Pilot A Ce en
Pilot A Ce en
Pilot A Ce en
MANUAL
TECHNICAL MANUAL
PILOT A, CE
PiloteCE_A_enTDM.fm 1
4.2.2 Use within the framework of the departmental order ............................................... 33
4.3 Checks .................................................................................................. 35
4.3.1 Test access .............................................................................................................. 35
4.3.2 Visual check ............................................................................................................. 35
4.3.3 Running time and servicing inspection date............................................................. 36
4.3.4 Indicator lights check................................................................................................ 36
4.3.5 Keyboard check ....................................................................................................... 37
4.3.6 Checking the battery voltage.................................................................................... 38
4.3.7 Testing the last 10 alarms ........................................................................................ 38
4.3.8 Total running time check .......................................................................................... 39
4.3.9 TTL serial link test .................................................................................................... 39
4.3.10 RS 232 serial link check ........................................................................................... 39
4.3.11 Checking the software version. ................................................................................ 40
4.3.12 Checking the ADC.................................................................................................... 40
4.3.13 Checking the position sensor ................................................................................... 40
4.3.14 Buzzer test ............................................................................................................... 41
4.3.15 Displaying the calibration values .............................................................................. 41
4.3.16 Checking the syringe clamp ..................................................................................... 41
4.3.17 Checking the syringe group number ........................................................................ 42
4.3.18 Checking the disengagement................................................................................... 42
4.3.19 Checking the anti-siphon arm .................................................................................. 42
4.3.20 Checking backpressure............................................................................................ 42
4.3.21 Checking the end of infusion pre-alarm ................................................................... 43
4.3.22 Checking the linearity ............................................................................................... 44
4.3.23 Checking mains/battery operation............................................................................ 44
4.3.24 Battery autonomy test .............................................................................................. 45
4.3.25 Continuity test .......................................................................................................... 45
4.3.26 Regular inspection sheet.......................................................................................... 47
4.4 Flow rate control .................................................................................. 49
4.4.1 Measurement with a computer ................................................................................. 49
4.4.2 Measurement with scales......................................................................................... 51
4.4.3 Measurement using a test tube................................................................................ 53
4.5 Cleaning and disinfection ................................................................... 55
4.6 Storage.................................................................................................. 56
5 Diagnosis......................................................................................................... 57
5.1 Troubleshooting................................................................................... 57
5.2 Error messages.................................................................................... 59
6 Operation sheets............................................................................................. 61
N°1, Procedure: Display and central unit boards .................................... 63
N°2, Procedure: Syringe clamp ................................................................. 65
N°3, Procedure: Syringe detection system .............................................. 67
N°4, Procedure: Motor + Opto + Disk........................................................ 69
N°5, Procedure: Dynamometer sensor ..................................................... 71
N°6, Procedure: Plunger advance control potentiometer....................... 77
22 PiloteCE_A_enTDM.fm
N°7, Procedure: Plunger cover and/or disengagement lever
+ anti-siphon arm ........................................................................................ 79
N°8, Procedure: Power supply board........................................................ 81
N°9, Procedure: Battery holder and battery ............................................. 83
N°10, Procedure: Rear plug support ......................................................... 85
N°11, Procedure: Ribbon cable winding kit.............................................. 87
N°12, Procedure: Centering ring kit .......................................................... 91
N°13, Procedure: Flex circuit and tube kit ................................................ 95
N°14, Procedure: Upper and lower cases ............................................... 101
7 Calibration..................................................................................................... 105
7.1 Calibration procedure ........................................................................ 105
7.1.1 Calibration access .................................................................................................. 105
7.1.2 EtA.4 Calibration of the 3 battery voltage levels................................................ 106
7.1.3 EtA.6 Calibration of the position sensor. ........................................................... 106
PiloteCE_A_enTDM.fm 3
44 PiloteCE_A_enTDM.fm
1 Overview
1.1 General
The Pilot A is a syringe pump intended for the infusion of intravenous agents at a accurate,
low flow rate. The wide choice of syringes, the use of the universally recognised control
symbols and the synoptic display of the alarms contribute to making the Pilot A easy to use.
The adjustable occlusion detection, the correct positioning detection and the overall syringe
protection system guarantee optimum safety.
Its technical characteristics, the flow range from 0.1 to 120 ml/hr and its excellent accuracy
(+1% on the equipment) contribute to making the Pilot A, the ideal instrument for medical
services.
Overview
01_007a_en.fm 5
1.2 Overview diagram
Ext
12/15 V
15 W
Step by
DC-DC Motor
230 V Power supply ON / OFF step
converter driver motor
Battery
EPROM
EPROM Motor speed
128K x 8 sensor
RAM
RAM
8K x8 Syringue
Motor body
speed
sensor
EEPROM
EEPROM
512
Syringue
Antisiphonbody
sensor
BUS Antisiphon
sensor
SPI
CPU
CPU
CPU
Watch
dog Displacement
transducer
Interface Opto
UART
bus interface
Switchcall
Nurse
occlusion
(option)
ADC
Switch
Displacement
occlusion
transducer
Disengagement
switch
Nurse call
Disengagement
(option)
switch
LCD LCD
RS 232 driver display
Keyboard Disengagement
switch
LED LED
Master
Buzzer driver display
Overview
6 01_007a_en.fm
1.3 Precautions to be taken before use
The symbol ! in the Quick Start Guide recommends that the manual should be read in full
in accordance with standard EN 60601-1.
Fresenius Vial may in no case be held responsible for medical problems or any other
problems resulting from inadequate use of the equipment.
Refer to the User’s Instructions for further details.
1.5.1 Electrical
n Power supply: 230 V - 50-60 Hz.
n Max. consumption : 23 VAC.
n Fuse F2: 100 mAT 250 V IEC 127.
n Battery: 6 V - 1.1 Ah.
n External power supply: 12 - 15 V DC - 15 W.
1.5.2 Electronics
The Pilot syringe pump is fitted with 3 circuit boards:
n Motor power supply and control board.
n CPU board.
n Keyboard display board.
01_007a_en.fm 7
1.5.4 Conformity, standards
n In compliance with the European Directive 93/42 EEC related to Medical Equipment:
CE0459.
n In compliance with the European Directive 89/336 EEC: Electromagnetic compatibility.
n Compliant with the standards EN 60601.2 and PrEN60601-2-24.
n Protection against leakage currents: CF type.
n Protection against electric shock: Class II.
n IP34 : Protection against ingress of liquid
Overview
8 01_007a_en.fm
2 Description and operation
2.1 Physical description
Plunger disengagement
control
Syringe clamp
Upper case
02_007a_en.fm 9
2.1.1 The display board and the front panel
The display board is mounted under the front control panel and is fitted with all the items
required for man-machine interaction.
n Input keyboard.
n Control lamps and overview diagrams.
n 7-segment display units.
J1
J3
J2
Pin Description
1 SEG1 display matrix Line 1
2 SEG2 display matrix Line 2
3 SEG3 display matrix Line 3
4 SEG4 display matrix Line 4
5 SEG5 display matrix Line 5
6 SEG6 display matrix Line 6
7 SEG7 display matrix Line 7
8 SEG8 display matrix Line 8
9 COL1 display matrix Column 1
10 COL2 display matrix Column 2
11 COL3 display matrix Column 3
12 diode FAIL control Fail
13 COL/DIG 9 LED type control "
14 LIG1 keyboard interface Line 1
15 LIG2 keyboard interface Line 2
16 LIG3 keyboard interface Line 3
17 LDSECT lighting control Mains LED
Description and operation
10 02_007a_en.fm
J2 connector to keyboard
Pin Description
1 Column 1
2 Column 2
3 Column 3
4 Column 4
5 Column 5
6 Column 6
7 Line 1
8 Line 2
9 Line 3
10 TON
11 TOFF
12 Gnd power supply
Pin Description
1 TON on key
2 TOFF off key
3 SPI SI bus
4 SPI CLK bus
5 SPI CSLCD bus
6 BUZZER buzz control
7 VBAT power supply
8 Gnd power supply
02_007a_en.fm 11
2.1.2 CPU board
The CPU board holds an 80C32 microcomputer. It is mounted and connected to the display
board by connectors J4 and J5.
A ribbon cable connects this to a power supply board by means of a connector J3.
J2
J4
J3 J5
CPU board
J2 connector to fin detection switch and syringe detection opto-electronic sensor
Pin Description
1 Opto anode diode +5V
2 Common points between cathode diode and opto 1 and opto 2 transistor
emitters.
