Unicel DxI Operator's Guide
Unicel DxI Operator's Guide
Unicel DxI Operator's Guide
C73199-AA
March 2021
Printed in U.S.A.
© 2021 Beckman Coulter, Inc.
Beckman Coulter, Inc. • 250 S. Kraemer Blvd. • Brea, CA 92821 • U.S.A.
Publication Notes
This manual, C73199-AA, release date 2021-03, supports UniCel DxI published system software version 5,
including full versions 5.1, 5.2, 5.3.0, 5.3.1, 5.5.0, 5.6.0, and 5.7.0.
Revision History:
C48057-AA, August 2020 UniCel DxI software versions 5.1, 5.2, 5.3.0, 5.3.1, 5.5.0, and 5.6.0
C38351-AB, March 2019 UniCel DxI software versions 5.1, 5.2, 5.3.0, 5.3.1, and 5.5.0
C38351, January 2019 UniCel DxI software versions 5.1, 5.2, 5.3.0, 5.3.1, and 5.5.0
C00548-AA, October 2016 UniCel DxI software versions 5.1, 5.2, 5.3.0, and 5.3.1
B13863F, March 2016 UniCel DxI software versions 5.1, 5.2, and 5.3.0
B13863E, June 2015 UniCel DxI software versions 5.1 and 5.2
B13863D, February 2015 UniCel DxI software version 5.1.
Beckman Coulter, Inc. grants a limited non-exclusive license to the UniCel DxI system owner or operator to
make a copy of all or a portion of this book solely for laboratory use.
Beckman Coulter, the stylized logo, and the Beckman Coulter product and service marks mentioned herein are trademarks or registered
trademarks of Beckman Coulter, Inc. in the United States and other countries.
ProClin™ is a trademark of The Dow Chemical Company ("Dow") or an affiliated company of Dow.
Contrad is a trademark of Decon Laboratories, Inc.
Citranox is a trademark of Alconox, Inc.
Monovette is a trademark of Sarstedt A.G. & Co.
All other trademarks are the property of their respective owners.
UniCel DxI Operator's Guide Table of Contents
Table of Contents
Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Preface-1
2 Supplies. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2.1 Supply and Supply Conditions Overview . . . . . . . . . . . . . . . . . . 2-3
2.2 Substrate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-16
2.3 Wash Buffer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-27
2.4 Reaction Vessels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-36
2.5 Liquid Waste . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-41
2.6 Solid Waste . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-48
2.7 Reagent Supplies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-53
2.8 Reagent Inventory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-70
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Index-1
Preface
About This Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Preface-2
About This This UniCel DxI Operator's Guide provides the same content delivered in the Help
Manual system Operator's Procedures book, but in a manual format.
NOTE
Most screen, window and report examples in this manual are based on the
UniCel DxI 800 system, and may show up to four reagent pipettors. The
UniCel DxI 600 system has two reagent pipettors.
UniCel DxI The UniCel DxI system includes the following documentation.
System • The UniCel DxI Help system, which provides detailed operating and
Documentation reference information, linked to the system software.
• The Instructions for Use, which contains information intended to be used
after you have become familiar with the UniCel DxI system.
• The LIS Vendor Information document, which provides interface information
for your laboratory information system vendor.
• The Operator's Guide and Reference Manual, which provide the same
content delivered in the Help system Operating Procedures and Reference
Information books. These manuals are available in the English language.
Manuals, Instructions for Use, and other supporting documents for your UniCel DxI
system are available on the Beckman Coulter website. Examples of available
documents include:
• UniCel DxI Instructions for Use
• UniCel DxI Operator's Guide & Reference Manual
• Access 2/UniCel DxI LIS Vendor Information Document
• Access assay instructions for use
• Safety Data Sheets for Access assays and consumables
• Certificates of analysis
To view or download electronic copies of UniCel DxI system documentation, visit the
Beckman Coulter website at techdocs.beckmancoulter.com. To receive email alerts
when new or updated UniCel DxI systemdocuments are released on the website,
register for the "My Technical Documents" notification tool.
Using the To make information easier for you to locate within a procedure or topic, UniCel DxI
Manual system documentation uses the following presentation styles.
Text Formatting
To call your attention to system software elements or indicators on the instrument, the
system documentation uses capital letters or bold type.
Example
If you are getting low on supplies, the Bulk Supplies system status button turns
yellow and the Amber status indicator on the instrument is lit. Select the Bulk
Supplies button and review the fields on the Bulk Supplies screen. If you need
more information, select the Help button.
Example
Both required and optional steps are illustrated below.
1. This is the first required step in a procedure.
(Optional) This is an optional step in a procedure.
2. This is the second required step in a procedure.
Modes
The system modes in which a procedure can be performed are included in a mode
statement, located before the first step of the procedure.
Example
The mode indicators and the mode statement format are illustrated below.
System Mode: Ready Running Paused
Ready Not Ready
The Help The UniCel DxI Help system provides context sensitive assistance for using the
System system software. In addition to the summary information included in the Instructions
for Use manual, the Help system contains expanded procedures for using the
UniCel DxI system. The Help system is available in the same languages as the system
software.
Information in the Help system is organized into main books, which are further
subdivided into books of related categories of information.
Using the You can access the Help system from most screens and windows in the UniCel DxI
Help System system software. Help information is presented in Help windows in small chunks
called topics or procedures.
From any Help system topic, you can link to other topics using navigational features
such as the link buttons, linked text highlighted in green, the Button Bar, or the tabs in
the Help Topics window.
Help
Description
Buttons
When you select the Help button from a software screen, the Help
system displays information about the screen. The Help window
includes a button at the top of the window for linking to the screen
illustration, and a column of arrow buttons on the left for linking to
related topics.
Help
Description
Buttons
When you select Help F12 from a window, the Help system displays
procedural information related to the window you are in. The Help
window often includes buttons at the top of the window for linking to
the window illustration or other related topics.
Help windows are displayed over the system software screens and windows. Different
types of windows display different types of information in designated areas of the PC
monitor. The window placement is intended to make it possible for you to navigate
from one topic to another without losing sight of a window until you are finished
reviewing the topic.
If you select anything in the system software while the Help system is available, the
Help system becomes hidden from view. When you select a Help button again, you
may see windows that were previously displayed.
You can exit the Help system by selecting Exit from the Help Button Bar.
NOTE
Select Exit only when you are completely done with the Help system. When you
select Exit, you exit the Help system and all Help windows close.
Button Description
(See Also button) Select to display the Topics Found window, which
contains a list of topics related to the text that you are viewing. To
view a related topic, select it and then select the Display button.
(Flowchart button) Select to display a flowchart of the procedure or
process that you are viewing.
Button Description
Many buttons are preceded or followed by green text. You can select either the text or
the button to link to the related topic.
2167A.bmp
Help Window Menu The Help window menu contains features that can
(Do Not Use) change the appearance and performance of the Help.
Beckman Coulter has optimized the Help system for
your use. Do not use the Help window menu.
Find (Button) Displays the Find view of the Help Topics
window, which you can use to search a database for a
word. For more information, see the Help Find Feature
topic.
Help Topics (Button) Displays the Contents view of the Help Topics
window. For more information, see the Help Topics
Window topic.
Back (Button) Select to return to the last displayed topic in the
same window position on the PC monitor.
This button is only available if a topic was previously
displayed in the same window position.
Print (Button) Select to print the displayed topic.
Glossary (Button) Select to display the Help system glossary.
This button is available only in some Help windows.
Exit (Button) Select to exit the Help system.
2168A.bmp
Contents (Tab) Select to display the table of contents for the Help
system. From the table of contents, you can:
• Select a topic page (the white icon with a ? symbol
inside) to display the Help about the topic.
• Select a book to list more available Help topic pages and
books related to this book.
Index (Tab) Select to locate information by typing the first few
letters of a word, and then by selecting one of the available
topics.
Find (Tab) Select to search a database for a word. For more
information, see the Help Find Feature topic.
The first time you use the Find feature, the Find Setup Wizard window is displayed to
assist you in setting up the search database. Select from the following options.
• Minimize database size - The search results in a list of the main topics
containing the word you entered. Beckman Coulter recommends that you
select this option.
• Maximize search capabilities - The search results in a list of every topic
containing the word you entered. It may take longer to get the results of the
search than if you selected to minimize the database size.
• Customize search capabilities - You select the characteristics of the search
in response to questions provided by the Find Setup Wizard.
After you select a database option, select Next and then select Finish. The Help
system creates the database and displays the Find window.
Notes, System documentation uses notes, cautions, and warnings to draw your attention to
Cautions, and important information, instrument operation instructions, and safety information.
Warnings
Graphic and Text Description
Technical For technical assistance with the UniCel DxI Access Immunoassay System:
Support • In the U.S.A. or Canada, contact Beckman Coulter Technical Support by
phone at 1-800-854-3633 or online at www.beckmancoulter.com. Before
using online support the first time, you will need to register online.
• Outside the U.S.A. and Canada, contact your technical support
representative.
NOTE
Be prepared to provide your system ID.
1 System Overview
1.1 System Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
Warranty. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
Reports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-31
The UniCel DxI system provides continuous, random access processing. While the
system is running tests, you can load routine or STAT samples, replenish supplies,
and change waste containers.
The system processes patient, quality control, and calibration samples in individual,
disposable plastic reaction vessels (RVs), using reagents from up to 50 on-board
reagent packs.
The system schedules tests as defined by the assay protocol file (APF) for each assay.
If there are not enough supplies to complete a test, system indicators inform you, so
you can attend to the supply condition and the test can be scheduled. With a large
sample storage area, a large capacity for on-board supplies, and multiple independent
reagent pipettors, the system schedules sample processing for optimum performance,
to produce up to 400 test results per hour on the UniCel DxI 800 instrument, and up to
200 test results per hour on the UniCel DxI 600 instrument.
For more information about the UniCel DxI system, instrument, and assay
technology, see Chapter 1: System Technology in the Reference Manual.
The UniCel DxI system consists of two major subsystems: the instrument, which
performs all sample processing functions, and the system console, which is the human
interface to the instrument software and consists of the external computer (PC),
touchscreen monitor, keyboard, mouse, printer, and PC bar code reader.
21 7
1
20 8
2
3
9 10
19
11
4
18 5
12
6
13
14
15
17 16
2055A.eps
19 7
1
18 8
2
3
9 10
4
17 5
11
6
12
13
14
16 15
2055B.eps
• Substrate Bar Code Reader - The substrate bar code reader is used to scan
the bar code labels on substrate bottles after you load them on the instrument.
For more information, see the Scanning the Substrate Bar Code Label topic
in Section 2.2: Substrate.
• Lower right door and instrument power switch - The instrument power
switch and the air filter are located behind the lower right door.
• Substrate Load Area - The substrate load area holds two bottles of
chemiluminescent substrate. You can replace an empty bottle while the
instrument continues processing with the other bottle. For more information,
see Section 2.2: Substrate.
• Wash Buffer Supply Drawer - The wash buffer supply drawer holds two
bulk wash buffer containers. You can replace a bulk wash buffer container
that is empty while the instrument continues processing with the second
container. For more information, see Section 2.3: Wash Buffer.
• Solid Waste Door - The solid waste door provides access to the solid waste
container. You empty the solid waste container while the instrument
continues processing in-progress tests. The instrument will temporarily
suspend disposal of waste products, and will not start any new tests until the
container is replaced. For more information, see Section 2.6: Solid Waste.
• Liquid Waste Drawer - If your DxI system does not use an external liquid
waste drain, the liquid waste drawer holds two reusable containers for liquid
waste produced during processing. You can empty a bulk liquid waste
container that is full while the instrument continues processing with the
second container. For more information, see Section 2.5: Liquid Waste.
• Sample Presentation Unit - The sample presentation unit (SPU) receives
sample racks loaded by an operator and moves them from the onload area to
the sample aliquot station for aliquoting. When sufficient sample volume is
aliquoted for all scheduled tests, the SPU moves the sample racks to the
offload area. For more information, see Section 3.4: Loading and Unloading
Racks.
• Side Offload Area - The side offload area receives racks from the SPU so
they can be removed while the instrument is running. For more information,
see Section 3.4: Loading and Unloading Racks.
• Vessel Hopper - The vessel hopper receives RVs loaded by an operator,
automatically orients them properly, and dispenses them to the internal
processing areas of the instrument. For more information, see
Section 2.4: Reaction Vessels.
• Left and Right Main Upper Covers - The left and right main upper covers
provide access to the mechanisms inside the instrument, for preventive
maintenance. For more information, see Chapter 7: Routine Maintenance.
Status The four status indicator lights are arranged vertically on the front panel of the
Indicator instrument. The Red, Green, and Blue indicator lights designate a specific instrument
Lights operating mode. The Amber indicator light informs you when a supply level requires
your attention.
2091A.eps
1 Red: Not ready. For more information, see the Red Status Indicator
Light topic.
2 Amber: Supplies required. For more information, see the Amber Status
Indicator Light Topic.
3 Green: Running. For more information, see the Green Status Indicator
Light topic.
4 Blue: Ready. For more information, see the Blue Status Indicator Light
topic.
The system console displays the Not Ready system mode. For more information, see
the System Modes topic in Section 1.3: Software Overview.
Installation The UniCel DxI Access Immunoassay System must be installed by a qualified
Beckman Coulter technical support representative. Do not remove the instrument
from the shipping crate until a technical support representative is present.
Warranty The UniCel DxI Access Immunoassay System is covered by and subject to the
provisions of the warranty included in your contractual agreement for the system or its
reagents.
Safety The UniCel DxI Access Immunoassay System is designed to meet U.S. and
Features international safety standards. Safety labels are affixed to the instrument to alert you
to safety considerations. Interlock switches stop the movement of the pipettors and
robotic modules to protect you from injury if you open the instrument covers while
the system is running.
Interlock Switches
The UniCel DxI system is equipped with interlock switches to protect you from
injury. If you open the covers of the instrument, the interlock switches stop all moving
parts. If you open the covers while the system is processing samples, the system
cancels all tests in progress.
Safety Certain areas of the UniCel DxI instrument present a risk of personal injury or
Symbols damage to the instrument if proper safety procedures are not followed. These areas are
marked with one or more safety symbols to identify the hazard. These symbols are
defined in the Access Immunoassay Systems Symbol Glossary, available at
techdocs.beckmancoulter.com.
Laser Symbol
The laser symbol indicates areas of the instrument where laser light is used. Do not
stare into the laser beam.
2208A.eps
Figure 1-5 Laser Symbol
The UniCel DxI laser products conform with the provisions outlined in Code of
Federal Regulations Title 21 (subchapter J, section 1040.12), and with IEC 60825-1.
CAUTION
Do not remove the laser warning labels or the protective housings that shield
the lasers. The lasers are accessible if the protective housings are removed.
Only a trained Beckman Coulter technical support representative should
service the lasers. Do not stare directly into a laser beam if the instrument
cover is opened or removed.
2368A.eps
1 This label is attached to the SPU bar code reader and the reagent storage
bar code reader.
CAUTION
VORSICHT
11-110008-01
LASER RADIATION
LASERSTRAHLUNG, NICHT IN DEN
DO NOT STARE INTO BEAM LICHTSTRAHL BLICKEN. SICHTBARE
CLASS 2 LASER PRODUCT LASERSTRAHLUNG.
670nm 1.0mW 101~186µs
LASERKLASSE 2
IEC 60825-1 Ed. 2 (2007)
CAUTION - CLASS 2 LASER RADIATION WHEN OPEN-DO NOT STARE INTO THE BEAM.
PRODUCT COMPLIES WITH 21 CFR 1040.10 AND 1040.11 EXCEPT FOR DEVIATIONS
PURSUANT TO LASER NOTICE NO. 50, DATED JUNE 24, 2007
2369A.eps
Figure 1-6 Laser Warning Label Placement Behind the UniCel DxI Sample
Processing Unit (SPU)
2 These labels are attached to the housing that shields the reagent storage
bar code reader.
2370A.eps
&
2371A.eps
Figure 1-6 Laser Warning Label Placement Behind the UniCel DxI Sample
Processing Unit (SPU)
2372A.eps
2373A.eps
Figure 1-7 Laser Warning Label Placement on the Back of the UniCel DxI
Instrument
2
PRODUCT COMPLIES WITH
21 CFR CHAPTER 1, SUBCHAPTER J
2374A.eps
Figure 1-7 Laser Warning Label Placement on the Back of the UniCel DxI
Instrument
Safety The following statements describe general safety concerns and provide information
Statements about attention symbols with no accompanying text.
WARNINGS
• The UniCel DxI instrument has moving parts which present an injury
hazard. Do not operate the UniCel DxI instrument with the covers or doors
open.
• Reagents, calibrators, and controls used with the system may contain small
quantities of sodium azide preservative. Sodium azide preservative may
form explosive compounds in metal drain lines. Refer to National Institute
for Occupational Safety and Health Bulletin: Explosive Azide Hazards
(8/18/76).
• Always plug the UniCel DxI system into a grounded three-conductor outlet.
DO NOT bypass the grounding prong on the plug.
• Do not defeat the safety interlock switches on the covers.
CAUTION
Replace substrate bottles only with the top cover closed to avoid spillage into
the instrument.
Regulatory The UniCel DxI Access Immunoassay System meets the requirements of a variety of
Symbols and domestic and international regulatory agencies, standards, and directives. This
Statements compliance is indicated by symbols and marks on the instrument. These symbols are
defined in the Access Immunoassay Systems Symbol Glossary, available at
techdocs.beckmancoulter.com.
CAUTIONS
• This equipment has been designed and tested to CISPR 11 Class A. In a
domestic environment it could cause radio interference, in which case
you may need to take measures to mitigate the interference.
• Prior to operation of this device, the electromagnetic environment
should be evaluated. Do not use this device in close proximity to sources
of strong electromagnetic radiation (for example, unshielded intentional
RF sources) as these could interfere with proper operation.
• If you suspect interference between the UniCel DxI system and other
equipment, you must take whatever action is required to correct the
interference. Beckman Coulter suggests the following actions:
- Move the equipment so there is a greater distance between the equipment
and the UniCel DxI system.
- Re-orient the equipment with respect to the UniCel DxI system.
- Be sure that the equipment is operating from a different power service
connector than that of the UniCel DxI system.
Other In addition to safety and regulatory symbols, other symbols are placed on the
Symbols instrument to identify functional elements such as network jacks and the air filter.
These symbols are defined in the Access Immunoassay Systems Symbol Glossary,
available at techdocs.beckmancoulter.com.
The UniCel DxI system software consists of a real-time operating system in the
instrument and a PC-based user interface (UI). The instrument software controls all of
the processing functions. The UI software allows you to interact with and direct the
instrument software. It also contains sample, calibration, and quality control
databases, as well as the assay protocol file (APF). The APF contains assay-specific
information about sample processing and test result calculations. The UI software is
located on the hard drive of the external computer (PC). The instrument software, UI
software, and APF are installed on the system through the CD-ROM drive on the PC.
NOTE
Your ability to perform a function can depend on the system mode. For example,
it is not possible to delete test results or perform a system backup when the
instrument is in the Running mode. For more information about system modes,
see Table 1-13.
The system also includes a laboratory information system (LIS) interface, which an
external LIS can use to communicate with the instrument through a standard EIA-232
serial connection. Test requests entered at the LIS are transmitted to the instrument for
processing. Once processing is complete, the test results are sent back to the LIS. For
more information about the LIS interface, see Chapter 1: System Technology in the
Reference Manual. For technical information about the LIS interface, see the LIS
Vendor Information CD. For information about the processing of LIS test requests,
see Section 3.2: LIS Requests and Work Pending.
Running Only the UniCel DxI system software should be installed on the external computer.
Other Do not install or remove any software except as directed by a Beckman Coulter
Applications technical support representative.
Do not use the external computer to view documentation other than the system Help.
Screens and The UI is organized by function into a set of screens that have common elements.
Windows From any screen, you can pause or stop the instrument, monitor the system mode,
open the Help system, view the Event Log, select a function or an option to operate
the system, or go to another screen.
The system will sometimes display a smaller window on top of a screen to prompt
you to perform an action, confirm that you want the system to perform an action,
display a message, or allow you to enter information.
System The UniCel DxI system provides three navigation methods that you can use in any
Navigation combination to select an item on a screen or in a window:
• Touching a part of the screen using the tip of your finger or fingernail
• Using the mouse to point the cursor and clicking the left mouse button to
select an item
• Pressing a keyboard equivalent such as a function key or the tab key
Keyboard Equivalents
The following table lists the equivalent keystrokes you can use to perform common
functions on the UniCel DxI system.
The word enter is used throughout the system documentation to signify putting
information into a field. Depending on the options available for the field, you can
enter information using one of the following methods:
• Type the information on the keyboard
• Select from available options or items in a list
• Scan bar coded information with the handheld bar code reader
Main Menu The Main Menu is the first screen displayed when the system is installed or
initialized. To get to this menu from any screen, press [F9] Main Menu on the
keyboard.
You can navigate almost everywhere through the UniCel DxI user interface beginning
from the eight function buttons across the bottom of the Main Menu. These buttons
are direct routes to functions you perform frequently.
The Main Menu function buttons are available from other screens when you use the
Menu tab. The Menu tab is displayed along the right edge of every screen except the
Main Menu.
2001C.bmp
Sample Manager Select to display the Sample Manager screen. Use this screen to:
F1 • View sample status information on one of four screen views
• Request new tests for QC and patient, calibration, or maintenance
samples
• Edit the test requests on a rack
• Display specific details for a selected rack or sample
For more information, see Chapter 3: Sample Manager.
Test Results Select to display the Test Results screen. Use this screen to:
F2 • View or print test results
• Troubleshoot flagged test results
• Send test results to the LIS
• Copy test results to a disk
• Delete test results
For more information, see Chapter 4: Test Results.
Supplies Select to display the Supplies menu. Use this menu to:
F3 • View the status of system supplies and waste containers, or change the
in-use containers
• View details about on-board reagent packs and to unload reagent packs
For more information, see Chapter 2: Supplies.
Quality Control Select to display the Quality Control screen. Use this screen to:
F4 • Set up quality controls
• View the chart and data for a selected quality control, or compare the
charts for two or three quality controls
• Print a QC Review report
For more information, see Chapter 6: Quality Control.
Calibration Select to display the Calibration screen. Use this screen to:
F5 • Add, edit, and delete calibrators
• View calibration data
• Print calibration reports
For more information, see Chapter 5: Assay Calibration.
Maintenance Review Select to display the Maintenance Review screen. Use this screen to:
F6 • Review system temperatures
• Decontaminate substrate
• Run, enable, or disable the Utility routine
• Review system check results
For more information, see Chapter 7: Routine Maintenance.
Diagnostics Select to display the Diagnostics menu. Use this menu to:
F7 • Initialize the system
• Prime the fluidic system
• View information about or calibrate various instrument sensors
• Perform diagnostic or troubleshooting procedures as instructed in
system documentation or by Technical Support
For more information, see Chapter 4: Diagnostics in the Reference
Manual.
Configure Select to display the Configure menu. Use this menu to:
F8 • Set up the system
• Set up tests and results
• Display configuration information about the system
• Perform administrative tasks, such as backing up data and shutting
down the system
For more information, see Chapter 2: System Administration and see
Chapter 3: Configuring Tests and Results in the Reference Manual.
Main Menu Select one of the Main Menu function buttons to display an associated screen or
Workflow menu. Across the bottom of a new screen is another row of function buttons. Select
one of these buttons to perform an action or to display a menu with additional
function buttons.
For an illustration of the workflow for each of the Main Menu function buttons, see
Figure 1-10.
F1 In Progress
F1 Views F2 Requested F1 Rack ID
F3 Daily F2 Delete Sample
F4 Exceptions F3 New Request
F4 Test Results
F5 Request QC
F2 Edit Request (Requested and Daily views) F6 Cancel Request
Clear Exceptions (Exceptions view) F7 Clear All Samples
Sample Manager F8 More Options
F1 F1 Patient/QC Requests
F3 New Request F2 Calibration
F1 Rack ID
F3 Maintenance
F3 New Request
F4 Test Results (N/A for
F4 Test Results F1 Filter maintenance requests)
F5 Work Pending F2 Sample Manager F6 Cancel Request
F6 Find F6 Delete
F8 Details F7 Print
F8 Options
F1 Filter
F2 Find
F3 Rerun Test
F4 Sample Details
F5 Send to LIS
Test Results
F6 Copy to Disk F1 Report
F2
F7 Print F2 Completed Samples
F1 Refresh Screen
F8 More Options F2 Configure Screen
F3 Delete Data
F8 Reagent Inventory
Supplies
F3
F1 Reagent Supplies
F1 Bulk Supplies
F2 Unload Reagent Pack
F2 Unload Reagent Packs
F5 Delete Pack
F3 Unload All Reagent Packs
F7 Print
F6 Details
F2 Reagent Supplies F8 Filter
F7 Print
F8 Reagent Inventory
F1 Filter F1 Filter
F2 Review Chart F5 Delete Point
and Data F6 Copy to Disk
F7 Print
Quality Control
F4 F3 Compare Charts F1 Save Comparison F1 Chart 1
F4 Delete Comparison F2 Review Chart and Data F2 Chart 2
F4 Edit Date Range F3 Chart 3
F1 Filter F1 Active
F2 Review Data F1 Select Curve/Limits F2 Previous Active
F2 Switch Active Curve F3 Last Run
F3 Comment
Calibration F7 Print
F5
3003C.wmf
Figure 1-10 Main Menu Workflow, Page 1 of 2
F1 RV Feeder
F1 Initialize System F2 RV Mixer
F2 Prime Fluidics F3 Pick and Places
F3 Subsystem F4 Reagent Storage
Diagnostics F5 Volume Checks
F1 Home All
F1 Mechanics F2 Alignments
F1 Pressure
F2 Wash Buffer
F2 Analog Devices F1 Analog Data F3 Liquid Waste
F2 Sensor Calibration F4 Wash Buffer Flow
F4 Device Diagnostics
F3 Digital Devices F1 Panel Lights
F2 Peristaltic and
F1 Read
Diagnostics
Vacuum Pumps
F2 Load/Read Vessel
F7 F3 Digital Inputs
F3 Save LED Reading
F4 Digital Outputs
F4 Luminometer F4 Save High Voltage
Control
F5 Dark Count Check
F8 Cancel
F1 Ultrasonic Settings
F5 Ultrasonics F2 Ultrasonics Adjustments F1 SPU
F6 Exerciser
F2 LAS
F7 RF Level Sense
3004J.wmf
Figure 1-11 Main Menu Workflow, Page 2 of 2
Common The UniCel DxI system displays certain elements on every screen so you can
Screen continuously monitor the system status. The labels in the example screen are
Elements described in the topics that follow.
2197B.bmp
1 System mode
2 System status buttons
3 Help button
4 Scroll bar (not displayed on all screens)
5 Back tab (not displayed on the Main Menu)
6 Menu tab (not displayed on the Main Menu)
7 Function buttons
8 Status line
9 System command buttons
System Modes
The UniCel DxI system operates in one of four system modes. The current mode is
displayed in the upper left corner of each screen. When the system is in the Running
mode, the estimated completion time for the scheduled tests is displayed as a text line
above the three system command buttons.
Not Ready The system is not ready to process samples. The system
requires initialization, or it is checking the status of
subsystems, initializing motors, or homing movable
parts.
System Status
Description Button Colors
Button
2233A.bmp
System Status
Description Button Colors
Button
2009A.bmp
The system cannot start tests until the
underlying instrument condition has
been resolved. Certain conditions
provide a Help button in the Status
column that link directly to the
corresponding Help procedure topic for
rectifying the condition.
NOTE
The icon color reflects the
condition currently displayed on
the screen with the highest concern
level. If there is at least one
instrument condition that meets a
red icon state, that condition takes
precedence and the Supplies
Required icon displays red.
System Status
Description Button Colors
Button
2010A.bmp
Help Button
Select the Help button to display a topic with information about the screen you are on,
a View Screen link to a picture with descriptions of the screen, and a list of related
topics. From the screen topic, you can navigate to anywhere in the Help system. The
Help button also provides links to procedures for critical instrument conditions
displayed on the Supplies Required screen. For more information, see Chapter 2:
Supplies.
2011A.bmp
Figure 1-15 Help Button
Scroll Bar
The scroll bar is displayed along the right side or bottom of some screens. Use the
scroll bar to view content that is out of the viewing area of the screen. Select the arrow
buttons at either end of the scroll bar to move through the screen incrementally, or
drag the scroll box inside the scroll bar to move through the screen rapidly.
2190A.bmp
Figure 1-16 Scroll Bar
Back Tab
The Back tab is displayed on all screens except the Main Menu. Select the Back tab
to display the previous screen.
0649A.bmp
Figure 1-17 Back Tab
Menu Tab
The Menu tab is displayed on all screens except the Main Menu. Select the Menu tab
to display the Main Menu function buttons.
0650A.bmp
Figure 1-18 Menu Tab
2138A.bmp
Function Buttons
The system displays a row of eight function buttons across the bottom of each screen.
Select a function button or press the corresponding function key to perform an action
or to display a menu of additional function buttons.
Function buttons vary from screen to screen, according to the operations you can
perform on each screen.
0812A.bmp
Information Fields
Three types of information fields may be displayed on screens and windows:
• Data entry fields, in which you can enter or edit pertinent information
NOTE
Do not enter a quotation mark (") in an available field. Although some fields may
accept quotation marks, characters to the right of the quotation mark may not
print on reports.
Data Entry Field When a field has a white background, you can
enter or change information in the field.
0814A.bmp
3005A.bmp
List Field A field with a down arrow on the right has a list
of several options for you to select from. Only
selections from the list can be entered in the field.
2234A.bmp
2. Press [Ctrl+C] (press and hold the [Ctrl] key while you press the C key).
3. Select the field into which you intend to place the text.
4. Press [Ctrl+V] (press and hold the [Ctrl] key while you press the V key).
Status Line
The status line displays the instrument ID, the software version number, the date, and
the time. Depending on the operation being performed, the status line may also
display other information pertinent to that operation.
Because the filters for each screen are different, a separate procedure is provided for
applying each filter.
Common The UniCel DxI system windows provide information, or prompt you to enter
Window information, to perform an action, or to confirm that you want the system to perform
Elements an action. Illustrations of many windows are included in the system documentation.
2209B.bmp
Op
Common Buttons
The OK F1, Done F1, Cancel F8, and Help F12 buttons are common buttons and
appear in many different windows.
Reports You can print reports as well as the displayed screen or window from the UniCel DxI
system software.
NOTE
If information about a sample or patient continues onto another page, the system
repeats the sample or patient ID at the top of that page.
Report header information is set up when you configure your system. For more
information, see Section 2.2: System Setup in the Reference Manual.
4 5
3 Laboratories, Inc.
Laboratory A
6
Report Name
123 Lake Street
Townsville, ST 3333
7
UniCel DxI 800, Access Immunoassay System 101-555-2323 ext. 109 8
S/N 654321, Version 4.4 Jane Smith
1 2
9
10
QNS = sample quantity not sufficient, ORL = result below reference range, ORH = result above reference range,
LEX = reagent pack lot expired
11
13 12
Page 1 of 1
Technologist Printed 02/03/11 02:43 PM
3006B.eps
2 Supplies
2.1 Supply and Supply Conditions Overview . . . . . . . . . . . . . . . . . . . . 2-3
The UniCel DxI system uses reagent and bulk supplies to process samples. Reagent
supplies are the on-board reagent packs. Bulk supplies are the supplies the system
consumes and the waste it generates during processing. There are three consumable
bulk supplies:
• Substrate
• Wash buffer
• Reaction vessels (RVs)
The system uses a liquid waste drain (or optional on-board bulk liquid waste
containers) and a solid waste container to collect the waste generated during sample
processing. To prevent interruptions for needed supplies, monitor the status indicators
and respond by adding or changing supplies promptly. For more information, see the
Monitoring Supply Status topic.
Several status indicators inform you when a bulk supply condition occurs, or if a
reagent pack or calibration are required to continue processing tests. If a supply is
required, the system processes as many tests as possible, but does not schedule
additional tests. You can override expired supply or calibration conditions to continue
processing tests.
For information about ordering supplies, see the Instructions for Use manual.
Monitoring The system displays and continuously updates its supply and operating status during
Supply Status processing. Three different system indicators present the supply information:
• The Amber status indicator light
• The system status panel (UniCel DxI 800 instrument only)
• The Bulk Supplies and Supplies Required system status buttons
• Work Pending
• Supplies Required
• Quality Control
• Event Log
The Console panel icon blinks if the Event Log button turns red.
2033A.bmp
1 RVs icon
2 Substrate icons
3 Console icon
4 Wash Buffer icons
5 Solid Waste icon
6 Liquid Waste icons
When either button changes color, select it to display the Bulk Supplies or Supplies
Required screens for detailed information. For more information, see
Section 1.3: Software Overview.
2026A.bmp
NOTE
If the bar code was not read, or if a bar code error occurs, a question mark (?) is
displayed in the field.
NOTE
If a substrate bottle is disabled, this field will not turn yellow or red until the
bottle is enabled.
Wash Buffer: Bulk Wash Buffer Containers
Status (Icon) Displays the wash buffer level in the two on-board wash buffer positions. As
the level of wash buffer in the in-use container decreases, the level displayed in the
corresponding icon also decreases.
Status
(Field) There are five possible status values:
• In Use - The system is using the container to process tests.
• Present - A container is present, but is not in use.
• Not Present - There is no container in the wash buffer position. The icon for a
not present container is gray on the screen.
• Empty - The container is empty, and must be changed. The icon for an empty
container is white on the screen. For more information, see the Changing an
Empty Bulk Wash Buffer Container in Section 2.3: Wash Buffer.
• Disabled - The container position is disabled and cannot be used. The icon for a
container in a disabled position is overlaid with a circular crossout symbol. For
more information, see Section 2.3: Supplies Setup of the Reference Manual.
Wash Buffer: Estimated % Full
Estimated % Full (Field) Displays a percentage estimate of the available wash buffer in the loaded
bulk wash buffer containers.
• This field turns yellow if one of the on-board containers is low.
• This field turns red if all on-board containers are low. If there is only one
on-board container, the field turns red when the container has 10% of its volume
available for processing.
NOTE
If you cannot change the in-use bottle for any of the three reagents, its button is
not available.
Reagent Inventory (Button) Select to display the Reagent Inventory screen. For more information, see
F8 Section 2.8: Reagent Inventory.
Supplies The Supplies Required screen lists the supplies or calibrations required to complete
Required your test requests. Use this screen to:
Screen • Determine which supplies require immediate attention
• Determine which assays require calibration
• Determine if an instrument condition requires attention, such as a disabled
pipettor
• Link to information in the Help system related to instrument conditions that
require attention
• Override certain supply and calibration conditions and continue processing
• Print the Supplies Required Report
2034B.bmp
Racks to Process (Field) Displays the total number of racks in the SPU that are waiting to be
processed. If the SPU door is open, the message SPU Open is displayed.
Required (Column) Displays the action or quantity required to relieve the supply, instrument,
or calibration condition. If there are no racks in the SPU that are ready to process,
the message No racks to process is displayed in the first row of the Supplies
Required table.
Supply (Column) Displays the name of the supply required and the corresponding icon. If
the supply required is curve-related (no curve, or an expired curve), the name of the
assay that requires calibration is displayed if it is different than the reagent pack
name. For example, if fTSH2 is ordered, the fTSH2 assay is displayed rather than
its associated pack name TSH.
Lot Number (Column) Displays the reagent pack lot number only when a reagent pack is
expired or needs a calibration. Otherwise, the entry for this column is blank.
Status (Column) Displays the calibration or expiration status of the supply. For critical
instrument check conditions, a Help button is displayed that links directly to the
procedure for rectifying the condition.
Otherwise, the entry for this column is blank.
Reagent Supplies (Button) Select to display the Reagent Supplies screen. For more information, see
F1 Section 2.7: Reagent Supplies.
Reagent Inventory (Button) Select to display the Reagent Inventory screen. For more information, see
F2 Section 2.8: Reagent Inventory.
Print (Button) Select to print the Supplies Required Report. For more information, see
F7 the Printing the Supplies Required Report procedure.
Override (Button) Select to continue sample processing with all of the expired supplies and
F8 expired or missing calibrations displayed on this screen. This button is only
available when you can override a supply or calibration condition. For more
information, see the Overriding Supply Conditions procedure.
Overriding You can start sample processing using expired calibrations, substrate, reagent packs,
Supply or reagent lots, and diluent packs or lots by overriding the current conditions. The
Conditions system applies appropriate flags to the affected test results and logs a corresponding
event in the Event Log.
NOTES
• You can use this procedure to run tests with a reagent pack that expires while
it is onboard the system. However, if a pack expires while it is offboard the
system, performing this procedure will not allow you to use the pack, even if
another pack from the same lot is onboard the system when the expiration
occurs.
• If you override a reagent pack that is not calibrated, the sample is processed
but the results are not reported.
• If you are recalibrating an expired calibration and running STAT tests for the
assay at the same time, override the expired calibration. The system
recalibrates the assay and then runs the STAT tests. Until you select
Override F8, the STAT test requests stay in the Supply Wait status.
Example
You may have an expired Free T4 reagent pack and a Cortisol pack with an
expired calibration, and you only intend to override the Cortisol calibration.
Unload the expired Free T4 pack and load an unexpired Free T4 pack. Then
return to the Supplies Required screen and perform the override procedure.
Cortisol will be removed from the list on the Supplies Required screen and the
subsequent Cortisol test results will be flagged.
Resolving Use this procedure as general guidance for correcting conditions displayed on the
Supplies Supplies Required screen.
Required The most critical instrument check conditions are listed toward the top of the table
Conditions within the Supplies Required screen, followed by Bulk Supplies Required, and then
Reagent Supplies Required.
Printing the The Supplies Required report is a printout of the entries on the Supplies Required
Supplies screen. To print a copy of the report, select Print F7 on the Supplies Required screen.
Required System Mode: Any Mode
Report
Laboratories, Inc.
Supplies Required Report Laboratory A
UniCel DxI 800, Access Immunoassay System 123 Lake Street
Townsville, ST 33333
S/N 654321, Version 4.4 101-555-2323 ext. 109 Jane Smith
Racks to be Scanned: 0
Page 1 of 1
Technologist ____________________ Printed 02/08/11 04:03 PM
2052D.bmp
Figure 2-5 Supplies Required Report Example
2.2 Substrate
The UniCel DxI system uses substrate to measure the amount of analyte in a
processed test. For more information, see Section 1.2: Assay Technology in the
Reference Manual.
A substrate bottle is stored refrigerated to maintain its stability. Prior to use, a bottle is
equilibrated to room temperature in the substrate equilibration area located
underneath the reagent load door. The equilibration area of the UniCel DxI 800
instrument is partitioned, and holds up to four bottles of substrate (see Figure 2-10).
The equilibration area of the UniCel DxI 600 instrument holds up to two bottles (see
Figure 2-11). An unopened substrate bottle can remain at room temperature for 14
days.
When the bottle has equilibrated, you can load it on the system. You should have two
bottles on board the instrument at all times. When one bottle is empty, the system
automatically switches to the second bottle, and you can change the empty one. In
some circumstances, it may be necessary to change the in-use substrate bottle. You
can change the in-use bottle through the user interface on the Bulk Supplies screen.
Once you change an in-use substrate bottle it must be discarded.
All substrate bottles are labeled with a bar code. The system uses the information on
the bar code label to track lot number, serial number, expiration date, and the number
of tests remaining in the bottle.
You can use the following system indicators to monitor your on-board substrate
supply levels:
• The Amber status indicator light
• The system status panel (UniCel DxI 800 instrument only)
• The Bulk Supplies button
• The Bulk Supplies screen
Monitoring The following table summarizes the different methods available to monitor substrate
the Substrate supply conditions.
Supply • Amber status indicator light (see Figure 1-3 in Section 1.2: Instrument
Overview)
• Substrate panel icon (UniCel DxI 800 instrument only - see Figure 2-2 in
Section 2.1: Supply and Supply Conditions Overview)
• Bulk Supplies status button (see Figure 1-12 in Section 1.3: Software
Overview)
• Substrate Status and Tests Left fields on the Bulk Supplies screen (see
Figure 2-3 in Section 2.1: Supply and Supply Conditions Overview)
Changing an If there is a sufficient substrate supply, the system continues to process tests while you
Empty or change the empty or expired bottle. To change a bottle before it is empty or expired,
Expired see the Changing an In-Use Substrate Bottle procedure.
Check the Substrate panel icons (see Consult the Bulk Supplies screen to
Figure 2-2 in Section 2.1: Supply and determine whether a lot is near
Supply Conditions Overview). expiration or if a bottle is empty. For
Blinking: Change all empty or expired
more information, see
bottles. Section 2.1: Supply and Supply
Conditions Overview.
Solid: Consult the Bulk Supplies screen
to determine whether a lot is near
expiration or if a bottle is empty. For
more information, see
Section 2.1: Supply and Supply
Conditions Overview.
• If the substrate lot is nearing expiration, change the bottle within three days.
• If a bottle is empty, change the bottle that is not in use. If you cannot determine
the in-use bottle, consult the Bulk Supplies screen.
• If there is only one on-board bottle, load two new bottles.
CAUTIONS
• In the following step, do not force the substrate door open if it does not
raise automatically. Consult the system status panel (UniCel DxI 800
instrument only) or Bulk Supplies screen to confirm which bottle to
change.
• When the substrate door is raised the system drains substrate from the
lines back into the substrate bottle. This contaminates the remaining
volume of substrate. If you raise the door to the in-use substrate bottle
unintentionally, you cannot reload the in-use bottle. Always discard the
in-use bottle and load a fresh one.
2. Do not grasp the substrate load door handle. Instead, place a finger underneath the
handle and push up against the pressure until the door lifts approximately
1/4 inch. At this height, the system automatically raises the door.
3. Push inward on the front of the substrate load tray, then release quickly. The
system releases the tray approximately one inch. Grasp the sides of the tray and
pull out gently.
