Pwi Iom
Pwi Iom
Pwi Iom
STDS-PWI-IOM-TM-0.0 1/26
8885. Monroe St. Houston, Texas USA 77061
Phone: (713) 956-2002
Fax(713) 956-2141
TABLE OF CONTENTS
TABLE OF FIGURES
FIGURE 1 ....................................................................................................................................................... 7
FIGURE 2 ....................................................................................................................................................... 9
FIGURE 3 ..................................................................................................................................................... 26
TABLE OF TABLES
TABLE 1 ....................................................................................................................................................... 13
TABLE 2 ....................................................................................................................................................... 19
TABLE 3 ....................................................................................................................................................... 24
TABLE 4 ....................................................................................................................................................... 25
STDS-PWI-IOM-TM-0.0 2/26
8885. Monroe St. Houston, Texas USA 77061
Phone: (713) 956-2002
Fax(713) 956-2141
The PumpWorks 610 PWI pump is a single stage, vertical in-line centrifugal pump manufactured
in accordance with the American Petroleum Institute (API) Standard 610.
Our PWI-BB (OH3) design features a vertical bearing housing system to take the radial and
axial forces exerted upon the pump, and connects the pump and driver with a flexible spacer
coupling. The standard PWI (OH4) design relies on the bearings in the vertical driver to take the
forces, and utilizes a rigid spacer coupling design. Both systems allow for the removal of the
mechanical seal without removing the driver or suction and discharge piping.
NOTE: All reference numbers (provided in parentheses) in this manual refer to the typical
sectional drawing, Figure 3.
Inspection
Upon receipt, carefully inspect the unit for damage and validate against the bill of lading. Report
any damaged or missing items to the carrier's local representative and submit a copy of the
report to PumpWorks 610. While removing packaging do not discard any small accessories that
may be attached.
Handling
Lift the complete unit using the provided lifting lugs on the pump driver stand (3160) with proper
lifting techniques.
1. (PWI-BB only) Prior to storage, remove the oil vent plug and fill the bearing housing with one
quart of vapor emitting oil. Reinstall vent plug.
STDS-PWI-IOM-TM-0.0 3/26
8885. Monroe St. Houston, Texas USA 77061
Phone: (713) 956-2002
Fax(713) 956-2141
2. Coat all unpainted exterior machined surfaces liberally with a light petroleum grease or
equivalent rust preventative.
3. Rotate pump shaft (2210) 3.5 turns (180º from original position).
4. Remove the protective covers on the suction and discharge flanges. Remove any visible rust
on internal surfaces and coat with a light petroleum grease or equivalent rust preventative.
Replace protective covers on the flanges.
NOTE: Accumulation of condensation in the unit must be avoided. Store the unit away from
climatic extremes.
Section 3 - Installation
General
Support and anchor suction and discharge piping independently near the pump so strain will not
be transmitted to the pump casing when flange bolting is tightened. Piping must independently
align with the pump flanges; never force piping into place at suction or discharge flanges.
If an expansion joint or non-rigid pipe coupling is used, install a pipe anchor as near to the
suction and/or discharge flanges as possible. Proper installation of the pipe anchor will eliminate
any undesirable forces on the pump.Use large-radius elbows wherever possible. Flush all
piping thoroughly to remove any foreign matter before connecting to the pump. Per the
Hydraulic Institute, there should be at least five to ten pipe diameters of straight pipe length
coming into the suction before the first fitting, strainer, or valve is installed, and it is preferable to
have the same conditions after the discharge.
Location
The unit should be positioned as close to the suction source as practical, in order to optimize
suction conditions, while ensuring sufficient clearance around the unit to allow for cooling and
maintenance accessibility.
