper case.
n Unscrew the 6 Phillips head screws (ref. 1) located at the bottom of the lower case, which
link this to the upper case.
n Keep the Pilot assembled and turn it over onto the lower case.
n Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.

Hands must not come into contact with the CPU boards.

1 1
1

1
Operation sheets

06.0_00110a_en.fm 99
Dismounting
n Press the syringe pump disengagement lever towards the back of the Pilot as far as
possible.
n Maintain this position and slide the whole unit to the back left.
n Remove the plunger guide.
n Disconnect the black connectors (ref. 2 and ref. 3).

When electronic components are handled, it is recommended to wear an antistatic


wriststrap linked to ground and to work on an antistatic foam mat.

n Unscrew the flat jumper tongue (ref. 4) without removing it.


n Remove the mechanical plunger unit from its slot.
n Disconnect the blue connector (ref. 5) from the flat jumper.
n Remove the mechanical plunger unit.
n Fully unscrew and remove the guiding tongue (ref. 4).

4
5

2&3

Remounting
Fitting the flexible tongue:
n Remove the clip (ref. 6).
n Stick the flexible tongue (ref. 7) with the adhesive facing the ribbon cable.
n Reposition the clip making sure the ribbon cable and the flexible tongue are centred on the
mechanical unit sprocket.
n Lift both ends of the flexible tongue slightly.

7
Operation sheets

100 06.0_00110a_en.fm
Fitting the guiding flange:

There are several types of flange tongues. If you have an older version, replace it with that
provided with the kit.

n Position the new guiding flange (ref. 4) using the two screws.

additional screw

n Disconnect the blue connector (ref. 5) from the flat jumper.


n Position the whole mechanical plunger unit making sure the ribbon cable is centred under
the flange.
n Perform a few round cycles to check the ribbon cable winds up correctly.
n Connect the black connectors (ref. 2 and ref. 3).
n Replace the plunger guide.
Carry out the same procedures in reverse to remount the upper case.

When fitting the upper case, ensure the joint is perfectly positioned in its slot after
reassembly.

Perform the regular servicing tests (see "Regular servicing sheet").

Operation sheets

06.0_00110a_en.fm 101
Operation sheets

102 06.0_00110a_en.fm
Operation sheets

N°13, Procedure: Centering ring kit

Safety:
For safety reasons, the technician should not carry out any maintenance when the device is
connected to the 230 V mains supply voltage.
Disconnect the mains power supply cable.

Equipment required:
n 1 Posidriv Z1 screwdriver.
n 1 cutter.
n GEB type 100% silicone grease.

Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).

Procedure:

Access
n Turn the Pilot over onto the upper case.
n Unscrew the 6 Phillips head screws (ref. 1) located at the bottom of the lower case, which
link this to the upper case.
n Keep the Pilot assembled and turn it over onto the lower case.
n Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.

Hands must not come into contact with the CPU boards.

1 1
1

1
Operation sheets

06.0_00130a_en.fm 107
n Press the syringe pump disengagement lever towards the back of the Pilot as far as
possible.
n Maintain this position and slide the whole unit to the back left.
n Remove the plunger guide.
n Disconnect the black connectors (ref. 2).
n Unscrew the flat jumper mounting lug (ref. 3) without removing it.
n Remove the mechanical plunger unit from its slot.
n Disconnect the blue connector from the flat jumper (ref. 4).
n Remove the mechanical plunger unit.
4

n Press the syringe pump disengagement lever towards the back of the Pilot as far as
possible.
n Maintain this position and slide the whole unit to the back right.

Dismounting
n Unscrew the 2 slotted head screws (ref. 5) which link the centering ring to the mechanical
end shield.
n Using the cutter, cut the input bearing and remove it from the plunger tube.

5
Operation sheets

108 06.0_00130a_en.fm
Remounting
n Mount the new slotted input bearing onto the plunger tube by twisting it.
n Laying it flat on a table, cut the O-ring using a cutter.
n Place the O-ring around the tube.
n Fit the O-ring into the centering ring slot (horizontal cut, so as to avoid lining it up with the
centering ring cut).
n Mount the stainless steel slotted plate onto the pin, by twisting it.
n Place the silicone grease cord onto the plate around the passage of the pin.
n Place the plate on the input bearing.
n Position the whole assembly against the mechanical end shield and screw it on using the
two slotted screws (ref. 5).
n Check the plunger slides correctly.