4. Lift the empty or expired bottle out of the tray and discard it.
4
5
2049A.eps
1 Substrate bottle
2 Substrate load door handle, raised
3 Substrate load door, closed
4 Substrate load door handle, closed
5 Substrate load tray
3
2
2049B.eps
1 Substrate bottle
2 Substrate load tray
3 Substrate load door handle, closed
4 Substrate load door, closed
5 Substrate load door handle, raised
5. Lift up on the reagent load door (see Figure 2-35 in Section 2.7: Reagent
Supplies). If the reagent tray in-use light is green, wait until the light turns off
before opening the door.
4
3
2
1
2050A.eps
Figure 2-10 UniCel DxI 800 Instrument Substrate Bottle Equilibration Area
4
3
2
1
2050B.eps
Figure 2-11 UniCel DxI 600 Instrument Substrate Bottle Equilibration Area
8. Remove the cap from the new bottle and discard it.
9. Place the new bottle in the load tray with the bar code label facing out.
11. Keep your fingers away from the open door. The system automatically lowers the
door after a three second delay.
12. Scan the bar code label through the window in the substrate load door. For more
information, see the Scanning the Substrate Bar Code Label procedure.
13. Remove a new, unequilibrated substrate bottle from storage and write the current
date and time on the label.
14. Lift up on the reagent load door. If the reagent tray in-use light is green, wait until
the light turns off before opening the door.
Place an unequilibrated
bottle in the substrate
equilibration area
2017C.wmf
Figure 2-12 Changing the Substrate Bottle Flowchart
Scanning the Use this procedure to scan the substrate bar code label with the substrate bar code
Substrate Bar reader.
Code Label
NOTE
You have 30 seconds to scan the substrate bar code label after the system closes
the substrate load door. If you do not scan the bar code within 30 seconds, a
substrate timeout error occurs. The substrate load door opens automatically and a
message is sent to the Event Log. For more information, see the Resolving
Substrate Bar Code Errors procedure.
2. Remove the substrate bar code reader from its holder in front of the reagent load
area. For the location of the bar code reader, see Figure 1-1 (UniCel DxI 800
instrument) or see Figure 1-2 (UniCel DxI 600 instrument) in Section 1.2:
Instrument Overview.
3. Rotate the bar code reader approximately 90 degrees until the trigger is horizontal
to the floor.
4. Aim the bar code reader at the bar code label that is visible through the substrate
load door. Hold the reader approximately 1 to 8 inches (2.5 to 20.3 cm) from the
label.
5. Squeeze and hold the trigger, center the red light (indicating the scanning area)
within the lines of the bar code, and then release the trigger.
The bar code reader beeps if the bar code is read successfully and updates the
substrate information on the Bulk Supplies screen. The Substrate panel icon returns
to its unlit state in approximately five seconds.
NOTE
If the system reads the bar code successfully but does not accept it, a substrate bar
code scan error occurs. For more information, see the Resolving Substrate Bar
Code Errors procedure.
Resolving Two substrate bar code errors can occur while you change a substrate bottle.
Substrate Bar • Timeout errors for bar codes not successfully scanned within 30 seconds of
Code Errors closing the substrate load door.
• Scan errors for bar codes that the system cannot read.
2200A.bmp
Open substrate door (Option) Select to open the substrate load door. If a
bar code timeout error has occurred, this option is not
available.
Rescan bottle (Option) Select to rescan the substrate bar code label.
Enter bar code (Option) Select to manually enter the substrate bar
code using the keyboard.
Figure 2-13 Substrate Bar Code Error (Scan View) Window Descriptions
• If a timeout error occurred because you did not scan a visible bar code label
within 30 seconds, select the Rescan Bottle option and scan the bottle again.
• If a bar code scan error occurred, select the Enter Bar Code option and type the
bar code from the keyboard.
NOTE
If you have not returned the substrate bar code reader to its holder, do so now.
Changing an You cannot change the in-use substrate bottle if the second bottle is empty or was not
In-Use loaded, if the second position is disabled, or if the bar code was not scanned.
Substrate
NOTE
Bottle
When you change an in-use substrate bottle, the system cancels all of the tests
remaining in the bottle. The bottle cannot be used again and must be discarded.
4. The system changes the in-use bottle and updates the Status field on the Bulk
Supplies screen.
CAUTION
To avoid contaminating the substrate supply, always discard partial bottles
of substrate. Do not combine partial bottles or try to load a partial bottle on
another instrument.
5. Change and discard the old bottle. For more information, see the Changing an
Empty or Expired Substrate Bottle procedure.
WARNING
Wash buffer contains a preservative which may cause sensitization by skin
contact. After contact with skin, wash immediately with soap and water. Wear
suitable gloves.
The wash buffer supply drawer is on the lower right as you face the instrument. For
the location of the wash buffer supply drawer, see Figure 1-1 (UniCel DxI 800
instrument) or see Figure 1-2 (UniCel DxI 600 instrument) in Section 1.2: Instrument
Overview. The drawer holds two 10-liter bulk wash buffer containers, which is
enough for at least three hours of continuous processing.
You should have two containers on board the instrument at all times. The system
draws wash buffer from one container. When the container is empty, the system
automatically switches to the second, and you can change the empty one. In some
circumstances, it may be necessary to change the in-use wash buffer container. You
can change the in-use container through the user interface on the Bulk Supplies
screen.
You can use the following system indicators to monitor your on-board wash buffer
supply levels:
• The Amber status indicator light
• The system status panel (UniCel DxI 800 instrument only)
• The Bulk Supplies button
• The Bulk Supplies screen
Monitoring The following table summarizes the different methods available to monitor wash
the Wash buffer supply conditions.
Buffer Supply • Amber status indicator light (see Figure 1-3 in Section 1.2: Instrument
Overview)
• Wash Buffer panel icons (UniCel DxI 800 instrument only - see Figure 2-2
in Section 2.1: Supply and Supply Conditions Overview)
• Bulk Supplies status button (see Figure 1-12 in Section 1.3: Software
Overview)
• Wash Buffer Status and Estimated % Full fields on the Bulk Supplies
screen (see Figure 2-3 in Section 2.1: Supply and Supply Conditions
Overview)
Changing an If there is a sufficient wash buffer supply, the system continues to process tests while
Empty Bulk you change the empty bulk wash buffer container. To change a wash buffer container
Wash Buffer before it is empty, see the Changing an In-Use Bulk Wash Buffer Container
procedure.
Container
Use this procedure to change an empty bulk wash buffer container.
WARNING
Wash buffer contains a preservative which may cause sensitization by skin
contact. After contact with skin, wash immediately with soap and water. Wear
suitable gloves.
Check the Wash Buffer panel icons Consult the Bulk Supplies screen (see
(see Figure 2-2 in Section 2.1: Supply Section 2.1: Supply and Supply
and Supply Conditions Overview). Conditions Overview).
Blinking: Change all low wash buffer If the Wash Buffer Estimated % Full
containers. If there is only one on-board field is red, change all low wash buffer
container, load two new containers. containers. If there is only one on-board
Solid: Change the wash buffer
container, load two new containers.
container that is not in use. If the Wash Buffer Estimated % Full
• If you cannot determine the in-use field is yellow, change the wash buffer
container, consult the Bulk Supplies container that is not in use.
screen. For more information, see
Section 2.1: Supply and Supply
Conditions Overview.
2. Pull out the wash buffer supply drawer. For the location of the wash buffer supply
drawer, see Figure 1-1 (UniCel DxI 800 instrument) or see Figure 1-2
(UniCel DxI 600 instrument) in Section 1.2: Instrument Overview.
3. Remove the perforated cardboard panels at the top and the sides of a new wash
buffer container.
4. Gently mix the new wash buffer container.
5. Place the new wash buffer container on the floor adjacent to the open supply
drawer and the container you are replacing.
6. Grasp the cap on the new container and pull upward until the plastic neck is
completely extended.
7. Remove the cap and inner seal from the new container.
8. Determine if the wash buffer container is fitted with the original or redesigned
assembly (see Figure 2-16).
2
1
2
5
3
3
2378A.eps 2376A.eps
9. If the wash buffer container is fitted with an original assembly (includes a tubing
disconnect button), continue with step 10 of this procedure.
If the wash buffer container is fitted with a redesigned assembly (no disconnect
button), perform the Changing an Empty Wash Buffer Container with a Redesigned
Cap and Tube Assembly procedure which immediately follows this procedure.
10. Press the disconnect button to release the fitting and tubing from the cap/draw
tube assembly on the empty container. If the fitting does not release, pull up
gently on the fitting while pressing the disconnect button.
11. Lift the empty container from the supply drawer and replace it with the new
container.
CAUTION
To avoid contaminating the wash buffer supply, do not touch the draw tube
in the following steps. Handle the cap/draw tube assembly by the cap only.
12. Unscrew the cap/draw tube assembly on the empty container and carefully
remove it from the container. Be sure to lift the draw tube completely out of the
container before moving it (see Figure 2-17).
1
4
3
T
H
IS
E
N
D
U
P
F
R
T
O
H
N
IS
T
E
N
D
U
P
5
FR
ON
T
2056A.eps
1 Draw tube
2 Disconnect button
3 Bulk wash buffer containers
4 Cap/draw tube assembly
5 Cap fitting and tubing
Figure 2-17 Changing the Bulk Wash Buffer Container (Original Cap/Draw
Tube Assembly With a Disconnect Button)
13. Lower the cap/draw tube assembly into the new container and tighten the cap.
14. Connect the fitting into the cap/draw tube assembly on the new container.
15. Carefully push in the drawer. Be sure that none of the wash buffer tubing is
pinched between the drawer and the wall.
NOTE
Do not discard the shim after you remove it from the empty container. You will
need the shim in step 2 of this procedure.
2. Slide the shim into place underneath the extended plastic neck of the new wash
buffer container.
3. Lift the empty container from the wash buffer supply drawer and set it beside the
new container.
CAUTIONS
• To avoid contaminating the wash buffer supply, do not touch the wash
buffer supply or sensor tubes in the following steps. Handle the cap and
tube assembly by the cap only. Do not set the cap and tube assembly on
any surface after you remove it from the empty container.
• Handle the cap and tube assembly carefully in the following steps.
Pulling on the sensor electrical wires too vigorously or too often can
damage the wires, or disconnect them from the instrument or the
assembly.
4. Unscrew the cap and tube assembly on the empty container and carefully remove
it from the container. Lift the wash buffer supply and sensor tubes completely out
of the container before moving the cap and tube assembly (see Figure 2-18).
2377A.eps
Figure 2-18 Changing the Bulk Wash Buffer Container (Redesigned Cap
and Tube Assembly With No Disconnect Button)
5. Check the condition of the sensor electrical wires, the sensor and wash buffer
supply tubes, and the clips that hold both tubes in place.
a. Be sure that the sensor electrical wires are not stretched too far, and that the
connectors are not broken or disconnected from the instrument or the
assembly.
b. Be sure that the sensor and wash buffer supply tubes are not bent or broken.
c. Be sure that the clips are not missing or broken.
Contact Technical Support if the sensor electrical wires, the sensor tube, or the wash
buffer supply tube are damaged or broken, or if the clips are broken or missing.
6. Lower the cap and tube assembly into the new container and gently tighten the
cap.
7. Set the new wash buffer container in the empty position inside the supply drawer.
Route the tubing alongside the wash buffer container, between the container and
the drawer wall.
8. Carefully push in the drawer. Be sure that none of the wash buffer tubing is
pinched between the drawer and the instrument.
9. Discard the empty container.
Is there a
disconnect on the No
cap and tube
assembly?
2018D.vsd
Figure 2-19 Changing an Empty Bulk Wash Buffer Container Flowchart
Changing an You cannot change the in-use bulk wash buffer container if the second wash buffer
In-Use Bulk container is empty or was not loaded, or if the second position is disabled.
Wash Buffer Use this procedure to change the in-use bulk wash buffer container.
Container
System Mode: Any Mode
1. From the Bulk Supplies screen, select Change In-Use Bottle F3.
3. The system changes the in-use wash buffer container and updates the Status field
on the Bulk Supplies screen.
WARNING
Wash buffer contains a preservative which may cause sensitization by skin
contact. After contact with skin, wash immediately with soap and water. Wear
suitable gloves.
4. Change the old container. For more information, see the Changing an Empty Bulk
Wash Buffer Container procedure.
The vessel hopper is located above the sample presentation unit (SPU) on the left side
of the instrument. For the location of the vessel hopper, see Figure 1-1
(UniCel DxI 800 instrument) or see Figure 1-2 (UniCel DxI 600 instrument) in
Section 1.2: Instrument Overview. The hopper holds up to two bags of approximately
1,000 RVs each. When the hopper is full, there are enough RVs for at least three hours
of continuous processing.
You can use the following system indicators to monitor your on-board RV supply
level:
• The Amber status indicator light
• The system status panel (UniCel DxI 800 instrument only)
• The Bulk Supplies button
• The Bulk Supplies screen
Monitoring The following table summarizes the different methods available to monitor RV supply
the RV Supply conditions.
• Amber status indicator light (see Figure 1-3 in Section 1.2: Instrument
Overview)
• RVs panel icon (UniCel DxI 800 instrument only - see Figure 2-2 in Section
2.1: Supply and Supply Conditions Overview)
• Bulk Supplies status button (see Figure 1-12 in Section 1.3: Software
Overview)
• Estimated RVs Left field on the Bulk Supplies screen (see Figure 2-3 in
Section 2.1: Supply and Supply Conditions Overview)
Adding RVs Use this procedure to add RVs to the vessel hopper.
WARNING
To avoid contaminating the RV supply, add full, previously unopened RV bags
to the hopper whenever it is possible. If you add a previously opened bag of RVs
to the hopper, be sure the bag was closed to protect the RVs from dust and other
contaminants.
Check the RVs panel icons (see Consult the Bulk Supplies screen to
Figure 2-2 in Section 2.1: Supply and determine how many RVs to add (see
Supply Conditions Overview). Section 2.1: Supply and Supply
Blinking: Add enough RVs to fill the
Conditions Overview).
vessel hopper (nearly two bags). If there are fewer than 100 RVs, add
Solid: Add one bag of RVs.
enough RVs to fill the vessel hopper
(nearly two bags).
If there are between 100 and 1000 RVs,
add one bag of RVs.
2088A.eps
3. Carefully open a bag of RVs by breaking the seal just underneath the top of the
bag.
CAUTION
In the following step, only add UniCel DxI system RVs to the hopper. If you
add other RVs or containers, the instrument will malfunction.
4. Carefully pour the RVs into the hopper using a side-to-side motion. This prevents
the RVs from accumulating in only one area of the hopper, and keeps the RV
level relatively even.
2019C.wmf
Figure 2-23 Adding RVs Flowchart
The UniCel DxI system uses a liquid waste drain or two on-board bulk liquid waste
containers to manage liquid waste. Liquid waste is generated when:
• The pipettors are washed between processing steps
• Unreactive fluids are washed from the RVs
WARNINGS
• Consider the liquid waste to be potentially infectious. Handle and dispose of
liquid waste according to proper laboratory procedures. Proper hand, eye,
and facial protection is required.
• Reagents, calibrators, and controls used with the system may contain small
quantities of sodium azide preservative. Sodium azide preservative may
form explosive compounds in metal drain lines. Refer to National Institute
for Occupational Safety and Health Bulletin: Explosive Azide Hazards
(8/18/76).
If your system uses on-board bulk liquid waste containers, they are housed in the
liquid waste drawer on the lower left as you face the instrument. For the location of
the liquid waste drawer, see Figure 1-1 (UniCel DxI 800 instrument) or see Figure 1-2
(UniCel DxI 600 instrument) in Section 1.2: Instrument Overview. The drawer on the
UniCel DxI 800 instrument holds two 9.5-liter reusable bulk liquid waste containers.
The drawer on the UniCel DxI 600 instrument holds two 5-liter reusable bulk liquid
waste containers. For either instrument, the liquid waste containers collect the liquid
waste generated during three hours of continuous processing. The system dispenses
liquid waste into one container at a time. When the container is full, the system
automatically switches to the second, and you can change the full one. In some
circumstances, it may be necessary to change the in-use liquid waste container. You
can change the in-use container through the user interface on the Bulk Supplies
screen.
You can use the following system indicators to monitor your on-board liquid waste
levels:
• The Amber status indicator light
• The system status panel (UniCel DxI 800 instrument only)
• The Bulk Supplies button
• The Bulk Supplies screen
Monitoring The following table summarizes the different methods available to monitor liquid
the Liquid waste supply conditions if your system uses on-board bulk liquid waste containers.
Waste Supply • Amber status indicator light (see Figure 1-3 in Section 1.2: Instrument
Overview)
• Liquid Waste panel icons (UniCel DxI 800 instrument only - see Figure 2-2
in Section 2.1: Supply and Supply Conditions Overview)
• Bulk Supplies status button (see Figure 1-12 in Section 1.3: Software
Overview)
• Liquid Waste Status and Estimated % Empty fields on the Bulk Supplies
screen (see Figure 2-3 in Section 2.1: Supply and Supply Conditions
Overview)
• The Amber status indicator light • All on-board containers are almost
blinks full
• Both of the Liquid Waste panel • There is only one on-board container,
icons blink (UniCel DxI 800 and it has 9% or less of its volume
instrument only) available for collecting waste
• The Bulk Supplies button turns red • No containers are present
• On the Bulk Supplies screen, the
Liquid Waste Estimated % Empty
field turns red
Changing a Use this procedure only if your DxI system uses on-board bulk liquid waste
Full Bulk containers.
Liquid Waste If there is sufficient volume available for collecting waste, the system continues to
Container process tests while you change a bulk liquid waste container. Use this procedure to
change a full bulk liquid waste container. To change a liquid waste container before it
is full, see the Changing an In-Use Bulk Liquid Waste Container procedure.
WARNINGS
• You will come in contact with potentially infectious materials during this
procedure. Handle and dispose of biohazard materials according to proper
laboratory procedures. Proper hand, eye, and facial protection is required.
• Thoroughly flush any decontaminant from the liquid waste container before
you store the container. If you reuse a container without flushing it, any
remaining decontaminants may react with chemicals dispensed into the
waste container. The resulting chemical reactions may create gases harmful
to you or the instrument.
Check the Liquid Waste panel icons Consult the Bulk Supplies screen (see
(see Figure 2-2 in Section 2.1: Supply Section 2.1: Supply and Supply
and Supply Conditions Overview). Conditions Overview).
Blinking: Change all liquid waste If the Liquid Waste Estimated %
containers. If there is only one on-board Empty field is red, change all liquid
container, load two new containers. waste containers. If there is only one
Solid: Change the liquid waste
on-board container, load two new
container that is not in use. containers.
2. Pull out the liquid waste drawer. For the location of the liquid waste drawer, see
Figure 1-1 (UniCel DxI 800 instrument) or see Figure 1-2 (UniCel DxI 600
instrument) in Section 1.2: Instrument Overview.
3. Place an empty liquid waste container on the floor adjacent to the open liquid
waste drawer and the container you are replacing.
4. Press the disconnect button to release the fitting and tubing from the cap
assembly. If the fitting does not release, pull up gently on the fitting while
pressing the disconnect button.
5. Lift the full container from the drawer and replace it with the empty container.
6. Connect the fitting into the cap assembly on the empty container.
7. Carefully push in the drawer. Be sure that none of the liquid waste tubing is
pinched between the drawer and the wall.
8. Decontaminate the contents of the full bulk liquid waste container according to
proper laboratory procedures before disposing of the liquid waste.
10. Rinse the just-emptied container thoroughly by filling it with tap water and
discarding the contents to remove any traces of the decontaminant.
WARNING
Thoroughly flush any decontaminant from the liquid waste container before you
store the container. If you reuse the container without flushing it, any remaining
decontaminants may react with chemicals dispensed into the waste container.
The resulting chemical reactions may create gases harmful to you or the
instrument.
2
3
1 4
2087A.eps
DxI 800
DxI 600
Check the Liquid Waste
Check the Bulk Supplies
panel icons for container
screen for container status
status
2020C.wmf
Figure 2-27 Changing a Full Bulk Liquid Waste Container Flowchart
Changing an Use this procedure only if your DxI system uses on-board bulk liquid waste
In-Use Bulk containers.
Liquid Waste If there is sufficient volume available for collecting waste, the system continues to
Container process tests while you change a bulk liquid waste container. Use this procedure to
change a bulk liquid waste container before it is full. To change a full liquid waste
container, see the Changing a Full Bulk Liquid Waste Container procedure.
You cannot change the in-use bulk liquid waste container if the second liquid waste
container is full or was not loaded, or if the second position is disabled.
3. The system changes the in-use liquid waste container and updates the Status field
on the Bulk Supplies screen.
4. You can now change the old container. For more information, see the Changing a
Full Bulk Liquid Waste Container procedure.
The solid waste container collects the empty reagent packs, used RVs, and
condensation generated by the UniCel DxI system during processing.
WARNING
Consider the solid waste to be potentially infectious. Handle and dispose of solid
waste according to proper laboratory procedures. Proper hand, eye, and facial
protection is required.
The solid waste container is a large plastic bin lined with a biohazardous waste bag.
There is only one container for solid waste. The container is stored behind the
pull-down door located between the liquid waste and wash buffer supply drawers. For
the location of the solid waste door, see Figure 1-1 (UniCel DxI 800 instrument) or
see Figure 1-2 (UniCel DxI 600 instrument) in Section 1.2: Instrument Overview.
When the container is empty, it can hold the solid waste generated during three hours
of continuous processing. When the container is full, the system continues processing
tests but does not process any new test requests.
CAUTION
Do not overfill the solid waste container. Excess solid waste may obstruct the
waste chute.
You can use the following system indicators to monitor your on-board solid waste
level:
• The Amber status indicator light
• The system status panel (UniCel DxI 800 instrument only)
• The Bulk Supplies button
• The Bulk Supplies screen
Monitoring The following table summarizes the different methods available to monitor solid
the Solid waste supply conditions.
Waste Supply • Amber status indicator light (see Figure 1-3 in Section 1.2: Instrument
Overview)
• Solid Waste panel icon (UniCel DxI 800 instrument only - see Figure 2-2 in
Section 2.1: Supply and Supply Conditions Overview)
• Bulk Supplies status button (see Figure 1-12 in Section 1.3: Software
Overview)
• Solid Waste Estimated % Empty field on the Bulk Supplies screen (see
Figure 2-3 in Section 2.1: Supply and Supply Conditions Overview)
Changing a The system continues processing tests while you change the full solid waste container,
Full Solid but does not initiate any new test requests. The system does not dispose of any solid
Waste waste until the empty container is in place again.
Container Use this procedure to change the solid waste container bag.
WARNING
You will come in contact with potentially infectious materials during this
procedure. Handle and dispose of biohazard materials according to proper
laboratory procedures. Proper hand, eye, and facial protection is required.
CAUTIONS
• If you change a full solid waste container while the system is in the
Running or Paused mode, be sure to complete the procedure within five
minutes. Solid waste may accumulate and obstruct the waste chute if it
takes longer than five minutes to change the container.
• Do not open the solid waste door while the system is initializing, running
a Daily Clean System routine, or running a Special Clean routine.
Opening the solid waste door will cancel the routine and the system will
enter the Not Ready mode. For more information, see the Initializing the
System procedure in Section 4.2: Initialize System of the Reference
Manual, or see Running Clean Routines in Section 7.2: Daily
Maintenance.
System Mode: Any Mode
1. Check the solid waste status.
Check the Solid Waste panel icon (see Consult the Bulk Supplies screen (see
Figure 2-2 in Section 2.1: Supply and Section 2.1: Supply and Supply
Supply Conditions Overview). Conditions Overview).
Blinking: Change the solid waste If the Solid Waste Estimated % Empty
container. field is red, change the solid waste
Solid: The solid waste container is
container.
almost full, change the container soon. If the Solid Waste Estimated % Empty
field is yellow, the solid waste container
is almost full, change the container
soon.
2. Open the solid waste door by pulling down on the handle. For the location of the
wash solid waste door, see Figure 1-1 (UniCel DxI 800 instrument) or see
Figure 1-2 (UniCel DxI 600 instrument) in Section 1.2: Instrument Overview.
2089A.eps
4. Seal the biohazard bag lining the container by tying two ends of the bag together
in a knot.
6. Line the container with a new biohazard bag. Expand the bag so it reaches the
sides of the container.
8. Press the green reset button mounted on the left side of the storage area.
NOTE
If you do not press the reset button, the system will not reset the system
indicators. If you changed a full container, the system will not process any new
test requests until you press the reset button.
2021C.wmf
Figure 2-31 Changing a Full Solid Waste Container Flowchart
The reagent supplies are on-board reagent packs. The UniCel DxI system uses two
types of reagent packs:
• Assay reagent packs
• Diluent packs
WARNING
Consider the assay-specific reagents inside the reagent packs to be potentially
infectious. Handle and dispose of reagent packs according to proper laboratory
procedures. Proper hand, eye, and facial protection is required.
You can monitor reagent supplies on the Reagent Supplies screen. The screen
contains 50 buttons that correspond to the 50 slots in the reagent storage chamber. An
empty button indicates an available slot in the storage chamber. The buttons are sorted
alphabetically down a column, and then numerically left to right.
During sample processing, the system monitors reagent dispense volume to assure
accurate delivery of reagents. If a reagent dispense error occurs, the system records an
event with troubleshooting information in the Event Log.
If three or more reagent dispense errors occur for a single reagent pack, the system
records an event in the Event Log and the reagent pack cannot be used. For this
reagent pack, any test requests which are not already in process are assigned to
another pack. If no other pack is available, the test requests are assigned to
Work Pending. For related troubleshooting information, see Section 5.4:
Event-Specific Troubleshooting in the Reference Manual.
The reagent loading area is located to the right of the sample presentation unit. For the
location of the reagent loading area, see Figure 1-1 (UniCel DxI 800 instrument) or
see Figure 1-2 (UniCel DxI 600 instrument) in Section 1.2: Instrument Overview.
You can load up to four packs at a time on the loading tray of the UniCel DxI 800
instrument (up to two packs at a time on the UniCel DxI 600 instrument). When you
shut the reagent load door, the instrument scans the reagent bar code and moves the
packs either to the reagent pipetting station or the reagent storage chamber. The
system moves or stores packs as they are needed to process tests.
You can unload up to four packs at a time on the Reagent Supplies screen for the
UniCel DxI 800 instrument (up to two packs at a time on the UniCel DxI 600
instrument), or unload them one at a time on the Reagent Inventory screen. For more
information, see Section 2.8: Reagent Inventory.
When a pack does not contain enough reagent to process the requested tests, or if the
pack lot or open pack expiration has expired, the system assigns those tests the
Supply Wait status. For more information, see Section 2.1: Supply and Supply
Conditions Overview.
You can use the following system indicators to monitor your on-board reagent supply
levels:
• The Amber status indicator light
• The system status panel (UniCel DxI 800 instrument only)
• The Supplies Required button
• The Supplies Required screen
Monitoring The following table summarizes the different methods available to monitor reagent
the Reagent pack supply conditions. You can monitor supply conditions for individual packs on
Pack Supply the Reagent Supplies screen.
• Amber status indicator light (see Figure 1-3 in Section 1.2: Instrument
Overview)
• Console panel icon (UniCel DxI 800 instrument only - see Figure 2-2 in
Section 2.1: Supply and Supply Conditions Overview)
• Supplies Required status button (see Figure 1-12 in Section 1.3: Software
Overview)
• Supplies Required screen (see Figure 2-4 in Section 2.1: Supply and Supply
Conditions Overview)
To get to this screen from the Bulk Supplies screen, select Reagent Supplies F1.
2028C.bmp
Reagent Pack For each on-board reagent pack, the button displays the following information:
Button Test Name
(Button label) Displays the reagent inventory pack name. In the case of primary
and secondary tests, such as Folate and its secondary test RBC Folate, the primary
test name is displayed.
A reagent pack button with a red background and a white test name cannot be used.
For this reagent pack, any test requests which are not already in process are
assigned to another pack. If no other pack is available, the test requests are
assigned to Work Pending. See the Event Log for related troubleshooting
information.
A diluent pack button has a blue inventory pack name which begins with the letter
"d". This pack contains diluent and is used for onboard dilution tests only.
Lot
(Field) Displays the reagent pack lot number.
• This field turns yellow if the pack lot number or calibration expires in three days
or sooner.
• This field turns red if the pack lot number or calibration is expired, there is no
calibration, or if the pack bar code was not read.
NOTE
If you disable the test for a reagent pack, no expiration information is
provided. For more information, see Section 3.3: Test Setup in the Reference
Manual.
If the bar code was not read, a question mark (?) displays in the field.
Qty (for assay reagent packs only)
(Field) Displays the number of tests remaining in the reagent pack. The number
changes as the system uses the reagent.
• This field turns yellow if there are 10 tests or fewer remaining for all on board
packs with the same lot number, or if the open pack stability expires in three
days or sooner.
• This field turns red if there are no tests remaining in the pack, or if the open pack
stability has expired.
Button, continued (Padlock) If the system is processing samples and the reagent pack is
scheduled for use, the system displays a padlock icon and disables the
Unload Reagent Packs F2 button to prevent you from removing the
pack. You cannot remove in-use packs until the system aspirates the reagent for the
current run.
The system updates the Reagent Pack buttons as you load and unload packs.
Bulk Supplies (Button) Select to display the Bulk Supplies screen. For more information, see the
F1 Bulk Supplies Screen topic in Section 2.1: Supply and Supply Conditions
Overview.
Unload Reagent (Button) Select to unload individual reagent packs. You can unload up to four
Packs reagent packs at a time on the UniCel DxI 800 instrument (up to two packs at a
F2 time on the UniCel DxI 600 instrument). For more information, see the Unloading
a Reagent Pack procedure.
Unload All Reagent (Button) Select to unload all of the reagent packs from the reagent storage
Packs chamber. For more information, see the Unloading All Reagent Packs procedure.
F3
Details (Button) After selecting a Reagent Pack button, select to display more detailed
F6 information about the reagent pack. This information includes:
• Lot number
• Serial number
• Expiration date
• For a diluent pack, the number of dilutions remaining for each test.
For more information, see the Displaying and Printing Reagent Pack Details
procedure.
Print (Button) Select to print the Reagent Supplies Report for all on-board reagent packs.
F7 For more information, see the Displaying and Printing Reagent Pack Details
procedure.
Reagent Inventory (Button) Select to display the Reagent Inventory screen. For more information, see
F8 Section 2.8: Reagent Inventory.
Displaying The Reagent Supplies screen displays summary information about each on-board
and Printing reagent pack. The Reagent Supplies Report provides summary information about all
Reagent Pack on-board reagent packs. This report does not print information for individual reagent
packs.
Details
Use this procedure to display detailed information for one reagent pack, or to print the
Reagent Supplies Report.
5. To print the reagent pack details listed on the Reagent Supplies Report, select
Print F7.
Laboratories, Inc.
Reagent Supplies Report Laboratory A
UniCel DxI 800, Access Immunoassay System 123 Lake Street
Townsville, ST 33333
S/N 654321 101-555-2323 ext. 109 Jane Smith
Test Name Lot No. Serial No. Tests Days Pack Calib. Slot Instrument
Left Open Expiration Expiration
CK-MB 013055 231 2 8 04/27/04 10/15/03 1 654321
CK-MB 013055 232 2 8 04/27/04 10/15/03 2 654321
CK-MB 013040 234 2 8 04/27/04 None 3 654321
CK-MB 013055 235 25 6 02/28/04 10/15/03 4 654321
CK-MB 013055 236 25 6 02/28/04 10/15/03 5 654321
CK-MB 013040 481 25 4 02/28/04 None 6 654321
CK-MB 013055 483 50 0 01/31/04 10/15/03 7 654321
CK-MB 013055 484 50 0 01/31/04 10/15/03 8 654321
CK-MB 013040 841 10 12 04/20/04 None 9 654321
CK-MB 013055 842 50 0 01/31/04 10/15/03 10 654321
Ferritin 105011 2297 17 5 04/27/04 10/30/03 11 654321
Ferritin 125002 2296 27 4 04/28/04 None 12 654321
Ferritin 125002 2298 38 3 04/29/04 None 13 654321
Ferritin 105011 3011 17 5 04/27/04 10/30/03 13 654321
Ferritin 125002 3012 27 4 04/28/04 None 15 654321
Ferritin 125002 3010 38 3 04/29/04 None 16 654321
TSH 085673 3013 5 12 04/06/04 17 654321
TSH 10/20/03
fTSH2 None
TSH 085673 3454 5 12 04/06/04 18 654321
TSH 10/20/03
fTSH2 None
MYO 093738 3461 47 7 04/25/04 10/15/03 19 654321
MYO 093738 3462 47 7 04/25/04 10/15/03 20 654321
Page 1 of 1
Technologist ____________________ Printed 10/10/2003 14:36
2053D.bmp
Figure 2-34 Reagent Supplies Report Example
Loading a If there is sufficient reagent available, the system continues to process tests while you
Reagent Pack load another reagent pack.
WARNINGS
• You will come in contact with potentially infectious materials during this
procedure. Handle and dispose of biohazard materials according to proper
laboratory procedures. Proper hand, eye, and facial protection is required.
• Do not load partial packs from another UniCel DxI system onto your
instrument.
NOTES
• The system prevents you from accidentally loading empty or expired reagent
packs. You can use a pack that expires while it is onboard the instrument by
overriding the expired condition. However, selecting to override an expired
condition does not permit you to load a pack that expires while it is offboard,
even if another pack from the same lot is onboard when the expiration occurs.
For more information, see the Overriding Supply Conditions procedure in
Section 2.1: Supply and Supply Conditions Overview.
• Loading a reagent pack does not automatically enable its assay. If the assay is
disabled, you must enable it on the Tests Setup screen. For more information,
see Section 3.3: Test Setup of the Reference Manual.
Green: The system is processing tests and the door is locked. Wait until the light turns
off before loading reagent packs.
2085B.eps
2. Mix the new, unpunctured packs by gently inverting them several times to
dislodge particles from the seal or sides of the well. It is not necessary to
completely resuspend the particles.
NOTES
• Do not invert packs that have been punctured.
• Do not vigorously shake packs or use a vortex mixer, except for the Access
25(OH) Vitamin D Total assay. For the Access 25(OH) Vitamin D Total
assay, see the reagent IFU for pack mixing instructions.
WARNING
In the following step, do not load partial packs from another system onto your
instrument.
CAUTION
To avoid damaging the reagent packs in the following step, be sure they are
properly seated in the tray positions.
4. Place up to four reagent packs in the tray on the UniCel DxI 800 instrument (up to
two packs on the UniCel DxI 600 instrument). Place the packs in the numbered
tray positions so the narrow end is at the back of the tray and the bar code label
faces up.
1
2 3
4
3
2
2086A.eps
1 Reagent pack
2 Reagent pack position
3 Reagent load tray
1 2
3
4
3
2
1
2086B.eps
1 Reagent pack
2 Reagent pack position
3 Reagent load tray
6. Carefully review the Reagent Supplies screen to make sure the information for
the reagent pack just loaded is correct.
WARNING
If you loaded a used pack, make sure the number of tests displayed for the pack
is less than 50. Erroneous, but believable, results may be obtained if a used pack
is incorrectly identified as a new pack containing 50 tests. If a used pack is
identified as a new pack, unload it and use this procedure to load a fresh pack.
Yes
2022D.wmf
Figure 2-38 Loading a Reagent Pack Flowchart
Unloading a The system automatically unloads empty reagent packs and places them in the solid
Reagent Pack waste container. Use this procedure to manually unload a reagent pack. All
information for the unloaded pack remains in the database until you manually delete it
from inventory. For more information, see Section 2.8: Reagent Inventory.
WARNING
You will come in contact with potentially infectious materials during this
procedure. Handle and dispose of biohazard materials according to proper
laboratory procedures. Proper hand, eye, and facial protection is required.
NOTES
• You can unload up to four reagent packs at a time on the Reagent Supplies
screen for the UniCel DxI 800 instrument (up to two packs on the
UniCel DxI 600 instrument).
• You can unload only one reagent pack at a time on the Reagent Inventory
screen.
• If a reagent pack is in-use (identified by the in-use [padlock] icon on the
Reagent Supplies screen), you cannot unload it from the system.
NOTE
If one or more of the selected packs cannot be unloaded, the Unload Reagent
Pack F2 button is not available.
3. When the reagent tray in-use light turns off, open the load door by lifting up on
the bottom edge (see Figure 2-35).
NOTES
• If you unload packs while the system is in the Running mode, it may take
longer than normal for the system to unlock the reagent load door.
• If you do not open the door within one hour, the system returns the pack to
the reagent storage chamber.
NOTE
If you close the door before removing each pack from the tray, the remaining
packs are returned to the reagent storage chamber.
6. If you intend to use an unloaded pack again, store it upright in the refrigerator.
Use the pack before it expires.
(Any Screen)
Bulk Supplies
2023C.svg
Figure 2-39 Unloading a Reagent Pack Flowchart
Unloading All Use this procedure to unload all reagent packs from the system. The system unloads
Reagent the packs, but maintains all pack information in the database.
Packs
WARNING
You will come in contact with potentially infectious materials during this
procedure. Handle and dispose of biohazard materials according to proper
laboratory procedures. Proper hand, eye, and facial protection is required.
2030A.bmp
2. When the reagent tray in-use light turns off, open the reagent load door by lifting
up on the bottom edge (see Figure 2-35).
3. Remove the packs from the reagent tray and close the reagent door.
The system deletes the corresponding Reagent Pack buttons from the Reagent
Supplies screen, and begins moving the next set of reagent packs from the reagent
storage chamber to the tray.
NOTES
• The system waits until the reagent load door is closed to continue unloading
reagent packs.
• If you do not open the door within one hour, the system returns the packs to
the reagent storage chamber and cancels the operation.
4. Repeat steps 2 and 3 until all packs are removed from the instrument.
(Any Screen)
Bulk Supplies
Reagent Supplies F1
Yes
Select Exit F1
2024B.svg
Figure 2-41 Unloading All Reagent Packs Flowchart
The UniCel DxI system keeps an inventory of all on-board and off-board reagent
packs. When you load or unload a reagent pack, the instrument scans the bar code
label and adds or removes an entry for the pack to the Reagent Inventory screen. The
system stores the pack inventory information until it is deleted.
The system tracks pack activity in inventory as processing occurs and as time until
expiration passes for each pack. When a pack is empty, the system discards it
automatically and deletes the pack from inventory. When a pack is expired, the
system notifies you of the condition. Unload the pack from the system and delete it
from inventory. For more information about assay specific open pack stability and
expiration times, see the reagent instructions for use. You can set up the system to
override an expired reagent pack. For more information, see Section 2.3: Supplies
Setup in the Reference Manual.
You can remove a reagent pack from the instrument, store it, and reload it when the
pack is needed. For more information, see Section 2.7: Reagent Supplies.
If necessary, you can manually delete full or partially full packs from the database.
For more information, see the Deleting a Reagent Pack procedure.
NOTE
Once a reagent pack is deleted from the inventory, it cannot be loaded again.
To get to this screen from the Bulk Supplies screen, select Reagent Inventory F8.
2031A.bmp
Test Name (Column) Displays the reagent pack test names that meet the filter criteria.
If a pack is used for more than one test, the system displays the primary and
secondary tests for that pack as two separate entries. When sorted, the primary and
secondary tests stay together. For example, the Folate pack is used to run a primary
test (Folate) and a secondary test (RBC Folate). The system displays both of the
tests, primary and secondary, underneath the primary test (Folate).
Lot No. (Column) Displays each pack lot number.
Serial No. (Column) Displays the unique pack serial number.
Tests Left (Column) Displays the number of tests remaining in each reagent pack.
NOTE
This column is not updated in real-time. To refresh the information on the
screen, exit the screen and redisplay it.
Days Open (Column) Displays the number of days since the system first punctured each
reagent pack.
Pack Expiration (Column) Displays the expiration date for each reagent pack. The expiration date is
the lot expiration date or the open pack expiration date, whichever is earliest.
The open pack expiration date is the date the system first punctured the reagent
pack, plus the number of days defined in the APF.
• This entry turns yellow if the open pack stability expires in three days or sooner.
• This entry turns red if the open pack stability is expired.
Calib. Expiration (Column) For on-board packs, displays the expiration date of the active calibration.
For off-board packs, displays Unloaded.
• This entry turns yellow if the calibration expires in three days or sooner.
• This entry turns red if the calibration is expired.
NOTE
This column is not updated in real-time. To refresh the information on the
screen, exit the screen and redisplay it.
Reagent Supplies (Button) Select to display the Reagent Supplies screen. For more information, see
F1 Section 2.7: Reagent Supplies.
Unload Reagent (Button) Select to unload one selected reagent pack. If the selected reagent pack is
Pack off board this button is not available. For more information, see the Unloading a
F2 Reagent Pack procedure in Section 2.7: Reagent Supplies.
The system updates the screen once the pack is unloaded.
NOTE
This column is not updated in real-time. To refresh the information on the
screen, exit the screen and redisplay it.
Delete Pack (Button) Select to delete an off-board reagent pack from inventory. When you
F5 delete a reagent pack, the system removes pack information from screens and
reports, but retains the information in the database. If the selected pack is on-board
the instrument, this button is not available. For more information, see the Deleting
a Reagent Pack procedure.
Print (Button) Select to print the Reagent Inventory Report for all reagent packs that
F7 meet the filter parameters. For more information, see the Printing the Reagent
Inventory Report procedure.
Filter (Button) Select to change the filter parameters and sort order to apply to the
F8 Reagent Inventory screen. The applied filter and sort order displays in the Filter
field. For more information, see the Applying a Filter to the Reagent Inventory
Screen procedure.
Applying a A filter is a set of parameters you can apply to increase or decrease the number of
Filter to the reagent packs that display on the Reagent Inventory screen.
Reagent Use this procedure to apply a different filter to the Reagent Inventory screen.
Inventory
Screen System Mode: Any Mode
1. From the Reagent Inventory screen, select Filter F8 to display the Filter window.
2032B.bmp
NOTES
• If you select the All Packs filter, the Slot option is not available.
• If you select the Off-Board Packs filter, the Slot and Instrument sort order
options are not available.