STDS-PWI-IOM-TM-0.0 4/26
8885. Monroe St. Houston, Texas USA 77061
Phone: (713) 956-2002
Fax(713) 956-2141
Suction Piping
1. The nominal suction pipe diameter must be the same size as the nominal suction flange size.
2. Ensure that suction lines are sealed to avoid leakage and air pockets.
3. Per the Hydraulic Institute, reducers, if used, should be of the conical type and sufficiently
long to prevent fluid turbulence. Contour type reducers are not recommended.. Eccentric or
concentric reducers may be used when the liquid source is above the pump and the suction
piping is sloping upwards toward the source.
4. Start-up, cone-type suction strainers should be used and must have a net free area of at least
three times the suction pipe area; they should point upstream into oncoming flow.
5. Installing a gate valve in the suction line permits closing the line for pump inspection and
maintenance.
CAUTION: Never throttle the pump with the suction valve; while pump is in operation, this valve
should always be fully open.
Discharge Piping
1. A check valve and gate valve should be installed in the discharge piping. The check valve,
placed between the unit and gate valve, prevents pumpage from running back through the unit.
The gate valve is used in priming, starting, and shutdown of the unit and to control discharge
flow.
2. If increasers are used in the discharge piping, they should be located between the check
valve and the unit.
STDS-PWI-IOM-TM-0.0 5/26
8885. Monroe St. Houston, Texas USA 77061
Phone: (713) 956-2002
Fax(713) 956-2141
General
When pumps and drivers are received from the factory with the motor installed and pinned on
the pump driver stand, they have been accurately aligned before shipment. Realignment is
necessary after the suction and discharge piping have been installed.
Alignment should be checked after working on the mechanical seal or removing the driver to
ensure minimal repair and maintenance of the unit after re-assembly.
Driver Lubrication
Motor bearings should be serviced and lubricated as outlined in the motor manufacturer's
instructions. For the PWI (OH4), all pump hydraulic thrust loads are taken by the motor
bearings.
The bearing housing incorporates an oil flinger and pumping ring to circulate oil to the bearings.
Fill the bearing housing to the proper level (Centerline of the bull’s eye) with ISO/ASTM VG 32
oil. For lubrication changing intervals, refer to Section 7 Operational Checks.
The constant level oiler bottle will be found already mounted to the bearing housing. Adjustment
should be made as detailed below:
Fill the bearing housing with proper grade of oil through the constant level oiler cup to 1/4" (6.35
millimeters) below the level mark. Then fill the oiler bottle with oil and place it in the oiler cup.
The housing is filled when oil remains in the oiler bottle.
If the oil level is not at the proper level, remove bottle and reset level adjuster (See Figure 1).
Replace bottle (repeat operation until proper level is obtained). Thereafter, it is only necessary
to keep bottle filled with oil.
STDS-PWI-IOM-TM-0.0 6/26
8885. Monroe St. Houston, Texas USA 77061
Phone: (713) 956-2002
Fax(713) 956-2141
Figure 1
Mechanical Seals
As a standard, the PWI pump is equipped with a cartridge-type mechanical seal (4200), which
complies with the service requirements outlined on the pump data sheet. The seals are factory-
installed, and no adjustment is required.
NOTE: Seal setting devices must be disengaged prior to start-up, but are required to be
installed when removal of the coupling spacer is necessary. The mechanical seal is not
designed to withstand the entire weight of the rotating assembly.
Rotation
The driver rotation must be verified prior to start-up. To check rotation, remove the coupling
guard and coupling spacer to disengage the pump from the driver. Make momentary contact
with the motor starter, also known as a “bump test”, to verify correct rotation.
STDS-PWI-IOM-TM-0.0 7/26
8885. Monroe St. Houston, Texas USA 77061
Phone: (713) 956-2002
Fax(713) 956-2141
WARNING: Keep clear of rotating components when checking the motor rotation. Be sure that
tools and loose hardware are clear before starting motor.