3 Opto 1 transistor collector
4 Opto 2 transistor collector
Pin Description
1 +5V controlled power supply
2 Gnd power supply
3 +VBAT power supply
4 Gnd power supply
5 Phase A motor control
6 Phase B motor control
7 Phase C motor control
8 Phase D motor control
9 I signal motor control
10 BOOST signal motor control
11 Sopt1 opto rotation motor output
12 Sopt2 opto anti-siphon
13 n.u.
Description and operation
12 02_007a_en.fm
J3 connector to power supply board
Pin Description
21 DEB/OFF disengagement signal
22 RTS request to send Line 2
23 OCC/OFF stoppage signal
24 BUZ nurse call relay control
25 EOC end of adc conversion
26 CSADC spi adc bus select
27 CLK spi adc bus clock
28 SI data in spi adc bus
29 SO data out spi adc bus
30 CDANA analogue sensor power supply control
31 Rx2 receive data ttl Line 2
32 Tx2 transmit data ttl Line 2
33 Txd1 transmit data ttl Line 1
34 Rxd1 receive data ttl Line 1
35 TON on key
36 TOFF off key
37 +VBAT power supply
38 Gnd
39 +5V
40 Gnd
J4 connector to keyboard
Pin Description
1 SEG1 display matrix Line 1
2 SEG2 display matrix Line 2
3 SEG3 display matrix Line 3
4 SEG4 display matrix Line 4
5 SEG5 display matrix Line 5
6 SEG6 display matrix Line 6
7 SEG7 display matrix Line 7
8 SEG8 display matrix Line 8
9 COL1 display and keyboard matrix Column 1
10 COL2 display and keyboard matrix Column 2
11 COL3 display and keyboard matrix Column 3
12 FAIL LED control fail
13 RSCRT display power reduction control
14 LIG1 keyboard interface Line 1
15 LIG2 keyboard interface Line 2
16 LIG3 keyboard interface Line 3
Description and operation
02_007a_en.fm 13
J5 connector to display board
Pin Description
1 TON on key
2 TOFF off key
3 SI bus spi
4 CLK bus spi
5 CSLCD bus spi
6 BUZZ buzzer control
7 VBAT power supply
8 Gnd power supply
Description and operation
14 02_007a_en.fm
2.1.3 Power supply board and battery
The power supply board is mounted on the bottom of the lower case. It supplies all electronic
parts using the mains 230 V AC or the external 12 V DC. It also provides electrical supply to
the 1.1 or 1.2 Ah battery.
J6
J7
J8
J9
J5 F2
J3
J4 J1
J2 F1
Pin Description
1 Neutral
2 Phase
Pin Description
1 +5V controlled power supply
2 Gnd power supply
3 +VBAT power supply
4 Gnd power supply
5 Phase A motor control
6 Phase B motor control
7 Phase C motor control
8 Phase D motor control
9 I signal motor control
10 BOOST signal
11 Sopt1 opto rotation output module
12 Sopt2 opto anti-siphon output module
13 N.U
14 Cdopt1 opto rotation control module
Description and operation
02_007a_en.fm 15
J2 connector to CPU board
Pin Description
22 RTS request to send
23 OCC/OFF occlusion signal active at 0
24 BUZ nurse call relay control
25 EOC end of ADC conversion
26 CSADC SPI ADC bus select
27 CLK SPI ADC bus clock
28 SI data IN SPI ADC bus
29 SO data out SPI ADC bus
30 CDANA analogue sensors power supply control
31 RX2 TTL receive data Line 2
32 TX2 TTL transmit data Line 2
33 TXD1 TTL transmit data Line 1
34 RXD1 TTL receive data Line 1
35 TOFF OFF key
36 TON ON key
37 +VBAT power supply
38 Gnd
39 +5V
40 Gnd
J3 connector to potentiometer
Pin Description
1 VREF
2 Centre point
3 Gnd
Pin Description
1 + battery
2 - battery
J5 connector to motor
Pin Description
1 +VBAT
2 +VBAT
3 Phase D
4 Phase C
5 Phase B
Description and operation
6 Phase A
7 Opto rotation anode diode /+5V
8 Opto rotation cathode diode
9 Opto rotation transistor collector
10 GND/ opto rotation transistor emitter
16 02_007a_en.fm
J6 connector to RS232 and Master plugs
Pin Description
1 TX1 transmit data Line 1
2 +5V
3 RX1 receive data Line 1
4 Gnd
5 Interface validation
6 Nurse call relay common point
7 Nurse call relay normally open
8 Nurse call relay normally closed
9 CD ON external on
10 CD OFF external off
11 I-OPTON motor output control
12 I-SECT mains LED
13 +VBAT external power supply plug
14 RX2 receive data Line 2
15 TX2 receive data Line 2
16 Gnd
17 CTS
18 RTS
Pin Description
1 ± external power
supply
2 ± external power
supply
Pin Description
1 +VREF + internal gauge bridge
2 E1 Internal gauge bridge input / occlusion ON
3 E2 Internal gauge bridge input / occlusion ON
4 Gnd – internal gauge bridge
5 Opto anti-siphon LED cathode
6 Opto anti-siphon LED anode / +5V
7 Opto anti-siphon transistor collector
8 Disengagement micro-switch ON
9 Disengagement micro-switch OFF
Description and operation
10 Gnd
Remember to dismount the ribbon cable holder on the power supply board before
extracting the mechanical assembly from the case (the ribbon cable may break).
02_007a_en.fm 17
J9 connector, test points
Pin Description
1 gnd
2 Position sensor output
3 Battery discharge control output
4 Amplified force sensor output
5 n.u.
6 Motor control optical switch output
7 Force and position sensor reference voltage
8 Piston head detection optical switch output
18 02_007a_en.fm
2.2 Functional description
From a functional point of view, the Pilot A is composed of three sub-assemblies :
n A syringe position control and maintenance sub-assembly.
n A motorisation sub-assembly.
n An external connection sub-assembly.
02_007a_en.fm 19
Description and operation
20
02_007a_en.fm
3 Description of the menus
3.1 Configuration menu of the basic operation parameters
The configuration menu enables users to adapt the Pilot to specific needs. It provides access
to the menus allowing for customisation of the parameters associated with basic operation
modes.
Fresenius Vial recommends users to implement the selected configuration procedures in the
presence of qualified personnel or an employee of the technical department.
It is possible to exit the configuration mode at any time by pressing the OFF key.
03.2_007a_en.fm 21
3.1.1 Access to menus
Keys used
Key Function
ON, is used to switch the machine on.
OFF, is used to switch the machine off and, when pressed for over three
seconds, to exit the configuration mode.
SILENCE ALARME, is used to access the configuration mode of the
current operation parameters.
The selection keys allow to scroll the figures and letters on the tens,
units, tens segments etc.
22 03.2_007a_en.fm
The modification can be cancelled by pressing the "STOP" key.
It is possible to exit the configuration mode at any time by pressing the "OFF"
key.
When mode SEL3 is selected, and if there is a choice of more than one syringe, the
Pilot automatically moves onto the configuration of the list of syringes that can be
selected PAr4 when the machine is next switched on.
It is possible to exit the configuration mode at any time by pressing the "OFF"
key.
Syringe type Min. flow rate (ml/hr) Max. flow rate (ml/hr)
50/60 cc 0,1 120
20 cc 0,1 60
Description of the menus
03.2_007a_en.fm 23
o Press "VALIDATION"
It is possible to exit the configuration mode at any time by pressing the "OFF"
key.
It is possible to exit the configuration mode at any time by pressing the "OFF"
key.
24 03.2_007a_en.fm
3.1.6 PAr5, configuration of the compulsory priming
This configuration enables users to choose whether or not draining is compulsory after the
syringe is selected.
n PAr4, press "VALIDATION"
o If PurG, compulsory draining, pressing "BOLUS"
during start-up is compulsory to be able to select the flow
rate.
o If noPu, draining is not compulsory, the flow rate may
be selected upon start-up straight after the syringe is
validated.
It is possible to exit the configuration mode at any time by pressing the "OFF"
key.
In order to control a Pilot with a PC, users need to transmit the request using the
appropriate command (see the RS232 link protocol for the Pilote computer interface).
It is possible to exit the configuration mode at any time by pressing the "OFF"
key.