It should slide uniformly across the entire centering ring and slight friction is due to the O-
ring scraping against the tube.

Perform the regular servicing tests (see "Regular servicing sheet").

Operation sheets

06.0_00130a_en.fm 109
Operation sheets

110 06.0_00130a_en.fm
Operation sheets

N°14, Procedure: Flex circuit and tube kit

Safety:
For safety reasons, the technician should not carry out any maintenance when the device is
connected to the 230 V mains supply voltage.
Disconnect the mains power supply cable.

Equipment required:
n 1 Posidriv Z1 screwdriver.
n 1 antistatic wriststrap.

Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).

Procedure:

Access
n Turn the Pilot over onto the upper case.
n Unscrew the 6 Phillips head screws (ref. 1) located at the bottom of the lower case, which
link this to the upper case.
n Keep the Pilot assembled and turn it over onto the lower case.
n Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.
n Disconnect the CPU board flat jumper.

Hands must not come into contact with the CPU boards.

1 1
1

1
Operation sheets

06.0_00141a_en.fm 113
Dismounting
n Press the syringe pump disengagement lever towards the back of the Pilot as far as
possible.
n Maintain this position and slide the whole unit to the back left.
n Remove the plunger guide.
n Disconnect the black connectors (ref. 2).
n Unscrew the flat jumper mounting lug (ref. 3) without removing it.
n Remove the mechanical plunger unit from its slot.
n Disconnect the blue connector from the flat jumper (ref. 4).
n Remove the mechanical plunger unit.
4

n Press the syringe pump disengagement lever towards the back of the Pilot as far as
possible.
n Maintain this position and slide the whole unit to the back right.

Welding and unwelding operations are carried out using a soldering iron fitted with a tip in
good condition, constantly tapered and kept clean. The temperature of the iron should be
between 315°C and 340°C.

n Unweld the flex circuit (ref. 5) ensuring the disengagement switch is not damaged in the
process.
n Unscrew the 2 slotted head screws (ref. 6) which link the centering ring to the mechanical
end shield.
n Unscrew the 2 Phillips head screws (ref. 7) which link the centering ring to the mechanical
end shield.
n Remove the flex circuit clip (ref. 8).

5 7
Operation sheets

6
8

114 06.0_00141a_en.fm
n Remove the plunger guide.
n Unscrew the 3 Phillips head screws (ref. 9) located at the bottom of the plunger end shield
which links this to the plunger cover.
n Remove the plunger end shield.

9
9

n Remove the backpressure adjustment button (ref. 10).


n Remove the disengagement lever and its spring (ref. 11) as well as the protective plunger
film (ref. 12).
n Remove the plunger cover clip (ref.13).

12
11

10

13
Operation sheets

06.0_00141a_en.fm 115
n Unscrew the 2 Phillips head screws (ref. 14 and 15) which link the plunger support to the
plunger cover.
n Unscrew and remove the Phillips head screw and the washer (ref. 16) which attach the
anti-siphon cam to the anti-siphon arm.
n Remove the cam/arm unit as well as the plunger cover.
14

15 16

n Remove the retaining ring that holds the anti-siphon opto in place (ref. 17).
n Unscrew the 6 hexagon socket screws (ref. 18).

18
17

17
Operation sheets

116 06.0_00141a_en.fm
n Remove the backpressure micro support/spring leaf (ref. 19).
n Pull off the ribbon cable from the double-sided board

19

ribbon cable stuck


to the double-sided
board

Welding and unwelding operations are carried out using a soldering iron fitted with a tip in
good condition, constantly tapered and kept clean. The temperature of the iron should be
between 315°C and 340°C.

n Unweld the sensor wires:

When dismounting the force sensor, ensure the welding blocks are not damaged and the
square is not removed in the process.

o Add more weld to facilitate the unwelding process.


o Heat and pull on the wires one by one.

n Remove the flex circuit and tube kit from the plunger support.