A spacer type coupling is used between the pump and driver. To align the spacer coupling, first
replace the mechanical seal setting devices per the seal manufacturer’s guidelines, loosen the
set screws fastening the seal drive collar to the shaft, remove the spacer between the pump and
driver, then remove the pump coupling hub and mechanical seal. Either construct solid
brackets or use magnetic clamps, as shown in Figure 2, in order to hold both dial indicators at
the required locations.
NOTE: All T.I.R. measurements are to be taken perpendicularly to each face marked in Figure
2.
Position 1: Fix one bracket/clamp to the motor coupling hub horizontal face and assemble the
dial indicator so that it is located and oriented on the seal chamber bore. Fix the other
bracket/clamp to the pump seal chamber horizontal face and assemble the dial indicator so that
it is located and oriented on the outer diameter of the motor coupling hub flange as shown.
Maximum total run-out readings on both dial indicators shall be per coupling manufacturer
recommendation, and not exceed 0.005” T.I.R. If the runout exceeds this value, loosen the
motor mounting bolts and realign the motor. Once the motor is realigned, carefully tighten the
motor mounting bolts and repeat this procedure as many times as necessary.
NOTE: It is not necessary to have two dial indicators. Position 1 can be done as a two-step
process with a single indicator.
Position 2: Using the first bracket mounted to the motor hub, align the dial indicator on the back
face of the case cover plate. The runout should not exceed 0.002”.
Position 3: Mount a bracket on the back of the case cover face, seen as the left bracket in the
figure, and align the indicator to the inside edge of the motor hub. The runout should not exceed
0.0005”.
Position 4: Using the same bracket mounting face location from the third position, check the
runout of the motor hub by indicating on the face as shown. This runout should also not exceed
0.0005”.
Position 5: Do a final check on the pump shaft by fully assembling the coupling and setting up a
bracket on the case cover face and indicating on the shaft. This runout should not exceed
0.002”.
STDS-PWI-IOM-TM-0.0 8/26
8885. Monroe St. Houston, Texas USA 77061
Phone: (713) 956-2002
Fax(713) 956-2141
Figure 2
WARNING: Bring the pump up to the operating temperature slowly before start up. Severe
thermal shock can damage the pump and cause injury.
STDS-PWI-IOM-TM-0.0 9/26
8885. Monroe St. Houston, Texas USA 77061
Phone: (713) 956-2002
Fax(713) 956-2141
Section 5 – Operation
Operating Capacity
Centrifugal pumps should not be operated at greatly reduced capacity or with a closed
discharge valve because the energy required to drive the pump is converted to heat and the
temperature of the liquid may reach its bubble point. If this occurs, the rotating parts are
exposed to vapor with no lubrication and damage to internal parts will occur.
1. Add a liquid temperature sensor to shut the pump down if the pumpage temperature
exceeds a predetermined level. See the High Temperature Operation passage of
Section 4 for some guidelines.
2. Add a flow meter to the suction with an alarm level set for the minimum continuous safe
flow (MCSF) ability of the pump, which can be found on the pump data sheet. Operating
below the MCSF increases the radial loading on the system, which can lead to
shortened bearing life, increases the NPSH of the pump, and can lead to internal
cavitation.
3. Installing a constant open minimum-flow by-pass between the pump discharge and the
suction source should unexpected blockage occur in the pipeline.
4. Adding a suction pressure sensor upstream from the pump to shut the pump down if the
suction pressure drops below a predetermined level.
5. Adding a vibration sensor to shut the pump down if the vibration level exceeds a
predetermined value.
NOTE: Never throttle pump on suction side. The suction valve should always be fully open
during pump operation.
Freezing
During cold weather when the pump is not in operation, the pump should be drained to prevent
the liquid inside from freezing. Proper storage procedures should be followed while the pump is
not in use.
WARNING: If the pumpage is toxic, flammable, or corrosive, take proper precautions for
handling pumpage before draining the pump.
Shut-Down
To avoid water hammer, it is recommended that the discharge valve be partially closed to the
minimum flow point just prior to shut-down of the pump.