03.2_007a_en.fm 25
3.1.8 PAr9, configuration of the RS232 communication speed
This configuration enables the user to select the communication speed of the RS232 link.
n PAr9, press "VALIDATION"
o If 19K2, speed at 19,200 Bauds.
It is possible to exit the configuration mode at any time by pressing the "OFF"
key.
It is possible to exit the configuration mode at any time by pressing the "OFF"
key.
Description of the menus
26 03.2_007a_en.fm
3.1.10 Syringe/name match table
03.2_007a_en.fm 27
Description of the menus
28
03.2_007a_en.fm
Configuration, calibration and inspection
03.3_002a_en.fm 35
Configuration, calibration and inspection
36
03.3_002a_en.fm
Configuration, calibration and inspection
The Service tests may also be performed more easily and more quickly using a PC
with installed maintenance software (consult our After Sales Service).
The service test menu enables users to perform a series of 15 tests or checks:
n tSt1: Displays the running time and the maintenance date.
n tSt2: Indicator lights test.
n tSt3: Keyboard test.
n tSt4: Displays the battery voltage.
n tSt5: Displays the codes of the last 10 alarms.
n tSt6: Displays the total running time.
n tSt7: TTL serial link test.
n tSt8: RS 232 serial link test.
n tStA: Displays the software version.
n tStb: Displays the analogue input.
n tStC: Displays the plunger position.
n tStd: Buzzer test.
n tStE: Displays the calibration values.
n tStf: Displays the syringe type.
n tStG: Displays the syringe group.
03.4_007a_en.fm 31
Configuration, calibration and inspection
32
03.4_007a_en.fm
4 Preventive maintenance
4.1 Recommendations
The Pilot syringe pump can only be inspected, serviced or repaired by Fresenius Vial or by
an authorised and appointed service. The qualified technicians in your establishment and our
After Sales Service should be notified of any abnormal operation of the device.
If a repair is necessary, send the instrument in its original packaging if possible with a precise
description of the observed fault, to the official dealer for Fresenius Vial.
For further information concerning troubleshooting or the usage procedure, please contact
our After Sales Service or our Sales Department.
Fresenius Vial is not liable for loss or damage to the equipment during transport to our After
Sales Service.
Frequency Name
12 months Carry out a servicing inspection.
3 years Replace the battery (see "Battery-holder and battery" operation sheet).
Frequency Name
1st inspection in Perform the first servicing inspection.
the 3rd year
Then every 2 Perform a servicing inspection.
years
3 years Replace the battery (see "Battery-holder and battery" operation sheet).
Preventive maintenance
04.1&2_001a_en.fm 41
Preventive maintenance
42
04.1&2_001a_en.fm
Preventive maintenance
4.3 Checks
In order to perform equipment follow-up by means of preventive maintenance, a regular
servicing inspection is recommended every 12 months (see “Regular servicing sheet”).
To ensure the check procedure is carried out efficiently, recharge the battery
beforehand (16 hours).
Keyboard description.
Key Function
ON, is used to switch the machine on.
OFF, is used to switch the machine off when pressed for over three
seconds.
SILENCE ALARME, is used to access the test mode.
The selection keys allow to scroll the figures and letters on the tens,
units, tens segments etc.
04.3_007a_en.fm 35
4.3.3 Running time and servicing inspection date
This test allows for n tSt.1, press the "VALIDATION" key.
display of the Pilot o If xxx.H: number of hours of use, 999 hours max.
running time since
o If xxx.J: number of days of use, 999 hours max.
its last serving
inspection. It also o If xxx.M : number of months, 999 months max.
allows for display (average duration of one month considered as 30 days).
and modification of
the “last” servicing n Press "VALIDATION" again to display the date of the last
inspection date. servicing inspection.
When the servicing o xx.d, day of the servicing inspection date.
inspection date is o Validate once again,
modified, the
s xx.m, month of the servicing inspection date.
running time is
reset. o By validating once again,
s xxx.x, year of the servicing inspection date.
Each time this information is read, the month, day and year of the
servicing inspection date may be modified by using the tens and
units "- and +" keys. This date will be stored in the EEPROM and
the running time will be reset.
The test may be stopped at any time by pressing the "STOP" key, and a
different test may be selected.
The test may be stopped at any time by pressing the "STOP" key, and a
different test may be selected.
Preventive maintenance
36 04.3_007a_en.fm
4.3.5 Keyboard check
This test allows to n tSt.3, press "VALIDATION".
check that all o The message CLAV is displayed.
keyboard keys
o Keep each key pressed down, one by one,
function correctly.
s Check the name of the key displayed on the display
unit.
StoP STOP
VAL VALIDATION
boLu BOLUS
S-.. + tens
S.- + units
S..- + tens
S_.. - tens
S._. - unit
S_.. - tens
If two or more keys are pressed simultaneously, the display unit shows Err and three
beeps are heard, indicating an error.
The "OFF" key is not included in the keyboard test.
The test may be stopped at any time by pressing the "VALIDATION" key
for over 3 seconds. Another test can then be selected.
If the display is faulty, replace the display board (see “Display and CPU boards”).
Preventive maintenance
04.3_007a_en.fm 37
4.3.6 Checking the battery voltage
This test enables n tSt.4, press the "VALIDATION" key.
users to display the o The voltage is displayed in Volts.
battery voltage in
Volts and tenths of
o By validating once again, a different test may be selected.
a Volt.
38 04.3_007a_en.fm
4.3.8 Total running time check
This test enables n tSt.6, press the "VALIDATION" key.
users to display the o If xxx.H: number of hours of use, 999 hours max.
total running time
o If xxx.J: number of days of use, 999 days max.
of the Pilot. It is
not possible to o If xxx.M : number of months, 999 months max.
modify this time (average duration of one month considered as 30 days).
manually. s By keeping the "tens" keys pressed down, the number
of times the equipment has been switched on is
displayed.
It is impossible to carry out this test when the PC is in communication with the Pilot. In this
case, “OPEN” is displayed.
Preventive maintenance
04.3_007a_en.fm 39
4.3.11 Checking the software version.
Run this test to n tSt.A, press the "VALIDATION" key.
display the o ex. V07.0 is displayed:
software version
and revision s 07, software version number.
numbers. s .0, revision number
o Press on the tens keys to display the Eprom checksum.
Ex.: 738.C.
o Press on the other keys to display the software version
index. Ex.: b.
o By validating once again, a different test may be selected.
Use the "-" and "+" selection keys to move from one input to another.
If the value is out of limits, recalibrate the position sensor (see “EtA.6 Position sensor
calibration.”).
Preventive maintenance
40 04.3_007a_en.fm
4.3.14 Buzzer test
Run this test to n tSt.d, press the "VALIDATION" key.
check that the
o The buzzer rings non stop and bEEp is displayed.
buzzer is working.
o By validating once again, a different test may be selected.
Display Name
bat.1 Alarm and pre-alarm battery voltage 6.3 V
Capacities which are non-existent or non-configured in the EEPROM are displayed in the
form of ---.-.
04.3_007a_en.fm 41
4.3.17 Checking the syringe group number
This test allows to n tSt.G, press the "VALIDATION" key.
display the list of o The syringe group number programmed EEPROM is
syringes for which
displayed, e.g. G01.
the equipment has
been configured.
o By validating once again, a different test may be selected.
n Press "ON".
o Lift the disengagement lever.
s Check the mechanical disengagement alarm (red
LED at the end of the syringe diagram).
o Fit the 50 cc syringe onto the equipment, ensuring the fin
and plunger are in position.
o Release the disengagement lever.
s Check that there is no mechanical lever release
alarm.
s Check that the plunger is locked.
This check must be carried out for every syringe provided with the equipment.
Check that no alarm goes off with spacers T301519 and T301518.
n Press "ON".
o Fit the 50 cc syringe on the equipment and lock it into
Preventive maintenance
position using the clamp, ensuring the fin and plunger are
correctly positioned.
o Place the manometer (or any other pressure
measurement equipment) at the syringe outlet.
o Select a 50 ml "B-D PLASTIPAK" syringe by pressing
"VALIDATION".
o Check the position of the backpressure adjustment button .
42 04.3_007a_en.fm
o Select a maximum flow rate of 120ml/h and initiate the
infusion by pressing the “VALIDATION” key.
s Ensure there is no acoustic and visual alarm
(backpressure LED off).
s Check that the infusion LEDs flash correctly.
s Ensure the alarm is triggered for a value of 0.7 bar +
0.1 bar.
o Stop the infusion cycle by pressing the "ARRET
PERFUSION" ("STOP INFUSION") key.