5. Select or clear the Ascending option.
6. Select OK F1.
The Reagent Inventory screen displays the reagent inventory according to the applied
filter.
Deleting a Use this procedure to delete off-board reagent packs from inventory.
Reagent Pack
System Mode: Any Mode
NOTES
• You cannot reload a reagent pack after you delete it.
• The system automatically unloads and deletes empty reagent packs from
inventory.
1. From the Reagent Inventory screen, select the off-board reagent pack you intend
to delete.
(Optional) If you cannot locate a pack quickly, apply a filter that displays the
pack near the top of the inventory list. For more information, see the Applying a
Filter to the Reagent Inventory Screen procedure.
2. Confirm that the reagent pack you selected is off board by verifying that the Slot
and Instrument fields are blank.
(Optional) If the reagent pack is on board the instrument, unload the reagent pack.
For more information, see the Unloading a Reagent Pack procedure
in Section 2.7: Reagent Supplies.
3. Select Delete Pack F5.
The system displays a confirmation window.
4. Select OK F1.
The system deletes the reagent pack.
(Any Screen)
Bulk Supplies
Reagent Inventory F8
Reagent Inventory
Yes
Delete Pack F5
OK F1
2025C.svg
Figure 2-44 Deleting a Reagent Pack Flowchart
Printing the Use this procedure to print the Reagent Inventory Report, containing all selected
Reagent packs or all packs in the applied filter.
Inventory System Mode: Any Mode
Report 1. From the Reagent Inventory screen, confirm that the applied filter includes the
packs you intend to print, and that the packs are in the correct sort order. If not,
apply the appropriate filter and sort order. For more information, see the
Applying a Filter to the Reagent Inventory Screen procedure.
(Optional) To print the Reagent Inventory Report for one or more selected packs,
select the packs you intend to print on the screen.
Laboratories, Inc.
Reagent Inventory Report Laboratory A
UniCel DxI 800, Access Immunoassay System 123 Lake Street
Townsville, ST 33333
S/N 654321 101-555-2323 ext. 109 Jane Smith
Page 1 of 1
Technologist ____________________ Printed 10/10/2003 14:36
2054C.bmp
Figure 2-45 Reagent Inventory Report Example
3 Sample Manager
3.1 Sample Manager Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Sample processing begins when the UniCel DxI system queries the laboratory
information system (LIS) for sample test requests, downloads test requests from the
LIS, or when a test request is entered on the Test Requests screen. If necessary, an
operator can edit test request information on the Test Requests screen.
When the system downloads a test request from the LIS, or if the system cannot
process a test, it adds the test to the work pending list on the Work Pending screen.
For more information, see Section 3.2: LIS Requests and Work Pending.
You place samples with assigned test requests in racks, and then load the racks onto
the sample presentation unit (SPU) (UniCel DxI 800 instrument: see Figure 1-1 in
Section 1.2: Instrument Overview; UniCel DxI 600 instrument: see Figure 1-2). The
system moves the racks to the sample aliquot station, where an aliquot of each sample
is removed and stored for processing. After aliquoting the samples, the system moves
the rack to the offload area.
Some assays require temperature restrictions. See Appendix A of the UniCel DxI
Instructions for Use manual for information on these restrictions.
The system displays sample status information on the Sample Manager screen. Test
status information is displayed on the Test Results screen.
If a sample error occurs, the system turns the Rack Exceptions button yellow. The
associated rack is displayed in the Exceptions field on the Sample Manager screen,
and the sample is displayed on the Exceptions view. The system also displays a
specific error message in the Status field of each Sample Manager view.
If the system cannot process a test because a supply level or calibration condition
exists, one or more system indicators present the necessary information.
Query the LIS for test Download test requests from Manually enter test requests
requests the LIS
Sample Manager
(Any Screen)
New Request F3
Patient/QC Requests F1
No
3002C.wmf
Figure 3-1 Sample Processing Overview Flowchart
2035A.bmp
Figure 3-2 Sample Manager Screen, In Progress View
2166A.bmp
Sample Manager (Buttons) There are four Screen View buttons. Each button corresponds to a
Screen View specific screen view. For more information, see the Sample Manager Screen Views
Buttons topic.
Select a button to display one of the four Sample Manager screen views. For more
information, see the Selecting a Sample Manager Screen View procedure.
Rack (Column) Displays the ID of the rack on which the sample is loaded.
Rack icon (Requested view only)
(Indicator) Displays the rack ID and the position of each sample
container in the rack.
Pos (Column) Displays the position of the sample in the rack, from 1-4.
Description (Column) Displays a description of the associated sample error.
(Exceptions view
only)
Sample ID (Column) Displays the unique value that identifies one sample from another.
• For patient samples, the sample ID is displayed.
• For QC samples and calibrators, the control or calibrator name is displayed.
• For maintenance samples, the maintenance routine sample type is displayed.
Patient ID (Column) Displays an optional value that identifies the sample being tested.
• For patient samples, the optional patient ID is displayed.
• For QC samples and calibrators, the lot number is displayed.
• For maintenance samples, the maintenance routine sample type is displayed.
NOTE
This column is not displayed on the Exceptions view.
Status (Column) Depending upon the selected screen view, displays the status
information for individual samples, or all samples in a rack. For more information,
see the Sample Status or see the Test Status topics in Section 3.5: Monitoring
Sample Progress.
NOTE
This column is not displayed on the Requested or Exceptions views.
Complete (Column) Displays the estimated completion time for each sample.
NOTE
This column is not displayed on the Requested or Exceptions views.
Creation Time (Column) Displays the time the associated sample error occurred.
(Exceptions view
only)
Sample Aliquot (Field and Icon) Displays the racks that are currently in the sample aliquot station.
Station The aliquot station processes up to four racks at a time. As a rack moves out of the
station, it is removed from the field display and is replaced by another rack. If there
are no racks in the sample aliquot station, the value Empty is displayed in the field.
NOTE
The field displays the order and direction of racks in the aliquot station. It does
not represent the actual location of the racks in station.
Exceptions (Field and Icon) Displays the racks that contain samples with one or more
exception errors. A sample container with an associated exception is highlighted in
red on the rack icon.
NOTES
• Only containers with an associated exception appear on the rack icon.
• This field is displayed if there is a rack with an associated sample
exception. If there are no exceptions, this area of the Sample Manager
screen is empty.
If there are more than four racks with exception errors, a scroll bar is displayed
next to the field. Racks with exceptions remain in the field for three days. After this
time, they are automatically removed from the field.
You can view the exception information for a specific rack of samples by selecting
the rack icon. The system displays the Exceptions view, and highlights the samples
on the screen. For more information, see the Selecting a Sample Manager Screen
View procedure.
Views (Button) Select to display one of the four Sample Manager screen views. For more
F1 information, see the Selecting a Sample Manager Screen View procedure.
Edit Request (Button) Select to edit a request on the selected rack. The Test Requests screen that
F2 corresponds to the type of request you are editing (patient/QC, calibration,
maintenance) is displayed. From this screen you can edit the test requests for the
entire rack, or for an individual sample. For more information, see
Section 3.6: Editing Test Requests.
NOTE
This button is only available on the Requested or Daily views.
Clear Exceptions (Button) Select to remove a sample from the Exceptions view. For more
F2 information, see the Clearing an Exception from the Exceptions View procedure.
(Only available on
the Exceptions
view)
New Request (Button) Select to request a new test request. There are three different types of test
F3 requests.
Patient/QC Requests F1
Select to request tests for patient samples and quality controls. For more
information, see the Manual Test Requests procedure in Section 3.3: Patient Test
Requests, or see the Entering QC Test Requests procedure in Section 6.3: Running
Quality Control Tests.
Calibration F2
Select to request an assay calibration. For more information, see the Entering
Calibration Requests procedure in Section 5.3: Running a Calibration.
Maintenance F3
Select to request a maintenance routine. For more information, see the Requesting
Maintenance Routines procedure in Section 7.1: Maintenance Overview.
Test Results (Button) Select to display the Test Results screen. Test requests begin to display on
F4 the Test Results screen when the samples are aliquoted by the system. For more
information, see Section 4.1: Test Results Overview.
Work Pending (Button) Select to display the Work Pending screen. For more information, see
F5 Section 3.2: LIS Requests and Work Pending.
Find (Button) Select to locate a specific sample ID on the Sample Manager screen. For
F6 more information, see the Finding a Sample ID procedure.
NOTE
This button is not available on the Exceptions view.
Details (Button) Select to display detailed information about the tests for a selected
F8 sample. For more information, see the Displaying Test Details procedure.
NOTE
This button is only available on the In Progress or Daily views.
Sample The Sample Manager screen features four screen views. Each view arranges and
Manager presents sample status information in a specific way. You can select a new screen
Screen Views view at any time. For more information, see the Selecting a Sample Manager Screen
View procedure.
In Progress View
The status information for in progress samples is displayed on the In Progress screen
view. Samples are removed from this view after the instrument generates their test
results.
Requested View
Samples waiting to be processed are displayed on the Requested screen view. The
samples on this view are grouped into their associated racks. Racks are displayed after
you enter them manually, or after the instrument scans each rack. Racks are removed
from this view after all samples in the rack have begun processing.
NOTES
• This view includes a Rack icon for each rack on the screen. The icon displays
the rack ID and position of each sample container. For more information, see
the Sample Manager Screen topic.
• This view does not include the Status and Complete columns.
Daily View
Samples that were processed within a 24-hour time period are displayed on the Daily
screen view. You can change the time period by entering a new date in the field that is
displayed on the Daily button. For more information, see the Entering a Date on the
Daily View procedure.
Exceptions View
Samples with any associated processing errors are displayed on the Exceptions screen
view. Any of the following conditions will generate an exception error.
• Duplicate rack ID
• Duplicate sample ID (on the same rack)
• QNS error for a sample container
• Rack bar code error
• Sample container bar code error
NOTE
Samples with an exception have not been aliquoted by the instrument.
Samples remain on the Exceptions view for three days. After this time, they are
automatically removed from the view.
Selecting a Use this procedure to select one of the four screen views.
Sample
System Mode: Any Mode
Manager
From the Sample Manager screen, select the appropriate Screen View button located
Screen View above the rows of sample information. For more information, see the Sample
Manager Screen Views topic.
NOTES
• You can access the Exceptions view from any screen by selecting the Rack
Exceptions button.
• On the Sample Manager screen, racks with one or more exception errors
appear as icons in the Exceptions field. If you select a rack icon, the
Exceptions view is displayed and the samples are highlighted on the screen.
Entering a Use this procedure to display a specific 24-hour time period on the Daily view of the
Date on the Sample Manager screen.
Daily View System Mode: Any Mode
1. From the Sample Manager screen, select the Daily view button to display the
Daily view.
2. Enter the date you wish to view in the field that is displayed on the Daily view
button, and press [Enter].
NOTE
When you enter the date, use the date format that you configured for the system.
For more information, see the Setting the Date and Time Formats procedure in
Section 2.2: System Setup of the Reference Manual.
The status information for the tests requested on that day is displayed.
Clearing an Use this procedure to clear a sample exception from the Exceptions view of the
Exception Sample Manager screen.
from the System Mode: Any Mode
Exceptions 1. From the Sample Manager screen, select the Exceptions button to display the
View Exceptions view.
2. To limit the number of exceptions you will clear, select one sample exception on
the screen, or press and hold the [Ctrl] key as you select more than one exception.
3. Select Clear Exceptions F2.
4. In the Clear Exceptions window, select an option from the Clear What list.
Finding a Use this procedure to find a specific sample ID on the Sample Manager screen.
Sample ID System Mode: Any Mode
NOTES
• The find function is not case sensitive.
• The find function is limited to the samples displayed on the current screen
view. If you are having difficulty finding a specific sample ID, change the
view.
1. From the Sample Manager screen, select Find F6 to display the Find window.
2. In the Find Sample ID field, enter either the complete or partial sample ID of the
sample you intend to find.
3. Select Find F1.
The system highlights the matched sample ID. If no match is found, a message is
displayed.
NOTE
Confirm that the sample ID you entered was correct. If the ID is correct, change
the screen view.
(Optional) If the system did not match the sample you intend to find, select
Find Next F1 to find the next match to your query. The next match becomes the
selected sample ID.
(Optional) To find a new sample ID, repeat steps 2 and 3.
4. When you are finished looking for specific samples, select Close F8 to exit the
window.
Displaying Use this procedure to display test details for a selected sample.
Test Details System Mode: Any Mode
NOTES
• You can only view test details on the In Progress or Daily views.
• If you have difficulty locating a sample, change the screen view or use the
Find function to find the sample ID.
1. From either the In Progress or Daily views of the Sample Manager screen, select
the sample whose test detail information you intend to view.
2. Select Details F8 to display the Test Details window.
2126A.bmp
3. When you are finished viewing the test details, select OK F1.
(Optional) To view the test details for another sample, repeat steps 1-3.
Resolving Use this procedure to resolve a sample error. You can view all of the samples with an
Sample Errors associated sample error on the Exceptions view of the Sample Manager screen.
1. From the Sample Manager screen, select the Exceptions button to display the
Exceptions view.
3. Take the necessary action to resolve the sample error. For more information, see
Section 3.7: Patient Sample Troubleshooting.
4. Once you have performed the necessary troubleshooting steps, select the sample
on the Exceptions view and then Clear Exception F2.
The sample and error message are removed from the Exceptions view.
LIS requests are processed by host query or by downloading from the laboratory
information system (LIS). Select the method for your system on the LIS Setup
window during system configuration. For more information, see Section 2.5: LIS
Setup in the Reference Manual. For downloaded requests, the system displays the
pending tests on the Work Pending screen. The tests remain on the screen until you
load the missing samples onto the instrument.
NOTES
• A test request with an associated sample dilution factor will not be placed on
the Work Pending screen. The system records an event in the Event Log,
which instructs you to request the sample manually and add the correct
dilution factor from the Test Request screen. For more information, see
Section 3.3: Patient Test Requests.
• If you use a Patient ID in a manual test request, the UniCel DxI system will
evaluate the manually requested results against default ranges even if your
laboratory has defined different ranges (reference, critical, and/or LIS) based
on demographics. For more information, see Section 3.4: Ranges Setup in the
Reference Manual.
The Work Pending screen lists all the tests that the UniCel DxI system cannot process
because samples are missing. The system turns the Work Pending button yellow
when it adds a test to the work pending list. You resolve items on the list by loading
samples or manually deleting tests from the list.
Pending tests for the following kinds of missing samples are displayed on the Work
Pending screen.
• A missing sample for a test downloaded from the LIS or for a manual test
request. For more information, see Section 3.3: Patient Test Requests.
• A sample without enough reserve volume to process a reflex test or a test
rerun from the Test Results screen. For more information, see Section 3.7:
Reflex Test Setup of the Reference Manual.
• A sample without enough reserve volume to process an automatic rerun test
request.
• A sample that the system did not process because the sample had expired.
• A sample that the system did not aliquot because there was not enough
volume.
• A sample that the system did not aliquot because it contained a clot or some
other type of obstruction.
• A sample that is missing or that the system did not aliquot for an unspecified
reason.
The Work Pending screen is updated in real time when a test on the screen is assigned
to a sample, or when test requests are downloaded from the LIS. You can refresh the
screen by selecting the Work Pending button.
2092D.bmp
NOTE
Apply the LIS Request filter option to view the number of tests in the LIS
Workpool. If the filter is not applied, the number in this field will always be 0.
For more information, see the Applying a Filter to the Work Pending Screen
procedure.
Sample ID (Column) Displays the sample ID for each sample in the work pending list.
Patient ID (Column) Displays the required patient ID for LIS requests, or the lot number for
QC requests.
Requested (Column) Displays the date and time of the original test request.
Tests Pending (Column) Displays the names of the tests that are pending. Each test name is
displayed in a separate row.
Origin (Column) Displays the origin of the test request in the work pending list.
(Rack/Instrument) • Reflex - The system initiated a reflex test and there was insufficient reserve
volume to process the request.
• Manual Rerun - The operator requested that a test be rerun and there was
insufficient reserve volume to process the request.
• Automatic Rerun - The system requested that a test be rerun and there was
insufficient reserve volume to process the request.
• LIS - The LIS requested a test or reflex test and there was not enough sample to
process the request, or the sample was not on board.
• Sample Expired - The system did not process the sample because the sample was
expired.
• QNS - The system did not aliquot the sample because there was not enough
volume.
• Obstruction - The system did not aliquot the sample because there was a clot or
some other type of obstruction.
• Other - The test is pending for any reason other than those defined above.
The rack ID associated with the sample is displayed in parentheses. If the sample is
not associated with a rack, the Rack ID is left blank. The Instrument ID, in
parentheses with the Rack ID, is not used by the system and will always display
Unknown.
Filter (Button) Select to change the filter parameters and sort order for the Work Pending
F1 screen. The applied filter and sort order displays in the Filter field. For more
information, see the Applying a Filter to the Work Pending Screen procedure.
Sample Manager (Button) Select to display the Sample Manager screen. For more information, see the
F2 Sample Manager Screen topic in Section 3.1: Sample Manager Overview.
Delete (Button) Select to delete tests from the work pending list. For more information, see
F6 the Deleting Tests from Work Pending procedure.
Print (Button) Select to print the Work Pending Report. For more information, see the
F7 Printing the Work Pending Report procedure.
Options (Button) Select to display the Options window. From the Options window, you can
F8 configure the system to automatically rerun tests that are cancelled due to system
errors. For more information, see the Configuring the Automatic Rerun Feature
procedure.
Downloading Use this procedure to process test requests downloaded from the LIS. When you load
LIS Test bar coded sample containers for the tests listed on the Work Pending screen, the
Requests system matches the downloaded test requests to the samples and runs the tests.
NOTES
• When downloading test requests, be sure that LIS host query is not enabled.
For more information, see Section 2.5: LIS Setup in the Reference Manual.
• When setting up new tests to run on the instrument, enable the tests before
downloading test requests from the LIS. The system rejects test requests for
tests that are not enabled. For more information, see Section 3.1: Configuring
Tests and Results Overview in the Reference Manual.
1. From any screen, select the Work Pending button to display the Work Pending
screen.
2. Review the LIS Workpool field and the list of tests displayed on the screen for
test requests the system has downloaded from the LIS.
NOTE
Be sure that LIS requests are included in the applied filter. For more information,
see the Applying a Filter to the Work Pending Screen procedure.
3. Place the bar coded sample containers for the samples and tests listed on the
screen in racks. For more information, see the Placing Sample Containers in
Racks procedure in Section 3.4: Loading and Unloading Racks.
NOTES
• Place only containers for samples listed on the screen in a rack. If the system
cannot match a sample ID bar code to an ID in the workpool, the system will
not remove an aliquot for processing.
• If you are placing sample containers in a rack that is designated for reserve
volume, be sure that the sample containers contain enough sample volume to
provide the reserve volume. For more information, see the Calculating
Minimum Sample Volume procedure in Section A.2: Sample Containers.
4. Load the racks. For more information, see the Loading a Rack for Routine
Processing procedure. Load any racks that contain STAT samples first. For more
information, see the Loading a STAT Rack procedure in Section 3.4: Loading and
Unloading Racks.
5. The system scans the racks and sample container bar code labels, and updates the
Sample Manager screen. For more information, see Section 3.1: Sample Manager
Overview.
(Any Screen)
Work Pending
2095C.wmf
Figure 3-5 Downloading LIS Test Requests Flowchart
Querying the Use this procedure to query the laboratory information system (LIS) for test requests.
LIS for Test
System Mode: Ready Running Paused
Requests
NOTES
• Be sure that LIS host query is enabled. For more information, see
Section 2.5: LIS Setup in the Reference Manual.
• When setting up new tests to run on the instrument, enable the tests before
querying the LIS for test requests. The system rejects an LIS test request if
the associated test is not enabled. For more information, see Section 3.1:
Configuring Tests and Results Overview in the Reference Manual.
1. Place the bar coded sample containers in racks. For more information, see the
Placing Sample Containers in Racks procedure in Section 3.4: Loading and
Unloading Racks.
NOTE
If you are placing sample containers in a rack that is designated for reserve
volume, be sure that the sample containers contain enough sample volume to
provide the reserve volume. For more information, see the Calculating Minimum
Sample Volume procedure in Section A.2: Sample Containers.
2. Load the racks. For more information, see the Loading a Rack for Routine
Processing procedure in Section 3.4: Loading and Unloading Racks. Load any
racks that contain STAT samples first. For more information, see the Loading a
STAT Rack procedure.
3. The system scans the bar code labels for each sample container in the rack,
queries the LIS for the associated test requests, and then waits for the LIS to send
the requests.
NOTES
• You set the length of time the system waits for a request in the LIS Setup
window. For more information, see Section 2.5: LIS Setup in the Reference
Manual.
• If the system does not receive a test request within the specified time period,
the system does not remove an aliquot for processing and moves to the next
container in the rack. The container with no assigned test requests is not
placed in the work pending list. The container must be reloaded after the
system aliquots the remaining samples in the rack.
When the system receives the test requests from the LIS, it updates the Sample
Manager screen and scans the bar code label on the next sample container in the rack.
Applying a The work pending filter is a set of parameters that you can apply to designate the
Filter to the kinds of test requests to be displayed on the Work Pending screen.
Work Pending Use this procedure to apply different filter parameters or a different sort order to the
Screen Work Pending screen.
NOTE
The Work Pending button turns yellow when a test request sample that meets the
applied filter parameters is added to the work pending list.
2139C.bmp
2. Select one or more of the pending test request options to include in the filter, or
clear to remove a selection from the filter.
(Optional) Select a Sort Order option.
(Optional) Select or clear the Ascending option.
3. Select OK F1. The Work Pending screen displays all pending tests that meet the
selected filter parameters.
Deleting Tests Use this procedure to delete pending tests displayed on the Work Pending screen.
from Work
System Mode: Any Mode
Pending
1. On the Work Pending screen, locate and select the tests you intend to delete from
the work pending list.
(Optional) To reduce or expand the test requests that are displayed on the Work
Pending screen, change the filter parameters. For more information, see the
Applying a Filter to the Work Pending Screen procedure.
2. Select Delete F6 to display the Delete window.
The Current Test option is selected in the Delete what list.
2140A.bmp
Delete (Option) Select the tests that you will delete from the Work
What Pending screen.
• Current Test - Select to delete the test you selected on the
work pending list.
• Selected Tests - If you selected more than one test on the
work pending list, select to delete each of the tests. This
option is not available if you only selected one test on the list.
• All Tests in Filter - Select to delete all of the tests that meet
the applied filter parameters.
4. Select OK F1 to delete the test or tests from the work pending list.
A confirmation window is displayed.
Printing the Use this procedure to print the Work Pending Report, which lists the pending test
Work Pending requests that meet the applied filter parameters.
Report System Mode: Any Mode
1. On the Work Pending screen, confirm that the tests you intend to print are
displayed in the applied filter.
(Optional) To reduce or expand the kinds of tests to print, or to change the sort
order, apply a new filter to the Work Pending screen. For more information, see
the Applying a Filter to the Work Pending Screen procedure.
Laboratories, Inc.
Work Pending Report Laboratory A
UniCel DxI 800, Access Immunoassay System 123 Lake Street
Townsville, ST 33333
S/N 654321 101-555-2323 ext. 109 Jane Smith
Filter: Reflex Tests, Manual Rerun, LIS Requests, Sample Expiration, Insufficient Sample Volume, Other
Tests Requested: 6
LIS Workpool: 4
Page 1 of 1
Technologist ____________________ Printed 10/10/2003 14:36
2147D.bmp
Figure 3-8 Work Pending Report Example
A patient test request includes a unique sample ID, sample information such as patient
ID and processing priority, and an assigned test, or list of tests, to be run on the
UniCel DxI system.
In most cases, the system queries a laboratory information system (LIS) for patient
sample and test request information, or downloads the information from the LIS. For
more information, see Section 3.2: LIS Requests and Work Pending. Patient sample
and test request information that does not originate from an LIS is manually entered
on the Test Requests screen. If necessary, the test list or patient sample information
can also be edited on this screen. For more information, see Section 3.6: Editing Test
Requests.
System errors can cause the cancellation of tests and prevent the system from
calculating a result. The system can be configured to automatically rerun some
cancelled patient and control tests. For more information, see the Configuring the
Automatic Rerun Feature topic.
Calibration, QC, and maintenance test requests are entered manually on the Test
Requests screen. For more information, see Section 5.3: Running a Calibration, see
Section 6.3: Running Quality Control Tests, or see Section 7.1: Maintenance
Overview.
NOTE
Manually requested patient test results are evaluated against default ranges, even
if your laboratory has established reference, critical, or LIS ranges. For more
information, see Section 3.4: Ranges Setup in the Reference Manual.
2068D.bmp
Rack Button (Button and Field) Displays the rack ID, and the position and sample ID of each
and container in the rack.
Enter ID Select the Rack button to enter or change the rack ID. The Enter ID field is
displayed.
Auto Sample ID (Indicator) When the Auto Sample ID option is on, the indicator displays above the
sample information. When the Auto Sample ID option is off, the indicator is not
displayed. For more information, see the Using Auto Sample ID procedure.
Batch Request (Indicator) When the Batch Request option is on, the indicator displays above the
sample information. When the Batch Request option is off, the indicator is not
displayed. For more information, see the Using Batch Request procedure.
Position (Buttons) Select a numbered position to enter or edit sample information for the
sample in that position in the rack. The positions in a rack are arranged numerically
from 1-4, as displayed on the Rack button.
The sample information fields for a specific sample are available for entry when its
corresponding Position button is selected. You can enter or change the information
in any field until the system scans the rack.
Sample ID (Field) Displays the unique value that identifies one sample from another.
• For manual test requests, enter the sample ID in this field.
• For QC test requests, select Request QC F5 to enter a control ID in this field.
The tests requested for the sample are displayed beneath the sample ID. If the test
list is too long to fit in the field, ... is displayed.
Patient ID/Lot (Field) Displays an optional value that identifies the patient whose sample is being
Number tested. If you enter a quotation mark ("), characters to the right of the quotation
mark may not print on reports.
• For manual test requests, enter the patient ID in this field.
NOTE
A patient ID is required for LIS test requests.
• For QC test requests, the system enters the control lot number in this field when
you select a control from the Request QC window.
Dilution (Field) Displays the dilution factor for a manually diluted sample. The default
value is 1. The acceptable dilution range is from 1 (no dilution, 1:1) to 999.
• For manual test requests, change the default dilution factor if necessary.
• For QC test requests, this field is not available.
Sample Type (Field) Displays the sample fluid type.
• For manual test requests, change the sample type if necessary.
• For QC test requests, the system displays the sample type entered during setup.
Figure 3-9 Test Requests Screen Descriptions for Patient/QC Samples (continued)
STAT (Box) Displays the sample priority. The default value is routine priority (box not
selected).
• For manual test requests, select the box to process a patient sample with STAT
priority (before all other samples), or clear the box to process the sample with
routine priority.
NOTE
Confirm that all tests requested for a STAT sample have an active assay
calibration listed on the Calibration screen. For more information, see the
Calibration Screen topic in Section 5.1: Assay Calibration Overview.
The system displays sample comments on test result reports and on the Quality
Control screen and reports.
Tests Requested (Field) Displays a list of all the tests requested for a selected sample (X is the
for Sample X selected sample position). Test replicates are displayed as a number in parentheses
after the test name. For quality control samples, the system displays the reagent lot
number in parentheses after the test name if you use the Change Reag. Lot button
to specify a reagent lot for the control.
For manual and QC test requests, you can use the Test buttons to add or delete tests
from the test list. For more information, see Section 3.6: Editing Test Requests.
Test (Field) Displays the test ID for the selected test.
• To add a replicate, press the [+] key.
• To remove a replicate, press the [-] key.
Figure 3-9 Test Requests Screen Descriptions for Patient/QC Samples (continued)
Test Buttons (Buttons) Displays the test name and ID for each available test or test panel, sorted
by test number. Select a button to add one test, or one replicate of each test in a test
panel, to the Tests Requested for Sample X list.
The test buttons that display depend on the currently selected sample tube. If
available, test panel button names are displayed in bold blue, and onboard dilution
test button names are displayed in blue.
Remove (Button) Select to remove one replicate of the selected test in the Tests Requested
for Sample X list.
Rack ID (Button) Select to enter a new rack ID, or change the current rack ID. The Enter ID
F1 field is displayed on the Rack button. For more information, see the Manual Test
Requests procedure or see Section 3.6: Editing Test Requests.
Delete Sample (Button) Select to delete the sample selected on the Test Requests screen. For more
F2 information, see Section 3.6: Editing Test Requests.
New Request (Button) Select to exit this request and to request a new test request. There are three
F3 different test request types.
Patient/QC Requests F1
Select to request tests for patient samples and quality controls. For more
information, see the Manual Test Requests procedure, or see the Entering QC Test
Requests procedure in Section 6.3: Running Quality Control Tests.
Calibration F2
Select to request an assay calibration. For more information, see the Entering
Calibration Requests procedure in Section 5.3: Running a Calibration.
Maintenance F3
Select to request a maintenance routine. For more information, see the Requesting
Maintenance Routines procedure in Section 7.1: Maintenance Overview.
Test Results (Button) Select to display the Test Results screen. Test requests begin to display on
F4 the Test Results screen when the samples are aliquoted by the system. For more
information, see Section 4.3: Reviewing Test Results.
Request QC (Button) For quality control samples, select to display the Request QC window.
F5 For more information, see Section 6.3: Running Quality Control Tests.
If the selected sample position contains a patient sample, the system displays an
error message.
Cancel Request (Button) Select to cancel the test request and to return to the Sample Manager
F6 screen. For more information, see Section 3.6: Editing Test Requests.
Clear All Samples (Button) Select to clear the sample information for each sample on the selected
F7 rack.
Figure 3-9 Test Requests Screen Descriptions for Patient/QC Samples (continued)
Figure 3-9 Test Requests Screen Descriptions for Patient/QC Samples (continued)
Manual Test Use this procedure to manually enter patient test requests that do not originate from
Requests the LIS.
NOTES
• Do not enter a quotation mark (") in an available field. If you do, characters
to the right of the quotation mark may not print on reports.
• Manually requested patient test results are evaluated against default ranges,
even if your laboratory has established reference, critical, or LIS ranges. For
more information, see Section 3.4: Ranges Setup in the Reference Manual.
System Mode: Any Mode
1. From the Sample Manager screen (see Figure 3-3 in Section 3.1: Sample
Manager Overview), select New Request F3.
2. Select Patient/QC Requests F1 to display the Test Requests screen.
3. Enter the rack ID in the Enter ID field and press [Enter].
NOTE
The system will reject the rack ID if the ID is not set up for your system. For more
information, see the Setting Up Rack IDs procedure in Section 2.3: Supplies
Setup of the Reference Manual.
NOTES
• The sample ID can be up to 15 characters long. The system accepts the
following characters: 0-9, A-Z, a-z, /, +, and -. The system does not accept
spaces, punctuation marks, or special characters in this field.
• When the rack is loaded, the system compares the manually requested ID to
the sample container bar code, if present. If the sample IDs do not match, the
NOTE
The patient ID can be up to 15 characters long. The system accepts any character
in this field.
5. Select the tests that you will run on the sample by selecting their corresponding
Test buttons. To run a replicate of a test, select the Test button again.
The number of additional replicates are displayed in parentheses to the right of the test
name in the Test Requested for Sample X list.
(Optional) If you are requesting confirmatory tests, use one of the following
methods:
• Request the screening and confirmatory tests at the same time, so that the samples
for each test are aspirated from the same sample container. For more information
about confirmatory tests, see Section 1.5: Confirmatory Assay Theory in the
Reference Manual.
• Request the screening test only. However, before doing this, enable an adequate
reserve volume and set up a reflex condition to automatically run the
confirmatory tests if the test result is reactive. For more information, see the
Setting Up Reserve Volume procedure in Section 3.7: Reflex Test Setup of the
Reference Manual.
(Optional) If you will run the same tests on subsequent samples, turn the Batch
Request option on. For more information, see the Using Batch Request
procedure.
(Optional) To remove a requested test or one of its replicates, select the test in the
Tests Requested for Sample X list and then select Remove.
6. Review the values in the following fields and change them if necessary.
a. Dilution: Enter a dilution factor for a manually diluted sample.
NOTES
• To change the dilution factor, enter a whole number. The acceptable dilution
range is from 1 (no dilution, 1:1) to 999 (1:999 dilution). The system does not
accept other entries, such as 1/3 or 3:1.
• You cannot enter a manual dilution factor for an onboard dilution test.
b. Sample Type: If the sample type does not match the sample type set up for
the test, select a type from the option list.
c. STAT: Select the box to assign the STAT priority to the sample.
NOTE
If you assign the STAT priority to a sample, confirm that each requested test has
an active calibration on the Calibration screen For more information, see the
Calibration Screen topic in Section 5.1: Assay Calibration Overview.
d. Comment: Select the Pen or the Notepad button to display the Comment
window. For more information, see the Adding or Deleting Sample
Comments procedure.
(Optional) To enter additional test requests for the same rack, select a new sample
row and repeat steps 4-6.
7. Confirm that the sample and test request information entered on the Test Requests
screen is correct.
(Optional) If you requested STAT samples for the rack, place the containers in
the rack and load the rack immediately. For more information, see the Placing
Sample Containers in Racks and see the Loading a STAT Rack procedures in
Section 3.4: Loading and Unloading Racks.
NOTE
If you are placing sample containers in a rack that is designated for reserve
volume, be sure that the sample containers contain enough sample volume to
provide the reserve volume. For more information, see the Calculating Minimum
Sample Volume procedure in Section A.2: Sample Containers.
NOTE
If you are placing sample containers in a rack that is designated for reserve
volume, be sure that the sample containers contain enough sample volume to
provide the reserve volume. For more information, see the Calculating Minimum
Sample Volume procedure in Section A.2: Sample Containers.
Load the racks. For more information, see Section 3.4: Loading and Unloading Racks.
Sample Manager
New Request F3
Patient/QC Requests F1
Test Requests
Yes Yes
2096C.svg
Figure 3-10 Manual Test Requests Flowchart
Using Auto The Auto Sample ID option automatically assigns sequential sample IDs during
Sample ID manual test request entry. This feature is typically used when your laboratory needs to
perform a correlation study or method comparison for an updated assay.
Examples
• If the first sample ID ends with a numeric character, Sample1, the subsequent
sample IDs are Sample2, Sample3, and so on.
• If the first sample ID ends with a non-numeric character, SampleA, the
subsequent sample IDs are SampleA1, SampleA2, and so on.
Use this procedure to turn the Auto Sample ID option on and off.
CAUTION
When using the Auto Sample ID option, the first sample ID must be less than
the allowed 15 characters. The sequential characters added for an Auto
Sample ID are included in the 15 character maximum. If a sample ID
exceeds 15 characters, a system error will occur when the rack is loaded.
2. Select More Options F8, and then select Turn Auto Sample ID On F2.
3. Select the next sample position. The system automatically enters the next
incremental sample ID in the Sample ID field.
(Optional) To change the sample ID sequence, select the Sample ID field and
enter a new ID.
5. When you are done entering test requests, exit the Test Requests screen.
The system automatically turns the Auto Sample ID option off.
Using Batch The Batch Request option automatically assigns the same test list from the previous
Request sample to the next sample.
Use this procedure to turn the Batch Request option on and off.
2. Select More Options F8, and then select Turn Batch Request On F1.
3. Select the next sample position and enter the sample ID.
The system automatically assigns the same test list when you begin entering
information in another field.
5. When you are done entering test requests, exit the Test Requests screen.
The system automatically turns the Batch Request option off.
Requesting Your technical support representative may use service assays for instrument
Service diagnostics.
Assays
NOTE
Use this function only as directed by your technical support representative.
Reusing Patient samples are identified by a unique sample ID. For 30 days after a patient test
Sample IDs request is entered on the system, a request that includes the same sample ID is
automatically associated with the sample in the original test request. If the sample ID
is inactive for 30 days or more, the system accepts the ID as a unique identifier again.
If you need to reuse a sample ID within 30 days, all of the test request information
about the previous sample ID must be removed from the system. Use this procedure to
remove the sample ID information.
2. Be sure that each test you request for the reused sample ID is enabled. For more
information, see Section 3.3: Test Setup in the Reference Manual.
3. Delete all work pending tests associated with the sample ID. For more
information, see the Deleting Tests from Work Pending procedure in Section 3.2:
LIS Requests and Work Pending.
Adding or Use this procedure to enter, edit, or delete sample comments on the Test Requests
Deleting screen.
Sample System Mode: Any Mode
Comments 1. Select the Pen or the Notepad button to display the Sample Comment window.
2. In the Comment field, enter or edit comments. Or, to delete an existing comment,
select it and press the [Backspace] key.
Configuring Use this procedure to configure the system to automatically rerun patient and control
the Automatic tests that are cancelled due to system errors characterized by the fatal flags listed
Rerun Feature below. For more information, see the Test Result Flags and Troubleshooting topic in
Section 4.3: Reviewing Test Results.
• CLT • QNS • QSB • QSD • QSS
• RLU • SYS • TRI • TRS • TRW
NOTES
• The automatic rerun feature is not available on integrated workstations.
• Tests for manually diluted samples are rerun only if sufficient reserve volume
is onboard when the automatic request is made. If sufficient reserve volume
is not onboard, the request is not added to Work Pending.
• If a test is cancelled due to a reagent pipettor error, the system attempts to
automatically rerun the test using a different pipettor. If the test is configured
for only one pipettor, the test is run on the same pipettor.
• The automatic rerun feature does not apply to confirmatory tests. A cancelled
confirmatory test and the qualitative test it is paired with must be requested
manually.
2352A.bmp
3. Select the Rerun cancelled tests automatically box to enable the automatic
rerun feature, or clear the box to disable the automatic rerun feature.
4. Select OK F1.
The UniCel DxI system uses four different rack sizes. Each rack holds a specific type
of sample container. Racks include rack ID and sample container type labels to help
identify each type of rack. You configure the types of sample containers that are
accepted for a range of rack IDs on the System Setup screen. For more information,
see Section 2.2: System Setup in the Reference Manual.
You place sample containers in the appropriate racks, and then load the racks in the
onload area of the sample presentation unit (SPU). The onload area of the
UniCel DxI 800 instrument holds approximately 30 racks. The onload area of the
UniCel DxI 600 instrument holds approximately 15 racks. The system moves the
racks to the sample aliquot station, where an aliquot of each sample is removed for
processing. When all samples in a rack are aliquoted, the system moves the rack to the
offload area of the SPU. You can unload racks from the offload area at any time.
WARNING
You will come in contact with potentially infectious materials when following
these procedures. Handle and dispose of biohazard materials according to
proper laboratory procedures. Proper hand, eye, and facial protection is
required.
You can monitor the status of samples in a rack on the Sample Manager screen. For
more information, see Section 3.5: Monitoring Sample Progress.
Attaching Use this procedure to attach a bar code label on a sample container.
Sample
WARNING
Container Bar
Code Labels You will come in contact with potentially infectious materials during this
procedure. Handle and dispose of biohazard materials according to proper
laboratory procedures. Proper hand, eye, and facial protection is required.
NOTE
The internal bar code reader automatically discriminates between symbologies,
so that samples with bar code labels using different symbologies can be included
on the same rack. You may need to enable parameters for the bar code
symbologies that you use. For more information, see Section 2.4: SPU Bar Code
Reader Setup in the Reference Manual.
1. Place the bar code label on the sample container so that the label is at least
0.4 inches (10 mm) from the top and bottom of the container. Be sure that it is
securely fastened.
2. When you place the sample container in the rack, be sure the sample bar code
label is centered in the opening slot of its rack position for accurate scanning.
1
ABC123456789
ABC123456789
2199A.eps
CAUTION
Remove caps from all sample containers before loading on the UniCel DxI
system.
NOTE
Do not place an insert cup in a rack. Place insert cups inside a sample container.
For more information, see the Using Insert Cups with Sample Tubes procedure.
NOTE
If you are transferring sample to a new sample container, carefully pipette the
sample into the sample container or insert cup. Avoid creating air bubbles.
WARNING
Use the rack identifier, and the container type and rack ID labels to locate the
correct rack in the following step. If you place a sample container in an incorrect
rack, you may damage the system or compromise the integrity of your test
results.
2. Locate the appropriate rack for the sample container (as indicated by the rack ID
number and label). If you do not intend for the system to draw a reserve, be sure
to place the container in a rack identified with the No Reserve Volume label. For
more information, see the Setting Up Rack IDs procedure in Section 2.3: Supplies
Setup of the Reference Manual.
WARNING
In the following step, if you are manually entering test requests for sample
containers with no bar code labels, be sure to place the sample containers in the
positions you specified on the Test Requests screen. If the containers are not
placed in the designated positions, the system will report incorrect test results.
3. Place the sample container in the rack. Be sure the sample bar code label is in
good condition, and is properly affixed to the container. When you place the
container in a rack, the label should be visible through the slot.
NOTE
Be sure to use supported sample bar code symbologies, and to configure the
internal bar code reader appropriately. For information, see Section 2.4: SPU Bar
Code Reader Setup in the Reference Manual.
2
ABC123456789
3
2
1
2152A.eps
R
1 Sample container
2 Sample container bar code label
3 Rack slot
4. After you have placed all of the sample containers in the rack, you can load the
rack onto the onload area of the SPU for either routine or STAT processing. For
more information, see the Loading a Rack for Routine Processing or see the
Loading a STAT Rack procedures.
Using Insert An insert cup has a smaller dead volume than a sample tube. You may need to transfer
Cups with sample from a sample tube to an insert cup to meet the total sample volume required
Sample Tubes for testing.
Use this procedure if you are using an insert cup on the instrument.
WARNING
You will come in contact with potentially infectious materials during this
procedure. Handle and dispose of biohazard materials according to proper
laboratory procedures. Proper hand, eye, and facial protection is required.
1. Transfer the required sample volume from the sample tube into an insert cup. Be
sure that there is a sufficient volume of sample to process all of the requested
tests. To calculate the correct sample volume, see Section A.2: Sample
Containers.