STDS-PWI-IOM-TM-0.0 10/26
8885. Monroe St. Houston, Texas USA 77061
Phone: (713) 956-2002
Fax(713) 956-2141
Start-up Checklist
Priming
NOTE: The pump casing, suction line and seal flush must be filled with liquid and fully vented
BEFORE the pump is started.
Starting
A centrifugal pump usually requires less power to operate with the discharge valve closed than
with the discharge valve open. For this reason, it is recommended that this valve be partially
closed when starting. However, ensure the pump is not operating below the minimum
stable flow conditions at any time.
Start the pump and immediately bring to operating speed. Slowly open the discharge valve as
soon as there is a discharge pressure indication. Continue opening the discharge valve until
rated capacity and discharge pressure are obtained. If the rated conditions cannot be obtained,
refer to the Trouble Checklist in Section 8.
STDS-PWI-IOM-TM-0.0 11/26
8885. Monroe St. Houston, Texas USA 77061
Phone: (713) 956-2002
Fax(713) 956-2141
First 60 Minutes
It is recommended that the following parameters be recorded at 10 minute intervals during the
first 60 minutes of operation and periodically thereafter:
Satisfactory operation is indicated by correct capacity, discharge pressure, and low vibration
levels. Keep records for future troubleshooting reference and performance trend analysis.
A periodic check of the items listed above will ensure the pump is maintained in its best
operating condition.
Mechanical Seal
Check mechanical seals for leakage during the first hours of operation. Minor leakage through
the seal usually stops after a short operation period, but, if leakage continues, shut down the
pump and investigate the cause. Excessive leakage past the seal generally indicates worn or
broken parts requiring replacement.
Lubrication
Driver Lubrication
Driver bearings should be serviced and lubricated according to the motor manufacturer's
instructions. Consult lubrication directions on nameplate and literature provided in the final data
package.
Great care should be exercised to keep the housing clean and only clean lubricants should be
used. Foreign solids or liquids within the bearing housing can completely ruin the bearings in a
short time. Keep the oiler bottle filled with the correct grade of oil (See Section 4). Under normal
conditions the oiler will maintain proper oil level. A routine check of the oil level will verify proper
working order of the oiler.
After the initial start-up, change the oil after 24 hours of running time. Thereafter, change the oil
at the following intervals:
STDS-PWI-IOM-TM-0.0 12/26
8885. Monroe St. Houston, Texas USA 77061
Phone: (713) 956-2002
Fax(713) 956-2141
When changing the oil, flush the inside of the housing with clean oil to remove any accumulated
contaminants.
Due to rolling friction and drag of the bearing races, heat is generated within the bearings and
they will operate at temperatures above the surrounding atmosphere unless cooled. Oil
lubricated ball bearings can safely be operated up to 180°F (82.2°C) and bearing temperatures
of 160°F (71.1°C) are normal. Do not use the human hand as a thermometer. Determine the
temperatures accurately by placing a contact type thermometer against the bearing housing.
Record this temperature on a regular basis and maintain a log of the reading. A stable
temperature indicates normal operation.
Check to see that oil is of proper viscosity and the oil level is neither too high nor too low.
The unit should also be checked for unstable hydraulic operation and unnecessary mechanical
loads, such as shaft misalignment.
NOTE: Stop the pump immediately if bearing housing temperature exceeds 190° F (88°
C). Inspect for possible problems such as those stated above before restarting the
pump.
NOTE: Bearing housings equipped with water cooling coils are to be adjusted to
operate so as to maintain an oil temperature no less than 25°F above the
surrounding ambient temperature to prevent condensation.
STDS-PWI-IOM-TM-0.0 13/26
8885. Monroe St. Houston, Texas USA 77061
Phone: (713) 956-2002
Fax(713) 956-2141
STDS-PWI-IOM-TM-0.0 14/26
8885. Monroe St. Houston, Texas USA 77061
Phone: (713) 956-2002
Fax(713) 956-2141
Section 9 - Maintenance
General
The procedures given below are general and apply to all PWI pumps. For procedures specific to
a particular unit, refer to the drawings and instructions supplied in the final data package with
that unit.