The backpressure button should not be raised once the backpressure alarm threshold
has been adjusted with the adjustment screw.
n Press "ON".
o Fit the syringe on the equipment, ensuring the fin and
plunger are in position.
o Check that the correct syringe is detected, e.g. 50c.c.
o Select a 50 m/l "B-D PLASTIPAK" syringe filled to 20 ml.
o Select a flow rate of 120 ml/h.
s For normal flow rates, the pre-alarm is activated 5
minutes before end of infusion.
s Flow rate example: > 72 ml/h, the pre-alarm is
activated when the remaining volume equals 10% of
the total syringe capacity.
s Ensure the end of infusion pre-alarm is present.
o Press the "SILENCE ALARME" and "TENS"
simultaneously.
s The acoustic alarm is silenced and the visual signal is
maintained.
o Measure the “hard height” at “end of infusion”.
s 18.5 < x < 19.5.
Top view
For accurate checking of the "hard height", do not move the plunger when measuring.
Preventive maintenance
If the "hard height" reading is out of limits, recalibrate the position sensor (voir “EtA.6
Position sensor calibration.”).
04.3_007a_en.fm 43
4.3.22 Checking the linearity
To carry out this operation, exit the test mode, press “OFF",
n Equipment required: Chronometer, calliper square, BD
Plastipak 50 ml. syringe.
n Press "ON".
o Fit the equipment with the "B-D PLASTIPAK" 50 ml
syringe, ensuring the fin and plunger are in starting
position.
o Measure the distance X in mm.
Top view
For accurate checking of the linearity do not move the plunger when measuring.
o Press "VALIDATION"
o Select a flow rate and validate.
44 04.3_007a_en.fm
o Reduce the test power supply voltage until the battery
discharge is triggered.
s Ensure this is triggered between 5.8 V and 6 V.
If the results obtained during the tests differ from the values indicated, recalibrate the battery
voltage levels (see “EtA.4 Calibration of the 3 battery voltage levels.”).
04.3_007a_en.fm 45
Preventive maintenance
46
04.3_007a_en.fm
Preventive maintenance
Conformity
N° Procedure Resulting value
Yes No
1 n Check the general condition of the case and its labels.
2 n Display total running time, tSt1
(in hours, days or months):***************************************** ..........................
3 n Display the last servicing inspection date, tSt1
(in days, months or years):***************************************** ..........................
4 n Check all indicator lights, tSt2.
5 n Check the keyboard, tSt3.
6 n Display the total running time, tSt6
(in hours, days or months):***************************************** ..........................
7 n Check the position sensor, tStC:
o High position with T300940 spacer, check that the displayed
value is 115.0 + 0.5 mm: ********************************** ..........................
o Low position with T300775 spacer, check that the displayed
value is 20.0 + 0.5 mm: ************************************ ..........................
8 n Check the syringe clamp, tStF:
o Syringe clamp in high position, check that the displayed
value is ---.- :******************************************** ..........................
o Position the 50cc syringe or the T301521 spacer, 50cc mini/maxi.
Check that he value displayed is 50c.c :**************** ..........................
o Position a 20cc syringe or a T301521 spacer, 20cc mini/maxi.
Check that he value displayed is 20c.c :**************** ..........................
o Syringe clamp in low position, check that the displayed
value is ---.- :******************************************** ..........................
9 n Check the anti-siphon arm:
o Free travel without end play.
o Presence of the alarm in high and low position.
o No alarm in presence of 20 cc and 50 cc syringe piston or T301519 et
T301518 spacers.
10 n Check the backpressure (use a Fresenius Vial dynamometer):
o Position of adjustment knob
Preventive maintenance
fiche_006a_en.fm 47
Conformity
N° Procedure Resulting value
Yes No
12 n Check the end of infusion alarm:
o With a 60 cc BD, check that the hard height range
is 18.5 < x mm < 19.5: ****************************************** ..........................
13 n Check the linearity (60 cc BD Plastipack):
o Measure the plunger starting position range, XI mm:********* ..........................
o Measure the plunger range after 50 mn pump running, X2 mm: ......................
o Check: 74.96 < X1-X2 mm < 76.47: *************************** ..........................
14 n Check the battery autonomy:
o Recharge the battery for 15 hours.
o Operate the Pilot for 1 hr at a flow rate of 120 ml/hr:********* ..........................
15 n Carry out the electrical tests according to standard EN 60601-1
Comments:
Preventive maintenance
48 fiche_006a_en.fm
Diagnosis
In order to purchase the ISDébit flow rate control software, contact the Fresenius Vial
customer service.
The test procedure defined below must be carried out using a 50 ml or 20 ml syringe.
Equipment required
n Scales coupled to a microcomputer:
Flow rate value Scales sensitivity
x < 5 ml/hr 1/10000th
5 ml/hr < x < 30ml/hr 1/1000th
x > 30 ml/hr 1/100th
n Multi scales acquisition programme.
n Test tube or beaker with 1 ml graduating
n Liquid: distilled water and oil
n Luer lock type plastic syringe (50 or 20 ml).
n Catheter extension with Luer Lock (length 100 cm, inside diameter 2,5 mm).
n Needle:
Flow rate value Needle type
x < 30ml/hr G26
x > 30 ml/hr G18 or G21
Installation
n According to the installation drawings shown below.
1 cm
Cord
Anti vibration measurement table
RS232
04.4_001a_en.fm 59
n Fill the syringe with 50 ml of distilled water.
o Prime to eliminate any air bubbles.
n Secure the female Luer Lock end piece of the catheter
extension onto the syringe and the male Luer lock end piece
onto the needle.
n Fit the syringe onto the device.
n Fill the test tube ensuring the needle is dipped in the liquid
(> 1 cm).
o Add several drops of oil to create a greasy film on the
surface of the liquid. This way the user will avoid any
measurement error due to evaporation of the liquid.
n Place the test tube in the centre of the scales platform.
n Place the needle inside the test tube.
The infusion line (needle/catheter extension) must not rest on the scales/test tube
assembly.
Operating mode
The software works following the operating mode described in the PrEN-60-601-2-24
standard project for infusion pumps.
60 04.4_001a_en.fm
4.4.2 Measurement with scales
Equipment required
n Stop clock
n Scales
Flow rate value Scales sensitivity
x < 5 ml/hr 1/10000th
5 ml/hr < x < 30ml/hr 1/1000th
x > 30 ml/hr 1/100th
n Test tube or beaker with 1 ml graduating
n Liquid: distilled water and oil
n Luer lock type plastic syringe (50 or 20 ml)
n Catheter extension with Luer Lock (length 100 cm, inside diameter 2,5 mm).
n Needle:
Flow rate value Needle type
x < 30ml/hr G26
x > 30 ml/hr G18 or G21
Installation
n According to the installation drawings shown below.
1 cm
04.4_001a_en.fm 61
The infusion line (needle/catheter extension) must not rest on the scales/test tube
assembly.
Operating mode
n Select a flow rate.
For low flow rates (< 5 ml/hr), validate and wait for the infusion to stabilise for 1 hour. For
higher flow rates, wait for 10 to 30 minutes after infusion.
1 gram = 1 ml.
62 04.4_001a_en.fm
4.4.3 Measurement using a test tube
Equipment required
n Stop clock.
n Test tube or beaker with 1 ml graduating
n Liquid: distilled water and oil
n Luer lock type plastic syringe (50 or 20 ml)
n Catheter extension with Luer Lock (length 100 cm, inside diameter 2,5 cm).
n Needle:
Flow rate value Needle type
x < 30ml/hr G26
x > 30 ml/hr G18 or G21
Installation
n According to the installation drawings shown below.
1 cm
04.4_001a_en.fm 63
Operating mode
n Select a flow rate.
For low flow rates (< 5 ml/hr), validate and wait for the infusion to stabilise for 1 hour. For
higher flow rates, wait for 10 to 30 minutes after infusion.
64 04.4_001a_en.fm
Diagnosis
n ¨TRICHLOROETHYLENE-DICHLOROETHYLENE:
o AMMONIA.
o AMMONIUM CHLORIDE
o CHLORINE AND AROMATIC HYDROCARBON.
o ETHYLENE DICHLORIDE-METHYLENE CHLORIDE
o CETONE based cleaning products.