NOTE
Carefully pipette small sample volumes into the insert cup to prevent air bubbles
from forming at the meniscus.
2. Place the insert cup back into the original sample tube.
3. Place the sample tube into a rack. For more information, see the Placing Sample
Containers in Racks procedure.
2
ABC123456789
2141A.eps
1 Insert cup
2 Sample tube
Loading a Use this procedure to load a rack containing routine samples onto the instrument. If
Rack for you are loading a rack containing STAT samples, see the Loading a STAT Rack
Routine procedure. You load racks onto the onload area of the SPU.
Processing WARNINGS
• You will come in contact with potentially infectious materials during this
procedure. Handle and dispose of biohazard materials according to proper
laboratory procedures. Proper hand, eye, and facial protection is required.
• Do not place your hand inside of the sample aliquot station.
NOTES
• Each loaded rack must have a rack ID that is set up for your system. A rack
loaded with an ID that was not set up will pass through the system without
being processed. For more information, see the Setting Up Rack IDs
procedure in Section 2.3: Supplies Setup of the Reference Manual.
• Be sure that the rack ID bar code label is in good condition and is properly
affixed to the rack. For more information, see Section 2.3: Supplies Setup in
the Reference Manual.
• You can reuse a rack after it is aliquoted by the system. You cannot have two
racks with the same rack ID in the onload area of the SPU.
• Do not load a rack while the utility routine is running.
WARNING
In the following step, if your sample containers do not have bar code labels, be
sure to place the sample containers in the positions you specified on the Test
Requests screen. If the containers are not placed in the designated positions, the
system will report incorrect test results.
1. Place the sample containers for the requested tests in a rack. For more
information, see the Placing Sample Containers in Racks procedure.
On: The system is pushing one or more racks into position and the cover is locked.
Select the Routine button and wait until the system unlocks the cover.
The system aliquots the current sample container, moves the onload pusher to the
home position, and unlocks the SPU cover.
3
1 4
2
LOAD LOAD
ROUTINE STAT
5
2182C.eps
1 SPU cover
2 Routine button
3 Load lights
4 STAT button
5 Onload pusher motion indicator
CAUTION
In the following step, be sure the bar code label is on the right side of the
rack. If the rack is loaded incorrectly, it may jam on the SPU or pass through
without being aliquoted.
4. Place the rack in the onload area of the SPU. Place the rack to the right of any
racks in the onload area. Be sure the bar code label is on the right side of the rack,
and the rack arms rest over the guide rails on either side of the onload area.
(Optional) If you have additional racks to load, repeat step 4.
4
6
3
2
5
1
51
0
2
4
3
1
2150B.eps
1 Guide rails
2 Rack correctly loaded, with bar code facing to the right
3 Rack ID label side of the rack, with the rack arm on the guide rail
4 Front onload area of the SPU
5 End of the onload sensor
6 Presentation tray gate in the sample presentation area (gate is up)
Note: The tray gate shown is for the UniCel DxI 800 instrument.
7 Sample aliquot station (with the sample presentation tray inside)
Yes
3000C.wmf
Figure 3-17 Loading a Rack for Routine Processing Flowchart
Loading a Use this procedure to load a rack that contains a STAT sample. You can load a STAT
STAT Rack rack ahead of racks previously loaded for routine sample processing.
WARNINGS
• You will come in contact with potentially infectious materials during this
procedure. Handle and dispose of biohazard materials according to proper
laboratory procedures. Proper hand, eye, and facial protection is required.
• Do not place your hand inside of the sample aliquot station.
NOTES
• Each loaded rack must have a rack ID that is set up for your system. A rack
loaded with an ID that was not set up will pass through the system without
being processed. For more information, see the Setting Up Rack IDs
procedure in Section 2.3: Supplies Setup of the Reference Manual.
• Be sure that the rack ID bar code label is in good condition and is properly
affixed to the rack. For more information, see Section 2.3: Supplies Setup in
the Reference Manual procedure.
• You can reuse a rack after it is aliquoted by the system. You cannot have two
racks with the same rack ID in the onload area of the SPU.
• Confirm that all tests requested for a STAT sample have an active assay
calibration listed on the Calibration screen. For more information, see the
Calibration Screen topic in Section 5.1: Assay Calibration Overview.
• Do not load a rack while the utility routine is running.
WARNING
In the following step, if your sample containers do not have bar code labels, be
sure to place the sample containers in the positions you specified on the Test
Requests screen. If the containers are not placed in the designated positions, the
system will report incorrect test results.
1. Place the sample containers for the requested tests in a rack. For more
information, see the Placing Sample Containers in Racks topic.
NOTES
• The system assigns the Requested status to interrupted LIS test requests, and
lists the requests on the Work Pending screen (see Figure 3-5 in Section 3.2:
LIS Requests and Work Pending). The tests are removed from the work
pending list when the system scans the rack again.
• The system assigns the Requested status to interrupted manual test requests,
and lists the requests on the Requested view of the Sample Manager screen.
4. Gently push any racks to the right, moving them from the sample aliquot station
to the onload area (see Figure 3-16).
NOTE
Do not remove the racks from the instrument unless it is necessary.
CAUTION
In the following steps, be sure the bar code label is on the right side of the
rack. If the rack is loaded incorrectly, it may jam on the SPU or pass through
without being aliquoted.
5. Place the rack containing the STAT sample ahead of the other racks in the onload
area. Be sure the bar code label is on the right side of the rack, and the rack arms
rest over the guide rails on either side of the onload area (see Figure 3-16).
6. If you interrupted sample processing to load the STAT rack, place the rack
containing the interrupted samples to the right of any STAT racks and to the left
of the routine racks in the onload area of the SPU.
NOTE
If the Console icon on the system status panel and the Amber system indicator
light begin to blink, there may be a rack remaining in the sample aliquot station.
Open the SPU cover, remove the rack, and close the cover again.
3001C.wmf
Figure 3-18 Loading a STAT Rack Flowchart
Unloading a After each sample in a rack is aliquoted, the system moves the rack to the offload area
Rack of the SPU. The offload area of the UniCel DxI 800 instrument holds up to 31 racks,
filling across the front of the instrument and then to the side. The offload area of the
UniCel DxI 600 instrument holds up to 15 racks. If the offload area becomes full, the
system records an event in the Event Log. The system continues to process tests, but
will not process any new test requests. You must unload racks to continue sample
processing.
WARNING
You will come in contact with potentially infectious materials during this
procedure. Handle and dispose of biohazard materials according to proper
laboratory procedures. Proper hand, eye, and facial protection is required.
5
1
2151A.eps
2. If there are racks present in one or both offload areas, open the appropriate
covers.
• SPU cover - Open this cover to unload racks from the front off load area. If the
onload pusher motion indicator (see Figure 3-15) is lit, the system is working
with a rack and the SPU cover is locked. You can unload racks when the system
unlocks the cover. If the SPU cover does not open, troubleshoot according to the
information displayed in the Event Log. For more information, see Section 5.4:
Event-Specific Troubleshooting in the Reference Manual.
• Side offload area cover - Open this cover to unload racks from the side offload
area cover. You can unload racks from this area at any time.
3. Unload all racks from the front and side offload areas.
During sample processing, the UniCel DxI system provides the following status
information.
• Individual sample status, or the status of all samples in a rack
• Test status
The system displays rack and sample status information in the Status column of two
Sample Manager screen views:
• In Progress view
• Daily view
The system displays the test status in the Results column of the Test Results screen.
For more information, see Section 4.1: Test Results Overview.
Sample You can monitor the status of individual samples in the Status column on the In
Status Progress or Daily views of the Sample Manager screen. For more information, see the
Sample Manager Screen topic in Section 3.1: Sample Manager Overview.
Test Status You can monitor individual test status until a result is generated by looking at the
Result column on the Test Results screen. For more information about this screen, see
Section 4.3: Reviewing Test Results.
While a manually entered rack is in the onload area of the sample presentation unit
(SPU), you can access and edit the test request information on the Test Requests
screen. From the Test Requests screen, you can modify manually entered test requests
in the following ways:
• Cancel a request
• Edit the sample information fields
• Add and remove tests
• Delete a selected sample or a rack of samples
When a rack is in the sample aliquot station, you can no longer edit the test requests.
NOTE
Do not edit test requests downloaded from a laboratory information system (LIS)
on the Test Requests screen. Edit these test requests at the LIS instead.
Cancelling the Use this procedure to cancel the test requests in a manually entered rack that has not
Test Requests been scanned.
on a Rack System Mode: Any Mode
1. From the Requested view of the Sample Manager screen, select the rack that
contains the test requests that you will cancel.
(Optional) From the Daily view, select an individual sample from the rack that
contains the test requests that you will cancel.
2. Select Edit Request F2 to display the rack on the Test Requests screen.
3. Select Cancel Request F6.
A confirmation window is displayed.
4. Select Yes F1 to cancel the entire rack of test requests and return to the Sample
Manager screen.
Editing a Use this procedure to edit a manual test request for a patient sample in a rack that has
Manual Test not been scanned.
Request System Mode: Any Mode
1. From the Requested view of the Sample Manager screen, select the rack that
contains the test requests that you intend to edit.
(Optional) From the Daily view, select an individual sample from the rack that
contains the test requests that you intend to edit.
2. Select Edit Request F2 to display the rack on the Test Requests screen.
NOTE
If you edit the sample ID, the system deletes all of the sample information and the
requested tests.
5. Edit the test list using the Test buttons. The system adds or deletes tests in
increments of one.
• To add a test, select a Test button.
• To delete a test, select the test in the Tests Requested for Sample X list and then
select Remove.
(Optional) To edit additional test requests on the same rack, repeat steps 3-5.
Sample Manager
From the Requested view,
select the rack that contains
the test request that you
intend to edit
Edit Request F2
Test Requests
No
Done editing the
rack?
Yes
Test Requests
2097C.svg
Figure 3-22 Editing a Manual Test Request Flowchart
Deleting a When you delete a manually entered sample from the Test Requests screen, all of the
Sample from sample information for the container is deleted from the screen. You can delete a
the Test sample until the system scans its sample container.
Requests Use this procedure to delete a manually entered sample from the Test Requests
Screen screen.
2. Select Edit Request F2 to display the rack on the Test Requests screen.
• To delete all of the samples in the rack, select Clear All Samples F7.
Sample Manager
Test Requests
Deleting all
No Select the sample that you
samples in the
intend to delete
rack?
Yes
Test Requests
2098C.svg
Figure 3-23 Deleting a Sample from the Test Requests Screen Flowchart
Changing a Use this procedure to change a rack ID for a manually entered rack.
Rack ID
NOTE
You cannot use a rack ID that is assigned to another rack in the onload area of the
SPU.
NOTE
If you are changing the rack ID for a calibration that requires more than four
containers, both Rack buttons display on the screen. Select Rack ID F1, and then
select First F1 to change the first rack ID, or select Second F2 to change the
second rack ID.
Unexpected or abnormal patient sample results can occur for various reasons. To rule
out and fix the most significant problems first, troubleshoot each failure in the
following order:
1. Identify event log errors. Contact Technical Support if you need help
troubleshooting event log errors. For more information, see Chapter 5: Event Log
in the Reference Manual.
2. Troubleshoot test result flags. For more information, see Section 4.3: Reviewing
Test Results.
3. Check the quality control (QC) results. If the results failed, troubleshoot
according to the instructions documented in Section 6.7: QC Troubleshooting
Overview.
If you are unable to identify the cause of your problem, or if the problem persists after
performing the specified troubleshooting procedures, contact Technical Support for
assistance.
Troubleshooting Use the following table to troubleshoot patient sample results that do not seem
Unexpected correct.
Patient Sample
Results
Possible Causes Corrective Action
Incorrect patient sample type Refer to the reagent instructions for use for the
appropriate sample types for each assay.
Patient sample handled incorrectly Be sure that sample is stored and handled according to
proper laboratory procedures. Refer to individual
assay reagent instructions for use for proper sample
handling instructions.
Microclots present in the serum 1. Spin the sample again or use a serum filter to
sample separate the serum from any microclots.
2. Repeat the test.
Quantity of patient sample not 1. Calculate the correct volume of patient sample
sufficient for testing (QNS result flag needed for the number of replicates and type of
and event) sample container used (see Section A.2).
2. Pipette the amount of patient sample calculated in
step 1 into the appropriate sample container.
3. Be sure the rack and rack ID are correct (see
Section 2.3 in the Reference Manual).
4. Repeat the test.
5. If the test fails again, contact Technical Support
for assistance.
Expired reagent pack 1. Check the reagent pack expiration date on the
Reagent Inventory screen (see Section 2.8).
2. Unload the expired reagent pack and load a new
one (see Section 2.7).
• If the lot number of the new reagent pack is
different than the expired pack, recalibrate the
assay.
3. Repeat the test.
Reagent pack unstable or 1. Unload the reagent pack and load a new reagent
contaminated due to improper pack (see Section 2.7).
handling • If the lot number of the new reagent pack is
different than the removed pack, you must
recalibrate the assay.
2. Repeat the test.
Expired assay calibration 1. Check the Reagent Pack button on the Reagent
Supplies screen (see Section 2.7). If the Lot
Number field is red, the calibration has expired.
2. View the test results (see Section 4.3). Any test
results calculated with an expired calibration will
be flagged.
3. Recalibrate the assay.
4. Repeat the test.
Partial bottles of substrate combined 1. Change the in-use substrate bottle and prime the
substrate (see Section 2.2).
2. Repeat the test
Wrong sample container and/or rack Load the samples using appropriate sample containers
and racks. Be sure that the sample containers are in a
rack with the correct rack ID (see Section 2.3:
Supplies Setup in the Reference Manual).
Sample container missing or placed in 1. Be sure that the sample containers are in the
incorrect position in the rack proper order in the rack.
2. Repeat the test.
Troubleshooting Use the following table to troubleshoot patient sample results that demonstrate poor
Patient Sample precision, value shifts, or values that increase over the course of a run.
Results with
Increased
Possible Causes Corrective Action
Variability
Expired reagent pack 1. Check the reagent pack expiration date on the
Reagent Inventory screen (see Section 2.8).
2. Unload the expired reagent pack and load a new
one (see Section 2.7).
• If the lot number of the new reagent pack is
different than the expired pack, recalibrate the
assay.
3. Repeat the test.
Expired assay calibration 1. Check the Reagent Pack button on the Reagent
Supplies screen (see Section 2.7). If the Lot
Number field is red, the calibration has expired.
2. View the test results (see Section 4.3). Any test
results calculated with an expired calibration will
be flagged.
3. Recalibrate the assay.
4. Repeat the test.
4 Test Results
4.1 Test Results Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Test results are data that the UniCel DxI system creates and stores for every test the
system runs. To view test results, display the Test Results screen. To view a specific
result or a set of specific results, apply a test result filter to the screen or use the find
feature to limit or change the results displayed on the screen.
For each test result, you can view sample details such as patient demographics,
sample comments, and flag descriptions. For results with flags, you can view
troubleshooting information. For more information, see the Viewing Sample Details
procedure or the Test Result Flags and Troubleshooting procedure in Section 4.3:
Reviewing Test Results.
Upon reviewing a patient or quality control result, you might decide to rerun the test
from the Test Result screen. After reviewing any result, you can manually delete the
result from the system database. However, most results are deleted using the
Auto-Delete function. For more information, see the Setting Up the Auto-Delete
Feature procedure in Section 2.6: Data Management of the Reference Manual.
Test results can be saved for your laboratory archives in several ways:
• Send results to the LIS
• Copy results to a disk and import them into a PC-based spreadsheet program
• Print results in one of several report formats
View results
(Optional)
Configure screen (to present
results according to your lab
needs)
(Optional)
Define filter (to present
results according to your lab
needs)
(Optional)
Is the result No
Archive, delete, or print
flagged?
results
Yes
(Optional)
Archive, delete, or print
results
2080B.wmf
Figure 4-1 Test Results Overview Flowchart
2036B.bmp
Filter (Field) Displays the selected filter and sort order. Each time you change the filter, it
becomes the default filter.
NOTE
If you display the Test Results screen from the Test Requests screen or from the
Daily View or Exceptions View of the Sample Manager screen, the Daily View
filter is the default filter.
Entries (Field) Displays the number of test results that meet the selected filter parameters.
Displays a visual cue (+) when more results are available since the filter was applied.
To see the results, refresh the screen.
For more information, see the Refreshing the Test Results Screen procedure.
Selected Result (Row) Displays an Arrow button at the beginning of the selected row. The selected
row is highlighted. If more than one row is selected, all rows are highlighted and the
Arrow button is displayed in the last row selected. For more information, see the
Selecting Test Results procedure.
Patient ID (Optional column) Displays the patient ID, if one exists, for the patient sample.
Sample ID (Required column) Displays the sample ID that is associated with the result.
Rack (Optional column) Displays the number of the rack that the sample is in.
Test Name (Required column) Displays the test name that is associated with the result.
Result (Required column) Displays the result of the test, including the value, units, and
interpretation (if applicable).
• Results described below are displayed with bold, red type.
- Patients:
• Results associated with any of the following flags: ORL, ORH, CRL, CRH, GRY,
OVR
• Results preceded by < (Results below the assay’s lowest reportable result)
• Qualitative, confirmatory, or semi-quantitative results interpreted as Positive,
Equivocal, Reactive, Confirmed, or Class 1, 2, 3, 4, 5, or 6
- Controls:
• Results associated with an OVR or QCF flag
• Results preceded by < (Results below the assay’s lowest reportable result)
- Calibrators: All Reactive calibrators
• If a test result is below the lowest reportable result, the result is displayed as <x,
where x is the lowest reportable result. Semi-quantitative, onboard dilution results
below the lowest reportable result are reported as, No Interpretation <x.
• For onboard dilution test results, the lowest reportable result is defined in the APF,
and is typically 85% of the undiluted parent assay’s highest calibrator
concentration. In cases where the test result is below the lowest reportable result,
run the undiluted parent test instead. For more information, see Onboard Dilutions
in Section 1.3: Sample Processing Theory of the Reference Manual.
• If test processing is incomplete, the test status is displayed. For more information,
see the Test Status topic in Section 3.5: Monitoring Sample Progress.
Pipettor (Optional column) Displays the reagent pipettor on which the test was run.
Flags (Required column) Displays the flags associated with the result. If there are no flags,
the field is blank.
LIS (Optional column) Displays the LIS status of the test result. Values are:
• <blank> - The result has not yet been calculated, or the result has just been
calculated, but the system has not updated the LIS status yet. Alternatively, the
result is not intended for the LIS (a calibrator or maintenance result).
• Sending - The system is sending the result to the LIS.
• Accepted - The system sent the result and the LIS received it.
• Not Sent - The system did not send the result to the LIS because the result did not
meet the Auto-Verify criteria. Alternatively, the system did not send the result to
the LIS because one or more of the following settings in the LIS Setup window
have been made:
- The Local LIS Interface field is set to Off.
- The Global LIS Interface field is set to Off.
- The Auto Send to LIS field is set to Off.
• Failed - The system sent the result, but the LIS did not receive it. Resend the result.
For more information, see the Sending Test Results to the LIS procedure in Section
4.4: Managing Test Results.
• Idle - The system attempted to send the result to the LIS while the global LIS
interface was offline. Set the global LIS interface to On to automatically send
results with this status to the LIS. For more information, see the Setting Up the LIS
Interface procedure in Section 2.5: LIS Setup of the Reference Manual.
Rgt. Lot (Optional column - not shown) For systems running with software version 5.2 or
higher, this column displays the reagent pack lot number that was used for generating
the test result.
Pack S/N (Optional column - not shown) For systems running with software version 5.2 or
higher, this column displays the serial number of the reagent pack lot that was used for
generating the test result.
RLU (Optional column - not shown) Displays the RLUs for the test result.
Comp. Time (Optional column - not shown) Displays the date and completion time of the test.
Verify (Optional column - not shown) Displays the status of the test results. When the LIS is
set up to send only auto-verified results to the LIS, the system compares the test
results to the LIS ranges you defined. For more information, see Section 2.5: LIS
Setup. Also see Section 3.4: Ranges Setup in the Reference Manual. Values are:
• <blank> - The system has not compared the test result to the set LIS ranges. If the
column is blank, it usually means the result has not been calculated yet.
• Pass - The test result was within the set LIS ranges.
• Fail - The test result was not within the set LIS ranges and the system cannot send it
to the LIS. The system displays the word Fail in red.
• Not Verified - The LIS is not set up to verify results.
Instrument (Optional column - not shown) Displays the ID of the instrument on which the test
was run.
Filter (Button) Select to apply, edit, or delete test result filters. The system displays the
F1 selected filter and sort order in the Filter field. For more information, see
Section 4.2: Test Result Filters.
Find (Button) Select to search the filtered results for a specific test result. For more
F2 information, see the Finding a Test Result procedure in Section 4.3: Reviewing Test
Results.
Rerun Test (Button) Select to run the selected test again. If sufficient sample to rerun the test
F3 remains on the instrument, the test is automatically assigned and runs at the next
opportunity. For more information, see the Rerunning a Patient Test procedure in
Section 4.3: Reviewing Test Results, or see the Rerunning a QC Test procedure in
Section 6.7: QC Troubleshooting Overview. If there is not enough sample on the
instrument to rerun the test, the system displays the test request on the Work Pending
screen. For more information, see Section 3.2: LIS Requests and Work Pending.
Sample Details (Button) Select to display sample related patient demographic information, comments
F4 entered by the operator, test result flags, and flag definitions. For results with flags,
troubleshooting information can be accessed from the Sample Details window. For
more information, see the Viewing Sample Details procedure in Section 4.3:
Reviewing Test Results.
Send to LIS (Button) Select to send test results to the LIS. For more information, see the Sending
F5 Test Results to the LIS procedure in Section 4.4: Managing Test Results.
Copy to Disk (Button) Select to copy all test results or the selected test results to a disk. For more
F6 information, see the Copying Test Results to a Disk or Flash Drive procedure
in Section 4.4: Managing Test Results.
Select to manually print the Continuous Sample Report for tests that are completed,
but not yet printed.
For more information, see Section 4.5: Test Result Reports.
More Options (Button) Select to display a menu of the following options:
F8 Refresh Screen F1
Select to update the information displayed on the Test Result screen and display new
results that match the filter parameters. For more information, see the Refreshing the
Test Results Screen procedure.
Configure Screen F2
Select to change the column order or the columns displayed on the Test Result screen.
For more information, see the Configuring the Test Results Screen procedure.
Delete Data F3
Select to remove test result data from the system database. For more information, see
the Deleting a Test Result procedure in Section 4.4: Managing Test Results.
NOTE
You need the system password to use this feature.
Refreshing When a plus sign (+) is displayed in the Entries field of the Test Results screen,
the Test refresh the screen.
Results System Mode: Any Mode
Screen 1. From the Test Results screen, select More Options F8.
Configuring Use this procedure to add or rearrange result information on the Test Result screen.
the Test The Sample ID, Test Name, Result, and Flags columns are required, but you can
Results change their order on the screen.
Screen The Patient ID column is displayed by default for most filters. However, for the
Failed Auto-Verify and Not Sent to LIS filter, the system displays the Verify column
and does not display the Patient ID column.
System Mode: Any Mode
1. From the Test Results screen, select More Options F8, then Configure
Screen F2.
The Configure Screen window displays the columns in the order they display on the
Test Results screen.
Column Order
Buttons
0602A.bmp
Column Order (Row) Displays the columns currently displayed on the Test Results screen, in their
respective order.
Optional Columns (Boxes) Select the optional columns to include on the Test Results screen.
Column Order (Buttons) Select the right or left Arrow button to rearrange a selected column in the
Buttons Column Order row. If you do not select a column before selecting an Arrow button,
the first column is selected by default.
2. In the Optional Columns list, select the columns you want to add to the Test
Result screen from the following options: Patient ID, Rack, Verify, Comp. Time,
LIS, Instrument, RLU, and Pipettor. For systems running with software
version 5.2 or higher, you can also select the Rgt. Lot and Pack S/N options.
You can add optional columns until you reach the maximum width of the screen.
(Optional) To rearrange the order of the columns, select a column in the Column
Order row and use the right or left column order button to move it.
3. Select OK F1.
The Test Result screen displays the columns in the order you selected.
Selecting Test Select individual results to limit the data that prints on a report, to rerun a specific test,
Results to limit the data that is copied to a disk or to an LIS, to limit the data that is deleted, or
to investigate information associated with specific results.
Follow the guidelines below to select one or more test results on the Test Result
screen.
System Mode: Any Mode
1. From the Test Results screen, locate a single test result or a group of test results.
To locate a single test result, see the Finding a Test Result procedure in Section 4.3:
Reviewing Test Results. To locate a group of test results, see the Applying a Test
Result Filter procedure in Section 4.2: Test Result Filters.
NOTES
• To view the results that are out of the current viewing area, use the scroll bar.
• To go to the beginning or end of the test result list, press the [Ctrl] key while
you press the [Home] or [End] key, respectively.
2. Select the result or the group of results by taking one of the following actions.
• To select one result:
Select a row by touching it. By default, one result is selected on the Test Result
screen, as indicated by the Arrow button to the left of the row. When you select a new
result, the Arrow button is displayed to the left of the new row you selected and is
cleared from the previous selection.
• To select multiple contiguous results:
Select the first result, press and hold the [Shift] key, and then select the last of the
contiguous results.
Example
To select the first 5 results, select the first result, press and hold the [Shift] key,
and then select the fifth result. Release the [Shift] key. All five results are
selected.
Example
To select the first, second, and fourth results shown on the Test Results screen,
select the first result, then press the [Ctrl] key and select the other results. Release
the [Ctrl] key. Each of the rows is selected.
NOTES
• If you select a group of results and then select Rerun Test F3 or
Sample Details F4, only the result indicated with the Arrow button is rerun
or displayed in the Details window.
• If you select a group of results and then select Send to LIS F5 or Copy to
Disk F6, you must select Current Selection in the window to send or copy
all selected results. Otherwise, only the result with the Arrow button is sent or
copied.
(Optional) To remove a selection from a group of selected results, press the [Ctrl]
key while you select the result you want cleared.
You can create and edit test result filters for the Test Results screen, or apply
system-defined filters.
To create new filters, you can define all of the parameters for the filter, or you can edit
an existing filter. If you edit a system-defined filter, and you want to save the changes,
you must save the new filter with a different name. To view a list of the
system-defined filters with their default sort orders, see the Applying a Test Result
Filter procedure.
For more information, see the Filters and Sort Order topic in Section 1.3: Software
Overview.
Applying a Use this procedure to apply one of the available test result filters.
Test Result
The following test result filters are system defined:
Filter
Filter 1st Sort 2nd Sort
2037A.bmp
Examples
• A filter applied to results according to Rack ID in ascending order displays
results from rack 1 at the top of the screen, followed by results from rack 2
and other racks in ascending order.
• A filter applied to results according to Test Name in ascending order
displays Ferritin results closer to the top of the screen than TPOAb results.
Description (Field) Displays a description of the selected filter.
New Filter (Button) Select to define a new filter.
F2
Edit Filter (Button) Select to edit the parameters of the selected filter. You can then save the
F3 changes as a new filter or apply the edited filter without saving it.
Delete Filter (Button) Select to delete the laboratory-defined filter selected in the Filter list.
F4
NOTE
You cannot delete a system-defined filter.
(Optional) To change the sort order, select an option from each of the sort order
lists.
(Optional) To display the results in descending order, clear the Ascending box.
3. Select OK F1.
The system displays all test results that meet the selected filter parameters on the Test
Results screen. The applied filter becomes the default filter until you apply a different
saved filter.
NOTE
If you select the system-defined filters All Patient Samples Loaded Between...
or All Samples Loaded Between..., or a laboratory-defined filter that asks for
the date and time, the system prompts you to enter the date range.
Defining or Use this procedure to define a new filter or edit an existing filter and apply it to the
Editing a Test Test Results screen.
Result Filter
NOTE
The New Filter and Edit Filter windows are the same, but the New Filter fields
default to all and New Filter check boxes are selected, except for the Ask for
Date & Time box.
NOTE
Every new filter has the default sort orders Sample ID and Test Name.
0605B.bmp
Filter Name (Field) Displays the name of the filter selected in the Filter window. You can delete
the name and enter a new name to create a new filter.
Description (Field) Displays the description of the filter selected in the Filter window. You can
delete the description and enter a new description.
Ask for Date & (Box) Select to have the system ask for date and time parameters each time you
Time apply the filter.
Instruments (Box) Displays your instrument box selected by default. Includes your instrument
in the filter when the box is selected.
Test Result Status (Boxes) Select one or more of the test result statuses to include in the filter.
LIS State (List) Select the LIS state to include in your filter.
Example
If you select > from the list, and enter 1011 in the field, then the filter includes
results with sample IDs > 1011.
Test Name (List) Select the enabled tests to include in the filter. The list contains all enabled
test names.
Auto-Verify (List) Select the status of the test results to include in the filter.
Values are: unverified, pass, fail, or all results.
Rack ID (List) Select the rack IDs to include in the filter. The list contains all racks shown
in the Requested View of the Sample Manager screen.
Patient ID (List) Select a relationship operator from the list.
(Field) For list values other than (all), enter the patient ID that completes the
relational statement for the samples to include in the filter.
Example
If you select = from the list, and enter 12345678 in the field, then the filter
includes results for patient ID 12345678.
Result (List) Select a relationship operator from the list.
(Fields) For values other than (all), enter a numeric value that completes the
relational statement for the results to include in the filter in the first field. For the
outside and between list values, enter a numeric value in the second field.
Example
If you set the value in the first box to >, and the value in the next box is 4.00,
then the filter includes results > 4.00.
NOTE
Because numerical results are rounded and truncated for display, the =
relationship operator is not often useful in a result filter. The result
1.000020 ng/ml is displayed as 1.00 ng/ml, but is not included by using the
filter = 1.00.
Population (Boxes) Select one or more of the result populations to include in the filter.
OK (Button) Select to save the edited filter with the same name and overwrite the
F1 previous filter. This is only allowed with laboratory-defined filters.
Save As (Button) Select to save the edited filter as a new filter with a new name. For
F2 laboratory-defined filters, you can keep the same name and overwrite the previous
filter. When you save the filter, it is applied.
Clear Parameters (Button) Select to clear all of the fields and reset the parameters to the current filter
F3 parameters.
3. Locate and change the parameters you need to enter or edit. For descriptions of
the window parameters, see Figure 4-6.
4. Select OK F1.
The system prompts you to save the filter.
NOTE
If you select Cancel F8, your changes are not saved and the filter is not applied.
6. If you selected the Ask for Date & Time box in the Edit Filter window, the
Specify a Date Range window is displayed.
Enter the appropriate information. The date format follows the format you set up on
the System Setup screen. The system displays a message if the incorrect format is
used. For more information, see the Setting the Date and Time Formats procedure in
Section 2.2: System Setup of the Reference Manual.
7. Select OK F1.
The system displays the test results that meet the filter parameters on the Test Results
screen.
Test Results
Filter F1
Filter
Filter/Edit Filter
OK F1
Yes
Save the filter? Yes F1
OK F1
Specify the date and time (if
required), the system
applies the filter,
but does not save it
The system saves and
applies the filter
2065B.wmf
Figure 4-7 Defining or Editing a Test Result Filter Flowchart
Deleting a Use this procedure to delete laboratory-defined filters. You cannot delete
Test Result system-defined filters.
Filter System Mode: Any Mode
1. From the Test Results screen, select Filter F1 to display the Filter window.
3. Select Delete Filter F4. This button is not available for system-defined filters.
The system displays a confirmation message.
The Test Results screen provides a wide variety of information about test results. To
examine the available information about test results, first locate the results. If you
cannot immediately locate a result, use one or more of the following features:
• Filter
• Find
• Refresh Screen
Generally, more information is available about results than can fit on the Test Result
screen at one time. The screen can be configured to display the information you
consider most important. For more information, see the Configuring the Test Results
Screen topic in Section 4.1: Test Results Overview.
Flags are codes that the UniCel DxI system associates with test results when a special
circumstance requires attention. Flags, flag explanations, and sample details can be
viewed from the Sample Details window.
When reviewing results for a particular sample, you can rerun any test previously run
on that sample.
NOTE
Manually requested patient test results are evaluated against default ranges, even
if your laboratory has established reference, critical, or LIS ranges. For more
information, see Section 3.4: Ranges Setup in the Reference Manual.
Finding a Test Use this procedure to find a specific test result on the Test Result screen.
Result
NOTE
This procedure finds only test results that fit the current filter parameters. If you
have difficulty finding a specific test result, apply a different filter.
2. In the Find window, enter information in one of the Sample ID, Patient ID, Rack
ID, or Test Name fields for the result you want to find.
NOTE
For the Sample ID and Patient ID field, you do not need to enter complete
information. The Find feature locates results based on the first characters you
enter in the field.
(Optional) To change the direction of the search, select the Up or Down option.
If you select Down, the system begins the search at the currently selected result and
continues down the Test Result screen. If you select Up, the system begins the search
at the currently selected result and continues up the Test Result screen.
0609A.bmp
Sample ID (Field) Enter all or part of the sample ID to find a result using only its sample ID.
Example
If you are searching for sample ID 123456789, you can enter 1234. The system
finds the first sample ID starting with 1234.
Rack ID (List) Select from a list of rack IDs for which there are results to find a result using
only its rack ID.
Patient ID (Field) Enter all or part of the patient ID to find a result using only its patient ID.
Example
If you are searching for patient ID 123456789, you can enter 1234. The system
finds the first patient ID starting with 1234.
Test Name (List) Select from a list of test names or enter a test name to find a result using only
its test name.
Direction (Option) Select Up or Down to change the direction of the search. If you select
Down, the system begins the search at the currently selected result and continues
down the Test Result screen. If you select Up, the system begins the search at the
currently selected result and continues up the Test Result screen.
When a search reaches the top or bottom of the Test Result screen, the search
continues from the opposite end.
Find Next (Button) Select to find the next result that matches the information in the field.
F1
3. Select Find Next F1. The closest match to the search parameters you entered
becomes the selected result.
NOTES
• If the system does not find a match, a message is displayed in the status line
of the Test Result screen.
• If any of the filtered results have been deleted, a message is displayed that
directs you to refresh the screen. For more information, see the Refreshing
the Test Results Screen procedure in Section 4.1: Test Results Overview.
Repeat step 3.
(Optional) To find the next result that matches the information in the field, select
Find Next F1.
Test Results
Find F2
Find
(Optional)
Select a search direction, up
or down
Find Next F1
Is the result No
found?
Yes
Cancel F8
2066B.wmf
Figure 4-9 Finding a Test Result Flowchart
Viewing Use this procedure to view explanations about result flags or details about the sample
Sample a result is associated with, including demographic information and comments.
Details System Mode: Any Mode
1. From the Test Results screen, locate and select a single test result.
For more information on locating a result, see the Finding a Test Result procedure, or
the Applying a Test Result Filter procedure in Section 4.2: Test Result Filters. For
more information on selecting a result, see the Selecting Test Results procedure in
Section 4.1: Test Results Overview.
NOTE
If you select a group of results and then select Sample Details F4, only the result
indicated with Arrow button is displayed in the Details window.
2043B.bmp
Patient ID (Field) Displays the patient ID for the patient sample. The Patient ID can be
entered on the Test Request screen or sent from the LIS. The Patient ID is required
if it is sent from the LIS.
First Name (Field) Displays the first name of the patient. This information is sent from the LIS.
Last Name (Field) Displays the last name of the patient. This information is sent from the LIS.
Sex (Field) Displays the sex of the patient. This information is sent from the LIS.
Birth Date (Field) Displays the birth date of the patient. This information is sent from the LIS.
Doctor Name (Field) Displays the name of the patient's doctor. This information is sent from the
LIS.
Sample ID (Field) Displays the sample ID.
Sample Type (Field) Displays the type of fluid in the sample container.
Offboard Dilution (Field) Displays the dilution factor for sample dilutions prepared manually.
Onboard Dilution (Field) Displays the dilution factor for onboard dilution assays and special dilution
assays.
For more information, see Onboard Dilutions in Section 1.3: Sample Processing
Theory of the Reference Manual.
Test Name (Field) Displays the name of the test that was run on the patient sample.
Reference Range (Field) Displays the reference range used to evaluate the test. For more
information, see Section 3.4: Ranges Setup in the Reference Manual.
Result (Field) Displays the results of the test run on the sample.
Sample Comment (Field) Displays any comments about the sample entered from the Test Request
screen.
Flags (List) Displays any flags associated with the result and a description of each flag.
Troubleshoot (Button) Displays troubleshooting Help about flags.
F2
NOTE
This button is not available for results without flags.
3. Select OK F1.
Test Result The following tables provide troubleshooting information associated with test result
Flags and flags that are fatal (no calculated result) or non-fatal (calculated result, but a condition
Troubleshooting exists for that result). Use the information to identify and correct assay related
problems. This information is also provided online when you select Troubleshoot F2
from the Sample Details window. For more information, see the Viewing Sample
Details procedure.
AEX The aliquoted sample cannot be used for 1. Review the Event Log for error events with a similar
one or more of the following reasons: date and time to this event (see Section 5.2 in the
• An error occurred when the aliquot Reference Manual). If events are found, troubleshoot
was dispensed into the vessel. according to available information or contact Technical
Support.
• The aliquot has been on board the
instrument for up to three hours and 2. Make sure that required supplies are on board before
has expired. running the test again.
• The last test taken from the aliquot 3. Check the test status in the Result column of the Test
was completed one hour ago. Result screen and take one of the following actions:
• For calibrator and maintenance samples with the test
status Cancelled: Request the test again and load a
fresh quantity of sample.
• For patient and QC samples with the test status
Requested: Load a fresh quantity of sample. The test
is run automatically.
CCR A result could not be calculated because: 1. Take one of the following actions:
• One of the tests included in a • For results other than derived results, skip to step 2.
derived result formula (see • For derived results, review each test result used in the
Section 3.6: Derived Result Setup in derived result formula. If a result failed, troubleshoot
the Reference Manual) did not according to the flag for that result.
produce a result.
2. Review the Event Log for error events with a similar
• A confirmatory test result could not date and time to this event (see Section 5.2 in the
be calculated. Usually this occurs Reference Manual). If events are found, troubleshoot
because the samples for the according to available information or contact Technical
qualitative and confirmatory tests Support.
were not aspirated at the same time
3. Repeat the test.
(see Section 1.5: Confirmatory
Assay Theory in the Reference • For a derived result, repeat all tests included in the
Manual). derived result formula.
• Another error prevented the system • For a confirmatory result, repeat the confirmatory and
from calculating a result. qualitative tests.
CLT An obstruction was detected in the sample Check the test status in the Result column of the Test Result
tube before aliquoting or in the RV during screen.
processing. • For patient and QC samples with the test status
Requested:
1. Take the necessary steps to remove or disperse the
obstruction, or obtain a fresh quantity of sample.
2. Load the sample. The test is run automatically.
3. If the problem persists, contact Technical Support.
• For samples with the test status Cancelled:
1. Take the necessary steps to remove or disperse the
obstruction, or obtain a fresh quantity of sample.
2. Determine if the Automatic Rerun feature is enabled
(see Configuring the Automatic Rerun Feature in
Section 3.3) and take one of the follow actions.
- If the Automatic Rerun feature is not enabled,
proceed to step 3.
- If the Automatic Rerun feature is enabled, the test
has been reordered automatically. Proceed to step 4.
3. Request the test again.
4. Load the sample.
Note: Do not load the sample for automatic rerun
requests unless the test has been added to work
pending. If sufficient reserve volume is onboard the
instrument, and all other supplies are acceptable, the
test runs automatically.
5. If the problem persists, contact Technical Support.
IND • For sandwich assays, which use 1. For IND flagged results which meet the following
positive slope calibration curves, the criteria, dilute and rerun the sample:
result is at the low end of the • The result is on a competitive assay, and
concentration curve and cannot be
• Sample dilution is allowed for the assay, and
distinguished from a system failure
because the RLU reading is too low. • The sample RLU is low.
• For competitive assays, which use See the reagent instructions for use to determine the
negative slope calibration curves, the assay type and whether sample dilution is allowed.
result is at either: 2. Rule out a system problem by reviewing the Event Log
• The high end of the concentration for error events with a date and time shortly before this
curve and cannot be distinguished event. Troubleshoot accordingly.
from a system failure because the 3. If you have ruled out a system problem and the IND flag
RLU reading is too low, or event is the only issue, recalibrating the affected assay
• The low end of the concentration may resolve the problem.
curve and cannot be distinguished • Recalibrate the assay using a new reagent pack and a
from a system failure because the new set of calibrators.
RLU reading is too high. • Run QC.
• Repeat the test on the affected sample.
If IND flags persist, contact Technical Support.
NCR No calibration data existed for the reagent 1. Calibrate the assay (see Section 5.3).
lot when the patient or QC result was 2. Repeat the test.
processed.
QNS The sample volume is insufficient in the Check the test status in the Result column on the Test Result
sample container or in the RV during screen.
processing. • For patient and QC samples with the test status
If the flag is applied because of Requested:
insufficient volume in the RV, and 1. Identify event log errors. Contact Technical Support if
Automatic Rerun is enabled, the test will you need help troubleshooting event log errors (see
be reordered. Otherwise, additional tests Section 5.1 in the Reference Manual).
will not be scheduled for this sample.
2. Pipette sufficient sample volume into the sample
Tests already scheduled will be completed.
container (see Section A.2) and make sure the rack is
appropriate for the sample container (see Section A.1).
3. Load the sample. The test is run automatically.
4. If the problem persists, contact Technical Support.
• For samples with the test status Cancelled:
1. Identify event log errors. Contact Technical Support if
you need help troubleshooting event log errors (see
Section 5.1 in the Reference Manual).
2. Determine if the Automatic Rerun feature is enabled
(see Configuring the Automatic Rerun Feature in
Section 3.3) and take one of the follow actions.
- If the Automatic Rerun feature is not enabled,
proceed to step 3.
- If the Automatic Rerun feature is enabled, the test
has been reordered automatically. Take one of the
following actions.
• If sufficient reserve volume is not onboard the instrument,
the test is added to work pending. Proceed to step 4.