WARNING: Be sure that power to the motor is off and locked out before starting maintenance
procedures.
Pump Disassembly
The disassembly procedure applies generally to the PWI series supplied with a cartridge-type
seal. Refer to the specific drawings supplied with your unit to augment this procedure.
NOTE: Typical sectional drawings are for disassembly and assembly purposes only and should
not be used for specific detailed dimensions.
WARNING: Be sure the power to the motor is turned off and locked in the off position
before beginning disassembly.
2. Close suction and discharge gate valves and any auxiliary valves.
NOTE: The back pull-out design of this pump allows the complete Back Pull-Out unit to
be removed without disturbing the suction and discharge piping or the driver.
WARNING: If the pumpage is toxic, flammable or corrosive take proper precautions for
handling pumpage before draining the pump.
4. Disconnect and remove all seal flush piping after the pump is completely drained.
5. If the pump is equipped with an auxiliary seal flush reservoir, drain the reservoir and
disconnect the seal flush piping.
NOTE: Always cap off open ports and lines to prevent dirt from entering.
STDS-PWI-IOM-TM-0.0 15/26
8885. Monroe St. Houston, Texas USA 77061
Phone: (713) 956-2002
Fax(713) 956-2141
7. Replace seal setting devices per seal manufacturer’s recommendation, then loosen
set screws fastening seal drive collar to shaft (2110). Remove nuts (6580.2) attaching
seal (4200) to casing cover (4100).
8. Disassemble and remove coupling spacer. Remove pump coupling hub and coupling
key (6700.2).
9. (PWI ONLY) Remove mechanical seal (4200) from casing cover (4100). Refer to seal
manufacturer’s guidelines for proper mechanical seal maintenance.
10. With lifting strap or hook attached to driver eyebolt or lifting lug, remove cap screws
that secure driver to driver stand (3160) and lift driver off of the driver stand.
11. Remove cap screws (6577) that secure driver stand (3160) to pump casing (1110) or
casing cover (4100) and lift driver stand off of pump casing or casing cover.
WARNING: Do not work under a heavy suspended object unless there is a positive
support under it which will protect personnel should a hoist or sling fail.
CAUTION: The back pull-out unit is heavy. Proper lifting equipment must be used to
avoid injury.
12. Remove nuts (6580.1) attaching casing cover (4100) to pump casing (1110) and
remove casing cover by tightening jack bolts in the tapped holes in casing cover to
facilitate removal of the rotating assembly. Remove casing gasket (4510).
(PWI-BB ONLY) WARNING: The back pull out unit with the bearing housing is very
heavy. Proper lifting equipment must be used to avoid injury.
After the complete back pull-out unit has been taken to a clean work area, the unit can be fully
dismantled by following the instructions given below.
NOTE: The impeller locknut loosens in a clockwise direction. It has a left hand thread.
2. Remove impeller (2200) and impeller key (6700.1). The impeller is a slip fit but may
require the use of a puller for removal.
CAUTION: Do not bend or damage the impeller when removing with a puller.
3. To disassemble casing/casing cover wear rings (1500.1, 1500.2) remove set screws
holding the wear ring. Remove the wear ring. Note that some wear rings are fuse-welded
in place; in this case, grind off the fuse welds and remove the wear rings.
STDS-PWI-IOM-TM-0.0 16/26
8885. Monroe St. Houston, Texas USA 77061
Phone: (713) 956-2002
Fax(713) 956-2141
4. To remove the sleeve bearing (3000), use a press. Examine casing cover (4100) to
ensure there is no damage that will prevent installation of a new sleeve bearing.