These aggressive agents could damage the plastic parts and lead to apparatus malfunctions.
Also beware of ALCOHOL SPRAYS (20% -40% alcohol) that tarnish and crack the
plastic and fail to provide the cleaning action required prior to disinfection.
For further information, please contact the competent department in your hospital for supply
of the appropriate cleaning and disinfecting products.
Diagnosis
04.5&6_001a_en.fm 65
4.6 Storage
In case of prolonged storage, it is advisable to disconnect the battery using the battery access
door on the bottom of the device. This operation should be done by an experienced
technician.
The equipment must be stored in a dry and cool place.
n The recommended environmental temperature conditions for storage of the equipment are
between 0°C and 40°C.
n Relative humidity tolerated: max. 85%, no condensation.
Fully recharge the battery before using the equipment to avoid any risks caused by micro
power cuts in the mains supply and to ensure maximum autonomy.
Diagnosis
66 04.5&6_001a_en.fm
5 Diagnosis
5.1 Troubleshooting
05.1_003a_en.fm 57
Diagnosis
58 05.1_003a_en.fm
Diagnosis
05.2_004a_en.fm 59
Error code Description Recommended action
Calculation parameter errors (motor and flow rate)
Er14 n Motor period calculation n Check the ADC (see “Checking the
error. ADC”).
n Check the position sensor (see
Er24 n Motor rotation direction
“Checking the position sensor.”).
error.
60 05.2_004a_en.fm
6 Operation sheets
This chapter lists the set of dismounting/remounting sheets.
Operation sheets
06.0_001a_en.fm 71
Operation sheets
72 06.0_001a_en.fm
Operation sheets
Safety:
For safety reasons, the technician should not carry out any maintenance when the device is
connected to the 230 V mains supply voltage.
Disconnect the mains power supply cable.
Equipment required:
n 1 Posidriv Z1 screwdriver.
n 1 antistatic wriststrap.
Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).
Procedure:
Access
n Turn the Pilot over onto the upper case.
n Unscrew the 6 Phillips head screws (ref. 1) located at the bottom of the lower case, which
link this to the upper case.
n Keep the Pilot assembled and turn it over onto the lower case.
n Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.
n Disconnect the CPU board flat jumper.
Hands must not come into contact with the CPU boards.
1 1
1
1
Operation sheets
06.0_0010a_en.fm 73
Dismounting
n Unscrew the 5 Phillips head screws (ref. 2) located at the display board, which link this to
the upper case.
n Remove the board insulator (ref. 3) located on the left.
n Lift the display board slightly and remove the display unit flat jumper (ref. 4).
n Disconnect the black connector (ref. 5).
2
2
Remounting
A specific type of board corresponds to each Pilot "CPU and display board" ; It is
important to avoid reversing the references between each Pilot and order the part
number corresponding to your device.
Carry out the same procedures in reverse to reassemble the parts.
When mounting the display board, it is important to reduce the torque in the plastic inserts
so as to avoid causing damage to these.
Operation sheets
When fitting the upper case, ensure the joint is perfectly positioned in its slot after
reassembly.
Recalibrate the Pilot (see "Calibrations") then carry out the regular servicing tests (see
"Regular servicing sheet").
74 06.0_0010a_en.fm
Operation sheets
Safety:
For safety reasons, the technician should not carry out any maintenance when the device is
connected to the 230 V mains supply voltage.
Disconnect the mains power supply cable.
Equipment required:
n 1 Posidriv Z1 screwdriver.
n 1 Posidriv Z1 flat screwdriver.
Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).
Procedure:
Access
n Turn the Pilot over onto the upper case.
n Unscrew the 6 Phillips head screws (ref. 1) located at the bottom of the lower case, which
link this to the upper case.
n Keep the Pilot assembled and turn it over onto the lower case.
n Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.
n Disconnect the CPU board flat jumper.
Hands must not come into contact with the CPU boards.
1 1
1
1
Operation sheets
06.0_0020a_en.fm 75
Dismounting
n Remove the spring retaining ring (ref. 2) located on the syringe clamp shaft, which holds
this to the CPU support.
n Remove the obturator and the spring (ref. 3).
n Remove the syringe clamp.
Reassembly
Carry out the same procedures in reverse to remount the unit.
When fitting the upper case, ensure the joint is perfectly positioned in its slot after
reassembly.
76 06.0_0020a_en.fm
Intervention sheets
Safety:
For safety reasons, the technician should not carry out any maintenance when the device is
connected to the 230 V mains supply voltage.
Disconnect the mains power supply cable.
Tools required:
n 1 Posidriv Z1 screwdriver.
n 1 antistatic wriststrap.
n 1 soldering iron
n "RADIEL Sn60Pb RI 1" welding wire (cleaning not required for rewelding) or equivalent.
n Silicone ref. 161249.
Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).
Access
n Rotate the Pilot on the upper case.
n Unscrew the 6 Phillips head screws (ref. 1) Located at the bottom of the lower case, which
tighten the latter to the upper case.
n Keep the Pilot assembled and turn it over onto the lower case.
n Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.
n Disconnect the CPU board flat jumper.
Hands must not come into contact with the CPU boards.
1 1
1
1
Intervention sheets
06.0_0031a_en.fm 67
Dismounting
n Disconnect the connector (ref. 2).
n Unscrew the Phillips head screw (ref. 3) which holds the conductor to the upper case.
n Unscrew the 2 Phillips head screws (ref. 5) which hold the opto PCB in position.
n Remove the opto assembly.
5
Green
Black
2 White
Red
Reassembly
n Place the new opto assembly on the PCB support.
n Screw the 2 Phillips head screws (ref. 5) which hold the “PCB and opto” assembly in
position.
When mounting the "PCB and opto", it is important to reduce the torque in the plastic
inserts so as to avoid causing damage to these.
When fitting the upper case, ensure the joint is perfectly positioned in its slot after
reassembly.
68 06.0_0031a_en.fm
Operation sheets
Safety:
For safety reasons, the technician should not carry out any maintenance when the device is
connected to the 230 V mains supply voltage.
Disconnect the mains power supply cable.
Equipment required:
n 1 Posidriv Z1 screwdriver.
n 1 antistatic wriststrap.
Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).
Procedure:
Access
n Turn the Pilot over onto the upper case.
n Unscrew the 6 Phillips head screws (ref. 1) located at the bottom of the lower case, which
link this to the upper case.
n Keep the Pilot assembled and turn it over onto the lower case.
n Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.
n Disconnect the CPU board flat jumper.
Hands must not come into contact with the CPU boards.
1 1
1
1
Operation sheets
06.0_0041a_en.fm 83
Dismounting
n Press the syringe pump disengagement lever towards the back of the Pilot as far as
possible.
n Maintain this position and slide the whole unit to the back left.
n Remove the plunger guide.
n Disconnect the black connectors (ref. 2).
n Unscrew the flat jumper mounting lug (ref. 3) without removing it.
n Remove the mechanical plunger unit from its slot.
n Disconnect the blue connector from the flat jumper (ref. 4).
n Remove the mechanical plunger unit.
Reassembly
Carry out the same procedures in reverse to remount the unit.
Operation sheets
When fitting the upper case, ensure the joint is perfectly positioned in its slot after
reassembly.
84 06.0_0041a_en.fm
Operation sheets
Safety:
For safety reasons, the technician should not carry out any maintenance when the device is
connected to the 230 V mains supply voltage.
Disconnect the mains power supply cable.
Equipment required:
n 1 Posidriv Z1 screwdriver.
n 1 hexagon socket key (2.5).
n 1 soldering iron.
n "RADIEL Sn60Pb RI 1" welding wire (cleaning not required for re-welding) or equivalent.
Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).
Procedure:
Access
n Turn the Pilot over onto the upper case.
n Unscrew the 6 Phillips head screws (ref. 1) located at the bottom of the lower case, which
link this to the upper case.
n Keep the Pilot assembled and turn it over onto the lower case.
n Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.
Hands must not come into contact with the CPU boards.
1 1
1
1
Operation sheets
06.0_0051a_en.fm 85
Dismounting
n Press the syringe pump disengagement lever towards the back of the Pilot as far as
possible.
n Maintain this position and slide the whole unit to the back right.
n Remove the plunger guide.
n Unscrew the 3 Phillips head screws (ref. 2) located at the bottom of the plunger end shield
which links this to the plunger cover.
n Remove the plunger end shield.