Operation sheets

06.0_00141a_en.fm 117
Reassembly
n Remove the sticky tape holding the flex circuit to the tube.
n Mount the new flex circuit onto the plunger support.
n Weld the ribbon cable to the new pressure switch, ensuring the mounting procedure is
carried out in order:
o Taper the 3 welding points.
o Weld the 3 points.
o Check that the welding has been carried out correctly. It should not form a ball, but
should lie against the length of the component snubbers.

n Repeat the operation for the anti-siphon opto (4 welding points).


n Stick the bottom side of the ribbon cable to the double-sided board.
n Repeat the operation for the disengagement switch. (3 welding points)
Tube D. 12

ube

Switch
w disengagement
Switch
backpressure
Anti-siphon opto
Lower side
(connection specific to Pilot C)
stuck to the double-sided board on the plunger support

Carry out the same procedures in reverse to remount the unit.

Do not forget to impregnate the cam/arm clamping screw (rep. 16) with weak loctite.

When fitting the upper case, ensure the joint is perfectly positioned in its slot after
reassembly.

Recalibrate the force sensor (see "Calibrations") then carry out the regular servicing tests
(see "Regular servicing sheet").
Operation sheets

118 06.0_00141a_en.fm
Operation sheets

N°15, Procedure: Upper and lower cases

Safety:
For safety reasons, the technician should not carry out any maintenance when the device is
connected to the 230 V mains supply voltage.
Disconnect the mains power supply cable.

Equipment required:
n 1 Posidriv Z1 screwdriver.
n 1 antistatic wriststrap.
n 1 soldering iron.
n "RADIEL Sn60Pb RI 1" welding wire (cleaning not required for rewelding) or equivalent.

Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).

Procedure:

Access
n Turn the Pilot over onto the upper case.
n Unscrew the 6 Phillips head screws (ref. 1) located at the bottom of the lower case, which
link this to the upper case.
n Keep the Pilot assembled and turn it over onto the lower case.
n Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.
n Disconnect the CPU board flat jumper.

Hands must not come into contact with the CPU boards.

1 1
1

1
Operation sheets

06.0_00150a_en.fm 119
Dismounting the upper cover
n Dismount the central unit and display boards (see corresponding sheet).
n Dismount the syringe clamp (see corresponding sheet).
n Dismount the syringe detection system (see corresponding sheet).

Remounting the upper cover


n Remount the syringe detection system (see corresponding sheet).
n Remount the syringe clamp (see corresponding sheet).
n Remount the central unit and display boards (see corresponding sheet).
o If the original lower case is fitted with a removable buzzer bell, you can use your “CPU
and display board” kit without having to adapt it.
o If the original lower case has a cylindrical guide for accommodating the buzzer, the buz-
zer from the “CPU and display board” kit must be replaced with the buzzer supplied with
the spare parts as the latter is fitted with snubbers and a separate foam joint.
Follow the mounting diagram shown below to carry out this adaptation procedure.

Welding and unwelding operations are carried out using a soldering iron fitted with a tip in
good condition, constantly tapered and kept clean. The temperature of the iron should be
between 315°C and 340°C.
o Unweld the snubbers from the buzzer:
s add more weld to facilitate the unwelding process.
s Heat and pull on the buzzer pins one by one.

Clean the buzzer slot surfaces using a cloth soaked in rubbing alcohol.

o Weld the snubbers of the new buzzer:

+
Buzzer with snubbers

Display board
s taper the 2 buzzer pins to be welded.
s place the buzzer onto the board following the
mounting diagram.
s weld to buzzer onto the board.
s check that the welding has been carried out correctly.
It should not form a ball, but should lie against the
length of the wire.
Operation sheets

Continue with the remounting procedure, referring to the “central unit and display board”
sheet.

Perform the regular servicing tests (see "Regular servicing sheet").

120 06.0_00150a_en.fm
Dismounting the lower cover
n Dismount the power supply board (see corresponding sheet).
n Dismount the battery holder (see corresponding sheet).
n Dismount the rear plug support (see corresponding sheet).

Remounting the lower cover


Carry out the remounting procedure in reverse.

Perform the regular servicing tests (see "Regular servicing sheet").