• If sufficient reserve is onboard the instrument, and all other
supplies are acceptable, the test runs automatically.
Proceed to step 6.
3. Request the test again.
4. Pipette sufficient sample volume into the sample
container (see Section A.2) and make sure the rack is
appropriate for the sample container (see Section A.1).
5. Load the sample.
Note: Do not load the sample for automatic rerun
requests unless the test has been added to work
pending. If sufficient reserve volume is onboard the
instrument, and all other supplies are acceptable, the
test runs automatically.
6. If the problem persists, contact Technical Support.
QSB A substrate dispense failure occurred 1. Prime the substrate (see Section 4.3 in the Reference
during processing. Manual) for four cycles.
2. Determine if the Automatic Rerun feature is enabled (see
Configuring the Automatic Rerun Feature in Section 3.3)
and take one of the follow actions.
- If the Automatic Rerun feature is not enabled,
proceed to step 3.
- If the Automatic Rerun feature is enabled, the test
has been reordered automatically. Proceed to step 4.
3. Request the test again.
4. Load the sample.
Note: Do not load the sample for automatic rerun
requests unless the test has been added to work pending.
If sufficient reserve volume is onboard the instrument,
and all other supplies are acceptable, the test runs
automatically.
5. If the problem persists, contact Technical Support.
QSD Insufficient sample volume or reagent 1. Review the Event Log. Contact Technical Support if you
volume was dispensed into an RV. need help troubleshooting event log errors (see
Section 5.1 in the Reference Manual).
2. Determine if the Automatic Rerun feature is enabled (see
Configuring the Automatic Rerun Feature in Section 3.3)
and take one of the follow actions.
- If the Automatic Rerun feature is not enabled,
proceed to step 3.
- If the Automatic Rerun feature is enabled, the test
has been reordered automatically. Proceed to step 4.
3. Request the test again.
4. Load the sample.
Note: Do not load the sample for automatic rerun
requests unless the test has been added to work pending.
If sufficient reserve volume is onboard the instrument,
and all other supplies are acceptable, the test runs
automatically.
5. If the problem persists, contact Technical Support.
QSS An insufficient volume of sample or 1. Review the Event Log. Contact Technical Support if you
reagent was withdrawn for the test. need help troubleshooting event log errors (see
Section 5.1 in the Reference Manual).
2. Determine if the Automatic Rerun feature is enabled (see
Configuring the Automatic Rerun Feature in Section 3.3)
and take one of the follow actions.
- If the Automatic Rerun feature is not enabled,
proceed to step 3.
- If the Automatic Rerun feature is enabled, the test
has been reordered automatically. Proceed to step 4.
3. Request the test again.
4. Load the sample.
Note: Do not load the sample for automatic rerun
requests unless the test has been added to work pending.
If sufficient reserve volume is onboard the instrument,
and all other supplies are acceptable, the test runs
automatically.
5. If the problem persists, contact Technical Support.
RLU The relative light units (RLUs) are outside 1. Review the Event Log. Contact Technical Support if you
the acceptable luminometer measuring need help troubleshooting event log errors (see
range. Section 5.1 in the Reference Manual).
2. Determine if the Automatic Rerun feature is enabled (see
Configuring the Automatic Rerun Feature in Section 3.3)
and take one of the follow actions.
- If the Automatic Rerun feature is not enabled,
proceed to step 3.
- If the Automatic Rerun feature is enabled, the test
has been reordered automatically. Proceed to step 4.
3. Request the test again.
4. Load the sample.
Note: Do not load the sample for automatic rerun
requests unless the test has been added to work pending.
If sufficient reserve volume is onboard the instrument,
and all other supplies are acceptable, the test runs
automatically.
5. If the problem persists, contact Technical Support.
SYS A device error occurred during processing. 1. Review the Event Log. Contact Technical Support if you
need help troubleshooting event log errors (see
Section 5.1 in the Reference Manual).
2. Determine if the Automatic Rerun feature is enabled (see
Configuring the Automatic Rerun Feature in Section 3.3)
and take one of the follow actions.
- If the Automatic Rerun feature is not enabled,
proceed to step 3.
- If the Automatic Rerun feature is enabled, the test
has been reordered automatically. Proceed to step 4.
3. Request the test again.
4. Load the sample.
Note: Do not load the sample for automatic rerun
requests unless the test has been added to work pending.
If sufficient reserve volume is onboard the instrument,
and all other supplies are acceptable, the test runs
automatically.
5. If the problem persists, contact Technical Support.
TRI The temperature of the analytical module 1. Check the wash carousel temperature on the
was outside the acceptable limits when the Maintenance Review screen. If the temperature is
test was being incubated. outside the acceptable limits, periodically monitor the
wash carousel temperature on this screen until it is
within the acceptable limits.
• If the instrument was restarted or instrument covers
were recently opened, you may need to wait up to 10
minutes for the temperature to normalize.
• If the system was powered down for an extended
period of time, you may need to wait up to one hour for
the temperature to normalize.
2. When the wash carousel temperature is within the
acceptable limits, determine if the Automatic Rerun
feature is enabled (see Configuring the Automatic Rerun
Feature in Section 3.3) and take one of the follow
actions.
• If the Automatic Rerun feature is not enabled, proceed
to step 3.
• If the Automatic Rerun feature is enabled, the test has
been reordered automatically. Proceed to step 4.
3. Request the test again.
4. Load the sample.
Note: Do not load the sample for automatic rerun
requests unless the test has been added to work pending.
If sufficient reserve volume is onboard the instrument,
and all other supplies are acceptable, the test runs
automatically.
5. If the temperature does not normalize, or if the problem
persists, contact Technical Support.
TRS The temperature of the substrate was 1. Check the substrate temperature on the Maintenance
outside the acceptable limits when the Review screen. If the temperature is outside the
substrate was dispensed. acceptable limits, periodically monitor the substrate
temperature on this screen until it is within the
acceptable limits.
• If the instrument was restarted or instrument covers
were recently opened, you may need to wait up to 10
minutes for the temperature to normalize.
• If the system was powered down for an extended
period of time, you may need to wait up to one hour for
the temperature to normalize.
2. When the substrate temperature is within the acceptable
limits, determine if the Automatic Rerun feature is
enabled (see Configuring the Automatic Rerun Feature
in Section 3.3) and take one of the follow actions.
• If the Automatic Rerun feature is not enabled, proceed
to step 3.
• If the Automatic Rerun feature is enabled, the test has
been reordered automatically. Proceed to step 4.
3. Request the test again.
4. Load the sample.
Note: Do not load the sample for automatic rerun
requests unless the test has been added to work pending.
If sufficient reserve volume is onboard the instrument,
and all other supplies are acceptable, the test runs
automatically.
5. If the temperature does not normalize, or if the problem
persists, contact Technical Support.
TRW The temperature of the analytical module 1. Follow the corrective actions provided for the TRI flag.
was outside the acceptable limits when the
reaction vessel was in the wash carousel.
CEX The calibration curve or cut-off value is 1. Recalibrate the assay (see Section 5.3).
expired. 2. Repeat the test.
CLX The calibrator lot is expired. • For patient and QC tests: Review the Calibration
Note: For patient and QC tests, the Data screen to determine whether replicates of the
calibrator lot may have expired after a active calibration are associated with the CLX flag
successful calibration. (see Section 5.4). Then take one of the following
actions:
• If the calibration is not associated with the flag,
the patient or QC test result is a valid result. No
corrective action is necessary.
• If the calibration is associated with the flag, and
you did not intend to run the QC or patient test
using this calibration, calibrate the assay again
with a calibrator lot that has not expired. Request
the QC or patient test again.
• If the calibration is associated with the flag, and
you intended to run the QC or patient test using
this calibration, no corrective action is
necessary.
• For calibrations: If you did not intend to calibrate
with an expired lot, run another calibration with a
calibrator lot that has not expired. Otherwise, no
corrective action is necessary.
CRH The patient test result is above the upper This is a valid test result. No corrective action is
limit of the critical range. necessary.
CRL The patient test result is below the lower This is a valid test result. No corrective action is
limit of the critical range. necessary.
DEX The open pack stability time has expired 1. Make sure sufficient unexpired diluent is available
for the diluent pack. The system measures (see Section 2.7).
open pack stability separately by well. 2. Repeat the test.
Note: For an LIS attached to a DxI
instrument, the PEX flag is sent to the LIS
in place of the DEX flag.
EXS The substrate is expired. 1. Review the expiration date on the Bulk Supplies
screen (see Section 2.1).
2. If necessary, change the substrate bottle (see
Section 2.2).
3. Repeat the test.
GRY For qualitative assays, the patient or QC This is a valid test result. No corrective action is
result is within the specified gray zone., necessary.
also known as the equivocal zone.
LEX The reagent or diluent pack lot is expired. 1. Unload all of the packs from the expired reagent
lot and load a pack from a new lot (see
Section 2.7).
2. Repeat the test.
LOW The patient or QC result is lower than the No corrective action is necessary.
minimum reportable result value defined
in the APF.
LRH The patient result is above the upper limit 1. Review the result.
of the LIS range. 2. Take one of the following actions:
Notes: • Send the result to the LIS manually (see
• If the Auto-Send to LIS option is set to Section 4.4).
Verify, the system does not • Delete the result (see Section 4.4) and repeat the
automatically send results with this flag test.
to the LIS.
• This flag is for quantitative assays,
semi-quantitative assays, and derived
results only.
LRL The patient result is below the lower limit 1. Review the result.
of the LIS range. 2. Take one of the following actions:
Notes: • Send the result to the LIS manually (see
• If the Auto-Send to LIS option is set to Section 4.4).
Verify, the system does not • Delete the result (see Section 4.4) and repeat the
automatically send results with this flag test.
to the LIS.
• This flag is for quantitative assays,
semi-quantitative assays, and derived
results only.
ORH The patient result is above the upper limit No corrective action is necessary.
of the reference range.
ORL The patient result is below the lower limit No corrective action is necessary.
of the reference range.
OVR The calculated concentration is above the 1. Review the Event Log for error events with a
highest or most concentrated calibrator. similar date and time to this event (see Section 5.2
This flag is only used for quantitative and in the Reference Manual).
semi-quantitative assays. • If events occurred, troubleshoot (see Section 5.3
in the Reference Manual).
2. Take one of the following actions:
• If events occurred, and you performed
troubleshooting procedures, run controls and
then repeat the test.
- If the controls are in range, and the test result
is reported as greater than the value of the
highest calibrator (>X), you may be able to
dilute the sample. To identify whether the
assay allows dilutions, see the reagent
instructions for use.
- If the controls are not in range, follow the QC
troubleshooting instructions (see Section 6.7).
• If no events occurred, take one of the following
actions:
- If the test result is reported as >X, you may be
able to dilute the sample. To identify whether
the assay allows dilutions, see the reagent
instructions for use. If dilutions are allowed,
dilute and repeat the test.
- If dilutions are not allowed, no further action
is necessary.
3. If you have questions about the result, or if the
problem persists, contact Technical Support.
PEX The open pack stability time has expired 1. Unload the expired reagent pack and load a new
for the reagent pack. The system starts one (see Section 2.7).
measuring open pack stability when it first • If the lot number of the new reagent pack is
punctures the pack. different than the expired pack, recalibrate the
Note: For an LIS attached to a DxI assay.
instrument, the PEX flag is sent to the LIS 2. Repeat the test.
in place of the DEX flag.
QCF The quality control result violates one or 1. Display the QC Chart and Data screen to review
more QC rules. which criteria is not met.
2. Follow the QC troubleshooting instructions (see
Section 6.7).
QEX The quality control lot is expired. 1. Add a new, unexpired quality control lot (see
Section 6.2). Make sure that expected means and
SDs are established according to laboratory
guidelines.
2. Repeat the test.
RFX The patient sample result is from a reflex No corrective action is necessary for this flag alone.
test.
TRA The temperature of the sample wheel was 1. Check the sample wheel temperature on the
outside the acceptable limits when the Maintenance Review screen (see Section 7.1). If
sample aliquot was in the sample wheel. the temperature is outside the acceptable limits,
periodically monitor the sample wheel temperature
on this screen until it is within the acceptable
limits.
• If the system was restarted or instrument covers
were opened recently, you may need to wait up
to 10 minutes for the temperature to normalize.
• If the system was powered down for an extended
period of time, you may need to wait up to one
hour for the temperature to normalize.
2. When the sample wheel temperature is within the
acceptable limits, repeat the test.
3. If the temperature does not normalize, or if the
problem persists, contact Technical Support.
TRR The temperature of the reagent storage 1. Check the reagent storage chamber temperature on
chamber was outside the acceptable limits the Maintenance Review screen (see Section 7.1).
when reagents were pipetted. If the temperature is outside the acceptable limits,
periodically monitor the reagent storage chamber
temperature on this screen until it is within the
acceptable limits.
• If the system was restarted or the instrument
storage chamber cover was recently opened, you
may need to wait up to 10 minutes for the
temperature to normalize.
• If the system was powered down for an extended
period of time, you may need to wait up to one
hour for the temperature to normalize.
2. When the reagent storage chamber temperature is
within the acceptable limits, repeat the test.
3. If the temperature does not normalize, or if the
problem persists, contact Technical Support.
NOTE
If the test status is Requested, and the sample container is bar coded, the test has
already been added to the Work Pending list and the system will prevent you from
requesting it again. Load the bar coded sample again and the test will run
automatically.
To rerun a quality control test, see the Rerunning a QC Test procedure in Section 6.7:
QC Troubleshooting Overview.
System Mode: Any Mode
1. From the Test Results screen, locate and select the result you want to rerun.
For more information on locating results, see the Finding a Test Result procedure or
the Applying a Test Result Filter procedure in Section 4.2: Test Result Filters.
NOTE
Only select a single result. If you select a group of results and then select
Rerun Test F3, only the result indicated with the Arrow button is rerun.
To store results offline before they are deleted from the system, copy test results to a
disk. After copying results to a disk or flash drive, transfer them to a spreadsheet
program for viewing.
If you have an LIS, you can configure the system to automatically send test results
with their associated flags to the LIS when sample processing is complete. For
information on configuring the LIS, see Chapter 2: System Administration in the
Reference Manual. However, occasionally you must manually send results to the LIS.
For more information, see the Sending Test Results to the LIS procedure.
All results remain on the system until they are deleted. Automatic deletion is the
preferred method for deleting results. For information on configuring automatic
deletion, see the Setting Up the Auto-Delete Feature procedure in Section 2.6: Data
Management of the Reference Manual.
You can also manually delete results from the Test Results screen, but expect
performance delays. For more information, see the Deleting a Test Result procedure.
Copying Test Use this procedure to copy test results to a USB flash drive or a 3.5-inch high density
Results to a disk.
Disk or Flash The UniCel DxI system writes test result data into a comma-separated field format
Drive that can be imported into a PC-based spreadsheet program. The following data is
presented in the order indicated:
NOTE
If a plus sign (+) is displayed in the Entries field of the Test Results screen,
refresh the screen to be sure you are searching the most current data. For more
information, see the Refreshing the Test Results Screen procedure in Section 4.1:
Test Results Overview.
(Optional) To locate a specific test result to copy, select Find F2. For more
information, see the Finding a Test Result procedure in Section 4.3: Reviewing
Test Results.
(Optional) Select a single test result or a group of results. For more information,
see the Selecting Test Results procedure in Section 4.1: Test Results Overview.
NOTE
If you select a group of results and then select Copy to Disk F6, you must select
the Current Selection option in the window to copy all selected results.
Otherwise, only the result with the Arrow button is copied.
NOTE
For UniCel DxI instruments set up to perform system backups using a USB flash
drive, the backup drive might already be in one of the USB ports on the back of
the PC. Use a different flash drive to copy data. Do not disturb the flash drive
used for backups when inserting and removing the flash drive for copying data.
• To copy data to a 3.5-inch disk, insert a disk into the 3.5-inch disk drive of
the PC.
3. Select Copy to Disk F6. A default file name and extension are provided in the
File Name field of the Copy to Disk window.
2107B.bmp
File Name (Field) Displays the default file name and extension of the test
results to be copied. You can delete the name and enter a new
name.
Drive (List) Select the drive to which data should be copied.
Current (Option) Select to copy one selected test result to the storage
Result media.
Current (Option) Select to copy one or more selected test results to the
Selection storage media.
All (filtered (Option) Select to send all the test results in the current filter to
results) the storage media.
Format Disk (Box) Select to format a disk. Formatting a disk prepares it for
reading and writing data. This option is not available for flash
drives.
4. Confirm that the correct drive is selected from the Drive list, or select a different
drive.
NOTE
Unless directed by Technical Support, do not copy data to the C: Console PC
drive or any drive labeled BACKUPDRIVE.
(Optional) To change the default file name, enter a name and file extension for
the data file in the File Name field of the Copy to Disk window.
• The system supplies a default file name based on the date and time and a .csv
(comma delimited) extension.
• If the file name already exists on the disk or flash drive, the system displays a
warning message. To overwrite the existing file, select Yes F1. To rename
the new file, select Cancel F8.
5. Select a Copy what option.
• For the Current Result and the Current Selection options, one or more
results must be selected on the Test Results screen.
(Optional) If the 3.5-inch disk is not formatted, select the Format disk box.
6. Select OK F1.
A progress window is displayed.
• To stop the copy process at any time, select Cancel F8.
7. If the flash drive or disk becomes full before the copy process is complete, the
system displays a status message.
• For a USB flash drive, clear sufficient space on the drive, or copy the files to
a different flash drive.
• For a 3.5-inch disk:
- Remove the disk and label it with the data file names.
- Insert another disk, and select OK F1.
The system automatically names the file on the new disk the same as the file on
the first disk, and adds a disk number to the end.
Example
If two disks are required to copy a test result file, and you name disk
one mydata.csv, the system names disk two mydata(2).csv.
8. When the copy process is complete, the progress window closes. Remove the
flash drive or disk from the PC and label it.
NOTE
If you will be using one flash drive or disk to copy multiple files, you can leave it
in place for the next copy operation. Be sure to keep track of the file names so you
can correctly label the flash drive or disk when you remove it.
Test Results
Are
results to be No Do you want a No
copied specific result?
displayed?
Yes Yes
Apply filter
(Optional) Enter one of the four options
and sort orders
Select one or more results
Find Next F1 OK F1
Copy to Disk F6
Copy to Disk
• Confirm or change
the file name
• Select a drive from
the Drive list
• Select a Copy What option
(Optional)
For 3.5-inch disks
select Format disk
OK F1
2038C.wmf
Figure 4-14 Copying Test Results to a Flash Drive or Disk Flowchart
Sending Test Use this procedure to manually send test results to the LIS if one or more of the
Results to the following circumstances applies:
LIS • The Auto Send to LIS option is set to Off or to Send Only Auto-Verified
Results. For more information, see Section 2.5: LIS Setup in the Reference
Manual.
• The LIS fails to receive the test results that were sent.
• The test results were on hold while the LIS was offline, and now the LIS is
back online.
(Optional) To locate a specific test result to send, select Find F2. For more
information, see the Finding a Test Result procedure in Section 4.3: Reviewing
Test Results.
2. (Optional) Select a single test result or a group of results. For more information,
see the Selecting Test Results procedure in Section 4.1: Test Results Overview.
NOTES
• You do not need to select results to send all of the results or all of the failed
results in the applied filter.
• If you select a group of results and then select Send to LIS F5, you must
select the Current Selection option in the window to send all selected
results. Otherwise, only the result with the Arrow button is sent.
2108A.bmp
Current (Option) Select to send one selected test result to the LIS.
Result
Current (Option) Select to send one or more selected test results to the
Selection LIS.
Resend LIS (Option) Select to resend all of the test results with an LIS
Fail status of Fail in the current filter to the LIS.
All (filtered (Option) Select to send all the test results in the current filter to
results) the LIS.
4. Select an option from the Send what list in the Send to LIS window.
• For the Current Result and the Current Selection options, one or more
results must be selected.
• For the Resend LIS Fail or the All (filtered results) options, the system
sends the results from the current filter.
5. Select OK F1.
NOTE
If tests continue to be added or removed from the filter, you may need to refresh
the screen before results can be sent. For more information, see the Refreshing the
Test Results Screen procedure in Section 4.1: Test Results Overview. Then repeat
this procedure.
6. Monitor the status of the test results sent to the LIS by viewing the LIS column of
the Test Results screen.
• If the LIS successfully receives results, the LIS status changes to Accepted.
• If the LIS does not successfully receive the results, the LIS status changes to
Failed. The system records the event in the Event Log and the Event Log
button turns red. Contact Technical Support for assistance.
Test Results
Are results
to be sent
No Do you want a No
specific result?
displayed?
Yes Yes
Find F2 Filter F1
(Optional)
Select one or more results
Find Filter
OK F1
Monitor transmission
2070B.wmf
Figure 4-16 Sending Test Results to the LIS Flowchart
Deleting a Use this procedure to delete patient and QC test results from the database.
Test Result
NOTES
• You need the system password to use this feature. If you do not know the
password, contact your lab supervisor. For more information about
passwords, see the Changing the System Password procedure in Section 2.2:
System Setup of the Reference Manual.
• Deleting test results removes the data permanently from the database.
• You cannot delete calibrator or maintenance results with this procedure.
• User interface performance is diminished while tests are being deleted using
this procedure. To delete test results automatically, use the Auto-Delete
feature. For more information, see the Setting Up the Auto-Delete Feature
procedure in Section 2.6: Data Management of the Reference Manual.
(Optional) To locate a specific test result to delete, select Find F2. For more
information, see the Finding a Test Result procedure in Section 4.3: Reviewing
Test Results.
2. (Optional) Select one or more results. For more information, see the Selecting
Test Results procedure in Section 4.1: Test Results Overview.
NOTE
You do not need to select results to delete all filtered results.
4. In the Password field of the password window, enter the system password. Then
select OK F1.
5. Select an option from the Delete what list of the Delete Data window.
• To delete a single selected test result, select the Current Result option.
• To delete a group of selected test results, select the Selected Results option.
• To delete all filtered test results, select the All (filtered results) option.
6. Select OK F1.
The system displays a warning message.
Test Results
Are results
to be deleted
No Do you want a No
specific result?
displayed?
Yes Yes
Find F2 Filter F1
(Optional)
Select one or more results
Find Filter
FInd Next F1 OK F1
Password
OK F1
Delete Data
OK F1
2069B.wmf
Figure 4-17 Deleting a Test Result Flowchart
Result information is available on four different test result reports. All of the reports
can be printed manually. Some of the reports can be set to print automatically as well.
You can customize the test result information you obtain on a report in a number of
ways. You can configure the Test Results screen to display the type and order of result
information you want on the report. You can also filter and sort test results by various
criteria to obtain the desired report. Finally, you can select which results to print on
the report.
Examples
• To print a report of all TSH results, sort the results by test name, select the
desired TSH results, and print the Selected Test Results Report.
• To print a report of all tests completed in the past 24 hours, apply the All
Samples Loaded Between... filter to the test results, select the Past 24
hours option for the date and time, and print the Selected Test Results
Report.
Test Result Four test result reports are available on the UniCel DxI system.
Report
Descriptions NOTE
If present, dilution factors displayed on the Sample Report, Patient Report, and
Continuous Sample Report are manual dilution factors typed during a manual test
request. Onboard dilution and Special Dilution information is not included on test
result reports. For more information, see Onboard Dilutions in Section 1.3:
Sample Processing Theory of the Reference Manual.
Sample Report
The Sample Report (see Figure 4-18) contains all test result information associated
with a sample ID.
If two or more tests associated with the same sample ID are selected on the Test
Results screen, and you manually print the Sample Report, the report prints only once.
For more information, see the Printing Test Result Reports procedure.
If set to print, STAT sample results automatically print in this format. For more
information, see the Setting Up Reports procedure in Section 2.2: System Setup of the
Reference Manual.
Patient Report
The Patient Report (see Figure 4-19) contains all test result information associated
with a patient ID.
If two or more tests associated with the same patient ID are selected on the Test
Results screen and you manually print the Patient Report, the report prints only once.
For more information, see the Printing Test Result Reports procedure.
This report format mirrors the configuration of the Test Results screen. To print this
report, see the Printing Test Result Reports procedure.
If set to print, the report prints automatically when there are enough completed test
results to fill a page, when all processing is complete, or when all tests for a STAT
sample tube are complete.
NOTE
Requested test results that cause reflex tests are printed on two Continuous
Sample Reports. The first report prints only the requested test results. When the
reflex tests are complete, the report prints a second time with the reflex results
and the requested results.
Alternatively, you can manually print the report as samples are completed. For more
information, see the Printing the Continuous Sample Report procedure.
Calibration and maintenance results are not included on this report. For more
information, see the Printing a Calibration Data Report in Section 5.4: Reviewing
Calibration Data or see the Displaying and Printing System Check Result Data
procedure in Section 4.4: System Check Routines in the Reference Manual.
Laboratories, Inc.
Laboratory A
123 Lake Street
Sample Report Townsville, ST 33333
UniCel DxI 800, Access Immunoassay System 101-555-2323 ext. 109
S/N 610240 B. Beaulieu
**STAT**
Instrument: 501240
Test Name Result Ref. Range Flags Dilution Completed
CK-MB >293 ng/mL 0.6-6.3 OVR 1 10-10-02 07:30 PM
MYO No Value 17.4-105.7 QNS 1 10-10-02 07:27 PM
MYO(2) 39.6 ng/mL 17.4-105.7 1 10-10-02 07:16 PM
Instrument: 501240
Test Name Result Ref. Range Flags Dilution Completed
FreePSA% 16.6
freePSA 1.58 ng/mL 1 08-08-02 02:45 PM
PSA-Hyb 9.53 ng/mL 0.000-4.000 1 08-08-02
02:15 PM
Comments: Cloudy sample.
2040B.bmp
Figure 4-18 Sample Report Example
Laboratories, Inc.
Laboratory A
123 Lake Street
Patient Report Townsville, ST 33333
UniCel DxI 800, Access Immunoassay System 101-555-2323 ext. 109
S/N 610240 R. Goldman
**STAT**
Sample ID: 23345128Dil Sample Type: Plasma Dilution: 10
Test Name Result Ref. Range Flags Completed
CK-MB 939.0 ng/mL 0.5-5.0 ORH 10-10-02 08:18 PM
Comments: Sample slightly cloudy.
**STAT**
Sample ID: 23345128 Sample Type: Plasma
Test Name Result Ref. Range Flags Completed
CK-MB >293 ng/mL 0.5-5.0 OVR 10-10-02 07:30 PM
MYO No Value 14.3-65.8QNS 10-10-02 07:27 PM
MYO(2) 125.8 ng/mL 14.3-65.8ORH 1 0-10-02 07:27 PM
2039B.bmp
Figure 4-19 Patient Report Example
Laboratories, Inc.
Laboratory A
123 Lake Street
Selected Test Results ReportUniCel DxI 800, Townsville, ST 33333
Access Immunoassay System 101-555-2323 ext. 109
S/N 610240 V. Flores
Sample ID Rack Test Name Result Flags LIS Pipettor Comp. Time
13501350 48 Ferritin 404.1 ng/mL Accepted 1 10/10/02 06:40
13501350 48 FOL2 4.34 ng/mL Accepted 3 10/10/02 06:32
13501350 48 VitB12 354 pg/mL Accepted 4 10/10/02 06:38
19953581 504 freePSA Supply Wait 1
19953581 504 PSA-Hyb Supply Wait 3
23345128 106 CK-MB >293 ng/mL OVR Accepted 4 10/10/02 19:30
23345128 106 MYO 39.6 ng/mL Accepted 2 10/10/02 19:27
23345128 106 MYO(2) No Value QNS Accepted 1 10/10/02 19:27
25688888 48 VitB12 103 pg/mL Failed 2 10/10/02 06:41
39422493 505 MYO 0.00 ng/mL Failed 2 10/10/02 14:03
78910116 502 FOL2 Cancelled SYS 3 10/10/02 10:46
2041B.bmp
Figure 4-20 Selected Test Results Report Example
4-56
Laboratory A
123 Lake Street
Continuous Sample Report Townsville, ST 33333
UniCel DxI 800, Access Immunoassay System 101-555-2323 ext. 109
S/N 601240 Jane Smith
___________________________________________________________________________________________________________________________________
Sample ID Type
4.5: Test Result Reports
Rack/Pos Patient/Lot ID Dilution Test Result Ref. Range Flags RLUs Completion
________________________________________________________________________________________________________________________________
122/1 23345128 Plasma
**STAT** 555-56-9898 1 CK-MB >293 ng/mL 0.000 - 4.000 OVR 18445200 10-10-02 06:01 PM
MYO 39.6 ng/mL 14.3-65.8 337183
10-10-02 05:58 PM
MYO(2) No Value 14.3-65.8 QNS No Value
10-10-02 05:48 PM
Dig No Value NCR 853012
10-10-02 06:07 PM
C73199-AA
2042B.bmp
UniCel DxI Operator's Guide 4.5: Test Result Reports
(Optional) To locate a specific test result to print, select Find F2. For more
information, see the Finding a Test Result procedure in Section 4.3: Reviewing
Test Results.
(Optional) Select one or more results. For more information, see the Selecting
Test Results procedure in Section 4.1: Test Results Overview.
0623B.bmp
3. Select the type of report you want to print from the report list. For more
information, see the Test Result Report Descriptions topic.
5. Select OK F1.
The system sends the report to the printer. A progress window is displayed. Select
Cancel F8 to stop the print job at any time.
Test Results
Are results
to be printed
No Do you want a No
specific result?
displayed?
Yes Yes
FInd Next F1 OK F1
Print
OK F1
2071B.wmf
Figure 4-23 Printing Test Result Reports Flowchart
Printing the Use this procedure to manually print the Continuous Sample Report. For a description
Continuous of the Continuous Sample report, see the Test Result Report Descriptions topic.
Sample Most operators will wait for the Continuous Sample Report to print automatically. To
Report set up the report to print automatically, see the Setting Up Reports procedure in
Section 2.2: System Setup of the Reference Manual. However, use this procedure if
you need to print the report immediately for completed samples not yet printed.
NOTES
• Incomplete tests do not print on the Continuous Sample Report.
• Results print on the Continuous Sample Report only once. If a result was
printed manually or automatically on the Continuous Sample Report and you
want to reprint the result, you must choose one of the other three report
formats.
5 Assay Calibration
5.1 Assay Calibration Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Assay calibrations establish values that the system uses to calculate test results for
patient and quality control (QC) samples. Calibrations are generated from tests run on
samples with stated concentrations of analyte. The samples used to calibrate the assay
are known as calibrators.
NOTE
Whether you use calibrators or controls to calibrate an assay, the UniCel DxI
system and this documentation use the term calibrator to refer to these
components.
Calibrations either pass or fail according to the acceptance criteria defined in the assay
protocol file (APF). When a calibration passes, it becomes the active calibration for
processing patient and QC test requests. An active calibration remains active, even if
it is expired, until you run another calibration that passes, you switch to a previous
active calibration, or you delete the calibration. If there is no active calibration for an
assay, RLUs are reported, but no patient or QC test results are obtained.
Calibrations expire according to parameters defined for the assays in the APF. If a
calibration does not exist or is expired and you request a test for the assay, the system
prompts you with the following indicators to calibrate the assay before running the
test:
• The Supplies Required button turns yellow
• The Amber status indicator light blinks
• The Console panel icon is lit on the system status panel (UniCel DxI 800
instrument only)
You can process tests with expired calibrations by overriding the condition, however
the results will be flagged. For information about overriding, see Section 2.1: Supply
and Supply Conditions Overview.
When a calibrator lot expires, it remains available for calibration until you delete the
lot from the system. You can request a calibration using an expired calibrator lot. You
can also request QC and patient tests using an active calibration curve based on a
calibrator lot that was or has since expired. When an expired lot is used, results for
calibration, QC, and patient tests report with the non-fatal CLX flag. For more
information on result flags, see Section 4.3: Reviewing Test Results.
For failed calibrations, the failure code is displayed on the Calibration Data screen
and a caution event is reported on the Event Log. For help on how to interpret or
troubleshoot failed calibrations, see Section 5.5: Troubleshooting Failed Calibrations.
For information about assay calibration theory, see Section 1.4: Assay Calibration
Theory in the Reference Manual.
Set up calibrators
Run a calibration
• Prepare samples
• Request the calibration
• Load racks
Yes
2127C.wmf
Figure 5-1 Assay Calibration Overview Flowchart
2128B.bmp
Review Data (Button) Select to review the Calibration Data screen for a selected Assay button.
F2 The data for the active calibration is displayed. If no calibration has passed the
acceptance criteria, the system displays the last failed calibration. For more
information, see Section 5.4: Reviewing Calibration Data.
Calibrator Setup (Button) Select to display the Calibrator Setup screen. From this screen, you can
F5 set up, edit, or delete calibrators. For more information, see Section 5.2: Setting Up
Calibrators.
Delete (Button) Select to delete a selected calibration and its Assay button. All of the
F6 calibration data for the selected test and reagent lot is removed from the system.
For more information, see the Deleting Calibrations procedure.
Print (Button) Select to print the Calibration Report, which lists assay calibrations
F7 according to the applied filter parameters. For more information, see the Printing
the Calibration Report procedure.
Applying a Use this procedure to apply a different filter or sort order to the Assay buttons on the
Calibration Calibration screen.
Filter System Mode: Any Mode
1. From the Calibration screen, select Filter F1.
0850C.bmp
Filter (Options) Select either the filter that displays all assay
calibrations or the filter that displays only those assay
calibrations with reagent on board the instrument.
3. Select the sort order for the filtered calibrations in the Ascending box.
Calibrations are sorted by test name, then by reagent lot.
• To sort in ascending order, select the box.
• To sort in descending order, clear the box.
4. Select OK F1 to apply the filter and display the calibrations that meet the filter
parameters on the Calibration screen.
Deleting Use this procedure to delete the assay calibration data for a specific reagent lot. Delete
Calibrations calibrations that are no longer needed, such as expired calibrations for reagents that
will not be calibrated again. To save calibration data before you delete it, see Copying
Test Results to a Disk or Flash Drive in Section 4.4: Managing Test Results.
NOTE
If you have no further need for the calibrator lot or for the assay, delete the
calibrator lot instead of the calibration. For more information, see the Deleting a
Calibrator Lot procedure in Section 5.2: Setting Up Calibrators.
Printing the The Calibration Report is a list of assay calibrations, sorted first by test name and then
Calibration by reagent lot. Use this procedure to print the report.
Report System Mode: Any Mode
1. On the Calibration screen, confirm that the calibrations you intend to print are
included in the current filter. If needed, apply a different filter. For more
information, see the Applying a Calibration Filter procedure.
Laboratories, Inc.
Laboratory A
123 Lake Street
Calibration Report Townsville, ST 33333
UniCel DxI 800, Access Immunoassay System 101-555-2323 ext. 109
S/N 610240 Jane Smith
Filter: All Calibrations - Ascending Sorted by Test Name and Reagent Lot
Page 1 of 1
Technologist__________________________________________ Printed 03/03/10 12:43 PM
2131C.bmp
Figure 5-4 Calibration Report Example
Before you can request a calibration, the calibrator must be set up on the system. A
calibrator is identified on the system by the test name and the lot number.
NOTE
Whether you use calibrators or controls to calibrate an assay, the UniCel DxI
system and this documentation use the term calibrator to refer to these
components.
Most calibrators are multi-level. They contain more than one vial, and each vial
contains a different concentration or level of analyte. When you add a multi-level
calibrator to the system, all calibrator levels are added as one calibrator lot.
To correct calibrator information that is in error, you can edit some calibrators on the
system. If you cannot edit a calibrator that you need to change, delete the calibrator
and add it again.
When a calibrator lot is expired, your supply has run out, or you have no further need
for the assay on the system, delete the calibrator. All data associated with the
calibrator lot is removed from the system.
2134B.bmp
Calibrator List (List) Displays information about all of the calibrator lots set up on the system. For
each calibrator, the following information is displayed:
• Test - Displays the assay test name.
• Calibrator Lot - Displays the calibrator lot number.
• Expiration - Displays the expiration date for the calibrator lot.
Select a calibrator to edit or delete the calibrator lot.
Add Calibrator (Button) Select to add a calibrator lot to the system.
F1
Edit Calibrator (Button) For quantitative and semi-quantitative assays only. Select to edit the
F2 calibrator level information for one or more calibrator levels for a selected
calibrator lot. For more information about assay types, see Section 1.4: Assay
Calibration Theory in the Reference Manual.
Delete Calibrator (Button) Select to delete a selected calibrator lot and to remove all data associated
F6 with it from the system.
NOTE
You need the system password to use this feature.
Adding a Use this procedure to add a calibrator lot to the system for an individual test. If you
Calibrator Lot will be running a secondary test with this calibrator, add the calibrator lot again with
the second calibration card.
NOTE
This procedure refers to quantitative, semi-quantitative, and qualitative assays.
For more information about these types of assays, see Section 1.4: Assay
Calibration Theory in the Reference Manual.
2. From the Calibrator Setup screen, select Add Calibrator F1 to display the Add
Calibrator window.
2162A.bmp
Bar Code (Field) Enter the bar code information from the calibrator card using the PC bar
code reader. Always scan the bar code at the top of the card first. For some
calibrators, all of the calibrator information is automatically added to the window.
For other calibrators, you must scan each bar code on the card.
NOTES
• If you cannot scan the bar code, type the characters printed underneath the
bar code and press [Enter].
• This is the only field in which you can enter information in this window
(and the only field that can be edited in the Edit Calibrator window).
Calibrator Name (Field) Displays the name of the calibrator, based on the bar code entry.
Lot Number (Field) Displays the calibrator lot number, based on the bar code entry.
Expiration Date (Field) Displays the calibrator expiration date, based on the bar code entry.
Calibrator Level (List) Displays the following information for each calibrator level after bar code
List entry:
• Test Name - Displays the assay test name.
• Level - Displays the calibrator level.
• Stated Concentration - For quantitative and semi-quantitative assays only,
displays the concentration of analyte material in the calibrator level, as stated on
the calibrator card.
• Units - For quantitative and semi-quantitative assays only, displays the units of
measure for the calibrator level.
For more information about quantitative and semi-quantitative assays, see
Section 1.4: Assay Calibration Theory in the Reference Manual.
3. Use the PC bar code reader to scan the bar code at the top of the calibrator card.
The scanned information displays in the Bar Code field. For more information,
see the Scanning the Calibrator Card procedure.
For some multi-level calibrators (usually qualitative assays), the system automatically
adds all of the calibrator levels with their stated analyte concentrations and units to the
calibrator level list.
NOTE
If you cannot scan the bar code at any step of this procedure, you can type the
characters printed underneath the bar code and press [Enter].
4. If the calibrator is multi-level and the information for each level is not displayed
in the calibrator level list (usually quantitative assays), scan each bar code on the
card.
Calibration
Calibrator Setup F5
Calibrator Setup
Add Calibrator F1
Add Calibrator
When typing bar
code information, Scan the calibrator
press [Enter] after bar code information
each bar code
Is the
calibrator No
information
correct?
Yes
OK F1
2135B.wmf
Figure 5-7 Adding a Calibrator Lot Flowchart
Scanning the All Access Immunoassay System calibrator supplies include calibrator cards with the
Calibrator bar coded information required for adding or editing calibrator setup information. Use
Card this procedure to scan a calibrator card with the PC bar code reader.
2. Remove the PC bar code reader from the holder attached to the PC
(UniCel DxI 800 instrument: see Figure 1-1; UniCel DxI 600 instrument: see
Figure 1-2).
3. Set the calibrator card on a flat surface, such as the area to the right of the system
status panel.
4. Hold the PC bar code reader approximately 1-8 inches (2.5-20.3 cm) above the
card and aim the reader straight down at the card.
5. Starting with the bar code at top of the card, squeeze and hold the trigger.
6. Center the red light (indicating the scanning area) within the lines of the bar code
and release the trigger. When a bar code is read successfully, the bar code reader
beeps and the scanned information is displayed in the Bar Code field.
7. Confirm that the information in the Add or Edit Calibrator window is correct.
• If the information is not correct, and the calibrator card is not damaged in any
way, scan the bar code again.
• If the calibrator card is damaged or the bar code is not readable, type the
characters printed underneath the bar code and press [Enter].
Editing a Use this procedure to correct a problem with the calibrator level information for a
Calibrator quantitative or semi-quantitative assay.
Level
NOTES
• To edit the calibrator information for a qualitative assay, or to change any
information that cannot be edited, delete the calibrator and then add the
calibrator again.
• For more information about quantitative, semi-quantitative, or qualitative
assays, see Section 1.4: Assay Calibration Theory in the Reference Manual.
3. Select Edit a Calibrator F2. The selected calibrator is displayed in the Edit
Calibrator window.
NOTE
The Edit Calibrator window is identical to the Add Calibrator window.
4. Use the PC bar code reader to scan the bar code into the Bar Code field from the
calibrator card. For more information, see the Scanning the Calibrator Card
procedure.
NOTE
If you cannot scan the bar code at any step of this procedure, type the characters
printed underneath the bar code and press [Enter].
5. To correct more than one level of a multi-level calibrator, scan the bar code for
each level individually.
Deleting a Use this procedure to delete a calibrator lot from the system. All data associated with
Calibrator Lot the calibrator lot, including calibrations, is removed from the system.
NOTES
• You need the system password to use this feature. If you do not know the
password, contact your lab supervisor. For more information about
passwords, see Section 2.2: System Setup in the Reference Manual.
• You cannot delete a calibrator lot that is included in a calibration request.
2. From the Calibrator Setup screen, select the calibrator to be deleted from the
system.
5. Confirm your selection, and then select Yes F1 to delete the calibrator.
Each assay reagent lot must be calibrated on the instrument before it can be used to
analyze samples. The first calibration for an assay adds an Assay button to the
Calibration screen. Separate Assay buttons are added for each reagent lot. For more
information about Assay buttons and the Calibration screen, see Section 5.1: Assay
Calibration Overview.
Before calibrating an assay, be sure that the following conditions are met:
• The test is enabled on the system. For more information, see Section 3.3: Test
Setup in the Reference Manual.