CAUTION: Sleeve bearings are typically made of a soft carbon material. Too much
force or impulse will cause the sleeve bearing to crack or shatter.
After the complete back pull-out unit has been taken to a clean work area, the unit can be fully
dismantled by following the instructions given below and referring frequently to the pump
sectional drawing Figure 4.
1. Remove impeller locknut set screw, if installed, and impeller locknut (821-1). NOTE:
The impeller locknut loosens in a clockwise direction. It has a left hand thread. Also,
some lock nut designs do not require set screws.
2. Remove impeller and impeller key. The impeller is a slip fit but may require the use of
a puller for removal. (805, 811-1)
CAUTION: Do not bend or damage the impeller when removing with a puller.
4. Engage seal locating devices. Loosen the cartridge seal drive collar set screws. Refer
to seal drawing in final data package.
5. Unscrew cap screws holding bearing bracket (849) to casing cover (802) and pull
casing cover and seal from remaining rotating assembly.
CAUTION: When separating the casing cover from the bearing bracket, be careful not to
damage the mechanical seal.
6. Unbolt and remove the cartridge seal assembly and seal chamber gasket from casing
cover. Refer to the manufacturer’s instructions for seal maintenance.
7. To disassemble casing/cover wear rings (808-1,2), remove set screws holding the
wear ring. Remove the wear ring. NOTE: Some Wear Rings Are Fuse Welded In Place.
8. To remove the throat bushing (847), grind off welds between cover and bushing and
remove bushing.
STDS-PWI-IOM-TM-0.0 17/26
8885. Monroe St. Houston, Texas USA 77061
Phone: (713) 956-2002
Fax(713) 956-2141
After dismantling the casing cover, the bearing bracket can be disassembled.
2. Remove the fan guard (812), inboard heat sink (829-2), and fan (853) as applicable.
3. Loosen capscrews and remove outboard end cap/bearing seal assembly (813-2, 829-
2), and plastic shims (815-3).
5. Remove ball bearing locknut and lockwasher (821-2 and 831) from fan shaft.
6. Using an arbor press with appropriate fixturing, remove thrust bearings (881-2) and
radial bearing (881-1) from the pump shaft.
7. Remove oil rings (817) from shaft or loosen set screws and remove oil flinger.
Impeller
Inspect impeller (2200) passages and vane surfaces for evidence of erosion. Replace if
excessively worn or corroded. The impeller is dynamically balanced at the factory, and balance
must be maintained for proper operation of the pump.
Impeller wear rings (1500.1, 1500.2, 2300.1, 2300.2) are always provided on the front of the
impeller, and may or may not be on the back side depending on the specific process conditions
(2200). These rings allow a small clearance to be maintained between the rotating impeller and
the stationary casing (1110). For proper hydraulic performance, rings should be replaced when
excessive increase in clearance begins to affect performance.
STDS-PWI-IOM-TM-0.0 18/26
8885. Monroe St. Houston, Texas USA 77061
Phone: (713) 956-2002
Fax(713) 956-2141
NOTE: For materials with known galling tendencies, and for all materials operating at
temperatures above 500º F, add 0.005"(127 micrometers) to the above diametrical clearances.
Pumps utilizing special non-galling wear materials may use smaller wear ring clearances. The
clearances referenced above act as a guideline, with the final clearances determined by
PumpWorks 610 for each process condition.
Seals
Inspect all mechanical seals (4200) for irregularities or damage. Consult the seal manufacturer's
data for seal reconditioning and service. Seal faces, "O" rings and the seal sleeve must be in
perfect condition. Replace all worn parts.
Shaft
Inspect the shaft (2110) for damage and straightness. Dress minor damage and polish areas
where the shaft contacts a seal. Support the shaft in rollers or V-blocks about ¼ of the overall
length from each end and check run-out. Run-out must not exceed .002 inch TIR on all
diameters.