2
2
5
4
6
Operation sheets
86 06.0_0051a_en.fm
n Unscrew the 2 Phillips head screws (ref. 7 and 8) which link the plunger support to the
plunger cover.
n Unscrew and remove the Phillips head screw and the washer (ref. 9) which attach the anti-
siphon cam to the anti-siphon arm.
n Remove the cam/arm assembly.
n Remove the plunger cover fitted with the bonding pad.
8
7 9
n Remove the retaining ring that holds the anti-siphon opto in place (ref. 10).
n Unscrew the 6 hexagon socket screws (ref. 11).
11
10
Operation sheets
06.0_0051a_en.fm 87
n Remove the backpressure micro support/spring leaf (ref. 12).
n Pull off the ribbon cable from the double-sided board
12
Welding and unwelding operations are carried out using a soldering iron fitted with a tip in
good condition, constantly tapered and kept clean. The temperature of the iron should be
between 315°C and 340°C.
When dismounting the force sensor, ensure the welding pellets are not damaged.
88 06.0_0051a_en.fm
Reassembly
n Weld the ribbon cable to the new pressure switch, ensuring the mounting procedure is
carried out in order:
o Taper the 3 welding points.
o Weld the 3 points.
o Check that the welding has been carried out correctly. It should not form a ball, but
should lie against the length of the component snubbers.
n Repeat the operation for the anti-siphon opto (4 welding points).
n Stick the bottom side of the ribbon cable to the double-sided board.
Anti-siphon opto
w
Backpressure switch
Lower side
(connection specific to Pilot C)
stuck to the double-sided board on the plunger support
n Impregnate with weak loctite and screw the 6 hexagon socket screws (ref. 11), which hold
the spring leaf/backpressure micro support in position.
n Position the opto and its retaining ring (ref. 10).
n Mount the plunger cover onto the plunger support. Impregnate the screw with weak loctite
(rep. 9).
n Position and screw the cam/anti-siphon arm unit using the Phillips head screw and the
washer (ref. 9).
n Screw the 2 Phillips head screws (ref. 8 and 7) which link the plunger support to the
plunger cover.
n Position the plunger cover clip (ref. 6).
Operation sheets
06.0_0051a_en.fm 89
n Cover the plunger with the protective film, mount the disengagement lever, its spring and
the adjustment button. Screw the plunger end shield back on (ref. 5, 4 and 3).
n Position the plunger cover and screw it back on (ref. 2).
Carry out the same procedures in reverse to remount the unit.
When fitting the upper case, ensure the joint is perfectly positioned in its slot after
reassembly.
Recalibrate the pressure sensor (see "Calibrations") then carry out the regular servicing tests
(see "Regular servicing sheet").
Operation sheets
90 06.0_0051a_en.fm
Operation sheets
Safety:
For safety reasons, the technician should not carry out any maintenance when the device is
connected to the 230 V mains supply voltage.
Disconnect the mains power supply cable.
Equipment required:
n 1 Posidriv Z1 screwdriver.
n 1 flat screwdriver.
n 1 flat key (12).
n 1 potentiometer dismounting tool ref. T300869.
Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).
Procedure:
Access
n Turn the Pilot over onto the upper case.
n Unscrew the 6 Phillips head screws (ref. 1) located at the bottom of the lower housing
case, which link this to the upper case.
n Keep the Pilot assembled and turn it over onto the lower case.
n Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.
n Disconnect the CPU board flat jumper.
Hands must not come into contact with the CPU boards.
1 1
1
1
Operation sheets
06.0_0060a_en.fm 89
Dismounting
n Press the syringe pump disengagement lever towards the back of the Pilot as far as
possible.
n Maintain this position and slide the whole unit to the back left.
n Remove the plunger guide.
n Disconnect the black connectors (ref. 2).
n Unscrew the flat jumper mounting lug (ref. 3) without removing it.
n Remove the mechanical plunger unit from its slot.
n Disconnect the blue connector from the flat jumper (ref. 4).
n Remove the mechanical plunger unit.
n Unscrew the potentiometer lock nut (ref. 5).
n Remove the tooth lockwasher and the potentiometer.
Reassembly
n Dismount the end shield from the reducer frame.
n Mount the potentiometer onto the end shield (screw the nut onto a thread).
n Position the end shield in equipment T 300 869 and lock it into position using the knurled
screw.
n Position the potentiometer in the device and bring it up against the end shield.
n Tighten the potentiometer.
n Extract the end shield from device T 300 869.
n Mount the pinion on the potentiometer (match up the indexing plane).
o The larger diameter of the pinion must be flattened against the potentiometer.
n With the potentiometer facing you, turn the pinion anticlockwise until it blocks, then turn it
¼ of a turn in the opposite direction.
n Mount the moving mechanical assembly onto the reducer frame.
n Slide the end shield onto the guides and rack.
o Check the position of the input bearing which must be on the plunger side.
o Ensure the flex circuit is not damaged when mounting (folding).
n Secure the end shield using the three M3x3 TC screws.
n Secure the input bearing using the two M3x3 TC screws.
Operation sheets
When fitting the upper case, ensure the joint is perfectly positioned in its slot after
reassembly.
Recalibrate the position sensor (see "Calibrations") then carry out the regular servicing tests
(see "Regular servicing sheet").
90 06.0_0060a_en.fm
Operation sheets
Safety:
For safety reasons, the technician should not carry out any maintenance when the device is
connected to the 230 V mains supply voltage.
Disconnect the mains power supply cable.
Equipment required:
n 1 Posidriv Z1 screwdriver.
Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).
Procedure:
Access
n Press the syringe pump disengagement lever towards the back of the Pilot as far as
possible.
n Maintain this position and slide the whole unit to the back right.
Dismounting
n Unscrew the 3 Phillips head screws (ref. 1) located at the bottom of the plunger end shield
which links this to the plunger cover. Remove the plunger end shield (ref. 2).
n Remove the backpressure adjustment button (ref. 3), then the disengagement lever and its
spring (ref. 4) as well as the protective plunger film (ref. 5).
n Unscrew and remove the Phillips head screw and the washer (ref. 6) which attach the anti-
siphon cam to the anti-siphon arm.
n Remove the cam/anti-siphon arm assembly (ref. 7).
4
6-7
3
5 2
1
Operation sheets
1 1
06.0_0071a_en.fm 93
Reassembly
Carry out the same procedures in reverse to remount the unit.
n Impregnate the screw with weak loctite (ref. 6) before mounting the cam and arm
assembly.
When fitting the upper case, ensure the joint is perfectly positioned in its slot after
reassembly.
94 06.0_0071a_en.fm
Intervention sheets
Safety:
For safety reasons, the technician should not carry out any maintenance when the device is
connected to the 230 V mains supply voltage.
Disconnect the mains power supply cable.
Tools required:
n 1 Posidriv Z1 screwdriver.
n 1 small flat screwdriver.
n 1 antistatic wriststrap.
Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).
Procedure:
Access
n Rotate the Pilot on the upper case.
n Unscrew the 6 Phillips head screws (ref. 1) located at the bottom of the lower case, which
tighten the latter to the upper case.
n Keep the Pilot assembled and turn it over onto the lower case.
n Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.
n Disconnect the CPU board flat jumper.
Hands must not come into contact with the CPU boards.
1 1
1
1
1
Intervention sheets
06.0_0082a_en.fm 81
Dismounting
n Press the syringe pump disengagement lever towards the back of the Pilot as far as
possible.
n Maintain this position and slide the whole unit to the back left.
n Remove the syringe pump guide.
n Disconnect the black connectors.
n Unscrew the flat jumper mounting lug (ref 2) without removing it.
n Remove the mechanical plunger unit from its slot.
n Disconnect the blue connector from the flat jumper (ref. 3).
n Remove the mechanical plunger unit.
n Disconnect the connectors (ref 4, ref 5, ref. 6, ref. 7, ref. 8)
n Unscrew the 3 Phillips head screws (ref. 9) located on the power supply board, which
tighten the latter to the lower case.
n Unscrew the 2 Phillips head screws (ref. 10) located on lower case, which tighten the latter
to the power supply board.
n Remove the power supply board.
9
8
3
9
6
5
4
7
10
Reassembly
Carry out the same procedures in reverse to remount the unit.
When fitting the upper case, ensure the joint is perfectly positioned in its slot after
Intervention sheets
reassembly.
