Introduction
The laser level sensor (AL40/AL100) is a non-contact, level measuring instrument designed for
granular solid materials and opaque liquids.
Operating Principal
The AL40/AL100 uses a high speed laser pulse to measure distance. The laser light is emitted
towards the surface and some of it reflects back to the instrument where it is detected by a
sensitive optical receiver. The time it takes for the light to travel to the surface and back to the
instrument is directly proportional to the distance between the instrument and the surface. Using
a time-of-flight calculation, and knowing the height of the vessel, the AL40/AL100 accurately
measures the distance to the target surface using the equation below:
Level = height - speed of light x time-of-flight
Range
The AL40/AL100 is able to measure the level in silos because of the inherent long distance
capability of laser technology. The laser has a natural advantage because it gets strong, clear
signals from most types of granular solid materials. As the laser beam doesn’t spread out and
lose strength as it travels, there is little signal loss with increased distance.
AL40 & AL100 Display & push button
Key Features
• Non Contact Level Measurement
• Fast setup, configuration & commissioning
• Accuracy 0.1% of full range
• Power supply 90-260Vac and 24Vdc
• 4-20mA output (Namur Compliant)
• IP65 rated enclosure
• 2 x Relay Outputs
• NC/COM/NO 5A at 250V AC
• Local Display : 4 Digit, 7-segment led display 0.10-40.00/100
• High Resolution ( Up to 0.01meters)
• Fast Refresh Rate ( 10 readings per second)
• Low Power Consumption ( 200mA)
• CE approval
• Optional Stainless Steel Enclosure
Page 3
Applications
Suitable for:
Petrochemicals
Chemical Processes
Food & Beverage Industries
Mining Industries
Metal Industries
Grain, Seed & Feed
Cement & Aggregates
Pulp & Paper
Positioning
Page 4
Quick Setup
Press the MENU key again and select yes to save your settings
and exit programming mode
n.sig on the display means No Signal. This may happen when the unit is face down on the
bench while configuring, for example.
At any time when in the programming mode, should you wish to EXIT, press the MENU button.
EHIT will appear for one second on the screen and then it will ask you YES or NO. Confirm
YES or NO using the UP or DOWN buttons, and then press ENTER
By selecting YES the instrument will first save all values changed and then exit the
programming mode back to the RUN mode. If NO is selected all changes will be disregarded
During the programming MODE the relays will not change state and mA output will be constant
on the last updated value.
When you press the MENU button on the display, it will display PASS on the screen for 1
second then it will show zero.
PASSWORD
On the display it will show PASS
If you want to enter and change the different parameters, select the UP button to change the
password to 40 and press ENTER.
FULL
On the display it will show Full then 40.00
The full distance is the full range to be measured. To change this distance use the UP and
DOWN button. When complete press the ENTER button. The units of the FULL distance is in
meters.
The maximum range that can be selected is 40.00m 100.0m (model dependent) and the
minimum is 0.1m. (DEFAULT = 40.00m or 100.0m)
Page 6
Menu Set-up
SPAN
On display it will show SPAN
The SPAN is the distance between the FULL distance and 100% (20mA). The SPAN is
changeable using the UP or DOWN buttons. It is used to calculate the 4-20mA range of the
instrument. When done press the ENTER button.
RELAY 1
On display it will show r1,on
This is to activate the Relay 1 output. The parameter is changeable from OFF to ON by pressing
the UP and DOWN buttons. After selection press the ENTER button. (DEFAULT = OFF)
When selected to the ON mode r1,ST will appear on the screen. (Switch Point).
This is the distance where the relay must be switched ON. The unit is in meters and to change
the value use the UP or DOWN button. After the correct value is selected press the ENTER
button.
The maximum range that can be selected is 40.00m or 100.0m (depending on model) and the
minimum is 0.01m. (DEFAULT = 10.00m)
This is the distance where the relay must be switched OFF. The unit is in meters and to change
the value use the UP or DOWN button. After the correct value is selected press the ENTER
button.
The maximum range that can be selected is 40.00m or 100.0m (depending on model) and the
minimum is 0.01m. (DEFAULT = 10.00m)
Page 7
RELAY 2
On display it will show r2,on
This is to activate the Relay 2 output. The parameter is changeable from OFF to ON by pressing
the UP and DOWN buttons. After selection press the ENTER button. (DEFAULT = OFF)
When selected to the ON mode r2,ST will appear on the screen. (Switch Point).