Operation sheets

06.0_00150a_en.fm 121
Operation sheets

122 06.0_00150a_en.fm
7 Calibration
7.1 Calibration procedure

The calibration menu is for trained personnel only.

7.1.1 Calibration access

Keyboard description.
Key Function
ON, is used to switch the machine on.
OFF, is used to switch the machine off when pressed for over three
seconds.
SILENCE ALARME, is used to access the calibration mode.

VALIDATION, is used to validate a choice.

BOLUS, is used to access the calibration mode.

The select keys allow to scroll the figures and letters on the LCD
screen, on the units, tens, tens segments etc.

Activate calibration
n Press the "SILENCE ALARME" and
"BOLUS" keys simultaneously.

n Maintain this position while pressing "ON".

n When EtA. is displayed, release the “SILENCE ALARME”


and “PURGE/BOLUS” keys, then validate within three
seconds by pressing the “VALIDATION" key.
n The equipment initiates the EtA.4calibration by default
n Scroll the different display unit calibration states using the "+
or -" select keys.

o EtA.4: calibration of the 3 battery voltage levels.


o EtA.6: calibration of the position sensor.
Calibration

07.1_005a_en.fm 105
7.1.2 EtA.4 Calibration of the 3 battery voltage levels.
This menu enables n EtA.4, press "VALIDATION".
the user to store o Bat.1: Supply the equipment with 6.3 V+ 0.05 using a
the three Bat1,
stabilised power supply.
Bat2 and Bat3
s Press "VALIDATION".
battery voltage
The voltage is read and stored in the EEPROM.
values in an
o Press "VALIDATION".
EEPROM.
o Bat.2: Supply the equipment with 5.9 V+ 0.05 using a
stabilised power supply.
s Press "VALIDATION".
The voltage is read and stored in the EEPROM.
o Press "VALIDATION".
o Bat.3: Supply the equipment with 5.7 V+ 0.05 using a
stabilised power supply.
s Press "VALIDATION".
The voltage is read and stored in the EEPROM.

o By validating once again, it is possible to select another


calibration.

7.1.3 EtA.6 Calibration of the position sensor.


This menu enables n EtA.6, press "VALIDATION".
users to store both o Hig.h: Place a spacer measuring 115 mm + 0.05 ref.
high and low
T 300940, into the fin groove.
displacement limit
s Position the plunger in contact with the spacer.
values in the
s Keep the plunger disengaged and press
EEPROM.
“VALIDATION”.
The position value is read and stored in the EEPROM.
o Press "VALIDATION".
o Lou: Place a spacer measuring 20 mm + 0.05 ref. T
300775, into the fin groove.
s Position the plunger in contact with the spacer.
s Keep the plunger disengaged and press
“VALIDATION”.
The position value is read and stored in the EEPROM.

Once both high and low values have been stored in the EEPROM, the equipment
indicates the number of LSB in decimals between the two measurement points.
This value should be 772 + 10 LSB.

o If the value is out of limits, carry out the calibration


procedure once again.
o By validating once again, it is possible to select another
calibration.
Calibration

106 07.1_005a_en.fm
8 Spare parts catalogue
8.1 Upper case

99

100
101

.....
.....
.....

.....
.....
..... .....
.....
.....
..... .....
.....

.....

..... ..... : see next page

Mark Qty Reference Name


1 167246 LCD transparent pilot window
1 167629 6 diam injected bumper
2 199560 Female M3x12 hybrid spacer
1 167067 PCB 500 protective film
1 167296 Adhesive flat jumper protective film
1 167632 Buzzer foam
Spare parts catalogue

1 167636 Buzzer foam bell


99 1 167744 17.5 diam Pilot label
100 1 167476 Injected syringe clamp
101 1 199214 Upper case kit DIN version
1 199218 Upper case kit NL version
1 199222 Upper case kit IS version

08.1_007a_en.fm 107
.....

.....
.....