Recalibrate the force sensor (see "Calibrations") then carry out the regular servicing tests
(see "Regular servicing sheet").
82 06.0_0082a_en.fm
Operation sheets
Safety:
For safety reasons, the technician should not carry out any maintenance when the device is
connected to the 230 V mains supply voltage.
Disconnect the mains power supply cable.
Equipment required:
n 1 Posidriv Z1 screwdriver.
Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).
Procedure:
Access
n Turn the Pilot over onto the upper case.
Dismounting
n Unscrew the 2 Phillips head screws (ref. 1) to be found in the lower case battery holder kit,
which link this to the lower case.
n Swivel the battery holder and remove the battery.
n Disconnect this.
Reassembly
Carry out the same procedures in reverse to remount the unit.
Operation sheets
It is recommended to perform a full loading and unloading cycle to ensure the battery is in
working condition.
06.0_0090a_en.fm 95
Operation sheets
96 06.0_0090a_en.fm
Intervention sheets
Safety:
For safety reasons, the technician should not carry out any maintenance when the device is
connected to the 230 V mains supply voltage.
Disconnect the mains power supply cable.
Tools required:
n 1 Posidriv Z1 screwdriver.
Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).
Procedure:
Access
n Rotate the Pilot on the upper case.
n Unscrew the 6 Phillips head screws (ref. 1) located at the bottom of the lower case, which
tighten the latter to the upper case.
n Keep the Pilot assembled and turn it over onto the lower case.
n Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.
n Disconnect the CPU board flat jumper.
Hands must not come into contact with the CPU boards.
1 1
1
1
1
Intervention sheets
06.0_00103a_en.fm 85
Dismounting
n Disconnect the two black connectors (ref 2).
n Unscrew the 2 Phillips head screws (ref. 3) located on the cabled plug support, which link
this to the lower case.
n Remove the cabled plug support.
3
Reassembly
Carry out the same procedures in reverse to remount the unit.
If the number of rear door connector points is different to the number of power
supply board points, make the connections as described below.
86 06.0_00103a_en.fm
Operation sheets
Safety:
For safety reasons, the technician should not carry out any maintenance when the device is
connected to the 230 V mains supply voltage.
Disconnect the mains power supply cable.
Equipment required:
n 1 Posidriv Z1 screwdriver.
Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).
Procedure:
Access
n Turn the Pilot over onto the upper case.
n Unscrew the 6 Phillips head screws (ref. 1) located at the bottom of the lower case, which
link this to the upper case.
n Keep the Pilot assembled and turn it over onto the lower case.
n Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.
Hands must not come into contact with the CPU boards.
1 1
1
1
Operation sheets
06.0_00110a_en.fm 99
Dismounting
n Press the syringe pump disengagement lever towards the back of the Pilot as far as
possible.
n Maintain this position and slide the whole unit to the back left.
n Remove the plunger guide.
n Disconnect the black connectors (ref. 2 and ref. 3).
4
5
2&3
Remounting
Fitting the flexible tongue:
n Remove the clip (ref. 6).
n Stick the flexible tongue (ref. 7) with the adhesive facing the ribbon cable.
n Reposition the clip making sure the ribbon cable and the flexible tongue are centred on the
mechanical unit sprocket.
n Lift both ends of the flexible tongue slightly.
7
Operation sheets
100 06.0_00110a_en.fm
Fitting the guiding flange:
There are several types of flange tongues. If you have an older version, replace it with that
provided with the kit.
n Position the new guiding flange (ref. 4) using the two screws.
additional screw
When fitting the upper case, ensure the joint is perfectly positioned in its slot after
reassembly.
Operation sheets
06.0_00110a_en.fm 101
Operation sheets
102 06.0_00110a_en.fm
Operation sheets
Safety:
For safety reasons, the technician should not carry out any maintenance when the device is
connected to the 230 V mains supply voltage.
Disconnect the mains power supply cable.
Equipment required:
n 1 Posidriv Z1 screwdriver.
n 1 cutter.
n GEB type 100% silicone grease.
Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).
Procedure:
Access
n Turn the Pilot over onto the upper case.
n Unscrew the 6 Phillips head screws (ref. 1) located at the bottom of the lower case, which
link this to the upper case.
n Keep the Pilot assembled and turn it over onto the lower case.
n Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.
Hands must not come into contact with the CPU boards.
1 1
1
1
Operation sheets
06.0_00130a_en.fm 107
n Press the syringe pump disengagement lever towards the back of the Pilot as far as
possible.
n Maintain this position and slide the whole unit to the back left.
n Remove the plunger guide.
n Disconnect the black connectors (ref. 2).
n Unscrew the flat jumper mounting lug (ref. 3) without removing it.
n Remove the mechanical plunger unit from its slot.
n Disconnect the blue connector from the flat jumper (ref. 4).
n Remove the mechanical plunger unit.
4
n Press the syringe pump disengagement lever towards the back of the Pilot as far as
possible.
n Maintain this position and slide the whole unit to the back right.
Dismounting
n Unscrew the 2 slotted head screws (ref. 5) which link the centering ring to the mechanical
end shield.
n Using the cutter, cut the input bearing and remove it from the plunger tube.
5
Operation sheets
108 06.0_00130a_en.fm
Remounting
n Mount the new slotted input bearing onto the plunger tube by twisting it.
n Laying it flat on a table, cut the O-ring using a cutter.
n Place the O-ring around the tube.
n Fit the O-ring into the centering ring slot (horizontal cut, so as to avoid lining it up with the
centering ring cut).
n Mount the stainless steel slotted plate onto the pin, by twisting it.
n Place the silicone grease cord onto the plate around the passage of the pin.
n Place the plate on the input bearing.
n Position the whole assembly against the mechanical end shield and screw it on using the
two slotted screws (ref. 5).
n Check the plunger slides correctly.
It should slide uniformly across the entire centering ring and slight friction is due to the O-
ring scraping against the tube.
Operation sheets
06.0_00130a_en.fm 109
Operation sheets
110 06.0_00130a_en.fm
Operation sheets
Safety:
For safety reasons, the technician should not carry out any maintenance when the device is
connected to the 230 V mains supply voltage.
Disconnect the mains power supply cable.
Equipment required:
n 1 Posidriv Z1 screwdriver.
n 1 antistatic wriststrap.
Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).
Procedure:
Access
n Turn the Pilot over onto the upper case.
n Unscrew the 6 Phillips head screws (ref. 1) located at the bottom of the lower case, which
link this to the upper case.
n Keep the Pilot assembled and turn it over onto the lower case.
n Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.
n Disconnect the CPU board flat jumper.
Hands must not come into contact with the CPU boards.
1 1
1
1
Operation sheets
06.0_00141a_en.fm 113
Dismounting
n Press the syringe pump disengagement lever towards the back of the Pilot as far as
possible.
n Maintain this position and slide the whole unit to the back left.
n Remove the plunger guide.
n Disconnect the black connectors (ref. 2).
n Unscrew the flat jumper mounting lug (ref. 3) without removing it.
n Remove the mechanical plunger unit from its slot.
n Disconnect the blue connector from the flat jumper (ref. 4).
n Remove the mechanical plunger unit.
4
n Press the syringe pump disengagement lever towards the back of the Pilot as far as
possible.
n Maintain this position and slide the whole unit to the back right.
Welding and unwelding operations are carried out using a soldering iron fitted with a tip in
good condition, constantly tapered and kept clean. The temperature of the iron should be
between 315°C and 340°C.
n Unweld the flex circuit (ref. 5) ensuring the disengagement switch is not damaged in the
process.
n Unscrew the 2 slotted head screws (ref. 6) which link the centering ring to the mechanical
end shield.
n Unscrew the 2 Phillips head screws (ref. 7) which link the centering ring to the mechanical
end shield.
n Remove the flex circuit clip (ref. 8).
5 7
Operation sheets
6
8
114 06.0_00141a_en.fm
n Remove the plunger guide.
n Unscrew the 3 Phillips head screws (ref. 9) located at the bottom of the plunger end shield
which links this to the plunger cover.
n Remove the plunger end shield.
9
9
12
11
10
13
Operation sheets
06.0_00141a_en.fm 115
n Unscrew the 2 Phillips head screws (ref. 14 and 15) which link the plunger support to the
plunger cover.
n Unscrew and remove the Phillips head screw and the washer (ref. 16) which attach the
anti-siphon cam to the anti-siphon arm.
n Remove the cam/arm unit as well as the plunger cover.