• An unexpired calibrator is set up for the assay. For more information, see
Section 5.2: Setting Up Calibrators.
• Sufficient unexpired reagent is loaded on the instrument. For more
information, see Section 2.7: Reagent Supplies.
Some assays require temperature restrictions. See Appendix A of the UniCel DxI
Instructions for Use manual for information on these restrictions.
All calibrator samples in a calibration request are run using the same reagent pipettor.
The system selects that pipettor from the pipettors assigned to the test. For more
information, see Section 3.3: Test Setup in the Reference Manual.
You can monitor the status of a calibration request on the Sample Manager screen.
For more information, see Section 3.5: Monitoring Sample Progress.
You can view calibration results on the Calibration Data screen. For more
information, see Section 5.4: Reviewing Calibration Data.
2161B.bmp
Rack Button (Button and Field) Displays the rack ID, and the position and sample ID of each
and container in the rack.
Enter ID Select the Rack button to enter or change the rack ID. The Enter ID field is
displayed.
Calibrator (Field) Displays the name of the calibrator selected for this calibration in the
Request Calibration window (see Figure 5-9).
Calibrator Lot (List) Displays the calibrator lot selected for this calibration. If more than one lot
exists for the calibrator, you can select an alternate lot from the list.
Reagent (Fields) Displays the reagent name and lot number of the oldest reagent pack on
and board that is not empty, expired, or included in another calibration request. The
Reagent Lot number of test replicates is displayed in parentheses to the right of the reagent
name.
Test (Field) Displays the test ID of the assay to be calibrated.
Change Reag. Lot (Button) Select to change the reagent lot to another on-board reagent lot. The Set
Reagent Lot window is displayed. For more information, see the Changing the
Reagent Lot for a Calibration procedure.
Rack ID (Button) Select to enter a new rack ID, or change the current rack ID. The Enter ID
F1 field is displayed on the Rack button. For more information, see the Entering
Calibration Requests procedure or see Section 3.6: Editing Test Requests.
For multi-level calibrator samples that require two racks for calibration, select to
display a menu of available Rack buttons.
First F1
Select to enter or change the first rack ID for the first four calibrator samples.
Second F2
Select to enter or change the second rack ID.
New Request (Button) Select to exit this request and to request a new test request. There are three
F3 different test request types.
Patient/QC Requests F1
Select to request tests for patient samples and quality controls. For more
information, see Section 3.3: Patient Test Requests or see Section 6.3: Running
Quality Control Tests.
Calibration F2
Select to request an assay calibration. For more information, see
Section 5.3: Running a Calibration.
Maintenance F3
Select to request a maintenance routine. For more information, see
Section 7.1: Maintenance Overview.
Test Results (Button) Select to display the Test Results screen. The Test Results screen begins
F4 to display test requests when the samples are aliquoted by the system. For more
information, see Section 4.3: Reviewing Test Results.
Cancel Request (Button) Select to cancel the test request and to return to the Sample Manager
F6 screen. For more information, see Section 3.6: Editing Test Requests.
Entering Use this procedure to request a calibration, which can consist of up to seven calibrator
Calibration samples.
Requests
NOTES
• Before you can enter a calibration request, the calibrator must be set up, the
test must be enabled, and sufficient reagent must be in inventory. For more
information, see Section 5.2: Setting Up Calibrators, see Section 3.3: Test
Setup of the Reference Manual, or see Section 2.8: Reagent Inventory.
• You can request a calibration using an expired calibrator lot.
2164B.bmp
Calibrator List (List) Displays the calibrator name, the calibrator lot
number, and the lot expiration date for each calibrator set
up on the system. Expired calibrator lots are included in
the list until they are deleted.
Select a calibrator lot from the list for this calibration
request.
2. Select the calibrator lot for this calibration request, and then select OK F1.
3. On the Test Requests screen, enter the rack ID in the Enter ID field and press
[Enter].
(Optional) To enter a second rack ID, select Rack ID F1, then Second F2, and
then repeat step 3.
(Optional) To change the calibrator lot, select another lot from the list.
(Optional) To change the reagent lot for this calibration request, select Change
Reag. Lot. For more information, see the Changing the Reagent Lot for a
Calibration procedure.
4. Place the calibrator samples in the racks and order indicated on the Rack button of
the Test Requests screen. For more information, see the Placing Sample
Containers in Racks procedure in Section 3.4: Loading and Unloading Racks.
NOTE
If calibrator samples are not placed in the racks in the proper order, the calibration
will fail and must be requested again.
5. When you are done entering calibration requests, exit the Test Requests screen.
6. Load the racks on the instrument. For information about loading racks, see
Section 3.4: Loading and Unloading Racks.
Sample Manager
New Request F3
Calibration F2
Request Calibration
OK F1
Test Requests
2101B.wmf
Figure 5-10 Entering Calibration Requests Flowchart
Changing the When you enter a calibration request, the system assigns the oldest reagent pack on
Reagent Lot board that is not empty, expired, or included in another calibration request. Use this
for a procedure to change the assigned reagent lot.
2. From the Set Reagent Lot window, select the reagent lot to calibrate. You can
only select from reagent lots that are on board the instrument.
Each type of calibration differs in the way it calculates and reports test results. For
more information about calibration types, see Section 1.4: Assay Calibration Theory
in the Reference Manual.
Depending on the type of assay calibration and the parameters defined in the APF, the
system generates either a calibration curve or cutoff. A calibration curve plots the
measured RLUs of each replicate to the concentration of analyte. A calibration cutoff
establishes the point at which a result is determined clinically significant, according to
information provided in the reagent instructions for use.
NOTE
Do not use calibrator RLU values to predict test results of other samples.
You can view or print the most recent calibration data by reagent lot. Regardless of
the lot number, the most recent accepted calibration for an assay becomes the only
active calibration. For quantitative and semi-quantitative assays only, you can switch
between active and previously active calibrations.
The system stores all calibration data until you manually delete it. For more
information, see the Deleting Calibrations procedure in Section 5.1: Assay Calibration
Overview or see the Deleting a Calibrator Lot procedure in Section 5.2: Setting Up
Calibrators.
The Most Although all calibration data is stored on the system until you delete it, you can only
Recent view the most recent calibrations on the Calibration Data screen. For calibrations that
Calibration produce a curve, you can view data for the active, the previous active, and the last run
calibrations. For calibrations with a cutoff, you only can view data for the active and
Data
last run calibrations. Active and previous active curves can be switched. For more
information, see the Switching the Active with the Previous Active Calibration
procedure.
Active The data for the most recent calibration that passed.*
Only active calibrations are used to evaluate patient samples.
Previous The data for the calibration that was previously the active
Active calibration. You can switch the previous active calibration with
the active calibration.
Last Run The results of the most recent calibration request, which either
passed or failed.
If the last run calibration passed, the same data is displayed for
both the active and the last run calibration.*
NOTE
The format of the Calibration Data screen differs for calibration results that are
expressed as a curve or a cutoff. Information about both formats is included with
the screen descriptions.
2130A.bmp
Units (Field) Displays the units of measure for the calibrator sample type defined in the
APF. The calibrator sample type cannot be changed.
• For calibration curves, the units for the calibrator sample type can be edited for
the test. For more information, see Section 3.3: Test Setup in the Reference
Manual.
• For calibration cutoffs, the units are defined in the APF and cannot be changed.
Curve (Field) Displays which of the recent calibrations is displayed on the screen. For
more information, see The Most Recent Calibration Data topic.
Status (Field) Displays the status of the calibration request:
• Passed - The calibration results are within the acceptance criteria defined in the
APF. The calibration is either active or was previously active.
• Failed - The calibration results are not within the acceptance criteria defined in
the APF. The previous calibration remains active and this calibration is the last
run.
Reason (Field) Displays the failure code for assay calibrations with the Failed status. For a
list and explanation of failure codes, see Section 5.5: Troubleshooting Failed
Calibrations.
Cal Lot (Field) Displays the lot number of the calibrator used to calibrate the assay.
Data Table (Table) Displays a list of the calibrator samples run, the results specific to each test,
and the mean result information for test replicates.
NOTE
For some qualitative assays, three replicates are run and the results of one
replicate may be excluded when determining whether the calibration passes or
fails.
Stated Conc.
Acceptance Limits (Table) Displays the upper and lower RLU measurement limits defined in the APF
(RLUs) for each calibrator level of this assay. RLU readings beyond these limits result in
(for cutoffs only) failed calibrations.
• Level - The calibrator level.
• Replicate Low - The lowest acceptable RLU reading for a test replicate.
• Replicate High - The highest acceptable RLU reading for a test replicate.
• Mean Low - The lowest acceptable calculated RLU mean.
• Mean High - The highest acceptable calculated RLU mean.
Select Curve (Button) Select to display a menu of the most recent calibrations for this reagent
F1 lot.
Active F1
or Displays the active calibration for the assay.
Previous Active F2
Select Limits
F1 Available for curves only. Displays the previous active calibration for the assay.
Last Run F3
Viewing Use this procedure to view detailed calibration data, along with the generated
Calibration calibration curve or cutoff.
Data
NOTE
For some qualitative assays, three replicates are run and the results of one
replicate may be excluded when determining whether the calibration passes or
fails.
• Previous Active F2 - Available for curves only. Displays the previous active
calibration for the assay.
• Last Run F3 - Displays the last run calibration.
For more information about active, previous active, and last run calibrations, see The
Most Recent Calibration Data topic.
Switching the For quantitative and semi-quantitative assays only, use this procedure to switch the
Active with active calibration with the previous active calibration. For more information on active
the Previous and previous active calibrations, see The Most Recent Calibration Data topic.
Active NOTE
Calibration
This procedure will not change calibration expiration dates or recalculate results
for patient samples that have already been run.
4. Confirm your selection and then select OK F1 to switch the active with the
previous active calibration.
Entering, Use this procedure to enter, view, or edit a calibration comment. Comments print on
Viewing, or the Calibration Data Report.
Editing a System Mode: Any Mode
Calibration 1. From the Calibration screen, select the Assay button for the assay calibration you
Comment intend to comment on.
(Optional) To delete a comment, select all of the text in the Calibration Comment
window and press [Delete].
Printing a Use this procedure to print the information from the Calibration Data screen
Calibration manually. You can also set the Calibration Data Report to print automatically after a
Data Report calibration is run. For more information about report setup, see Section 2.2: System
Setup in the Reference Manual.
System Mode: Any Mode
1. On the Calibration screen, locate the Assay button for the calibration data you
intend to print. If needed, apply a different filter. For more information, see the
see the Applying a Calibration Filter topic in Section 5.1: Assay Calibration
Overview.
Laboratories, Inc.
Laboratory A
Calibration Data Report 123 Lake Street
tUniCel DxI 800, Access Immunoassay System Townsville, ST 33333
S/N 610240 101-555-2323 ext. 109 Jane Smith
Stated Conc. ng/mL RLU 1SD %CV Calculated Conc. ng/mL Flags
S0 0.0 11238 0.0
S0 0.0 10730 No Value
Mean 10984.0 359.2 3.3 0.0
S1 10.0 129201 10.3
S1 10.0 127697 10.2
Mean 128449.0 1063.5 0.8 10.3
S2 50.0 586338 49.3
S2 50.0 597980 50.3
Mean 592159.0 8232.1 1.4 49.8
S3 200.0 2202860 197.6
S3 200.0 2239570 201.2
Mean 2221215.0 25957.9 1.2 199.4
S4 500.0 4795090 495.6
S4 500.0 4869380 505.6
Mean 4832235.0 52531.0 1.1 500.6
S5 1500.0 9894830 1535.7
S5 1500.0 9660040 1464.2
Mean 9777435.0 166021.6 1.7 1500.0
Page 1 of 1
Technologist ______________________________________________ Printed 03/03/10 11:05 AM
2132B.bmp
Figure 5-13 Calibration Data Report (Curve) Example
Assay calibrations can fail for various reasons. To rule out and fix the most significant
problems first, troubleshoot each failure in the following order:
1. Troubleshoot event log errors. For more information, see Section 5.3:
Troubleshooting Caution and Warning Events in the Reference Manual.
2. Troubleshoot test result flags. For more information, see Section 4.3: Reviewing
Test Results.
3. Run all System Check routines. If results are out of range, contact Technical
Support. For more information, see Section 4.4: System Check Routines in the
Reference Manual.
4. Troubleshoot according to calibration failure codes and assay calibration
troubleshooting tables.
Calibration When a calibration fails, the system displays a failure code in the Reason field on the
Failure Codes Calibration Data screen and the Calibration Data Report.
Bad Fit The curve does not meet the acceptance criteria defined in the
APF. Possible causes are poor precision or a curve that is too
flat or too steep.
CV Std 0 The %CV of the S0 (zero) calibrator replicates does not meet
the acceptance criteria defined by the APF.
Insuff Data The system did not have enough data to perform the
necessary calibration calculations. This failure occurs when
two or more replicates are not calculated, usually due to a
short sample or an instrument error.
Limits The RLUs of either a replicate or the mean of the replicates
are outside the accepted ranges defined in the APF.
Max Iterate The system was unable to create a curve before performing
the maximum number of calculation iterations. The system
attempts to calculate the curve 100 times before it generates
this error.
No Fit The system was unable to fit a curve to the data points.
Possible causes include math rule violations, such as dividing
by zero or calculating the log of a negative number.
Resp Delta The total response of the curve, which is the difference
between the highest RLU reading and the lowest RLU
reading, is too small.
Temp. Out The upper cabinet temperature changed more than 6ºC while
the system processed the curve replicates.
Assay When a calibration fails, look for and troubleshoot system errors. For more
Calibration information, see the introduction to this section (see Section 5.5: Troubleshooting
Problems Failed Calibrations).
After ruling out or resolving system errors, use the following table to help
troubleshoot remaining calibration problems.
Although precision Multi-level calibrators aliquoted out of 1. Place the calibrators in the rack in the order
is good, the order, resulting in: displayed on the Test Requests screen (see
calibration fails for • A calibration curve that does not ascend Section 5.3).
any reason other or descend smoothly 2. Repeat the calibration.
than Insuff Data
• A calibration cutoff that is opposite of
the expected result
Expired calibrator lot (CLX result flag) 1. Find the expired calibrator lot on the Calibrator
Setup screen (see Section 5.2).
2. Delete the calibrator lot, discard all calibrator sets
of this expired lot, and set up a new, unexpired
calibrator lot.
3. Repeat the calibration using the same reagent lot,
if available.
Calibrator set beyond its stability date 1. Compare the date of thaw or the date of expiration
recorded on the calibrator vials to the product
stability information provided in the reagent
instructions for use.
2. If the calibrator vial is beyond its stability date,
discard the calibrator.
3. Repeat the calibration using the same calibrator
lot and reagent lot, if available.
Calibrator unstable or contaminated due to Repeat the calibration. If possible, repeat with a new
improper storage or handling set of the same calibrator lot.
Reagent pack unstable or contaminated 1. Unload the reagent pack and load a new reagent
due to improper storage or handling pack (see Section 2.7).
2. Repeat the calibration.
Although precision A rack of patient samples was processed 1. Unload the racks (see Section 3.4) and look for
is good, the as a rack of calibrators because the racks duplicate rack IDs.
calibration fails for had the same rack ID (when processing 2. If a patient sample rack has the same rack ID as
any reason other LIS requests with host query) the calibration request, process the patient samples
than Insuff Data again (see Section 3.2), making sure that:
(continued)
• Adequate sample is in the container for
processing (see Section A.2).
• The rack has a unique rack ID.
3. Request the calibration again, making sure that
each rack has adequate sample for processing and
that calibrators are loaded in racks with unique
rack IDs.
Although precision Wrong calibrator placed in rack 1. Place the correct calibrators in the rack.
is good, the 2. Repeat the calibration.
quantitative curve
is flat or the
qualitative result is
No Value;
calibration fails for
any reason other
than Insuff Data
Routine maintenance not performed 1. Perform routine maintenance (see Section 7.1).
2. Repeat the calibration.
Poor precision, Problems with the pipettor, substrate, or 1. If the results of all System Check routines are
and calibration RV wash system within the expected ranges, these systems are not
fails for any reason the cause of the calibration failure.
other than Insuff
Data 2. If the results of any of the System Check routines
are not within the expected range, contact
Technical Support for assistance in
troubleshooting.
Reagent gone because it leaked out of the 1. Unload the reagent pack and load a new reagent
pack during off-board storage or a partial pack (see Section 2.7).
pack from another Access Immunoassay 2. Repeat the calibration.
System was loaded on the UniCel DxI
system
RLUs are too low Some reagent loss during off-board Repeat the calibration with the open reagent pack.
at one end of the storage (first few test replicates between Note:
calibration curve, 6,000-9,000 RLU, remaining replicates During processing, the instrument attempted to aliquot
and calibration acceptable) the lost reagent from the reagent pack and aspirated air
fails for any reason for the first few test replicates, resulting in low RLUs.
other than Insuff Once the pipettor reached the actual level of reagent in
Data the pack, the RLUs returned to acceptable levels.
Calibration fails Quantity of calibrator not sufficient for 1. Calculate the correct volume of calibrator needed
for the reason testing (QNS result flag and event) for the number of replicates and type of sample
Insuff Data container used (see Section A.2 for the equation).
2. Pipette the amount of calibrator calculated in
step 1 into the appropriate sample container.
3. Be sure the rack and rack ID are correct (see
Section 2.3 in the Reference Manual).
4. Repeat the calibration.
5. If the calibration fails again for the same reason,
contact Technical Support for assistance.
Two or more replicates not calculated due 1. Review the Event Log messages (see Section 5.2
to instrument error in the Reference Manual) for device errors prior to
the calibration failure.
2. Review the Test Results screen for flags (see
Section 4.3).
3. Troubleshoot the device errors. If necessary,
contact Technical Support for assistance.
4. After resolving the device errors, repeat the
calibration.
Calibrator sample missing from required 1. Place the sample containers in the rack in the
position order displayed on the Test Requests screen (see
Section 5.3).
2. Repeat the calibration.
6 Quality Control
6.1 Quality Control Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Quality controls are samples of known concentration that are used to monitor system
performance and verify calibration. Quality control (QC) samples can be run
separately, or in a rack with patient samples.
Quality control operations for the UniCel DxI Access Immunoassay System can be
grouped into four main processes (see Figure 6-1):
• Setting up quality controls, in which you provide information about the QC
samples, including expected ranges of results, rules for flagging results that
are outside of the expected ranges, and pipettors to use for QC tests
• Running QC tests
• Reviewing the results, comparing results from different QC test runs, printing
QC reports, and so on
• Troubleshooting out-of-range QC results
When the system flags a result because of a violation of an applied QC rule, the
Quality Control button turns red until it is selected. For information about reviewing
test results and flags, see Chapter 4: Test Results. For help on how to interpret failed
quality controls, see Section 6.7: QC Troubleshooting Overview.
Yes
Troubleshoot out-of-range
Are all results in No QC, correct problem, or
range? adjust QC ranges, as
indicated
Yes
2048B.wmf
Figure 6-1 Quality Control Flowchart
2062B.bmp
QC Buttons (Buttons) Select to view the chart and data for a particular quality control, or to
compare charts for two or three quality controls.
One button is displayed for each assay that has been run successfully at least once
on a specific quality control lot. QC buttons are labeled with the QC name, lot
number, and test name.
QC buttons turn red to indicate out-of-range results. Select a red QC button to view
the results. Once selected, the QC button returns to neutral.
No QC buttons can be selected if a Comparison button is selected.
Filter (Button) Select to change the filter and sort order parameters for quality controls.
F1 Comparison buttons are always displayed before QC buttons.
Review Chart and (Button) Select to view the accumulated data for the selected QC button. This
Data button is not available if more than one QC button is selected, or if a Comparison
F2 button is selected.
Compare Charts (Button) Select to view the charts for two or three quality controls. This button is
F3 available only when two or three QC buttons are selected, or when a Comparison
button is selected.
Delete Comparison (Button) Select to delete one or more saved comparisons.
F4
QC Setup (Button) Select to display the QC Setup screen to add, edit, or delete quality
F5 controls.
Print (Button) Select to print a report of the quality controls that are currently set up. For
F7 more information, see the Printing the QC Review Report topic.
Applying A quality control filter is a set of parameters you can change to reduce or expand the
a Quality number of quality controls displayed on the Quality Control screen.
Control Filter Use this procedure to apply a different filter or sort order to the Quality Control
screen.
0758A.bmp
(Optional) Select the first and second sort order you want to apply.
3. Select OK F1.
The Quality Control screen displays all quality controls that meet the selected
parameters.
NOTE
Comparison buttons are always displayed, and always precede QC buttons.
Comparison buttons are always sorted alphanumerically by comparison name, in
ascending or descending order, as determined by the setting of the first sort order
Ascending box.
Westgard QC Westgard QC rules are interpretive rules developed to avoid excessive false rejection
Rules rates, to improve quality monitoring, and to decrease subjectivity of data analysis.
Each laboratory determines the QC rules to apply and the course of action to take
when a result violates a rule. Understand how your laboratory uses QC rules before
you set up controls.
Application of a QC rule consists of screening quality control test results for the
number of standard deviations (SD) above or below the mean, and for the number of
consecutive occurrences. Data points are colored red to mark violations of rules. For
each combination of control lot and reagent, you can apply one or more of these QC
rules:
• 12s (1-2s) — The system marks a data point if it is more than 2 SD from the
mean.
• 22s (2-2s) — The system marks a data point if two consecutive points of a
given quality control are more than 2 SD from the mean in the same
direction. The system only marks the second point.
• 13s (1-3s) — The system marks a data point if it is more than 3 SD from the
mean.
• 41s (4-1s) — The system marks a data point if four consecutive points of a
given quality control are more than 1 SD from the mean in the same
direction. The system only marks the fourth point.
• 10x (10x) — The system marks a data point if ten consecutive points of a
given quality control are on the same side of the mean (for example, ten
points are above the mean). The system only marks the tenth point.
QC Status In addition to being provided on screens that are specific to quality controls, QC status
Indicators and information is provided in a number of other places.
Flags Quality Control Button
The Quality Control system status button appears in the row of system status buttons
at the top of every screen, and turns red when a QC test produces results that violate
one or more Westgard QC rules. Select the Quality Control button to go directly to
the Quality Control screen from any screen. If the Quality Control button is red, it
returns to neutral when selected.
2010A.bmp
Figure 6-4 Quality Control System Status Button
QC Buttons
The QC buttons on the Quality Control screen turn red when a test produces results
that violate one or more Westgard QC rules. Select the button to view the chart and
data for that QC. Once selected, a red QC button returns to neutral.
QCF Indicates that the QC test violated one or more Westgard QC rules
QEX Indicates that the QC lot has expired
Printing the Use this procedure to print the QC Review Report. This report lists all of the quality
QC Review controls in the current filter of the Quality Control screen. To print the report, select
Print F7.
Report
System Mode: Any Mode
Laboratories, Inc.
Laboratory A
QC Review Report 123 Lake Street
tUniCel DxI 800, Access Immunoassay System Townsville, ST 33333
S/N 610240 101-555-2323 ext. 109 Jane Smith
Page 1 of 1
Technologist ____________________________________________ Printed 10/10/03 02:43 PM
2081B.bmp
Figure 6-5 QC Review Report Example
Each new lot of a quality control must be set up before it can be selected in a QC test
request.
To set up a quality control lot, enter the name of the control, the lot number, the
expiration date, and the sample type, then select which assays can be run with this
control lot. For each assay, provide the expected mean and standard deviation for
results, and choose one or more Westgard QC rules for flagging out-of-range results.
A quality control lot can be configured for up to 50 assays.
To set up QC for onboard dilution tests, use QC samples with high concentrations of
analyte. Onboard dilution tests have a lowest reportable result that is slightly less than
the highest calibrator for the assay. For more information about onboard dilution tests,
see the Onboard Dilutions topic in Section 1.3 of the Reference Manual.
Establish an acceptable range of values (mean and standard deviation) for quality
controls specific for each UniCel DxI instrument. For each new lot, use the
manufacturer's suggested ranges from the directional insert as a guideline for the
expected mean and standard deviation (SD), until you have processed enough quality
control samples to determine a mean and SD specific for the instrument you use. For
more information, see Section 6.6: Establishing QC Ranges.
2063B.bmp
Delete Control (Button) Select to permanently remove the selected quality control and all
F6 associated data points from the system.
NOTE
This function requires you to enter the system password.
Adding or Use this procedure to add a new quality control lot or to edit an existing quality
Editing a control lot.
Quality System Mode: Any Mode
Control Lot 1. From the Quality Control screen, select QC Setup F5 to display the QC Setup
screen.
NOTE
The Add Control window is similar to the Edit Control window except that:
• You can view other quality controls and lot numbers from the Edit Control
window.
• You cannot enter the expiration date or the sample type from the Edit Control
window.
2181B.bmp
Control (List or Field) Select a quality control to edit, or enter the name of a quality control
to add.
Lot Number (List or Field) Select the lot number of a quality control to edit, or enter the lot
number for a quality control you are adding. Lot numbers must be unique.
Expiration (Field) Displays the expiration date for the quality control. If you are adding a
quality control, enter the expiration date.
Sample Type (List or Field) Displays the sample type for the quality control. If you are adding a
quality control, select a sample type from the list.
Test Name (List) Select a test name from the list of tests enabled on the instrument. If a test
has been disabled, the Test Name field cannot be changed. Also, if the selected
control row for an assay has already been defined (that is, selections were made for
all available fields), the Test Name field cannot be changed.
For more information about enabling and disabling a test, see the Disabling a Test
topic in Section 3.3: Test Setup of the Reference Manual.
Mean (Field) Enter the expected mean for the test. If a test has been disabled, the Mean
field cannot be changed.
SD (Field) Enter the expected standard deviation for the test. If a test has been
disabled, the SD field cannot be changed.
Unit (Field) Displays the units of measure used for the test results, as defined for the
selected test and sample type.
Westgard QC (Boxes) Select the rules to apply for the new control. If a test has been disabled, the
Rules Westgard QC Rules fields cannot be changed.
Your laboratory determines the Westgard QC rules to apply and the course of
action you need to take when a result violates a rule. Understand how your
laboratory uses QC rules before you add controls. For more information, see the
Westgard QC Rules topic in Section 6.1: Quality Control Overview.
Designated (Field) Displays the reagent pipettors designated to process the test. Each
Pipettors designated pipettor runs one replicate of each requested test. If a test has been
disabled, the designated pipettors fields cannot be changed.
Clear Row (Button) Select to clear the information from a selected row.
F3
Designate Pipettor (Button) Select to designate one or more reagent pipettors to use for the selected
F4 test. For more information, see the Designating Reagent Pipettors for a Quality
Control topic. If a test has been disabled, the Designate Pipettor F4 button is gray
and not available.
Figure 6-7 Add Control and Edit Control Window Descriptions (continued)
3. Enter or select the name of the quality control. The name is limited to
15 characters. Allowable characters are 0-9, A-Z, a-z, /, +, and -.
4. Enter or select the lot number for the quality control. Lot numbers must be unique
and are limited to 12 characters. Allowable characters are 0-9, A-Z, and a-z.
NOTE
Some multi-level quality controls have the same lot number for each level. To
make the lot number unique, add the level (such as 1, 2, or 3), or some other
designator, to the lot number.
5. If you are adding a new quality control lot, enter the expiration date for the
quality control.
The expiration date format must follow the date format selected on the System Setup
screen. For more information about the date format, see Section 2.2: System Setup in
the Reference Manual.
6. If you are adding a new quality control lot, select a sample type from the Sample
Type list.
7. If you are adding a test to a new or existing quality control, select an assay from
the Test Name list.
NOTE
During the initial setup of a new quality control lot, use the ranges (mean and
standard deviation) suggested by the manufacturer in the directional insert, then
establish QC ranges for each instrument. For more information, see
Section 6.6: Establishing QC Ranges.
9. Enter the expected standard deviation (SD) for the selected assay.
(Optional) To add or edit Westgard QC rules to apply to QC test results for the
selected assay, select or clear one or more boxes. For more information, see the
Westgard QC Rules topic in Section 6.1: Quality Control Overview.
(Optional) To designate one or more reagent pipettors for the selected assay,
select Designate Pipettor F4. For more information, see the Designating
Reagent Pipettors for a Quality Control topic.
10. To add another assay to a multi-analyte quality control, select a blank row and
repeat the preceding steps, beginning with step 7.
11. To remove an assay from the quality control, select the row for that assay, then
select Clear Row F3.
(Optional) To edit an additional quality control without leaving the Edit Control
window, select it with the Control or Lot Number list. You will be prompted to
save the current quality control if you have made any changes to it.
12. To save the changes and exit the window, select OK F1.
The system adds any new quality controls to the QC Setup screen.
(Optional) To document your changes, press the Print Screen key on the
keyboard. The system prints the screen image in a report format including the
date and time.
Designating Use this procedure to designate one or more reagent pipettors to be used for a QC test,
Reagent or to allow the test to randomly use any assigned pipettor.
Pipettors for a
NOTES
Quality
• When you designate pipettors to use for a QC test, the test runs once on each
Control
designated pipettor for each requested test and replicate.
• A pipettor can only be designated for a QC test if it has been assigned to that
assay on the Test Setup screen. For more information, see Section 3.3: Test
Setup in the Reference Manual.
• To perform this procedure from the Add Control window, begin with step 4.
2224B.bmp
Currently (Fields) Displays a check mark under the number of each reagent pipettor that has
Assigned for been assigned to this assay. For more information, see Section 3.3: Test Setup in
[assay] the Reference Manual.
Designate (Option and Boxes) Select one or more assigned reagent pipettors to be used for the
Pipettors for this quality control. The quality control will be run with each designated pipettor.
Control Boxes are not available for pipettors that are not assigned to this assay.
Any Assigned (Option) Select to allow the system to use any assigned pipettor when the quality
Pipettor control is run.
5. To designate one or more pipettors to be used each time this quality control test is
run, select the Designate Pipettors option, then select the boxes for pipettors to
use. Only pipettors that are assigned to this assay can be designated for a QC test.
NOTE
If a pipettor that is designated for a QC test is disabled or is no longer assigned to
the assay when the QC test is requested, the replicates for that pipettor will not
run. The Supplies Required button turns yellow until the pipettor is enabled and
assigned. For more information, see the Enabling and Disabling Reagent
Pipettors topic in Section 2.3: Supplies Setup in the Reference Manual. Also see
the Assigning Pipettors to Tests topic in Section 3.3: Test Setup in the Reference
Manual.
(Optional) To use any assigned pipettor for the test, select the Any Assigned
Pipettor option.
(Optional) To document your changes, print the QC Review Report. For more
information, see the Printing the QC Review Report topic in Section 6.1: Quality
Control Overview.
Deleting a Use this procedure to delete a quality control and all the data points associated with
Quality that quality control lot.
Control
NOTES
• Before deleting a quality control, you may want to print the data or copy it to
a disk or flash drive.
• You need the system password to use this feature. If you do not know the
password, contact your lab supervisor. For more information about
passwords, see Section 2.2: System Setup in the Reference Manual.
Run quality control (QC) tests on QC lots that are set up on the system. (For setup
information, see Section 6.2: Setting Up Quality Controls.) You can run QC tests
separately, or in a rack with patient samples.
To best simulate the characteristics of patient samples, use quality control samples
prepared from material similar to the patient samples being tested. For example, if
testing serum, use serum-based quality control samples.
QC test results are reported, with any test result flags, on the Test Results screen. (For
more information, see Chapter 4: Test Results.) QC test results are also stored in a QC
database, and can be reviewed collectively on the QC Chart and Data screen. For
more information, see Section 6.4: Reviewing Quality Control Charts and Data.
When a QC test result violates one or more Westgard QC rules, the Quality Control
status button turns red, and on the Quality Control screen, the button for that quality
control also turns red. On the QC Chart and Data screen, data points that violate
Westgard QC rules are displayed in red, and in the data table, the rules that were
violated are displayed in red, with asterisks on both sides. To troubleshoot a QC result
that violates Westgard QC rules, see Section 6.7: QC Troubleshooting Overview. For
more information on Westgard QC rules, see Section 6.1: Quality Control Overview.
When a QC test fails to produce results, nothing is added to the QC database, and one
or more fatal flags are displayed on the Test Results screen, indicating the nature of
the problem. To troubleshoot a failed QC test, see Section 6.7: QC Troubleshooting
Overview.
Entering QC Use this procedure to enter a test request for a quality control sample or a multi-level
Test Requests set of samples.
NOTES
• You can only order tests that are defined for the selected control lot. For
information on setting up a quality control lot, see Adding or Editing a
Quality Control Lot.
• You can run tests on patient samples and QC samples in the same rack, even
if different tests are requested for each sample. For information on entering
patient test requests, see Section 3.3: Patient Test Requests.
System Mode: Any Mode
1. From the Main Menu, select Sample Manager F1 to display the Sample Manager
screen.
2. Select New Request F3, then select Patient/QC Requests F1 to display the Test
Requests screen. For more information about this screen, see Section 3.3: Patient
Test Requests.
3. Enter the rack ID in the Enter ID field and press [Enter].
4. Select Request QC F5.
The Request QC window is displayed.
2047C.bmp
6. Select the first quality control that you requested in the previous step.
NOTE
If you cannot find a Test button, it might not be enabled. For information on
configuring and enabling tests, see Section 3.3: Test Setup in the Reference
Manual.
2060C.bmp
Reagent (Column) Lists the reagents that will be used to run the selected tests. When you
select a reagent from the list, the associated test number is displayed in the Test
field.
Reagent Lot (Column) Displays the lot number that has been assigned for each reagent, for the
selected sample position. If no reagent lot has been assigned, nothing is displayed,
and the instrument will use any available lot. For more information, see the
Reagent Pack Usage topic in Section 1.2: Assay Technology of the Reference
Manual.
Test (Field) Displays the test number associated with the selected reagent.
Change Reag. Lot (Button) Select to display a list of on-board lots of the reagent selected in the
Reagent field. Select a reagent lot from the displayed list to assign a reagent lot to
use for the QC test.
Remove (Button) Select to remove one replicate of a test from a selected sample position. If
more than one test is assigned to that sample position, select the associated reagent
from the Reagent list, or enter the number of the test in the Test field.
Test Buttons (Buttons) Select to assign a test or test panel to the selected sample position. The
Test buttons that are displayed depend on the selected sample.
Select a test more than once to run replicates. Select multiple tests for multi-analyte
quality controls.
8. Repeat steps 6 and 7 for each quality control test requested in step 5.
(Optional) Assign an on-board reagent lot to use for a particular sample. For more
information, see the Assigning a Reagent Lot for QC Test Requests procedure.
10. Place the QC samples in the rack entered in step 3, and load the rack on the
instrument. For information on placing sample containers in racks and loading
racks on the UniCel DxI instrument, see Section 3.4: Loading and Unloading
Racks.
NOTE
If you are placing the sample containers in a rack that is designated for reserve
volume, be sure that the sample containers contain enough sample to provide the
reserve volume. For more information, see the Calculating Minimum Sample
Volume procedure in Section A.2: Sample Containers.
Sample Manager
New Request F3
Patient/QC Requests F1
Test Requests
Request QC F5
Request QC
OK F1
Test Requests
2057B.wmf
Figure 6-11 Entering QC Test Requests Flowchart
Editing QC Use this procedure to edit a QC test request that has been entered, but not yet run.
Test Requests
NOTE
You can edit QC test requests any time before the instrument reads the rack and
sample container bar codes, or while the sample status is Requested. For more
information about sample status, see Section 3.5: Monitoring Sample Progress.
(Optional) To change the rack ID, select the Rack button or Select Rack ID F1,
and enter the new rack ID.
(Optional) To assign a different QC lot number, select the new lot number from
the Patient ID/Lot Number list.
NOTE
Only QC lots that have been set up on the instrument can be selected. For more
information, see Section 6.2: Setting Up Quality Controls.
Assigning a Use this procedure to assign an on-board reagent lot to use for a particular QC sample.
Reagent Lot
System Mode: Any Mode
for QC Test
1. From the Test Requests screen, select the sample position for which you want to
Requests assign a reagent lot.
3. Select Change Reagent Lot to display the Set Reagent Lot window.
4. Select the reagent lot to assign, or select Any to allow the instrument to use any
available lot, according to its reagent pack usage rules. For more information, see
the Reagent Pack Usage topic in Section 1.2: Assay Technology of the Reference
Manual.
The number of the selected lot is displayed in the Reagent Lot column. All replicates
of a test for a particular sample position will use the same reagent lot, but if the same
test is assigned to more than one sample position, each position can be assigned a
different reagent lot.
NOTE
You can only select reagents that are loaded on the instrument.
5. Select OK F1 to save the change and return to the Test Requests screen.
The UniCel DxI system keeps a QC database of all test results for each quality
control, and performs statistical calculations on the information stored in the database.
The statistical calculations for the selected quality control are performed on all data
points within the active date range, and are recalculated each time new results are
added, data points are omitted, or the active date range is changed.
The following statistical data is displayed on the QC Chart and Data screen:
• Date range used for calculations and data displays
• Calculated mean of the results
• Calculated standard deviation (SD) of the results
• Calculated coefficient of variation (%CV) of the results
• The number of points used in the calculations
2058B.bmp
Reagent Lot (Field) Displays the lot number of the reagent pack that is selected in the filter
and displayed on the chart and in the data table. For information about viewing
QC data for a specific reagent lot, see the Applying a QC Chart and Data Filter
topic.
Start (Field) Displays the first date of the date range included on the chart and table.
By default, date range is the past 31 days. For information about viewing QC
data for a specific date range, see the Applying a QC Chart and Data Filter topic.
End (Field) Displays the last date of the date range included on the chart and table. By
default, date range is the past 31 days. For information about viewing QC data
for a specific date range, see the Applying a QC Chart and Data Filter topic.
Mean (Field) Displays the calculated mean of the points within the date range. The
value displays with the number of decimal places defined for the assay, plus one
additional decimal place, up to a maximum of four decimal places.
Omitted points are not included in the calculations.
For information about defining the decimal places for an assay, see Section 3.3:
Test Setup in the Reference Manual.
SD (Field) Displays the calculated standard deviation (SD) for the points within the
date range. The value displays with the number of decimal places defined for the
assay, plus one additional decimal place, up to a maximum of four decimal
places. Omitted points are not included in the calculations.
%CV (Field) Displays the calculated coefficient of variation (%CV) of the points
within the date range. The value always displays with two decimal places.
Omitted points are not included in the calculations.
Points (Field) Displays the number of data points within the date range used to calculate
the mean, SD, and %CV. Omitted points are displayed on the chart with a slash
through them, but are not included in the calculations.
Point Cursor (Chart item) Displays as a blue vertical line on the graph, and marks the selected
point. Data for the selected point is displayed in the table beneath the chart. To
change the selected point, use the left and right Arrow buttons beneath the chart,
or select a point in the data table. Also see the Levey-Jennings Charts topic.
Calibration Line (Chart item) Displays as a dotted vertical line on the graph, and marks the first
result calculated after a change in calibration. Also see the Levey-Jennings
Charts topic.
Y-axis Left Scale (Chart item) Displays the seven y-axis coordinates expressed in result values.
These are the stated values for the mean and 1SD, 2SD and 3SD above and
below the mean. The value displays with the number of decimal places defined
for the assay, plus one additional decimal place, up to a maximum of four
decimal places. Also see the Levey-Jennings Charts topic.
Y-axis Right Scale (Chart item) Displays the seven y-axis coordinates expressed in standard
deviations. These mark the mean and 1SD, 2SD and 3SD above and below the
mean. Also see the Levey-Jennings Charts topic.
X-axis (Chart item) Displays the dates for the chart. Also see the Levey-Jennings Charts
topic.
Points (Chart item) Displays the plotted numerical value of the result. A slash through
the point indicates an omitted point. Points greater than 3 SD above the mean are
indicated as solid triangles on the +3 SD line, and points greater than 3 SD below
the mean are indicated as solid triangles on the -3 SD line. Points that are flagged
because they violate QC rules are displayed in red. Also see the Levey-Jennings
Charts topic.
Arrow Buttons (Buttons) Select to move the point cursor on the chart left or right to the next data
point, and to select the corresponding row of the data table. The chart and table
scroll as necessary to display the new selection.
Value (Column) Displays the test result and the unit of measure.
SDI (Column) Displays the deviation of the result from the expected mean, expressed
in standard deviation intervals, based on the standard deviation specified during
QC setup. The value always displays with one decimal place. For more
information, see Section 6.2: Setting Up Quality Controls.
Completion (Column) Displays the date and time of the result.
Westgard QC Rules (Column) Displays the five available Westgard QC rules. The rules applied to
this quality control are displayed in bold black characters. The rules violated by
this result are displayed in red, with asterisks on both sides.
Comment (Button) Select the button to read or enter a comment for a quality control result.
• Select the Pen button to enter a new comment.
• Select the Notepad button to read or edit a previous comment.
Omit (Box) Select to omit the result from QC calculations. The point remains on chart,
but has a slash through it.
Pipettor (Column) Displays the number of the reagent pipettor used for the QC test.
Filter (Button) Select to limit displayed data to certain reagent lots, date ranges, and
F1 reagent pipettors. For more information, see the Applying a QC Chart and Data
Filter topic.
Delete Point (Button) Select to delete the currently selected point and all data associated with
F5 that point from the system. For more information, see the Deleting a QC Data
Point topic.
NOTE
You need a password to use this feature.
Copy to Disk (Button) Select to copy QC data. For more information, see the Copying QC
F6 Data to a Disk or Flash Drive procedure.
Print (Button) Select to print the QC chart and data for the currently selected quality
F7 control and date range. For more information, see the Printing a QC Chart or
Data Report topic.
The Y-axis of the chart indicates test results and their deviation from the expected
mean. The Y-axis right scale is graduated in standard deviation intervals from +3 SD
to -3 SD, and the Y-axis left scale shows the corresponding test result values. The
Y-axis label values display with the number of decimal places defined for the assay,
plus one additional decimal place, up to a maximum of four decimal places.
Points that are above +3 SD are displayed as solid triangles on the +3 SD line. Points
that are below -3 SD are displayed as solid triangles on the -3 SD line.