122
111
112

104
125
126 114
127
123
128 178
103

124

102 ..... : see previous page

Mark Qty Reference Name


102 1 167361 5 diam retaining ring
103 1 167310 Syringe clamp compression spring
104 1 167452 Injected PCB support (opto)
111 1 167670 Syringe list label
112 1 167640 A IEC pilot front panel
114 1 167550 Display board
1 167713 Keyboard (active part)
122 1 167458 Syringe clamp shaft
123 1 167932 C.I. opto cabled
Spare parts catalogue

124 1 167459 Pilot 20/60 ml obturator


178 1 167190 CPU board

108 08.1_007a_en.fm
Spare parts catalogue

8.2 Lower case


109

115
116

108
117

118

107
119

120
121

Mark Qty Reference Name


1 167059 Pilot guide rail
1 167299 Buzzer adjustment washer
1 167297 Spring washer
1 167923 HE13 battery connector
1 167122 Injected flexible PCB flange
167370 Flexible tongue
1 167355 Mains fibre gasket
1 170416 Mains cord
107 1 167093 Buzzer bell
108 1 167967 Pilot A wired plug support
109 1 177201 Transformer
1 167432 Buzzer adjustment button
115 1 170228 Fuse F2
116 1 167078 Pilot A power supply board
117 1 167942 Wired mains socket
Spare parts catalogue

118 1 199201 Lower case kit DIN version


1 199202 Lower case kit NL version
1 199203 Lower case kit IS version
119 6 167249 Black stop piece
120 1 174019 6V 1.3 Ah battery
121 1 199169 Pilot battery holder kit

08.2_007a_en.fm 109
Spare parts catalogue

110
08.2_007a_en.fm
Spare parts catalogue

8.3 Plunger unit

.....
..... .....
.....
..... .....
200
201 ..... ..... .....
202 ..... .....
224
225
203
204
205
206 .....
207 ..... ..... .....

208 220
221 223 .....
209 .....
..... .....
222 .....
.....

..... .....
.....
210

..... : see next page

Mark Qty Reference Name


1 167264 Anti-siphon spring
200/207 2 167360 3.5mm diam retaining ring
201 1 167465 Disengagement spring follower
202 1 167469 Disengagement follower
203 1 167460 Injected disengagement cam
204 1 167471 Injected upper mechanism nut
205 1 167464 Disengagement spring shaft
206 1 167472 Injected lower mechanism nut
208 1 167466 Mechanical flange
209 1 167281 Injected slug + mechanical block
210 1 167275 Flexible PCB clip
220 1 167271 Flex circuit
221 1 167317 Full disengagement shaft
222 1 167240 12 diam tube (20/60 ml version)
Spare parts catalogue

223 1 167403 Slotted injected input bearing


1 177203 O-ring (to put in the bearing)
1 177204 Stainless steel plate (to be mounted between the bearing and the end
shield)
224 1 167291 Pilot anti-siphon arm
225 1 199252 Pilot A plunger kit

08.3_007a_en.fm 111
249
229 260
261
..... 253 250
..... 230 254 251
..... 252
..... 231
.....
.....
.....
.....
..... 270
..... 232 271 272
..... .....
..... ..... 255
..... ..... 256
..... 257
258
259
226
..... 227
.....

273
..... : see next page
Mark Qty Reference Name
226 1 167259 Plunger cover clip
227 1 167442 Injected M 0.5 rack
229 1 167237 Bonding pad
230 1 167411 Backpressure spring leaf
231 1 167273 Backpressure micro support (plunger)
232 1 173408 OMRON micro-switch
249 1 162311 Photo switch (type RP I 131)
250 1 167492 Backpressure spring
251 1 167270 Backpressure spring leaf adjusting screw
252 1 167269 Backpressure adjustment button
253 1 167282 Plunger bracket
254 1 167493 Protective plunger film
256 1 167488 Disengagement finger
257 1 167487 Sintered disengagement shaft bearing
258 2 167298 Injected flexible PCB protector
259 1 167272 Injected clamping collar
260 1 190714 Retaining ring
Spare parts catalogue

261 1 167385 Anti-siphon cam


270 1 167260 Disengagement lever
271 1 167245 Disengagement lever spring
272 1 167055 PILOT A plunger end shield
273 1 199103 Flexible circuit + tube kit