14
15 16
n Remove the retaining ring that holds the anti-siphon opto in place (ref. 17).
n Unscrew the 6 hexagon socket screws (ref. 18).
18
17
17
Operation sheets
116 06.0_00141a_en.fm
n Remove the backpressure micro support/spring leaf (ref. 19).
n Pull off the ribbon cable from the double-sided board
19
Welding and unwelding operations are carried out using a soldering iron fitted with a tip in
good condition, constantly tapered and kept clean. The temperature of the iron should be
between 315°C and 340°C.
When dismounting the force sensor, ensure the welding blocks are not damaged and the
square is not removed in the process.
n Remove the flex circuit and tube kit from the plunger support.
Operation sheets
06.0_00141a_en.fm 117
Reassembly
n Remove the sticky tape holding the flex circuit to the tube.
n Mount the new flex circuit onto the plunger support.
n Weld the ribbon cable to the new pressure switch, ensuring the mounting procedure is
carried out in order:
o Taper the 3 welding points.
o Weld the 3 points.
o Check that the welding has been carried out correctly. It should not form a ball, but
should lie against the length of the component snubbers.
ube
Switch
w disengagement
Switch
backpressure
Anti-siphon opto
Lower side
(connection specific to Pilot C)
stuck to the double-sided board on the plunger support
Do not forget to impregnate the cam/arm clamping screw (rep. 16) with weak loctite.
When fitting the upper case, ensure the joint is perfectly positioned in its slot after
reassembly.
Recalibrate the force sensor (see "Calibrations") then carry out the regular servicing tests
(see "Regular servicing sheet").
Operation sheets
118 06.0_00141a_en.fm
Operation sheets
Safety:
For safety reasons, the technician should not carry out any maintenance when the device is
connected to the 230 V mains supply voltage.
Disconnect the mains power supply cable.
Equipment required:
n 1 Posidriv Z1 screwdriver.
n 1 antistatic wriststrap.
n 1 soldering iron.
n "RADIEL Sn60Pb RI 1" welding wire (cleaning not required for rewelding) or equivalent.
Maintenance level:
Level 2, specialised technician (see documentation on biomedical training).
Procedure:
Access
n Turn the Pilot over onto the upper case.
n Unscrew the 6 Phillips head screws (ref. 1) located at the bottom of the lower case, which
link this to the upper case.
n Keep the Pilot assembled and turn it over onto the lower case.
n Remove the upper case making sure the CPU board flat jumper is not accidentally
removed in the process.
n Disconnect the CPU board flat jumper.
Hands must not come into contact with the CPU boards.
1 1
1
1
Operation sheets
06.0_00150a_en.fm 119
Dismounting the upper cover
n Dismount the central unit and display boards (see corresponding sheet).
n Dismount the syringe clamp (see corresponding sheet).
n Dismount the syringe detection system (see corresponding sheet).
Welding and unwelding operations are carried out using a soldering iron fitted with a tip in
good condition, constantly tapered and kept clean. The temperature of the iron should be
between 315°C and 340°C.
o Unweld the snubbers from the buzzer:
s add more weld to facilitate the unwelding process.
s Heat and pull on the buzzer pins one by one.
Clean the buzzer slot surfaces using a cloth soaked in rubbing alcohol.
+
Buzzer with snubbers
Display board
s taper the 2 buzzer pins to be welded.
s place the buzzer onto the board following the
mounting diagram.
s weld to buzzer onto the board.
s check that the welding has been carried out correctly.
It should not form a ball, but should lie against the
length of the wire.
Operation sheets
Continue with the remounting procedure, referring to the “central unit and display board”
sheet.
120 06.0_00150a_en.fm
Dismounting the lower cover
n Dismount the power supply board (see corresponding sheet).
n Dismount the battery holder (see corresponding sheet).
n Dismount the rear plug support (see corresponding sheet).
Operation sheets
06.0_00150a_en.fm 121
Operation sheets
122 06.0_00150a_en.fm
7 Calibration
7.1 Calibration procedure
Keyboard description.
Key Function
ON, is used to switch the machine on.
OFF, is used to switch the machine off when pressed for over three
seconds.
SILENCE ALARME, is used to access the calibration mode.
The select keys allow to scroll the figures and letters on the LCD
screen, on the units, tens, tens segments etc.
Activate calibration
n Press the "SILENCE ALARME" and
"BOLUS" keys simultaneously.
07.1_005a_en.fm 105
7.1.2 EtA.4 Calibration of the 3 battery voltage levels.
This menu enables n EtA.4, press "VALIDATION".
the user to store o Bat.1: Supply the equipment with 6.3 V+ 0.05 using a
the three Bat1,
stabilised power supply.
Bat2 and Bat3
s Press "VALIDATION".
battery voltage
The voltage is read and stored in the EEPROM.
values in an
o Press "VALIDATION".
EEPROM.
o Bat.2: Supply the equipment with 5.9 V+ 0.05 using a
stabilised power supply.
s Press "VALIDATION".
The voltage is read and stored in the EEPROM.
o Press "VALIDATION".
o Bat.3: Supply the equipment with 5.7 V+ 0.05 using a
stabilised power supply.
s Press "VALIDATION".
The voltage is read and stored in the EEPROM.
Once both high and low values have been stored in the EEPROM, the equipment
indicates the number of LSB in decimals between the two measurement points.
This value should be 772 + 10 LSB.
106 07.1_005a_en.fm
8 Spare parts catalogue
8.1 Upper case
99
100
101
.....
.....
.....
.....
.....
..... .....
.....
.....
..... .....
.....
.....
08.1_007a_en.fm 107
.....
.....
.....
122
111
112
104
125
126 114
127
123
128 178
103
124
108 08.1_007a_en.fm
Spare parts catalogue
115
116
108
117
118
107
119
120
121
08.2_007a_en.fm 109
Spare parts catalogue
110
08.2_007a_en.fm
Spare parts catalogue
.....
..... .....
.....
..... .....
200
201 ..... ..... .....
202 ..... .....
224
225
203
204
205
206 .....
207 ..... ..... .....
208 220
221 223 .....
209 .....
..... .....
222 .....
.....
..... .....
.....
210
08.3_007a_en.fm 111
249
229 260
261
..... 253 250
..... 230 254 251
..... 252
..... 231
.....
.....
.....
.....
..... 270
..... 232 271 272
..... .....
..... ..... 255
..... ..... 256
..... 257
258
259
226
..... 227
.....
273
..... : see next page
Mark Qty Reference Name
226 1 167259 Plunger cover clip
227 1 167442 Injected M 0.5 rack
229 1 167237 Bonding pad
230 1 167411 Backpressure spring leaf
231 1 167273 Backpressure micro support (plunger)
232 1 173408 OMRON micro-switch
249 1 162311 Photo switch (type RP I 131)
250 1 167492 Backpressure spring
251 1 167270 Backpressure spring leaf adjusting screw
252 1 167269 Backpressure adjustment button
253 1 167282 Plunger bracket
254 1 167493 Protective plunger film
256 1 167488 Disengagement finger
257 1 167487 Sintered disengagement shaft bearing
258 2 167298 Injected flexible PCB protector
259 1 167272 Injected clamping collar
260 1 190714 Retaining ring
Spare parts catalogue
112 08.3_007a_en.fm
298
08.3_007a_en.fm 113
Spare parts catalogue
114
08.3_007a_en.fm
Spare parts catalogue
315
314
312
311 313
308
309
310
296
300
301
302
303
304
08.4_007a_en.fm 115
Spare parts catalogue
116
08.4_007a_en.fm
Spare parts catalogue
8.5 Labels
08.5_007a_en.fm 117
Spare parts catalogue
118
08.5_007a_en.fm
Useful addresses
SALES DEPARTMENT
Fresenius Vial
Le Grand Chemin, 38590 Brézins
CUSTOMER SERVICE Tel.: 00 33 (0) 4 76 67 10 81
or 00 33 (0) 4 76 67 10 54
or 00 33 (0) 4 76 67 11 08
Fax: 00 33 (0) 4 76 67 11 34
TRAINING DEPARTMENT
Fresenius Vial Tel.: 04 76 67 10 76
Le Grand Chemin, 38590 Brézins Fax: 04 76 67 11 22
This document may not be reproduced in part or in whole without the written permission of
Fresenius Vial S.A.