Points that are flagged because they violate Westgard QC rules are displayed in red.
For more information, see the Westgard QC Rules topic in Section 6.1: Quality
Control Overview.
2059B.bmp
Figure 6-13 Levey-Jennings Chart
Use the Levey-Jennings chart to observe the range of quality control test results over a
period of time. QC results that are within an acceptable range form a normal
distribution pattern with equal numbers of points above and below the mean. A
normal distribution has approximately:
• 70% of the points between +1 SD and -1 SD
• 25% between +1 SD and +2 SD or between -1 SD and -2 SD
• 5% between +2 SD and +3 SD or between -2 SD and -3 SD
Applying a QC Use a QC chart and data filter to control the amount of data displayed on the QC Chart
Chart and and Data screen and printed in reports. The filter limits data according to the
Data Filter following parameters:
• By date range - You can display data for the last week, the last month, or
you can specify a starting and ending date.
• By reagent lot - You can display data from a single reagent lot or all reagent
lots.
• By reagent pipettor - You can display data from any combination of the
available reagent pipettors.
Use the following procedure to apply a filter to the QC Chart and Data screen:
2225A.bmp
NOTES
• When viewing charts for date ranges greater than 31 days, the last date
displayed on the chart will be the date of the most recent point in the date
range.
• When printing charts for date ranges greater than 31 days, the charts will
span multiple pages, with 31 days represented on each page, even if no data
exists for a particular 31-day period.
(Optional) Select a single reagent lot to view, or select All to view data for all lots
of the reagent for the selected test.
(Optional) Select one or more reagent pipettors to view only the data associated
with those pipettors.
3. Select OK F1 to save your changes and exit the window.
Viewing QC Use this procedure to view the QC data for a selected test.
Charts and System Mode: Any Mode
Data 1. From the Quality Control screen, locate and select the quality control you want to
view.
NOTE
To limit or reorganize the quality controls displayed on the Quality Control
screen, apply a filter. For more information, see the Applying a Quality Control
Filter topic in Section 6.1: Quality Control Overview.
2. Select Review Chart and Data F2. The QC data for the past 31 days is displayed
on the QC Chart and Data screen.
If no data is visible, or if the data points on the chart are too close together, adjust the
date range. For more information, see the Applying a QC Chart and Data Filter topic.
NOTE
Quality control tests may be completed while you are reviewing this screen. To
refresh the information displayed, exit the screen and display it again.
(Optional) Apply a QC chart and data filter to display data for certain date ranges,
certain reagent lots, or certain reagent pipettors. For more information, see the
Applying a QC Chart and Data Filter topic.
Entering Use this procedure to enter, edit, or delete comments for quality control test results on
Quality the QC Chart and Data screen.
Control System Mode: Any Mode
Comments 1. From the Quality Control screen, locate and select the quality control to
comment.
3. From the QC Chart and Data screen, select the Pen or Notepad button next to the
appropriate result to display the QC Comment window.
The pen indicates that no comments are associated with the result.
0727A.bmp
The notepad indicates that comments are associated with the result.
These comments were entered in the QC Comment window.
0728A.bmp
Omitting a QC Use this procedure to omit a point from the QC calculations for the mean, SD, and
Data Point %CV. Omitted points continue to display on the screen and are included in reports.
NOTE
If a data point is omitted by the system due to a fatal flag or other error condition,
the omission is permanent and the Omit box cannot be cleared.
Deleting a QC Use this procedure to permanently delete a selected point from the screen, reports, and
Data Point QC calculations.
NOTE
You need the system password to use this feature. If you do not know the
password, contact your lab supervisor. For more information about passwords,
see Section 2.2: System Setup in the Reference Manual.
Printing a QC Use this procedure to print a QC Chart Report, a QC Data Report, or both reports.
Chart or Data
System Mode: Any Mode
Report
1. From the Quality Control screen, locate and select the quality control you want to
print.
(Optional) Adjust the date range to include the points you want to print. For more
information, see the Applying a QC Chart and Data Filter topic.
NOTE
When printing charts for date ranges greater than 31 days, the charts will span
multiple pages, with 31 days represented on each page, even if no data exists for a
particular 31-day period.
2090A.bmp
Report Type (Boxes) Select to print the chart, the data table, or both
reports. For an example of the data report format, see
Figure 6-16. For an example of the chart report format, see
Figure 6-17.
Print Entire Date (Box) Select to print the entire date range selected in the
Range QC Chart and Data screen. If the date range is more than 31
days, charts continue on additional pages.
Clear to print only the data on the screen. Chart reports are
limited to 31 days, and data reports print only four lines.
Also see the Applying a QC Chart and Data Filter topic.
Figure 6-15 Print (QC Chart and Data) Window Descriptions (continued)
(Optional) To print the entire date range selected in the QC Chart and Data
screen, select Print entire date range. For more information, see the Applying a
QC Chart and Data Filter topic.
NOTE
If you do not select Print entire date range, only the data visible on the screen is
printed:
5. Select OK F1.
The system sends the report to the printer.
Laboratories, Inc.
Laboratory A
123 Lake Street
QC Data Report Townsville, ST 33333
UniCel DxI 800, Access Immunoassay System 101-555-2323 ext. 109
S/N 654321 Jane Smith
Page 1 of 1
Technologist _____________________________________________ Printed 10/10/10 05:32 PM
2082B.bmp
Figure 6-16 QC Data Report Example
C73199-AA
QC Chart Report Townsville, ST 33333
UniCel DxI 800, Access Immunoassay System 101-555-2323 ext. 109
S/N 654321, Version 4.4 Jane Smith
________________________________________________________________________________________________________________________________
Test: PSA-Hyb Start Date: 06/23//10
Control: BRLevel2 End Date: 06/30/10
UniCel DxI Operator's Guide
2083C.bmp
6-41
6.4: Reviewing Quality Control Charts and Data UniCel DxI Operator's Guide
Copying QC Use this procedure to copy QC data to a 3.5-inch high density, DOS-formatted disk or
Data to a Disk a USB flash drive. The system copies the QC data into a comma-separated field
or Flash Drive format that you can import to a PC-based spreadsheet application. The QC data fields
are:
• Control lot headers
• Control lot data
• Data point headers
• Rows of data point values
2. Select Review Chart and Data F2 to display the QC Chart and Data screen.
(Optional) Adjust the date range to include the points you want to save to disk.
For more information, see the Applying a QC Chart and Data Filter topic.
• To copy data to a USB flash drive, insert a flash drive into a port on the front
of the PC. If the front of the PC does not contain a USB port, use a port on the
back of the PC.
NOTE
For UniCel DxI instruments set up to perform system backups using a USB flash
drive, the backup drive might already be in one of the USB ports on the back of
the PC. Use a different flash drive to copy data. Do not disturb the flash drive
used for backups when inserting and removing the flash drive for copying data.
• To copy data to a 3.5-inch disk, insert a disk into the 3.5-inch disk drive of
the PC.
4. Select Copy to Disk F6. A default file name and extension are provided in the
File Name field of the Copy to Disk window.
5. Confirm that the correct drive is selected from the Drive list, or select a different
drive.
NOTE
Unless directed by Technical Support, do not copy data to the C: Console_PC
drive or any drive labeled BACKUPDRIVE.
(Optional) To change the default file name, enter a file name and extension for
the data file in the File Name field.
• If you do not enter a file name or extension, the system supplies a file name
based on the date and time, with a .csv (comma delimited) extension.
• If the file name already exists on the disk or flash drive, the system displays a
warning message. Select Yes F1 to overwrite the existing file, or select
Cancel F8.
(Optional) If the 3.5-inch disk is not formatted, select the Format disk box.
6. Select OK F1.
The copy process begins and a progress window is displayed.
7. If the disk or flash drive becomes full before the copy process is complete, the
system displays a status message.
• For a USB flash drive, clear sufficient space on the drive, or copy the files to
a different flash drive.
• For a 3.5-inch disk:
- Remove the disk and label it with the data file names.
- Insert another disk, and select OK F1.
The system automatically names the file on the new disk the same as the file on
the first disk, and adds a disk number to the end.
Example
If two disks are required to copy a test result file, and you name disk
one QCdata.csv, the system names disk two QCdata(2).csv.
8. When the copy process is complete, the progress window closes. Remove the disk
or flash drive from the PC and label it.
NOTE
If you will be using one disk or flash drive to copy multiple files, you can leave it
in place for the next copy operation. Be sure to keep track of the file names so you
can correctly label the disk or flash drive when you remove it.
You can display charts for up to three QC test runs simultaneously in order to
compare the data. You can also save comparisons for later review. Saved comparisons
are updated with new data as QC tests are run.
2064B.bmp
Chart Name (Field) Displays the test name, quality control name, quality control lot number, and
system ID for data on the screen.
Date Range (Field) Displays the start and end dates for the chart data.
Mean (Field) Displays the calculated mean of the points within the date range. The value
displays with the number of decimal places defined for the assay, plus one additional
decimal place, up to a maximum of four decimal places.
Omitted points are not included in the calculations. For information about defining the
decimal places for an assay, see Section 3.3: Test Setup in the Reference Manual.
SD (Field) Displays the calculated standard deviation (SD) for the points within the date
range. The value displays with the number of decimal places defined for the assay,
plus one additional decimal place, up to a maximum of four decimal places. Omitted
points are not included in the calculations..
%CV (Field) Displays the calculated coefficient of variation (%CV) of the points within the
date range. The value always displays with two decimal places. Omitted points are not
included in the calculations..
Points (Field) Displays the number of data points within the date range used to calculate the
mean, SD, and %CV. Omitted points are displayed on the chart with a slash through
them, but are not included in the calculations.
Levey-Jennings (Chart) Displays a graph of the plotted points from each control. For more information
Charts about charts, see the QC Chart and Data Screen and Levey-Jennings Charts topics in
Section 6.4: Reviewing Quality Control Charts and Data.
Save (Button) Select to save the comparison. The system creates a Comparison button
Comparison F1 which can be selected to view the comparison.
Review Chart (Button) Select to display the QC Chart and Data screen for one of the charts. A menu
and Data F2 of chart numbers is displayed. Charts are numbered sequentially from the top.
Edit Date (Button) Select to change the range of dates for which QC data is displayed. For more
Range F4 information, see the Changing the Date Range topic.
Comparing Use this procedure to view QC charts for two or three quality controls at the same
QC Charts time.
System Mode: Any Mode
1. From the Quality Control screen, locate and select two or three quality controls
for comparison, or select a previously saved comparison to review.
To remove a button from your selection, select the button again.
NOTE
You can select a comparison button only if no quality controls are selected, and
you can select quality controls only if no comparison button is selected.
(Optional) To change the date range of the displayed data, select Edit Date
Range F4 and then enter the date range you want to view. For more information,
see the Changing the Date Range topic.
(Optional) To save the comparison for viewing again, select Save
Comparison F1. For more information about how to save comparisons, see the
Saving QC Comparisons topic.
(Optional) To view one of the charts in more detail, select Review Chart and
Data F2, then select the button (Chart-1 F1, Chart-2 F2, or Chart-3 F3)
corresponding to that chart. Chart-1 is the top chart in the display.
Quality Control
Compare Charts F3
Compare Charts
Save the No
comparison?
Yes
Is correct date Yes
range displayed?
Save Comparison F1
No Save Comparison
2103B.wmf
Figure 6-19 Comparing QC Charts Flowchart
Changing the Use this procedure to change the date range used for QC chart comparisons.
Date Range
System Mode: Any Mode
1. From the Compare Charts screen, select Edit Date Range F4 to display the Edit
Date Range window.
2094A.bmp
2. To set the date range to the past seven days, select the Last Week option.
3. To set the date range to the past 31 days, select the Last Month option.
5. Select OK F1.
NOTE
When viewing charts for date ranges greater than 31 days, the last date displayed
on the chart will be the date of the most recent point in the date range.
Saving QC Use this procedure to save a comparison displayed in the Compare Charts screen.
Comparisons
System Mode: Any Mode
1. From the Compare Charts screen, select Save Comparison F1.
2. Enter one or two lines of text for the Comparison button (notes or information
about the comparison) in the Line 1 and Line 2 fields of the Save Comparison
window.
Deleting QC Use this procedure to delete a saved comparison. The Comparison button is removed
Comparisons from the Quality Control screen.
3. Select OK F1.
A confirmation window is displayed.
As you set up quality controls for use on the instrument, record appropriate product
data and results.
NOTES
• Use fresh quality control materials for best results.
• Do not re-use quality controls by pooling them into sample containers or
storage containers.
Collecting Use this procedure over a 15-day period to gather sufficient data to establish quality
Data to control ranges for a specific UniCel DxI instrument.
Establish QC Required Materials
Ranges
• Access Immunoassay System assay calibrators (2 lots, if available)
• Access Immunoassay System reagent packs (2 lots, if available)
• Quality control materials, prepared according to the manufacturer's
instructions
Procedure
1. Calibrate the assay at the beginning of each 5-day period. Use different
combinations of calibrator and reagent lots, if available. For information about
running calibrations, see Chapter 5: Assay Calibration.
2. Set up the quality controls, using the manufacturer's suggested mean and standard
deviation from the directional insert. For more information, see
Section 6.2: Setting Up Quality Controls.
NOTE
If you are using quality controls with no available suggested mean and standard
deviation values, use the values from a previous lot. If you are establishing ranges
for this quality control for the first time, assign reasonable values. The values
entered here will keep the data points from the first few tests within the visible
range of the chart. These values will be replaced by new values after more data
points are generated.
3. Run one or two replicates of each quality control test one time each day. For more
information, see Section 6.3: Running Quality Control Tests.
• Vary the sequence in which the QC samples are processed.
• Vary the reagent lots used to process the QC samples.
• After the second assay calibration, alternate between active and previous
active calibrations. For more information, see the Switching the Active with
the Previous Active Calibration topic in Section 5.4: Reviewing Calibration
Data.
• Continue processing QC samples for 15 days, until you have obtained a
minimum of 15 values per quality control level on at least three different
calibration curves (45 values minimum).
4. When you have collected the required amount of data, evaluate the data and set
the QC ranges. For more information, see the Evaluating QC Data and
Establishing Ranges topic.
Evaluating QC Use this procedure to evaluate accumulated QC data and establish acceptable ranges
Data and of values for each level of a tri-level quality control set on a specific UniCel DxI
Establishing instrument.
Ranges
NOTES
• This procedure requires a minimum of 15 test results per quality control
level. For more information, see the Collecting Data to Establish QC Ranges
topic.
• Depending on the values first used to establish the mean and the SD of the
quality control, many or all data points may appear to be out of range, and
may violate Westgard rules. This does not indicate a failure. After
performing the following steps, the data points should be evenly distributed
within acceptable range tolerances.
1. From the Quality Control screen, locate and select the quality control lot for
which you are establishing ranges, then select Review Chart and Data F2 to
display the QC Chart and Data screen.
2. Set the date range to the 15-day period during which the QC data was collected.
For more information, see the Applying a QC Chart and Data Filter topic in
Section 6.4: Reviewing Quality Control Charts and Data.
3. Write down the Mean and SD values displayed to the left of the Levey-Jennings
chart.
4. From the QC Chart and Data screen, select Back, then select QC Setup F5. Select
the quality control lot for which you are establishing ranges, then select Edit
Control F2 to display the Edit Control window.
5. Edit the quality control lot setup information, replacing the previous mean and
SD with the mean and SD from step 3. For more information, see the Adding or
Editing a Quality Control Lot topic in Section 6.2: Setting Up Quality Controls.
6. Return to the QC Chart and Data screen. The plotted points will now be evenly
distributed above and below the centerline on the Levey-Jennings chart.
7. Locate data points that are above +3 SD or below -3 SD (outliers). These points
are displayed on the chart as solid triangles on the +3 SD or the -3 SD lines.
NOTE
If no outliers are present, skip to step 10, using the current mean and SD values.
8. Remove outliers from the calculations by omitting any data points that are above
+3 SD or below -3 SD. For more information, see the Omitting a QC Data Point
topic in Section 6.4: Reviewing Quality Control Charts and Data.
As you omit outliers, the Mean and SD fields to the left of the Levey-Jennings chart
change to new values.
10. Return to the Edit QC window and edit the quality control lot setup information
with the mean and SD from the previous steps.
11. If the %CV is less than 6, multiply the SD by 1.5 and enter the result in the SD
field.
13. Periodically recalculate the mean and SD over the life of the quality control lot.
QC results can fail for a variety of reasons including past-due maintenance, cold
substrate, QC material stability, calibrator stability, reagent stability, and instrument
hardware or software problems.
1. Identify event log errors. Contact Technical Support if you need help
troubleshooting event log errors. For more information, see Chapter 5: Event Log
in the Reference Manual.
2. Troubleshoot test result flags. For more information, see Section 4.3: Reviewing
Test Results.
3. Verify that the sample container has sufficient volume, and has been loaded in the
correct rack and the correct sample container position. If not, correct the problem
and repeat the test.
4. Verify that routine maintenance has been performed. If it has not, perform the
required maintenance tasks and then repeat the QC test. For more information,
see Chapter 7: Routine Maintenance.
5. Verify that the substrate was equilibrated to room temperature before being
loaded. If it was not equilibrated, repeat the QC test after the substrate has
equilibrated. For the recommended temperature and equilibration time, see the
substrate instructions for use.
NOTE
Statistically, even with an appropriate mean and 2SD range, 1 of 20 QC results
will be out of range, and 1 of 333 QC results will be out of the 3SD range.
Quality control material stability 1. Prepare new quality control according to the
procedure provided by the manufacturer.
2. Load the freshly prepared quality control and repeat
the test.
Wrong quality control loaded onto 1. Load the correct quality control.
the sample rack 2. Repeat the test.
Wrong quality control lot number Repeat the test using the correct lot number.
selected for the test
Wrong mean and/or standard 1. Review the QC information using Edit Control F2
deviation information entered when from the QC Setup screen.
setting up the quality controls 2. If necessary, edit the information (see Section 6.2).
3. Repeat the test.
WARNING
System Check Solution contains ProClin™ 300 preservative, which may cause
sensitization by skin contact. After contact with skin, wash immediately with
soap and water. Wear suitable gloves.
8. Run the System Check routines. If results are out of range, contact Technical
Support. For more information, see Section 4.4: System Check Routines in the
Reference Manual.
9. Identify any reagent pipettor that is producing failed QC results for all assays that
run on that pipettor.
a. If QC failures are coming from one pipettor, clear the assignment of that
pipettor in the test setup for each failing assay. For more information, see the
Assigning Pipettors to Tests topic in Section 3.3: Test Setup of the Reference
Manual.
b. If the pipettor that became unassigned in step a was the only pipettor assigned
to a test, assign a different pipettor to that test.
c. Repeat the QC test by entering a new QC test request.
NOTE
Because the reagent pipettor that was used for the failed QC test is no longer
assigned to that test, you cannot rerun the test by using the Rerunning a QC Test
procedure.
d. If QC results from other pipettors are satisfactory, disable the failing pipettor
on the Supplies Setup screen until repairs can be made. For more
information, see the Enabling and Disabling Reagent Pipettors topic in
Section 2.3: Supplies Setup of the Reference Manual.
10. Recalibrate the assay to compensate for subtle changes in the instrument or
reagent. For more information, see Chapter 5: Assay Calibration.
11. If none of the preceding steps resolves the problem, contact Technical Support.
Rerunning a If a quality control test produces questionable results, use this procedure to rerun the
QC Test test. The system uses the same reagent pipettor and reagent pack to eliminate random
errors as a cause of the questionable result.
NOTE
Only select a single result. If you select a group of results and then select
Rerun F3, only the result indicated with the Arrow button is rerun.
2. Select Rerun Test F3. A message informs you that the test has been reordered,
and the test is added to the work pending list.
NOTE
After the test has been added to the work pending list, the test will be applied
automatically to the next request for that QC.
3. Go to the Sample Manager screen and select New Request F3, then select
Patient/QC Requests F1 to display the Test Requests screen.
6. Select the quality control to rerun. The test is automatically removed from the
work pending list, and is added to the test list for the quality control.
8. Place the QC sample in the rack entered in step 4, and load the rack on the
instrument. For information on placing sample containers in racks and loading
racks on the UniCel DxI instrument, see Chapter 3: Sample Manager.
NOTE
If the reagent pipettor or the reagent pack are unavailable, the sample status will
be changed to Supply Wait and the test will not rerun until both the reagent
pipettor and the reagent pack are available. When the supply condition has been
cleared, the test will rerun automatically.
7 Routine Maintenance
7.1 Maintenance Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3
Inspecting the Reagent Pipettors and Cleaning the Wash Towers. . . . . . . . . 7-53
Routine maintenance for the UniCel DxI system includes the following maintenance
procedures:
• Daily maintenance
• Special weekly maintenance
• Maintenance after a 5,000 test interval
• Maintenance after a 10,000 test interval
• The Utility routine
Daily maintenance consists of procedures for preparing the system to process samples
each day. When the 10,000 and 5,000 test interval procedures are scheduled on the
same day, always end by performing daily maintenance. For more information, see
Section 7.2: Daily Maintenance.
Special weekly maintenance is only for laboratories that run tests for HIV antibody
levels. Special weekly maintenance consists of a procedure for installing clean
aspirate probes. For more information, see Section 7.3: Special Weekly Maintenance.
Maintenance after a 5,000 test interval consists of a procedure for replacing the duck
bill valve. For more information, see Section 7.4: Test Interval Maintenance: 5,000
Tests.
Maintenance after a 10,000 test interval consists of cleaning and closely inspecting
internal components and tubing connections. The 10,000 test interval procedures
include a step to perform 5,000 test interval maintenance before the 10,000 interval
procedures are complete. For more information, see Section 7.5: Test Interval
Maintenance: 10,000 Tests.
The Utility routine primes the reagent pipettors, aspirate probes, and dispense probes.
The system automatically runs the routine every four hours if the system is not
processing samples. You can also run the Utility routine manually. The Utility routine
should always be enabled. If the Utility routine is disabled, you should enable the
routine and prime fluidics before running a clean routine as part of your daily
maintenance procedures. For more information about enabling the Utility routine, see
the Enabling and Disabling the Utility Routine procedure.
Begin 10,000 test maintenance Begin the remaining 10,000 1 - Remove and inspect the dispense
at the Diagnostics menu test interval maintenance probes
procedures ** 2 - Inspect pump connections at the
wash buffer reservoir
Device Diagnostics F4 3 - Inspect the upper aspirate peristaltic
pump connections
Close the right 4 - Inspect the reagent pipettors and
Device Diagnostics main upper cover clean the wash towers
5 - Inspect for fallen RVs
Install clean*
aspirate probes **
1 - Back up the system
(weekly if testing for HIV
levels or at 10,000 test Perform daily maintenance ** 2 - Record the test count
interval maintenance) 3 - Shake down the solid waste
container
4 - Run a clean routine
Replace the duck bill valve **
Run QC and evaluate any
(5,000 test interval
out-of-range results to verify
maintenance)
system performance **
* Clean the used aspirate probes after ** Record that the procedure is
you complete all other maintenance complete on the Maintenance Log
procedures
2270G.wmf
Figure 7-1 Routine Maintenance Overview Flowchart
2170D.bmp
Zone (Fields) Displays the current temperature reading in degrees Celsius for nine
Temperature (°C) thermal zones.
Temperature readings are updated in real-time. If a temperature reading is outside
of the range listed in the Limits (°C) field, the Temperature (°C) field for the
corresponding thermal zone turns red.
Limits (°C) (Fields) Displays the expected temperature range in degrees Celsius for each
thermal zone. If a temperature reading is outside of the range listed in the
Limits (°C) field, the Temperature (°C) field for the corresponding thermal zone
turns red.
Pipettor (Button) Select to display the Pipettor Matching menu with the following options:
Matching F1 Reag. Pipettor Matching F1
Use this function only as directed by Technical Support.
Low Volume Matching F2
Use this function only as directed by Technical Support.
System Checks F2 (Button) Select to display the System Checks screen. For more information, see
Section 4.4: System Check Routines in the Reference Manual.
Carryover F3 (Button) Select to display the DxI Carryover screen. From the DxI Carryover
screen you can review and print the results of the DxI carryover test. For more
information, see Section 4.7: Carryover Testing in the Reference Manual.
Decontaminate (Button) Select to decontaminate the substrate system and any components that
Substrate F4 come in direct contact with the substrate. Use this function only as directed by
Technical Support. For more information, see Section 6.4: Fluidics Support
Procedures in the Reference Manual.
Run Utility (Button) Select to run the Utility routine manually. Run this routine only if
Routine F5 requested to do so by Technical Support, or if it is a required step in a
troubleshooting procedure. For more information, see the Requesting the Utility
Routine Manually procedure.
Disable (Button) Select to disable or enable the automatic Utility routine. If enabled, the
or Utility routine runs when the instrument has been idle for four hours. The button
Enable name switches between Disable Utility Routine and Enable Utility Routine each
Utility Routine F6 time you select it. For more information, see the Enabling and Disabling the Utility
Routine procedure.
Print F7 (Button) Select to print the Temperature Report. For more information, see the
Printing the Temperature Report procedure.
Requesting Use this procedure to manually request the Utility routine. Request the Utility routine
the Utility when instructed to do so by Technical Support, or when additional cleaning is
Routine required for troubleshooting. The Event Log prompts you to manually request a
Utility routine whenever processing is interrupted. The Utility routine takes
Manually
approximately three minutes to complete.
NOTE
Do not load samples while the Utility routine is running.
Enabling and Enabling the Utility routine sets up the instrument to run the routine every four hours
Disabling the when it is not processing samples. The system records an event in the Event Log each
Utility Routine time you enable or disable the Utility routine.
Use this procedure to enable or disable the Utility routine on the Maintenance Review
screen.
2. If you are enabling the Utility routine, and the routine has been disabled for more
than four hours, prime all of the fluidic components before continuing with
normal operation. For more information, see Section 4.3: Prime Fluidics in the
Reference Manual.
Printing the The Temperature report is a printout of the thermal zone temperatures displayed on
Temperature the Maintenance Review screen. To print a copy of the report, select Print F7 on the
Report Maintenance Review screen.
Requesting You can request and load maintenance routines while the system is processing
Maintenance samples. Requested routines run after calibrator, QC, and patient sample processing is
Routines complete.
WARNING
You will come in contact with potentially infectious materials during this
procedure. Handle and dispose of biohazard materials according to proper
laboratory procedures. Proper hand, eye, and facial protection is required.
2176B.bmp
NOTES
• If you request a clean routine when one or more reagent pipettors are
disabled, the disabled pipettors will not be cleaned.
• Do not select the Run Matching, Verify Matching, or Low Volume
Matching routines unless you are directed to do so by a technical support
representative.
• Do not run patient, QC, or calibrator samples while System Check routines
are running. The Unwashed routine may fail. Wait until the System Check
routines are complete before loading patient, QC, or calibrator samples.
3. Select OK F1.
4. On the Maintenance Requests screen, enter the rack ID in the Enter ID field and
press [Enter].
5. Prepare a rack as instructed on the Maintenance Requests screen, and place the
sample containers in the rack. For more information, see the Placing Sample
Containers in Racks procedure in Section 3.4: Loading and Unloading Racks.
6. Load the rack. For more information, see the Loading a Rack for Routine
Processing procedure in Section 3.4: Loading and Unloading Racks.
NOTE
If you load a rack with a System Check routine test request at the same time as a
rack with a clean routine request, the system runs the clean routine before the
System Check routine.
2177B.bmp
Rack Button (Button with field) Displays the rack ID, and the required position, contents, and
and volume of each container in the rack.
Enter ID Select the Rack button to enter or change the current rack ID. The Enter ID field is
displayed.
Contents (Column) Displays the name of the requested maintenance routine, and the
required contents and volume of each sample container in the rack.
Position (Indicator) Indicates the position of each sample container in the rack. The
positions in a rack are arranged numerically from 1-4, as displayed on the Rack
button.
Rack ID F1 (Button) Select to enter a new rack ID, or change the current rack ID. The Enter ID
field is displayed on the Rack button. For more information, see the Changing a
Rack ID procedure in Section 3.6: Editing Test Requests.
New Request F3 (Button) Select to exit this request and to make a new request. There are three
different request types.
Patient/QC Requests F1
Select to request tests for patient samples and quality controls. For more
information, see the Manual Test Requests procedure in Section 3.3: Patient Test
Requests, or see the Entering QC Test Requests procedure in Section 6.3: Running
Quality Control Tests.
Calibration F2
Select to request an assay calibration. For more information, see the Entering
Calibration Requests procedure in Section 5.3: Running a Calibration.
Maintenance F3
Select to request a maintenance routine. For more information, see the Requesting
Maintenance Routines procedure.
Cancel Request F6 (Button) Select to cancel the requested maintenance routine and return to the
Sample Manager screen. For more information, see Section 3.6: Editing Test
Requests.
Beckman Coulter recommends that you perform daily maintenance once every
24 hours. Perform daily maintenance prior to beginning your daily workload. If your
laboratory runs the instrument continuously, perform maintenance at the end of the
longest period of inactivity. Even if the system is not used routinely, perform daily
maintenance each day.
Perform daily maintenance to prepare the instrument to process samples for the day.
• Back up the system
• Record the test count
• Shake down the solid waste container
• Run a clean routine
Perform one of two clean routines each day to clean the sample, reagent, and aspirate
probes. Run the Special Clean routine daily if your laboratory has run the Vitamin B12
assay during the last 24-hour time period. Otherwise, run the Daily Clean System
routine.
NOTE
You can load patient sample racks while a clean routine is in process.
Maintenance Requests
Record the instrument test
count on the Maintenance
Log
Enter the rack ID and press
[Enter]
Time to Completion
The system begins the clean
Daily Clean System: Approximately 10 minutes
routine *
Special Clean: Approximately 13 minutes
New Request F3
2105F.svg
Figure 7-5 Daily Maintenance Flowchart
Backing Up Use this procedure to back up the system each day to save your system data files.
the System
System Mode: Ready Not Ready
1. Perform an immediate system backup, or confirm that a scheduled system backup
successfully ran today. Recent backup information is displayed on the PC Admin
screen. For more information, see Section 2.6: Data Management in the
Reference Manual.
2. Record that the daily backup is complete on the Maintenance Log. Then continue
with the Recording the Test Count procedure.
Recording the The 10,000 and 5,000 test interval maintenance procedures are scheduled after the
Test Count UniCel DxI system runs a specific number of tests.
Use this procedure to record the test count. Schedule the test interval maintenance
procedures based on the daily instrument test count.
3. Find the test count listed in the Run Hours : Test Count field.
NOTE
The test count is the second number listed in this field.
5. Compare the test count to the test count value recorded when the 5,000 and
10,000 test interval procedures were last performed.
(Optional) Schedule maintenance if maintenance is due.
6. Select OK F1 to exit the Details window. Then continue with the Shaking Down
the Solid Waste Container procedure.
Shaking As the UniCel DxI system processes samples, waste accumulates in the solid waste
Down the container in a conical pattern. If the accumulated waste grows too high, it may
Solid Waste obstruct the waste chute and prevent the system from dispensing solid waste into the
container. To prevent this from occurring, use this procedure to shake down the
Container
container each day.
WARNING
You will come in contact with potentially infectious materials during this
procedure. Handle and dispose of biohazard materials according to proper
laboratory procedures. Proper hand, eye, and facial protection is required.
CAUTIONS
• Do not press the green reset button after shaking down the solid waste
container. Pressing the green button resets the solid waste indicators. An
inaccurate solid waste count could cause the solid waste to overflow the
container or obstruct the waste chute.
• Do not open the solid waste door while the system is initializing, running
a Daily Clean System routine, or running a Special Clean routine.
Opening the solid waste door will cancel the routine and the system will
enter the Not Ready mode. For more information, see the Initializing the
System procedure in Section 4.2: Initialize System of the Reference
Manual, or see the Running Clean Routines in Section 7.2: Daily
Maintenance.
System Mode: Any Mode
1. Open the solid waste door by pulling down on the handle (see Figure 2-30 in
Section 2.6: Solid Waste).
2. Reach into the storage area and grasp the solid waste container. Shake gently until
the accumulated waste is distributed evenly in the container.
4. Record that the waste container procedure is complete on the Maintenance Log.
Then continue with the appropriate clean routine. For more information, see the
Running Clean Routines procedure.
Running Clean routines keep the sample, reagent, and aspirate probes clean and working
Clean properly. Run one of two clean routines each day as a part of daily maintenance.
Routines Use this procedure to run a clean routine. Once you prepare and load the rack, it takes
approximately ten minutes to complete the Daily Clean System routine, and
approximately thirteen minutes to complete the Special Clean routine.
CAUTION
Do not open the solid waste door while either clean routine is in progress.
Opening the solid waste door will cancel the routine and the system will enter
the Not Ready mode. For more information, see Shaking Down the Solid
Waste Container, or see Changing a Full Solid Waste Container in Section
2.6: Solid Waste.
NOTE
You can load patient sample racks while a clean routine is in process. Samples are
aspirated after the clean routine rack is aspirated, and are held in the sample
wheel until the clean routine is complete.
Required Materials
• Sample containers that hold at least 4 mL of solution
• Sample rack with the appropriate rack ID for the container used
• UniCel DxI wash buffer
• Contrad 70 cleaning solution
• 1:5 dilution of Citranox cleaning solution - mix 1 part Citranox and 4 parts
deionized water
• 70% ethanol solution — mix 7 parts of 95% ethanol and 3 parts wash buffer
(alternatively, you can substitute methanol for the 95% ethanol)(Special
Clean routine only)
• Maintenance Log
WARNINGS
• You will come in contact with potentially infectious materials during this
procedure. Handle and dispose of biohazard materials according to proper
laboratory procedures. Proper hand, eye, and facial protection is required.
• Citranox cleaning solution is acidic and may cause eye or skin irritation.
Handle according to proper laboratory procedures. See the manufacturer's
label for details.
• Contrad 70 cleaning solution is alkaline and may cause severe eye irritation
or mild skin irritation. Handle according to proper laboratory procedures.
See the manufacturer's label for details.
• Ethanol and methanol are extremely flammable. Do not use near heat or
flame. Do not ingest. Avoid contact with eyes, skin, and clothing. Use with
adequate ventilation.
CAUTION
Racks are configured to accept only one type of sample container. The
sample containers used must match the ID configured for the rack. Placing
an incorrect sample container in a rack may damage the system. You can
find the rack ID ranges set up for each type of sample container on the
System Setup screen. For more information, see Section 2.2: System Setup in
the Reference Manual.
Procedure Steps
System Mode: Ready
1. Confirm that the Utility routine is enabled. For more information, see the
Enabling and Disabling the Utility Routine procedure in Section 7.1:
Maintenance Overview. If the routine is not enabled, enable the routine and prime
the fluidics before you continue. For more information, see the Priming the
Fluidics procedure in Section 4.3: Prime Fluidics of the Reference Manual.
2. From the Sample Manager screen, select New Request F3, and then select
Maintenance F3 to display the Request Maintenance window.
3. Select either the Daily Clean System or the Special Clean option.
NOTE
Run the Special Clean routine if your laboratory has run the Vitamin B12 assay
during the last 24-hour time period. Otherwise, run the Daily Clean System
routine.
7. If you are running the Special Clean routine, also place an empty sample
container in position 3 of the rack.
WARNING
In the following steps, you will pipette solutions that may cause eye or skin
irritation. Handle according to proper laboratory procedures.
WARNING
Ethanol and methanol are extremely flammable. Do not fill sample tubes on the
instrument or in a rack positioned in the sample presentation unit. Remove
sample tubes from the instrument immediately after completing the Special
Clean routine.
10. If you are running the Special Clean routine, pipette 4 mL of a 70% ethanol or
methanol solution into the sample container in position 3.
12. Load the rack. For more information, see the Loading a Rack for Routine
Processing procedure in Section 3.4: Loading and Unloading Racks.
13. Record that the procedure is complete on the Maintenance Log. For more
information, see Section 7.6: Maintenance Log.
If your laboratory runs tests for HIV antibody levels, Beckman Coulter recommends
that you perform special weekly maintenance. Special weekly maintenance consists
of a procedure for installing clean aspirate probes. For more information, see the
Installing Clean Aspirate Probes procedure in Section 7.5: Test Interval Maintenance:
10,000 Tests.
Laboratories that install clean aspirate probes during special weekly maintenance are
not required to install clean probes during 10,000 test interval maintenance.
After you complete the special weekly maintenance procedure, perform daily
maintenance and verify system performance. For more information,
see Section 7.2: Daily Maintenance and see the Verifying System Performance
procedure in Section 7.4: Test Interval Maintenance: 5,000 Tests.
WARNINGS
• Never wear exposed metal that can come into contact with electrical
components of the instrument.
• You will come in contact with potentially infectious materials while
performing special weekly maintenance. Handle and dispose of biohazard
materials according to proper laboratory procedures. Proper hand, eye, and
facial protection is required.
Device Diagnostics
2301D.wmf
Figure 7-6 Special Weekly Maintenance Overview Flowchart
Beckman Coulter recommends that you replace the duck bill valve after each
5,000 test interval. The duck bill valve prevents wash buffer from entering the RVs in
the wash carousel during the aspirate probe wash.
If the 10,000 and 5,000 test interval maintenance procedures are scheduled for the
same day, begin with the 10,000 test interval maintenance procedures. The 10,000 test
interval procedures include a step to perform 5,000 test interval maintenance. Perform
the daily maintenance procedures after you complete the 10,000 and 5,000 test
interval procedures.
Device Diagnostics
2300D.wmf
Figure 7-7 Test Interval Maintenance (5,000 Tests) Overview Flowchart
Replacing the Use this procedure to replace the duck bill valve. Replace the duck bill valve 5,000
Duck Bill tests after the valve was last replaced, or as instructed by the system documentation or
Valve by a technical support representative.
WARNINGS
• Never wear exposed metal that can come into contact with electrical
components of the instrument.
• You will come in contact with potentially infectious materials during this
procedure. Handle and dispose of biohazard materials according to proper
laboratory procedures. Proper hand, eye, and facial protection is required.
CAUTIONS
• Opening the covers abruptly shuts off power to the upper cabinet, and
may cause damage if the system is not in the proper mode and the user
interface is not displaying the correct screen. Follow this procedure
carefully.
• Before you begin this procedure, confirm that you have a replacement
valve in your CARE kit. If there is no replacement valve, do not perform
this procedure. Order a new set of valves and replace the duck bill valve
at your earliest opportunity.
NOTE
If you are continuing to this procedure from 10,000 test interval maintenance,
begin at step 8.
1. From the Diagnostics menu, select Device Diagnostics F4 to display the Device
Diagnostics screen.
2. Open the right main upper cover. For more information, see Section 6.2:
Instrument Cover Procedures in the Reference Manual.
3. Locate the dispense probe plate. The dispense probe plate is the lower plate,
which contains the D1-D3 dispense probes (see Figure 7-8).
2187C.eps
CAUTION
In the following step, do not lift up on the dispense probe plate at the D1
dispense probe position. The dispense probe plate can bend when lifted at
this position.
4. Locate the D2 dispense probe. At the D2 dispense probe position, place your
fingers underneath the dispense probe plate and lift up gently.
CAUTION
In the following step, be careful not to brush against the D1 dispense probe
or tubing while you position the sample rack. The probe will not function
5. Carefully slide a sample rack beneath the dispense probe plate. Position the rack
directly in front of the D3 dispense probe.
CAUTION
In the following two steps, handle the A1 aspirate probe and tubing with
extreme care. The probe is fragile and will not function properly if it is bent,
or if the tubing is routed incorrectly, kinked, punctured, or torn.
6. Gently grasp the A1 aspirate probe fitting, which is located just above the aspirate
probe plate. Turn the fitting counterclockwise until the probe is loose.
7. Carefully lift the A1 aspirate probe out of the aspirate probe plate and set the
probe down on several clean, lint-free tissues.
8. Locate the wash collar in the dispense probe plate (see Figure 7-9), which is
located to the right of the D2 dispense probe.
9. Grasp the collar from beneath the dispense probe plate, and turn it clockwise until
the plastic shoulder is aligned with the key slot in the dispense probe plate.
10. Carefully pull the wash collar down through the key slot. The duck bill valve is
on the underside of the collar.
11. Turn the retaining nut counterclockwise to remove the nut from the wash collar.
12. Lift the duck bill valve out of the retaining nut. Discard the valve in your
biohazard waste.
A1
D2
2241C.eps
2303A.eps
2. To attach the nut to the wash collar, turn the retaining nut clockwise until it is
finger-tight.
WARNING
If the retaining nut is not finger-tight, the duck bill valve could fall into the
instrument during sample processing. This may affect your results.
3. Inspect the duck bill valve. The valve should be completely closed, or open no
more than a few thousandths of an inch (no more than the width of the valve on
either side of the opening).
4. If the valve is open too far, remove the duck bill valve and replace with a new
valve. Discard the faulty valve in your biohazard waste.
5. Grip the wash collar from beneath the dispense probe plate and align the plastic
shoulder with the key slot. Then carefully push the wash collar through the
opening.
6. Turn the collar counterclockwise to secure the plastic shoulder on the dispense
probe plate.
7. Grip the A1 aspirate probe by the probe fitting and lift the probe off the lint-free
tissue. Carefully lower the probe through the opening in the aspirate probe plate.
Turn the fitting clockwise until it is finger-tight. Be careful not to overtighten the
fitting.
8. Verify the A1 aspirate probe deflection. Grip the aspirate probe above the probe
fitting and lift the probe approximately 1/4 inch (6.35 mm). Then release the
probe. The probe should snap back into position.
• If the probe does not snap back, the wash collar may not be positioned correctly
in the key slot. Check the wash collar position and correct if necessary (see
Figure 7-9). Verify the probe deflection again.
• If the wash collar is not the problem, the aspirate probe could be bent. It cannot be
straightened. Replace the aspirate probe with another clean probe. Then verify the
probe deflection. For more information, see the Installing Clean Aspirate Probes
procedure in Section 7.5: Test Interval Maintenance: 10,000 Tests.
• If the probe does not snap back after verifying the wash collar position and
replacing the probe, contact Technical Support for assistance.