General
All parts should be cleaned before assembly. This is especially important at "O" ring grooves,
threads, cylindrical fits and gasket surfaces. Any burred edge must be removed before part is
installed into the pump. Coat all parts with light oil and cover with protective cloth if the pump is
not immediately reassembled.
STDS-PWI-IOM-TM-0.0 19/26
8885. Monroe St. Houston, Texas USA 77061
Phone: (713) 956-2002
Fax(713) 956-2141
Assembly
When assembling this pump it is recommended that only genuine PumpWorks 610 parts be
used. Always use new O-rings, gaskets and lock washers. Assembled parts must be clean and
free of dust or dirt.
The pump unit can be reassembled by following the instructions given below.
1. If removed during disassembly, press sleeve bearing (3000) into casing cover (4100).
CAUTION: Sleeve bearings are typically made of a soft carbon material. Too much
force or impulse will cause the sleeve bearing to crack or shatter.
3. Lubricate pump shaft (2110) and slide cartridge seal (4200) onto shaft with seal
setting devices in place.
CAUTION: Care must be taken not to damage seal sleeve gasket or seal faces.
5. Slide casing cover (4100) over pump shaft (2110) from the impeller side.
6. Install casing gasket (4510) over casing cover (4100) pilot outer diameter.
7. Install impeller key (6700.1) and impeller (2200). Due to the small clearance between
the impeller bore and mating shaft (2110) diameter, gentle force with a soft mallet may
be required to ensure impeller is fully seated against shaft shoulder.
8. Lift back pull-out unit (impeller, shaft, casing cover, and seal) with shaft (2110) axis
oriented vertically and install casing cover (4100) on casing (1110) and tighten casing
stud nuts (6580.1) evenly, alternating between diametrically opposed nuts in a circular
pattern.
9. Using the pilot outer diameter on the casing (1110) or casing cover (4100), install the
driver stand (3160) on the casing or casing cover. Tighten cap screws (6577) mating the
driver stand to the casing or casing cover.
10. If motor was removed from driver stand (3160), install the motor on the driver stand.
STDS-PWI-IOM-TM-0.0 20/26
8885. Monroe St. Houston, Texas USA 77061
Phone: (713) 956-2002
Fax(713) 956-2141
12. Tighten seal gland nuts (6580.2) evenly until metal to metal contact is realized
between seal gland and cover.
13. Check shaft alignment for proper coupling performance. Refer to Preparation For
Start-Up.
The bearing bracket can be reassembled by following the instructions given below and by
referring to the pump drawing Figure 4.
NOTE: The bearing bracket is common for both the mechanical seal and Seal Reservoir
arrangements.
1. Press inboard bearing seal (829-1) and firmly seat into bearing bracket (849) if
removed during disassembly.
NOTE: Bearing seal oil return slot must be located such that it faces down with the
bearing bracket in its normally mounted position.
2. Assemble oil rings, if applicable, (817) on shaft (820), installing each in the groove in
the shaft.
3. Assemble oil flinger, if applicable, (817) on shaft (820) against the shoulder and lock
in place with set screws.
4. Lubricate thrust bearing seat on shaft. Slide thrust bearing (881-2) on shaft (820) as
far as possible by hand. Place sleeve over shaft, being sure it rests against inner race
only. Press sleeve evenly until bearing is seated firmly against shaft shoulder.
NOTE: Heating the bearings using an induction bearing heater or hot oil bath can be
used to obtain a slip fit assembly. Be sure the bearing is not magnetized if induction
heating is used. Do not heat bearings above 250°F.
NOTE: Install thrust bearings back to back. Generally this means that the markings on
the outer race are installed together. Bearings must always be replaced in pairs.
STDS-PWI-IOM-TM-0.0 21/26
8885. Monroe St. Houston, Texas USA 77061
Phone: (713) 956-2002
Fax(713) 956-2141
5. Assemble thrust bearing lockwasher (831) and locknut (821-2). Crimp tab of
lockwasher into bearing locknut slot.