112 08.3_007a_en.fm
298

Mark Qty Reference Name


298 1 199135 Mechanical kit

Spare parts catalogue

08.3_007a_en.fm 113
Spare parts catalogue

114
08.3_007a_en.fm
Spare parts catalogue

8.4 Mechanical gear box

315
314
312
311 313
308

309
310

296

300
301
302
303

304

Mark Qty Reference Name


2 190714 Retaining ring
296 1 167930 Wired potentiometer connector
300 1 167128 Motor rotation switch
301 1 167145 Standard motor opto support
302 1 167111 Motor rotation disk
304 1 167156 Reducer
303 to 1 167948 Motor reducer Pilot A wired blue connector
304
303 to 1 167964 Motor reducer Pilot A wired black connector
304
308 1 167117 Mechanical end shield
309 1 167140 Intermediate end shield
310 1 167158 Standard reducer end shield
Spare parts catalogue

311 1 167142 T M6 2 x 100 threaded rod


312/313 3 167143 Ø 6 guide pin
314 1 167144 Pinion (64 teeth)
315 1 167157 Spacer

08.4_007a_en.fm 115
Spare parts catalogue

116
08.4_007a_en.fm
Spare parts catalogue

8.5 Labels

Mark Qty Reference Name


1 167970 Instruction guide label DIN
1 167770 Instruction guide label IS/BS
1 167827 Instruction guide label NL
1 167773 Instruction guide label IT
1 167855 Instruction guide label SP
1 167890 Label DIN, IS/BS, NL, IT, SP
1 167742 Buzzer adjustement label DIN, IS/BS, NL, IT, SP
1 167979 Danger selection label DIN
1 167776 Danger selection label IS/BS
1 167811 Danger selection label NL
1 167778 Danger selection label IT
1 167858 Danger selection label SP
1 167734 Battery door label DIN, IS/BS
1 167812 Battery door label NL, IT
1 167734 Battery door label SP
1 167731 0.63 m AT label (inside) DIN, IS/BS, NL, IT, SP
1 167978 Danger mains label DIN
1 167725 Danger mains label IS/BS
1 167814 Danger mains label NL
1 167800 Danger mains label IT
1 167725 Danger mains label SP
1 167648 Keyboard (front panel) DIN
1 167644 Keyboard (front panel) IS
1 167658 Keyboard (front panel) NL
1 167652 Keyboard (front panel) IT
1 167655 Keyboard (front panel) SP
1 167713 Keyboard active part DIN, IS, NL, IT, SP

Spare parts catalogue

08.5_007a_en.fm 117
Spare parts catalogue

118
08.5_007a_en.fm
Useful addresses
SALES DEPARTMENT
Fresenius Vial
Le Grand Chemin, 38590 Brézins
CUSTOMER SERVICE Tel.: 00 33 (0) 4 76 67 10 81
or 00 33 (0) 4 76 67 10 54
or 00 33 (0) 4 76 67 11 08
Fax: 00 33 (0) 4 76 67 11 34

SALES MANAGEMENT Tel.: 00 33 (0) 4 76 67 10 81


Fax: 00 33 (0) 4 76 67 11 34

TRAINING DEPARTMENT
Fresenius Vial Tel.: 04 76 67 10 76
Le Grand Chemin, 38590 Brézins Fax: 04 76 67 11 22

AFTER SALES SERVICE


FRANCE After Sales Service Division GRENOBLE Tel.: 04 76 67 10 34
Fresenius Vial Le Grand Chemin, 38590 Brézins Fax: 04 76 67 11 22
FRANCE

BELGIUM Fresenius Vial NV/SA Tel.: 32/388.07307


Vial Médical Division Fax: 32/388.05007
Molenberglei 7
2627 Schelle
BELGIUM

GERMANY Fresenius MCM


AM-Neuen Berg
63749 ALZENAU
GERMANY

This document may contain errors or typing mistakes.


Modifications may be made in future revisions.
COPYRIGHT © 1998, Fresenius Vial S.A.

This document may not be reproduced in part or in whole without the written permission of
Fresenius Vial S.A.

Fresenius Vial S.A. Headquarters: Le Grand Chemin - 38590 Brézins (FRANCE)


Public Company with board of directors and advisory council, and with a capital of
90,128,000FF
SIREN Grenoble B 408 720 282.
Design and development: SONOVISION-ITEP

Technical Manual Pilot CE, A: ed. A-09.2001

You might also like