9. Record that the procedure is complete on the Maintenance Log. For more
information, see Section 7.6: Maintenance Log.
10. If the 10,000 test interval maintenance procedures are scheduled for today,
continue with the Removing and Inspecting the Dispense Probes procedure next.
For more information, see Section 7.5: Test Interval Maintenance: 10,000 Tests.
11. If the 10,000 test interval maintenance procedures are not scheduled, perform the
following steps:
a. Confirm that the dispense probe tubing is routed correctly (see Figure 7-15
and see Figure 7-16). If the tubing is routed correctly, skip to step c.
• The D1 dispense probe tubing should be routed above the aspirate probe plate
and A1 aspirate probe support arm.
• The D2 and D3 dispense probe tubing should be routed behind and not beneath
the A1 and A2 aspirate probe support arms, respectively.
b. If the tubing is not routed correctly, remove the dispense probes and inspect
for kinked, punctured, or torn tubing. For more information, see the
Removing and Inspecting the Dispense Probes procedure.
• If the tubing is damaged, contact Technical Support for assistance.
• If the tubing is not damaged, return the dispense probes to their proper
positions and route the tubing correctly.
c. Remove the sample rack from beneath the dispense probe plate.
d. Close the right main upper cover. For more information, see Section 6.2:
Instrument Cover Procedures in the Reference Manual.
e. Exit the Device Diagnostics screen.
f. Initialize the system. For more information, see Section 4.2: Initialize System
in the Reference Manual.
g. Perform daily maintenance and then verify system performance. For more
information, see Section 7.2: Daily Maintenance and see the Verifying
System Performance procedure.
For 5,000 test maintenance, begin this If continuing from 10,000 test
procedure at the Diagnostics menu maintenance, begin the procedure here
Gently lift the dispense probe Record that the procedure is Perform daily maintenance
Insert a new valve into
plate and slide a sample rack complete on the and verify system
the retaining nut
beneath the plate Maintenance Log performance
Yes
Inspect the duck bill valve
2240K.wmf
Figure 7-11 Test Interval Maintenance (5,000 Tests): Replacing the Duck Bill Valve Flowchart
Verifying Verify system performance after completing maintenance on the duck bill valve or the
System aspirate and dispense probes, and before you process any samples. If you are
Performance performing other maintenance procedures, verify performance after you finish all
scheduled maintenance.
NOTE
If you just completed daily maintenance, you have already completed this step.
Do not run a second clean routine.
2. Run QC for all assays you use to analyze patient samples. Evaluate out-of-range
QC results. For more information, see Chapter 6: Quality Control.
NOTE
If your laboratory runs tests for HIV antibody levels, you will install clean
aspirate probes each week as part of special weekly maintenance instead of
installing clean probes during 10,000 test interval maintenance.
The 10,000 test interval maintenance procedures include a step to perform 5,000 test
interval maintenance before the 10,000 interval procedures are complete. Perform the
daily maintenance procedures after you complete the 10,000 and 5,000 test interval
procedures.
Install clean Open the left Close the left and right
aspirate probes * Perform daily maintenance *
main upper cover main upper covers
(unless your lab installs
clean probes weekly
for HIV testing)
2260G.svg
Figure 7-12 Test Interval Maintenance (10,000 Tests) Overview Flowchart
Installing As the system processes samples, paramagnetic particles can accumulate inside the
Clean aspirate probe or on the probe exterior. A dirty aspirate probe can affect system
Aspirate performance.
Probes Use this procedure to install clean aspirate probes 10,000 tests after clean aspirate
probes were last installed.
NOTE
If your laboratory runs tests for HIV antibody levels, you will install clean
aspirate probes each week as part of special weekly maintenance instead of
installing clean probes during 10,000 test interval maintenance.
Required Materials
• An empty sample rack
• Several clean, lint-free tissues
• Three clean aspirate probes
WARNINGS
• Never wear exposed metal that can come into contact with electrical
components of the instrument.
• You will come in contact with potentially infectious materials during this
procedure. Handle and dispose of biohazard materials according to proper
laboratory procedures. Proper hand, eye, and facial protection is required.
CAUTIONS
• Opening the covers abruptly shuts off power to the upper cabinet, and
may cause damage if the system is not in the proper mode and the user
interface is not displaying the correct screen. Follow this procedure
carefully.
• Handle each aspirate probe and its tubing with extreme care. The probes
are fragile and will not function properly if they are bent, or if the tubing
is routed incorrectly, kinked, punctured, or torn.
• Remove the aspirate probes one at a time. Do not replace all three
probes at once.
• Remove aspirate probes from the aspirate probe plate before
disconnecting the tubing fittings from the manifold. Fluid may leak out
of the probe and onto the analytical module cover if a fitting is
disconnected while an aspirate probe is still in the plate.
• Use extreme care when removing the A3 aspirate probe from the probe
plate. The probe is difficult to reach and could bend easily if it is not
handled carefully.
• Be sure that no tubing is routed between the aspirate and dispense probe
plates. Tubing routed between the probe plates will be pinched and may
tear open during normal operation.
System Mode: Ready Not Ready
1. From the Diagnostics menu, select Device Diagnostics F4 to display the Device
Diagnostics screen.
2. Open the right main upper cover. For more information, see Section 6.2:
Instrument Cover Procedures in the Reference Manual.
3. Locate the dispense probe plate. The dispense probe plate is the lower plate,
which contains the D1-D3 dispense probes (see Figure 7-8).
CAUTION
In the following step, do not lift up on the dispense probe plate at the D1
dispense probe position. The dispense probe plate can bend when lifted at
this position.
4. Locate the D2 dispense probe. At the D2 dispense probe position, place your
fingers underneath the dispense probe plate and lift up gently.
CAUTION
In the following step, be careful not to brush against the D1 dispense probe
or tubing while you position the sample rack. The probe will not function
properly if it is bent, or if the tubing is routed incorrectly, kinked,
punctured, or torn.
5. Carefully slide a sample rack between the dispense probe plate and the incubator
cover. Position the rack directly in front of the D3 dispense probe (see
Figure 7-8).
6. Locate the three aspirate probes on the aspirate probe plate. The probes are
labeled A1, A2, and A3 in sequence from right to left (see Figure 7-8).
CAUTION
For steps 7 to 12, remove only one aspirate probe at a time. Do not remove all
three probes at once.
7. Gently grasp one of the probe fittings and turn it counterclockwise until the probe
is loose (see Figure 7-13).
2264B.eps
1 Probe fitting
8. Lift the aspirate probe out of the aspirate probe plate and set the probe down on
several clean, lint-free tissues.
9. Locate the aspirate and dispense probe manifold (see Figure 7-14) by following
the aspirate probe tubing to the manifold. The aspirate and dispense probe
manifold is an upright plate positioned near the aspirate and dispense probe
plates. The manifold holds the probe tubing in a stationary position with a fitting
connection and routing clamps.
10. Find which routing clamp secures the probe tubing, and then lift up gently on the
tubing to release it from the clamp.
NOTES
• The A1 aspirate probe has two lines of tubing. Both lines are secured by
routing clamps located on the front manifold plate.
• The A2 and A3 aspirate probes each have one line of tubing. Each line of
tubing is secured by a routing clamp located on the rear manifold plate.
A3 A2 A1
D3 D2 A1W
D1
1
A3
D3
A2
A1W
D2
D1
A1
3
2
A1W
D2
D1
A1
2353A.eps
CAUTION
In the following step, be sure that you remove the aspirate probe from the
aspirate probe plate before disconnecting the tubing fitting from the
manifold. Fluid may leak out of the probe and onto the analytical module
cover if a fitting is disconnected while its aspirate probe is still in the plate.
11. Turn the aspirate probe tubing fitting counterclockwise until it unlocks. After
unlocking the fitting, gently pull it toward you to disconnect the aspirate probe
and tubing from the manifold.
NOTES
• The A1 aspirate probe fittings disconnect from positions A1 and A1W.
• The A2 and A3 aspirate probe fittings disconnect from positions A2 and A3,
respectively.
• The line of tubing for the probe that disconnects from position A1W is
identified with a yellow sleeve. The tubing for the probes that disconnect
from positions A1, A2, and A3 is identified with a white sleeve.
12. Install a clean aspirate probe. For more information, see the Installing Clean
Aspirate Probes procedure.
13. Repeat steps 7-12 to remove and install the remaining aspirate probes.
WARNING
You will come in contact with potentially infectious materials during this
procedure. Handle and dispose of biohazard materials according to proper
laboratory procedures. Proper hand, eye, and facial protection is required.
CAUTIONS
• Handle each aspirate probe and its tubing with extreme care. The probes
are fragile and will not function properly if they are bent, or if the tubing
is routed incorrectly, kinked, punctured, or torn.
• Use extreme care when installing the A3 aspirate probe. The opening in
the probe plate is difficult to reach, and the probe could bend easily if it
is not handled carefully.
• Be sure that no tubing is routed between the aspirate and dispense probe
plates. Tubing routed between the probe plates will be pinched and may
tear open during normal operation.
1. Follow steps a and b to connect the tubing fitting to the aspirate and dispense
probe manifold:
a. Locate the white or yellow colored sleeve attached to the aspirate probe
tubing, and check the probe number printed on the sleeve.
NOTES
• The A1 aspirate probe fittings connect to positions A1 and A1W.
• The A2 and A3 aspirate probe fittings connect to positions A2 and A3,
respectively.
• The A1, A2, and A3 tubing is identified with a white sleeve. The A1W tubing
is identified with a yellow sleeve.
b. Insert the aspirate probe tubing fitting into its corresponding position on the
aspirate and dispense probe manifold. Insert the tubing fitting into the
position on the manifold which corresponds to the probe number printed on
the sleeve. To connect the tubing fitting to the probe manifold, turn the fitting
clockwise until it locks in place (see Figure 7-14).
2. Follow steps a through c to secure the aspirate probe tubing to its routing clamp:
a. For the A1, A1W, A2, and A3 probe tubing, center the colored sleeve over
the appropriate routing clamp. Be sure the routing clamp corresponds to the
probe number printed on the sleeve.
NOTES
• Secure the A1 and A1W aspirate probe tubing to the routing clamps located
on the front manifold plate.
• Secure the A2 and A3 aspirate probe tubing to the routing clamps located on
the rear manifold plate.
b. Press down gently on the colored sleeve to secure it to the clamp.
c. Align the lower section of the A1 and A1W tubing over the lower routing
clamps on the front manifold plate, then press down gently on the tubing to
secure it to the clamp. The lower section of the A1 and A1W tubing does not
have a colored sleeve.
NOTE
The front manifold plate contains two clamps. Secure the tubing to both clamps.
CAUTION
In the following steps, do not return the A1 aspirate probe to its position in
the probe plate if you are performing your 5,000 test interval maintenance
procedures next. Let the probe rest on several clean, lint-free tissues until
you are instructed to return the probe during 5,000 test interval
3. Correctly route the aspirate probe tubing above the aspirate probe plate (see
Figure 7-15 and see Figure 7-16). Any tubing routed between the aspirate and
dispense probe plates will be pinched and may tear open during normal operation.
2354A.eps
4. Hold the aspirate probe by the probe fitting, and then lower the probe through the
opening in the aspirate probe plate.
5. Turn the probe fitting clockwise until the fitting is finger-tight. Be careful not to
overtighten the fitting.
6. Be sure you do not route the aspirate tubing beneath the aspirate probe plate. Any
tubing routed between the aspirate and dispense probe plates will be pinched and
may tear open during normal operation (see Figure 7-16).
2315A.eps
1 A1 aspirate probe plate support arm
2 Incorrect: D1 dispense probe tubing routed beneath the A1
aspirate probe plate support arm
3 Incorrect: D2 dispense probe tubing routed beneath the A1
aspirate probe plate support arm
7. Verify the aspirate probe deflection. Grip the aspirate probe above the probe
fitting and lift the probe approximately 1/4 inch (6.35 mm). Then release the
probe. The probe should snap back into position.
• If the probe does not snap back, the probe will not function properly and should
be removed. Replace the aspirate probe with another clean probe. Then verify the
probe deflection.
• If the probe does not snap back after it is replaced, contact Technical Support for
assistance.
8. Remove the next aspirate probe. For more information, see the Removing the
Aspirate Probes procedure.
9. Repeat steps 1 through 8 to install another clean aspirate probe. Continue until
each probe is clean.
10. Record that the procedure is complete on the Maintenance Log. For more
information, see Section 7.6: Maintenance Log.
11. If you are only installing clean aspirate probes, or are installing clean aspirate
probes as part of special weekly maintenance, end the procedure with the steps
listed below. Otherwise, continue with step 12 of this procedure.
a. Confirm that the dispense probe tubing is routed correctly (see Figure 7-15
and see Figure 7-16).
• The aspirate and dispense probe tubing should not be routed beneath the
aspirate probe plate or aspirate probe support arms.
• Place two fingers underneath the aspirate probe plate, then gently move the
plate up and down to be sure there is no interference from incorrectly
routed aspirate or dispense probe tubing.
If the tubing is routed correctly, skip to step c.
b. If the tubing is not routed correctly, remove the dispense probes from the
dispense probe plate and inspect for kinked, punctured, or torn tubing. For
more information, see the Removing and Inspecting the Dispense Probes
procedure.
• If the tubing is damaged, contact Technical Support for assistance.
• If the tubing is not damaged, return the dispense probes to their proper
positions and route the tubing correctly.
c. Remove the sample rack from beneath the dispense probe plate.
d. Close the right main upper cover. For more information, see Section 6.2:
Instrument Cover Procedures in the Reference Manual.
e. Exit the Device Diagnostics screen.
f. Initialize the system. For more information, see Section 4.2: Initialize System
in the Reference Manual.
g. If you are not continuing with either the 5,000 or 10,000 test interval
maintenance procedures, perform daily maintenance and then verify system
performance. For more information, see Section 7.2: Daily Maintenance and
see the Verifying System Performance procedure.
12. If you scheduled the 5,000 test interval maintenance procedures for today, replace
the duck bill valve now. Begin at step 8 of the Removing the Used Duck Bill
Valve procedure. For more information, see Section 7.4: Test Interval
Maintenance: 5,000 Tests.
13. If the 5,000 test interval maintenance procedures are not scheduled, continue with
the 10,000 test interval procedures. Perform the Removing and Inspecting the
Dispense Probes procedure next.
14. Clean the removed aspirate probes after you complete all other maintenance
procedures. For more information, see the Cleaning the Aspirate Probes
procedure.
Yes
Are you
Are you performing Is 5,000 test Continue with
only installing No special weekly No No Yes Remove and inspect the
maintenance 10,000 test
clean aspirate maintenance? dispense probes
scheduled? maintenance?
probes?
Yes Yes No
Exit the Device Diagnostics
screen and initialize
the system
For more Confirm that the dispense Remove the sample rack
Close the right
information, see probe tubing is routed from beneath the dispense
main upper cover
Figure 7-15 correctly probe plate
2266K.wmf
Figure 7-17 Test Interval Maintenance (10,000 Tests): Installing Clean Aspirate Probes Flowchart
Cleaning the Use this procedure to clean the aspirate probes. Clean the aspirate probes you
Aspirate removed from the instrument at your earliest opportunity. Store the clean probes in
Probes your CARE kit for the next time you install clean aspirate probes.
CAUTION
Handle the aspirate probes and tubing with extreme care. The probes are
fragile and will not function properly if they are bent, or if the tubing is
routed incorrectly, kinked, punctured, or torn.
Required Materials
• Proper hand, eye, and facial protection
• Small beakers (2)
• Contrad 70 cleaning solution
• Disposable aspirate probe brush
• Deionized water
• Syringe and syringe fitting assembly
• Alcohol wipe
WARNINGS
• You will come in contact with potentially infectious materials during this
procedure. Handle and dispose of biohazard materials according to proper
laboratory procedures. Proper hand, eye, and facial protection is required.
• The disposable aspirate probe brush is considered a biohazard after it has
been used to clean from 1 to 3 aspirate probes. Discard the brush with your
biohazard waste. Do not save a used brush for future use.
1. Fill a small beaker with approximately 20 mL of Contrad 70 cleaning solution.
2. Dip the aspirate probe brush in the Contrad 70 cleaning solution.
WARNING
In the following two steps, inserting and removing the aspirate probe brush can
disperse biohazard materials that are lodged inside the probe. Work carefully to
prevent scattering small droplets of these materials into the air.
3. Insert the aspirate probe brush into the tip of the aspirate probe and gently push
the brush inside the probe until you feel it come to a stop.
4. Remove the brush.
5. Repeat steps 2-4 several times on the same probe, or until no orange colored
residue is visible on the brush when it is removed from the probe.
NOTE
For the A1 aspirate probe, be sure both tubing fittings are submerged.
3. Attach the syringe fitting assembly to the syringe (see Figure 7-18).
0400e.eps
1 Syringe
2 Syringe fitting assembly
4. Push the aspirate probe tip into the open end of the tubing on the syringe fitting
assembly.
CAUTION
In the following step, only draw deionized water through the probe tubing
fitting. Do not push water from the syringe through the aspirate probe,
because this could damage the probe and tubing fitting.
5. Draw deionized water through the probe tubing fitting and into the syringe.
6. Remove the syringe from the syringe fitting assembly and push the deionized
water out of the syringe and into a sink.
7. Reattach the syringe to the syringe fitting assembly and repeat steps 5-6 two more
times.
8. Remove the aspirate probe fitting from the beaker of deionized water.
11. Remove the syringe from the fitting assembly, and then gently remove the
assembly from the aspirate probe tip.
12. Position the probe upright on absorbent paper for about 10 minutes to drain any
residual fluid.
2267C.wmf
Figure 7-19 Cleaning an Aspirate Probe Flowchart
Removing Use this procedure to remove and inspect all three dispense probes 10,000 tests after
and the probes were last inspected.
Inspecting the As the system processes samples, crystalline deposits can accumulate on the dispense
Dispense probe exterior and affect system performance.
Probes
WARNINGS
• Never wear exposed metal that can come into contact with electrical
components of the instrument.
• You will come in contact with potentially infectious materials during this
procedure. Handle and dispose of biohazard materials according to proper
laboratory procedures. Proper hand, eye, and facial protection is required.
CAUTIONS
• Handle each dispense probe and its tubing with extreme care. The
probes are fragile and will not function properly if they are bent, or if
the tubing is routed incorrectly, kinked, punctured, or torn.
• Do not disconnect the dispense probe tubing fitting from the aspirate
and dispense probe manifold. All wash buffer in the line will leak out
into the instrument if the fitting is disconnected.
• Use extreme care when handling the D3 dispense probe from the
dispense probe plate. The probe is difficult to reach and could bend
easily if it is not handled carefully.
• Remove and inspect the dispense probes one at a time. Do not try to
remove and inspect all three probes at once.
• Be sure that no tubing is routed between the aspirate and dispense probe
plates. Tubing routed between the probe plates will be pinched and may
tear open during normal operation.
3. Locate the three dispense probes. The probes are labeled D1, D2, and D3 in
sequence from right to left (see Figure 7-8).
CAUTION
In the following steps, do not disconnect the dispense probe tubing fitting
from the aspirate and dispense probe manifold. All wash buffer in the line
will leak out into the instrument if the fitting is disconnected.
4. Visually inspect the dispense probes for kinked, punctured, or torn tubing, and for
crystalline deposits or corrosion at the tubing connections and fittings.
Kinked, punctured, or torn tubing can affect system performance. Deposits or
corrosion can indicate that a connection or fitting is leaking. Contact Technical
Support for assistance if you find damaged tubing, or if you find deposits or corrosion
on the probe fittings.
5. Gently grasp the probe fitting located on the dispense probe plate and turn it
counterclockwise until the probe is loose (see Figure 7-13).
6. Lift the dispense probe out of the dispense probe plate. Do not disconnect the
dispense probe tubing from the routing clamps located on the aspirate and
dispense probe manifold (see Figure 7-14).
If the dispense probe tubing is loose, secure the tubing to the appropriate routing
clamp by centering the colored sleeve over the routing clamp. Be sure the probe
number printed on the yellow sleeve corresponds to the routing clamp. Then, press
down gently on the tubing to hold it in place.
7. Inspect the dispense probe for crystalline deposits on the probe exterior.
8. If there are deposits on the probe, moisten a lint-free tissue with deionized water
and wipe down the probe exterior.
9. Correctly route the dispense probe tubing above or behind the aspirate probe plate
support arms (see Figure 7-15 and see Figure 7-16).
NOTES
• Route the D1 dispense probe tubing above the aspirate probe plate and A1
aspirate probe support arm.
• Route the D2 and D3 dispense probe tubing behind and not beneath the A1
and A2 aspirate probe support arms, respectively.
• Each dispense probe is fitted with a strain relief, which helps route the
dispense probe tubing toward the aspirate and dispense probe manifold. Be
sure the strain relief is not routed beneath the aspirate probe plate (see
Figure 7-15). The strain relief will be pinched and may tear open during
normal operation if it is routed between the aspirate and dispense probe
plates.
10. Hold the dispense probe by the probe fitting, and then lower the probe through the
opening in the dispense probe plate.
11. Turn the probe fitting clockwise until the fitting is finger-tight.
12. Repeat steps 5 through 11 to individually remove and inspect each remaining
dispense probe.
13. Record that the procedure is complete on the Maintenance Log. For more
information, see Section 7.6: Maintenance Log.
14. Confirm that the dispense probe tubing is routed correctly (see Figure 7-15 and
see Figure 7-16).
• The aspirate and dispense probe tubing should not be routed beneath the aspirate
probe plate or aspirate probe support arms.
• Place two fingers underneath the aspirate probe plate, then gently move the plate
up and down to be sure there is no interference from incorrectly routed aspirate or
dispense probe tubing.
If the tubing is routed correctly, skip to step 16.
15. If the tubing is not routed correctly, remove the dispense probes from the
dispense probe plate and inspect for kinked, punctured, or torn tubing.
• If the tubing is damaged, contact Technical Support for assistance.
• If the tubing is not damaged, return the dispense probes to their proper positions
and route the tubing correctly.
16. Remove the sample rack from beneath the dispense probe plate.
17. If you are continuing with 10,000 test interval maintenance, perform the
Inspecting the Pump Connections at the Wash Buffer Reservoir procedure next.
18. If you are not continuing with 10,000 test interval maintenance, perform the
following steps:
a. Close the right main upper cover. For more information, see Section 6.2:
Instrument Cover Procedures in the Reference Manual.
b. Exit the Device Diagnostics screen.
c. Initialize the system. For more information, see Section 4.2: Initialize System
in the Reference Manual.
d. Verify system performance. For more information, see the Verifying System
Performance procedure in Section 7.4: Test Interval Maintenance: 5,000
Tests.
Device Diagnostics
Is the
Moisten a lint-free tissue with
probe free of No
deionized water and wipe
crystalline
down the probe exterior
deposits?
Yes
Yes
Continue with
10,000 test
No Close the right
main upper cover
Record that the procedure is maintenance?
complete on the
Maintenance Log
Yes
Exit the Device Diagnostics
screen and Initialize
Confirm that the dispense the system
Inspect pump connections at
probe tubing is routed
the wash buffer reservoir
correctly
2268J.svg
Figure 7-20 Test Interval Maintenance (10,000 Tests): Removing and Inspecting the Dispense
Probes Flowchart
Inspecting the Use this procedure to visually inspect the pump connections at the wash buffer
Pump reservoir. The wash buffer reservoir is located beneath the left main upper cover.
Connections
WARNINGS
at the Wash
• Never wear exposed metal that can come into contact with electrical
Buffer
components of the instrument.
Reservoir
• You will come in contact with potentially infectious materials during this
procedure. Handle and dispose of biohazard materials according to proper
laboratory procedures. Proper hand, eye, and facial protection is required.
2
1
4
2124A.eps
Figure 7-21 Inspecting the UniCel DxI 800 Wash Buffer Reservoir Pump
Connections
1 2
4
2124B.eps
Figure 7-22 Inspecting the UniCel DxI 600 Wash Buffer Reservoir Pump
Connections
3. If you find any kinked tubing, deposits, or corrosion contact Technical Support
for assistance.
Kinked tubing can affect precision. Deposits or corrosion can indicate that a
connection or fitting is leaking.
4. If you are performing 10,000 test interval maintenance, record that the inspection
is complete on the Maintenance Log. Perform the Inspecting the Upper Aspirate
Peristaltic Pump Connections procedure next.
Inspecting the Use this procedure to visually inspect the upper aspirate peristaltic pump connections
Upper and tubing. The upper aspirate peristaltic pump is located beneath the right main
Aspirate upper cover and to the right of the wash buffer reservoir.
Peristaltic WARNINGS
Pump
• Never wear exposed metal that can come into contact with electrical
Connections components of the instrument.
• You will come in contact with potentially infectious materials during this
procedure. Handle and dispose of biohazard materials according to proper
laboratory procedures. Proper hand, eye, and facial protection is required.
2. Visually inspect the following locations (see Figure 7-23) for kinked tubing, and
for crystalline deposits or corrosion at the tubing connections and fittings.
• Pump channels (three on either side of the pump)
• Pump manifold
NOTE
UniCel DxI systems are configured with one of the perstaltic pump
configurations shown in Figure 7-23. This procedure is appropriate for use with
either configuration.
1 1
2
2
3
3
2261B.eps 3009A.eps
3. If you find any kinked tubing, deposits, or corrosion, contact Technical Support
for assistance.
Kinked tubing can affect system performance. Deposits or corrosion can indicate that
a connection or fitting is leaking.
4. If you are performing 10,000 test interval maintenance, record that the inspection
is complete on the Maintenance Log. Perform the Inspecting the Reagent
Pipettors and Cleaning the Wash Towers procedure next.
Inspecting the Use this procedure to inspect and clean the reagent pipettors and the wash towers. The
Reagent pipettors and towers are located beneath the left upper main cover. It is not necessary
Pipettors and to inspect or clean the sample pipettor.
2. Locate the reagent pipettors. The reagent pipettors are mounted on separate
pipettor assemblies (for the UniCel DxI 800 instrument, see Figure 7-24; for the
UniCel DxI 600 instrument, see Figure 7-25).
3. Visually inspect the reagent pipettor fluidic lines for kinked tubing and for
crystalline deposits or corrosion at the tubing connections, and inspect the tubing
near the top of the pipettor for leaks.
4. If you find any kinked tubing, deposits, corrosion, or leaks, contact Technical
Support for assistance.
Kinked tubing can affect precision. Deposits or corrosion can indicate that a
connection or fitting is leaking.
5. Locate the wash towers. The reagent pipettor wash towers are underneath the
reagent pipettors on the lower right side. The sample pipettor wash tower is
underneath the sample pipettor on the lower left side of the instrument.
1 4 2
5 4 3
2330B.eps
1 Vessel hopper
2 Reagent pipettor assemblies (4)
3 Reagent pipettor wash towers (4)
4 Circuit board (1), beneath or to the side of the vessel hopper
5 Sample pipettor wash tower
1
4 2
5 4 3
2244G.eps
1 Vessel hopper
2 Reagent pipettor assemblies (2)
3 Reagent pipettor wash towers (2)
4 Circuit board (1), beneath or to the side of the vessel hopper
5 Sample pipettor wash tower
6. Check inside the wash tower overflow reservoirs for liquid or excessive
crystalline deposits (see Figure 7-26).
8. Wet a lint-free swab (or equivalent fiber-free applicator) with deionized water
and wipe any deposits from the reservoirs and exterior of the wash towers.
2262B.eps
9. If you are performing 10,000 test interval maintenance, record that the inspection
is complete on the Maintenance Log. Perform the Inspecting for Fallen RVs
procedure next.
Inspecting for Use this procedure to visually inspect for RVs that have fallen and are loose inside of
Fallen RVs the instrument.
WARNINGS
• Never wear exposed metal that can come into contact with electrical
components of the instrument.
• You will come in contact with potentially infectious materials during this
procedure. Handle and dispose of biohazard materials according to proper
laboratory procedures. Proper hand, eye, and facial protection is required.
WARNINGS
• The carriages contain many mechanical parts which could puncture the
skin. Proper hand, eye, and facial protection is required. Always be careful
when reaching for an RV.
• When reaching for an RV, be careful not to touch the circuit board beneath
or to the side of the vessel hopper. You can disturb important electronic
connections if you brush against the board.
3. If you find any fallen RVs, carefully remove them from the instrument and
dispose of them with your biohazard waste.
4. If you cannot remove an RV that is contributing to a system problem, contact
Technical Support for assistance.
5. Close both main upper covers. For more information, see Section 6.2: Instrument
Cover Procedures in the Reference Manual.
6. Exit the Device Diagnostics screen.
7. Initialize the system. For more information, see Section 4.2: Initialize System in
the Reference Manual.
8. If you are performing 10,000 test interval maintenance, record that the inspection
is complete on the Maintenance Log. If your DxI system uses on-board bulk
liquid waste containers, perform the Inspecting the Liquid Waste Drawer
procedure next. If your system uses a liquid waste drain, perform the Inspecting
the Wash Buffer Supply Drawer procedure next.
Inspecting the If your system uses on-board bulk liquid waste containers, use this procedure to
Liquid Waste visually inspect the liquid waste drawer for leaks. The liquid waste drawer is located
Drawer on the lower left as you face the instrument. If your system uses a liquid waste drain,
skip this procedure.
WARNING
You will come in contact with potentially infectious materials during this
procedure. Handle and dispose of biohazard materials according to proper
laboratory procedures. Proper hand, eye, and facial protection is required.
2. Check for any liquid outside of the bulk liquid waste containers, either pooled on
top of one or both of the bulk liquid waste containers, or at the bottom of the
liquid waste drawer.
Liquid waste can indicate that there is a leak in the fluidic drawer located above the
liquid waste drawer.
3. If there is any liquid outside of the bulk liquid waste containers, handle and
decontaminate the waste according to proper laboratory procedures. Then contact
Technical Support.
4. Check the connections at the back of the drawer for crystalline deposits or
corrosion at the tubing connections.
5. If you find any deposits or corrosion, contact Technical Support for assistance.
The presence of deposits or corrosion can indicate that a connection or fitting is
leaking.
7. If you are performing 10,000 test interval maintenance, record that the inspection
is complete on the Maintenance Log. Perform the Inspecting the Wash Buffer
Supply Drawer procedure next.
Inspecting the Use this procedure to visually inspect the wash buffer supply drawer. The wash buffer
Wash Buffer supply drawer is located on the lower right as you face the instrument.
Supply
WARNING
Drawer
Wash buffer contains a preservative which may cause sensitization by skin
contact. After contact with skin, wash immediately with soap and water. Wear
suitable gloves.
2. Check the condition of both bulk wash buffer containers. Be sure the containers
are in good condition and are not beginning to collapse.
A collapsing container can indicate that the wash buffer cap filter is clogged and is
creating a vacuum.
4. Check for any wash buffer outside of the bulk wash buffer containers, either
pooled on top of one or both of the containers, or at the bottom of the wash buffer
supply drawer.
The presence of wash buffer can indicate that there is a leak in the tubing, or that the
cap and tube assembly is loose or connected incorrectly.
5. If you find that the cap and tube assembly is loose, reconnect the assembly. For
more information, see the Changing an Empty Bulk Wash Buffer Container
procedure in Section 2.3: Wash Buffer.
6. If you find a leak in the wash buffer tubing, contact Technical Support for
assistance.
7. Check the tubing and connections at the back of the drawer for kinks, or
crystalline deposits and corrosion at the tubing connections.
8. If you find any kinks, deposits, or corrosion, contact Technical Support for
assistance.
Deposits or corrosion can indicate that a connection or fitting is leaking.
10. If you are performing 10,000 test interval maintenance, record that the inspection
is complete on the Maintenance Log. Perform the Cleaning the Air Filter
procedure next.
Cleaning the Use this procedure to clean the air filter. The air filter is located behind the lower right
Air Filter door as you face the instrument.
WARNING
Use extreme caution when handling the air filter. The filter edges are sharp and
can puncture the skin if the air filter is handled carelessly.
CAUTION
Do not run the instrument if the air filter is not in place. Running without the
air filter may result in dust buildup on critical instrument components.
Required Materials
• Hand-held vacuum device
2. Locate the air filter near the bottom of the cabinet (see Figure 7-27).
3. Grasp the filter by its metal frame and slide the filter out of the filter housing.
4. Turn the filter over so the screen cage faces the floor, and then lay the filter down
on a solid surface.
5. Vacuum the filter until most of the dust and dirt is removed.
6. Turn the filter so the screen cage faces up and slide the filter into the filter
housing.
7. When the air filter is inside the housing, gently close the lower right door.
8. If you are performing 10,000 test interval maintenance, record that the filter is
clean on the Maintenance Log. Then perform the following steps:
a. Perform daily maintenance. For more information, see Section 7.2: Daily
Maintenance.
b. Verify system performance. For more information, see the Verifying System
Performance procedure.
1 2
3
2243A.eps
Use the Maintenance Log as a checklist to confirm that you perform each routine
maintenance procedure. The log also serves as a daily record of the instrument test
count. One month of routine maintenance can be recorded on the log.
Review the instrument test count daily and schedule the test interval maintenance
procedures as needed, based on the number of tests you run on the instrument. If your
laboratory runs tests for HIV antibody levels, review the log and schedule special
weekly maintenance accordingly.
Use the Maintenance Log in the Instructions for Use manual as an original and make
photocopies for use with each UniCel DxI system in your laboratory. You can also
print copies of the Maintenance Log from the Help system. Beckman Coulter
recommends that you save completed Maintenance Logs in your Maintenance and
Service Log binder for reference.
C73199-AA
(Turn form to record number)
5,000 TEST INTERVAL MAINTENANCE On Date On Date On Date On Date On Date On Date
7-63
This page may be reproduced for laboratory use.
7.6: Maintenance Log UniCel DxI Operator's Guide
Sample Containers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . -8
The container type label includes an illustration of the only sample container type you
can use with the rack. The rack bar code label includes the bar code and rack ID
number. The Rack ID label distinguishes one sample rack from another, and also
identifies the accepted sample container type for the rack. The Reserve Volume label
identifies the type of reserve volume designated for the rack.
NOTE
You can find the rack ID ranges set up for each type of sample container from the
System Setup screen. For more information, see Section 2.3: Supplies Setup in
the Reference Manual.
The instrument recognizes the type of sample containers held in the sample rack when
it scans the rack bar code label.
For information about attaching labels to racks, see Section 2.3: Supplies Setup in the
Reference Manual.
Rack Sizes The following table lists the different sizes of racks, and the accepted container types
and for each rack size. For more information about sample containers, see
Containers Section A.2: Sample Containers.
CAUTION
A rack is configured to accept only one type of sample container. Do not mix
types of containers within a rack. You can find the rack ID ranges set up for
each type of container from the System Setup screen. For more information,
see Section 2.3: Supplies Setup in the Reference Manual.
Cleaning You can clean your sample racks in a dishwasher or an autoclave. You can also wash
Sample Racks the racks by hand using a brush and a mild detergent.
NOTE
You can load patient and QC samples in the same rack as long as the samples are
in the same type of container. A rack that contains calibrators cannot include
patient or QC samples.
Place most containers directly in a rack. If you use insert cups, place the cups into the
appropriate containers or rack inserts before placing them in a rack. For more
information about loading racks, see Section 3.4: Loading and Unloading Racks.
WARNINGS
• Because the system software contains specific parameters for each of the
valid sample containers, only use the recommended containers on the
UniCel DxI instrument. Other containers may damage the system and
compromise the integrity of your test results.
• A rack is configured to accept only one type of sample container. Placing an
incorrect container in a rack may damage the system and compromise the
integrity of your test results. You can find the rack ID ranges set up for each
type of container from the System Setup screen. For more information, see
Section 2.2: System Setup in the Reference Manual.
CAUTION
Remove caps from all sample containers before loading on the UniCel DxI
system.
Calculating The sample volume in a container must be sufficient to process the requested tests,
Minimum and can be increased to accommodate reflex tests or patient and QC reruns. For more
Sample information about setting the reserve volume, see Section 3.7: Reflex Test Setup in
the Reference Manual.
Volume
Use this equation whenever you need to calculate the minimum sample volume for a
container in a rack.
A B C D E
Minimum sample
Sample assay + System dead + Reserve volume + Sample pipettor + Sample container = volume required
volume volume (if set up) overdraw dead volume
2223B.wmf
The sample assay volume is the sum of the sample volumes for each requested test. To find the
A
sample volume for each test, see the corresponding reagent instructions for use.
The system dead volume is the amount of sample that cannot be drawn from an RV inside of the
B
instrument. The system dead volume is 60 mL for each RV.
(If set up) The reserve volume is the amount of sample the system stores for additional testing. For
C more information, see the Setting Up Reserve Volume topic in Section 3.7: Reflex Test Setup of the
Reference Manual.
The sample pipettor overdraw is 20 mL or 5% of the volume of sample in the RV (A + B + C),
D
whichever is greater.
The sample container dead volume is the amount of sample that cannot be aliquoted from the sample
E
container. To find the dead volume for a specific sample container, see Table A-2.
NOTE
If the sum of the sample and reserve volumes (A + C) is greater than 500 mL, the
system aliquots the reserve volume into 1-2 additional RVs. For each additional
RV, include the system dead volume and the calculated amount of sample
pipettor overdraw.
Example
Sample volume aliquot in one RV - The PSA assay has a sample volume of
25 mL, and the fPSA assay has a sample volume of 25 mL. The system is set up to
draw 200 mL of reserve volume. Here is how to calculate the minimum sample
volume required for these two assays if you are using a 2 mL cup.
Example
Sample volume aliquot in two RVs - The PSA assay has a sample volume of
25 mL, and the fPSA assay has a sample volume of 25 mL. The system is set up to
draw 500 mL of reserve volume. Here is how to calculate the minimum sample
volume required for these two assays if you are using a 2 mL cup.
.
Sample The following table lists the sample containers that are accepted on the UniCel DxI
Containers instrument, along with their dead volume requirements.
WARNING
Racks are configured to accept only one type of sample container. Sample
containers must match the ID configured for the racks you use. Placing an
incorrect container in a rack may damage the system and compromise the
integrity of your test results. You can find the rack ID ranges set up for each type
of container from the System Setup screen. For more information, see
Section 2.2: System Setup in the Reference Manual.
CAUTION
Remove caps from all sample containers before loading on the UniCel DxI
system.
Glass
• Dead volume: 500 mL
• Dead volume: 200 mL
12/13
x75
12/13
x75 13x100
16x75
Insert
• Dead volume: 350 mL • Dead volume: 400 mL
1 mL in 1 mL in
13x75 13x100
0.5 mL cup 2 mL cup
• Dead volume: 80 mL • Dead volume: 150 mL
2 mL
2 mL in
16x100 3 mL
Aliquot PED
Table A-2 Sample Containers and Their Dead Volume Requirements (continued)
Index
electrical warning 7-33
Symbols handling with caution 7-33
10,000 test maintenance HIV testing and maintenance 7-19
overview 7-3 installing (flowchart) 7-41
procedures 7-31 installing clean 7-36
5,000 test maintenance loosening fittings 7-34
overview 7-3 positions 7-34
procedures 7-21 potentially infectious 7-33, 7-42
removing 7-33
A replacing 7-32
Access Immunoassay Systems syringe assembly 7-43
family 1-2 verifying deflection 7-28, 7-39
UniCel DxI 1-2 assay calibrations
active calibrations 5-27 active 5-27
switching 5-32 calibration data report 5-33
calibration report 5-8
AEX result flag 4-27
calibration screen 5-5
air bubbles
calibration screen filter 5-6
prevent when pipetting 3-43
calibrator setup 5-10
air filter comments 5-33
cleaning 7-60 curves 5-26
location 7-61 cutoffs 5-26
sharp edges 7-60 deleting 5-7
amber light excluded replicates 5-32
monitoring supply status 2-4 expired 5-3
status indicator 1-8, 2-4 failed qualitative 5-38
APF failure codes 5-36
see assay protocol file flowchart 5-4
applications last run 5-27
running other 1-15 most recent data 5-27
aspirate and dispense probe manifold 7-35 overview 5-3
poor precision 5-38
aspirate probes
previous active 5-27
10,000 test maintenance 7-32
problems with 5-37
brushing interior 7-42
reagent lot assignment 5-25
cleaning 7-42
running a calibration 5-19
cleaning (flowchart) 7-44
switching 5-32
I L
IND result flag 4-28 labels
insert cups also see bar code labels
dead volume requirements A-8 also see safety symbols
in sample tubes 3-43 rack container type A-7
loading 3-43 sample racks A-2
pediatric A-3 laboratory information system (LIS)
preventing air bubbles 3-43 also see LIS
rack container type labels A-8 software overview 1-14
rack sizes A-3 laser safety symbol 1-9
inspections laser warning labels
dispense probes 7-45 location 1-9
fallen RVs electrical warning 7-56 last run calibrations 5-27
for fallen RVs 7-56 LED Safety 1-13
liquid waste drawer 7-58
Levey-Jennings charts 6-33
pump connections at wash buffer
reservoir 7-49 LEX result flag 4-34
reagent pipettors 7-53 lights
upper aspirate peristaltic pump amber status indicator 1-8
connections 7-51 blue status indicator 1-8
wash buffer supply drawer 7-59 green status indicator 1-8
installation red status indicator 1-7
aspirate probes 7-32 liquid waste
duck bill valve 7-27 changing container (flowchart) 2-46
DxI system 1-8 changing full container 2-43
Instructions for Use changing in-use container 2-47
Instrument - Preface-2 disposal 2-44
Instrument in-use container 2-47
Instructions for Use - Preface-2 monitoring levels 2-42
panel icon (DxI 800 only) 2-42
instrument
potentially infectious 2-41
10,000 test maintenance 7-31
removal 2-43
5,000 test maintenance 7-21
supply overview 2-41
daily maintenance 7-12
DxI 600 illustration 1-4 liquid waste container
DxI 800 illustration 1-3 flushing 2-45
overview 1-3 liquid waste drawer
special weekly maintenance 7-19 inspection 7-58
temperatures 7-5 location 2-41
instrument software potentially infectious 7-58
overview 1-14 LIS