6. Lubricate radial bearing seat on shaft. Slide radial bearing (881-1) on shaft (820) as
far as possible by hand. Place sleeve over shaft, being sure it rests against inner race
only. Press sleeve evenly until bearing is firmly seated against shaft shoulder.
NOTE: Heating the bearings using an induction bearing heater or hot oil bath can be
used to obtain a slip fit assembly. Be sure the bearing is not magnetized if induction
heating is used. Do not heat bearings above 250°F.
8. Install outboard bearing end cap/bearing seal assembly (813-2,829-2), and plastic
shims (815-3) onto bearing bracket.
9. Check axial end play of shaft. End play should be adjusted to 0.003” to 0.005” by
adding or removing shims.
10. Install inboard bearing seal by pressing or lightly tapping into bearing bracket. Rotate
shaft to check for possible rubs on inpro seals and adjust if necessary.
12. Install inboard heat sink (829-2), outboard fan (817), and fan shroud (812) as
required.
The back pull-out unit can be reassembled by following the instructions given below and
referring to the appropriate pump sectional drawing Figure 4.
2. If removed during disassembly press throat bushing (847) into casing cover (802).
3. If removed during disassembly, press casing or cover wear rings (808-1,2) into casing
or cover and fusion weld in four locations, equispaced. Check rings for proper clearance
to the impeller rings.
4. Lubricate pump shaft and slide cartridge seal assembly and if part of your unit, the
inboard heat sink (829-2), onto shaft.
STDS-PWI-IOM-TM-0.0 22/26
8885. Monroe St. Houston, Texas USA 77061
Phone: (713) 956-2002
Fax(713) 956-2141
CAUTION: Care must be taken not to damage seal sleeve gasket or seal face.
6. Slide casing cover (802) over pump shaft (820) and seal. Insert and tighten cap
screws between bearing bracket (849) and cover.
7. Draw gland nuts up evenly until metal to metal contact is realized between gland and
cover.
8. Install impeller key (811-1) and impeller (805). The impeller is a very snug slip fit,
therefore the use of gentle force may be required to fully seat it to the shaft shoulder.
9. Tighten the seal drive collar set screws and packing sleeve follower if applicable.
Disengage seal locating devices and check rotation.
10. Check shaft and make sure it is free to rotate. Re-engage seal locating device prior
to storing or installing back pull-out unit in casing.
The pump back pull-out unit which includes the impeller (805), the casing cover (802), the
bearing bracket (849), and the shaft (820) is now ready for reassembly to the pump in the field.
Follow the instructions given below and refer to the pump sectional drawing Figure 4.
3. Slide Back Pull-Out unit into casing (800) and tighten casing stud nuts evenly.
CAUTION: Check shaft that it is free to rotate and does not bind.
STDS-PWI-IOM-TM-0.0 23/26
8885. Monroe St. Houston, Texas USA 77061
Phone: (713) 956-2002
Fax(713) 956-2141
STDS-PWI-IOM-TM-0.0 24/26
8885. Monroe St. Houston, Texas USA 77061
Phone: (713) 956-2002
Fax(713) 956-2141
SECTIONAL
PUMP CRITICAL
REFERENCE DESCRIPTION START-UP
RECONDITION SERVICE
NUMBER
4510 Casing Gasket 1 1 2
4200 Cartridge Seal 1 1 1
2200 Impeller - 1 1
2912 Impeller Locknut - 1 1
2110 Shaft - 1 1
2300.1 Impeller Wear
- 2 2
2300.2 Rings
1500.1
Casing Wear Rings - 2 2
1500.2
3000 Sleeve Bearing - 1 2
Table 4
STDS-PWI-IOM-TM-0.0 25/26
8885. Monroe St. Houston, Texas USA 77061
Phone: (713) 956-2002
Fax(713) 956-2141
Figure 3
STDS-PWI-IOM-TM-0.0 26/26