This is the distance where the relay must be switched ON. The unit is in meters and to change
the value use the UP or DOWN button. After the correct value is selected press the ENTER
button.
The maximum range that can be selected is 40.00m or 100.0m (depending on model) and the
minimum is 0.01m. (DEFAULT = 10.00m)
This is the distance where the relay must be switched OFF. The unit is in meters and to change
the value use the UP or DOWN button. After the correct value is selected press the ENTER
button.
The maximum range that can be selected is 40.00m or 100.0m (depending on model) and the
minimum is 0.01m. (DEFAULT = 10.00m)
mA
On display it will show AnA.o
This is the time delay setting for when the unit will go into FAIL SAFE mode. This is normally
when NO signal is received. Could be out of Range.
The units displays in seconds and is changeable using the UP or DOWN keys. When selected
press the ENTER button.
The maximum range that can be selected is 300sec and the minimum is 0sec. (DEFAULT =
10sec)
When this FAIL SAFE time has been met, the unit will output the selected Fail Safe as below.
FAIL SET
On display it will show F.SET
The FAIL SAFE setting is when the unit does not receive a signal and the fail time has counted
down. The unit will change the mA output to the selected Namur NE43 setting. The units are in
mA.
Page 8
This is selectable between 4 different settings. The settings are selectable using the UP or
DOWN buttons. When selected press the ENTER button.
PARAMETER 8 – RATE
On display it will show rATE
This is the update rate at which the unit will update the measured distance. It can be activated
using the UP or DOWN buttons. To activate it select YES and press the ENTER button.
It will then display SET on the screen. This is measured in meters and changeable using the UP
or DOWN buttons. When the desired update rate is selected press the ENTER button.
Example:
When the current distance is 1m and the next distance is measured at 2m, with the RATE
selected in off mode the distance will increase instantly. With the Rate selected in ON mode
and with a setting of 0.1m, the distance will update every second adding only 0.1m until it
reaches 2m.
The maximum range that can be selected is 1.00m and the minimum is 0.01m. (DEFAULT =
1.00m)
PARAMETER 9 – APPLICATION
On display it will show APP
The application can be selected for either LEVEL mode or DISTANCE mode.
LEVEL mode is used in level applications. To change the application mode select between LEV
and DIST using the UP or DOWN buttons. When the mode is selected press the ENTER
button.
Note: Distance is measured from the face of the Laser to the top of the product, Level is
measured from the bottom of the bin to the top the product measured.
Page 9
DUST
On display it will show duST
To enable or disable, use the UP or DOWN buttons. When selected press the ENTER button.
Selectable settings: ON or OFF. (DEFAULT = OFF)
Note: Enabling Dust mode will slow down your response time
FACTORY RESET
On display it will show FAC.r
To reset press the UP or DOWN button and select YES. Press the ENTER button.
Now to EXIT, press the MENU button. EHIT will appear for one second on the screen and then
it will ask you YES or NO. Confirm YES or NO using the UP or DOWN buttons, and then press
ENTER
By selecting YES the instrument will first save all values changed and then exit the
programming mode back to the RUN mode. If NO is selected all changes will be disregarded
Page 10
Performance DataSheet
Range 0.1-40m/100m (model dependent) distance and level format
Resolution 0.01m
Accuracy 0.1% of full range
Update rate 10 readings per second
DISPLAY
LOCAL LED 4 Digit 7-segment led display 0.10-40.00/100
SETUP
Local setup Manual setup using buttons and local display
TECHNICAL DATA
Power supply 90-260Vac and 24Vdc
Current consumption 200mA
Output 4-20mA NAMUR compliant
Relays 2 x relay NC/COM/NO 5A at 250vac
DIMENSIONS
Diameter 100mm (Base mounting 140mm)
Height 185mm
Flange optional 4” flange with dust tube
Material Powder coated aluminium (Stainless Steel Optional)
OPTICAL DATA
Optical aperture 55mm
Beam angle Less than 1 degree
Lens Impact resistant acrylic
Laser safety Class 1M. Do not view laser directly
ENVIRONMENT DATA
Operating temperature -20°C to +60°C
Pressure Atmospheric
Enclosure rating IP65
Page 11
Page 12
Laser Level Sensor Drawings
Terminal Connections
Face